About this Manual
Notes and Safety
This Operator’s Manual contains warning symbols which alert the user to check for hazardous conditions. These appear throughout this manual where applicable, and are defined below. To ensure the safety of operating perfor­mance of this instrument, these instructions must be adhered to.
Warning, refer to accompanying documents.
Caution, risk of electric shock.
Information
This instrument is designed to prevent accidental shock to the operator when properly used. However, no engineering design can render safe an instru­ment which is used carelessly. Therefore, this manual must be read carefully and completely before making any measurements. Failure to follow direc­tions can result in a serious or fatal accident.
Warranty
SIMPSON ELECTRIC COMPANY warrants each instrument and other ar­ticles manufactured by it to be free from defects in material and workmanship under normal use and service, its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one (1) year after delivery of such instrument or other article of equipment to the original purchaser be returned intact to it, or to one of its authorized service centers, with transportation charges prepaid, and which its examina­tion shall disclose to its satisfaction to have been thus defective; this warranty being expressly in lieu of all other warranties expressed or implied and of all other obligations or liabilities on its part, and SIMPSON ELECTRIC COM­PANY neither assumes nor authorizes any other persons to assume for it any other liability in connection with the sales of its products.
This warranty shall not apply to any instrument or other article of equipment which shall have been repaired or altered outside the SIMPSON ELECTRIC COMPANY factory or authorized service centers, nor which has been subject to misuse, negligence or accident, incorrect wiring by others, or installation or use not in accord with instructions furnished by the manufacturer.
SHOCK
Requirements tation, a shock hazard shall be considered to exist at any part involving a
and
Returns
HAZARD: As defined in American National Standard, C39.5, Safety
for
Electrical & Electronic Measuring &
Controlling
2
Instrumen-
www.GlobalTestSupply.com
Page 3
potential in excess of 30 volts RMS (sine wave) or 42.4 volts DC or peak and where a leakage current from that part to ground exceeds 0.5 milliampere, when measured with an appropriate measuring instrument defined in Sec­tion 11.6.1 of ANSI C39.5.
3
www.GlobalTestSupply.com
Page 4
1. INTRODUCTION ……………………………………………………
1.1 General ……………………………………………………………….. 5
1.2 Suppliers & Accessories ……………………………………………. 5
1.3 Safety Consideration ……………………………………………… 5
1.4 Technical Data ………………………………………………………... 5
2.
INSTALLATIO
2.1 Unpacking & Inspection …………………………………………….. 7
2.2 Warranty ………………………………………………………………. 7
2.3 Shipping ……………………………………………………………….. 8
2.4 Operating Procedure …………………………………………………. 8
N ………………………………………………………. 7
3.
CONTROLS,
3.1 Panel Description …………………………………………………….. 8
CONNECTORS & INDICATORS …………………….. 8
4.
OPERATIO
4.1 Safety Precaution …………………………………………………….. 9
4.2 Zero Adjustment ........................................................................ 10
4.3 Measuring DC Voltages, 250 Millivolt Range .......................... 10
4.4 Measuring DC Voltages, 1 V Range .......................................... 11
4.5 Measuring DC Voltages, 2.5 Through 500 V Ranges ............... 11
4.6 Measuring DC Voltages, 1,000 Volt Range Only ...................... 12
4.7 Measuring AC Voltages, 2.5 Through 500 Volt Range ............. 12
4.8 Measuring AC Voltages, 1000 Volt Range Only ......................... 13
4.9 Measuring Decibels ................................................................... 13
4.10 Measuring Resistance ............................................................... 13
4.11 Direct Current Measurement ...................................................... 14
4.12 Measuring Direct Currents, 50 mA Range ................................. 14
4.13 Measuring Direct Current, 0-1 Through 0-500 mA Ranges ....... 15
N …………………………………………………………. 9
5.
BATTERY REPLACEMENT ...........................................................
5
15
4
www.GlobalTestSupply.com
Page 5
1.
INTRODUCTION
1.1
General
The Simpson 160 Volt-Ohm-Milliammeter (hereafter referred to as the 160 or the Instrument) is a compact, easy-to-operate instrument used for measuring electrical characteristics of circuits and circuit components. The Instrument has a taut-band movement suspension with diode overload protection to pro­vide long, trouble-free service. The 100 degree mirrored dial arc and knife edge pointer provide excellent readability and eliminate parallax errors.
