SKC AirChek2000 Operating Instructions Manual

Page 1
Operating Instructions
SKC Inc.
863 Valley View Road
Eighty Four, PA 15330 USA
Form #37740 Rev 0806
Page 2
Table of Contents
Description ...................................................................................................................................................1
Performance Profi le .................................................................................................................................... 2
Battery Operation
Charging the Battery Pack ................................................................................................................... 4
Battery Charge Level Indicator ............................................................................................................5
Replacing the Battery Pack .................................................................................................................5
Battery Eliminator ................................................................................................................................6
Introduction
Pump Display ...................................................................................................................................... 7
Keypad Basics .....................................................................................................................................8
Pump Activation ...................................................................................................................................8
Pump Operating Modes ....................................................................................................................... 9
Setup
Verifying Battery Charge Level .......................................................................................................... 10
Entering and Navigating the User Interface ....................................................................................... 10
Using User Interface - Level One ......................................................................................................10
Using User Interface - Level Two ......................................................................................................11
Resetting Run Time Data ..................................................................................................................12
Setting Pump Flow Rate .................................................................................................................... 12
Setting Sampling Time ......................................................................................................................13
Setting a Delayed Start ...................................................................................................................... 14
Deleting a Delayed Start .................................................................................................................... 14
Deleting a Sampling Time .................................................................................................................15
Deleting a DataTrac 2000 Program ................................................................................................... 15
Calibration
Flow Rate and Volume Display .........................................................................................................16
Verifying Flow Rate Using a Primary Standard .................................................................................16
Calibrating Using CalChek ................................................................................................................18
CalChek Error Chart ..........................................................................................................................22
Calibrating for Low Flow (5 to 500 ml/min) Sampling Applications .................................................... 23
Sampling
Sampling (1000 to 3250 ml/min) ........................................................................................................ 24
Low Flow Sampling (5 to 500 ml/min) ...............................................................................................24
Flow Fault .......................................................................................................................................... 25
Maintenance
Replacing the Inlet Port Housing and Filter .......................................................................................26
Service Policy .................................................................................................................................... 27
Accessories ...............................................................................................................................................28
DataTrac 2000 Software ............................................................................................................................30
UL Certifi cate .............................................................................................................................................31
Warranty .....................................................................................................................................................32
Index ...........................................................................................................................................................33
Indicates a reminder or note.
Notice: This operating instruction may not address all safety concerns (if any) associated with this product and its use. The user is responsible for determining and following the appropriate safety and health practices and regulatory limitations (if any) before using the product. The information contained in this document should not be construed as legal advice, opinion, or as a fi nal authority on legal or regulatory procedures.
Indicates a warning or caution.
Page 3
AirChek 2000 Quick Guide
Operation
Star Button Ò
Scrolls through run time data, display options, and sampling parameters during pump setup
Up and Down Arrow Buttons ST
Increase or decrease sampling parameters and toggle between display choices in setup
Button Sequence
TÒ = press buttons individually [ST] = press simultaneously when bracketed ÒSTÒ = security code, always press in sequence
Pump Activation
Press any button.
Mode Change
Press [ST] to toggle between Run and Hold.
Pump Setup Options
User Interface Level One—Set Up Flow Rate
To enter fl ow rate setup, pump must be in Hold. Press [ST] to Run pump and press ÒSTÒ within 10 seconds.
• Change and calibrate fl ow rate:
Flow rate and SET fl ash. Press S or T to change fl ow rate. Press Ò when fi nished; ADJ
displays and fl ashes. Press S or T to adjust the fl ow until pump and calibrator agree. When fi nished, press Ò until End displays. Press [ST] to save new setting.
User Interface Level Two—Set Up Sampling and Display
To enter sampling and display setup, pump must be in Run. Press [ST] to Hold pump and press ÒSTÒ within 10 seconds. Select from the following options:
Security Code ÒSTÒ
Must be pressed within 10 seconds of previous command
• Clear accumulated data:
Press Ò until CLr displays and then press [ST]. Press Ò until End displays and then
press [ST].
• Change temperature scale (F/C):
Press Ò until temperature displays. Press S or T to switch units. Press Ò until End
displays and then press [ST] to save new setting.
• Change atmospheric pressure scale (mm Hg/ins Hg/millibars):
Press Ò until pressure displays. Press S or T to switch units. Press Ò until End
displays and then press [ST] to save new setting.
• Change time scale (12 Hr/24 Hr):
Press Ò until 12 Hr or 24 Hr displays. Press S or T to switch units. Press Ò until End
displays and then press [ST] to save new setting.
• Change time of day:
Press Ò until time displays. Press S or T to change fl ashing hour. Press Ò to move to
minutes and press S or T to change. Press Ò until End displays and then press [ST] to save new setting.
• Change sampling time:
Press Ò until SГ 00 displays. Press S or T to change fl ashing digits. Press Ò until End
displays and then press [ST] to save new setting.
To exit without saving changes in User Interface Level Two, scroll to Esc and press [ST].
Page 4
Description
®
The SKC AirChek that combines lightweight compact design, computer-compatible circuitry, and a patented* isothermal fl ow sensor. AirChek 2000 can be used with optional DataTrac
®
2000 So ware and your PC to expand pump programmability and recordkeeping options. The result of extensive research and development, the AirChek 2000 Pump exemplifi es SKC’s commitment to quality and innovation in air sampling equipment.
2000 Pump is an advanced programmable sample pump
* U.S. Patent No. 5,892,160
Battery Status Icon
Battery Pack
Inlet Port with Protective Filter
Flashing LED Run Indicator
Liquid Crystal Display (LCD)
Keypad with Operating Buttons
AirChek 2000 shown at 75% of actual size
1
Not shown: Beltclip (back)
Battery Charging Jack (back)
Computer Interface Port (back)
Page 5
Performance Profi le
Flow Range: 1000 to 3250 ml/min
(5 to 500 ml/min requires optional low fl ow adapter kit)
Compensation Range: 3000 ml/min at 15 inches water back pressure 2000 ml/min at 30 inches water back pressure 1000 ml/min at 30 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min) 1.0 1.5 2.0 2.5 3.0 Filter/Pore Size (μm)
25-mm MCE, 0.8 6 9 12 15 18 25-mm MCE, 0.45 14 22 28 35 40 37-mm MCE, 0.8 2 3 4 5 6 37-mm PVC, 5.0 1 1 2 2 2.5
Compare the information in this table to pump compensation range to determine appropriate applications.
Accuracies: Timing: 1 min/month at 25 C
Atmospheric Pressure: ± 0.3 ins Hg
Flow Rate: ± 5% of set-point after calibration
Battery Charge Level Indicator: Icon displays at full, mid, and low charge (see page 5)
Temperature Range: Operating: 32 to 113 F (0 to 45 C) Charging: 32 to 113 F (0 to 45 C) Storage: -4 to 113 F (-20 to 45 C)
Altitude: Do not use pump beyond 7500 ft
Run Time: With battery pack, run time is 10 hours at
Sampling Time (SГ) AirChek 2000 sampling time can be set up to 999 minutes Setting: from the pump keypad or up to 30 days (43,200 minutes)
Time Display: Time of day in hours and minutes (12 or 24-hour
Volume Display: Continually updated, based on corrected ow rate
multiplied by sampling time
Protect sample pump from weather when in use outdoors.
