SKC 590-300 User Manual

Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
Nitrous Oxide Passive Sampler
Cat. No. 590-300
Introduction
The SKC Nitrous Oxide Passive Sampler is a small sampler that provides accurate and reliable sampling of nitrous oxide in the 50 to 450 ppm range (12.75 to 890 ppm­hours dose range). The samplers are packaged as empty sampler housings that the user fi lls with cleaned/purged Molecular Sieve 5A sorbent supplied in glass vials with PTFE-lined caps. This method of packaging maintains low sorbent background. Samples are thermally desorbed and analyzed by gas chromatography with electron capture detector.
Collar Clip
Collar
Clip
Plug
Cover
Green Secondary Diffusion Barrier
Primary
Diffusion
Barrier
Vial
Adapter
Front View of Housing
Back View of Housing
Vial containing
purged sorbent
Performance Profi le
Sorbent/Amount: Housing Material: Nylon, non-sterile Diameter: 1.4 in (3.5 cm) Length (including clip): 2.5 in (6.3 cm) Depth: 0.6 in (1.5 cm) Concentration Range: 50 to 450 ppm (12.75 to 890 ppm-hrs dose) Analysis: Thermal desorption; gas chromatography (GC) with
Shelf-life: 6 months from date of manufacture. Store at < 39.2 F (4 C). Sample Storage: 3 weeks at < 39.2 F (4 C) or 1 week at ambient temperature Sample Time: Validated for 8-hour sampling Sampling Rate: 0.81 ml/min ± 12% RSD
Samplers are non-sterile.
Molecular Sieve 5A 60/80, 736 mg (supplied in glass vial)
electron capture detector (ECD)
www.skcinc.com
Sampling
2
2
3A
Open aluminized resealable pouch and
remove sampler, vial of sorbent, and green secondary diff usion barrier.
Use proper handling technique when working with samplers.
Write pertinent information, including
Sample ID#, on the lines provided on the pouch label.
In a clean, organic solvent-free environment,
remove plug from vial adapter on back of sampler housing (side without holes). Remove cap from sorbent vial. Immediately hold housing vial adapter down over the open end of the vial. Screw open end of vial into the adapter (see inset 3A). Rotate assembly until sorbent is draining from vial to sampler. Tap vial lightly before unscrewing and removing it from the adapter. Immediately replace plug on adapter.
Sorbent
Avoid exposing sorbent to air.
Sorbent Vial
Vial Adapter
Remove cover from face of sampler housing
(side with holes) immediately before sampling. Store cover for reuse.
Important!
Snap on green secondary diff usion barrier to
control sampling rate.
Clip sampler to a worker's clothing near the
breathing zone. The sampler must not be covered by clothing and the diff usion holes must face outward.
Samplers are non-sterile.
Do not cover sampler with clothing.
A er desired sampling period, unclip sampler
1111
10
1010
1111
from worker's clothing. Immediately remove green secondary diff usion barrier and dispose following appropriate procedures.
Seal sampler with cover immediately.
It is important that samplers are sealed tightly with cover. The samplers could potentially pick up moisture over time which will affect the recovery of N
O.
2
Write stop time on label on the pouch.
In a clean, organic solvent-free environment,
10
carefully remove plug from vial adapter on back of sampler housing. Immediately screw open end of sorbent vial into adapter (see inset 10A).
Vial Adapter
Rotate assembly until sorbent is draining from sampler to vial. Tap sampler lightly. Ensure all sorbent is recovered in the vial. Remove sampler housing from vial and immediately screw cap onto vial.
Avoid exposing sorbent to air.
Sorbent Vial
Mark sample vial(s) as appropriate and seal in
aluminized pouch. Package sample(s) and one blank sorbent vial in a padded envelope. Send
5:30
9:30
to a laboratory for analysis.
Storage
Before Sampling: Up to 6 months at < 39.2 F (4 C) in a clean environment
A er Sampling: 3 weeks at < 39.2 F (4 C) or 1 week at ambient temperature
in a clean environment
10A
Sorbent
Sorbent vials should be capped tightly and kept in foil pouches before and a er use.
Analysis
1. Remove cap from vial. Using the transfer funnel accessory, immediately pour sorbent from vial into a 0.25-inch OD (0.194-inch ID) x 3.5-inch L thermal desorption tube. Ensure end of tube facing downward contains a stainless steel screen and glass wool separator to hold sorbent in place.
2. A er sorbent is transferred to the tube, remove funnel and immediately place a stainless steel screen and glass wool separator into the open end of the tube.
3. Seal tube with PTFE caps or Swagelok
®
 ings.
4. Thermally desorb the sample.
5. Analyze samples using a gas chromatograph (GC) with electron capture detector (ECD).
* Thermal desorption tubes must be 0.25-inch O.D., 0.194-inch I.D., and a minimum length
of 3.5 inch (Perkin Elmer Thermal Desorber). Thermal desorption tube blanks that meet these specifi cations are available from SKC (Cat. No. P226350).
Validation
The SKC Nitrous Oxide Passive Sampler has been validated over a concentration range of 50 to 450 ppm (20 to 80% RH) and at time intervals from 15 minutes to 8 hours. This is described in detail in the validation report
1
. The average sampling rate over the dose range of 12.75 to 890 ppm-hours was 0.81 ml/min with a standard deviation of ± 12%. The sorbent was transferred to a Perkin Elmer thermal desorption tube and then thermally desorbed into a GC with electron capture detection.
References
1
Kuhlman, C. and Coyne, L., Validation of Nitrous Oxide Using SKC Passive Sampler 590-300,
SKC Inc., Mar. 2008, www.skcinc.com/pdf/1762.pdf
Ordering
Description Cat. No. Nitrous Oxide Passive Samplers include empty sampler housings*,
secondary diffusion barriers, and sealed glass vials of 736 mg cleaned/ purged Molecular Sieve 5A sorbent, pk/5; limited shelf-life 590-300 Replacement Sorbent Vial, glass vial with PTFE-lined cap fi lled with 736 mg cleaned/purged Molecular Sieve 5A sorbent 590-268
* Samplers are non-sterile.
Analysis Accessories Cat. No. Transfer Funnel, aluminum, facilitates transfer of sorbent from vial to
0.25-inch OD thermal desorption tube 590-264 Thermal Desorption Tube, Perkin Elmer, 0.25 x 3.5 inch, includes screens and end caps P226530