Getting Started ..................................................................................................... 3
Charging the Battery Pack
Battery Charging Notes and Cautions
Turning Pump Power On/Off
Interpreting the Display
Determining Battery Charge Status
Using the Touch Screen
Navigating Menus and Screens
Determining Pump Status
Modifying Device Settings
(Clock Menu: Date/Time, Units Menu: Display Units, Screen Menu: Dim/Security)
Operation (Sample Menu: Set Flow, Calibrate, and Sample) ........................13
To set pump fl ow rate
To set up a single sorbent tube train and calibrate – Constant Flow Mode
To set up a multiple sorbent tube train and calibrate – Constant Flow Mode
To set up a multiple sorbent tube train and calibrate – Constant Pressure Mode
To calibrate pump fl ow rate prior to bag sampling – Constant Flow Mode
To sample using a single sorbent tube, multiple sorbent tubes, or sample bag
To set up and run a manual sample
To set up and run a timed sample
To run a sample preset
Flow Fault Mode and Display
Viewing History
Using Pocket Pump TOUCH with DataTrac Pro for Pocket Pump TOUCH Software
Accessories/Replacement Part Ordering ............................................... 25
SKC Limited Warranty and Return Policy .............................................. 25
Appendix: Performance Profi le ............................................................... 26
www.skcinc.com
Form 40017 Rev 1906
Page 1 of 26
Page 2
Introduction
Checking Pump/Kit Contents
Use the following table to verify that you received all items associated with the Cat. No. ordered. If you are missing items, contact SKC
at 800-752-8472 (U.S. only) or 724-941-9701.
If you ordered Cat. No.Your Package Should Contain
220-1000TCPocket Pump TOUCH only* with Li-Ion battery and screwdriver, requires charger; see kits or charger below
220-1000TC-CPocket Pump TOUCH with Charger* includes pump as described above and single charger (cable and wall cube with
220-1000TC-SPocket Pump TOUCH Starter Kit* includes pump and single charger as described above, 3 feet (0.9 meter) of Tygon
220-1000TC-KSingle Pump Kit*
220-1000TC-K55-pack Pump Kit* includes 5 pumps as described above, 5 single chargers (cables and wall cubes with U.S. plug), and
220-1000TC-K5D5-pack Deluxe Pump Kit* includes 5 pumps as described above, 5 Type A single tube holders, 5 single charging cables
Pocket Pump TOUCH requires ¼-inch ID tubing.
* Pocket Pump TOUCH pumps contain Li-Ion ba eries and are subject to special shipping regulations.
† Single kits are also available in hard-sided Pelican case. Contact SKC.
U.S. plug) 100-240 V
tubing, and collar clip with cable tie 100-240 V
nylon carry case 100-240 V
5 Type A single tube holders, in a Pelican case 100-240 V
(no wall cubes), one 5-port USB charging hub with power cable (U.S. plug), and 1 DataTrac Pro USB Bluetooth Adapter
(software available via free download), in a Pelican case 100-240 V
†
includes pump and single charger as described above, and Type A single tube holder, in a soft-sided
Quick Guide
Charging the battery: Insert USB connector on cable into USB port on supplied wall adapter. Plug into 100 to 240-V outlet. Hover
magnetic charging connector on cable closely over contact on back of pump until engaged and LEDs on contact and cable light blue.
Full charge is achieved in approximately 10 hours. See Charging the Battery Pack on page 3 for details.
Turning on: Press Power On/Off button on top of pump; the
screen will light and Flow screen will display.
Using the touch screen: Use a fi ngertip to gently touch
screen buttons in the indicated active zones.
Accessing Main Menu: Touch
.
Navigating from Main Menu
Desired OperationButton to TouchResult
Access device settings
History, Clock, Units, and Screen sub-menus display. Touch
buttons to view and choose setting options within each.
View information about the pump
Firmware version/date of pump manufacture, and pump
serial number display.
Access fl ow rate, calibration, and sampling
Flow, Presets, and Advanced (constant pressure) sub-menus
display. Touch buttons to view and choose mode or options
under each.
Basic Pump Operation
Run pump: From Main Menu: Touch Sample >Flow >Run
Active Touch Screen Areas
Note: Pressing the Power On/Off button on top of pump will dim and lock screen. To undim and unlock, press Power On/Off
button again.
Pause pump: Touch Pause
Stop pump: Touch Stop
. Touch Run to resume.
. Sample Summary will display. Touch to return to Main Menu.
Turning off: Press Power On/Off button on top of pump; the screen will go dark. The pump will power off automatically after
fi ve minutes of no activity.
www.skcinc.com
Form 40017 Rev 1906
Page 2 of 26
Page 3
Getting Started
Charging the Battery Pack
AC Charging Options Portable Charging Options
Magnetic charging contact
Magnetic
Standard
100-240-V
wall outlet
Charging
cable/charger
Power cable
AC adapter
(wall cube)
with U.S. plug
Standard
100-240-V
wall outlet
Use up to 5 charging
cables/chargers
5-port USB hub
Figure 2. Charging Trains
1. Insert USB connector on charging cable into a USB port on a(n):
a. Adapter (wall cube, supplied with SKC Single USB Charger, Cat. No. 220-300). Plug adapter (wall cube) into a standard 100 to
240-volt wall outlet.
b. PC/laptop (not supplied by SKC) with a full charge or plugged into a wall outlet. Note: USB ports and PC/laptop se ings vary.
If using this charging option, verify that your pump is charging by checking that more bars have appeared on the pump ba ery
status icon midway through charging.
c. Portable power pack (not supplied by SKC) with a full charge and that is compatible with Pocket Pump TOUCH
charging contact
Charging
cable/charger
USB
connector
Portable power
pack or PC/Laptop
The blue LED on the charger will light.
