Congratulations for purchasing a SIGMA laboratory centrifuge. You have selected a device
which combines many advantages.
A wide spectrum of programming options and an electronic operation control allow a troublefree use of the centrifuge. With its 3-phase drive, maintenance-free, quiet operation without
any carbon dust pollution is guaranteed.
Your device is equipped with user-friendly options which make the operation and standard
settings easier for you. Built-in error-detecting functions keep the user from entering incorrect
values and check the complete operation.
A special advantage is the storage capacity the centrifuge offers. The instrument has a
program memory which can store up to 50 data sets and is capable of keeping the last run
program in its memory for an unlimited amount of time allowing the program to be restarted
at any time - even if the centrifuge was turned off in between. All important operation
parameters can be seen at a glance.
The settings are executed via the knob in the control panel which has a coated surface
protecting the device against moisture and dust. In addition, the interior of the centrifuge is
also easy to clean. We are able to offer you a device that combines functional variety with
practical applications.
We thank you for your confidence and wish you a successful application of the centrifuge.
Die nachfolgend bezeichnete Maschine wurde in Ü bereinstimmung mit den Richtlinien
73/23/EWG; 89/336/EWG und 98/37/EWG hergestellt und geprüft.
The following machine is manufactured and tested in compliance with directions 73/23/CEE;
89/336/CEE and 98/37/CEE.
La machine désignée ci-dessous est produit et examiné conforme aux directives 73/23/CEE;
89/336/CEE et 98/37/CEE
Bezeichnung der Maschine: Laborzentrifuge
Machine: Laboratory Centrifuge
Désignation de la machine: Centrifugeuse de laboratoire
Maschinentyp : 4 - 15
Type:
Type de la machine:
Bestell Nr. : 10750, 10751, 10752, 10753, 10730, 10731, 10732
Part No.: 10733
Réf. usine:
Normen: EN 61010-2-020
Standards: EN 61000-3-2 ; EN 61000-3-3
Normes : EN 61326
Sigma Laborzentrifugen
An der Unteren Sö se 50
D-37520 Osterode
01.02.2002 Geschä ftsführer
Managing Director
Directeur Gérant
3.3.1 Fastening of angle rotors with hermetically sealed lid
3.4 Initial start-up
3.4.1 Switching on of the centrifuge
3.4.2 Opening lid
3.4.3 Installation of a rotor
Operating Manual SIGMA 4-15C, page 7 of 95
03/06
Page 8
Operating Elements
4.1 Operating panel
4.1.1 Start-key
4.1.2 Stop-key
4.1.3 Lid-key
4.1.4 Knob
4.2 Display
4.2.1 Set
4.2.2 Speed
4.2.3 Relative Centrifugal Force (RCF)
4.2.4 Time
4.2.5 Program
4.2.6 Rotor
4.2.7 Parameters
4.2.7.1 Acceleration
4.2.7.2 Deceleration
4.2.7.3 Radius
4.2.7.4 Density
4.2.7.5 Start delay
4.2.7.6 Automatic lid opening after end of run
4.2.8 Configuration
Operation
5.1 Selection, display and alteration of program parameters
5.1.1 Selection and alteration of the parameters and activation of the start
delay and the automatic lid opening after end of run
5.1.2 Selection and alteration of the rotor part number
5.1.3 Alteration of the parameter values during the centrifuge run
5.1.4 Alteration of the configuration
5.1.4.1 Curves
5.1.4.2 Creation of curves for variable accelerations and decelerations
5.1.4.3 Alteration of existing curves
5.1.5 Alteration of the contrast
5.1.6 Imbalance monitoring
5.1.7 Shortrun and faststop
Programming
6.1 Load, save and delete programs
6.1.1 Load a program
6.1.2 Save a program
6.1.3 Delete a program
Operating Manual SIGMA 4-15C, page 8 of 95
03/06
Page 9
Notes for Centrifugation
7.1 Practical notes for centrifugation
7.2 Forbidden centrifuging operations
Care and Maintenance
8.1 Care and cleaning of centrifuge
8.2 Care and cleaning of accessories
8.3 Rotor pins
8.4 Glass breakage
8.5 not applicable
8.6 Sterilization and disinfection of rotor chamber and accessories
8.6.1 Autoclaving
8.7 Checks by operator
Appendix
9.1 Slope of the specified curves, linear curves
9.2 Quadratic curves
9.3 Entry limitations
9.4 Mathematical relations
9.4.1 Relative Centrifugal Force (RCF)
9.4.2 Density
9.5 Error correction
9.5.1 Centrifuge cannot be started
9.5.2 Centrifuge decelerates during operation
9.5.3 Lid cannot be opened/closed
9.5.4 Emergency lid release
9.5.5 Problems with the centrifuge?
9.6 Error codes
9.7 Speed-RCF-diagram
9.8 Declaration of decontamination/Return declaration
9.9 Form program data
9.10 Leaflet
Operating Manual SIGMA 4-15C, page 9 of 95
03/06
Page 10
Page 11
1.1 Technical Data
Manufactuer: S I G M A
Laborzentrifugen GmbH
D-37520 Osterode
Germany
Type: 4-15C
Electr. connection:
Protection class:
Power consumption (kVA):
Rated power (kW):
Max. current (A):
Power data:
Max. speed (rpm):
Max. capacity (l):
Max. gravitational field (x g):
Max. kin. energy (Nm):
Further parameters
Time range:
Programs:
Acceleration curves:
Deceleration curves:
Radius:
Rotor part no.:
Dimensions:
Depth (mm):
Width (mm):
Height (mm):
Weight (kg):
EMC (acc. to EN 55011):
Noise level (dBA):
The figures are valid for an ambient temperature of 23 °C +/- 2 °C and nominal voltage +/- 5 %.
(Allowable ambient temperature +4 °C - +40 ° C; max. humidity 80 %.)
Subject to technical alterations.
max. allowed data with rotor 11150:
radius edge 16.6 cm,
radius max. 16.0 cm,
radius min. 11.3 cm,
max. plate height 54 mm
max. allowed data with rotor 11150:
radius edge 16.6 cm,
radius max. 16.0 cm,
radius min. 10.0 cm,
max. plate height 56 mm, incl. plate holder
17979
bottle 150 or 200 ml e.g. 15175, approx.
Ø 61.5 x 140 mm, incl. adapter 13175,
suitable for 11150, max. radius 19.0 cm,
min. radius 8.7 cm,
suitable for 13231, 17347
x 70 - 90 mm, e.g. 15060, suitable for 11150
Ø 45 x 100 mm, e.g. 15100, 15102, 15103,
15106, incl. rubber cushion 16051, suitable
for 11150, max. radius 17.8 cm, min. radius
7.8 cm
50 ml 15151, suitable for 11150, max. radius
19.0 cm, min. radius 9.3 cm
Max. speed
(rpm)
4 100
4 500
4 500
4 300
4 500
Max.
gravitational
field (x g)
3 120
3 007
2 124
3 758
3 622
2 264
4 302
3 680
4 302
Operating Manual SIGMA 4-15C, page 13 of 95
03/06
Page 14
Part No.
13236 Bucket, aluminium, for 12 culture tubes
13350 Round bucket, aluminium, with thread, incl.
17134 Aluminium sealing cap for 13350
17135 Polysulfone sealing cap, clear, for 13350
17350 Round carrier for 24 reaction vials 0.5-
17351 Round carrier for 12 Monovettes, max.
17352 Round carrier for 25 RIA-tubes 5 ml, max.
17353 Round carrier for 16 reaction vials 1.5-
17354 Round carrier for 16 glass tubes 7 ml, max.
17355 Round carrier for 12 tubes with screw cap
17356 Round carrier for 16 Vacutainer/hemolyse/
Description
15 ml 15115, suitable for 11150, max. radius
18.8 cm, min. radius 9.6 cm
O-ring 482840, for the 500 ml bottles 15500,
15501 and for the round carriers ∅ 85 mm
17350 up to 17402, sealable with caps
17134, 17135, suitable for 11150, max.
radius 19.0 cm, min. radius 8.7 cm
0.75 ml, ∅ 7.9/10 x 28/31 mm, e.g. 15005,
polypropylene
∅ 15.5/18 x 50 - 75 mm, polypropylene
∅ 12.5 x 65 - 80 mm, e.g. 15060,
polypropylene
2.2 ml, max. ∅ 11 mm, e.g. 15008, 15040,
polypropylene
∅ 12.5 x 85 - 115 mm, e.g. 15007, 15027,
polypropylene
10-12 ml, max. ∅ 16.2/19 x 65 x 90 mm, e.g.
13026 plus 17126, 15000, 15010, 15039,
polypropylene
RIA-tubes 5-6 ml, max. ∅ 13.5/17.5 x 70 -
90 mm, polypropylene
Max. speed
(rpm)
4 500
4 500
Max.
gravitational
field (x g)
4 246
4 302
Operating Manual SIGMA 4-15C, page 14 of 95
03/06
Page 15
Part No.
