SKC 690-105 User Manual

Analysis Instructions
Containing Charcoal Sorbent (Solvent Extraction)
Cat. No. 690-105, 690-105-NB, or 690-205
Analysis
1. a. For Cat. No. 690-105 or 690-105-NB (prefilled sampler):
Remove to Step 2.
b. For Cat. No. 690-205 (sorbent in vial): Remove
vial and proceed to Step 4.
2. Lay the sampler on a fl at surface with the back of the sampler facing upward.
3. With the clip oriented to the le , remove the plug from the compartment containing the sample sorbent (on left side, immediately above the diff usion holes).
the
sampler from
Plug
the
resealable pouch and proceed
the
lid from
Sample sorbent compartment
(plug removed)
the
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
Calculations
[(SW) - (BW)] (24.45 x 106) C =
(DE) (MW) (SR) (MIN) (PT)
Where: C = Concentration of chemical (ppb) SW = Sample weight by analysis (μg) BW = Analyte weight in blank (μg) PT = Pressure/temperature correction (see below) DE = Desorption effi ciency (see below) MW = Molecular weight of chemical SR = Sampling rate (ml/min) MIN = Sampling time (minutes)
The equation above is correct for 25 C (298 K) and standard atmospheric pressure (760 mm Hg). To convert to other temperatures and pressures the correction factor is:
863 Valley View Road, Eighty Four, PA 15330 USA
Blank/correction sorbent
compartment (plug in place)
Clip
Sliding cover
(690-105-NB)
4. Hold the sampler or vial over a 3.7-ml glass vial. Use a larger vial if using more than 2 ml of solvent.
5. Transfer the sorbent to a 3.7-ml or larger vial by tipping the sampler or vial upside down over the vial. Gently tap to remove any remaining sorbent.
6. Add typically 2 ml of desorption solvent to the sorbent in the vial.
Add the solvent slowly.
7. Shake the sorbent for 30 minutes.
8. Analyze the sample by gas chromatography with the detector specifi ed in the method for the compound of interest.
9. Analyze the blank/correction sorbent, as needed, by repeating Steps 2 through 8 for the sampler and Steps 1 and 4 through 8 for the sorbent vial.
PT = (T
1/T2
1.5
)
(P2/P1)
Where: T
= Sampling site temperature (in Kelvin)
1
T
= 298 K
2
P
= Sampling site pressure (in mm Hg)
1
P
= 760 mm Hg
2
Desorption effi ciency should be determined and expressed as a decimal (e.g., 98% = 0.98).
For sampling rates and desorption solvents, go to www.skcinc. com/PassiveGuide/default.asp.