2. Hover the magnetic charging connector on the charging cable closely over the charging contact on the back of the pump (Figure 1)
until the LED behind the magnetic contact on the pump lights blue and pump status LEDs blink red in a repeating pa ern of right-
inlet-left.
Troubleshooting Charging
If the magnetic charging connector is seated on the pump charging contact, but the charger and charging contact LEDs
are not lit blue: Check that the charging cable is connected to an AC adapter and active outlet, PC/laptop connected to an active
power source, or a portable power pack with a full charge that is compatible with charging Pocket Pump TOUCH.
If only the charger LED lights blue: It means that there is no connection with the pump charging contact. Check and reconnect
the charging connector on the cable to the pump’s charging contact. Both the charger and charging contact LEDs should light blue
when connected properly.
3. Charge ba ery completely (approximately 10 hours to 100% or 4.5 hours to 80%). See ba ery charge status indicator descriptions below.
LED Activity DescriptionCharging Status
Charging contact LED blue/inlet and corner LEDs red, fl ashing in repeated pattern
of right-inlet-left
Charging contact LED blue/inlet LED green, solid/corner LEDs green, blinking,
battery status icon outline fl ashing
Battery charging
Battery charging complete
www.skcinc.com
Form 40017 Rev 1906
Page 3 of 26
Page 4
Battery and Charging Notes and Cautions
• Do not operate pump from or charge pump with charger in hazardous atmospheres.
• Use only the SKC-approved battery pack (Cat. No. P76303) and charging cable supplied in Cat. No. 220-300. Use of an
unapproved battery and/or charging cable could damage the pump and will void any warranty.
• Turn off pump before removing the battery pack.
• Tampering with the battery pack (opening, disassembling, short circuiting, crushing, or exposing the battery pack to fi re or
temperatures in excess of 212 F [100 C]) voids any warranty.
• Protect sample pump from rain and extended direct exposure to sun when in use outdoors
• Allow pump to equilibrate when moving it from one temperature extreme to another.
• Failure to follow warnings and cautions voids any warranty.
• Substitution of components may impair intrinsic safety. La substitution de composants peut compromettre la Sécurité Intrinsèque.
• To prevent ignition of hazardous atmosphere, batteries must only be charged (removed and replaced) in an area to be known
nonhazardous. Afi n de prévenir l’infl ammation d’atmosphères dangereuses, ne changer les batteries que dans des emplacements
désignés non dangereux.
• Maximum charge input voltage U
= 5.5 VDC.
m
• The battery used in this device may present a risk of fi re or explosion when heated above 212 F (100 C) or incinerated. Replace
battery with SKC Battery Pack Cat. No. P76303 only. Use of another battery may present a risk of fi re or explosion.
Turning Pump Power On/Off
Turn on: Press the Power On/Off bu on on the top of the pump (Figure 1); the screen will light.
Turn off : Press the Power On/Off bu on on the top of a non-running pump (Figure 1); the screen will go dark. The pump will automatically
power off after fi ve minutes of inactivity.
Interpreting the Display
Time
(12 or 24-hr display)
Constant display at the top of every screenDisplays while pump is running.
Date
(3 format options)
Battery status icon
(charge remaining)
Scrolling display of
accumulated sample
volume (L), ambient
temperature (F or C),
atmospheric pressure
(mbar, inHg, mmHg),
and back pressure
(inH2O)
Flow rate (ml/min
constant fl ow or inH2O
constant pressure)
Accumulated run time
FPO
Calibration correction
applied
Screen
locked to
prevent
accidental
tap
errors or
tampering
www.skcinc.com
Form 40017 Rev 1906
Page 4 of 26
Page 5
Determining Battery Charge Status
The ba ery status icon appears at the top of the screen at all times and contains bars that reduce in number as ba ery charge is used.
Following is information on interpreting the ba ery status icon.
Battery Status Icon DescriptionBattery Charge Remaining
Four bars
Full battery charge, approximately 75 to 100%
Three bars
Two bars
One bar
One bar (fl ashing)
Approximately 50 to 75%
Approximately 25 to 50%
Approximately 3 to 25%
Less than 3%; fl ow fault is imminent. Pump should be charged.
Using the Touch Screen
Use a fi ngertip to gently touch screen bu ons in the active zones indicated below.
Navigating Menus and Screens
Pocket Pump TOUCH operates through a series of menus and screens. When the pump is powered
on, the Flow screen displays, which allows you fast access to calibration and sampling. For more
details, start at To set pump fl ow rate.
Immediately below the display are four virtual navigation bu ons that access previous screen,
Main Menu, and increase/decrease values.
Back buttonUp Arrow buttonDown Arrow buttonMain Menu button
Returns to previous screen
Touching
displays the Main Menu from which you can access all options. Following is a Menu/Screen Overview.
Increases selected value or
moves up a list/range/display
Touch and hold to speed
increment of fl ow or pressure
settings.
Decreases selected value or
moves down a list/range/display
Touch and hold to speed
decrement of fl ow or pressure
settings.
www.skcinc.com
Returns to Main Menu
Form 40017 Rev 1906
Page 5 of 26
Page 6
Menu/Screen Overview
Main Menu
History Menu
• List of sample runs/summaries
Clock Menu
• Set Time
• Set Date
• Select Clock display
• Select Date display
Units Menu
• Select Temp display
• Select ATM display
Screen Menu
• Select Dim
• Select Secure Lock
• Select Auto Lock
• Firmware version number
• Lifetime run time and volume
• Pump serial number
• Pump manufacture date
Menus and screens can contain the following navigational touch bu ons.
ButtonGeneral Function
Check mark
Left and right movement
Saves a selected item
Allows horizontal movement on a scale (see right) or moves left or
right through fi elds, activating each for entry of a value.
