Applications
SKC patented* ULTRA Passive Samplers provide for low ppb to ppt detection of
organic vapors in ambient and indoor air.
Performance Profi le
Sampling Rates: Go to the SKC Passive Sampling Guide at
www.skcinc.com/catalog/passive-guide.php.
Analysis: Solvent extraction; gas chromatography
Shelf-life: Six months from date of manufacture. Store at ≤ 39.2 F (4 C).
See Analysis Instructions at www.skcinc.com/instructions/40127.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com

Sampling
1. Open pouch. Remove sampler.
a. Do not discard pouch.
b. Store pouch in a clean, organic solvent-free
environment.
2. Record the following:
a. Sample ID on pouch and sampler ID label
b. Sampling location on pouch
c. Date sample taken on pouch
Do not use Sharpie® markers or equivalent
to mark label or pouch as these can off-gas
VOCs.
3. Clip sampler(s) in appropriate location.
a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate
height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air
to fl ow freely). Hang shelter securely at the
appropriate height.
Ensure diffusion holes in sampler are not
obstructed in any way.
4. Start sampling: Slide sampler cover to open
position (diff usion holes visible). Write sample
start time on pouch label.
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sampler
I.D. label
Clip
Sliding cover
5. Stop sampling: Slide sampler cover to closed
position (arrow and SKC visible). Write sample
Sample compartment
stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent
to the laboratory.
Blank/
correction
compartment
Clip
Sliding cover

6. Place sampler in its resealable pouch
immediately after sampling.
Ensure sample ID on sampler matches sample
ID on pouch.
7. Seal pouch(es). Place pouch(es) in container with
cool pack.
8. Use expedited shipping to send sample(s),
pertinent information, and Chain of Custody
form(s) to a laboratory for analysis.
&
Storage
Before Sampling: Up to 6 months at ≤ 39.2 F (4 C) in a clean, organic solvent-free
After Sampling: 21 days at ≤ 39.2 F (4 C) in a clean, organic solvent-free
environment
environment
Ordering
ULTRA Passive Samplers, pk/5, prefi lled with: Cat. No.
Charcoal, 500 mg in each compartment 690-105
Limited shelf-life; see Storage above
Sampling Accessories
Stand for Indoor Sampling 690-302
Shelter for Outdoor Sampling 690-303