2000 ml/min and up to 30 inches water back pressure; dependent on sample media used. See Table 1.
from a PC using DataTrac 2000 Software. During sampling, the pump LCD will display remaining run time (will count time down from a set number of minutes) when sampling time is set from the keypad. The pump LCD will show elapsed time in minutes when the pump is programmed from a PC. If run time is longer than 9999 minutes (6.8 days), the sample time displayed on the LCD will roll over to 1 after reaching 9999.
clock) with AM and PM indicators
2
Page 6
Performance Profi le
Flow Fault: If the pump is unable to compensate for longer than
Battery Pack: Removable rechargeable 4-cell NiCad battery pack,
4.8 V, 2.0-Ah capacity.
Charge Time: 6 hrs using PowerFlex Charger
Size: 5.6 x 3 x 2.3 in (14.2 x 7.6 x 5.8 cm)
Weight: 22 oz (624 gm)
RFI/EMI Shielding: RFI/EMI-shielded case, CE marked
Intrinsic Safety: UL and cUL Listed
Use only SKC-approved parts to ensure reliable performance and to maintain the UL Listing for intrinsic safety and the SKC warranty.
15 seconds due to excessive back pressure, the pump enters fl ow fault mode. During fl ow fault, the pump HOLDs, the fault icon appears on the display during the length of the fault, and the accumulated run time display is frozen and retained. After 5 minutes in fl ow fault, auto-restart is attempted every 5 minutes up to 10 times.
Table 1. AirChek 2000 Run Time in Hours with
NiCad Battery
Following are typical run times achieved when using a fully charged Nickel­Cadmium (NiCad) battery pack. Data is sorted by type of sample media. All run times are listed in hours.
Mixed Cellulose (MCE) fi lter, 0.8-μm pore size
Filter Diameter
Flow Rate (L/min)
2.0 24.2 15.2
2.5 20.4 12.4
3.0 17.7 **
37 mm 25 mm
Polyvinyl Chloride (PVC) fi lter, 5.0-μm pore size
Filter Diameter
Flow Rate (L/min)
2.0 28.2 21.8
2.5 27.0 22.0
3.0 22.6 18.2
** Filter back pressure exceeded pump capability during testing
37 mm 25 mm
Increases in back pressure during sampling due to buildup of sample on the fi lter can decrease ba ery life.
3
Page 7
Battery Operation
Installing the Battery Pack
To enhance battery life, SKC ships battery packs separate from the pump. Once
installed, completely charge battery pack before operating pump.
1. Carefully align the ba ery jack on the ba ery pack with the ba ery ter- minal on the bo om of the pump base plate and push the ba ery pack into place.
2. Insert and tighten two security screws provided.
3. Charge ba ery completely before operating pump (see Charging the Ba ery Pack).
Charging the Battery Pack
Ensure the pump is not running. Do not operate pump while charger is a ached. To charge the ba ery, insert the charging plug from the charger into the ba ery charging jack on the back of the pump. Plug the charger into a standard wall outlet. The PowerFlex Charger will fully recharge the ba ery in ≤ 6 hours (see Accessories on page 28 for charger). Follow charger instructions.
Using a non-approved charger voids the SKC warranty.
Tampering with the battery pack voids the SKC warranty and the UL Listing for intrinsic safety.
Do not charge or operate pump from charger in hazardous locations.
Ensure that the computer interface port is covered before and during charging.
Computer Interface Port
Battery Charging Jack
Ensure proper orientation of charging cable before plugging it into the charging jack. Improper orientation/contact will short­circuit the battery. See photo at right. Note the keyed side of the charging cable plug must mate with the keyed side of the pump charging jack. Inserting the plug does not take much force. If force is required, the plug is not properly oriented.
Short-circuiting the battery pack will render it immediately inoperative.
4
Page 8
Battery Operation
Battery Charge Level Indicator
Three bars indicate a full charge (normally appears a er charging), approximately 75 to 100%.
Two bars indicate that the ba ery is charged enough to operate the pump, approximately 25 to 75%.
One bar indicates ba ery charge is low (charge ba ery), approximately 1 to 25%.
Low Battery Fault
No bars and a fl ashing outline indicate a Low Ba ery Fault (pump will go into HOLD, and then SLEEP mode in 10 seconds). Run time data can be displayed by pressing any bu on. The pump will return to SLEEP in 10 seconds. This may be repeated.
When the pump stops due to a low ba ery and is le to stand for a period of time, one ba ery bar may appear. This false “recovery” will fall quickly if the pump is operated without recharging it. RECHARGE THE PUMP BEFORE SAMPLING.
Replacing the Battery Pack
Ba ery Replacement Notes:
• To retain pump history, ensure the pump has been allowed to go to SLEEP a er the last run. Pump history will be lost if the ba ery pack or AC power (ba ery eliminator) is removed while the pump is running. SKC recommends that data be downloaded to a PC using DataTrac 2000 So ware prior to removal of the ba ery or power.
• The fi rst 8 program steps will be retained in memory. Programs should be reloaded using DataTrac 2000 So ware a er replacing the ba ery pack.
1. Release the ba ery pack by removing the two security screws located on the bo om of the ba ery pack. Pull the ba ery pack away from the pump body.
5
Page 9
Battery Operation
2. Carefully align the ba ery jack on the replacement ba ery pack with the ba ery terminal on the bo om
Ba ery
terminal
Pump base
plate
of the pump base plate and push the ba ery pack into place.
3. Replace and tighten the two security screws removed in Step 1.
Use of a repaired or rebuilt battery pack voids the SKC warranty and the UL Listing for intrinsic safety.
Do not charge or operate the pump from the charger in hazardous locations!
Use only an SKC-approved charger and battery pack designed for the AirChek 2000 Sample Pump to ensure reliable performance and to maintain the UL Listing for intrinsic safety and the SKC warranty.
Use of any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety.
Ba ery
jack
Ba ery
pack
For more information on battery care and maintenance, go to
¡
http://www.skcinc.com/instructions/1756.pdf.
Battery Eliminator
The Ba ery Eliminator is an accessory that converts alternating current (AC) to direct current (DC) from which the pump can operate indefi nitely. The
Ba ery Eliminator should be used in non-hazardous locations only. See page 28 for ordering.
To use the Ba ery Eliminator, the ba ery pack must be removed from the pump (see Replacing the Ba ery Pack on page 5). The Ba ery Eliminator is comprised of two pieces, (1) a wall cube that converts AC voltage to DC voltage and (2) a power adapter that reduces DC voltage. The wall cube ts into a standard wall outlet and its plug end is inserted into the power adapter. The power adapter is  ed on the pump in place of the ba ery pack.