17358 Round carrier for 12 tubes 10-15 ml, max.
17359 Round carrier for 9 culture tubes 15 ml,
17360 Round carrier for 10 culture tubes 15 ml,
17362 Round carrier for 5 sterilin tubes 30 ml,
17370 Round carrier for 5 tubes 25-30 ml, max.
17375 Round carrier for 3 tubes 50 ml, max.
17376 Round carrier for 4 tubes with screw cap
17377 Round carrier for 4 culture tubes 50 ml
17378 Round carrier for 4 culture tubes 50 ml with
17385 Round carrier for 1 tube with screw cap
Description
∅ 17.2/19.5 x 90 - 115 mm, e.g. 15015,
15020, 15022, 15023, 15024 and Monovettes
9 ml and 10 ml, polypropylene
max. ∅ cap 23 mm, e.g. 15115, Greiner or
Corning tubes, polypropylene
max. ∅ cap 22 mm, e.g. 15115, Greiner or
Corning tubes, polypropylene
graduated up to 20 ml, with skirt, incl. cap,
max. ∅ 25/31 x 65 - 95 mm, polypropylene,
see www.bibby-sterilin.co.uk, no. 03008
∅ 25.4/29 x 85 - 115 mm, e.g. 15025, 15026,
15029, 15030, 15032, 15033, polypropylene
∅ 35/38 x 90 - 110 mm, e.g. 15049, 15050,
15056, polypropylene
40-50 ml, max. ∅ 29/34 x 85 - 110 mm, e.g.
13055 plus 17054, 15051, 15052, 15054,
polypropylene
15151, polypropylene
skirt, polypropylene
78-85 ml, max. ∅ 38/40 x 85 - 115 mm, e.g.
13085 plus 17185, 15074, 15075, 15076,
15080, polypropylene
Max. speed
(rpm)
Max.
gravitational
field (x g)
Operating Manual SIGMA 4-15C, page 15 of 95
03/06
Page 16
Part No.
17390 Round carrier incl. rubber cushion 16051
17395 Round carrier for 1 bottle with screw cap
17400 Round carrier for 1 bottle with screw cap
17401 Round carrier incl. rubber cushion 16250
17402 Round carrier for 1 tube with pointed
17347 Round carrier for 1 bottle with screw cap
13215 Rectangular bucket, aluminium, sealable
11156 Swing-out rotor for 6 buckets for the
11140 Swing-out rotor for 4 buckets for the
Description
for 1 tube 100 ml, max. ∅ 45/50 x 85 -
110 mm, e.g. 15100, 15102, 15103, 15106,
polypropylene
125 ml, max. ∅ 51 x 90 - 115 mm, e.g. 15125,
polypropylene
200 ml, max. ∅ 57 x 90 - 115 mm, e.g. 15202,
15203, polypropylene
for 1 round bottom tube 175-250 ml, max.
∅ 57.5 x 90 - 137 mm, e.g. 15201, 15206,
15250, 15251, 15254, polypropylene
bottom 150-250 ml, max. Ø 60 x 120 -
140 mm, polypropylene
250 ml, max. Ø 61.5 x 90 - 125 mm, e.g.
13255, 15247, 15248, 15249 and for 15175
with adapter 13175 (suitable for bucket
13350 without screw cap 17135),
polypropylene
with cap 17112, for the system of
rectangular carriers, max. tube length
115 mm, suitable for rotor 11150, max.
radius 18.2 cm, min. radius 8.8 cm
following accessories,
data with 13115: max. radius 18.2 cm, min.
radius 9.5 cm
data with 13127: max. radius 19.1 cm, min.
radius 9.7 cm
following accessories
Max. speed
(rpm)
4 500
4 500
4 500
5 000
Max.
gravitational
field (x g)
4 120
4 120
4 324
4 668/4 416
Operating Manual SIGMA 4-15C, page 16 of 95
03/06
Page 17
Part No.
13115 Rectangular bucket, aluminium, suitable for
13127 Rectangular bucket, aluminium, incl.
18000 Rectangular carrier, undrilled, for tubes 85 -
18002 Rectangular carrier for 20 reaction vials
18003 Upper part for 18002 for 20 reaction vials
18005 Rectangular carrier for 20 RIA-tubes 5 ml,
18007 Rectangular carrier for 20 glass tubes 7 ml,
18009 Rectangular carrier for 20 hemolyse tubes,
18010 Rectangular carrier for 12 tubes with screw
18012 Rectangular carrier for 12 Vacutainer-tubes
18015 Rectangular carrier for 12 tubes 10-15 ml,
Description
the system of rectangular carriers, max.
tube length 110 mm
polysulfone sealing cap 17112, suitable for
the system of rectangular carriers, max.
tube length 115 mm
110 mm, polyallomer
1.5-2.2 ml, max. ∅ 11 mm, e.g. 15008, 15040,
polypropylene
1.5-2.2 ml, max. ∅ 11 mm, e.g. 15008, 15040
(can be used together with 18002),
polypropylene
max. ∅ 12.2 x 60 - 75 mm, flat and round
bottom, e.g. 15060, polyallomer
max. ∅ 12.3 x 80 - 105 mm, flat and round
bottom, e.g. 15007, 15027, polyallomer
max. ∅ 12.8 x 70 - 90 mm, polypropylene
cap 10-12 ml, max. ∅ 16.8/17.5 x 60 - 85
mm, e.g. 13026 plus 17126, 15000, 15010,
15039, polyallomer
∅ 13.5/18 x 65 - 90 mm, polypropylene
max. ∅ 17 x 90 - 105 mm, e.g. 15015, 15020,
15022, 15023, 15024, polyallomer
Max. speed
(rpm)
5 000
5 000
Max.
gravitational
field (x g)
4 416
4 668
Operating Manual SIGMA 4-15C, page 17 of 95
03/06
Page 18
Part No.
18016 Rectangular carrier for 4 culture tubes
18017 Rectangular carrier for 10 tubes 15 ml and
18025 Rectangular carrier for 5 glass tubes 25 ml,
18022 Rectangular carrier for 4 sterilin tubes
18030 Rectangular carrier for 5 tubes with screw
18050 Rectangular carrier for 2 tubes 50 ml, max.
18051 Rectangular carrier for 2 tubes with screw
18052 Rectangular carrier for 2 culture tubes
18053 Rectangular carrier for 2 culture tubes
18085 Rectangular carrier for 1 tube with screw
Description
15 ml 15115, polypropylene
tubes with cap, max. ∅ 17.2/18 x 80 -
112 mm, e.g. Monovettes 9 ml and 10 ml,
polypropylene
max. ∅ 24 x 85 - 105 mm, e.g. 15025, 15026,
polyallomer
30 ml, graduated up to 20 ml, with skirt,
incl. cap, max. ∅ 25/31 x 65 - 95 mm,
polypropylene, see www.bibby-sterilin.co.uk,
no. 03008
cap 27-30 ml, max. ∅ 25.4/27.5 x 80 -
110 mm, e.g. 15029, 15030, 15032,
polypropylene
∅ 35/38 x 85 - 110 mm, e.g. 15049, 15050,
15056, polyallomer
cap 40-50 ml, max. ∅ 29/35 x 80 - 110 mm,
e.g. 13055 plus 17054, 15051, 15052, 15054,
polypropylene
50 ml 15151, polypropylene
50 ml with skirt, max. ∅ 29.5/38 x 85 -
118 mm, polypropylene
cap 78-85 ml, max. ∅ 38/40 x 85 - 112 mm,
e.g. 13085 plus 17185, 15074, 15075, 15076,
15080, polyallomer
Max. speed
(rpm)
Max.
gravitational
field (x g)
Operating Manual SIGMA 4-15C, page 18 of 95
03/06
Page 19
Part No.
18100 Rectangular carrier for 1 tube 100 ml, max.
18105 Rectangular carrier for 20 RIA-tubes 5 ml,
18107 Rectangular carrier for 20 glass tubes 7 ml,
18115 Rectangular carrier for 12 tubes 10-15 ml,
18125 Rectangular carrier for 1 bottle with screw
18200 Rectangular carrier for 1 bottle with screw
13201 Round bucket for 1 glass tube 200 ml
11141 Swing-out rotor 48 x 15 ml complete,
11157 Swing-out rotor 72 x 15 ml complete,
Description
∅ 45.5/48 x 85 - 110 mm, e.g. 15100, 15102,
15103, 15106, polyallomer
max. ∅ 12.2 x 60 - 75 mm, flat and round
bottom, e.g. 15060, decantable, polyallomer
max. ∅ 12.3 x 80 - 105 mm, flat and round
bottom, e.g. 15007, 15027, decantable,
polyallomer
max. ∅ 17 x 90 - 105 mm, e.g. 15015, 15020,
15022, 15023, 15024, decantable,
polyallomer
cap 125 ml, max. ∅ 51 x 90 - 115 mm, e.g.
15125, polypropylene
cap 200 ml, max. ∅ 57 x 90 - 115 mm, e.g.
15202, 15203, polypropylene
15201, 15206, incl. rubber cushion 16250,
sealable with aluminium cap 17121
consisting of rotor 11140, 4 rectangular
buckets 13115, 4 carriers 18015 and
48 polystyrene tubes 15020, max. radius
15.7 cm, min. radius 6.1 cm
consisting of rotor 11156, 6 rectangular
buckets 13115, 6 carriers 18015 and 72
polystyrene tubes 15020, max. radius
18.2 cm, min. radius 9.5 cm
Max. speed
(rpm)
5 000
4 500
Max.
gravitational
field (x g)
4 416
4 120
Operating Manual SIGMA 4-15C, page 19 of 95
03/06
Page 20
Part No.
