SKC ULTRA 690-105-NB, ULTRA 690-106, ULTRA 690-101, ULTRA 690-103, ULTRA 690-104 Operating Instructions Manual

...
Page 1
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
ULTRA Passive Samplers with Charcoal Sorbent
Cat. No. 690-105
Clip
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Blank/
correction
compartment
Sliding cover in closed position
Sample sorbent
compartment
Sliding
cover
correction
compartment
Clip
Plugs
Blank/
sorbent
Back view
Applications
SKC patented* ULTRA Passive Samplers provide for low ppb to ppt detection of organic vapors in ambient and indoor air.
Sampling Rates: Go to the SKC Passive Sampling Guide at www.skcinc.com/catalog/passive-guide.php.
Analysis: Solvent extraction; gas chromatography
Shelf-life: Six months from date of manufacture. Store at 39.2 F (4 C).
See Analysis Instructions at www.skcinc.com/instructions/40127.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com
Page 2
Sampling
1
23a3b
4
5
1. Open pouch. Remove sampler. a. Do not discard pouch. b. Store pouch in a clean, organic solvent-free
environment.
2. Record the following: a. Sample ID on pouch and sampler ID label b. Sampling location on pouch c. Date sample taken on pouch
Do not use Sharpie® markers or equivalent to mark label or pouch as these can off-gas VOCs.
3. Clip sampler(s) in appropriate location. a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air to fl ow freely). Hang shelter securely at the appropriate height.
Ensure diffusion holes in sampler are not obstructed in any way.
4. Start sampling: Slide sampler cover to open
position (diff usion holes visible). Write sample start time on pouch label.
Diffusion holes (sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sampler
I.D. label
Clip
Sliding cover
5. Stop sampling: Slide sampler cover to closed
position (arrow and SKC visible). Write sample
Sample compartment
stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent to the laboratory.
Blank/
correction
compartment
Clip
Sliding cover
Page 3
7
8
6. Place sampler in its resealable pouch
6
immediately after sampling.
Ensure sample ID on sampler matches sample ID on pouch.
7. Seal pouch(es). Place pouch(es) in container with
cool pack.
8. Use expedited shipping to send sample(s),
pertinent information, and Chain of Custody form(s) to a laboratory for analysis.
&
Storage
Before Sampling: Up to 6 months at 39.2 F (4 C) in a clean, organic solvent-free
After Sampling: 21 days at 39.2 F (4 C) in a clean, organic solvent-free
environment
environment
Ordering
ULTRA Passive Samplers, pk/5, prefi lled with: Cat. No. Charcoal, 500 mg in each compartment 690-105
Limited shelf-life; see Storage above
Sampling Accessories Stand for Indoor Sampling 690-302 Shelter for Outdoor Sampling 690-303
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.
Page 4
Form 40126 Rev 1807