Use the battery eliminator to operate the pump in non-hazardous locations only.
Use of any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety.
6
Page 10
Introduction
Pump Display
AirChek 2000 LCD shown at 100%
HOLD: Flashes when the pump is in HOLD mode (see page 9)
SГ or S: Displayed when a sampling time is manually
VOL: Volume of air pumped
SET: Flashes when changing any setting
CLr: User Interface option that clears volume and run time
ESC: User Interface Level Two option that exits User
End: User Interface option that saves changes and exits
ADJ: Displayed during pump ow calibration (see page 11)
CAL: User Interface option that allows pump to be set up for
FLOW: Appears when ow rate is displayed
PROG: Displayed when a delayed start or a DataTrac 2000
programmed into pump memory (see page 13)
data (see page 12)
Interface without saving changes (see page 12)
User Interface (see page 12)
single or multiple-point calibration using the CalChek feature (see page 11)
program is loaded into pump memory
Display Icons
Flow Fault: Flashes during ow fault (see page 9)
Battery Charge Level: Shows battery charge level (see page 5)
7
Page 11
Introduction
Keypad Basics
The AirChek 2000 Pump operates by pressing various bu on sequences on the keypad located on the front of the pump housing.
S Up arrow
Ò Scrolls through run time data, display options, and sampling
parameters during pump setup
S Increases values and toggles between options
T Decreases values and toggles between options
[ST] When pressed simultaneously, displayed item is selected or
entered.
ÒSTÒ Security code that must be pressed in sequence within 10
seconds a er mode change to enter the User Interface. If the 10-second time limit is exceeded, the pump will remain in its current mode. Perform steps again to enter the User Interface (see Entering and Navigating the User Interface on page 10).
button
AirChek 2000 Keypad shown at 100%
Ò
Button
T Down
arrow
button
Pump Activation
• Press any bu on fi rmly to activate the LCD.
• Press [ST] to run the pump or to place a running pump in HOLD.
• Auto OFF switches pump to SLEEP a er fi ve minutes in HOLD with no
activity.
8
Page 12
Introduction
Pump Operating Modes
RUN
Pump is running; run time data is continuously updated in memory. The LCD shows real-time run time data. Press Ò to scroll through the parameters. Run
time
and volume stored in memory will continue to accumulate unless
reset (see page 12).
To change the pump from HOLD to RUN, press [ST]
HOLD
Pump is off and run time data is stored. Run time data readings are retained and displayed on the LCD.
To change the pump from RUN to HOLD, press [ST]
SLEEP
A er fi ve minutes in HOLD, the pump automatically enters SLEEP. The LCD shuts down and the electronic circuitry enters a low-power state.
Press any bu on to change pump from SLEEP to HOLD
Flow Fault
If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters fl ow fault mode. During fl ow fault, the pump HOLDs, the fault icon appears on the display during the length of the fault, and the accumulated run time display is frozen and retained. As part of the fl ow fault feature, the pump a empts to restart sampling a er fi ve minutes in fl ow fault and continues to a empt a restart every fi ve minutes therea er until the restricted fl ow is corrected or the pump has a empted a maximum of 10 restarts. The maximum number of restart a empts may be changed using a PC and DataTrac 2000 So ware.
9
Page 13
Setup
Verifying Battery Charge Level
The LCD displays an icon that shows the current ba ery charge level (see Ba ery Operation on page 5). A new ba ery should be fully charged before operation.
Entering and Navigating the User Interface
The AirChek 2000 User Interface features two levels:
Level One—Set Up Flow Rate allows the user to change fl ow rate, adjust fl ow rate to a primary standard, or calibrate the pump using the CalChek feature.
Level Two—Set Up Sampling and Display permits the user to change display options of temperature (F or C) and atmospheric pressure (ins, m, or mm), set a sampling time, set 12 or 24-hour clock or delayed start, set real-time clock, or clear accumulated run time data.
Entering:
• With pump in HOLD, press [ST] to place pump in RUN and enter
ÒSTÒ within 10 seconds. You are in Level One of the User Interface.
• With pump in RUN, press [ST] to place pump in HOLD and enter
ÒSTÒ within 10 seconds. You are in Level Two of the User Interface.
Navigating: Press Ò to scroll through parameters. Once LCD shows End,
parameters will repeat until the user exits the User Interface.
Exiting: Press Ò until End appears on the LCD. Press [ST]. Any changes
made to parameters will be saved and the pump will continue in its current mode. Level Two of the User Interface off ers the option of exiting without saving changes to parameters. Press Ò until ESC appears on the LCD. Press [ST].
Using User Interface - Level One
Flow Set Set fl ow by pressing S to increase
or T to decrease pump fl ow rate. Press Ò to move to next parameter or the User Interface can be exited by continuing to press Ò until End appears. Press [ST] to save new ow se ing and exit. See page 12.
10
Page 14
ADJ Used during calibration with primary
standard calibrator (not for use with CalChek feature). Press S to increase
or T to decrease fl ow adjustment until desired fl ow is indicated on an a ached primary standard. Press Ò to move to next parameter or the User Interface can be exited by continuing to press Ò until End appears. Press [ST] to save new fl ow and adjustment se ings and exit the User Interface. See page 16.
CAL Use for CalChek calibration feature
only. Press [ST] to initiate single­point calibration. Press S seven times to start a full calibration. See CalChek Calibration instructions on pages 18 to 22.
Using User Interface - Level Two
Temperature Press S or T to toggle between
Fahrenheit (F) and Celsius (C). Press Ò to move to next parameter.
Setup
Atmospheric Pressure Press S or T to toggle through
inches of mercury (ins), millimeters of mercury (mm), and millibars (m). Press Ò to move to next parameter.
Sample Time (SГ) Program a speci c run time. Press
S to increase or T to decrease the time in minutes (from 1 to 999). Press Ò to move to the next parameter. Sample Time is not an option when a DataTrac 2000 program is loaded into pump memory (PROG). See page 13.
12-hr/24-hr Clock and
Press S or T to toggle through
Delayed Start standard (12-hour) time, military
(24-hour) time, and delayed start (dELA). See Se ing a Delayed Start on page 14. Press Ò to move to next parameter.
11
Page 15
Setup
Time of day The hour will fl ash. Press S to increase
and T to decrease hour. Press Ò to advance to minutes. Once minutes are ashing, press S to increase or T to decrease minutes. Press Ò to move to next parameter.
CLr Press [ST] to reset accumulated
run time and volume from pump memory to zero. See below.
ESC Press [ST] to exit User Interface
Level 2 without saving changes to parameters.
End Press [ST] to save changes to
parameters and exit User Interface.
OFF Appears only when a DataTrac 2000
program or delayed start is loaded in pump memory. When OFF displays, press [ST] to delete and exit the User Interface. See page 14.