Flow Menu
• Set Flow
• Calibrate (constant fl ow mode)
• Set Duration (timer)
• Run button
Presets Menu
• Select presets P1-P4 (created in DataTrac Pro,
uploaded to pump)
Advanced Menu
• Set Pressure
• Set Duration (timer)
• Run button
Calibration/Flow Adjustment
Run (start)
Pause
Stop
Allows selection of an adjustment to fl ow during calibration
Runs the pump for sampling
Pauses a running pump. Elapsed time and volume accumulation pause. When Run is touched,
time and volume will continue to accumulate.
Stops a running pump and resets elapsed time and volume to zero. Run time information will
be available in Sample Summary and History.
www.skcinc.com
Form 40017 Rev 1906
Page 6 of 26
Page 7
Determining Pump Status
Pocket Pump TOUCH features status LEDs on the top corners of the pump, beneath the pump inlet, and behind the magnetic charging contact
on the back of the pump. These indicate pump status at a quick glance.
LED Activity DescriptionPump Status
Corner and inlet LEDs green, fl ashingPump running
Corner and inlet LEDs red, fl ashingPump is in fl ow fault
Inlet LED yellow-orange, fl ashingcorner LEDs green, blinkingPump in constant pressure mode,
working to achieve pressure setting. Inlet
LED will change to solid green when set
pressure is achieved.
www.skcinc.com
Form 40017 Rev 1906
Page 7 of 26
Page 8
Modifying Device Settings
Note about default se ings: Pocket Pump TOUCH is shipped with the following default se ings that may be changed by the user
from the Device sub-menus.
Dim: On, Auto Lock: Off , Secure Lock: Off , Temperature Units: F, Atmospheric Pressure Units: inHg, Time Format: 12 Hour, and Date
Format: mm/dd/yyyy
Clock Menu
To change time on pump
From Main Menu:
Touch Device > Touch Clock > Touch time displayed
Hour digits will fl ash.
Touch up/down arrow bu ons to increment/decrement hours.
Touch right arrow to advance to minutes (will fl ash) and up/down arrow bu ons to adjust minutes.
Touch right arrow to advance to AM/PM (will fl ash) and up/down arrow bu ons to toggle AM/PM.
Touch check mark to accept new time and return to Clock Menu. New time se ing will display.
To change clock display format
From Main Menu:
Touch Device > Touch Clock > Touch 12 or 24 Hour
12 Hour and 24 Hour bu ons display. Touch desired se ing bu on to select and return to Clock
Menu. New clock display format will display.
www.skcinc.com
Form 40017 Rev 1906
Page 8 of 26
Page 9
To change date and date display
From Main Menu:
Touch Device > Touch Clock > Touch Date
Month digits will fl ash.
Touch up/down arrow bu ons to increment/decrement month.
Touch right arrow to advance to day (will fl ash) and up/down arrow bu ons to increment/decrement
day.
Touch right arrow to advance to year (will flash) and up/down arrow buttons to increment/
decrement year.
Touch check mark to accept new date and return to Clock Menu. New date will display.
To change date display format
From Main Menu:
Touch Device > Touch Clock > Touch Date Display Format
Touch the desired date display format to select and return to Clock Menu. New date display format
will display.
www.skcinc.com
Form 40017 Rev 1906
Page 9 of 26
Page 10
Units Menu
To change temperature units
From Main Menu:
Touch Device > Touch Units > Touch Temp
°F and °C bu ons display. Touch desired bu on to select and return to Units Menu.
To change atmospheric pressure display units
From Main Menu:
Touch Device > Touch Units > Touch ATM
mbar, inHg, and mmHg bu ons display. Touch desired bu on to select and return to Units Menu.
Note: Changing the display units aff ects only the display of atmospheric pressure on the pump
screen and in Sample Summary and History. Back pressure (inlet pressure) will always display in
“inH2O” on the pump screen and in Sample Summary and History.
www.skcinc.com
Form 40017 Rev 1906
Page 10 of 26
Page 11
Screen Menu
To set Dim to on or off
From Main Menu:
Touch Device > Touch Screen > Touch Dim
Touch desired bu on to select and return to Screen Menu.
If Dim is set to On, the user can dim and lock the screen of a running pump by pressing the Power
On/Off bu on. If Auto Lock is set to On in combination with Dim, the screen will dim and lock
automatically when the pump is run. The screen can be undimmed and unlocked by pressing the
Power On/Off bu on. For more information on the Auto Lock and Secure Lock features, see below. This
se ing helps conserve ba ery usage.
If Dim is set to Off , the screen backlight will stay on during the entire sample run.
To set Auto Lock to on or off
From Main Menu:
Touch Device > Touch Screen > Touch Auto Lock
Touch desired bu on to select and return to Screen Menu.
If Auto Lock is set to On, the screen will lock (become inactive) as soon as the pump starts running
a sample. A lock icon and “Press top bu on to unlock” message will appear on the screen. If Dim
is set to On, the screen will both lock and dim when the pump starts running a sample.
If Auto Lock is set to Off , the screen will remain active. The screen may be locked (made inactive)
manually at any time during a sample run by pressing the Power On/Off bu on on top of the pump.
To unlock and reactivate the screen, press the Power On/Off bu on on top of the running pump.
This se ing helps to prevent accidental tap errors during sample runs.
www.skcinc.com
Form 40017 Rev 1906
Page 11 of 26
Page 12
To set Secure Lock to on or off and set passcode
From Main Menu:
Touch Device > Touch Screen > Touch Secure Lock
Secure Lock helps to prevent tampering during a sample run. Touch desired bu on. If Secure Lock was set previously to On and you select
Off , you will be prompted to enter the previously set passcode. Once the passcode is entered, you will be returned to the Screen Menu.
