SKC 690-105-NB User Manual

Page 1
Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
ULTRA III Passive Samplers without Blank,
with Charcoal Sorbent
Cat. No. 690-105-NB
Clip
Diffusion holes
(sample sorbent
compartment)
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Sliding cover in closed position
Sample sorbent
compartment
Sliding
cover
compartment
Clip
Plugs
Empty
Back view
Applications
SKC patented* ULTRA III Passive Samplers provide for low ppb to ppt detection of organic vapors in ambient and indoor air.
Performance Profi le
Sampling Rates: Go to the SKC Passive (Diff usive) Sampling Guide at www.skcinc.com/PassiveGuide/default.asp. Indoor (zero
Analysis: Solvent extraction; gas chromatography (GC)
Shelf-life: Six months from date of manufacture. Store at 39.2 F (4 C).
velocity) sampling rates are available at www.skcinc.com/ instructions/1811.pdf.
See Analysis Instructions at www.skcinc.com/instructions/40127.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com
Page 2
Sampling
1
23a3b
4
5
1. Open the pouch. Remove the sampler. a. Do not discard the pouch. b. Store the pouch in a clean, organic solvent-free
environment.
2. Record the following: a. Sample ID on the pouch and sampler
ID label
b. Sampling location on the pouch c. Date sample taken on the pouch
Do not use Sharpie® markers or equivalent to mark the label or pouch as these can off-gas VOCs.
3. Clip the sampler(s) in an appropriate location. a. Indoors: Clip the sampler(s) to a stand
accessory (Cat. No. 690-302) and locate at an appropriate height.
b. Outdoors: Clip the sampler(s) inside the
shelter (open on two ends and bo om to allow air to fl ow freely). Hang the shelter securely at the appropriate height.
Ensure the diffusion holes in the sampler are not obstructed in any way.
Sampler
I.D. label
4. Start sampling: Slide the sampler cover to the
open position (diff usion holes visible).
a. If performing for an extended sampling
time or sampling in an area of high
Diffusion holes (sample sorbent
compartment)
Clip
concentrations, install the Secondary Diff usion Barrier accessory (Cat. No. 690-
300); see Operating Instructions 39008.
Sliding cover
b. Write the sample start time on the pouch
label.
5. Stop sampling: Slide the sampler cover to the
closed position (arrow and SKC visible).
a. Remove the Secondary Diff usion Barrier, if
used, and then slide the cover to the closed
Sample compartment
Clip
position; see Operating Instructions 39008.
b. Write the sample stop time on the pouch label.
Sliding cover
Page 3
7
8
6. Place the sampler in its resealable pouch
6
immediately a er sampling.
Ensure the sample ID on the sampler matches the sample ID on the pouch.
7. Seal the pouch(es). Place the pouch(es) in a
container with a cool pack.
8. Use expedited shipping to ship the sample(s)
and Chain of Custody form(s) to a laboratory for analysis.
&
Storage
Before Sampling: Up to 6 months at 39.2 F (4 C) in a clean, organic solvent-
A er Sampling: 21 days at 39.2 F (4 C) in a clean, organic solvent-free
free environment
environment
Ordering
ULTRA III Passive Samplers, pk/5, prefi lled with: Cat. No. Charcoal, 500 mg 690-105-NB
Limited shelf-life; see Storage above
Sampling Accessories Stand for Indoor Sampling 690-302 Secondary Diffusion Barrier, 12 holes, lowers sampling rate for extended sampling time and for high concentrations 690-300 Shelter for Outdoor Sampling 690-303
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.asp.
Page 4
Form 40129 Issue 1306