Resetting Run Time Data
To reset accumulated volume and run time data to zero, enter User Interface Level Two by placing a running pump in HOLD and pressing ÒSTÒ within 10 seconds. Press Ò until CLr appears on the LCD. Press [ST] to clear the run time data. Press Ò until End appears, and then press [ST]. The pump is now in HOLD.
CLr does not clear previously set sampling time (SГ). See Deleting a Sampling Time on page 15.
Setting Pump Flow Rate
1. With the pump in HOLD, press [ST] to run the pump and enter ÒSTÒ within 10 seconds. You are now in User Interface Level One.
2. The fl ow rate on the LCD will fl ash. Press S to increase fl ow or T to decrease fl ow until the desired fl ow rate is displayed.
3. Press Ò repeatedly until End appears on the LCD.
4. Press [ST] to save the fl ow rate and exit the User Interface.
5. Press [ST] again to place the pump in HOLD if desired.
12
Page 16
Setup
Setting Sampling Time (S Г)
Program the AirChek 2000 from the keypad to run STEL, TWA, or any other run time from 1 to 999 minutes*. AirChek 2000 sampling time may also be programmed from a PC using DataTrac 2000 So ware*. To program a sampling time using the pump keypad:
1. With the pump running, press [ST] to place the pump in HOLD and enter ÒSTÒ within 10 seconds. You are in User Interface Level Two.
2. Repeatedly press Ò until appear on the display.
3. Set the sampling time by pressing S to increase it or T to decrease it to the desired time in minutes.
4. Press Ò repeatedly until End appears.
5. Press [ST] to save the sampling time and exit the User Interface. A fl ashing S will appear with the selected sampling time on the LCD.
6. Press [ST] to begin sampling. The time display will count down, the pump will go into HOLD, and the total sampling time will display when sampling is complete.
7. To delete a set sampling time, enter User Interface Level Two, scroll to until time appears as 00. Exit the User Interface by scrolling to End and pressing [ST].
SГ and a ashing 00
SГ, and press T
If a time still displays a er canceling a sample time, this display is the total run time since the data was last reset.
If a sampling time (SГ) has been programmed into the pump, a DataTrac 2000 program cannot be entered without deleting the sampling time fi rst (see page 15). Likewise, if a DataTrac 2000 program resides in pump memory, the sampling time (SГ) function cannot be selected until the DataTrac program is deleted (see page 15).
* Sampling Time (SΓ) Se ing: AirChek 2000 sampling time can be set up to 999 minutes from the pump keypad or up to 30 days (43,200 min­utes) from a PC using DataTrac 2000 So ware. During sampling, the pump LCD will display remaining run time (will count time down from a set number of minutes) when sampling time is set from the keypad. The pump LCD will show elapsed time in minutes when the pump is programmed from a PC. If run time is longer than 9999 minutes (6.8 days), the sample time displayed on the LCD will roll over to 1 a er reaching 9999.
13
Page 17
Setup
Setting a Delayed Start
The Delayed Start feature is available in pump version 2.59 or higher.
When se ing the pump manually for sampling to begin within a 12-hour timeframe, follow this procedure:
1. With the pump running, press [ST] to place the pump in HOLD and
enter ÒSTÒ within 10 seconds. You are in User Interface Level Two.
2. Press Ò until required to set a delayed start.
3. Press Ò until the display reaches the 12-hr/24-hr clock. Press S or T until the display shows a fl ashing dELA (delayed start).
4. Press Ò. Time will display with hours fl ashing. Press S to increase or T to decrease the hour. Press Ò to move to minutes. Press S to increase or T to decrease the minutes.
The time entered here will be the next occurrence of this time. There is no a.m. or p.m. designation.
5. Press Ò until End appears.
6. Press [ST].
7. The PROG icon and a fl ashing HOLD will appear in the upper le corner of the display. The pump is now set for delayed start.
SГ appears. Press S or T to set a sample time. This is
If se ing a delayed start using DataTrac 2000 So ware, refer to the DataTrac 2000 Operating Instructions (included on the so ware CD). A programmed delayed start can be viewed in the DataTrac 2000 Scheduler.
Deleting a Delayed Start
This is the same method used to delete a DataTrac 2000 program from the pump (see Deleting a DataTrac 2000 Program on page 15).
1. With the pump running (will only run in short bursts), press [ST] to place the pump in HOLD and enter ÒSTÒ within 10 seconds. You are
in User Interface Level Two.
2. Press Ò to scroll to the fl ashing OFF indicator and press [ST].
3. Press Ò until End displays.
4. Press [ST]. The PROG icon should be gone. If not, repeat process until it disappears.
14
Page 18
Setup
Deleting a Sampling Time (SГ)
To delete a sampling time, enter User Interface Level Two and press the Ò bu on to scroll to End appears. Press [ST].
Deleting a DataTrac 2000 Program
If a DataTrac 2000 program is in pump memory, the PROG icon will appear in the upper le corner of the display. To delete a DataTrac 2000 program, enter User Interface Level Two, scroll to the fl ashing OFF indicator, and press [ST]. This is the same method used to delete a delayed start from the pump. See Deleting a Delayed Start on page 14.
SГ. Press T until SГ 00 appears. Press the Ò bu on until
15
Page 19
Calibration
Flow Rate and Volume Display
Flow rate displayed on the pump LCD is the fl ow to which the pump has been calibrated. To maintain fl ow as displayed, the pump automatically adjusts fl ow during sampling for changes in temperature and atmospheric pressure that may diff er from the temperature and atmospheric pressure present at the time of calibration. The fl ow rate display does not change from the calibrated fl ow rate. The pump will fault if it is unable to maintain the calibrated fl ow rate.
Volume displayed on the pump LCD is ”corrected” in that it is the result of a continual calculation of corrected fl ow rate multiplied by sample time.
Verifying Flow Rate Using a Primary Standard
1. Connect the pump inlet to the outlet of the representative sampling medium using fl exible tubing.
To ensure sample integrity, tubing seals tightly onto the pump inlet lter cover. Use care when removing tubing to prevent cracking or breakage of the inlet fi lter cover.
2. Connect the inlet of the sampling media to a primary standard calibrator using another length of fl exible tubing.
3. From HOLD, press [ST] to run the pump and enter ÒSTÒ within 10 seconds. You are in User Interface Level One. SET fl ashes on the LCD.
4. The LCD will display a fl ashing fl ow rate from the last sample taken. If you do not wish to change the fl ow rate go to Step 6.
5. Press S to increase the fl ow rate or T to decrease the fl ow rate to the desired se ing.
6. Press Ò. The fl ashing ADJ appears on the LCD, replacing the fl ashing SET. The pump fl ow rate can now be calibrated using a primary standard calibrator.
The fl ow rate displayed on the calibrator will change as a result of this adjustment, not the fl ow rate displayed on the pump. The pump will display an adjustment factor (ADJ) only.
