Sierra Monitor Corporation 2050-00, 2050-10, 2060-00, 2060-10 User Manual

Page 1
GAS MONITORS
MODELS 2050-00 & 2050-10 MODELS 2060-00 & 2060-10
Page 2
Applicability & Effectivity
Sierra Monitor Corporation
1991 T arob Court, Milpit as, CA 95035
(408) 262-6611 (800) 727-4377
GAS MONITORS
MODELS 2050-00 & 2050-10 MODELS 2060-00 & 2060-10
APPLICABILITY & EFFECTIVITY
This manual provides instructions for the following Sierra Monitor products:
Model Description 2050-00 Ammonia Gas Monitor 2060-00 Halocarbon Gas Monitor
The instructions are effective for the above models as of January 1, 2004
Instruction Manual Part Number: T10013
Rev C
Applicability & Effectivity
Page 3
T ABLE OF CONTENTS
SECTION PAGE
1.PRODUCT DESCRIPTION 1
1.1 Introduction 1
1.2 Application 1
1.3 Configuration 1
1.4 T emperature Range 2
1.5 Sensor 2
1.6 Remote Alarm 2
2. QUICK ST ART 3
2.1 Overview 3
2.2 Wiring 3
3. OPERA TION 4
3.1 Introduction 4
3.2 Alarms 4
3.3 Outputs 4
4. ALARM CALIBRATION 5
4.1 Factory Alarm Calibration 5
4.2 Frequency of Alarm Calibration Check 5
4.3 Alarm Calibration Check Process 5
4.4 Equipment Required 6
4.5 Alarm Calibration Procedure 6
5. SERVICE
5.1 Sensor Replacement 7
6. INST ALLATION 8
6.1 Gas Sensor Module Locations 8
6.2 Mounting 8
6.3 Wiring 8
6.4 Power Supply 8
6.5 Alarm Devices 8
7. ALARM CONFIGURATION 10
8. SPECIFICA TIONS 11
9. ACCESSORIES AND REPLACEMENT P ARTS 11
10. LIMITED WARRANTY 11
Table of Contents
Page 4
Instruction Manual
1.0 PRODUCT DESCRIPTION
1.1 Introduction
The Model 2050 Ammonia and Model 2060 Refrigerant/ Halocarbon Gas Monitors are continuous duty , low cost stationary monitors for detection of high levels (50-500 PPM) of ammonia and halocarbon ("freons") refrigerant gases. See the specifications page for a full list of avail­able applicable halocarbons
This manual provides instructions for both the 2050 and 2060 series Gas Monitors, including the relay versions.
1.2 Application
These monitors are primarily used in and around freez­ers, refrigerators, and compressor rooms to warn of leaks and dangerous levels of these gases.
1.3 Configuration
An alarm is activated when the concentration of gas exceeds the user-adjustable level. An alarm condition activates the alarm light on the front of the monitor and generates an output signal. On Models 2050-00 and the 2060-00 the output signal a TTL (Transistor-Transis­tor-Logic signal) can activate a remote alarm through one of Sierra Monitor’s Alarm Panels. (The Model 2102 Dual Channel Alarm Panel can be connected up to 500 feet from the monitors. ) The models can be config­ured with a 0.5 amp, dry contact relay.
Model Gas TTL Relay
2050-00 Ammonia X X 2060-00 Halocarbons X X
Table 1.1 Gas Monitor Configurations
factory setpoint for R12 is 500 ppm while the setpoint for R134 a is 50 ppm. See the specifications for more details.
The Model 2050 / 2060 is cross sensitive to other com­bustible gases. If combustible gasses in excess of 50 ppm are expected in the area please contact Sierra Moni­tor to determine amount of cross-sensitivity expected for the specific combustible gas. It is not recommended to use this unit in areas with combustible gases present.
The monitors are housed in a polystyrene enclosure that is suit­able for installation in freezer, re­frigerator or other applications. Test connections and adjustment controls are located on the out­side of the enclosure and are eas­ily used, even when the operator is wearing hand protection.
The Model 2050 Ammonia moni­tor has a factory alarm setpoint of 100 ppm NH3. This setpoint is user adjustible in the field over a range of 50 to 500 ppm NH3. The Model 2050 calibration can be checked with Hydrogen gas.