A polarity-reversing switch and a one-knob Function/Range selector simplify operation of the Instrument. The internal batteries are easily obtained and replaced.
Most of the component parts are mounted on a printed circuit board. This helps ensure uniform performance, reduces maintenance and extends the useful life of the Instrument.
1.2 Supplies &
All supplies and accessories required for the operation of the 160 are fur­nished with each Instrument and listed in Table 1-2.
Accessories
1.3 Safety
This Operator’s Manual contains cautions and warnings alerting the user to
hazardous operating and service conditions. CAUTION or WARNING sym­bols are placed throughout this publication, where applicable, and are de­fined on the inside cover of the manual under the heading SAFETY SYM­BOLS. Adherence to these instructions will help ensure safety of operating and servicing personnel and protect the instrument.
Considerations
1
.4 Technical
Data
Table 1-1. Technical
Data
DC Volts
Ranges: 2.5, 10, 50, 250V; 0.25, 1.0 and 1000V on sepa­Sensitivity: 20,000 per volt
Rated Accuracy: Within 2% DC and 3% AC of full scale on all
rate jacks.
ranges.
Ranges: 2.5, 20, 50, 250, 500V; 1,000V on
Sensitivity: 5,000 per volt. Indication: Full-wave average-responding: calibrated in RMS
* Frequency Response: Rated accuracy to 100,000 Hz on all ranges
**
Rated
Circuit-
To-Ground
Voltage:
separate jack.
for sinusoidal waveforms. through 50V; to 20 kHz on 250V range. 2-A.
1,000V AC/DC max.
5
www.GlobalTestSupply.com
Page 6
* See typical Response Curves, Figure 1-1. ** Per ANSI C39.5 April 1974: “The maximum voltage with respect to ground, which may safely and continuously be applied to the circuit of an instrument.”
Direct Current: Ranges: 1, 10, 100, 250, 500 mA; 50 µA on separate jack. Rated Accuracy: Within ±2% full scale, all ranges
DC Resistance
Ranges:
(30 center)
000,3-01 x R R x 10 0-30,000 R x 100 0-300,000 R x 1K 0-3 M R x 10K 30 M Accuracy: 3° arc Max. Voltage or Current Delivered:
R x 10 5 mA short circuit, 1.5V open circuit R x 100 0.5 mA short circuit, 1.5V open circuit R x 1K 0.75 mA short circuit, 22.5V open circuit R x 10K 0.075 mA short circuit, 22.5V open circuit
dB Ranges: –20 dB to +10 dB on 2.5 VAC range
Movement: Taut-Band 100° arc, 50
Dial Arcs: One arc for ohms, one arc for DC, two arcs for
Scale Length: 3.0 inches
(30 K
(300 K
–8 dB to +22 dB on 10 VAC range +6 dB to +36 dB on 50 VAC range +20 dB to +50 dB on 250 VAC range Zero dB referenced to 1 milliwatt at 600
AC, one arc for dB.
6
center)
(300
(3 K center)
center)
center)
tiucric nepo V5.1 ,tiucric trohs Am 051 x R
A full scale
www.GlobalTestSupply.com
Page 7
Lead
Reversal: Polarity Reverse Switch; on DC and ohms 10.
Operating Position
Batteries
:
1.5 V AA penlight and AA 3.6V.
Movement & Protection
Test Op
era Range: Size: Weight
Probe Tip Test Leads 02055 Operator’s Manual 05-111658
Table 1-3.
Alligator Clip Leads 01927
2.
:
Leads: Custom molded elbow terminals; 3 ft. flexible
ting Temp
error over the range of +25 °F to +130 °F.
4-9/16 x 3-5/16 x 1-3/4 (inches)
:
Approx. 12 oz., complete
INSTALLATION
This section contains instructions for the installation and shipping of the 160. Included are unpacking and inspection procedures, warranty, shipping, and operating procedure.
:
Horizontal or vertical; rubber feet prevent slipping on
moderate slopes.