If you select On, proceed as follows.
You will be prompted to enter a four-digit passcode of your choosing. Touch the screen keypad to
enter the desired four-digit number combination. Note: Entered digits will display as ****.
You will be prompted to confi rm the passcode. Touch the screen keypad to enter the same four-
digit number combination. Upon entering the last digit, the passcode will be saved and you will
be returned to the Screen Menu.
If Secure Lock is set to On, the user can lock the screen of a running pump by pressing the Power
On/Off bu on. If Auto Lock is set to On concurrently, the screen locks when the pump is run.
To unlock and reactivate the screen, press the Power On/Off bu on on top of the running pump
and touch the screen keypad to enter the previously set four-digit passcode.
Master Unlock Feature: If you cannot remember the Secure Lock passcode, touch 1 2 3
This will override the Secure Lock, but will not disable it.
when prompted for the passcode.
www.skcinc.com
Form 40017 Rev 1906
Page 12 of 26
Page 13
Operation - Set Flow, Calibrate, and Sample
tube
Tubi
ng
Calibratoro
u
Sample Menu
To set pump fl ow rate
From Main Menu:
Touch Sample > Touch Flow > Touch fl ow display
Touch left/right arrow bu ons to set gross fl ow se ing. Touch up/down arrow bu ons to fi ne tune
se ing.
Note: A sustained touch on the up/down arrow bu ons will speed increment/
decrement fl ow se ing.
Touch check mark to accept selection and return to Flow Menu with new fl ow se ing displayed.
To set up a single sorbent tube train and calibrate – Constant Flow Mode
* Charge pump battery completely before calibration and sampling.
* Requires 222 Series Low Flow Holder accessory.
* Calibrate/verify pump fl ow rate before and after each sampling operation
using the tube holder and pump to be used for sampling.
1. Break tips off a representative sorbent tube and insert (arrow printed on
tube pointing toward the pump) into the rubber sleeve of a 222 Series
(non-adjustable) Low Flow Tube Holder (Figure 3).
2. Connect tubing from the 222 Series (non-adjustable) low fl ow tube holder
to the pump inlet (Figure 3).
3. Use a length of fl exible tubing to connect the exposed open end of the
representative sorbent tube in the holder to a calibrator outlet (Figure 3).
4. Set fl ow rate on pump (see To set pump fl ow rate above).
Pump will run as a result of touching the calibration icon. It is good practice to allow the pump to
run for 5 minutes before calibrating the fl ow rate.
Build a calibration train with sample media in line (Figure 3).
Touch up/down arrow bu ons to increment/decrement calibration adjustment.
www.skcinc.com
Form 40017 Rev 1906
Page 13 of 26
Page 14
The calibration adjustment value will display beside the calibration icon.
Sorbenttubes
Tubi
ng
Calibratoroutle
t
Note: The fl ow rate displayed on the calibrator will change as a result of this adjustment.
Touch check mark to accept the calibration adjustment value and return to the Flow Menu.
The fl ow rate displayed on the pump will remain unchanged.
6. Disconnect representative sorbent tube from the calibrator. Replace representative sorbent tube with a newly opened unexposed
method-specifi ed sorbent tube to complete the sampling train. See To sample on page 18.
To set up a multiple sorbent tube train and calibrate – Constant Flow Mode
* Charge pump battery completely before calibration and sampling.
* Use of an unapproved battery and/or charging cable could damage the pump and will void any warranty.
* Requires Constant Pressure Controller (CPC) and Multiple-tube Low Flow Holder accessories.
* See Adjustable Low Flow Holder Operating Instructions for details on holder operation.
* Calibrate/verify pump fl ow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
1.
Set fl ow rate on pump (see To set pump fl ow rate on page 13). Important! Set
pump fl ow rate to ≥ 15% higher than the sum of the fl ow rates through all tubes
to be used during the multiple-tube sample run to ensure adequate airfl ow. Note: The combined fl ow through all tubes cannot exceed 500 ml/min.
2. Press tubing from the outlet (unlabeled side) of a Constant Pressure
Controller (CPC) onto the pump inlet (Figure 4).
3. Press tubing from the 224 Series Multiple-tube Low Flow Holder onto the
CPC inlet (labeled “To Sample” - Figure 4).
4. Break tips off a representative sorbent tube and insert (arrow printed on
tube pointing toward the pump) into the rubber sleeve of a 224 Series
Multiple-tube Low Flow Holder (Figure 4). Repeat for additional tubes
up to the maximum number of holder ports or desired number of tube
samples. Important! Fill unused tube holder ports with unopened
sorbent tubes to keep ports clean during sampling.
5. Label all sorbent tubes and ports.
6. On the Multiple-tube Low Flow Holder, use a small fl at-head screwdriver
to turn counterclockwise the brass fl ow adjust screw directly beneath the fi rst port holding an open sorbent tube. This opens the port to airfl ow.
7. Use a length of fl exible tubing to connect the exposed open end of the fi rst
port sorbent tube to a calibrator outlet (Figure 4).
8. Calibrate pump fl ow rate through the tubes as follows:
a. Touch Run bu on. It is good practice to run the pump for 5 minutes before calibrating the fl ow
rate.
b. Use a small fl at-head screwdriver to turn the brass fl ow adjust screw beneath the port of the
tube to be calibrated counterclockwise to increase fl ow or clockwise to decrease fl ow until
the calibrator consistently displays the required fl ow rate for the fi rst port sorbent tube.
Note: The fl ow rate displayed on the calibrator will change as a result of adjusting the brass fl ow adjust screw on the holder; this is the fl ow
rate through the sorbent tube. The fl ow rate displayed on the pump will not change.
c. Touch the Stop bu on
to stop the pump. Touch the Back bu on
to return
to the Flow Menu.
