Battery Replacement Notes:
• To retain history, ensure the pump has gone to SLEEP after the last run. The battery pack should not be removed while the pump is running.
• Replacing the battery pack may erase scheduled runs programmed with DataTrac 2000 software. After replacing the battery pack, use the software to reprogram the scheduled runs in the AirChek 2000 pump.
1. To remove a battery pack, release the battery pack by removing the two security screws located on the bottom of the battery pack. Pull the battery pack away from the pump body.
4. Charge the new battery pack attached to or detached from the pump using only
the charging jack on the back of the battery pack.
Computer
interface
port
Covers
Ba ery
charging
jack
Ba ery a ached to pump Ba ery detached from pump
Cautions
• Use of a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.
• Do not charge or operate pump from charger in hazardous locations!
• Use only an SKC-approved charger and battery pack designed for the AirChek® 2000 Sample Pump to ensure reliable performance. Failure to do so voids intrinsic safety and any warranty.
• Failure to follow warnings and cautions voids any warranty.
Cover
Ba ery
charging
jack
2. To install a battery pack, carefully align the power connector on the replacement battery pack with the terminal on the bottom of the pump base plate and push the battery pack into place.
3. Replace and tighten the two security screws removed in Step 1.
Power
connector
Terminal
Ba ery
pack
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy , which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.
Pump base
plate
www.skcinc.com