Sizewise Rotate, Big Turn Quick Reference Manual

Rotate™ and Big Turn
Quick Reference Guide
Operating Instructions
1. Locate both power switches on left side of control unit.
2. Turn both switches to ON position.
Mattress Ination/Autorm
Press AUTOFIRM button to fully inate surface. Transfer patient to surface. AUTOFIRM times out after 10 minutes and resets to previous setting. To manually exit, press
AUTOFIRM a second time.
Rotation Therapy
1. Choose desired turn by pressing corresponding button:
• Right or Left Turn: maintains a turning cycle only to the selected side
• Full Turn: provides left, center, and right rotational therapy (CLRT)
2. After selecting desired turn, press CYCLE TIME -/+ buttons until desired time is reached. If using left or right turn as a holding cycle, press RIGHT or LEFT button until Hd shows on screen. Patient will be held on desired side.
3. After selecting therapy, pressing TRAINING MODE will slowly acclimate patient to therapy over a period of 60 minutes.
Pulsation Therapy
1. Press PULSATE button until desired time is reached (3, 5, or 10 min).
2. Set rmness using FIRM/SOFT buttons based on patient comfort and clinical discretion.
Static Therapy
1. Press STATIC button.
2. Set rmness using FIRM/SOFT buttons based on patient comfort and clinical discretion.
See reverse side for additional instructions
Other Features
Fowler: Increases airow when head section of bed is
elevated. Newer models include an Auto Fowler feature.
FOWLER
Firm/Soft: Increases (FIRM) or decreases airow (SOFT)
creating a rmer or softer surface.
Lockout: Locks control panel (auto after 3 min).
To unlock, press and hold for 3 sec.
Alarm Reset: Light remains illuminated until problem is
corrected.
CPR/Quick Deflation
1. CPR control is on front of control unit. Depending on model, push red button or turn dial clockwise to activate CPR. Alarm will sound.
2. Locate red CPR valve at head of mattress. Turn valve for quick air release.
-OR- Locate hose connection to control unit (all models). Disconnect for quick air release.
Patient Transport
1. Activate Autorm mode for 3-5 minutes to ensure mattress is completely inated.
2. Turn both power switches o (left side of control unit).
Unplug from wall.
3. Mattress will slowly deate. As patient sinks into mattress, patient will rest on foam transport pad.
This card is not meant to replace product user manual. For complete manual, contact Sizewise.