Simpson Electric S663 Instruction Manual

S663
Digital Preset / Totalizer Counter & Rate Meter
Operation Manual
SIMPSON ELECTRIC COMPANY
520 Simpson Avenue, Lac du Flambeau, WI 54538-0099
(715) 588-3311 FAX (715) 588-3326
Printed in U.S.A. Part No. 06-116548 Edition 8, 06/07
!
To the best of our knowledge and at the time written, the information contained in this docu­ment is technically correct and the procedures accurate and adequate to operate this instrument
Menu
Category
Parameter
Name
Choices /
Format
Description Tech
Note
in compliance with its original advertised specifications.
Notes and Safety Information
This Instruction Manual contains warning headings that alert the user to check for hazardous conditions. These appear throughout this manual where applicable, and are defined below. To ensure the safety of operating performance of this instrument, these instructions must be ad­hered to.
Caution, refer to accompanying documents.
Caution, risk of electric shock.
Technical Assistance
SIMPSON ELECTRIC COMPANY offers assistance Monday through Friday 8:00 am to 4:30 pm Central Time by contacting Technical Support or Customer Service at (715) 588-3311. Internet: http://www.simpsonelectric.com
SP4 000040*
rStPos 000000*
SP4
Set point #2 High.
Values: -99999 to
999999. Decimal point will appear according to the current Output 2
Source setting.
Reset Position
Count value is set to this
when an Auto or Manual
Reset event occurs. Values: -99999 to
999999. Decimal point will appear according to the user-programmed
Count DP position
4.6
4.6
Warranty and Returns
SIMPSON ELECTRIC COMPANY warrants each instrument and other articles manufactured by it to be free from defects in material and workmanship under normal use and service, its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one (1) year after delivery of such instrument or other article of equipment to the original purchaser be returned intact to it, or to one of its authorized service centers, with transportation charges prepaid, and which its examination shall disclose to its satisfaction to have been thus defective; this warranty being expressly in lieu of all other war­ranties expressed or implied and of all other obligations or liabilities on its part, and SIMPSON ELECTRIC COMPANY neither assumes nor authorizes any other persons to assume for it any other liability in connection with the sales of its products.
This warranty shall not apply to any instrument or other article of equipment which shall have been repaired or altered outside the SIMPSON ELECTRIC COMPANY factory or authorized service centers, nor which has been subject to misuse, negligence or accident, incorrect wiring by others, or installation or use not in accord with instructions furnished by the manufacturer.
Under the normal field usage there is no need to remove the front bezel of this product. The front bezel of this product should only be removed by a qualified technician.
Menu
Category
rESEt SEtuP
Parameter
Name
ArESEt diSAbL*
rStbtn enable*
PonrSt no*
Choices /
Format
At SP1 At SP2 At SP3 At SP4 AFtoP1 AFtoP2
diSAbL
YES
Description Tech
Auto Reset Mode
Selects when an auto reset function is to occur. Disabled at a set point or after output times out. See also the RSTPOS parameter in the Setpnt Setup category
Reset Button
Enable or disable front panel reset button.
Power On Reset Select
whether count reset event will occur at power-up.
Note
4.7
4.8
4.8
End
* = factory default settings
1 38
Exit Programming Menu
Menu
Category
Parameter
Name
Choices /
Format
Description Tech
Note
Contents
1 Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.1 General Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
SEtPnt SEtuP
SP1 000010*
SP2 000020*
SP3 000030*
SP1
Set point #1 Low.
Values: -99999 to
999999. Decimal point will appear according to the current Output 1
Source setting.
SP2
Set point #1 High.
Values: -99999 to
999999. Decimal point will appear according to the current Output 1
Source setting.
SP3
Set point #2 Low.
Values: -99999 to
999999. Decimal point
4.6
4.6
4.6
1.2 Part Number Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Option Module Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Hardware Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.1 Panel Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Removing / Installing Option Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 120/240 VAC Power Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 Standard Input Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Quadrature Input Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.6 Excitation Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.7 Single and Dual Relay Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 Display & Keypad Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1 Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.2 Display Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
3.3 What the Keys Do in Display Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4 What the Keys Do in the Programming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5 Special Start-Up Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
will appear according to the current Output 2 Source setting.
* = factory default settings
4 Programming Operations and Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
4.1 Password and Security Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2 A/B Channel Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
4.3 Scaling and Decimal Point Position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4 Rate (Frequency) Scaling and Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.5 Output Control Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
4.6 Set Point Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.7 Auto Reset Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
4.8 Miscellaneous Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Appendix A: Technical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
A.1 Functional Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
A.2 Electrical, Environmental and Mechanical Specifications. . . . . . . . . . . . . . . . . . . . . 29
Appendix B: Programming Quick Reference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
37 2
1 Product Description
Menu
Parameter
Choices /
Description Tech
1.1 General Description
The S663 preset counter and rate meter fits a 1/8 DIN standard cutout and is perfect for tight spaces, extending only 3.24” (82mm) behind the panel. The unit is UL listed. The unit is for indoor use at altitudes up to 2000m, temperatures between 0° and 40°C, and installation catego­ry III, pollution degree 2.