7. Compare the fl ow rate from the primary standard to that displayed by the pump. If the calibrator indicates a higher fl ow rate than the pump, press T until the pump and calibrator are in agreement (within 10 ml). If the calibrator indicates a lower fl ow rate than the pump, press S until the pump and calibrator are in agreement (within 10 ml).
16
Page 20
Calibration
8. Press Ò until End appears on the LCD. Press [ST] to save fl ow se ing and exit the User Interface. The pump will remain in RUN. Reset run time data (see page 12).
9. Place pump in HOLD by pressing [ST]. Disconnect the calibration train.
Pressing [ST] will place a running pump in HOLD, and a holding pump in RUN.
If a delayed start or DataTrac 2000 schedule has been programmed into the pump, it may remain in pump memory. PROG will display in the upper le corner of the pump display. To delete this program, enter User Interface Level Two, scroll to the fl ashing OFF, and press [ST].
17
Page 21
Calibration
Calibration Using the CalChek Automatic Calibration Feature
The CalChek automatic calibration feature is available when calibrating an AirChek 2000 Pump with a Defender Calibrator (Cat. No. 717-510M). A CalChek Communication Cable (Cat. No. 210-502) is required. Optional DataTrac 2000 So ware can be used to expand the documentation capabilities of this system. The CalChek feature provides two calibration options: single­point calibration allows se ing and verifying a fl ow rate at a single point before and a er sampling. Multiple-point (full) calibration corrects fl ow to a primary standard at multiple fl ow rates to cover the basic operational fl ow range of the pump. Both types of calibration bring fl ow to within 5%.
If using a pump version below 2.59, a pump upgrade is required for the CalChek feature. Contact SKC.
Do not perform single-point or multiple-point calibration until the pump has remained at ambient temperature for several hours.
Single-point Calibration Using CalChek
The CalChek feature provides correction at a single fl ow se ing and takes less than one minute to complete. Use it to set the desired fl ow rate before sampling and to verify fl ow a er sampling.
1. Use 1/4-inch tubing to connect the Defender Calibrator suction port to the inlet of a representative sampling medium and the outlet of the medium to the AirChek 2000 pump inlet.
2. Run the pump for 2 minutes before starting calibration. Leave the pump on.
3. Select the Defender data port:
a. Press and hold the Defender power bu on to turn on the calibrator. b. Press the right arrow to highlight Setup; press Enter. c. Press the right arrow to highlight Preferences; press Enter. d. Press the down arrow to navigate to Data Port. e. Press the le or right arrow to toggle to SKC. f. Press the down arrow to highlight Con rm; press Enter.
4. Enter Defender calibration mode:
a. Press the right arrow and then the down arrow to highlight Measure;
press Enter.
b. Press the right arrow to highlight Cont.; press Enter.
5. A ach the female end of the CalChek Communication Cable to the serial port (RS-232) on the back of the Defender Calibrator.
18
Page 22
Calibration
6. Insert the male end of the CalChek Communication Cable into the data port on the pump.
7. Place the pump in HOLD by pressing [ST]. Press [ST] again to start the pump running, and enter the security code ÒSTÒ in sequence on the pump keypad within 10 seconds. You are in User Interface Level
One.
8. Press S or T to set the pump to the desired fl ow rate.
9. Press Ò on the pump keypad until CAL appears on the pump display.
10. Press [ST] to select single-point calibration.
11. 1CAL will appear on the pump display. During calibration, the pump will display briefl y the fl ow rates that it is reading from the calibrator.
12. When calibration is complete, the pump display will either show End indicating a successful calibration, or it will show an error code of E4[x] (see CalChek Error Chart on page 22).
Note: To remove a CalChek error code from the LCD, press Ò.
13. Press [ST] twice to place the pump in HOLD. Disconnect the pump from the representative sampling medium and the calibrator.
To ensure sample integrity, tubing seals tightly onto the pump inlet lter cover. Use care when removing tubing to prevent cracking or breakage of the inlet fi lter cover.
14. Allow pump to go to SLEEP.
Data does not write to memory until pump has gone into SLEEP mode.
Successful single-point calibration will provide an entry in the pump history that can be viewed using DataTrac 2000 So ware (v. 3.59 or higher). If used with earlier DataTrac 2000 versions, the entry will appear as an undecipherable code; update DataTrac 2000 So ware. If an error occurs during calibration, the event will not be stored in history.
19
Page 23
Calibration
Full (Multiple-point) Calibration Using CalChek
This type of calibration provides fl ow correction across the basic operating range of the AirChek 2000 pump (750 to 3000 ml/min) in approximately 4 minutes. The operation calibrates each fl ow rate to a primary standard. It can also provide a record of calibration for maintenance and quality purposes if DataTrac 2000 So ware is used. SKC recommends that a full calibration be performed during pump maintenance and a er non-factory repairs.
Full calibration clears history, run time parameters, and the Scheduler in DataTrac 2000 Software.
Do NOT place sampling media in line for full calibration. Ensure the battery pack is completely charged before starting a full calibration.
Do not perform single-point or multiple-point calibration until the pump has remained at ambient temperature for several hours.
1. Use 1/4-inch tubing to connect the Defender Calibrator suction port to the AirChek 2000 pump inlet. Do NOT place sample medium in line.
2. Run the pump for 2 minutes before starting calibration. Leave the
pump on.
3. Follow Steps 3 through 9 from Single-point Calibration Using CalChek on pages 18 and 19.
4. Verify that the ba ery icon on the pump display shows at least two bars. If it does not, charge the ba ery before proceeding.
5. Press S on the pump keypad 7 times to place pump in full calibration mode.
6. The Defender Calibrator will begin to automatically calibrate the pump. Initial fl ow measurements are taken without fl ow from the pump and the pump fl ow rate is adjusted automatically. The pump will display FCAL, CS1, and a brief fl ow rate. The pump will continue to display CS2, then a fl ow rate, CS3, then a fl ow rate, etc. until calibration is completed at all ow rates between 750 and 3000 ml/min.
Full calibration begins at ows lower than 750 ml/min. At these lower ow rates, the calibrator piston will move slowly. This is normal; do
not interrupt calibration.
CalChek full calibration can be aborted by pressing [ST]. The pump will go to HOLD.
20
Page 24
Calibration
7. When calibration is completed, the pump will go to HOLD. If the calibration was successful, the pump LCD will revert to displaying pump run time as 0. If there was failure during the calibration process, an error code of E4[x] will appear (see CalChek Error Chart on page 22).
Note: To remove a CalChek error code from the LCD, press Ò.
8. Allow the pump to go to SLEEP mode to write calibration data to pump memory.
Data does not write to pump memory until pump has gone into SLEEP mode.
CalChek Full Calibration Data
Full calibration data can be viewed and printed by going to the DataTrac 2000 Pump Manager window and clicking on the View menu. Choose Calibration Info. This will display calibration results, pump serial number, and date of the last full calibration. A bu on allows this data to be printed. The printed report contains pump version, date printed, and a validation code.