9. Repeat Steps 6 through 8 for each active sorbent tube in the holder.
10. Disconnect representative sorbent tube from the calibrator. Replace representative sorbent tubes with newly opened unexposed methodspecifi ed sorbent tubes to complete the sampling train. Important! Ensure all unused ports remain plugged with unopened sorbent tubes
to keep ports clean during sampling. See To sample on page 18.
To set up a multiple sorbent tube train and calibrate – Constant Pressure Mode
(For Advanced Users Only)
* Charge pump battery completely before calibration and sampling.
* Requires Multiple-tube Low Flow Holder accessory
* See Adjustable Low Flow Holder Operating Instructions for details on holder operation.
* It is good practice to run the pump for 5 minutes before calibrating the fl ow rate.
* Calibrate/verify pump fl ow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
1. Determine back pressure to be set in Constant Pressure Mode:
a. Connect tubing from holder outlet to pump inlet. (Figure 5).
b. Break tips off a representative sorbent tube and insert (arrow printed on
tube pointing toward the pump) into the rubber sleeve of a 224 Series
Multiple-tube Low Flow Holder.
c. Use a small fl at-head screwdriver to turn counterclockwise (open) the
brass fl ow adjust screw on the Multiple-tube holder port containing
the tube.
d. Label sorbent tube and port. Plug all unused ports with unopened
tubes to keep ports clean during sampling.
e. Set method-specifi ed fl ow rate for the tube on the pump (see To set
pump fl ow rate on page 13). Important! Ensure all unused ports remain
plugged with unopened sorbent tubes.
i. Touch the Run bu on
to start the pump.
ii. Read and record back pressure shown on pump display (inH2O part
of scrolling display when pump is running – use up/down arrow
bu ons to speed scrolling display).
f. Repeat Steps 1.b through 1.e for remaining representative sorbent tubes.
Note: The combined fl ow rate through all tubes cannot exceed 500 ml/
min. Touch the Stop bu on
bu on
to return to the Flow Menu.
to stop the pump. Touch the Back
g. Review the recorded back pressures and determine the highest back
pressure. Add 2 or 3 inches to this value, round up to the integral
(whole) number, and add 1. This will be the back pressure se ing for Constant Pressure Mode in Step 2.
www.skcinc.com
Form 40017 Rev 1906
Page 15 of 26
Page 16
2. Set pressure to value obtained in Step 1.g as follows:
a. Touch left/right arrow bu ons to set gross pressure se ing. Touch up/down arrow bu ons
to fi ne-tune se ing.
Note: A sustained touch on the up/down arrow bu ons will speed increment/ decrement
pressure se ing.
Note: Set pressure value will blink and inlet LED will blink yellow-orange. When set
pressure is achieved, the pressure value will stop blinking and the inlet LED will turn
green. This may take up to 90 seconds.
b. Touch check mark to accept selection and return to Advanced Menu with new pressure
se ing displayed.
3. Use a length of fl exible tubing to connect the exposed open end of the fi rst port sorbent tube
to a calibrator outlet (Figure 5).
4. Touch Run bu on
on the pump touch screen.
5. Use a small fl at-head screwdriver to turn fl ow adjust screw on port (clockwise to decrease,
counterclockwise to increase) until calibrator displays the method-specifi ed fl ow rate.
Note: The fl ow rate displayed on the calibrator is the fl ow rate through the sorbent tube.
6. Repeat Steps 3 through 5 for all remaining open sorbent tubes.
7. Disconnect representative sorbent tube from the calibrator. Replace representative sorbent
tubes with newly opened unexposed method-specified sorbent tubes to complete the
sampling train.
Important! Ensure all unused ports remain plugged with unopened sorbent tubes to keep ports clean during sampling. See To sample on page 18.
Note: If the pump is run in constant pressure mode (Sample > Advanced > Pressure se ing) without media connected or with media connected
that has a pressure drop below the set pressure value, the pump will “rev “ in an a empt to achieve the desired pressure se ing. If it cannot,
it will fault.
www.skcinc.com
Form 40017 Rev 1906
Page 16 of 26
Page 17
To calibrate pump fl ow rate prior to bag sampling – Constant Flow Mode
Pump inlet Tubing Calibrator outlet
* Charge pump battery completely before calibration and sampling.
* Use of an unapproved battery and/or charging cable could damage the
pump and will void any warranty.
1. Set fl ow rate on pump (see To set pump fl ow rate on page 13).
2. Use fl exible tubing to connect pump inlet to the outlet of a calibrator
Pump will run. Allow pump to run for 5 minutes before performing calibration.
Touch up/down arrow bu ons to increment/decrement calibration adjustment.
adjustment value will display,
as shown below left.
A calibration
Note: The fl ow rate displayed on the calibrator will change as a result of this adjustment. The
fl ow rate displayed on the pump will not change.
Touch check mark to accept the
calibration adjustment value and
return to the Flow Menu. The calibration adjustment value will
display beside the calibration icon.
4. Disconnect calibrator and tubing from pump inlet. See To sample on page 18.
www.skcinc.com
Form 40017 Rev 1906
Page 17 of 26
Page 18
To sample using a single sorbent tube, multiple sorbent tubes, or sample bag
* Charge pump battery completely before sampling.
* Use of an unapproved battery and/or charging cable could damage the pump and will void any warranty.
* If using sample tubes as media, calibrate/verify pump fl ow rate before and after each sampling operation using the tube holder and
pump used for sampling.
1. Ensure pump is calibrated to the method-specifi ed fl ow rate. See To calibrate on page 13.
From Main Menu:
> Touch Run button in Flow Menu to
run pump in constant fl ow.