Indicator
Silicon double diode across movement terminals.
probe tips.
erature
:
75 °F for rated accuracy; less than 4% additional
Table 1-2. Items and Accessories Supplied
Description
Additional Accessories
Description
Padded Nylon Case 00836
with
This
Cat. Number
Cat. Number
Instrument
2.1 Unpacking &
Examine the shipping carton for signs of damage before unpacking. Unpack and inspect the Instrument for possible damage in shipment. If damaged, notify the carrier and supplier before using the Instrument. Check that all furnished items and accessories are included (Table 1-2). Save all shipping materials for future use.
Inspection
2.2 Warranty
The Simpson Electric Company warranty policy is printed on the inside front cover of this manual. Read it carefully before requesting a warranty repair. For all assistance, contact the nearest Authorized Service Center for instruc­tions. If necessary, contact the factory directly, give full details of the difficulty and include the Instrument model number and date of purchase. Service data or shipping instructions will be mailed promptly. If an estimate of charges for nonwarranty or other service work is required, a maximum charge estimate
7
www.GlobalTestSupply.com
Page 8
will be quoted. This charge will not be exceeded without prior approval.
2.3
Shipping
Pack the Instrument carefully and ship it prepaid and insured to the proper destination.
2.4 Operating
The Instrument may be operated in a horizontal or vertical position.
Procedure
3.
CONTR
OLS, CONNECTORS & IND
All operating and adjustment controls, connectors, and indicators are de­scribed in this section along with a list (Table 3-1) describing their functions. Become familiar with each item before operating the Instrument.
ICATORS
3.1 Panel
Table 3-1 lists all Controls, Connectors, and Indicators.
Description
Figure 3-1. Front
Panel
8
www.GlobalTestSupply.com
Page 9
Table 3-1. Controls, Connectors, and
1
.
Function and Switch
2.
Zero Ohms Control:
3.
Input Jacks: There are six input jacks: Four of these are on the
4.
Polarity Reversing Switch:
NOTE:
reversing switch must be set to the +DC position to obtain readings.
4.
OPERATION
Before operating the 160, review the SHOCK HAZARD definition printed on the inside front cover of this manual.
Range
:
Adjust
When making measurements of the 50 µA, 250 mV, 1V ranges, the
This switch, located in the lower center of the panel, is used to select the desired current, voltage, or re­sistance function and appropriate range.
This control, located at the lower left on the front panel, is used to obtain a “0” indication on the ohms scale when the test leads are shorted together. Dur­ing operation, the zero indication is checked each time the ohmmeter is to be used. This permits com­pensation for aging internal batteries, and allows them to be used for a longer period of time.
right side of the panel and two directly below the zero adjustment screw. The four jacks on the right are identified COM –, +, +1V, and +50 / +250 mV. The COM + jack is used for all ranges and functions with the exception of the 1V, 50 A, 250 mV, 1,000 VDC and 1,000 VAC ranges. The two jacks below the Instrument are identified 1,000 VDC and 1,000 VAC and are used to extend the 500 VDC and 500 VAC ranges
This switch, located above the ADJ control, allows simple lead polarity reversal when making DC or resistance measurements on any range except the 50 µA, 250 mV, or 1V positions. For normal opera­tion, set this switch to +DC position, using COM – as the reference. Conversely, negative po­larity signals can be measured without interchang­ing leads by setting the switch to the – DC position. When the VOM is set on any resistance range, this switch reverses the polarity of the measuring poten­tial in the same manner.
Indicators
4.1 Safety
Instruments of this type are intended for use only in low-power, consumer product type applications, such as TV or radio. Their use is not recommended in high-power circuits such as power plants, substations or high power trans­mitter circuits, where the likelihood of corona, together with sufficient energy
Precautions
the
9
www.GlobalTestSupply.com
Page 10
to sustain flash-over arcs, is a serious hazard.
1. Do not hold the instrument when working in circuits that might contain a shock hazard.
2. Inspect the test leads, probes, connectors, and insulating boots for dam­age or deterioration before each use. If defects are found, replace the leads immediately with leads designed for the Instrument. Do not use test leads inferior to those furnished with the Instrument.