CalChek Calibration Data Verifi cation in DataTrac 2000 Software
This feature requires pump version 2.62 or higher and DataTrac 2000 So ware version 3.62 or higher.
To ensure that printed data has not been tampered with, pull down the Tools menu and choose Confi rm Validation Code. Enter the data from the printed report, including the validation code. DataTrac 2000 will indicate whether the information is completely valid or if a parameter has been changed.
Clearing the history will not clear full calibration data. This data can only be cleared by performing another full calibration or by obtaining more than 36 pump history records.
When entering data to confi rm the validation number, enter the date in the following format: mmm dd, yyyy (e.g., Aug 18, 2008)
21
Page 25
Calibration
CalChek Error Chart
Single-point Calibration Errors
Error Problem Troubleshooting
E41 Correction required too large. A Perform a full calibration. If this fails mismatch of greater than 360 ml/min contact SKC Technical Support. between the fl ow setting on the pump and the reading generated by the Defender Calibrator has occurred.
E48 Could not get a successful single point Try the calibration again. If problem calibration within fi ve fl ow readings. persists, perform a full calibration.
Full (Multiple-point) Calibration Errors
Error Problem Troubleshooting
E44 First fl ow reading greater than 750 Check tubing between pump’s upper ml/min. The pump is fl owing faster pressure sensor and the top valve than it should, even though the assembly diaphragm to ensure it is not calibration routine delivered only a pinched or blocked, or contact SKC very small voltage to the pump. Technical Support.
E45 Pump unable to achieve fl ow rate of Check pump’s fl ow tube to ensure 3000 ml/min possibly due to a blocked it is not blocked, or contact SKC fl ow tube or an air leak inside the pump. Technical Support.
E46 Analysis error in the data (rare). Try full calibration again. If problem persists, contact SKC Technical Support.
E47 Less than two bars appear in the battery Recharge the battery. icon on the pump display indicating that the battery is too low. There must be at least two bars to begin a full calibration.
No Code At conclusion of full calibration, pump Pump not at ambient conditions for at does not verify to within 5%. least 2 hours. Retry calibration after pump has been at ambient conditions for 2 hours.
Pump not running for 2 minutes prior to calibration. Run pump for 2 minutes and retry calibration.
Errors That Can Occur During Both Calibration Modes
Error Problem Troubleshooting
E42 Unstable average. There is too Try the calibration again. If problem much variation in the fl ow readings. persists, contact SKC Technical Support.
E43 Serial time out. The calibrator is Check cable connection. If loose not communicating with the pump. or disconnected, connect properly.
22
Page 26
Calibration
Calibrating for Low Flow (5 to 500 ml/min) Sampling Applications Using a Constant Pressure Controller (CPC)
For use with CPC and Adjustable Low Flow Tube Holders.
The Constant Pressure Controller (CPC) is an accessory that allows low fl ow (5 to 500 ml/min) applications. In conjunction with an Adjustable Low Flow Tube Holder accessory, the CPC is used as a pressure regulator to maintain a constant 20 inches water back pressure. See Accessories on page 28.
Calibrate with Adjustable Low Flow Holder, CPC, and representative sample medium inline.
1. a. For single-tube applications, set the fl ow rate to 1500 ml/min. b. For multiple-tube applications, the pump fl ow rate must be set at
15% higher than the sum of the fl ow rates through all tubes.
Do not exceed 500 ml/min fl ow rate per tube for multiple-tube sampling.
To set the pump fl ow rate, from HOLD press [ST] to run the pump
and enter ÒSTÒ within 10 seconds. You are in User Interface Level One. SET and fl ow rate fl ash on the LCD.
2. The fl ow rate displayed is from the last sample taken. Press S to increase fl ow and T to decrease the fl ow to the desired se ing. Press Ò until End appears, and then press [ST] to save the change and exit the User Interface. The pump will remain running. If desired, press [ST] again to place the pump in HOLD.
3. Connect the pump inlet to the CPC outlet (the side of the CPC without a label) using a short length of Tygon tubing.
4. Connect the inlet side of the CPC (marked “to sample”) to the Adjustable Low Flow Holder.
5. Label tube(s) and port(s). Insert an opened representative tube into the rubber sleeve of each port on the Adjustable Low Flow Holder. Place
unopened tubes in any unused ports to “seal” them.
6. Loosen the brass fl ow adjust screw on the low fl ow holder. Connect the exposed end of one tube to a primary standard calibrator using another length of tubing. From HOLD, press [ST] to run the pump.
7. Turn the fl ow adjustment screw (needle valve) on the low fl ow holder until the calibrator indicates the desired fl ow rate.
The fl ow rate displayed on the calibrator will change as a result of this adjustment. The fl ow rate on the pump LCD will remain at the previously set fl ow rate, showing the fl ow rate through the pump, not through the sample tube.
8. Repeat Steps 6 and 7 to calibrate the ow rate through each tube. Seal unused ports during calibration with unopened tubes.
9. Place the pump in HOLD by pressing [ST]. Disconnect the calibration train and replace the representative tube(s) with the tube(s) to be used for sampling. Reset the run time data to zero (see page 12).
23
Page 27
Sampling
Sampling (1000 to 3250 ml/min Constant Flow)
1. Following calibration, replace representative sampling medium with a
new unexposed medium. Ensure run time data has been reset (see page 12).
Protect sample pump from weather when in use outdoors.
For multiple-tube sampling, seal unused holder ports with unopened tubes.
2. To begin sampling, press [ST] to run the pump. Record the start time.
3. Sample for the time specifi ed in the method used.
4. To stop sampling, press [ST] to place the pump in HOLD. Record the
stop time.
5. When sampling is complete, pump data is retained in memory for
recovery. Data can be viewed on the LCD by using the Ò bu on to scroll through it or on a PC using DataTrac 2000 So ware.
When using impingers, place an in-line trap between the pump and the impinger to protect the pump from harmful liquids or vapors. Failure to use the impinger trap voids the pump warranty.
If a delayed start or DataTrac 2000 schedule has been programmed into the pump, it may remain in pump memory. PROG will display in the upper le corner of the pump display. To delete this program, see page 14 or 15.
To ensure sample integrity, ensure tubing seals tightly onto the pump inlet fi lter cover. Use care when removing tubing to prevent cracking or breakage of the inlet fi lter cover.
Low Flow Sampling (5 to 500 ml/min Constant Pressure)
Low fl ow sampling requires a Constant Pressure Controller (CPC) and adjustable fl ow tube holder available as accessories or in the Low Flow Adapter Kit, Cat. No. 210-500.
When sampling with the CPC accessory and multiple tubes, the volume displayed on the pump is no longer representative of the volume of fl ow through the tubes due to the air bypass function of the CPC.
1. Following calibration (see page 23), replace the representative tube(s) with
a new unexposed tube(s).
Protect sample pump from weather when in use outdoors.
2. Press [ST] to RUN the pump. Sample for the time specifi ed in the
method used.
