SKC 303 User Manual

Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 www.skcinc.com
Soap Film Flowmeter
Cat. No. 303
SKC primary standard fl owmeters are designed, manufactured, and tested for accuracy; no calibration is needed. Read the operating instructions thoroughly before operating the fl owmeter.
Principle of Operation
The air sampling pump to be calibrated is connected to and pulls air through the fl owme- ter’s volumetric glass tube where a fl at soap bubble (fi lm) is interposed into the fl ow path. As the airfl ow causes the fi lm to move up the volume marks, travel time is measured using a stopwatch. Flow rate can be calculated using the travel time and known tube volume. The 303 Portable Flowmeter Kit is calibrated to a primary standard ± 2% of the volumes marked on the fl owmeter.
Assembly
SKC Model 303 is shipped ready to use. Film solution is supplied as well as a length of rubber tubing to connect the fl owmeter to the pump.
Preparing for Operation
1. Remove the fl owmeter from the case, and place it upright on a level surface.
2. Remove the large rubber stopper from the top of the fl owmeter.
3. Tip the fl owmeter at an angle, and pour a small amount of fi lm solution into the top of the glass tube. The liquid should run down into the rubber bulb at the bo om.
4. Continue adding fi lm solution until it completely fi lls the bulb and rises to just below the side arm inlet. Squeeze the rubber bulb occasionally to release trapped air.
5. Replace the stopper on top of the fl owmeter.
Wetting the Walls of the Tube
The interior walls of the fl owmeter must be wet for the fi lm to travel without rupture. There are two methods:
Method 1
1. Tip the fl owmeter to a horizontal position. Allow the fi lm solution to run out of the rub- ber bulb and into the tube.
2. Maintaining the horizontal position, rotate the fl owmeter so that fi lm solution coats the entire inside surface of the glass. Caution: Do not allow the solution to enter the metal tube in the rubber stopper at the top of the fl owmeter.
3. Return the fl owmeter to an upright position and allow the liquid to fl ow back down into the rubber bulb. Squeeze the bulb occasionally to release trapped air.
Method 2
1. Connect a pump to the fl owmeter (see Connecting a Pump).
2. Turn on the pump. Each successive fi lm that rises up the tube will wet the glass. The walls are suffi ciently wet when the fi lm bubbles successfully reach the highest volume mark on the fl owmeter.
Connecting the Pump
Note: If the pump to be calibrated draws air, connect it to the rubber tubing on the fl owmeter’s top rubber stopper. If the pump blows air, use rubber tubing to connect it to the lower side arm near the base of the fl owmeter.
1. Remove the rubber cap from the side arm inlet near the base of the fl owmeter.
2. Using rubber tubing, connect one end to the appropriate port on the fl owmeter and the other end to the inlet of the sampling medium to be used (if applicable).
3. Using fl exible tubing, connect the outlet of the sampling medium to the inlet of the pump.
Sample media outlet
Rubber tubing
Calibration chamber
Sample pump
Inlet
Tygon tubing
Flowmeter
Pump inlet
Outlet (upper port)
Inlet (lower side arm port)
Measuring the Flow
1. Turn on the pump.
2. Squeeze the rubber bulb on the fl owmeter to force the liquid above the inlet. Bubbles (fi lm) will form.
3. Using a stopwatch, start timing the fi lm as it passes the zero line and stop timing as the bubble reaches the 100 ml line. For best results, form several bubbles, fi ve to six seconds apart, and time the last bubble.
Caution: Do not squeeze the bulb continuously. This causes froth to form on the walls mak­ing timing diffi cult.
The fl ow rate can be determined from the Flow Chart on the back page, or calculated using the following equation:
Flow (ml/min) = 60 x (volume traveled) Time (in seconds)
Storage or Transport
Disconnect the pump and replace the rubber cap on the lower inlet of the fl owmeter. The instrument can be placed in any position without solution loss. SKC recommends that it be transported in its case.