11144 Swing-out rotor for microtiter plates, incl.
11118 Swing-out rotor for microtiter plates, incl.
17978 Plate holder for bucket 13218
12165 Angle rotor 6 x 78-85 ml for sealed tubes,
12166 Angle rotor 8 x 40-50 ml for sealed tubes,
12172 Angle rotor 12 x 27-30 ml for sealed tubes,
12168 Angle rotor 20 x 10-12 ml for sealed tubes,
12130 Angle rotor 30 x 1.5-2.2 ml for reaction
Description
2 buckets 13145 and plate holder 17977,
radius edge 12.85 cm,
radius max. 10.7 cm
radius min. 6.3 cm,
max. plate height 64 mm
2 buckets 13218 and plate holder 17978,
radius edge 15.15 cm,
radius max. 13.8 cm
radius min. 5.8 cm,
max. plate height 80 mm,
buckets sealable with cap 17108
incl. 4 clips 17118, for bucket 13218
e.g. 13085 plus 17185, 15074, 15075, 15076,
15080, incl. hermetic aluminium lid, max.
radius 9.8 cm, min. radius 2.7 cm, angle 25°
e.g. 13055 plus 17054, 15051, 15052, 15054,
incl. hermetic aluminium lid, max. radius
9.6 cm, min. radius 3.3 cm, angle 25°
e.g. 15029, 15030, 15032, incl. hermetic
aluminium lid, max. radius 11.0 cm, min.
radius 5.1 cm, angle 30°
e.g. 13026 plus 17126, 15000, 15010, 15039,
incl. hermetic aluminium lid, max. radius
9.8 cm, min. radius 5.8 cm, angle 25°
vials, e.g. 15008, 15040, incl. hermetic
aluminium lid, max. radius 10.0 cm, min.
radius 6.7 cm, angle 45°
Max. speed
(rpm)
5 000
5 500
12 500
12 500
12 500
12 500
13 500
Max.
gravitational
field (x g)
3 592
2 991
1 761
5 124
4 667
1 961
17 119
16 770
19 216
17 119
20 376
Operating Manual SIGMA 4-15C, page 20 of 95
03/06
Page 21
Part No.
11148 Swing-out rotor 32 x 1.5-2.2 ml, incl. 8
12200 Angle rotor 100 x 1.5-2.2 ml for reaction
12169 Angle rotor for 8 culture tubes 50 ml, e.g.
13060 Adapter for 1 culture tube 15 ml 15115,
12170 Angle rotor for 12 culture tubes 15 ml, e.g.
12256 Angle rotor for 6 bottles 250 ml, e.g. 13255,
Adapters, tubes and steel tubes
13000 Adapter , POM, for reaction vials 0.25-
13002 Adapter, POM, for reaction vials 0.5-0.75 ml,
13021 Adapter for PCR-tube 0.2 ml, Ø 5.85/6.95 x
Description
buckets, for reaction vials, e.g. 15008,
15040, incl. hermetic aluminium lid, max.
radius 8.4 cm, min. radius 4.4 cm
vials, e.g. 15008, 15040, incl. aluminium lid
17824, 2 lines, max. radius 16.3/15.18 cm,
min. radius 13.11/11.9 cm
15151, incl. hermetic aluminium lid, max.
radius 10.4 cm, min. radius 4.3 cm,
angle 25°
suitable for 12169, polypropylene
15115, incl. hermetic aluminium lid, max.
radius 10.0 cm, min. radius 4.6 cm,
angle 25°
15247, 15249, incl. aluminium lid 17801,
max. radius 14.5 cm, min. radius 3.9 cm,
angle 30°
0.4 ml 15014, suitable for 11148, 12130,
12200, 17353, 18002, 18003
15005, ∅ 7.9/10 x 28/31 mm suitable for
11148, 12130, 12200, 17353, 18002, 18003
20/23.4 mm, suitable for 11148, 12130,
12200, 17353, 18002, 18003
Max. speed
(rpm)
12 000
9 000
12 500
12 500
8 000
Max.
gravitational
field (x g)
13 523
14 761/13 747
18 168
17 469
10 375
Operating Manual SIGMA 4-15C, page 21 of 95
03/06
Page 22
Part No.
13079 Bottom adapter for 1 tube 40-42 ml 15051,
13080 Adapter for 1 culture tube 50 ml 15151,
13081 Adapter for 1 culture tube 15 ml 15115,
13082
13083
13084
15005
15008 Reaction vials 1.5 ml, 1 pack contains
15040 Reaction vials 2.2 ml, 1 pack contains
15014 Reaction vials 0.4 ml (Beckman system),
15060
13026
Description
15052, 15054, suitable for 12169,
polypropylene
suitable for 12165, polypropylene
suitable for 12165, polypropylene
Adapter for 1 tube 40-50 ml, max. ∅ 28.8 x
105 - 110 mm, e.g. 13055, 15051, 15052,
15054, suitable for 12165, polypropylene
Adapter for 1 tube 27-30 ml, max. ∅ 25.5 x
90 - 110 mm, e.g. 15029, 15030, 15032,
suitable for 12165, polypropylene
Adapter for 2 tubes 10-12 ml, max. ∅
16/17.5 x 75 x 90 mm, e.g. 15000, 15010,
15039, suitable for 12165, polypropylene
Reaction vials 0.5 ml, ∅ 7.9/10 x 28/31 mm,
1 pack contains 100 pcs., suitable for
13002, 17350
100 pcs., suitable for 11148, 12130, 12200,
17353, 18002, 18003
100 pcs., suitable for 11148, 12130, 12200,
17353, 18002, 18003
polypropylene, 1 pack contains 100 pcs.,
suitable for 13000
Polystyrene tube 5 ml (RIA-tube), ∅ 12 x
75 mm, suitable for 13233, 17352, 17356,
18005, 18009, 18105
Stainless steel tube 10 ml, ∅ 15.7 x 76 mm,
closeable with cap 17126, suitable for
12168, 17355, 18010
15027 ditto, graduated 0 - 5.5 ml in steps of 0.1 ml
15015
15024 ditto, graduated 0 - 10 ml in steps of 0.1 ml
Description
200 ml, Ø 61.5 x 139 mm, graduated up to
175 ml (N-3143-0175), suitable for 13231
with 13175, 17347 with 13175
190 ml, ∅ 56 x 112 mm, suitable for 17400,
18200
200 ml, ∅ 56 x 113 mm, suitable for 17400,
18200
cap 17256, Ø 61.4 x 125 mm, suitable for
12256, 17347
x 122 mm, suitable for 12256, 17347
ml, Ø 61.4 x 125 mm, suitable for 12256,
17347
cap, Ø 85 x 135 mm, suitable for 13350
Glass tube 7 ml, ∅ 12 x 100 mm, suitable
for 17354, 18007, 18107
Glass tube 10 - 12 ml, ∅ 16 x 100 mm,
suitable for 17358, 18015, 18115
Operating Manual SIGMA 4-15C, page 25 of 95
03/06
Page 26
Part No.
15022
15025
15026 ditto, graduated 5 - 25 ml in steps of 1 ml
15033
15050
15056 ditto, graduated 4 - 50 ml in steps of 1 ml
15100
15106 ditto, graduated 1 - 100 ml in steps of 1 ml
15201 Glass tube 200 ml, Ø 56 x 112 mm, suitable
15206 ditto, graduated 15 - 175 ml in steps of 5 ml
15250 Glass tube 230 ml, Ø 56 x 135 mm, suitable
15254 ditto, graduated 15 - 200 ml in steps of 5 ml
Further accessories
16018 Rubber adapter for 15 ml glass tube 15022,
16019 Rubber adapter for 15 ml glass tube 15022,
16030 Rubber adapter for 30 ml glass tube 15033,
Description
Special glass tube 15 ml, ∅ 17 x 110 mm,
suitable for 12166 with 16018 and for 12165,
with 16019, max. permitted speed 7 000 rpm
Glass tube 25 ml, ∅ 24 x 100 mm, suitable
for 17365, 18025
Special glass tube 30 ml, ∅ 24 x 105 mm,
suitable for 12166 with 16030 and for 12165
with 16031, max. permitted speed 7 000 rpm
Glass tube 50 ml, ∅ 34 x 100 ml, suitable for
17375, 18050
Glass tube 100 ml, ∅ 44 x 100 mm, suitable
for 17390, 18100
for 13201, 17401
for 17401
suitable for 12166
suitable for 12165
suitable for 12166
Operating Manual SIGMA 4-15C, page 26 of 95
03/06
Page 27
Part No.
16031 Rubber adapter for 30 ml glass tube 15033,
16906 Rubber plate for 18025
16907 ditto, only for 18050
16908
16909 Rubber plate for rectangular carriers
16250 Rubber cushion for 13201
17112 Rectangular polysulfone sealing cap, clear,
17118 Sealing clip for caps 17108, 17112
17801 Lid for rotor 12256
17919 Table for centrifuge made of varnished
17913 Fasteners for table 17919
Further accessories are available on request.
Description
suitable for 12165
ditto, only for 18100
incl. 2 clips 17118, for 13215 and spare for
13127
sheet steel with 2 doors and space for
accessories, movable on lock-type castors,
dimensions: w 490, d 635, h 490 mm
Maximum speed for tubes
Some tubes, e.g. centrifuges glass tubes, microtubes, culture tubes, Teflon tubes and
especially high volume tubes can be used in our rotors, buckets and adapters at
higher speeds than their breaking limit. We recommend to always fill up the tubes and
to follow the recommendations of the manufacturer.
Operating Manual SIGMA 4-15C, page 27 of 95
03/06
Page 28
Operating Manual SIGMA 4-15C, page 28 of 95
03/06
Page 29
1.3 Scope of Supply
The following belongs to the centrifuge:
1 Rotor wrench SW 13 Part No. 930 102
1 Square wrench “Emergency lid release“Part No. 80 054
1 Tube grease for rotor trunnion pins Part No. 70 284
20 ml slushing oil Part No. 70 104
Documentation:
1 Short operating instructions
1 Operating Manual
1 "Rotor and Accessories, Operation and Use"
1 EU-Statement of Conformity
1 Equipment Decontamination Certificate
Accessories according to your order, our order confirmation and our delivery
note.