Touch Sample > Touch Flow or Advanced
2. Ensure calibrator and tubing have been removed and representative
method-specifi ed sampling medium used for calibration has been replaced
with newly opened unexposed method-specifi ed sampling medium to
complete the sampling train.
> Touch Run button in Advanced Menu
to run pump in constant pressure.
For bag sampling
a. Use PTFE tubing to connect pump exhaust (Figure 1) to the bag fi tting. See bag operating instructions.
b. Unscrew cap on pump exhaust to remove it and screw in supplied exhaust fi tting or optional Quick-connect
Bag Sampling Adapter accessory.
c. Attach tubing from the bag to the fi tting or adapter on the pump exhaust.
3. On the pump, choose from a manual sample, timed sample, or sample preset (presets are uploaded to the pump from DataTrac Pro
Software). See To set up and run a manual sample below, To set up and run a timed sample on page 20, or To run a sample preset on page 21.
To set up and run a manual sample
1. Connect pump to sampling train. See To sample, Step 2, above.
2. Run sample as follows:
From Main Menu:
> Touch Run button in Flow Menu to
run pump in constant fl ow.
Touch Sample > Touch Flow or Advanced
www.skcinc.com
> Touch Run button in Advanced Menu
to run pump in constant pressure.
Form 40017 Rev 1906
Page 18 of 26
Page 19
Options on pump screen during a sample run:
Touch Stop to stop sampling, reset accumulated data
display, and view Sample Summary.
Touch Pause
to pause sampling and retain accumulated
data display. When touched, pause changes to Run button
.
Touch Run button to resume sampling and data accumulation.
Dim (set to On) dims a screen that has been locked by pressing
the Power On/Off button on a running pump or through Auto Lock
as soon as the pump starts running.
To resume normal backlighting, press the Power On/Off button
on top of the pump (Figure 1). See Modify Device Settings, Screen Menu on page 11.
This feature may be used concurrently with Auto Lock and Secure
Lock.
Auto Lock (set to On) locks (inactivates) the pump touch screen
when the pump starts running. A lock icon and “Press top button
to unlock” message appear on the screen.
If Auto Lock is set to Off, the screen remains active. The screen
may be locked (made inactive) at any time during sampling by
pressing the Power On/Off button on top of the pump.
To unlock and reactivate the screen, press the Power On/Off
button on top of the pump. This feature helps to reduce tap errors
during sample runs.
Secure Lock (set to On) locks (inactivates) the pump touch
screen when the Power On/Off button is pressed. A lock icon
and “Press top button to unlock” message appear on the screen.
To unlock Secure Lock and reactivate the screen, press the
Power On/Off button on top of the pump, and touch the screen
keypad to enter the previously set four-digit passcode.
If Auto Lock and Secure Lock are set to On concurrently, the
pump touch screen will lock when the pump is run, but will require
that the user enter a passcode to unlock the screen.
Flow fault may occur when there is a restriction in airfl ow (e.g.,
kinked tubing) that remains uncorrected and the pump can no
longer compensate fl ow. See Flow Fault Mode and Display on
page 22 for details on pump operation during fl ow fault.
www.skcinc.com
Form 40017 Rev 1906
Page 19 of 26
Page 20
3. When the required sampling period is completed (or the bag is properly infl ated - see bag operating instructions), touch Stop bu on
to stop sampling, which displays a Sample Summary and automatically resets the accumulated data.
Note: If pump is shut off or goes to sleep after a sample is completed and is powered on again, the initial display will be the Sample
Summary of the previous sample run.
For bag sampling: If the Quick-connect Bag Sampling Adapter accessory was used, remove tubing after
sampling by using fi ngertips to press center ring on adapter while pulling tubing away from the adapter.
To set up and run a timed sample
1. Set sample duration as follows:
From Main Menu:
> Touch Time button in Flow Menu to
set sample duration.
Touch Sample > Touch Flow or Advanced
Hour digit 1 will fl ash.
Touch up/down arrow bu ons to increment/decrement hour.
Touch right arrow to advance to hour digit 2 (will fl ash) and up/down arrow bu ons to adjust
hour digit 2. Repeat through minutes.
Touch check mark to accept new time and return to Clock Menu. New time se ing will display.
2. Connect pump to sample train. See To sample, Step 2, on page 18.
3. Run sample as follows:
> Touch Time button in Advanced
Menu to set sample duration.
www.skcinc.com
Form 40017 Rev 1906
Page 20 of 26
Page 21
From Main Menu:
> Touch Run button in Flow Menu to
run pump in constant fl ow.
Touch Sample > Touch Flow or Advanced
> Touch Run button in Advanced Menu
to run pump in constant pressure.
See Options on pump screen during a sample run on page 19.
4. When the Timed sampling period is complete, the pump will automatically stop sampling, which displays a Sample Summary and
automatically resets the accumulated data.
Note: If pump is shut off or goes to sleep after a sample is completed and is powered on again, the initial display will be the Sample Summary
of the previous sample run.
To run a sample preset (uploaded to pump from DataTrac Pro Software)
1. Connect pump to sampling train. See To sample, Step 2, on page 18.
2. Select sample preset as follows:
From Main Menu:
Touch Sample > Touch Presets> Touch desired Preset(programmed in DataTrac Pro Software and uploaded to pump)
Presets are created by the user in DataTrac Pro for Pocket Pump TOUCH Software and uploaded to the pump (see DataTrac Pro for
Pocket Pump TOUCH User Manual for details on creating presets). The start date for a preset with delayed start can be changed on the pump
by touching the date on the Preset screen. The pump can also be calibrated from this screen.
To run a preset sample with a delayed start, touch check mark to activate the preset.
www.skcinc.com
Note: The day of a preset with delayed
start can be changed from the pump
screen. Touch the delayed start date,
touch the up and down arrow buttons
to change to desired start day, and touch
check mark. This can be changed to a
future day only.
