5
134
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
www.skcinc.com
AirChek® XR5000 Sample Pumps
5. Orient battery holder properly with battery pack housing (black and red wires and battery terminal toward front of housing with SKC logo).
To avoid breakage, orient battery holder properly and do not force the battery holder into the battery pack housing.
Front of housing with
SKC logo
To retain display data, ensure pump is placed in Hold before disconnecting the battery pack. Display data will not be retained if battery is removed while pump is running.
Remove alkaline batteries from pump before storage to prevent damage from battery leakage.
Removing Existing Battery Pack
1. Remove two screws on bottom of battery pack housing.
2. Pull battery pack housing away from pump case.
3. Holding battery pack housing tightly in one hand, place a nger from the other hand through the loop on top of the battery pack. Pull upward fi rmly to remove the battery holder from the housing.
Installing New Batteries
4. If replacing existing batteries, fi rst remove batteries from holder. Place new batteries into the holder in the following polarity arrangement.
Resetting Pump to Manufacturer Settings
If the pump does not operate as expected, perform the following procedure:
1. Remove the battery pack (see Removing Existing Battery Pack).
2. On the pump keypad, press and hold the Ò key and the T key simultaneously while attaching the pump to the new battery pack. The LCD should display the software version number (525X).
Do not release hold on the two keys until the pump is rmly attached to the battery pack.
3. Release the Ò and T keys.
4. Press the Ò key 2 times. The LCD should read 0. If it does not, separate battery pack from pump and repeat Steps 1 through 4 until successful.
5. Replace and tighten two screws on the bottom of the battery pack housing.
For safe operation in hazardous locations, ensure the pump label contains the P85002. Use of any other battery pack (including alkaline) or device to power the pump voids the UL Listing for intrinsic safety.
Failure to follow warnings and cautions voids any warranty.
logo and the battery pack label contains Cat. No. P85004 or
Polarity is marked
inside the
ba ery holder.