Simrad AS shall not be liable for errors contained herein, or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
document.
No part of this work covered by the copyright hereon may be reproduced or otherwise
copied without prior permission from Simrad AS.
Simrad AS
Strandpromenaden 50
Box 111
N-3191 Horten
M A X I M I Z I N G Y O U R P E R F O R M A N C E A T S E A
Sections
1Application d escription
This chapter provides a very short description of the ER60 Scientific echo
sounder. Refer to page 1.
2Displayviews
ThevariousviewspresentedbytheER60Scientificechosounderareexplained.
Refer to page 2.
3Operational procedures
This chapter holds a selection of common procedures covering the key
functions of the ER60. Refer t o page 15.
4Menus
The ER60 is operated using a set of menus, short-cut menus and dialogue
boxes. This chapter lists the menus with page references to the applicable
dialogue boxes and/or functions. Refer to page 41.
Operator manual
5Dialogue boxes and functions
This chapter holds reference information describing the dialogue boxes and
functionsindetail.Thetopicsarelistedin alphabetical order. Refertopage46.
The ER60 display is organised as follows (from top):
• Main menu
• Tool bar
• A number of Channel windows
• Status bar
A display with two channels windows is shown as an example.
850-164692 / Rev.B
3
Simrad ER60 Operator manual
Main menu
The ER60 Main menu is located at the top of the echo sounder
window.
To operate, click on the menu name and observe the drop-down
menu. Select a new item on the drop-down menu by clicking on
the command.
Commands shown with light grey colour are unavailable in the
present configuration or operational mode.
A detailed description of the commands available from the main
menu is given in the Reference guide.
Related topics
→Main menu commands, page 41
4
850-164692 / Rev.B
Display views
Toolbar
The Toolbar is located directly underneath, or next to the main
menu. A number of dedicated buttons allow for easy access to
the most frequently used functions.
Two toolbars may be applied, Normal operation and Easyoperation. The toolbar selection is made by the Toolbar
command on the View menu. Note that Normal operation is
not available during replay.
The functions (from left) and their equivalent dialogue buttons
and controls are listed in the table below.
Easy operationEquivalentDialogue
Start pingingStartPing control
Perform single pingSingle stepPing control
Stop pingingStopPing control
Set ping ratePing ratePing control
Start raw data recordingSave raw dataFile output
Stop raw data recordingSave raw dataFile output
Initiate annotationAnnotation text inputAnnotation
Generate eventAnnotation
XX
Normal operationEquivalentDialogue
Select channelXXXX
Select modeModeNormal Operation
Related topics
→File Output, page 80
→Normal Operation, page 102
850-164692 / Rev.B
→Ping Control, page 105
→Toolbar, page 121
5
Simrad ER60 Operator manual
Status bar
The Status bar is located at the bottom of the display. It
presents messages from the echo sounder system as well as a
warning field at the right hand side.
Enable
The Status bar can be enabled or disabled with the Status bar
choice on the View menu, and configured by clicking the right
mouse button while the cursor is located within the Status bar
borders.
Information
(CD6974G)
When all information has been enabled, the Status bar will
display (from left):
• Current messages
• Current Pitch in degrees (P: 0.0°)
• Current Roll in degrees (R: 0.0°)
• Current Heave in meters (0.00 m)
• Event number
• Current Speed in knots (7.0 kts)
• Distance in nautical miles (1325.439 nmi)
• Geographical position (68°26’515 N...)
• Recording status (RECORD OFF)
• Current time (16:06:36)
Warnings and errors
The Warnings field on the far right side of the Status bar is a
button. When a warning has been issued, the button will be lit
with a colour and a text indicating the nature of the message. To
investigate further, click the button.
Related topics
→Configure Statusbar, page 60
→Warnings, page 131
6
850-164692 / Rev.B
Display views
Channel windows overview
The ER60 Channel window is the main information bearer on
the echo sounder display.
The Depth view comprises the horizontal view directly below
the Toolbar. The view displays the current depth.
If you click the right mouse button with the cursor position
within this view, a dedicated shortcut menu will appear. If you
press the middle mouse button, the Bottom Detection dialogue
box will open.
The Single Target Position view is normally located in the top
left corner.
This view shows the position within the beam
of the detected single echoes for the current
ping (largest circles) and the three previous
pings (smaller circles), all in the active layer.
The colours used to display this information
are determined from current target strength
(TS) value and the settings in the Colourscale.
If you place the cursor inside the view, a small
yellow label will appear to give you a detailed
readout of certain information. The Tooltip
dialogue defines which information that will
be shown.
(CD6974B)
If you click the right mouse button with the
cursor position within this view, a dedicated
shortcut menu will appear. If you press the
middle mouse button, the Single TargetDetection dialogue box will open.
The Single Target Histogram view is normally located in the
bottom left corner of the display presentation.