Form 40017 Rev 1906
Page 21 of 26
Page 22
The pump will display a screen showing a summary of the preset and a Cancel (X) bu on. Touch
the X if you wish to cancel the activated preset.
To run a preset sample without a set delayed start, touch the Run bu on
See Options on pump screen during a sample run on page 19.
3. When the Preset sampling period is complete, the pump will automatically stop sampling,
which displays a Sample Summary and automatically resets the accumulated data.
Note: If pump is shut off or goes to sleep after a sample is completed and is powered on again,
the initial display will be the Sample Summary of the previous sample run.
Flow Fault Mode and Display
During a sample run, overloaded sample media or kinked tubing can restrict airfl ow and cause
back pressure to build to a point at which the pump can no longer compensate flow within
± 5%. If this condition is sustained for 3 to 10 seconds, the pump will go into fl ow fault mode
as follows:
1. Pump stops running and status LEDs on pump fl ash red. Elapsed time stops.
2. An exclamation point icon appears on the display next to the fl ow rate.
3. After 20 seconds in fault, the pump will a empt to restart up to 5 times.
a. If full airfl ow is restored during the restart a empts, the pump will continue the sample
run.
b. If full airfl ow is not restored during 5 restart a empts within 5 minutes, the pump will
end the sample run and display the Sample Summary. The Sample Summary will indicate
the number of faults. The LEDs will fl ash red with decreasing frequency.
4. To clear a fl ow fault, touch any bu on on the display. Note: A fl ow fault will also be cleared when the ba ery charge is depleted.
.
www.skcinc.com
Form 40017 Rev 1906
Page 22 of 26
Page 23
Viewing History
History Menu
To review specifi c sample run summaries directly on the pump screen
From Main Menu:
Touch Device > Touch History > Touch arrows to scroll
> Touch checkmark to select > Touch arrows to scroll
selected summary info
Note: The last 16 sample runs can be viewed as Sample Summaries on the pump. A maximum of 4,416 data
entries can be held in pump memory. This is equal to approximately 70 hours of one-minute averages or 360
hours of 5-minute averages. To access this data, upload it to DataTrac Pro for Pocket Pump TOUCH Software
on your PC.
Using Pocket Pump TOUCH with DataTrac Pro for Pocket Pump TOUCH Software
Pocket Pump TOUCH communicates with a laptop/PC via a USB Bluetooth Adapter (Cat. No. 877-94) and DataTrac® Pro for Pocket Pump
TOUCH Software, which is available as a download (requires Internet connection). Access DataTrac Pro for Pocket Pump TOUCH Software
using the following procedure:
1. Check that PC meets DataTrac Pro system requirements.
Operating systemMicrosoft® Windows® 7, 8, 8.1, or 10
Required softwareDataTrac Installer (included with installation)
Minimum display resolution1024 x 768
Available portUSB 2.0
2. Install the USB Bluetooth Adapter or “dongle” (Cat. No. 877-94) on laptop/PC according to instructions included with adapter.
3. Browse URL provided in adapter instructions, and download and install the DataTrac Pro Installer on laptop/PC. (A fl ash drive option
is available for users without an Internet connection.)
4. Upon successful installation, DataTrac Pro will launch automatically and a empt to fi nd any active Pocket Pump TOUCH pumps
within the area.
For further instructions, see the DataTrac Pro for Pocket Pump TOUCH User Manual.
www.skcinc.com
Form 40017 Rev 1906
Page 23 of 26
Page 24
Maintenance
Change the Battery Pack Review Notes and Cautions on page 4 before proceeding.
1. Turn off the pump by pressing the Power On/Off bu on located on the top of the non-running pump (Figure 1); the screen will go
dark.
2. Use a 1.5-mm hex key (provided with pump) to loosen two safety screws on the back of the ba ery pack housing (Figure 1).
3. Orient the pump so that the belt clip faces upward and the pump inlet points left. Using your fi ngers, grip the ba ery pack housing
just above its top screw and just below its bo om screw and lift up against the belt clip. Once the housing clears the pump case, pull
it to the right, away from the pump case.
4. If replacing the ba ery pack with a new ba ery pack (Cat. No. P76303), dispose of the used ba ery promptly and according to all
state and local regulations.
Do not disassemble the battery pack. Do not dispose of in fi re. Dispose of used batteries promptly according to all state and local recycling
or waste regulations.
5. Install a new ba ery pack or re-install an existing ba ery pack.
a. Align ba ery pack housing with pump case.
b. Use the back of the ba ery pack housing to push up the belt clip until the ba ery pack is aligned with the pump case and presses
easily into place. Ensure the ba ery pack housing fi ts fl ush with the pump case.
c. Use a 1.5-mm hex key to tighten two screws on back of the ba ery pack housing. Tighten screws in an alternating fashion.
d. Charge the ba ery pack completely before use. See Charging the Ba ery Pack on page 3.
Note: When the ba ery is rea ached to the pump, the screen will light and then go black as the pump performs a fi rmware check. This
lasts approximately 20 seconds and the Main Menu will display as soon as the check is completed.
Clean the Touch Screen
Use a soft, clean, and dust-free cloth to remove fi ngerprints from the touch screen.
Do not use alcohol to clean the screen.
Troubleshooting
Troubleshooting Guide
IssuePossible Solutions
The display on the pump is frozen.1. Press the Power On/Off button to shut off the pump.
2. Remove the battery (see Maintenance above)
3. Reinstall the battery.
User may replace external components such as the inlet fi lter, battery, screen protector, and/or belt clip. Service must be performed by
SKC to maintain performance and intrinsic safety rating. Warranty is void if pumping compartment is opened by user.
www.skcinc.com
Form 40017 Rev 1906
Page 24 of 26
Page 25
Accessories/Replacement Part Ordering
Accessories Cat. No.
