SKC ULTRA 690-105-NB, ULTRA 690-106, ULTRA 690-101, ULTRA 690-103, ULTRA 690-104 Operating Instructions Manual

...
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
ULTRA Passive Samplers with Charcoal Sorbent
Cat. No. 690-105
Clip
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Blank/
correction
compartment
Sliding cover in closed position
Sample sorbent
compartment
Sliding
cover
correction
compartment
Clip
Plugs
Blank/
sorbent
Back view
Applications
SKC patented* ULTRA Passive Samplers provide for low ppb to ppt detection of organic vapors in ambient and indoor air.
Sampling Rates: Go to the SKC Passive Sampling Guide at www.skcinc.com/catalog/passive-guide.php.
Analysis: Solvent extraction; gas chromatography
Shelf-life: Six months from date of manufacture. Store at 39.2 F (4 C).
See Analysis Instructions at www.skcinc.com/instructions/40127.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com
Sampling
1
23a3b
4
5
1. Open pouch. Remove sampler. a. Do not discard pouch. b. Store pouch in a clean, organic solvent-free
environment.
2. Record the following: a. Sample ID on pouch and sampler ID label b. Sampling location on pouch c. Date sample taken on pouch
Do not use Sharpie® markers or equivalent to mark label or pouch as these can off-gas VOCs.
3. Clip sampler(s) in appropriate location. a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air to fl ow freely). Hang shelter securely at the appropriate height.
Ensure diffusion holes in sampler are not obstructed in any way.
4. Start sampling: Slide sampler cover to open
position (diff usion holes visible). Write sample start time on pouch label.
Diffusion holes (sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sampler
I.D. label
Clip
Sliding cover
5. Stop sampling: Slide sampler cover to closed
position (arrow and SKC visible). Write sample
Sample compartment
stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent to the laboratory.
Blank/
correction
compartment
Clip
Sliding cover