ULTRA III Passive Samplers without Blank
Cat. Nos. 690-101-NB, -103-NB, and -104-NB
Clip
Diffusion holes
(sample sorbent
compartment)
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Sliding cover in
closed position
Sample sorbent
compartment
Sliding
cover
compartment
Clip
Plugs
Empty
Back view
Applications
SKC patented* ULTRA® III Passive Samplers provide for low ppb to ppt detection
of organic vapors in ambient and indoor air.
Performance Profi le
Sampling Rates: Go to the SKC Passive (Diff usive) Sampling Guide at
Analysis: Thermal desorption (TD); gas chromatography (GC)
Shelf-life: 30 days from date of manufacture. Store at ≤ 39.2 F (4 C).
www.skcinc.com/PassiveGuide/default.asp. Indoor (zero
velocity) sampling rates are available at www.skcinc.com/
instructions/1811.pdf.
See Analysis Instructions at www.skcinc.com/instructions/40125.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com

Sampling
1. Open the pouch. Remove the sampler.
a. Do not discard the pouch.
b. Store the pouch in a clean, organic solvent-free
environment.
2. Record the following:
a. Sample ID on the pouch and sampler
ID label
b. Sampling location on the pouch
c. Date sample taken on the pouch
Do not use Sharpie® markers or equivalent to
mark the label or pouch as these can off-gas
VOCs.
3. Clip the sampler(s) in an appropriate location.
a. Indoors: Clip the sampler(s) to a stand
accessory (Cat. No. 690-302) and locate at an
appropriate height.
b. Outdoors: Clip the sampler(s) inside the
shelter (open on two ends and bo om to allow
air to fl ow freely). Hang the shelter securely at
the appropriate height.
Ensure the diffusion holes in the sampler are not
obstructed in any way.
Sampler
I.D. label
4. Start sampling: Slide the sampler cover to the
open position (diff usion holes visible).
a. If performing for an extended sampling time
Diffusion holes
(sample sorbent
compartment)
Clip
or sampling in an area of high concentrations,
install the Secondary Diffusion Barrier
accessory (Cat. No. 690-300); see Operating
Sliding cover
Instructions 39008.
b. Write the sample start time on the pouch label.
Sample compartment
5. Stop sampling: Slide the sampler cover to the
closed position (arrow and SKC visible).
Clip
a. Remove the Secondary Diff usion Barrier, if
used, and then slide the cover to the closed
position; see Operating Instructions 39008.
Sliding cover
b. Write the sample stop time on the pouch label.
6. Place the sampler in its resealable pouch
immediately a er sampling.
Ensure the sample ID on the sampler matches
the sample ID on the pouch.

7. Seal the pouch(es). Place the pouch(es) in a
container with a cool pack.
&
8. Use expedited shipping to ship the sample(s)
and Chain of Custody form(s) to a laboratory for
analysis.
Storage
Before Sampling: Up to 30 days at ≤ 39.2 F (4 C) in a clean, organic solvent-free
environment
A er Sampling: 21 days at ≤ 39.2 F (4 C) in a clean, organic solvent-free
environment
Ordering
ULTRA III Passive Samplers, pk/5, prefi lled with: Cat. No.
Anasorb
Chromosorb
Tenax
# Limited shelf-life; see Storage above
‡ Comparable to Carbopack B
¥ Go to www.osha.gov and search on ULTRA for additional information on sampling rates for Chromosorb 106.
Sampling Accessories
Stand for Indoor Sampling 690-302
Secondary Diffusion Barrier, 12 holes, lowers sampling rate
for extended sampling time and for high concentrations 690-300
Shelter for Outdoor Sampling 690-303
Analysis Accessory
Thermal Desorption Tube, Perkin Elmer, 0.25-inch OD x
3.5-inch length, includes screens and end caps, ea P226530
®
GCB1,#‡ 370 mg 690-101-NB
®
®
106,#¥ 285 mg 690-103-NB
TA,# 253 mg 690-104-NB