3. To stop sampling, press [ST] to place the pump in HOLD.
24
Page 28
Sampling
Long-duration Color Detector Tubes require a special tube cover that accommodates an in-line trap tube. The trap tube protects the pump from caustic fumes that are often released from detector tubes. Closely read all precautions when using these tubes. Failure to use the necessary traps will damage the pump and void the warranty.
To ensure sample integrity, tubing fi ts tightly onto the pump inlet lter cover. Use care when removing tubing to prevent cracking or breakage of the inlet fi lter cover.
Use of any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety.
Flow Fault
If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters fl ow fault mode. During fl ow fault, the pump HOLDs, the fault icon appears on the display during the length of the fault, and the accumulated run time display is frozen and retained. As part of the fl ow fault feature, the pump a empts to restart sampling a er fi ve minutes in fl ow fault and continues to a empt a restart every fi ve minutes therea er until the restricted fl ow is corrected or the pump has a empted a maximum of 10 restarts. The maximum number of restart a empts may be changed using DataTrac 2000 So ware.
25
Page 29
Maintenance
Replacing the Inlet Port Housing and Filter
1. Remove the three screws that secure the inlet port housing to the top of the pump.
2. Remove the inlet port housing and gasket.
3. Remove the O-ring.
4. Remove and discard the fi lter.
5. Insert a new fi lter (Replacement Cat. No. P40011).
6. Insert O-ring*. Replace the gasket.
7. Align the inlet port housing with the three screw holes and the LED.
8. Insert the three screws and tighten in an alternating fashion.
Do not overtighten screws. Overtightening can crack the inlet port.
* Replace with new O-ring only as needed.
26
Page 30
Maintenance
Service Policy
To return products to SKC for servicing:
1. Call 800-752-8472 (724-941-9701 for international customers) to obtain a Return Materials Authorization (RMA) number and Product Decontamination Form.
2. Carefully package the product. Mark the RMA number on any correspondence relating to the return and on the outside of the package.
3. Ship to SKC, freight prepaid, to the following address:
SKC Inc. National Service Center 863 Valley View Road Eighty Four, PA 15330
Package product carefully to prevent damage during transit. Include a contact name, phone number, shipping address, RMA number, and a brief description of the problem. For nonwarranty repairs, a purchase order number and billing address are also required. The Service Department will contact nonwarranty customers with an estimate before proceeding with repairs.
SKC Inc. will accept for repair any SKC product that is not contaminated with hazardous materials. Products determined to be contaminated will be returned unserviced.
Intrinsic safety and other approvals are void if SKC pumps are not repaired by SKC or authorized SKC repair centers. Use only SKC-approved parts to ensure reliable performance and to maintain the UL Listing for intrinsic safety and the SKC warranty.
27
Page 31
Accessories
Description Cat. No.
Defender Primary Standard Calibrator, 50 to 5000 ml/min, includes ba ery, charger (100-240 V), cable, and so ware 717-510M
CalChek Communication Cable, required for CalChek feature 210-502
Chargers
PowerFlex Charging System for SKC Personal Pumps 5-station, 100-240 V 223-1000 Single, 120 V 223-2000 Single, 100-240 V 223-2000B
PowerFlex Pump Cable for AirChek 2000 223-1001
Ba ery Eliminator, for continuous run using line power,
use voids the UL Listing for intrinsic safety
115 V 223-320 230 V 223-320B
Protective Pouches
Black 224-88 Red, for easy observation 224-96A Black, noise reducing 224-96C
DataTrac 2000 So ware Package
Includes so ware, adapter, and serial cable, requires Windows 98 or higher and serial port or appropriate adapter compatible with PC and operating system 877-91
Low Flow Adapter Kit (5 to 500 ml/min)
Includes Constant Pressure Controller (CPC), Adjustable Low Flow Tube Holder, and Type A Protective Cover 210-500
Constant Pressure Controller (CPC) for sampling in the 5 to 500 ml/min fl ow range. Use with Adjustable Low
Flow Holder (below). Available in Low Flow Adapter Kit above. 224-26-CPC
Adjustable Low Flow Tube Holders for Constant Pressure (Low Flow 5 to 500 ml/min) Applications
(requires separate tube cover listed below)
Single, available in Low Flow Adapter Kit above 224-26-01 Dual 224-26-02 Tri 224-26-03 Quad 224-26-04
Sample Tube Protective Covers
(for adjustable fl ow tube holders listed above) Type A (tubes 6-mm OD x 70-mm L) 224-29A Type B (tubes 8-mm OD x 110-mm L) 224-29B Type C (tubes 10-mm OD x 150-mm L) 224-29C Type T (tandem for color detector tubes up to 115-mm long and a trap tube) 224-29T
28
Page 32
Accessories
Description Cat. No.
Replacement Parts
Ba ery Pack P20136 Ba ery Pack (CE marked; for Cat. No. 210-2002Ex) P21113 Belt Clip P20139 Case P20137 Case with interface for PC board in place P20137A Charging Jack P20145 Cover, Ba ery Pack P20144 Filter (inlet)/O-ring (3) P20140 Filter Housing P20142 Filters, Inlet (50) P40011 Gasket Set P21273 Keypad P79361 Stack P20138 PC Board (includes top valve plate and stack inlet) P79519 PC Board Interface P79543 Cover, Port P20179 Keypad P79361 Screw/Gasket Kit P21002 Valve Assembly, Bo om P21272 Valve Assembly, Top P21322
Long-duration Detector Tube Accessories
Trap Tubes 222-3D-2 Tandem Protective Tube Cover 224-29T
29
Page 33
Accessories
DataTrac 2000 Software
With the optional DataTrac 2000 So ware accessory, the AirChek 2000 is programmable using a PC. DataTrac 2000 simplifi es chain-of-custody reporting by allowing users the option of programming a complete running sequence, delayed start, timed stop, and intermi ent sampling, all at diff erent fl ow rates. Time and sample volume are continuously updated in memory. There is no need to perform lengthy calculations; DataTrac does it for you. The advanced information retrieval system is specifi cally designed to store data and provide chain-of-custody information. Fault features allow storage of historical data in memory that can be retrieved days later as long as the ba ery pack is not completely discharged.
Features
• Program a sampling operation from a PC
• Calibrate the AirChek 2000 fl ow rate to a primary standard
• Display the AirChek 2000 operating state including fl ow rate,
temperature, run time, and ba ery status
• Create and save an AirChek 2000 program on a PC and upload to the
pump for operation in the fi eld.