3. Never disconnect the COMMON lead from an active circuit while the other lead is connected to an energized circuit. The COMMON lead becomes
unexpectedly “hot” in such a case and can be a shock hazard. Develop
safe habits by always turning off power to the measured circuit and dis­charging any capacitors before handling the test leads.
4. Become familiar with the circuit to be measured and locate any shock hazards before attempting measurements. Remember that high voltages might appear unexpectedly in a faulty circuit.
5. Measuring electricity is particularly hazardous in the presence of humid­ity or moisture. Hands, shoes, floor and workbench must be dry.
6. Avoid measuring in circuits where composite voltages can exceed the Instrument’s safe limits. When measuring DC voltages, the Instrument will not respond to (and thereby not indicate) the presence of AC compo­nents.
7. Be alert for corona in the measured circuit. Signs of corona include a buzzing sound, ozone odor and a pale blue emanation. Its presence in­dicates high voltage; an unexpected or unknown path might lead to a flash-over.
8. Do not work alone when measuring where a shock hazard can exist. Notify someone nearby of your intentions.
9. Do not connect the Instrument to an electrically energized circuit in a hazardous area. Do not use the Instrument to check electrical “blasting” circuits.
10. No general purpose VOM is to be used to make electrical measurements on blasting circuits or blasting caps. Use VOMs designed for this pur-
4.2 Zero
pose only.
Adjustment
Before measuring, check to see that the pointer indicates zero when the In­strument is in the operating position. If the pointer is off zero, make the re­quired correction by turning the screw located directly below the “Simpson 160” legend.
4.3 Measur
Use care when making measurements with the 160 on the 250 mV range. An excessive voltage applied when in this range can damage the Instrument.
ing DC
Voltages, 250 M
illivolt Range
10
www.GlobalTestSupply.com
Page 11
1. Connect the black test lead into the COM (–) jack and the red test lead into the +50 µA +250 mV jack.
2. Set polarity switch to the + DC position.
3. Set the Range Selector Switch to the 50 µA position (common to the 50 VDC position).
4. Connect the black test lead to the negative (–) side of the circuit to be measured and the red test lead to the positive (+) side of the circuit.
5. Read the voltage on the black arc marked DC. Use the figures marked 0-
2.5 and multiply reading by 100 for the millivolt reading. If the pointer moves to the left of zero, reverse the test lead connections, as the revers­ing switch must be kept in the + DC position for this range.
4.4 Measuring DC Voltages, 1 V
Range
1. Connect the black test lead to the COM (–) jack and the red test lead to
the + jack.
2. Set the polarity switch to the + DC position.
3. Set the range switch to the 1 VDC position (common to the 10 VDC posi­tion).
4. Connect the black test lead to the negative (–) side of the circuit to be measured and the red test lead to the positive (+) side.
5. Read the voltage on the black arc marked DC. Use the figures 0-10 and divide the reading by 10 to obtain voltage reading. If the pointer moves to the left of zero, reversing switch must be kept in the +DC position for this range.
4.5 Measuring DC Voltages, 2.5 Through 500 V
1. Connect the black test lead to the COM (–) jack and the red test lead to the + jack.
2. Set the range switch for any of the five DC volts range positions desired. These are marked 2.5, 10, 250, and 500 VDC. When in doubt as to which range to use, always start with the higher voltage range as a protection to the Instrument.
3. Connect the black lead to the negative (–) side of the circuit to be measured and the red test lead to the positive (+) side of the circuit.
4. Set the polarity switch to the + DC position. Turn the power on in the circuit to be tested. If the pointer deflects to the left of zero, the actual circuit polarity is the reverse of that anticipated. In this case, turn off power in the circuit to be tested, set the polarity switch to the – DC position and turn power on again.
5. Read the voltage on the black arc marked DC which is second from the top of the dial. If the voltage is within a lower range, the switch may be set for a lower range to obtain a more accurate reading.
2.5 VDC 10 and 50 VDC 250 VDC 500 VDC
range: Use the 0-2.5 scale and read the value directly.
ranges: Read the corresponding scale directly.
range: Use the 0-2.5 scale and multiply reading by 100. range: Use the 0-50 scale and multiply by 10.