Rotor Part No. Rotor No.
....................... .....................
....................... .....................
....................... .....................
....................... .....................
....................... .....................
....................... .....................
....................... .....................
1.4 Standards and Regulations
Please refer to the enclosed EU-Statement of Conformity.
Operating Manual SIGMA 4-15C, page 29 of 95
03/06
Page 30
1.5 Safety Instructions
regarding operation of centrifuges with rotors of different max. speed, e.g. angle
rotors and swing-out rotors.
According to the German trade association regulation BGR500 chapter 2.11
part 3 the owner of the instrument is advised to take care of the following points:
1. According to BGR500 the owner has to provide operating instructions based
on those of the manufacturer and to inform the employees accordingly.
2. For safety reasons these operating instructions must clearly state that the
max. speed engraved on the rotor and/or the bucket and the max. allowable
filling quantity must not be exceeded.
3. If the density of the material exceeds 1.2 g/cm3, the max. speed of the centrifuge must be reduced.
4. Operation of the centrifuge in hazardous locations is not allowed.
5. During operation the centrifuge must not be moved. Leaning against or resting
on the centrifuge is not allowed.
6. Do not spin explosive or highly inflammable materials.
7. Substances which could damage the material of the centrifuge, the rotors or
the buckets anyhow must not be centrifuged or only under consideration of
special safety measures. Infectious, toxic, pathogene or radioactive substances must be centrifuged in certified rotors only.
8. Keep a clearance of at least 30 cm around the centrifuge. Dangerous materials of any kind must not be put down or stored in that area.
9. Attention!
Keep your hand away from the danger zone when closing the centrifuge lid.
Risk of bruising!
10.Attention!
Defective lid relieving devices could cause the centrifuge lid to fall down
(contact Service). Risk of bruising!
Operating Manual SIGMA 4-15C, page 30 of 95
03/06
Page 31
1.6 Symbol Table
International symbols used for the centrifuge:
Symbol Title
Gefä hrliche elektrische Spannung
Achtung, Bedienungsanleitung beachten
Ein (Netzverbindung)
I
Aus (Netzverbindung)
O
Schutzleiteranschluß
Erde
Netzstecker ziehen
Vorsicht Quetschgefahr
Drehrichtungspfeil
→
Heiße Oberflä che
Dangerous voltage
Courant haute tension
Attention, consult accompanying documents
Attention, consulter les documents joints
On (Power)
Marche (mise sous tension)
Off (Power)
Arrê t (mise hors tension)
Protective earth (ground)
Liaison à la terre
Earth (ground)
Terre
Unplug mains plug
Tirer la fiche de prise
Caution! Risk of bruising
Attention! Danger de blessure
Arrow direction of rotation
Flèche sens de rotation
Hot surface
Surface chaude
Operating Manual SIGMA 4-15C, page 31 of 95
03/06
Page 32
2.1 General Outlay
The new generation of SIGMA laboratory centrifuges is equipped with two
microprocessors which guarantee independently the control of the rotor
recognition and the overspeed signal. A further optimization and increase of the
instrument’s safety could be maintained. The long-life asynchronous motor is
silent and brushless. The problem of carbon brush change is no longer existent,
and as there is no carbon dust pollution, operation in clean rooms is possible if
the appropriate accessories are used.
2.2 Construction and Constructive Safety Measures
The centrifuge is built into a sheet steel housing. The armoured chamber, the
sheet steel lid, the motorized lid lock device and the hinge system are providing
optimum safety. At the back the lid is secured by solid hinges and at the front
twice by a motorized lid lock. Due to these elements there is a solid safety case
around the rotor chamber.
The centrifuge stands on elastic feet.
2.3 Drive
The drive motor is a well dimensioned asynchronous motor.
2.4 Operation and Display
The grafical LCD display is hermetically sealed. A single knob only allows any
data input. The backlit display indicates any operating status and guides the
operator through the wide range of applications.
Option:
A connection for a serial interface is possible so that an external personal computer with printer can be connected for control or recording.
Operating Manual SIGMA 4-15C, page 32 of 95
03/06
Page 33
2.5 Electronic Control
The electronics controlled by two microprocessors allows extensive adaptations
of the centrifuge to the different tasks. The following parameters can be
programmed and displayed among others:
− Speed (by activation of FINE steps of 1 or 10 rpm possible)
− RCF in steps of 1 or 10 x g
− Time preselection (9 h, 59 min max.), in steps of 1 min or 1 sec
− Continuous operation
− Short-time operation
− Fixed deceleration and acceleration curves
− Free creation of deceleration and acceleration curves
− Saving, recalling and alteration of programs
− Input and measurement of the time integral
− Start delay
− Continuous self-monitoring and recognition of errors which are displayed and
saved for service
2.6 Safety Devices
Apart from the passive safety devices due to the instrument's mechanical design
there are the following active precautions for your safety:
2.6.1 Lid Lock, Cover Closing Device
The centrifuge can only be started when the power switch is switched to "ON“
and when the lid is correctly closed. After closing the lid the motorized lid locks
are automatically locked. Attention! Please do not leave your fingers between lid and upper edge of the centrifuge housing when closing the lid. The lid
can only be opened when the rotor has completely stopped. If the lid is opened
by the emergency release during operation, the centrifuge will immediately switch
off and decelerate brakeless up to standstill of the rotor. If the lid is open, the
drive is completely separated from the mains supply, that means starting of the
centrifuge is impossible (refer to chapter 9.5.4 "Emergency lid release").
2.6.2 Imbalance Monitoring System
In the event that uneven loading leads to imbalance, the drive is switched off and
an imbalance warning message is displayed.
Operating Manual SIGMA 4-15C, page 33 of 95
03/06
Page 34
2.6.3 Rotor Monitoring
During programming the rotor part no. and if required the bucket part no. must be
entered. Two microprocessors check, if the entered speed or the gravitational
field is allowed for the rotor. Input errors are impossible (refer to point 9.3 "Entry
limitations"). After starting, during the start-up phase, the computer additionally
checks the identification of the rotor.
Attention: Please take care to enter the correct bucket part no. as the max.
allowable speed could be exceeded by entering an incorrect bucket. This is
not allowed.
If the rotor doesn`t correspond to the programmed rotor no., STOP is carried out
and an error message is displayed. Restarting the centrifuge is only possible
after reset and when the correct rotor number has been selected.
2.6.4 Standstill Monitoring
Opening of the centrifuge lid is only possible, if the rotor is at standstill. This
standstill is checked by the computer and also by an additional hardware circuit.
2.6.5 System Check
An internal system check monitors data transmission and the sensor signals with
regard to plausibility. In the event of an error, malfunctions are recognized with
utmost sensitivity, displayed as error message together with an error number,
and saved for service.
2.6.6 Ground Wire Check
For ground wire check there is a ground screw at the rear panel of the centrifuge.
A ground wire check can be carried out using an appropriate measuring
instrument.
Operating Manual SIGMA 4-15C, page 34 of 95
03/06
Page 35
3.1 Unpacking of the Centrifuge
Open case. Take out the box containing accessories. Remove packaging
material. Lift centrifuge upwards with a lifting device or with several persons.
When lifting or carrying the centrifuge please always reach under the instrument
from the side.
Attention: The instrument is heavy!
Please keep case for possible transport of centrifuge later.
3.1.1 Transport Safety Device
The SIGMA 4-15C has a transport safety device which must be removed before
start-up. The transport safety device blocks the drive system.
The transport safety device screws are accessible from the outer underside.
Lift the left or front side of the centrifuge to put a suitable object – for example a
block of wood- between the table and the bottom of the centrifuge.
Attention: Danger of injuries!
The star grip transport safety device screws become apparent. Unscrew by hand
and totally remove.
After that, line centrifuge up and continue start-up.
The transport safety device screws should be stored for possible despatch
(service, repairs).
Operating Manual SIGMA 4-15C, page 35 of 95
03/06
Page 36
3.2 Installation
3.2.1 Site
All energy consumed by the centrifuge is converted into heat and emitted into the
ambient air. Therefore, sufficient ventilation is important. The air-ducts in the unit
must be open. Also, the centrifuge shouldn`t be positioned near radiators and
should not be directly exposed to sunshine.
A clearance of at least 30 cm around the centrifuge is necessary.
For normal operation the ambient temperature should not fall below 10 °C and
not exceed 35 °C. The max. humidity of air is 80 %. During transport from cold to
warmer places water will condensate inside the centrifuge. It is important that
there is enough time for drying before the centrifuge can be started again.
3.2.2 Connection/Fuse
The operating voltage on the name plate must correspond to the local supply
voltage!
SIGMA laboratory centrifuges are units of safety class I, DIN VDE 0700, and include a three wire power cord 2,5 m long with shockproof right angle plug. The
instrument has thermal fuses. In case of a disconnection through the thermal
fuses, allow a cool-down phase of two minutes, after which they could be
reactivated by a switch.
3.2.3 Fuses / Emergency Circuit Breaker on Site
The centrifuges must be protected typically with at least 16 A slow acting fuses.
An emergency circuit breaker to cut the power to the centrifuge in the event of a
malfunction is required on site. This switch should be located away from the
centrifuge, preferably outside the room where the centrifuge is used or at the exit
of this room.
Operating Manual SIGMA 4-15C, page 36 of 95
03/06
Page 37
3.3 Installation of Rotors and Accessories
1. Open centrifuge lid by pressing Lid-key.
2. Unscrew rotor tie-down screw from motor shaft (anticlockwise).
3. Lower the rotor straight down onto the motor shaft.
4. Tighten the tie-down screw (clockwise) with the rotor wrench so that the disc
In the event of frequent use the tie-down screw must be loosened by some
turns and fastened again. This should be done once a day or after approx.
