SKC 690-101, 690-102, 690-103, 690-104 User Manual

Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
ULTRA III Passive Samplers
Cat. Nos. 690-101, -103, and -104
Clip
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Sliding cover in open
(sampling) position
Sliding
cover
Clip
Sample
compartment
Blank/
correction
compartment
Sliding cover in closed position
Sample sorbent
compartment
Sliding
cover
correction
compartment
Clip
Plugs
Blank/
sorbent
Back view
Applications
SKC patented* ULTRA® III Passive Samplers provide for low ppb to ppt detection of organic vapors in ambient and indoor air.
Performance Profi le
Sampling Rates: Go to the SKC Passive (Di usive) Sampling Guide at
Analysis: Thermal desorption; gas chromatography
Shelf-life: 30 days from date of manufacture. Store at 39.2 F (4 C).
www.skcinc.com/PassiveGuide/default.asp. Indoor (zero velocity) sampling rates are available at www.skcinc.com/ instructions/1811.pdf.
See Analysis Instructions at www.skcinc.com/instructions/40125.pdf.
* U.S. Patent No. 6,607,581
www.skcinc.com
Sampling
1
2
3a
3b
456
1. Open pouch. Remove sampler. a. Do not discard pouch. b. Store pouch in a clean, organic solvent-free
environment.
2. Record the following: a. Sample ID on pouch and sampler I.D. label b. Sampling location on pouch c. Date sample taken on pouch
Do not use Sharpie® markers or equivalent to mark label or pouch as these can off-gas VOCs.
3. Clip sampler(s) in appropriate location. a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and bo om to allow air to fl ow freely). Hang shelter securely at the appropriate height.
Ensure diffusion holes in sampler are not obstructed in any way.
Sampler
I.D. label
4. Start sampling: Slide sampler cover to open
position (diff usion holes visible).
a. If performing for an extended sampling
time or sampling in an area of high concentrations, install Secondary Diff usion Barrier accessory (Cat. No. 690-300); see Operating Instructions 39008.
Diffusion holes
(sample sorbent
compartment)
Blank/
correction
sorbent
compartment
Clip
Sliding cover
b. Write sample start time on pouch label.
5. Stop sampling: Slide sampler cover to closed
Sample compartment
position (arrow and SKC visible).
a. Remove Secondary Diff usion Barrier, if used,
and then slide cover to closed position; see Operating Instructions 39008.
b. Write sample stop time on pouch label.
Blank/
correction
compartment
Clip
Sliding cover
6. Place sampler in its resealable pouch
immediately a er sampling.
Ensure sample ID on sampler matches sample ID on pouch.