Ò (star key) Scrolls through parameters in user setup functions.
ST (up/down arrow keys) Increase or decrease fl ow rate, timed run, and run delay time.
Key Sequences
TÒPress keys individually.
[ST] Press keys simultaneously. Toggles between Run and Hold
ÒSTÒSecurity code to access user setup functions. With pump in a
Operation
• Pump On Press and hold Ò.
• Pump Off Press and hold Ò through countdown. Auto-off will shut down
• Mode Change Press [ST] to toggle between Run and Hold.
• Keypad Lock Press T 5 times quickly to activate. Press T 5 times quickly
• Run/Hold With pump in a non-running state (no fl ashing blue LED),
Accessing User Setup Functions
•Entering User Setup Functions
With pump in a non-running state (no fl ashing blue LED), press ÒSTÒ.
•Exiting User Setup Functions
Press [ST]. Pump is ready. Press [ST] to run the pump or to start a run delay.
User Setup Functions
To navigate while in user setup functions, press Ò until the desired function displays.
and exits user setup functions.
non-running state (no fl ashing blue LED), press keys in
sequence.
pump after 5 minutes without activity.
to deactivate.
press [ST] to run pump. Press [ST] to Hold pump when
completed.
FunctionWhen LCD Displays User ActionResult
Clear Accumulated
Run Time
Function only available
when accumulated run
time exists.
Adjust Flow Rate*“---” and fl ashing
Set Timed Run
Set Run Delay
* Changing fl ow rate in user setup functions will not clear accumulated run time.
† Changing timed run and/or run delay settings in user setup functions will clear accumulated run time.
CLr and fl ashing
Hold
ADJ Flow
†
Flashing Set Timed
Run and min
†
Flashing Set Run
Delay and min
Press [ST].
Press S or T.
Press [ST] to exit
functions.
Press S or T.
Press [ST] to exit
functions.
Press S or T.
Press [ST] to exit
functions.
Clears run and run
time and exits functions. Press [ST] to
run pump.
Flow increases/
decreases. Press
[ST] to run pump.
Minutes increase/
decrease. Press
[ST] to run pump.
Minutes increase/
decrease. Press
[ST] to start run
delay. Blue LED
fl ashes. Pump starts
after delay elapses.
www.skcinc.com
Description
SKC AirChek XR5000 Sample Pumps are designed to offer users enhanced battery
power and easy operation in a lightweight pump that provides accurate airfl ows from
5 to 5000 ml/min.
• Three battery options provide fl exibility and economy for different applications
including long run times.
• The large three-button keypad and straightforward user setup functions offer userfriendly conveniences.
AirChek XR5000 pumps feature a patented* isothermal fl ow sensor that measures fl ow
directly and acts as a secondary standard, constantly maintaining the set fl ow rate. A
built-in sensor compensates for changes in temperature that occur after calibration.
* U.S. Patent No. 5,892,160
Inlet Port with
Protective Filter
Battery Charge
Level Indicator
Bright blue
pump status
LED
Easy-to-read
Liquid Crystal
Display (LCD)
Battery Pack
(3 options
available)
Not shown:
Beltclip (back)
Battery
Charging Jack
(back)
Large,
simple
Operating
Keypad
AirChek XR5000 Air Sampling Pump
1
Performance Profi le
Flow Range: 1000 to 5000 ml/min (5 to 500 ml/min requires optional
low fl ow adapter kit)
Compensation Range:5000 ml/min at 10 inches water back pressure
4000 ml/min at 20 inches water back pressure
2000 ml/min at 50 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)
For extended run times, the pump may be operated while
attached to the charger.
Timed Run, Run Delay,
and Continuous Run 1 to 9999 minutes (6.8 days). If run time exceeds
Display Range: 6.8 days, timer display rolls over.
Flow Fault: If pump is unable to compensate for > 15 seconds due to
excessive back pressure, the pump stops, displays fl ow
fault icon, and holds run time display. Auto-restart is
attempted every 15 seconds up to 5 times.
or
Standard Li-Ion (2 cell), rechargeable, 7.4 V, 2.2-Ah
or
Alkaline (6 cell), disposable, size AA, 1.5 V (nominal),
Battery Recharge Time:Standard Li-Ion (2 cell): approximately 4 hrs
(with SKC-approved chargers;
varies with battery capacity
and level of discharge)
Size: High-power Li-Ion
and alkaline models: 5.5 x 3 x 2.3 in (14 x 7.6 x 5.8 cm)
Weight: High-power Li-Ion: 21 oz (0.6 kg)
Case: Anti-static plastic
RFI/EMI Shielding: CE marked for RFI/EMI protection
Approvals: •
• RoHS compliant
Cautions:
• For safe operation in hazardous locations, ensure the pump label
contains the logo and the battery pack label contains Cat.
