Simpson MINI-MAX M245 User Manual

Mini-Max M245 Series Digital Panel Meter
Minimum Depth Indicator - Less than 2.5” (60mm) of Space Required Behind the Panel
Stackable Mounting Bracket for Easy Installation
LCD: 4-1/2 Digit, 0.5” (12.7mm) High LCD Display with Optional Negative Image, Bright Red Backlighting
DC Process
LED: 4-1/2 Digit, 0.4” (10.2mm) High Display
Limited Range Display Scaling and Adjustable Oset
Standard Screw Terminals for Easy Installation
Four Ranges: 4-20mA, 1-5VDC, 0-10 VDC, 0-100VDC
85-250VAC or optional 9-32VDC Power Supply
Scaled 0-100%
Simpson’s Mini-Max Process Indicators provide high quality accuracy and reliability in a compact, 60mm deep case.
LCD (Liquid Crystal Display) Units oer a 4-1/2 digit, 0.5” (12.7mm) LCD display with an optional bright red, negative image, backlight.
LED (Light Emitting Diode) Units oer a 4-1/2 digit, 0.4” (10.2mm) display.
Installation and Panel Cutout
LCD
LED
All units feature user-selectable decimal point, auto zero and limited scaling capabilities.
A unique mounting bracket is provided to allow for vertical or horizontal stacking of multiple indicators. All Mini-Max units feature a 3/64 DIN, high-impact plastic case. The LCD units have a clear viewing window, and the LED units have a red viewing window.
2.68” (68mm)
2.84" (72mm)
0.95"
(24mm)
0.16"
(4mm)
0.88" (22.2mm)
2.68"
(68mm)
2.36" (60mm)
0.13" (3.18mm) Max Panel Thickness
0.874””
(22.2mm)
Mounting Requirements
Display Scaling
TOP HALF
OF CASE
FRONT
OF METER
Using a screwdriver or thumbnail, spread the tabs on each side of the case to unlock the top half. Lift the rear of the top half and slide it away from the front of
the meter.
Scale Adjustment:
Mini-Max indicators have limited range
scaling. There are no optional connections required for these to function. The meter can be scaled down to 1/2 the value of the input, or scaled up to 1 times the value of the input, or a maximum reading of 1.9999, whichever is lower.
LOCKING
TAB
LCD VERSIONS
Scale Adjustment:
The "Coarse" adjustment R12 will allow a limited range of scaling values. The "Fine" adjustment R9 allows for an adjust­ment range of approximately 1% of the "Coarse" adjustment. Apply the full scale input to the meter. Adjust R12 to be within 1% of the desired result. Then use R9 to
The “Coarse” adjustment R7 will allow
ment. The “Fine” adjustment R6 allows for an adjustment range of approximately 1% of the “Coarse” adjustment.
on the 4-20 mA scale). Adjust R7 to
LED VERSIONS
Scale Adustment:
The "Coarse" adjustment RV1 will allow a limited range of adjustment values. The "Fine" adjustment RV2 allows for an adjust­ment range of approximately 1% of the "Coarse" adjustment.
Apply the full scale input to the meter. Adjust RV1 to be within 1% of the desired
R12
SCALE ADJ. COARSE
SCALE ADJ. FINE
Note: Any physical damage to the meter
during adjustment will void the warranty.
OFFSET ADJ. COARSE
R9
R7
R6
OFFSET ADJ. FINE
desired result.
The “Coarse” adjustment RV3 will allow
adjustment. The “Fine” adjustment RV4 allows for an adjustment range of approximately 1% of the “Coarse”
signal (e.g. 4mA on the 4-20mA scale). Adjust RV3 to within 1% of the desired
Stacking Features
The mounting brackets, included with every Mini-Max, can be connected together. Multiple units can be mounted in a single opening, allowing perfect alignment.
To punch one hole for multiple units, be sure to adjust the standard panel
in the hole.
remaining units.
Vertical
Standard cutout
0.88" (22.2mm)
2.68" (68mm)
Vertical
OFFSET ADJ. COARSE
OFFSET ADJ. FINE
Note: Any physical damage to the meter during adjustment will void the warranty.
SCALE ADJ. COARSE
SCALE ADJ. FINE
0.071" (1.8mm) Add to standard when stacking
Horizontal
Standard cutout
0.88" (22.2 mm)
2.68" (68mm)
Horizontal
0.16" (4.0 mm)
Add to standard when stacking