The histogram plot shows a visualisation of the
target strength (TS) distribution for the single
echoes detected in the Echogram view The colours
used to display this information are determined from
the current target strength (TS) value and the
settings in the Colour scale. The statistics are based
on the active layer and the current calculation
interval.
If you place the cursor inside the view, a small
yellow label will appear to give you a detailed
readout of certain information. The Tooltip dialogue
defines which information that will be shown.
(CD6974C)
If you click the right mouse button with the cursor
position within this view, a dedicated shortcut menu
will appear. If you press the middle mouse button,
the Single Target Detection dialogue box will open.
The Echogram view will normally dominate the echo sounder
display.
The Echogram contains information
about the acoustical values. The settings
in the Colour Scale is used to present
the information.
If you place the cursor inside the view, a
small yellow label will appear to give
you a detailed readout of certain
information. The Tooltip dialogue
defines which information that will be
shown.
If you click the right mouse button with
the cursor position within this view, a
dedicated shortcut menu will appear. If
you press the middle mouse button, the
Echogram dialogue box will open.
The Scope view is normally located on the right side of the
Echogram view.
The Scope view provides a logarithmic oscilloscope
visualization of the last ping. The presentation
corresponds to the current settings in the Echogram view.
The Scope view draws a range of horizontal symmetrical
colour lines. The distance from the vertical center axis and
the line colour reflects the received echo scatter.
If you place the cursor inside the view, a small yellow
label will appear to give you a detailed readout of certain
information. The Tooltip dialogue defines which
information that will be shown.
If you click the right mouse button with the cursor
position within this view, a dedicated shortcut menu will
appear.
The Colour scale view is normally located on the left side of
the Echogram view.
The Colour scale view visualizes the mapping of echo
strength into one out of 12 colours, light blue for weak
signals and dark brown for strong signals. Basically, each
discrete colour represents a 3 dB range of echo signal
strength implying that the next colour is selected every
time the echo strength doubles. However, as the number of
colours can differ, and dedicated colour schemes can be
used, this is not necessarily correct
A high numeric TVG value displays weak echo signals
properly while the stronger signals saturate into dark
brown colour. A low numeric value displays strong echo
signals properly. Weak signals below the lower limit of the
colour scale are not displayed (the display background
colour is used).
(CD6974F)
If you place the cursor inside the view, a small yellow
label will appear to give you a detailed readout of certain
information. The Tooltip dialogue defines which
information that will be shown.
If you click the right mouse button with the cursor
position within this view, a dedicated shortcut menu will
appear. If you click the middle button, the Colour Scale
dialogue box opens.
Use the following procedures to switch the ER60 echo sounder
on and off.
Power on
It is assumed that the echo sounder’s hardware and software are
properly installed and configured.
1Switch power on.
-The location of the power switches are individually
2Observe the hardware test messages and operating system
3Start the ER60 program.
4If the Login dialogue box appears, enter username and
assigned. The computer has its own power switch. The
transceiver(s) must be connected to separate power
supplies, and should have a remote power switch.
start-up messages.
password.
5On the Operation menu, select Normal.
6Once the echo sounder is running, start pinging by
pressing the “arrow” (") symbol on the toolbar.
Set-up
If no channels appear in the Normal Operation dialogue,
observe the following procedure:
1Open the Transceiver Installation dialogue box from the
Install menu
2Check that all frequency channels are properly installed.
3Open the Normal Operation dialogue box from the
Operation menu.
4Check the operational mode for each of the installed
channels. Adjust as required.
5Open the Window menu, and select New channel to open
a new channel window.
Power off
To switch off the ER60 echo sounder, observe the following
procedure.
16
1Select Exit on the File menu.
2Allow the computer to close all the ER60 software
applications.
- If desired, ER60 can be restarted using the desktop
icon.
850-164692 / Rev.B
Operational procedures
3Switch off the power on the General Purpose Transceiver
(GPT) and other peripherals (if any).
When the ER60 software has closed down, you need to close
down the computer as well.
1ClickontheStart button in the lower left corner of the
display.
2Exit the operating system in the normal matter.
3When notified, switch off the power.
This chapter presents a number of common procedures
frequently carried out on the ER60 echo sounder.
Changing the echogram settings
To change the echogram settings:
1Position the cursor in the Echogram view.
2Click the right mouse button.
3Select Echogram from the shortcut menu, and observe the
4Make the desired changes.
5Click Ok.
Changing the range
To change the range:
Echogram dialogue box.
1Position the cursor in the Echogram view.
2Click the right mouse button.
3Select Range on the shortcut menu.
4Observe either the Bottom Detection or Surface Range
dialogue box depending on the current setting.
5Make the required changes
6Click Ok.
Changing the vertical resolutio n
The vertical resolution of the echogram increases with a shorter
pulse duration.
Example: A pulse duration of 1.024 millisecond gives a vertical
resolution of 19.2 cm, whereas a pulse duration of 0.256
millisecond gives a vertical resolution of 4.8 cm. If the vertical
distance between two echoes is less than this, the two echoes
will be shown as one
Observe the following procedure to change the pulse duration:
1Click Operation on the main menu, and select Normal.
