Mounting Requirements
Insert the Mini-Max through the panel, and then slide the mounting brack­et on to the Mini-Max. The mounting bracket allows Mini-Max units to be stacked side-to-side or top-to-bottom and maintain the DIN standard panel arrangements in 24mm by 72mm multiples. Panel cutout instruc­tions for stacking multiple units are provided under “Stacking Features.”
Minimum Depth Indicator - Less Than 2.5” (60mm) of Space Required Behind the Panel
Stackable Mounting Bracket Included for Easy Installation
4-1/2 Digit, 0.5” (12.7mm) High LCD Display with Optional Negative Image, Bright Red Backlighting
Limited Range Display Scaling
Standard Screw Terminals for Easy Installation
Six Current Ranges: 200µA, 2mA, 20mA, 200mA, 2A, 5A
85-250VAC or 9-32VDC Power Supply
Mini-Max M245 Series Digital Panel Meter
Installation and Panel Cutout
Simpson’s Mini-Max Current Indicators provide high quality, accuracy and reliability in a compact, 60mm deep case. Units offer 4-1/2 digit, 0.5” (12.7mm) LCD display and are available with a bright red, negative image backlight option. All units feature user-selectable decimal point, auto zero and limited scaling capabilities.
A unique mounting bracket is provided to allow for vertical or horizontal stacking of multiple indicators. All Mini-Max units feature a 3/64 DIN, high-impact plastic case. The standard units have a clear viewing window, and the units with optional negative image, red backlighting have a red window.
DC Current
2.68"
(68mm)
Page 2
DISPLAY Type: 7-segment LCD Height: 0.5” (12.7mm) Decimal point: 3-position programmable Overrange indication: Most significant
digit = “1” Backlighting: Optional negative image, red LED backlighting
Polarity: Auto with “-” indication, “+” implied
POWER REQUIREMENTS AC Volt: 85-250VAC @40-440Hz DC Volt: 9-32VDC Power Consumption:
85-250VAC: 2.5VA min/4VA max 9-32VDC: 1.5VA min/3VA max
Rated Circuit to Ground Voltage: 750Vrms
ACCURACY @ 25°C
±(0.04% of reading + 1 count)
2A: ±(0.25% of reading + 1 count 5A: ±(0.5% of reading + 1 count)
ENVIRONMENTAL Operating Temperature: 0 to 55°C Storage Temperature: -10 to 60°C Relative Humidity: 0 to 85% non condensing
@ 40°C
Temperature Coefficient:
(0.02% of input 0.2 digits) / C
Warmup time: Less than 20 minutes
NOISE REJECTION NMRR: 60dB, 50/60Hz CMRR: (w/1Kunbalanced @ 60Hz): 90dB min
ANALOG TO DIGITAL CONVERSION Technique: Integrating
Rate: 3 samples/second-typical MECHANICAL Bezel: 0.95” x 2.84” (24mm x 72mm) Depth: 2.36” (60mm) Panel cutout: 0.88” x 2.68”
(22.2mm x 68mm)
Weight: 3.5oz (99.2g) Case Material:
94-0,UL-rated, glass-filled thermoplastic
Specifications
Wiring Display
Display Scaling
Mini-Max indicators have limited range coarse and fine adjustments for display scaling. There are no optional connections required for these to function. The “coarse” calibration R12 will allow a limited range of scaling val­ues. The meter can be scaled down to 1/2 the value of the input, but notscaled up. Maximum scaling is maximum input for a maximum reading of 1.9999, which ever is lower. Example: a 2 AMP input has a maxi­mum reading of 1.9999 counts, so cyou can’t double the 2 AMP’s, but you can make 1 AMP read 1.9999. the “fine” calibration R9 allows for an approximate range of 1% of the “coarse” calibration. Apply full scale input to the meter. Adjust R12 to be within 1% of the desired scaled value, then use R9 to obtain the final desired result.
Using a screwdriver or thumbnail, spread tab on each side of case to unlock top half. Lift rear top half and slide away from front of meter.
FRONT
OF METER
LOCKING
TA B
TOP HALF
OF CASE
Decimal Point: To select a decimal point, connect the appropriate DP pin (DP1, DP2, DP3 or DP4) to the DP COMMON (HOLD). Unused DP inputs
may remain unconnected (open).
Display Hold: Connect HOLD to DP COMMON (HOLD). If this feature is not required, the DP COMMON (HOLD) pin may remain unconnected
(open). Hold is optional on this model, and when equipped, replaces DP4. This feature is available from our modification centers.
Input Signal: Connect the IN HI and IN LO to the signal to be monitored.
Input Power: Connect power to the L1 and L2 terminals. For AC powered units, L1 and L2 are not polarized.
For 9-32 DC powered units, L1 must be positive with respect to L2.
(Input Signal) IN HI (Input Signal) IN LO
DP3 DP2 DP1
HOLD
L1 (POS) (Power) L2 (NEG) (Power) EX (NEG) EX (POS)
DP COMMON (HOLD)
These instruments are designed for maximum safety to the operator when mounted in a panel according to instructions. They
are not to be used unmounted or for exploratory measurements in unknown circuits.