Single Charger, USB, magnetic connector, supplied as cable and wall cube with U.S. plug100-240 V
5-port USB Hub with power cable (U.S. plug), requires up to 5 single charging cables available as Cat. No. P75739
100-240 V
Quick-connect Bag Sampling Adapter, installs on pump exhaust port for secure connection to tubing for bag sampling,
accepts 1/4-inch OD PTFE tubing
Single Kit Case, nylon, with shoulder strap
5-pack Kit Case, Pelican
DataTrac Pro for Pocket Pump TOUCH USB Bluetooth Adapter, required for free software download and use of
DataTrac Pro software
Low Flow Tube Holder with cover (non-adjustable)
with A size cover, 6-mm OD x 70-mm L
with B size cover, 8-mm OD x 110-mm L
with C size cover, 10-mm OD x 150-mm L
with D size cover, 10-mm OD x 220-mm L
Constant Pressure Controller for multiple-tube sampling in constant fl ow
Adjustable Multiple-tube Low Flow Holders
Dual (holds up to 2 tubes)
Tri (holds up to 3 tubes)
Quad (holds up to 4 tubes)
220-300
220-400
220-200
224-903
224-915
877-94
222-3-1
222-3L-1
222-3XL-1
222-3XD-1
224-26CPC-10
224-26-02
224-26-03
224-26-04
Tube Covers
A (6-mm OD x 70-mm L)
B (8-mm OD x 110-mm L)
C (10-mm OD x 150-mm L)
D (10-mm OD x 220-mm L)
Replacement Battery Pack,*‡ Li-IonP76303
Single Charging Cable, USB, without wall cube
Low Flow chek-mate Calibrator, 20 to 500 ml/min, includes a 9-volt alkaline battery and
NIST-traceable calibration certifi cate
‡ Replacing baƩ eries with non-approved or repaired/rebuilt baƩ ery packs voids any warranty and UL LisƟ ng for intrinsic safety.
* Li-Ion baƩ eries are subject to special shipping regulaƟ ons
* Li-Ion Battery Testing and Shipment
Rechargeable lithium-ion batteries for use with SKC sample pumps have been tested in accordance with the UN Manual and are proven to meet requirements of each test in the UN
Manual of Tests and Criteria, Part III, subsection 38.3. The batteries are rated below 100 watt-hours (Wh).
Consult with your carrier for more information on Lithium Battery Shipping Regulations UN 3480 and UN 3481 or visit SKC’s website for more information at
www.skcinc.com/catalog/pdf/instructions/1921.pdf
224-29A
224-29B
224-29C
224-29D
P75739
375-00205N
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive
remedy. To view the complete SKC Limited Warranty and Return Policy, browse h p://www.skcinc.com/warranty.
www.skcinc.com
Form 40017 Rev 1906
Page 25 of 26
Page 26
Appendix
Appendix: Performance Profi le
Flow Range in Constant
Flow Mode
Compensation Range in
Constant Flow Mode
Pressure Range in Constant
Pressure Mode
Flow Control System Corrects for changes in back pressure, temperature, and atmospheric pressure
PowerRemovable rechargeable Lithium-ion (Li-Ion) battery, 3.6 V, 2.6 Ah, 9.36 Wh or USB charger
Run Time (Li-Ion)20+ hrs at 500 ml/min up to 20 inches water back pressure
Charging Time
(varies with battery capacity
and level of discharge)
AccuracyConstant fl ow control: ± 5% of set-point or ± 3 ml/min, whichever is higher
Temperature RangesOperating: 32 to 113 F (0 to 45 C)
Humidity Range≤ 95% RH, non-condensing
AltitudeCorrects fl ow for changes in temperature (32 to 113 F [0 to 45 C]) and ambient pressure up to
Flow FaultAfter 3 to 10 seconds of restricted fl ow, pump stops running, elapsed time stops, status LEDs fl ash
Case MaterialIP 64-rated case, polycarbonate with rubberized anti-static overmolding
Communications with PCLow-energy Bluetooth, requires DataTrac Pro for Pocket Pump TOUCH USB Bluetooth Adapter
TubingRequires 1/4-in ID tubing
† Tested using 37-mm 0.8 μm MCE fi lter with new pump and ba ery. Pump performance may vary.
20 to 500 ml/min
20 to 500 ml/min up to 20 inches water back pressure
1 to 20 inches water
†
, extended run times available with
charger
≤ 10 hrs (≤ 4.5 hrs to 80%)
Atmospheric pressure: ± 0.3 in Hg
Temperature: ± 1.0 C
Real-time clock: ± 1%
Constant pressure mode: ± 0.5 inch water
Charging: 32 to 113 F (0 to 45 C)
Storage: -4 to 113 (-20 to 45 C)
15,000 feet (4572 meters) above and down to 2300 feet (701 meters) below sea level
red, and pump displays fault icon. After 20 seconds in fault, auto-restart is attempted up to 5 times
unless full airfl ow is restored prior to that. If full airfl ow is not corrected after 5 auto-restarts within
5 minutes, the pump ends the run and displays Sample Summary.
atmospheric pressure, back pressure, programmed run remaining time, and elapsed run time
Class I, Groups A, B, C, D; Class II, Groups E, F, G; Class III, T4; Class I, Zone 0, Group IIC T4;
Exia -20°C ≤ Ta ≤ 45°C
• EX II 1G Exia IIC T4 Ga
• IECEx UL 16.0113
• DEMKO 16ATEX 1600
• CE 0539
• RoHS compliant
Cat. No. 877-94
E62011
www.skcinc.com
Form 40017 Rev 1906
Page 26 of 26
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.