- Observe that the Normal Operation dialogue opens.
18
2Set the desired pulse duration for each of the frequency
channels.
3Click Ok.
A small value gives the best resolution, while larger values
provides you with a longer detection range.
850-164692 / Rev.B
Operational procedures
Setting minimum and maximum depth
Setting the minimum and maximum depth enables the echo
sounder to search for bottom lock.
NoteSetting both Minimum Depth and Maximum Depth to 0 m will
turn off bottom detection.
1Position the cursor over the depth information in the
Depth view.
2Click the right
mouse button.
3Select Bottom Detector from the shortcut menu.
4Set minimum and maximum depth to the desired values.
5Click Ok.
Use the following procedures to install, modify or delete
frequency channels from the echo sounder set-up.
Background
General Purpose Transceivers (GPT) physically connected to the
echo sounder’s ethernet interface are identified automatically by
the system. When you open the Transceiver Installation
dialogue box from the Install menu, a list will be provided.
A single frequency transceiver occupies one entry in the list.
This entry is identified as a Frequency Channel, and the line
displays the parameters for the channel. Entries in the frequency
channel list are shown in black, green, blue or red colour
identifying its current status.
To install a channel
1Select Transceiver on the Install menu.
-TheTransceiver Installation dialogue box opens, and
you are allowed to make changes.
2Click the desired entry (one of the black colour line
alternatives) in the Frequency Channel Selection list.
3Assign a transducer by selecting a transducer name in the
Transducer Selection list.
4Click OK.
To uninstall a channel
1Select Transceiver on the Install menu.
-TheTransceiver Installation dialogue box opens, and
you are allowed to make changes.
2Click the desired entry in the Frequency Channel
Selection list.
3Select the alternative NONE in the Transducer Selection
list.
4Click OK.
20
Related topics
→Transceiver Installation, page 123
850-164692 / Rev.B
Operational procedures
Data recording and playback
Background
You can set up the echo sounder to record unprocessed
transducer signals (sample data) and external sensor information
onto files. These files contain the necessary information to
reconstruct the situation during the real survey. The echo
sounder program reads these files during replay.
Recording
Prepar
ations
The recording is prepared as follows.
1Open the Output menu, and select File.
2Observe the Output File dialogue open.
3Select the tab Directory and push the Browse button if
you wish to change the direcory to which the files are
stored.
4Press the Raw Data tab to define how the raw data is to
be recorded:
- Save raw data: Start/stop recording of raw data.
- Range: Select the depth range to be recorded. This
range, referring to the transducer face, is independent
of any other range settings in the echo sounder.
- Echogram data: This is user defined excerpts of the
processed sample data (pixel data), ie the backscatter
value of the targets. The echograms are stored as time
tagged datagrams in separate files.
5If necessary, limit the maximum size of a single file.
- The current file will be closed and a new file will be
automatically generated when either the Max vessel
distance or the Max. file size limit is reached.
6Click OK to exit the dialogue.
850-164692 / Rev.B
Start and stop recording
Use either the Toolbar buttons or the Save raw data check box
in the File Output dialogue to start or stop recording of data.
Note on size of stored raw data
The size of the raw data files stored depends on several user
selections. From these selections you may estimate
approximately the total amount of raw data stored in a given
time period for each installed channel using the following
equation:
X = B bytes per sample * (8*R)/(c*tau) samples per ping * M
ping per sec * 24 * 3600 sec per day
21
Simrad ER60 Operator manual
where:
X = Total amount of stored raw data in bytes for one channel
B = 4 (Given by the resolution of the sample data)
R = Selected range in meters (User defined)
c = Sound speed in water in meters per sec (User defined)
tau = Pulse duration in seconds (User defined)
M = Ping rate in ping per seconds (User defined)
Thus, you can affect the amount of stored raw data by changing
e.g. the range, pulse duration, and ping rate settings.
Playback of a single file
The playback is started as follows:
1Open the Operation menu, and select Replay.
- Observe the Replay dialogue.
2Select the single file to be played.
- If no file is listed, click the Add button to select. Note
that only one replay file can be listed if this single file
shall be replayed.
3Click Loop if you wish to loop through the selected file
endlessly.
4Click OK.
5Click the Play buttonintheToolbar.
During playback, the ping rate is not limited by the speed of
sound in water. Hence, it is possible to select a higher ping rate
than during normal operation.
Playback of multiple files
The playback is started as follows:
1Open the Operation menu, and select Replay.
- Observe the Replay dialogue.
2Click the Add button to select playback files
- You can place any number of files in the list in the
Replay dialogue.
22
3Click the first file you wish to play back.
4Click Loop if you wish to loop through the the files
endlessly.
5Click OK.
6Click the Play buttonintheToolbar.
850-164692 / Rev.B
Loading...
+ 152 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.