The Model 2060 Halocarbon monitor can be calibrated with carbon monoxide and the factory setpoint varies for the specific type of halocarbon present. The
Page:
1
Figure 1.1 Outline View - 2050/2060 Series
2050/2060 Gas Monitors (5/04)
Page 5
1.4 Temperature Range
The Model 2050 and 2060 have temperature compen­sated circuitry to provide stabilization in temperatures ranging from -4o to 158oF.
However, for optimum operation the unit should be cali­brated at the normal temperature expected for the sensor location.
1.5 Semiconductor-T ype Sensor
A solid-state semiconductor-type sensor and associ­ated electronic circuitry ensure trouble-free, long-term operation. All the electronic circuitry needed to operate the monitor, except the DC input power , is contained in a compact unit. There are no pumps, filters, or chemical cells to replace or maintain. Except for periodic calibra­tion to verify the alarm setting, no attention is required after installation. The user may adjust the alarm level by applying a different calibration gas concentration.
Instruction Manual
A sensor self-check feature will flash the alarms on and off alternatively should the sensor fail (open circuit). The alarm flash includes LEDs, audible alarm, relay or alarm signal as applicable.
1.6. Remote Alarm A vailable
If a remote alarm is desired, Sierra Monitor’s Alarm Pan­els, the Model 2102 Dual Channel Alarm Panels can be connected up to 500 feet away from the monitors.
2050/2060 Gas Monitors (5/04)
Page:
2
Page 6
Instruction Manual
2.0 QUICK ST ART
2.1 Overview
The Gas Sensor Module has been supplied factory calibrated and ready for immediate installation and operation. An inst aller familiar with installation and operation of gas detection products can use this section to begin immediate use of the monitor.
2.2 Wiring
Each module requires four-conductor wiring (two wires for power and two wires for the signal).
2.2.1. Model 2050-2060 Series
The Model 2050-2060 series are designed to mount on any indoor vertical surface. Mount the monitor in the desired location.
- The installation must meet any hazard­ous environmental codes for AC/DC elec­trical instrumentation.
2.2.2 Wiring Connection
T erminal positions on the electronics board are as follows:
During warm-up the monitor will, first, cycle through safe/alarm/safe condition at one hertz. This will be followed by a short period of continu­ous alarm before warm-up is completed.
2.2.4 Configuration
The default configuration for each module is to operate with a buzzer and a normally operating open (NOO) relay . The user can change this con­figuration using the jumpers provided. Refer to table 6.2.
Terminal Function
TB1-1 + VDC (9 – 24) TB1-2 GND (0VDC) TB2-1 Relay NC (Normally Closed) or
Output to Model 2102 Safe TB2-2 Relay Comm on TB2-3 Relay NO (Normally Open) or
Output to Model 2102 Alarm
Table 2.1
2.2.3 Start-up & Operation
To begin operation of the Gas Sensor Module provide 9 - 24 VDC from a regulated power supply such as one of the Sierra Monitor Alarm Panels, 2102-XX. Each time the sensor mod­ule is powered up it will perform a warm-up for 2 - 60 minutes.
Page:
3
2050/2060 Gas Monitors (5/04)
Page 7
3.0 OPERA TION
Instruction Manual
3.1 Introduction
Under normal conditions the gas sensor module does not require operator or technician intervention. The following are conditions under which the module re­quires attention:
- Routine periodic calibration
- Sensor replacement on a planned schedule or when a sensor failure occurs.
- Periodic cleaning as necessary .
3.2 Alarms
Three alarm conditions are possible. These can be detected visually at the optional Sierra Monitor con­troller and at the module.
Warm-Up Alarm
Oscillating (On/Off) contact closure and “red/ green” visual indication when power is first con­nected to the module, followed by a continuous closure of approximately 2-60 minutes indicat­ing warm-up.
Buzzer Relay Relay
Fail Safe
On/Off Oscilating Oscilating
Gas Alarm
Trouble Alarm
Interrupted contact closure and “red/green” vi­sual indication (once the warm-up time is com­pleted) after unit has been in operation, indicat­ing either a failed sensor or calibration problems.