Volumetric Calibration of Hand Pumps
SKC Model 303 Flowmeter can be used to determine the volume of manual hand pumps such as the bellows or piston pumps used with detector tubes. Follow these procedures:
Bellows Pumps (i.e., Dräger®)
1. Connect a detector tube to the pump according to the manufacturer’s instructions.
2. Connect the inlet of the tube to the rubber hose on top of the fl owmeter.
3. Compress the bellows pump and release slowly. Squeeze the rubber bulb on the fl ow- meter. Bubbles (fi lm) will begin to form.
4. When the fi lm reaches the zero line on the fl owmeter, compress the bellows pump com- pletely and release. The fi lm will move up the fl owmeter and stop when the bellows pump is fi lled to capacity.
5. Record the point at which the fi lm stops. Most bellows pumps are designed to draw 100 ml. Lines on the fl owmeter are graduated at 90, 95, 100, 105, and 110 ml. Visually extrapolate the volume reading if the fi lm stops between lines.
6. Calculate and record the volume.
Note: Pumps should be repaired if they fall outside of the 90 to 110 ml range.
Piston Pumps (i.e., Gastec®, MSA, Kitagawa)
1. Connect a standard detector tube to the piston pump according to the manufacturer’s instructions.
2. Connect the inlet of the tube to the rubber hose on top of the fi lm fl owmeter.
3. While pulling the piston slowly back, squeeze the rubber bulb on the fl owmeter until bubbles (fi lm) form.
4. Watch the fi lm movement closely. When the fi lm is exactly at the zero line on the fl ow- meter, push the pump piston all the way in. Then, draw the piston out to full capacity and lock it in place. The fi lm will move up the fl owmeter and stop when the piston pump is fi lled to capacity.
5. Record the point at which the fi lm stops. Most piston pumps are designed to draw 100 ml. Lines on the fl owmeter are graduated at 90, 95, 100, 105, and 110 ml. Visually extrapolate the volume reading if the fi lm stops between lines.
6. Calculate and record the volume.
Note: Pumps should be repaired if they fall outside the 90 to 110 ml range.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.asp.
Flow Chart for 303 Calibrator
100 ml line 100 ml line 100 ml line ` 100 ml line ` Time Flow Time Flow Time Flow Time Flow (sec) (ml/min) (sec) (ml/min) (sec) (ml/min) (sec) (ml/min)
12 500 44 137 92 65 85 35 13 462 45 133 94 64 87 34 14 429 46 130 96 62 90 33 15 400 47 128 98 61 93 32 16 375 48 125 100 60 96 31 17 353 49 122 102 59 100 30 18 333 50 120 104 58 104 29 19 316 51 118 106 57 108 28 20 300 52 115 108 56 112 27 21 286 53 113 110 55 116 26 22 273 55 111 112 54 120 25 23 261 55 109 114 53 125 24 24 250 56 107 116 52 130 23 25 240 57 105 118 51 135 22 26 231 58 103 120 50 140 21 27 222 59 101 50 ml line 150 20 28 214 60 100 Time Flow 160 19 29 207 62 97 (sec) (ml/min) 170 18 30 200 64 94 60 50 180 17 31 193 66 91 61 49 190 16 32 187 68 88 62 49 200 15 33 182 70 86 63 47 210 14 34 176 72 83 65 46 225 13 35 171 74 81 66 45 250 12 36 167 76 79 68 44 275 11 37 162 78 77 70 43 300 10 38 158 80 75 72 42 330 9 39 154 82 73 74 41 365 8 40 150 84 71 76 39 425 7 41 146 86 70 78 38 500 6 42 143 88 68 81 37 600 5
43 139 90 67 83 36
Accessories and Spare Parts
Description Cat. No.
Film Solution, 1 pint, (473 ml) 302-4011 Precision Digital Stopwatch 303-01-1 Replacement Connecting Hose (2 pieces) P3032101 Replacement Squeeze Bulb P3032311 Replacement Flowmeter Sidearm Cap P3032312
www.skcinc.com
Form 3701 Rev 1210
Loading...