20 cycles (please refer to chapter 5.1.4 “Alteration of the Configuration“
– Cycles). This ensures a proper connection between rotor and shaft (please
refer to chapter 8.2 "Care and cleaning of accessories" as well).
5. Fill all positions of swing-out
rotors with buckets. Make
sure that all buckets are
inserted correctly.
6. Use only appropriate vessels
for the rotor (please refer to
chapter 1.2 "Suitable accessories" as well).
7. Fill vessels external to the centrifuge.
8. Put or screw on covers of vessels.
9. Opposite places of the rotors must always be loaded with same accessories
and same filling.
10. Attention when using microtiter rotors: It is important not to run the plate
holders without plates inserted.
11. In angle rotors the plastic vessels must always be totally filled to avoid cracks
of vessels and leakages or loosening of the caps in case of partial filling.
Attention, follow the special comments of chapter 1.5.
allowednot allowed
Universal grease
Operating Manual SIGMA 4-15C, page 37 of 95
03/06
Page 38
12. Attention: The centrifuge will absorb smaller differences in weight when
loading the rotors. But it is recommended to balance the vessels as
accurately as possible in order to ensure a run with minimal vibrations.
Should the centrifuge be operated with very uneven load, the imbalance
monitoring will switch off the drive. An imbalance warning would be displayed.
13. Rotors with lid should always be run with their lid. The rotor lid is tightened by
hand. Correct fastening must be ensured. Attention: The lid screw serves
for fastening of the lid onto the rotor only, not for fastening of the rotor
onto the drive! Before installation of the lid, the correct fastening of the rotor
fixing screw must always be checked using a wrench.
3.3.1 Fastening of Angle Rotors with Hermetically Sealed Lid
(Please refer to drawing on next page.)
1. Screw rotor lid (2) onto rotor and tighten it.
2. Lower rotor with lid (2) onto motor shaft.
3. Put rotor tie-down screw (1) onto motor shaft and tighten using the wrench.
4. The rotor can be run without lid (2) as well.
5. The rotor and lid seals (3) must be greased after cleaning.
6. Special instructions for the use of hermetically sealed rotors:
All rotors can be installed or removed with closed lid after loosening the rotor
tie-down screw. All rotors are autoclavable (refer to chapter 8.6 "Sterilization
and disinfection of rotor chamber and accessories").
To increase life of rotors and seals the rotors must be cleaned with slushing
oil and the seals and thread areas with vaseline or grease after cleaning.
Attention!
Please follow the special comments of chapter 1.5.
Operating Manual SIGMA 4-15C, page 38 of 95
03/06
Page 39
Operating Manual SIGMA 4-15C, page 39 of 95
03/06
Page 40
3.4 Initial Start-Up
Attention!
Before initial start-up please take care that your centrifuge is orderly installed
(refer to chapter 3.2 "Installation").
3.4.1 Switching on of the Centrifuge
Press mains switch (at the back of the centrifuge). When power is applied the
first time, default values will be displayed:
− The command panel illuminates.
− The speed display indicates "2000".
− The time display indicates "2“.
− The program display indicates "--", i.e.
- the centrifuge will accelerate to 2000 rpm,
- the centrifuge operation is terminated after 2 minutes,
- no program number has been allocated so far.
3.4.2 Opening Lid
Press Lid-key
− The lid opens.
3.4.3 Installation of a Rotor
Put a rotor onto the shaft and fasten it by screwing the rotor tie-down screw
clockwise onto the drive shaft. Please use the supplied rotor wrench (refer to
chapter 3.3 "Installation of rotor and accessories"). Please pay attention to the
fact that during tightening the disc spring of the rotor tie-down screw is pressed
together and the screw is tightened.
Operating Manual SIGMA 4-15C, page 40 of 95
03/06
Page 41
4.1 Operating Panel
Operating panel
The centrifuge can be operated via the operating panel. Keys can be pressed
when their LED is on.
Operating Manual SIGMA 4-15C, page 41 of 95
03/06
Page 42
4.1.1 Start-key
This key can be used for the following:
• starting centrifuge operation,
• terminating a previously started deceleration process and restarting
centrifuge,
• shifting into short-run at preselected speed. Pressing Start-key continuously
for longer than one second leads to acceleration to the maximum speed with
the maximum acceleration curve and after release deceleration with the
maximum deceleration curve.
The centrifuge can be started when
• the lid is closed
• the Start-key is illuminated.
4.1.2 Stop-key
This key can be used
• to early terminate a run: The centrifuge decelerates with the preset curve to a
complete stop. Deceleration can be terminated by pressing the Start-key
again.
• to carry out a faststop: Push Start-key for longer than one second. The
centrifuge decelerates with the maximum deceleration curve.
4.1.3 Lid-key
This key is used to open the lid. This can only be executed if
• the centrifuge has come to a complete stop
• the Lid-key is illuminated.
4.1.4 Knob
This knob can be used for selecting and alterating parameters and figures/
numbers.
Operating Manual SIGMA 4-15C, page 42 of 95
03/06
Page 43
4.2 Display
Default values
4.2.1 Set
If this area is illuminated, you are in the change mode which you can exit by
pressing the knob (here in combination with the speed).
4.2.2 Speed
Speed
In the upper section of the area the set speed of the centrifuge is displayed.
Underneath is the actual speed. The maximum speed values are rotor
dependent.
4.2.3 Relative Centrifugal Force (RCF)
RCF
The relative centrifugal force is the acceleration which the sample is exposed to.
The set value of this parameter is in the upper section of this area, underneath is
the actual value. The maximal RCF-values are rotor dependent.
(Refer to chapter 9.4 "Mathematical relations")
Operating Manual SIGMA 4-15C, page 43 of 95
03/06
Page 44
4.2.4 Time
Time
The set run time is displayed in the upper section of this area, underneath the
remaining run time is displayed. Time is defined as the period from the start of
the centrifuge to the beginning of deceleration, maximum value is 9 h 59 min.
The set time is underlined (here: 2 minutes).
The set value is indicated in hours, minutes, and seconds. The actual value has
the same units as the set value and is displayed in hours : minutes or in minutes :
seconds if the set value is below 10 minutes.
h:m
m:s
If the highest possible time of 9 h 59 min is exceeded or below the minium
adjustable time range, continuous operation is activated. The word "HOLD" is
displayed instead of the set value. After the start of a continuous run, the elapsed
time is displayed instead of the remaining run time. By entering a specific run
time the continuous mode is deactivated. It can be terminated by pressing the
Stop-key as well.
Continuous run
Operating Manual SIGMA 4-15C, page 44 of 95
03/06
Page 45
4.2.5 Program (PROG)
Program
In this area the number of the actual program is displayed. If the program is not
saved yet, "--" is displayed.
By activating this area you are able to load stored programs without calling the
selection list.
The program used last is automatically loaded after restarting.
Program Selection List ( )
Program selection list
The area marked with arrows for the program selection list presents the view of
the programs already saved. Storage space for fifty programs - No. 1 – 50 - from
which you can select and load a program is at your disposal. The actual loaded
program is indicated by "--".
Operating Manual SIGMA 4-15C, page 45 of 95
03/06
Page 46
4.2.6 Rotor (here: 11150/13215)
Rotor
In this area the actually selected rotor or a rotor/bucket combination are
displayed.
Rotor Selection List ( )
Rotor selection list
The area marked with arrows for the rotor selection list offers all available rotors
which can be selected via the knob.
4.2.7 Parameters (PARA)
Parameters
If this area has been selected, parameters can be changed and the start delay
and the automatic lid opening after end of run can be activated or deactivated.
Operating Manual SIGMA 4-15C, page 46 of 95
03/06
Page 47
4.2.7.1 Acceleration
Acceleration
The acceleration number selects an acceleration curve which the centrifuge will
follow. Two different versions are available:
0 - 9 linear
10 - 19 quadratic up to 1000 rpm, then linear
Their shape is further explained in chapter 9.1 "Slope of the specified curves,
linear curves" and 9.2 "Quadratic curves".
4.2.7.2 Deceleration
Deceleration
The deceleration number selects a deceleration curve that decelerates the
centrifuge down to standstill. The deceleration curves are inverted images of the
acceleration and are labelled with identical numbers. Curve no. 0 represents
brakeless deceleration.
4.2.7.3 Radius
Radius
The set radius will determine the displayed RCF-value. If the radius is not
changed, the max. RCF-value will be displayed.
Operating Manual SIGMA 4-15C, page 47 of 95
03/06
Page 48
4.2.7.4 Density
Density
If the density of a sample exceeds 1.2 g/cm3, the maximum final speed will
reduce (refer to chapter 9.4.2 “Density”). A value between 1.2 and 9.9 g/cm3 is
possible.
4.2.7.5 Start Delay
With this function a start delay can be set.
4.2.7.6 Automatic Lid Opening after End of Run
If this parameter has been set the lid will automatically open after standstill of the
rotor.
Operating Manual SIGMA 4-15C, page 48 of 95
03/06
Page 49
4.2.8 Configuration
Configuration
Selection of configuration (CONFIG) opens a menu where Code, Language,
Screen, Fine, Buzzer, Sensor, Info, Reset and the creation of freely
programmable curves can be chosen. Furthermore, the cycles and run times of
the individual rotors can be read.
Operating Manual SIGMA 4-15C, page 49 of 95
03/06
Page 50
5.1 Selection, Display and Alteration of Program Parameters
The value of each area can be changed as follows:
• This display shows the default values. No area is inverted.