Buzzer Relay Relay
Fail Safe
On/Off Oscilating Oscilating
3.3 Outputs
For the non-relay configurations a 5 volt DC TTL signal is provided to activate remote alarms such as the Sierra Monitor Alarm Panels. Relay configu­ration alarm conditions described above are trans­mitted to the annunciator as relay contact closures from the monitor.
The alarm outputs are non-latching closed contacts. As soon as the gas concentration falls below the alarm set-point, the alarm will stop and the alarm output will cause the relay to de-energize and the contacts will open.
During warm-up and failed sensor alarms the re­lays contact and the display LED's oscillate at ap­proximately one hertz.
Sustained contact closure and solid red light (once the warm-up time is completed) indicat­ing the presence of gas at, or above, the pre-set alarm limit. T able 2.1
Buzzer Relay Relay
On Closed Open
2050/2060 Gas Monitors (5/04)
=
Fail Safe
Page:
4
Page 8
Instruction Manual
4.0 ALARM CALIBRATION
4.1 Factory Alarm Calibration
The module has been factory calibrated to alarm as indicated in Table 4.1 and as marked on the cali­bration label shipped with the module.
Model Time Gas cc/min 2050-XX 100 sec. 100 ppm NH350 2060-XX 60 sec. 500 ppm R11 50
500 ppm R12 50 50 ppm R22 5 0 100 ppm R113 5 0 70 ppm R123 50 50 ppm R134a 50 40 ppm R141b 5 0 40 ppm R142b 5 0 40 ppm R500 50 70 ppm R502 50
T able 4.1
4.2 Frequency of Alarm Calibration Check
The manufacturer recommends that the calibration
of each gas sensor module be verified monthly dur­ing the first three months of operation and then quar­terly. More frequent checks are necessary during periods of extreme humidity and temperature changes. The monitor should have operated con­tinuously (uninterrupted) for at least 24 hours prior to calibration adjustment.
4.3 Alarm Calibration Check Process
The output signal of the gas sensor module is calibrated using a calibration gas mixture containing a known concentration of the gas of interest and a balance of air, or using a substitute gas. The concentration of the span gas must be equal to the alarm point desired.
The equivalent substitute gas is 80 ppm Hydrogen (H2) for the Model 2050 and 150 ppm Carbon Monoxide (CO) for the Model 2060.
Calibration requires application of the span gas to the sensor and adjustment of the sensitivity adjustment po­tentiometer.
Warning: During calibration the alarm will turn on and remote alarms connected to the alarm relays will be activated. Disable the remote alarm if necessary .
Page:
Fig. 4-1 Location of Calibration Voltmeter Connections and Adjustments
5
2050/2060 Gas Monitors (5/04)
Page 9
4.4 Equipment Required
The following tools and equipment will be required for calibration:
- Jewelers Screwdriver
- Calibration Gas
- Calibration gas delivery system
For accurate calibration use a gas mixture at the re­quired concentration mixed in an air balance, rather than with an inert gas like nitrogen. This gas and the required delivery equipment such as the Model 1200­26 Calibrator is available from Sierra Monitor Corpo­ration.
4.5 Alarm Calibration Procedure
1 . The monitor should be in the safe condition prior
to calibration (green LED “ON”).
- Be sure that the area is non-hazardous be­fore proceeding.
Instruction Manual
2 . With the monitor in operation, expose the sensor
to a sample of calibration gas. Apply calibration gas directly to the sensor insuring that the delivery cup loosely covers the sensor. Use 80 PPM H2 for the Model 2050 and 150 PPM CO for the Model
2060.
3. The red LED on the face of the monitor should turn "on". If the alarm occurs within 18-20 seconds and stops within one minute of the removal of the cali­bration gas, the monitor is in calibration and re­quires no further adjustment.
4 . If the monitor fails to alarm within 90 seconds of
the application of the gas adjust the potentiometer counter clockwise until the alarm turns "on".
5. If the monitor fails to stop alarming within one minute of the removal of the gas adjust the poten­tiometer clockwise until the alarm stops.