No. P85004 or P85002. Use of any other battery pack (including
alkaline) or device to power the pump voids the UL Listing for
intrinsic safety.
• Use onl y the charger and battery packs designed for the AirChek
XR5000 pump to ensure reliable performance. Failure to do so voids
any warranty.
• Remove alkaline batteries from pump before storage to prevent damage from battery leakage.
• Use onl y SKC-approved parts to ensure r elia ble performance and to
maintain the UL Listing for intrinsic safety . F ailure to do so voids any
warranty.
• Failure to follow warnings and cautions voids any warranty.
capacity, 32.6 Wh (Cat. No. P85004 for UL Listed pump)
capacity, 16.3 Wh (Cat. No. P85002 for UL Listed pump)
Cat. No. P75715 - not UL Listed for intrinsic safety
High-power Li-Ion (4 cell): approximately 8 hrs
Standard Li-Ion model: 4.3 x 3 x 2.3 in (10.9 x 7.6 x 5.8 cm)
Standard Li-Ion model: 16 oz (0.45 kg)
Alkaline model: 17 oz (0.48 kg)
for use in hazardous locations. Models that are
UL Listed for intrinsic safety contain the logo on
the label. These models must be used with battery pack
Cat. No. P85004 or P85002 to maintain the UL intrinsic
safety listing.
3
Setup
Charging the Lithium-Ion Battery Pack
For models containing a lithium-ion battery pack only.
Power
supply
Shown with single charger (Cat. No. 222-241). A
fi ve-station charger is available; see Accessories
on page 21. Follow charger instructions.
STOP!
Completely charge a new battery pack
before operating the pump. It may be
necessary to charge the battery a few
times before maximum battery capacity
is achieved.
Intrinsic safety circuitry inside the
battery causes the pump to selfdischarge during storage. Charge
Li-Ion pump battery completely
before calibration and sampling to
achieve optimum pump operation.
Jack
Charging
unit
AirChek XR5000 charging train
with single charger
Battery
charging
jack
1. For a complete charge, ensure the pump is not running. Insert plug on charging
unit into the battery charging jack on back of pump. Ensure plug is oriented so
that the arrow on the plug is facing upward.
Ensure proper orientation of charging cable before plugging it into
the charging jack. Improper orientation/contact will short circuit the
battery and voids any warranty.
Short circuiting the battery pack will render it immediately inoperative.
2. Insert plug on power supply into jack on charging unit.
3. Pull locking tab to side and insert appropriate
wall plug into power supply. Release locking
tab. Plug power supply into a wall outlet.
The standard 2-cell Li-Ion battery pack will
recharge in approximately 4 hours. The highpower 4-cell Li-Ion battery pack will recharge in
approximately 8 hours. Run pump for 5 minutes
after charging is complete.
Note
After charging the battery pack, it is good practice to run the pump for
approximately 5 minutes before calibrating. This ensures the battery is
in more steady-state conditions and improves the agreement in pre and
post-sampling calibrations.
4
Interchangeable wall plugs
insert into power supply.
Setup
Reading the Charging Status LED on the Single Charger
The Li-Ion Charging Unit (Cat. No. P22300) indicates battery charge status via
an LED on the unit that blinks in specifi c patterns. Observe the LED steadily for
> 5 seconds to read charge status.
LED ActionCharge Status
ON
Ò
steady
ON
Ò
2 sec
OFF
{
2 sec
OFF
{
.25 sec
ON
Ò
.25 sec
ON
(Repeats)Approximately
Ò
2 sec
OFF
(Repeats)Charge completed
{
2 sec
Charge in progress
80% charged
Power supply
jack
Charge status
LED
Note
The battery pack may be kept on SKC-approved Li-Ion battery chargers for
an indefi nite time.
Cautions:
• Do not charge or operate pump from charger in hazardous
locations.
• Use only the charger and battery packs designed for the AirChek
XR5000 pump to ensure reliable performance. Failure to do so will
void any warranty.
• For safe operation in hazardous locations, ensure the pump label
contains the
No. P85004 or P85002. Use of any other battery pack (including
alkaline) or device to power the pump voids the UL Listing for
intrinsic safety.
• Tampering with the battery pack or using a repaired or rebuilt
battery pack voids any warranty and UL Listing for intrinsic safety.
• Do not open, disassemble, short circuit, crush, incinerate, or
expose the battery to fi re or high temperatures.
• Remove alkaline batteries from pump before storage to prevent damage from battery leakage.
• Use only the SKC-approved charger for this pump. Use of an
unapproved charger may damage the battery and pump and voids
any warranty.
• Failure to follow warnings and cautions voids any warranty.
logo and the battery pack label contains Cat.
See http://www.skcinc.com/instructions/1756.pdf for more information
on SKC pump battery packs.
5