SKC chek-mate 375 series, chek-mate 375-0550, chek-mate 375-0550N, chek-mate 375-00205N, chek-mate 375-00205 Operating Instructions Manual

chek-mate Flowmeter
Cat. No. 375 Series
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA • Tel: 724-941-9701 • www.skcinc.com
Inlet
Low battery
indicator
LCD screen
On/off
button
Introduction
Front View
Outlet
CalChek interface socket (Cat. Nos. 375-0550N and
375-0550 only)
Battery
Figure 1. chek-mate Flowmeter Overview
Battery
compartment
cover
Back View
The chek-mate® Flowmeter (Figure 1) is designed specifi cally for use with air sample pumps and is available in medium fl ow and low fl ow models:
Medium flow chek-mate Flowmeters with CalChek Cat. Nos. 375-0550N and 375-0550 have a flow measurement range of
0.5 to 5 L/min and provides CalChek capability. Volumetric accuracy is 1% of reading for 0.75 to 5 L/min and 2.5% of reading for
0.5 to 0.75 L/min. This model has a CalChek serial interface that enables direct communication with CalChek-ready pumps in the chek-mate flow range for single-point (Single) and multiple-point (Full) automatic flow calibration.
Low Flow chek-mate Flowmeters Cat. Nos. 375-00205N and 375-00205 have a flow measurement range of 20 to 500 ml/min.
Volumetric accuracy is 1% of reading for 50 to 500 ml/min and 2.5% of reading for 20 to 50 ml/min.
Checking Package Contents
See below to verify that you received all items associated with the Cat. No. ordered. If you are missing items, contact SKC at 800-752-8472 (U.S. only) or 724-941-9701.
If You Ordered Cat. No. Your Package Should Contain
375-0550N Flowmeter with 9-volt alkaline battery and NIST-traceable calibration certi cate
375-0550 Flowmeter with 9-volt alkaline battery and UKAS-traceable calibration certi cate
375-00205N Flowmeter with 9-volt alkaline battery and NIST-traceable calibration certi cate
375-00205 Flowmeter with 9-volt alkaline battery and UKAS-traceable calibration certi cate
Required Equipment
; 1/4-inch ID tubing
www.skcinc.com
Form 39013 Rev 1905
Page 1 of 6
Getting Started
Notes and Cautions
• Allow chek-mate to equilibrate for at least 10 minutes after moving it from one temperature extreme to another before use.
• The chek-mate case is rated IP40, not as waterproof or splashproof; do not use it where water can enter the case.
• Ensure that fl uids do not enter either the inlet or outlet.
• Protect the chek-mate from direct exposure to sunlight to prevent it from heating the case.
• Avoid extended use of the chek-mate in areas with high levels of airborne particulates. If this cannot be avoided, use an external inlet fi lter with high collection effi ciency (i.e., one that collects the majority of all dust particles) and low back pressure. SKC does not supply this.
• Use only the specifi ed disposable or rechargeable battery types.
• Ensure that discharged batteries are not left in the battery compartment for long periods to prevent damage caused by electrolyte leakage.
• There are no user-serviceable parts in the chek-mate fl owmeter. Opening the chekmate case will void the product warranty and could affect the instrument calibration. An anti-tamper warning label is fi tted to the case to indicate that the case has been opened.
• Failure to follow these guidelines will void the warranty.
Turn On/Off
Allow the chek-mate to equilibrate for at least 10 minutes in the location where it will be used to ensure that it has stabilized to
ambient temperature.
To turn on the chek-mate, press the on/off bu on on the front. See Figure 1.
Read Battery Status on LCD
When the ba ery voltage drops below 8 volts as it nears the end of its life, LOW BATTERY will be displayed in the upper left corner of the LCD (Figure 1). Replace the ba ery. See Maintenance.
When the battery voltage drops below 7 volts, the LOW BATTERY message displayed on the LCD will fl ash on and off and the chek-mate
will turn off automatically. If the chek-mate is turned on again with the battery still in this condition, the LCD will display bAtt LO ten times and then turn off automatically.
Operation
Set/Calibrate Flow Rate
Allow the chek-mate to equilibrate for at least 10 minutes after moving it from one temperature extreme to another.
Protect the chek-mate from direct exposure to sunlight to prevent it from heating the case.
Ensure that uids do not enter either the inlet or outlet.
Avoid extended use of the chek-mate in areas with high levels of airborne particulates. See Notes and Cautions.
1. Turn on the fl owmeter by pressing the on/off bu on on the front. The LCD screen will cycle through the startup messages, “On” followed by the upper limit of the fl owmeter range (“5.0 L” for the medium ow model or “0.5 L” for the low ow model), and will then indicate the current fl ow rate or “_ _ _ _” if there is no airfl ow or the fl ow rate is below the minimum display value. Note: The minimum display value is dependent on the atmospheric
conditions, but at 68 F (20 C) and 1 atm (1013.25 mbar), the values are 5 ml/min for the low fl ow chek-mate or 0.3 L/min for the medium fl ow chek-mate. Minimum display values will be higher
when the fl owmeter is used at higher altitudes and temperatures.
Note: The chek-mate does not zero the fl ow reading at startup, so it is also possible to turn it on with
the sample train already connected and airfl ow passing through the fl owmeter without aff ecting the accuracy of the fl ow reading.
2. Connect the fl owmeter outlet to the inlet of the calibration train using a length of fl exible 1/4-inch ID tubing, and if required by the sampler, a calibration adapter. See Figure 2.
www.skcinc.com
Figure 2. Calibration Train
Form 39013 Rev 1905
Page 2 of 6