SKC Disposable Parallel Particle Impactor Operating Instructions Manual

Page 1
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
SKC User-loaded Disposable
Parallel Particle Impactors (PPI)
The patented† impaction-based SKC Disposable Parallel Particle Impactor (PPI collection effi ciency curves for respirable and thoracic dust specified by ISO 7708/CEN and adopted by ACGIH, CEN, and other occupational hygiene organizations. Performance data of the PPI Samplers relative to the ISO 7708/CEN criteria was published in the Journal of Physics, Conference Series 151, 2009, and the data was made part of the OSHA Docket used to develop the OSHA fi nal silica rule (fi nal rule page 16439). The thoracic model meets the requirements of NIOSH Method 5524 for metalworking fl uids and compounds with ACGIH Disposable PPI Samplers off er the convenience of single use to eliminate sampler assembly and cleaning; small size for worker comfort, even under helmets or other PPE; and a choice of fl ow rates for maximum fl exibility in pump options, sample duration, and contaminant concentration. These instructions are for PPI models that
are NOT preloaded with fi lters and supports; these models contain only impaction substrates.
Performance Profi le
Sampling Rate: 2 L/min respirable or thoracic and 4 or 8 L/min respirable Sample Pump: • Universal or AirChek
• Leland Legacy for 8 L/min
Sample Time: Dependent on method used. Note: SKC tests indicate that a particulate
Sample Media 37-mm, 5.0-μm PVC lter, 37-mm, 2.0-μm PTFE lter* (recommended): (NIOSH 5524), or 37-mm, 0.8-μm MCE lter
Impaction Four 3/8-in diameter pre-oiled porous plastic discs Substrate: (assembled in all Disposable PPIs) Analysis: Gravimetric and/or chemical Body Material: Conductive ABS plastic Dimensions: Height: 4.25 in (10.8 cm) - clip to exhaust Diameter: 1.8 in (4.6 cm) Depth: 1.2 in (3 cm) Weight: 1.1 oz (31.2 gm) Shelf-life: 18 mos from date of manufacture
† U.S. Patent No. 7,073,402 * Back pressure on PTFE fi lters can vary within the same lot.
®
) Samplers are designed to match precisely the
®
thoracic TLV®s.
®
Series for 2 and 4 L/min
mass of up to 6.8 mg on the four impaction substrates would not aff ect PPI performance. This amount is equivalent to sampling for 6 hours at 4 L/min in environments where respirable mass concentration is 4.76 mg/m
3
and equals 50% of total dust. However, labs have reported to SKC that they prefer no more than 2 mg on the fi lter for analytical reasons. Therefore, SKC recommends that you work with your lab to determine optimum sample times for your unique sampling conditions.
Use cellulose or stainless steel screen for support
Page 2
Principle of Operation
SKC Disposable PPI Samplers are impaction-based fi lter samplers that perform precise size-selection for either thoracic or respirable dust, depending on the model. PPI Samplers contain four small impactors in the inlet section of the device. Each impactor features a unique 50% cut-point to target a specifi c one-quarter segment of the ISO/CEN curve resulting in a precise fi t along the entire curve. A sample pump operating at 2, 4, or 8 L/min (2 L/min only for thoracic) pulls air through the inlet nozzle of each impactor in the inlet plate. Particles larger than each impactor’s 50% cut-point are scrubbed and retained by impaction onto the porous oiled impaction substrate contained in each impactor. Smaller particles continue to the standard 37-mm collection fi lter for analysis.
Dust enters here
Inlet plate (contains 4 impac­tors with unique cut-points preloaded with 4 impaction substrates and sonically welded)
Standard 37-mm collection lter (user installed)
Protective end cap
Exhaust to pump
Base plate
PPI Performance
Collection effi ciency of the 2, 4, and 8 L/min
respirable PPI Samplers compared to the
ISO respirable curve
37-mm support pad (user installed)
Collar clip
Collection effi ciency of the 2 L/min
thoracic PPI Sampler compared to the
ISO thoracic convention
2
Page 3
Media and Sampler Preparation
234
Media Preparation
Condition and weigh fi lters according to method used. Record the weight as the pre-sample weight.
