SKC 233 Series, 240 Series, 243 Series, 253 Series, 263 Series Operating Instructions Manual

Page 1
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
Single Stainless Steel or PTFE Fitted Bags
233, 240, 243, 253, and 263 Series Sample Bags
These sample bags feature a single stainless steel or PTFE fitting. The hose connection, shut-off valve, and syringe port with septum are all contained within the single-port  ing.
Guidelines for Bag Sampling
1. Ensure that the bag material and  ings are appropriate for the compounds to be sampled.
2. Use only PTFE tubing to connect the bag to the pump to prevent sample loss by adsorption on tubing walls.
3. Before using, fl ush the bag thoroughly with purifi ed air or nitrogen.
4. Analyze the sample within 24 to 48 hours after sampling. Long-term storage of air-contaminant mixtures in bags is not recommended.
5. Do not ship sample bags by air unless the cargo cabin is pressurized.
6. Infl ate bag properly (see Proper Bag In ation for details).
®
7. Due to the physical properties of FlexFoil creases may appear on the bags during SKC manufacturing, quality testing, and user handling. These marks only aff ect the aesthetics of the bag and not its performance. FlexFoil bags exhibiting minor crimps and creases are not subject to replacement or refund as functionality is not aff ected.
bag material, minor crimps and
Notes and Cautions
• SKC sample bags are designed for sampling air at atmospheric pressure only. A empting to pressurize the bag will result in bag rupture and sample loss. Do not ship bag samples by air freight in a non-pressurized cargo cabin. Bags can burst under such conditions.
• All federal and state packaging and transporting regulations apply.
• Do not use sample bags to sample compounds with boiling points > 249.8 F (121 C).
• SKC bag  ing valves are extremely durable but are not intended for use as handles or hanging devices. This type of handling may damage the seal, causing leakage. This is considered to be misuse and will void the SKC warranty.
• Failure to follow warnings and cautions voids any warranty.
• Sample bags are designed for single use only.
Page 2
12:42 03/15/2017
0.0 inH 2o
00:01:42
250
ml min
Operation
1. Flush the bag at least 3 times with purifi ed air or nitrogen before use.
2. To fi ll a bag, connect PTFE tubing from the exhaust port of an air sample pump to the stem on the bag  ing.
3. To open the valve, turn the knurled disc on the stem counterclockwise a maximum of 2 revolutions. Turn on the pump and sample as usual (see Proper Bag In ation).
Do not use sample bags to sample compounds with boiling points > 249.8 F (121 C).
4. When sampling is complete, turn off the pump and close the valve by turning the knurled disc clockwise until snug.
The outside diameter of the syringe needle should not be larger than 12 gauge (0.109 inch).
Do not allow the syringe needle to puncture the bag material when piercing the septum.
5. To withdraw samples using a needle and syringe, carefully insert the needle into the stem and pierce the septum centered in the base of the syringe port.
Proper Bag Infl ation
Avoid fi lling any bag to more than 80% of its maximum volume. At the end of the sampling period, turn off the pump and close the valve by turning the knurled disc clockwise until snug.
Incorrect
Infl ation
Correct
Infl ation
Page 3
SKC sample bags are available in Tedlar, Standard FlexFoil®, and FlexFoil PLUS with ttings made of stainless steel, polypropylene, or PTFE in sizes ranging from 0.5 to 100 liters. SKC also manufactures custom sample bags. Contact SKC at 724-941-9701 or skcorder@skcinc.com.
For sample bag stability reports,
visit www.skcinc.com/bags.asp.
Accessories
Replacement Septa, pk/10........................................................................................ Cat. No. 233-01-RS
PTFE Tubing, ts over bag tting and pump exhaust,
3/16-inch ID, 1/4-inch OD, 10 feet ................................................................................. Cat. No. 231-9-23
PTFE Tubing, ts inside bag tting, 1/16-inch ID, 1/8-inch OD, 10 feet .......................Cat. No. 231-9-21
PTFE Tubing, ts Vac-U-Chamber sample inlet port,
1/4-inch ID, 5/16-inch OD ......................................................................... 10 feet ......... Cat. No. 231-937
50 feet ......... Cat. No. 231-924
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the com­plete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/ warranty.asp.
Page 4
www.skcinc.com
Form 37147 Rev 1803