• Program up to ten sampling sequences, each with diff erent fl ow rates
• Download the AirChek 2000 run time data and history to your PC
• Create chain of custody information using the sample set-up feature
• Print a history fi le containing AirChek 2000 run time data
• Print a worker exposure profi le containing run time data and the AirChek
2000 history
• Document date of pump calibration and validate information when using
the CalChek Automatic Calibration feature (DataTrac 2000 version 3.59 or higher)
DataTrac 2000 Requirements
• Any IBM-compatible PC with a 386 processor or higher
• Hard drive with a minimum of 8 MB free disk space
• CD drive
• 4 MB RAM (8 MB preferred)
• VGA display or be er
• Available 9 or 25-pin serial port (i.e., a COM port not used by a mouse,
modem, or other device) or appropriate adapter compatible with PC and operating system
• Mouse
• Microso
DataTrac 2000 Software
Includes software on CD, adapter, and serial cable. (Requires Microsoft Windows 98 or higher and serial port or appropriate adapter compatible with
PC and operating system) .......................................................Cat. No. 877-91
®
Windows® 98 or higher
30
Page 34
31
Page 35
SKC INC.
LIMITED ONE YEAR WARRANTY
1. SKC warrants that its instruments provided for industrial hygiene, environmental, gas analysis, and safety and health applications are free from defects in workmanship and materials under normal and proper use in accordance with operating instructions provided with said instru­ments. The term of this warranty begins on the date the instrument is delivered to the buyer and continues for a period of one (1) year.
This warranty does not cover claims due to abuse, misuse, neglect, alteration, accident, or use in application for which the instrument was neither designed nor approved by SKC Inc. This warranty does not cover the buyer’s failure to provide for normal maintenance, or improper selection or misapplication. This warranty shall further be void if changes or adjustments to the instrument are made by other than an employee of the seller, or if the operating instructions furnished at the time of installation are not complied with.
2. SKC Inc. hereby disclaims all warranties either expressed or implied, including any implied warranties of merchantability or fi tness for a particular purpose, and neither assumes nor authorizes any other person to assume for it any liability in connection with the sale of these instruments. No description of the goods being sold has been made a part of the basis of the bar­gain or has created or amounted to an express warranty that the goods will conform to any such description. Buyer shall not be entitled to recover from SKC Inc. any consequential damages, damages to property, damages for loss of use, loss of time, loss of profi ts, loss of income, or other incidental damages. Nor shall buyer be entitled to recover from SKC Inc. any consequential damages resulting from defect of the instrument including, but not limited to, any recovery under section 402A of the Restatement, Second of Torts.
3. This warranty extends only to the original purchaser of the warranted instrument during the term of the warranty. The buyer may be required to present proof of purchase in the form of a paid receipt for the instrument.
4. This warranty covers the instrument purchased and each of its component parts.
5. In the event of a defect, malfunction, or other failure of the instrument not caused by any misuse or damage to the instrument while in possession of the buyer, SKC Inc. will remedy the failure or defect without charge to the buyer. The remedy will consist of service or replace­ment of the instrument. SKC Inc. may elect refund of the purchase price if unable to provide replacement and repair is not commercially practicable.
6. (a) To obtain performance of any obligation under this warranty, the buyer shall return the instrument, freight prepaid, to SKC Inc., at the following address:
SKC Inc., National Service Center 863 Valley View Road Eighty Four, PA 15330 USA
(b) To obtain return authorization information or for further information on the war­ranty performance you may telephone 724-941-9701 at the above address. See Service Policy section in operating manual (if applicable).
7. This warranty shall be construed under the laws of the Commonwealth of Pennsylvania
which shall be deemed to be the situs of the contract for purchase of SKC Inc. instruments.
8. No other warranty is given by SKC Inc. in conjunction with this sale.
Form #3755 Rev 0207
32
Page 36
Index
AC Charger ........................................4, 28
Accessories ............................................ 28
Activate Pump ..........................................8
ADJ Indicator ....................................... 7, 11
Adjustable Low Flow Tube Holders
Arrow Buttons ........................................... 8
Back Pressure .................................... 2, 23
Battery
Charging ............................................4
Eliminator ...........................................6
Icons ..............................................5, 7
Installing .............................................4
Replacing ...........................................5
Verifying Charge Level ....................10
Bracketed Button Sequence ....................8
Button
Arrow ................................................. 8
Star ....................................................8
Sequence ..........................................8
CAL Indicator ...............................7, 11, 19
CalChek Calibration ...............................18
CalChek Full Calibration Data ................ 20
Calibration
with CalChek Feature ......................18
Constant Flow .................................. 16
Constant Pressure ...........................23
Charging Battery ......................................4
Clearing Data Display ........................7, 12
Compensation Range ..............................2
Constant Flow ........................................16
Constant Pressure .................................23
Continuous Operation ..............................6
CPC .................................................. 23, 24
DataTrac 2000 Software ........................30
Delayed Start .........................................14
Deleting a DataTrac 2000 Program .......15
Deleting a Delayed Start ........................14
Deleting Sampling Time .........................15
Display
Data ...................................................7
Resetting Run Time Data ................12
Temperature ....................................11
Flow Fault ....................................... 7, 9, 25
Flow Rate
Changing ......................................... 12
HOLD ...................................................7, 9
Icons ................................................. 5, 7, 9
Battery ...........................................5, 7
Flow Fault ................................7, 9, 25
Inlet Port Housing/Filter,
Replacement of ...................................26
Keypad Basics .........................................8
Star button ........................................8
Up & Down Arrow Buttons .................8
LCD ..........................................................7
Maintenance ........................................... 26
Multiple-tube Sampling ............... 23, 24, 28
....23, 24, 28
Operating Indicators ................................. 7
ADJ ..............................................7, 11
CAL ........................................7, 11, 19
CLr ...............................................7, 12
End ..............................................7, 12
ESC .............................................7, 12
FLOW .......................................... 7, 10
HOLD .............................................7, 9
PROG ....................................7, 14, 15
Sampling Time .......................7, 11, 13
SET ..............................................7, 10
VOL ...................................................7
Operating Modes ...................................... 9
Flow Fault ..........................................9
HOLD .................................................9
RUN ................................................... 9
SLEEP ...............................................9
Performance Profi le .............................2, 3
Pressure Units ........................................ 11
ins Hg ..............................................11
mm Hg .............................................11
PROG Indicator ............................ 7, 14, 15
Replacing Battery Pack ............................ 5
Resetting Run Time Data ....................... 12
RUN .........................................................9
Sampling
Constant Flow .................................. 24
Constant Pressure ...........................24
Constant Pressure Controller ..........24
Sorbent Tube (Single) ...............23, 24
Sorbent Tube (Multiple) in Constant Pressure Mode
......................................... 23, 24, 28
Sorbent Tube (Multiple) using
Constant Pressure Controller .....24
Sampling Time .............................7, 11, 13
Setting .............................................13
Setting Delayed Start .............................14
Security Code ........................................... 8
Service Policy ......................................... 27
SET Indicator .....................................7, 10
SLEEP ...................................................... 9
Sorbent Tube Sampling (Single) ............24
Sorbent Tube Sampling (Multiple)
............................................... 24, 28
Star Button ...............................................8
Up and Down Arrow Buttons .................... 8
User Interface ......................................... 10
Entering ........................................... 10
Navigating ........................................10
Level 1 .............................................10
Level 2 .............................................11
Warranty ................................................. 32
33
Loading...