Default values
• One area can be activated by pressing the knob once. It is then inverted.
Being in the selection mode now, you can select other areas by turning the
knob.
• Activate a selected area by pressing the knob. You are now in the alteration
mode, "SET" and the selected area are inverted.
• Select the new value of the selected area by turning the knob.
• Confirm the entry by pressing the knob. You then leave the alteration mode.
"SET" and the selected area are deactivated. The alteration mode is left
automatically after 20 seconds. The areas are no longer inverted, the actual
values are taken over.
• Now, you can select other areas by turning the knob or you can operate the
start key. Then, the centrifuge starts with the selected values.
Notice:
All entry limits are automatically checked. When reaching a limit, the counting
operation is stopped (refer to chapter 9.3 "Entry limitations").
Operating Manual SIGMA 4-15C, page 50 of 95
03/06
Page 51
5.1.1 Selection and Alteration of the Parameters and Activation of the
Start Delay and the Automatic Lid Opening after End of Run
Selection and alteration of the parameters
• Select the area "Parameters" (PARA) by turning the knob (selection mode)
and confirm by pressing the knob.
• The parameters to be changed, e.g. acceleration, deceleration, radius, density
and the start delay and automatic lid opening after end of run can be selected
(here: acceleration).
• Confirm the selected parameter by pressing the knob (here: acceleration). The
parameter is activated. Select the desired acceleration curve by turning the
knob.
Deceleration, radius and density are changed in the same way.
Operating Manual SIGMA 4-15C, page 51 of 95
03/06
Page 52
Activation of the start delay and/or the automatic lid opening after end of run
• Select ”START DELAY” by turning the knob.
• Activate the “START DELAY” by pressing the knob.
• Select time delay (here: 10 seconds) by turning and confirm by pressing the
knob.
• Change the set time by turning the knob and deactivate the start delay by
pressing the knob.
The automatic lid opening after end of run is activated or deactivated in the same
way.
Operating Manual SIGMA 4-15C, page 52 of 95
03/06
Page 53
5.1.2 Selection and Alteration of the Rotor Part Number
• Select the area of the rotor selection list ( ) by turning the knob.
• After pressing the knob, you get a view over all available rotor types.
Rotor selection list
• Select the number of the actual rotor/bucket combination by turning the knob
and confirm the entry by pressing the knob.
Attention: Please take care that the correct rotor/bucket combination is
entered as otherwise the max. allowable speed could be exceeded. This
is not allowed.
Operating Manual SIGMA 4-15C, page 53 of 95
03/06
Page 54
5.1.3 Alteration of Program Parameters during the Centrifuge Run
During the run, the following values can be altered.
• Speed
• RCF
• Run time
• Switching into the continuous run mode/time mode
• Acceleration curve
• Deceleration curve
• Start delay
Alter the Set values as usual by activating the Alteration mode (refer to chapter
5.1 "Selection, Display and Alteration of Program Parameters").
You cannot change:
• Rotor
• Program
• Radius
• Density
The following functions can be activated/deactivated:
• Start delay
• Automatic lid opening after end of run
Operating Manual SIGMA 4-15C, page 54 of 95
03/06
Page 55
5.1.4 Alteration of the Configuration
In the configuration mode, several background functions can be changedand
data can be read.
• Select the configuration mode by turning the knob.
• After pressing the knob, the configuration menu appears.
• Select by turning the knob the desired area and activate the function by
pressing the knob.
• CODE:Protection of specific functionsby a code, deactivation of the protection and
changing of the code
• LANGUAGE: Selection of the language of the operation
• SCREEN: Magnification of the speed or the RCF display
• FINE: In this function it is possible to preselect the set speed in steps of 1 or 10
min-1. The set time can be preselected in steps of 1 min or 1 sec. Exeption:
Curve input in 1/10 sek.
• CYCLES: For each bucket/rotor combination cycles and run time are stored. The data
of the used rotors are displayed.
• BUZZER: An acoustic signal can be activated for a preselected time after termination
of a run or in the event of an imbalance or an error message. In the event
of "IMBALANCE" or "ERROR" the acoustic signal is already activated.
• SENSOR: The sensor menu displays different signals. In case of a failure of the
instrument this makes the diagnosis of errors and their fast repair through
the service easier. Values can neither be entered nor altered.
Operating Manual SIGMA 4-15C, page 55 of 95
03/06
Page 56
• INFO: The Info menu displays information like the type of centrifuge and the
EPROM version, the number of cycles, the total run time and the software
version and date. In case of a failure of the instrument this helps to find
reasons for the fault. Values can neither be entered nor altered.
• RESET: The "Reset"-function offers the possibility to delete all programs,
parameters and configurations to get the original settings again.
• CURVES: Creation of user defined accelerations and decelerations.
The procedure of alteration of the basic functions shall be explained by an example
(Alteration of the screen).
• You can choose between three display versions:
- RCF and speed at normal size
- Speed magnified -Zoom- (no RCF display)
- RCF magnified -Zoom- (no speed display)
• Select the configuration mode by turning the knob.
• After pressing the knob the configuration menu appears.
• Select "SCREEN" by turning the knob.
Operating Manual SIGMA 4-15C, page 56 of 95
03/06
Page 57
• By pressing the knob you get into the screen menu. The screen marked with a
cross is activated. If you want to select the screen "SPEED - (ZOOM)" you
proceed as follows:
• Select "SPEED - (ZOOM)" by turning the knob.
• Activate the Speed-Zoom-screen by pressing the knob.
• Select "EXIT" by turning the knob. Confirm it by pressing the knob. You get
into the configuration menu again.
Operating Manual SIGMA 4-15C, page 57 of 95
03/06
Page 58
• By pressing the knob you return to the main display. The areas "SPEED" and
"TIME" are displayed magnified. The area "RCF" disappears.
Operating Manual SIGMA 4-15C, page 58 of 95
03/06
Page 59
5.1.4.1 Curves
With the “curve function” user defined accelerations and decelerations can be
created. Curve numbers 20 – 29 are available. The intervals no. 1 – 10 of each
curve consist of fix points which are defined by time and speed. In the event that
the speed of one interval is higher than the set speed for the run, the curve speed
can be limited or allowed or starting can be prohibited.
CUT ⇒
Speed in one interval is automatically limited to the set speed for
the run.
ADMIT ⇒
CANCEL ⇒
Chosen speed in one interval is allowed.
A stop occurs. Starting is not possible without to a change of the
curve speed in one interval.
For interval no. 1 a linear (LIN) or quadratic (QUAD) acceleration can be chosen.
All further accelerations are linear.
Furthermore, run profiles can be created. The following conditions must be
fulfilled:
- The max. preset final speed of one interval corresponds to the set speed of
the run.
- The total time corresponds to the set time of the run.
- The final speed of the last interval is 0.
Operating Manual SIGMA 4-15C, page 59 of 95
03/06
Page 60
5.1.4.2 Creation of Curves for Variable Accelerations and
Decelerations
• Select the configuration mode by turning the knob.
• After pressing the knob the configuration menu appears.
• Select "CURVES" by turning the knob.
• After pressing the knob the curve menue appears.
Operating Manual SIGMA 4-15C, page 60 of 95
03/06
Page 61
• Select “0:00:00” by turning the knob. You are now creating interval no. 1.
Activate “0:00:00” by pressing the knob.
• Set the time by turning the knob.
• Save the time value in interval no. 1 by pressing the knob.
• Select “SPEED” by turning the knob.
Operating Manual SIGMA 4-15C, page 61 of 95
03/06
Page 62
• Activate speed value “0” of interval no. 1 by pressing the knob.
• Change the speed value by turning the knob. The created curve with the max.
curve speed appears.
• After pressing and turning the knob the time value of interval no. 2 appears.
• Activate “0:00:00” of the second interval by pressing the knob.
• Set the time by turning the knob. The extended curve appears.
Operating Manual SIGMA 4-15C, page 62 of 95
03/06
Page 63
• Save the time value no. 2 by pressing the knob.
• Select “SPEED” by turning the knob.
• Activate the speed value “0” by pressing the knob.
• Change the speed value by turning the knob.
• After pressing and turning the knob the time value of interval no. 3 appears.
Operating Manual SIGMA 4-15C, page 63 of 95
03/06
Page 64
• Activate “0:00:00” of the third interval by pressing the knob.
• Set the time by turning the knob. The extended curve appears.
• Save the time value no. 3 by pressing the knob.
Operating Manual SIGMA 4-15C, page 64 of 95
03/06
Page 65
• Select “SPEED” by turning the knob.
• Activate “0” by pressing the knob.
• Change the speed value by turning the knob. The curve appears and can be
programmed as acceleration or deceleration curve under no. 20.
• Select “EXIT” by pressing and turning the knob and leave the curve mode by
pressing the knob again.
• The created curve no. 20 is saved and can be recalled when programming
the parameters.
Operating Manual SIGMA 4-15C, page 65 of 95
03/06
Page 66
5.1.4.3 Alteration of Existing Curves
After activation of the area “CURVES” in the configuration menu you can
analogous to the procedure discribed in 5.1.4.2 select already existing curves
and alter them.
5.1.5 Alteration of the Contrast
If you press the knob for longer than a second, a dialogue window appears. Now
you can select the contrast by turning the knob. By pressing the knob again, you
confirm the entry, the new contrast remains.
5.1.6 Imbalance Monitoring
An imbalance dialogue window indicates an excessive imbalance during
operation.
Run cannot be continued (imbalance > cut-off limit). Run is terminated with max.
deceleration.
Reason:
Improper loading or malfunction during operation (e.g. glass breakage) resulting
in an uneven run.