Sampler Preparation
The PPI will arrive with the inlet plate shrink wrapped and base plate separated.
1. Remove shrink wrap from inlet plate.
2. Using forceps, insert fi rst a 37-mm support pad and then a conditioned and weighed 37-mm collection fi lter (if using gravimetric analysis) into the base plate.
3. Lay base plate on a fl at surface, align inlet plate with base, and press down rmly to achieve an even seal.
4. Write sample ID on sampling label. Adhere sampling label to bo om of base plate.
5. Remove protective end cap from exhaust.
Technical Tidbits:
Use forceps to carefully insert or remove collection lter and support pad. See Accessories for forceps.
3
Page 4
Calibration and Sampling
As the particle load on the fi lter increases during sampling, the pressure
Tip
drop will also increase. Therefore, use a compensating sample pump such as the AirChek Series or Leland Legacy depending on fl ow rate requirements.
Calibration
Calibrate pump fl ow rate with a representative Disposable PPI (support and fi lter loaded) in line. See pump and calibrator operating instructions.
1. Ensure the pump has run for 5 minutes before calibrating.
2. Ensure representative impactor is loaded with a support and collection fi lter and that it is fully assembled. See Sampler Preparation on page 3.
3. Align the bo om of the calibration adapter with the inlet plate of an assembled, loaded representative Disposable PPI and press down firmly until the calibration adapter’s O-ring is engaged and creates an even seal.
4. Use fl exible tubing to connect the exhaust of the Disposable PPI to the inlet of a sample pump.
5. Use fl exible tubing to connect the inlet of the calibration adapter to the suction port (outlet) of a calibrator.
6. Calibrate to 2 L/min for the 2 L/min respirable and thoracic model PPIs, 4 L/min for the 4 L/min respirable model, or 8 L/min for the 8 L/min respirable model. Follow the calibration instructions in the pump and calibrator operating instructions.
7. When calibration is completed, disconnect the tubing from the calibrator and calibration adapter.
8. Grasp the Disposable PPI with one hand and the calibration adapter with the other hand. Firmly pull/twist to remove the calibration adapter from the Disposable PPI inlet plate.
9. Replace the representative Disposable PPI used to set the fl ow with a new loaded Disposable PPI for sample collection.
Using excessive pressure to seal the calibration adapter to the Disposable PPI may make the calibration adapter diffi cult to remove.
Sample
pump
Calibration
adapter
Flexible
tubing
Disposable
PPI
Flexible
tubing
Calibrator
4
Page 5
Sampling
3
6-8
10
SKC tests indicate that a particulate mass of up to 6.8 mg on the four impaction substrates would not affect PPI performance. This amount is equivalent to sampling for 6 hours at 4 L/min in environments where respirable mass concentration is
4.76 mg/m that they prefer no more than 2 mg on the fi lter for analytical reasons. Therefore, SKC recommends that you work with your lab to determine optimum sample times for your unique sampling conditions.
3
and equals 50% of total dust. However, labs have reported to SKC
1. As per good industrial hygiene practice, replace representative sampler used for calibration with a new, unused loaded sampler. See Media and Sampler Preparation.
2. Record sample start time on label.
3. Clip Disposable PPI onto a worker’s collar or lapel in the breathing zone or in the area to be sampled.
4. Clip sample pump at the worker’s waist or close to the Disposable PPI.
5. Use fl exible tubing to a ach the Disposable PPI exhaust to the inlet of the sample pump.
6. Turn on pump and record pertinent sample data.
7. After the desired sample time has elapsed, turn off the pump and unclip sampler from sampling location.
8. Apply Post-use label to inlet plate to seal the sampler.
9. Record sample stop time on label on the bo om of sampler.
10. Disconnect sampler from pump and reinstall
10
protective end cap on the exhaust.