6. After adjustment, repeat steps 4 and 5 to verify cali­bration.
7 . When the calibration is complete, reconnect any
alarm equipment as necessary .
2050/2060 Gas Monitors (5/04)
Page:
6
Page 10
Instruction Manual
5.0 SERVICE
5.1 Sensor Replacement
The gas sensor needs replacement when:
- It is no longer possible to obtain correct calibration
- The failed sensor alarm (oscillating red/green LED) is on
- The sensor output signal is noisy, causing incorrect gas alarms.
T o replace the sensor:
1. Confirm that the system power has been removed.
2. Remove the cover from the main enclosure.
3. Unplug the sensor connector on the lower right hand corner of the electronics board.
4. Unscrew the sensor assembly from the end of the enclosure.
5. Reverse the preceding steps to install the new sen­sor assembly .
6 . Restore power and allow a minimum of 24 hours
for stabilization before recalibration. The calibra­tion should be re-verified after 6 days of continu­ous operation.
Page:
Figure 5-1 Model 2050/2060 Electronics Board Connections
7
2050/2060 Gas Monitors (5/04)
Page 11
6.0 INST ALLA TION
Instruction Manual
6.1 Gas Sensor Module Locations
The gas sensor module is a diffusion type sensor that should be located close to the anticipated source or destination of the gas hazard. For heavy gases such as H the ground. For lighter gases such as CO and com­bustible gases use a higher elevation.
After optimum locations are determined based on the above recommendations, consideration should be given to placing the sensors in locations that are accessible for calibration service. Slight adjustments to the location of the sensor may have little impact on effectivity but major effect on accessibility .
6.2 Mounting
T o mount the monitor , remove the four svrews in the cover plate and move the cover plate aside. Use screws through the monitoring holes visible in the four corners of the housing to mount the monitor to a wall.
- The installation must meed any hazardous envi­ronmental codes for AC/DC electrical instrumenta­tion.
S install the module within 24 inches of
2
6.4 Power Supply
The power supplied by the controlling device or an external power supply must meet the following speci­fications:
Volt age:9 - 24 VDC Current:250 mA
6.5 Alarm Devices
The internal relay is rated at 0.5 Amp, normally open dry contact rated at 100 VDC/130 VAC.
6.3 Wiring
Interconnect wiring from the controller to the module is by 4 conductor 22 AWG (or lower AWG) cable, conduit as necessary. Shielding is not required. For installations where the distance from the con­troller to the sensor is greater than 500 feet, 18 AWG cable is recommended.
Terminal Function
TB1-1 + VDC (9 – 24) TB1-2 GND (0VDC) TB2-1 Relay NC (Normally Closed) or
Output to Model 2102 Safe TB2-2 Relay Common TB2-3 Relay NO (Normally Open) or
Output to Model 2102 Alarm
Table 6.1
The terminal strip on the electronics board in the module. The wiring must be connected as indi­cated in Figure 6.1 depending upon the controller or relay configuration being used.
2050/2060 Gas Monitors (5/04)
Page:
8
Page 12
Instruction Manual
Page:
Figure 6.1
Interconnect Wiring T able
9
2050/2060 Gas Monitors (5/04)
Page 13
y Op
y Op
7.0 ALARM CONFIGURATION
Mo de l 20X or 200X C onfig ure d wi th Ala rm Rel a y O utpu t
Buzzer Active
Yes No
Default
Install Jumper
JP6
TB1-1 TB1-2 TB2-1 TB2-2
+ DCV
(9-24 V D C )
Mo del 20X or 200X Co nfigu re d for In terfa ce to M o del 2102 Alarm P an e l
Yes No
Default Default
Install Jumper
JP6
Remove Jumper
(Bes t for 20X)
Power S upply
Buzzer Active Relay Operation Interface Selection
Remove Jumper
(Bes t for 20X)
JP 6
0 DC
Common
JP 6
Configuration Jum pers
Relay Ope ration
Norma lly
Not E nergized
Default
Install Jumper
JP2 pins 1-2
W iring Termina tions