Note!
Additional information and a detailed description of errors and their correction is
given in chapter 9.5 “Error correction“.
Operating Manual SIGMA 4-15C, page 66 of 95
03/06
Page 67
5.1.7 Shortrun and Faststop
• Shortrun
By pressing the Start-key continuously the shortrun function is activated. The
instrument accelerates to the maximum speed with the maximum acceleration
curve and after release decelerates with the maximum deceleration curve until
standstill.
• Faststop
Pressing the Stop-key during operation for longer than a second leads to
maximum deceleration until standstill.
Operating Manual SIGMA 4-15C, page 67 of 95
03/06
Page 68
6.1 Load, Save and Delete Programs
What is considered a program?
A program contains all data that are required for a centrifuge run.
The advantage is that special sedimentation results can be repeated under equal
conditions without a change of data caused by entry errors.
Programs can be loaded, operated, altered, and deleted any time.
All programs can be protected against unauthorized use by a personal Code.
6.1.1 Load a Program
There are two possibilities of loading a program:
1. Loading by program number
• Select the program area (“PROG --“) by turning the knob, and activate it by
pressing the knob.
Activated program area
• By turning the knob, all programs saved and the actual program ("--") appear
one after the other.
Changed program
• Load desired program by pressing the knob.
Operating Manual SIGMA 4-15C, page 68 of 95
03/06
Page 69
2. Loading from the list of programs
• Select the program selection list ( ) by turning the knob and activate this area
by pressing the knob. You are able to see all programs in the memory, "--"
indicates the actual program.
Program selection list
• Select the program you want to load by turning the knob. After pressing the
knob a dialogue window appears. Select the instruction "LOAD" and confirm it
by pressing the knob.
Dialogue window
Operating Manual SIGMA 4-15C, page 69 of 95
03/06
Page 70
6.1.2 Save a Program
If you want to save an actual program:
• Select the program selection ( ) list by turning the knob and activate it by
pressing the knob to get a view over all programs in the memory and the free
positions in the list.
Program selection list
• Select the desired program position in the list by turning the knob.
• After pressing the knob a dialogue window appears. Select "SAVE" and
confirm it by pressing the knob.
Dialogue window
Operating Manual SIGMA 4-15C, page 70 of 95
03/06
Page 71
6.1.3 Delete a Program
If you want to delete a program from the list:
• Select the program selection list ( ) by turning the knob and activate this area
by pressing the knob to see the list of programs.
Program selection list
• Select the program you want to delete by turning the knob.
• A dialogue window appears after pressing the knob. Select "DELETE" and
confirm it by pressing the knob.
Dialogue window
Operating Manual SIGMA 4-15C, page 71 of 95
03/06
Page 72
7.1 Practical Notes for Centrifugation
1. Locate centrifuge horizontally on a level surface.
2. Ensure safe location.
3. Keep at least 30 cm free space around the centrifuge.
4. Provide for sufficient ventilation.
5. Tighten rotor firmly onto motor shaft.
6. Avoid imbalance.
7. Load opposite buckets with same
accessories.
8. Centrifugation with low capacity:
The vessels should be placed symmetrically so that the buckets and
their suspensions are loaded evenly.
Loading an inner or outer position of
the bucket only is not allowed (see
illustration). Even angle rotors should
be loaded symmetrically to same
weight.
Centrifugation with different tubes:
Working with different tube sizes is
possible. Opposite places must be
loaded with the same vessels (see
illustration).
9. Load all positions of swing-out
rotors.
10. Load vessels outside the centrifuge.
11. Please pay attention to the max. speed of
glass tubes. At speeds over 4 000 rpm there is an increased breaking glass
hazard.
12. Fill vessels carefully to same weight. Imbalances would result in increased
wear of bearings.
13. Grease joints of buckets and rotor pins in swing-out rotors.
14. Use perfect accessories only.
15. Avoid corrosion to accessories by careful maintenance.
Operating Manual SIGMA 4-15C, page 72 of 95
03/06
Page 73
16. Spin infectious material in sealed rotors and buckets only.
17. Do not spin explosive or highly imflammable materials.
18. Record all program data, refer to forms of appendix chapter 9.8.
19. When centrifuging substances with a density > 1,2 g/cm3 the allowable max.
speed must be reduced (refer to chapter 9.4.2 "Density").
Operating Manual SIGMA 4-15C, page 73 of 95
03/06
Page 74
7.2 Forbidden Centrifuging Operations
1. Operation of not carefully installed centrifuge.
2. Operation without front or back panels.
3. Operation by non authorized personnel.
4. Operation with rotor not installed properly (refer to chapter 3.3).
5. Operation with incompletely loaded drum rotor, swing-out rotor or angle rotor
with interchangeable buckets.
A rotor must always be loaded completely, empty places are not allowed!
Opposite buckets or carriers may nevertheless be empty. Mixed loading is
allowed, if opposite places are loaded with same buckets and carriers of
same weight.
6. Operation with overloaded rotors.
he load for a rotor is limited by its design and the max. speed (see
rotor/bucket engraving) and must not be exceeded. The rotors are intended
for liquids of max. homogeneous density of 1.2 g/cm
speed. If liquids of higher density are used, the speed must be reduced
accordingly (refer to chapter 9.4 "Mathematical relations").
7. Operation with rotors, buckets and carriers showing corrosion or other
defects.
8. Operation of very corrosive substances which can cause damages to material
and affect the mechanical strength of rotors, buckets and carriers.
9. Operation of rotors and accessories not allowed by the manufacturer. The
use of poor commodity goods is not recommended. At high speeds breaking
glass or bursting vessels can cause dangerous imbalances.
10. Operation in hazardous locations.
11. Operation with vessels of improper size.
12. Centrifugation of improper material.
13. Operation with partially filled plastic tubes in high-speed angle rotors.
14. Lifting or moving of the centrifuge during operation. Leaning against or resting
on the centrifuge is not allowed.
15. Do not place potential dangerous material - eg. glass vessels containing
liquids - near the centrifuge.
3
if centrifuged at max.
Operating Manual SIGMA 4-15C, page 74 of 95
03/06
Page 75
16. Attention: Do not open cover and/or reach into rotor chamber unless the rotor
is at standstill. Never attempt to override the lid interlock system while the
rotor is spinning.
17. Such materials are prohibited which chemically interact vigorously.
18. Do not spin explosive or inflammable materials.
19. Substances which could damage the material of the centrifuge, the rotors or
the buckets must not be centrifuged. Infectious, toxic, pathogene or
radioactive substances must be centrifuged in certified rotors and vessels
only and all necessary safety precautions are taken.
Operating Manual SIGMA 4-15C, page 75 of 95
03/06
Page 76
8.1 Care and Cleaning of the Centrifuge
Please use water-soluble, mild detergents for cleaning. Avoid corroding and
aggressive substances. Do not use alkaline solutions or solvents or agents with
abrasive particles. Before using detergents or decontamination agents which had
not been recommended by us, the user has to contact us to make sure that such
procedure would not damage the centrifuge.
Remove product particles from the rotor chamber using a cloth or paper towl. It is
recommended to open the cover when the centrifuge is not in use so that
moisture can evaporate. Increased wear of the motor bearings will thus be
avoided. If there is the risk of toxic, radioactive or pathogene
contamination, special safety measures must be kept.
Operating Manual SIGMA 4-15C, page 76 of 95
03/06
Page 77
8.2 Care and Cleaning of Accessories
For care of accessories special safety measures must be considered as these
are measures ensuring operational safety at the same time.
Buckets, trunnions and also synthetic buckets are produced exactly in order to
withstand the permanent high stress with high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating pressure and chemical reaction) can affect or destroy the metals. Hardly detectable
cracks on the surface expand and weaken the material without visible signs.
When detecting a visible damage of the surface, a crack, a mark or any other
change, also corrosion, the part (rotor, bucket etc.) must be replaced immediately.
In order to avoid corrosion, rotor incl. tie-down screw and cover seal, buckets and
carriers must be cleaned and greased regularly with the supplied slushing oil
(SIGMA part no.: 70104 for 20 ml slushing oil). Before using detergents or
decontamination agents which had not been recommended by us, the user has
to contact us to make sure that such procedure would not damage the centrifuge.
The rotor tie-down screw must be greased using the supplied grease (SIGMA
part no.: 70284).
Cleaning of accessories should be done outside of the centrifuge once a week or
preferably after every use. The rubber cushions should be removed from buckets
and carriers. After this the parts should be dried with a soft cloth or, alternatively,
in a drying chamber at approx. 50 °C. If there is the risk of toxic, radioactive
or pathogene contamination, special safety measures must be kept.
Especially aluminium parts are extremely corrosive. A neutral cleaning detergent
with a pH-value between 6 and 8 should be used for such parts. Alkaline agents
exceeding pH 8 must be avoided. Especially aluminium parts must be greased
regularly with slushing oil. This procedure essentially increases life time and
reduces corrosion.
Careful maintenance increases life time and avoids premature failure of the rotor.
Corrosion or resultant damages which are caused by insufficient care do not
constitute a warranty claim.
8.3 Rotor Pins
The trunnion pins of the rotor should always be greased as only this ensures
evenly swinging of buckets and thus quiet run of the centrifuge (part no. 70284
Grease).
Operating Manual SIGMA 4-15C, page 77 of 95
03/06
Page 78
8.4 Glass Breakage
In case of glass breakage all glass particles must be carefully removed. Rubber
inserts have to be cleaned carefully and possibly be replaced. If a problem has
occured, the following has to be considered:
Glass particles in the rubber cushion will cause glass breakage again.