The counter is powered from 120 or 240 VAC. The nonvolatile EEPROM retains all program­ming and count information when power is removed or interrupted.
Category
Name
tiMmEd <— LAtCHEd<— SP3=Lo <—
dELAY2 010.00*
untiL2 rESEt *
Format
—>At SP3 —>AtSP3* —>SP4=Hi
SP1 SP2 SP3 SP4
Output 2 Bindings
Reminder message indicates which and how
the set points are used for comparison. Which
message is displayed is
determined by the
Output 2 mode
selected.
Delay 2
Output 2 delay time.
Appears only if Output 2
mode set to timed.
Until 2
Output 2 latched until
parameter. Appears only
if MMode2 = latch and SRC2 = Count.
Note
4.5
4.5
4.5
Display scaling, preset values and all menu functions are easily programmed from the front panel, following easy-to-use word prompts. Front Panel Reset Disable and Password Lockout Protection features guard against unauthorized or accidental changes.
The counter accepts count rates up to 20 KHz and pulses from different types of sensors, in­cluding Quadrature, CMOS or TTL circuits and PNP or NPN devices.
The optional, field-replaceable, single/dual relay module enhances the counter from a passive display device to an integral control element for your application. Outputs may be controlled by count or rate values with independent reset operation. The counter has latching, boundary or timed (0.01 to 599.99 seconds) alarm output modes.
An optional 12 VDC (100mA) excitation output module can provide power for external sen­sors.
rStPoS
* = factory default settings
3 36
Menu
Parameter
Choices /
1.2 Part Number Identification
Description Tech
Category
oPut 2 SEtuP
Name
Format
untiL 1 rESEt*
SP1 SP2 SP3 SP4 rStPoS
m M odE2 diSAbL
tiMmEd LAtCH* bound
SrC 2 Count*
rAtE
Until 1
Output 1 latched until parameter. Appears only
if M m odE1 = LAtCH and
SrC1 = Count.
Output 2 Mode
Set the mode of operation for Output 2.
Can be disabled, timed, latched or boundary mode.
Output 2 Source
Select which Display
Value to be used for
Note
4.5
4.5
4.5
The following matrix indicates the configuration of your S663 counter.
Basic Unit
S663
Power Supply
120 VAC 240 VAC
1 2
Input
Standard
Quadrature
Output
1 2
One relay
Two Relays
None
0 1 2
Excitation
None 12VDC
0 1
1.3 Option Module Summary
Other
None
0
output 2 comparisons/ matchpoint. See also
SP3 and SP4.
* = factory default settings
1
Input Excitation Power
2
4
Supply
5
Output
Figure 1. Option Module Slots (Rear View)
The S663 is a modular product which uses field configuring slide-in modules. The modules slide easily into the rear of the counter. Figure 1 displays the functional assignments for each module position.
Table 1 describes available option modules for the S663.
Table 1. Option Module Summary
Module
See
Type P/N Description
Slot
Section 1 Input Standard Input Module 2.4 1 Input Quadrature Input Module 2.5 2 Ext. 45064 12 VDC Excitation Module 2.6 4 Power 120 VAC Power Module 2.3 4 Power 240 VAC Power Module 2.3 5 Output 45062 Dual Relay Module 2.7 5 Output 45063 Single Relay Module 2.7
35 4
2 Hardware Setup
Menu
Parameter
Choices /
Description Tech
2.1 Panel Installation
The S663 1/8 DIN counter requires a standard 1/8 DIN panel cutout of 1.77” (45mm) high by
3.62” (92mm) wide. To install the counter into a panel cutout, remove the clips from the side of the counter. Slide the counter through your panel cutout, then slide the mounting clips back on. Press evenly to ensure a proper fit.
3.62"
92mm
3.24"
82mm
1.77"
45mm
1.74"
44mm
2.04"
51.8mm
Category
Name
Format
SrC 1 Count*
rAtE
tiMmEd<— LAtCHEd<— SP1=Lo <—
—>At SP1 —>AtSp1* —>SP2=Hi
Output 1 Source
Select which Display Value to be used for output 1 comparisons / matchpoints. See also
SP1 and SP2.
Output 1 Bindings
Reminder message indicates which and how the set points are used for comparison. Which message is displayed is determined by the Output 1 Mode selected.
Note
4.5
4.5
ENTER
3.93"
.52"
13.2 mm
%
RESET
99.8 mm
Figure 2. Counter and Panel Cut-Out Dimensions
dELAY1 010.00*
* = factory default settings
Delay 1
Output 1 delay time. Appears only if Output 1 mode set to timed.
4.5
Figure 3. Panel Mounting Clips
5 34