Ranges
11
www.GlobalTestSupply.com
Page 12
4.6 Measuring DC Voltages, 1,000 Volt Range Only
Use extreme care when working in high voltage circuits. Even though the Instrument and test leads are well insulated for this voltage, do not handle when power is on in the circuit.
1. Set the range switch to the 1,000 VDC position (common with the 500 VDC position).
2. Connect the black test lead to the COM (–) jack and the red test lead to the 1,000 VDC jack.
3. Set the polarity switch to the + DC position.
4. Connect the black test lead to the negative (–) side of the circuit and the red test lead to the positive (+) side.
5. If the pointer deflects to the left side of zero, the actual circuit polarity is the reverse of that anticipated. (In this case see Paragraph 4.5, step d.)
6. Read the voltage, using the 0-10 scale on the black arc marked DC and multiply the reading by 100.
4.7 Measuring AC Voltages, Range
The 160 rectifier circuit responds to the full wave rectified average value of the AC voltage being applied. The Instrument dial, however, is calibrated in terms of RMS voltage, which will be correct for all sinusoidal waveforms.
NOTE
:
range an external blocking capacitor must be employed where measurements of AC superimposed on DC are encountered.
1. Connect the black test lead to the COM – jack and the red test leads to the + jack.
2. Set the range switch for any of the five VAC range positions desired. These are marked 2.5 VAC, 10 VAC, 50 VAC, 250 VAC, and 500 VAC. When in doubt as to which range to use, always start with the highest voltage range as a protection to the Instrument.
3. Turn the power on in the circuit to be tested. Read the voltage on the red arc marked AC.
0-2.5 VAC range:
10 and 50 VAC
250 VAC
500 VAC
Since the 160 will respond to DC voltage when set on any AC volt
Read the value directly on the special arc marked
2.5 VAC.
ranges: Read the red arc marked AC, and use the corre-
sponding black numbers immediately below the arc.
range: Read the red arc marked AC using the 0-2.5 figures
and multiply the reading by 100.
range: Read the red arc marked AC using the 0-50 figures
and multiply the reading by 10. If the voltage is within a lower range, the switch may be set for the lower range to obtain a more accurate reading.
2.5
Through 500
Volt
12
www.GlobalTestSupply.com
Page 13
4.8 Measuring AC Voltages, 1000
Volt
Range
Only
Use extreme care when working in high voltage circuits. Even though the Instrument and test leads are well insulated for this voltage, do not handle when power is on in the circuit.
1. Set the range switch at 1,000 VAC position (common with the 500 VAC position).
2. Turn off power in the circuit being measured.
3. Connect the black test lead to the COM (–) jack, and the red test lead to the 1,000 VAC jack.
4. Turn on power in the circuit being measured.
5. Read the voltage on the red arc marked AC using the 0-10 figures and multiply the reading by 100.
4.9 Measuring
The decibel scale at the bottom of the dial is numbered from –20 to +10. To measure decibels, proceed according to instructions for AC voltages, and read the dB arc. The dB scale is calibrated for direct reading on the 2.5V range. Scale factors for other ranges and dB reference at 0.006 watts into 500 ohms are given in the table below.
4.10
Measuring
When resistances are measured, the internal batteries of the 160 furnish power for the measuring circuit. Correction for battery deterioration over long peri­ods of time is provided by means of the Zero Adjust control which is part of the ohmmeter circuit.
1. Set the range switch at the desired resistance range position.
2. Connect the black test lead to the COM (–) jack, and the red test lead to the + jack.
3. Connect the contact ends of the test leads together.
4. Observe the Instrument indication. Look for a reading of “0” on the OHMS arc, which is at the top of the dial.
5. If the pointer does not read “0”, rotate the ZERO OHMS knob at the lower left on the front panel until it does. If the pointer cannot be brought up to the “0” mark, replace the appropriate battery.
NOTE:
Disconnect power from any resistor or circuit to be measured before measuring resistance. Do not apply any power until the measurements are completed and test leads are disconnected.