Range Resolution Voltage Max. Input
Drop (unfused)
200µA 10µA 200mV 10mA
2mA 100µA 200mV 40mA 20mA 1µA 200mV 100mA 200mA 10µA 200mV 400mA
2A 100µA 200mV 3A 5A 1mA 50mV 6A
INPUTS: DC Current
DP4/
NNoottee::
Any physcal damage to the meter during
calibration will void the warranty
Page 3
Stacking Features
The mounting brackets, included with every Mini-Max, can be connect­ed together. Multiple units can be mounted in a single opening, allow­ing perfect alignment.
To punch one hole for multiple units, be sure to adjust the standard panel cutout dimensions as shown here; otherwise the meters will not fit properly in the hole.
Mounting multiple units is quick and easy. Install the first meter (bottom unit first if stacking vertically). Position the next mounting bracket snug­ly against the first one, and slide the second meter into place. Repeat for remaining units.
2.68" (68 mm)
.071" (1.8 mm)
.88" (22.2 mm)
Standard cutout
Add to standard when stacking
.88" (22.2 mm)
2.68" (68mm)
Standard cutout
.16" (4.0mm)
Add to standard when stacking
Horizontal
Vertical
Vertical
Horizontal
Excitation Output
Excitation output is available in the Mini-Max for powering external transmitters or transducers. Excitation is available from the EX POS and EX NEG terminals. This source is isolated from the measurement input as well as the input power circuits. The voltages available are 12VDC or 24VDC, with a maximum load current of 25 milliamps. This feature elim­inates the need to mount an external DC power source for transducers or sensors used in your application.
Do not exceed an excitation
load current of 25mA!
EX NEG EX POS
Backlight Option and Decimal Point Selection
Negative image, bright red backlighting is avail­able in the Mini-Max. This illumination allows the unit to be read in low light areas. Backlighting power is supplied by the Mini-Max, so no addi­tional external power is required.
Decimal point selection is accomplished by con­necting the appropriate DP terminal to the DP Common terminal.
DP4
DP3
DP1
DP2
DP COMMON
321 G
(HOLD)
4*
DP 4 DP 1
DP 2 DP 3
*NOTE: If the Display Hold option is ordered, it replaces DP4 on the Terminal Connector.
Page 4
Application Example
Ordering Information Safety Symbols
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, practice, or the like, which, if not correctly adhered to, could result in damage to or destruction of part or all of the instrument.
A company needs to monitor the power supply voltage and load current of a 12VDC motor.
Voltage: A Mini-Max DC Volt meter, con­figured for the 20 volt range, is installed in parallel with the power source. The IN HI is connected to the positive lead of the power supply. The IN LO is connected to the negative lead of the power supply.
Current: A second Mini-Max DC Volt meter, configured for the 200 millivolt range, is used with a 100A/50mV shunt (sold separately) to measure the load cur­rent. The meter is first scaled to read
100.0 at 50mV by applying a 50mV input to the meter and adjusting R9 and R12 until the display reads 100.0. The meter is connected to the shunt, and the shunt is installed in series on the negative leg of the power supply. The positive lead of the shunt is connected to the IN HI terminal,
Non Backlight Negative Image Red
Basic Unit Display DPM Power Supply Range Excitation Output
M245
0 2
0 1
0 1 2
21 22 23 24 25 26
4-1/2 Digit Indicator
85-250VAC 9-32VDC
200µA 2mA 20mA 200mA 2A 5A
None 12VDC 24VDC
AC MOTO R
POWER SUPPLY
5 A AC UNIT
200 VAC UNIT
200Hz Frequency Unit
IN HI IN LO
IN HI IN LO
-
+
External shunts enable digital panel meters to indicate higher currents than can be provided with self-contained internal shunt meters. The shunt is installed in series with the load and source. The shunts produce a DCmV drop which is sent to the display unit. The Mini-Max can be scaled to display the actual current between the load and the source. Simpson manufactures portable and switchboard shunts.
Each shunt includes 5’ leads.
The Mini Max Voltage Indicator can be configured by making an entry for each box
NOTE: Display hold feature is configured at the factory only. Must specify at time of order.
25 DCmA max
output
and the negative lead connected to the IN LO terminal. The Mini-Max units will indi­cate the power supply voltage and load current of the motor.
Portable Shunts 50mV
Amps
1
5 10 15
25 30 50 75
100 150 200
Cat. Number
06700 06703 06704 06705
06707 06708 06709 06711
06713 06714 06715
Switchboard Shunts 50mV
Amps
100 150 200 250
300 400 500
Cat. Number
06500 06503 06504 06505
06506 06507 06508
Accessories
Portable
Switchboard
Ordering Information
SIMPSON ELECTRIC COMPANY 853 Dundee Avenue, Elgin IL 60120-3090 • (847) 697-2260 • FAX (847) 697-2272
www.simpsonelectric.com Printed in U.S.A. Part No. 6-116071, Edition 15, 03/05
NNoottee::
Special scaling is available from the factory at the time of ordering.