Particles on the rotor pins prevent buckets and carriers from swinging evenly
which will cause an imbalance.
Glass particles in the centrifuge chamber will cause metal abrasion due to the
strong air circulation. This dust will not only pollute the centrifuge chamber, the
rotor, the buckets, the carriers and the material to be centrifuged but also
damage the surfaces of the accessories, the rotors and the centrifuge chamber.
In order to totally remove the glass particles and the metal dust from the rotor
chamber, it is advisable to grease the upper part of the centrifuge chamber with
e.g. Vaseline. Then the rotor should rotate for some minutes at a moderate
speed. The glass and metal particles will now collect at the greased part and can
easily be removed with a cloth together with the grease. If necessary repeat this
procedure.
8.5 not applicable
8.6 Sterilization and Disinfection of Rotor Chamber and
Accessories
All usual disinfectants like e.g. Sagrotan, Buraton or Terralin (to obtain at
chemist’s shops) can be used. The centrifuges and the accessories consist of
different materials. A possible incompatibility must be considered. Before using
detergents or decontamination agents which had not been recommended by us,
the user has to contact us to make sure that such procedure would not damage
the centrifuge. For sterilization by steam resistance to temperature of the
individual material must be checked (refer to point 8.6.1 “Autoclaving“). Please
contact your laboratory safety officer regarding proper methods to use. If
dangerous materials are used, the centrifuge and the accessories must be
disinfected.
Principally we want to point out that for centrifuging of e.g. infectious material
certified and hermetically sealed accessories have to be used in order to avoid
that the centrifuge is contaminated.
Operating Manual SIGMA 4-15C, page 78 of 95
03/06
Page 79
8.6.1 Autoclaving
The life of the accessories essentially depends on the frequency of autoclaving
and use. When the parts are showing changes in colour or structure or in the
event of leaks etc., the accessories have to be replaced.
During autoclaving the caps of the tubes must not be screwed on to avoid
deformation of the tubes. It can not be excluded that plastic parts, e.g. lids or
carriers, would deform during autoclaving.
The operator has to ensure that no important part of the centrifuge is damaged.
This especially refers to:
1. Motor suspension
2. Concentricity of the motor shaft
3. Fastening of the trunnions in the rotor
4. Rotors and accessories have a limited life. For safety reasons a regular
check is recommended after 50.000 cycles. Any changes like e.g. corrosion,
cracks, material abrasion etc. require special attention.
5. Screw connections have to be tight.
Furthermore, the earth wire must be checked regularly.
Operating Manual SIGMA 4-15C, page 80 of 95
03/06
Page 81
9.1 Slope of Specified Curves, Linear Curves
− The slope of a specified acceleration curve is defined as the time required to
accelerate the rotor to 1000 rpm.
− With linear and quadratic slopes, curves are numbered in the direction of
increasing acceleration (from right to left).
− The deceleration curves are inverted images of the acceleration curves with
the same numbers (exception: curve 0).
− Curve 0 decelerates brakeless.
Comment: The acceleration time depends on the moment of inertia of the rotor.
Linear Curves (0 - 9)
Figure 2
Operating Manual SIGMA 4-15C, page 81 of 95
03/06
Page 82
The curve 9 is a special case compared with the other curves. The centrifuge
accelerates with max. torque. The acceleration time only depends on the moment of inertia of the rotor.
− The deceleration curves are inverted images of the acceleration curves with
the same numbers.
Quadratic Curves (10 - 19)
1.000
900
800
700
600
500
400
300
200
100
17
19
18
0
0
20
141516
40
13
60
80
100
120
140
160
180
200
101112
220
240
Figure 3
The curve 19 is a special case compared with the other curves. The centrifuge
accelerates with maximum torque. The acceleration time only depends on the
moment of inertia of the rotor.
Quadratic curve no. Time up to 1000 rpm Slope from 1000 rpm linear
10 240 120 sec./1000 rpm.
All entry limits are automatically checked. When reaching a limit the operation is
stopped.
Interdependent parameters are limited further with a change of the corresponding
parameter. With the alteration of a parameter, the corresponding parameter is
recalculated. If the changed parameter is selected, the recalculated value is
displayed.
9.4 Mathematical Relations
9.4.1 Relative Centrifugal Force (RCF)
The parameters speed, RCF and the diameter of the rotor are interrelated via the
following formula:
RCF = 11,18 x 10-6 x r x n2
If two values are given, the third value is determined by the equation. If the speed
or the rotation radius is changed, the resulting RCF will be recalculated. If the
RCF is altered, the speed under consideration of the radius is adapted
accordingly.
r = radius in cm
n = speed in rpm
RCF without dimension
9.4.2 Density
If the density of the liquid is higher than 1.2 g/cm3, the allowed maximum speed
of the centrifuge is calculated according to the following formula:
n = n
Rho = density in g/cm3
x )/2,1(Rho
max
Operating Manual SIGMA 4-15C, page 84 of 95
03/06
Page 85
Power cord plugged in and line
Imbalance dialogue window is
9.5 Error Correction
Most of the errors can be reset by power off/on. In the event of a short power
failure during a run, this run is interrupted and can be continued by pressing the
Start-key.
No indication on
the command panel:
− No power in the socket?
−
voltage present?
− Input fuse ok?
− Power switch on?
− Contrast?
Actions:
• Check fuse in mains supply.
• Plug in power cord correctly.
• Switch on thermal fuse.
• Switch on power.
• Change contrast.
9.5.1 Centrifuge cannot be Started
a) Start-key LED not illuminated:
b) Start-key LED illuminated:
• Close lid. Motorized lid locks must close.
• Power off/on. If error occurs again, call
service.
9.5.2 Centrifuge Decelerates during Operation
active:
− Rotor is loaded unevenly.
− Centrifuge tilted.
− Drive error (mechanical
damage).
− Centrifuge was moved during
operation.
− Centrifuge displays an error 73
to 77 after power on.
• Balance loading.
• Align centrifuge.
• Call Service.
• Restart after opening and closing lid.
• These error numbers indicate an error in
the internal program storage. This error is
e.g. generated when a power failure
occurs during a storage procedure.
Operating Manual SIGMA 4-15C, page 85 of 95
03/06
Page 86
9.5.3 Lid cannot be Opened/Closed
When first trying to open the lid the locks are not released. Press the Lid-key to
activate the motorized lid locks again (refer to chapter 9.6 “Error codes“, ERROR
79). Call Service.
9.5.4 Emergency Lid Release
In the event of a power supply failure it is possible to manually open the lid.
Remove stopper at the upper right side front panel using e.g. a screw driver.
Unlock the motorized lid lock using the supplied square wrench. Put in wrench
and turn to the right. The lid is unlocked and can be opened.
Attention!
The lid may only be unlocked and opened when the rotor is at standstill.
9.5.5 Problems with the Centrifuge?
Please contact your supplier for support or in the event of malfunctions and for
supply of spare parts.
Operating Manual SIGMA 4-15C, page 86 of 95
03/06
Page 87
9.6 Error Codes
Error no.Kind of errorMeasuresComment
1-62 Internal error
69-77 EEPROM error
78 Lid does not close
79 Faulty log identification
80 Lid electronics defective
81 Rotor spins with open cover
82-83 Lid does not open
84 Excess temperature heat
sink
85 Excess temperature centri-
fuge chamber
(> 50 °C)
87 Actual temperature > 45 ° C
resp. < -20 °C
90-96 Sensor defective
98 Faulty rotor identification
99 Incorrect rotor
• let slow down
• power off/on
• let slow down
• power off/on
• power off/on
• remove impurities
• press lid key
• close lid rapidly
• check log opening for
impurities
• contact Service
• let slow down
• close lid
• power off/on
• power off
• emergency lid release
• let cool down
• provide for better ventilation
• power off/on
• let cool down
• provide for better ventilation
• power off/on
• provide for better ventilation
• power off/on
• check rotor
• check rotor and set rotor
resp. bucket identification
ATTENTION: Saved
curves and programs
could have been
deleted.
Should it not be possible to repair the failure, please contact Service.
9.7 Speed-RCF-Diagram
An additional help is the enclosed Speed-RCF-Diagram.
9.8 Declaration of Decontamination / Return Declaration
The following declarations serve for keeping safety and health of our employees.
Fill in the forms and attach them when returning centrifuges, accessories and
spare parts. Please understand that we cannot carry out any work before we
have the declarations. (We recommend to make several copies of this page.)
Declaration of Decontamination of Centrifuges, Accessories and Spare Parts
This declaration may only be filled in and signed by authorised staff.
Repair Order dtd. :
Order No. :
Type of unit : Serial No. :
Type of unit : Serial No. :
Type of unit : Serial No. :
Type of unit : Serial No. :
Accessories :
Is the equipment free from harmful substances ? YES ΟNO Ο
If not, which substances have come into contact with the equipment?
Name of the substances : _________________________________________________________________
_________________________________________________________________
Remarks (e.g to be touched _________________________________________________________________
with gloves only) ; _________________________________________________________________
General characteristics of the substances :
Corrosive ΟExplosive Ο
Biologically hazardous ΟRadioactive Ο
Toxic Ο
In combination with which substances may hazardous mixtures develop?
Name of the substances :
Has the equipment been cleaned before shipment? YES ΟNO Ο
Is the equipment decontaminated and not harmful to health?YES ΟNO Ο
Prior to repair, radioactively contaminated components must be decontaminated according to the valid
regulations for radiation protection.
Legally Binding Declaration
I / we hereby declare that the information on this declaration are correct and complete.
Company / Institute :
Street :
Postcode, City :
Tel. : FAX :
Name :