11. Reinstate calibration train with representative Disposable PPI and verify fl ow rate. See Calibration.
5
Page 6
Sample Transport and Analysis
Package and transport samples and blanks to an accredited laboratory for gravimetric or chemical analysis.
References
Trakumas, S., Hall, P., Personal Respirable Sampler Containing Four Impactors Arranged in Parallel, Abstracts of 23rd Annual AAAR Conference, Atlanta, GA, 2004, p. 78
Trakumas, S., Salter, E., “Parallel Particle Impactor - Novel Size-selective Particle Sampler for Accurate Fractioning of Inhalable Particles,” Journal of Physics: Conference Series 151 (2009), 16 pp., 012060, www.skcinc.com/instructions/Parallel Particle Impactor Paper.pdf
Reference 2 is an author-created, non-copyrighted version of an article accepted for publication in the Journal of Physics; Conference Series 151. IOP Publishing Ltd. is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The defi nitive publisher authenticated version is available online. Go to h p://dx.doi.org, enter doi: 10.1088/1742-6596/151/1/012060.
Trakumas, S., “High-fl ow Personal Respirator Dust Sampler for Increased Sensitivity,” Poster 261, AIHce 2010, Denver, CO
Trakumas, S., “High-fl ow Personal Sampler to Monitor Exposure to Respirable Crystalline Silica at New Lower TLV,” IOHA 2010 8th Conference Book of Abstracts, Rome, p. 59
Trakumas, S., Salter, E., “High-Flow Personal Sampler to Monitor Exposure to Respirable Crystalline Silica at New Lower TLV” PowerPoint Presentation
OSHA Final Rule on Respirable Crystalline Silica, www.osha.gov/silica/
ISO 7708:1995 (2008), Air Quality — Particle Size Fraction Defi nitions for Health-related Sampling, www.iso. org, search on 7708
Stacey, P., Thorpe, A., and Echt, A., “Performance of High Flow Rate Personal Respirable Samplers When Challenged with Mineral Aerosols of Diff erent Particle Size Distributions,” Ann. Occup. Hyg., 60, 2016, pp. 479- 492, h p://annhyg.oxfordjournals.org/content/60/4/479.full.pdf
Görner, P., Simon, X., Boivin, A., Bau, S., “Sampling Effi ciency and Performance of Selected Thoracic Aerosol Samplers,” Annals of Work Exposure and Health, 2017, Vol. 61, No. 7, 784-796
6
Page 7
Ordering Information
User-loaded Disposable Plastic PPI Samplers, select the PPI for the desired convention. Designed for one-time use. Cat. No.
User-loaded Disposable PPI Samplers contain four porous
disc impaction substrates. Require collection fi lter and support; see information below and select based on application
Respirable PPI (red), 8 L/min, plastic Respirable PPI (orange), 4 L/min, plastic Respirable PPI (gold), 2 L/min, plastic Thoracic PPI (blue), 2 L/min, plastic
Recommended Collection Filters for User-loaded Disposable PPI,
required for sampling. Select a fi lter based on your application.
PVC Filters, 37 mm, 5.0-μm pore size, pk/100 225-5-37 PTFE Filters,* 37 mm, 2.0-μm pore size, with laminated PTFE support,
for metalworking uids (NIOSH 5524), pk/50 Contact SKC MCE Filters, 37 mm, 0.8-μm pore size, pk/100 225-5
Filter Supports, required for User-loaded Disposable PPI sampling.
Select either cellulose or stainless steel.
Support Pads, cellulose, 37 mm, pk/100 225-27 Support Pads, stainless steel, 37 mm, wide mesh, ea 225-26 Accessories Calibration Adapter for Disposable PPI 225-389 Forceps, stainless steel, non-serrated at tips 225-8371
* Back pressure on PTFE fi lters can vary within the same lot.
225-384 225-387 225-385 225-386
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.
7
Page 8
www.skcinc.com
8
Form 38055 Rev 1901
Loading...