Output Terminals
NC Comm on
Configuration Jum pers
Norm ally
Not E nergized
N/A N/A
Normally
Energized
Instruction Manual
Normally
Energize d
Install Jumper
JP2 pins 2-3
TB2-3
NO
Model 2102
Install Jum pers
JP3 pins 1-2 JP2 pins 1-2
W iring Termina tions
Power S upply
TB1-1 TB1-2 TB2-1 TB2-3
+ DCV
To Model 2102
Term in a l J5 -2
or J6-2
0 DC
To Model 2102
Term i na l J5 -3
or J6-3
Jumpe r Configuration and W iring Term inations
Safe
To Model 2102
Term i na l J5 -1
or J6-1
Table 6.2
Rela
Terminal J5 -4 or
eration
Nor m a lly No t En e r g ize d
TB2-1 NC TB2-2 Common TB2-3 NO No P o we r Ope n Sensor Fail Safe Open
Connec ted
Oscillating
Connec ted
Alarm Open
Normally Energized - Fail Safe
TB2-1 NC TB2-2 Common TB2-3 NO No P o we r Ope n Sensor Fail
Connec ted
Oscillating Safe Open Alarm Open
Connec ted
Table 6.3
Rela
erating States
Output Terminals
TB2-2
Alarm
To Model 2102
Not Used
J6-4
Connected
Connected
2050/2060 Gas Monitors (5/04)
Page:
10
Page 14
Instruction Manual
8.0 SPECIFICA TIONS
Indicators: Green LED for Monitor “ON”/Safe
Red LED for Alarm
Sensor Type: Solid-state metal oxide semiconduc-
tor
Alarm Setpoint::
2050 Series 100 ppm NH 2060 Series R1 1 500 ppm
R1 2 500 ppm R2 2 50 ppm R113 100 ppm R123 70 ppm R134a 50 ppm R141b 40 ppm R142b 40 ppm R500 40 ppm R502 70 ppm
User adjustable with calibration over range of 50 to 500 ppm
9.0 ACCESSORIES AND REPLACEMENT P ARTS
3
Calibration Gas: 2050 Series 80 ppm Hydrogen (H2) 2060 Series 150 ppm Carbon Monoxide (CO)
Power: 9 VDC, 250 mA Signal Output::
2050-00/2060-00 Nominal 5 volt DC, source 25 mA.
SAFE and ALARM Signals
0.5 AMP dry contact, 100 VDC, 130
VAC Response Time: Less than 90 seconds Operating T emperature Range:
-4o to 158oF (-20o to 70oC) Enclosure Material: Polystyrene Size: 5.6 x 3.1 x 2.1 in.
(14.2 x 7.9 x 5.3 cm)
Weight: 0.3 lb (0.15 Kg)
Alarm Panels
Model 2102-00 Alarm Panel - 2 channel Model 2102-01 Alarm Panel - 2 channel with audible
Calibration Accessories
1200-26 Gas Sensor Calibrator w/2 gas cylinders
(specify gas type/conc) 1290-10 Gas Cylinder - Carbon Monoxide 150 PPM 1290-11 Gas Cylinder - Hydrogen 80 PPM
10.0 LIMITED WARRANTY
SIERRA MONIT OR CORPORATION warrant s its products to be free from defects in workmanship or material under normal use and service for two year after date of shipment. SMC will repair or replace without charge any equipment found to be defective during the warranty period. Final determination of the nature and responsibility for defective or damaged equipment will be made by SMC personnel.
All warranties hereunder are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without SMC approval or which have been subjected to accident, improper maintenance, installation or application, or on which original identification marks have been re­moved or altered. This Limited Warranty also will not apply to interconnecting cables or wires, consumables (ie. calibra­tion gases, batteries), nor to any damage resulting from bat­tery leakage.
Replacement parts
SPF22073 Sensor for 2050-XX SPF22074 Sensor for 2060-XX
In all cases SMC’s responsibility and liability under this warranty shall be limited to the cost of the equipment. The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty provision and com­pliance with such instruction shall be a condition of this warranty .
Except for the express warranty stated above, SMC disclaims all warranties with regard to the products sold hereunder including all implied warranties of merchantability and fit­ness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of SMC for damages including but not limited to consequential damages arising out of/or in connection with the use or performance of the product.
Page:
11
2050/2060 Gas Monitors (5/04)
Loading...