Decibels
Range
1
2.5V direct –7 10 V +12 +5 50 V +26 +19 250 V +40 +33
mW @
600 6
mW @
Resistance
500
13
www.GlobalTestSupply.com
Page 14
6. Connect the test leads across the resistance which is to be measured. If there is a “forward” and “backward” resistance, such as with diodes, ob­serve polarity in the lead connections to control each direction of test. With the polarity switch in the + DC position, the + jack will provide a positive potential referred to COM – jack. Setting the switch to the – DC position will reserve this measuring potential.
7. Read the indication on the OHMS arc at the top of the dial. Note that the arc reads from right to left for increasing values.
8. Multiply the reading by the multiplier factor indicated at the switch posi-
tion; the result is the resistance value in ohms. “K” on the dial and panel stands for “times one thousand”.
NOTE:
The resistance of nonlinear components will measure as different val­ues on different ranges of the 160. For example, a diode could measure 80 the R x 1 range, and 300 on the R x 10 range. This is normal and is the result of the diode characteristic. The difference in readings does not indicate faulty operation of the ohmmeter circuit.
4.11
4.12
Direct Current
1. Do not switch the range setting of the Range or Polarity Switches while
the circuit under measurement is energized.
2. Never disconnect the test leads from the circuit under measurement while
the circuit is energized.
3. Always turn the power off and discharge all the capacitors before the
setting of the switches is changed, or the leads disconnected.
4. Never exceed the Circuit-To-Ground voltage of the Instrument (1,000 V
max: Table 1-1, Rated Circuit-Ground Voltage).
5. Always connect the Instrument in series with the ground side of the cir-
cuit.
6. In all direct current measurements make certain the power to the circuit
being tested has been turned off before connecting and disconnecting test leads or restoring circuit continuity.
Measuring
Measurement
Direct
Currents,
50 µA
Range
Never connect the test leads directly across any source of voltage when the 160 is used for current measurements. This will damage the Instrument.
1. Connect the black test lead to the COM – jack, and the red test lead to the
+ 50 mA jack.
2. Set the range switch at 50 mA (common with 50 VDC).
3. With the circuit power turned off, open the circuit at the point where cur-
rent is to be measured. Connect the Instrument in series with the circuit, observing proper polarities when making connection.
4. Turn on power to the circuit being measured. If the pointer is deflected to
the left of zero, the polarity is opposite to that anticipated. Turn power off
14
www.GlobalTestSupply.com
Page 15
and reverse the leads. The polarity switch must be kept in the + DC posi­tion on this range.
5. Read the current directly on the black arc marked DC, using the 0-50 scale. The current value is shown in microamperes.
6. Turn off power to the circuit. Remove the test leads and restore the circuit continuity.
4.13
Measuring Direct Current, 0-1 Through 0-500
mA
Ranges
1. Connect the black test lead to the COM (–) jack and the red test leads to the + jack. Set the polarity switch to the + DC position.
2. Set the range switch to any of the five mA direct current range positions, as desired. The switch positions are marked 1 mA, 10 mA, 100 mA, 250 mA, and 500 mA. When in doubt as to which range to use, always start with the highest ranges as a protection to the Instrument.
3. When the circuit power is turned off, open the circuit at the point where current is to be measured. Connect the Instrument in series with the cir­cuit, observing proper polarities.
4. Apply power to the circuit being measured. If the pointer deflects to the left of zero, the polarity is reversed. Turn off the power. Set the polarity switch to the –DC position and then reapply the power.
5. Read the current on the black scale marked DC, which is second from the top of the dial.
mA Range
1.0 0-10 Divide by 10 10 0-10 Read direct value 100 0-10 Multiply reading by 10 250 0-2.5 Multiply reading by 100 500 0-50 Multiply reading by 10
6. Turn off power to the circuit. Remove the test leads and restore circuit continuity.
Use
Scale Reading
5.
BATTERY
When the Ohms Adjust control cannot be adjusted for zero ohms (with shorted test leads), it is generally an indication that the battery must be replaced. Fail­ure to do so promptly can result in damage to the 160 due to chemical leak­age from the battery.
To replace the battery, remove the screw holding the back cover to front panel and separate the back cover from the front panel. Loosen the screw securing the battery contact plate located at the top of the 160 panel assembly, then rotate the contact plate enough to allow removal of the batteries. When in­stalling new batteries, note battery placement and polarity as indicated on the contact plate.
REPLACEMENT
15