Siemens Sitrans F FX300 User Manual

Page 1
Page 2
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of Siemens.
Subject to change without notice.
Copyright 2009 by Siemens
2
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 3
SITRANS FX300
CONTENTS
1 Safety instructions 6
1.1 Intended use ..................................................................................................................... 6
1.2 Certifications .................................................................................................................... 7
1.3 Safety instructions from the manufacturer ..................................................................... 8
1.3.1 Copyright and data protection ................................................................................................ 8
1.3.2 Disclaimer ............................................................................................................................... 8
1.3.3 Product liability and warranty ................................................................................................ 9
1.3.4 Information concerning the documentation........................................................................... 9
1.3.5 Warnings and symbols used................................................................................................. 10
1.4 Safety instructions for the operator............................................................................... 10
2 Device description 11
2.1 Scope of delivery............................................................................................................. 11
2.2 Instrument versions ....................................................................................................... 11
2.2.1 Devices with connection flange ............................................................................................ 12
2.2.2 Sandwich version .................................................................................................................. 12
2.2.3 Device description................................................................................................................. 13
2.2.4 Free air delivery measurement - FAD (optional) ................................................................. 13
2.2.5 Gross heat meter .................................................................................................................. 14
2.2.6 Dual seal ............................................................................................................................... 15
2.3 Nameplate ...................................................................................................................... 16
3 Installation 17
3.1 Notes on installation ......................................................................................................17
3.2 Storage ........................................................................................................................... 17
3.3 Transport ........................................................................................................................ 17
3.4 Installation condtitions...................................................................................................18
3.4.1 Measurement of liquids ........................................................................................................ 19
3.4.2 Measurement of vapours and gases .................................................................................... 21
3.4.3 Pipelines with control valve.................................................................................................. 22
3.4.4 Preferred mounting position ................................................................................................ 22
3.4.5 Turning the connection housing ...........................................................................................23
3.4.6 Turning the display ............................................................................................................... 24
3.4.7 Heat insulation ......................................................................................................................25
3.5 Inlet and outlet runs....................................................................................................... 26
3.5.1 Minimum inlet runs............................................................................................................... 26
3.5.2 Minimum outlet runs ............................................................................................................ 27
3.5.3 Flow straightener.................................................................................................................. 27
3.6 Installation...................................................................................................................... 28
3.6.1 General installation notes..................................................................................................... 28
3.6.2 Installing devices in flange design ....................................................................................... 29
3.6.3 Installing devices in sandwich design .................................................................................. 30
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
3
Page 4
CONTENTS
SITRANS FX300
4 Electrical connections 31
4.1 Safety instructions.......................................................................................................... 31
4.2 Connecting the signal converter .................................................................................... 32
4.3 Electrical connection of current and pulse output ........................................................ 33
4.3.1 Power supply......................................................................................................................... 34
4.3.2 Totalizer / pulse output......................................................................................................... 34
4.4 Grounding connections................................................................................................... 36
4.5 Protection category ........................................................................................................37
5 Start-up 38
5.1 Start ................................................................................................................................ 38
5.2 Start-up and control .......................................................................................................38
6 Operation 39
6.1 Display and operating elements .................................................................................... 39
6.2 Operating principles .......................................................................................................40
6.2.1 Functional description of the keys........................................................................................ 40
6.2.2 Switch from measuring mode to menu mode...................................................................... 40
6.2.3 Navigation within the menu structure.................................................................................. 41
6.2.4 Changing the settings in the menu....................................................................................... 41
6.2.5 Changing units ...................................................................................................................... 42
6.2.6 Measures in the event of faulty indications.......................................................................... 43
6.3 Overview of the most important functions and units..................................................... 44
6.4 Error messages.............................................................................................................. 46
6.5 Menu structure............................................................................................................... 47
6.5.1 Overview of firmware versions ............................................................................................. 47
6.5.2 Entering values in change mode .......................................................................................... 48
6.5.3 Character selection in change mode.................................................................................... 48
6.5.4 Menu item Quick Setup......................................................................................................... 49
6.5.5 Menu item Tests.................................................................................................................... 50
6.5.6 Menu item Setup (firmware version - basic) ........................................................................ 51
6.5.7 Menu item Setup (firmware version - steam) ...................................................................... 54
6.5.8 Menu item Setup (firmware version - gas)........................................................................... 58
7 Service 63
7.1 Replacing signal converter / LCD display...................................................................... 63
7.2 Spare parts availability...................................................................................................64
7.3 Availability of services .................................................................................................... 64
7.4 Returning the device to the manufacturer..................................................................... 64
7.4.1 General information.............................................................................................................. 64
7.4.2 Form (for copying) to accompany a returned device............................................................ 65
4
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 5
SITRANS FX300
CONTENTS
8 Technical data 66
8.1 Functional principle........................................................................................................ 66
8.2 Technical data................................................................................................................. 67
8.3 Dimensions and weights ................................................................................................ 70
8.3.1 Flange versions..................................................................................................................... 70
8.3.2 Sandwich version .................................................................................................................. 73
8.4 Flow tables ..................................................................................................................... 75
9 Notes 79
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
5
Page 6
1 SAFETY INSTRUCTIONS
1.1 Intended use
The vortex flowmeters are made to measure the flow of gases, steam and liquids.
The devices are particularly suitable for the measurement of:
Clean liquids with low viscosity (< 10 cP)
Hydrocarbons with low viscosity (< 10 cP)
Water
Chemicals with low corrosiveness
Saturated steam
Superheated steam, including CIP and SIP applications in the food industry
Industrial gases
The devices are rated for the following flow velocities:
Liquids: 0,3...7 m/s / 1.0...23 ft/s
Gases and steam: 2.0...80 m/s / 6.6...262 ft/s
DN15: 3.0...45 m/s / 9.8...148 ft/s; DN25: 2.0...70 m/s / 6.6...230 ft/s
SITRANS FX300
If there is a risk of waterhammers in steam networks, appropriate condensate separators must be installed. Suitable measures must be taken to avoid water cavitation if it is a possible risk.
CAUTION!
The operator himself bears the sole responsibility for the intended use of the device regarding the suitability and the corrosion resistance of the used materials against the measured fluid.
The sensors are made from Stainless Steel 316 L (1.4404) or Hastelloy
In your project planning, please observe the data given in the corrosion tables.
The pressure-bearing parts have been designed and rated for stationary operation taking into
account the maximum pressure and temperature.
External forces and moments, caused e.g. by pipe stresses, have not been taken into account.
Primarily, volumetric flow and temperature are measured, with pressure measurement as an option. From these parameters the measuring device calculates the mass flow or standard volumetric flow using pre-programmed density data and then exports the measured values via various communication interfaces.
®
C22.
6
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 7
SITRANS FX300
1.2 Certifications
CE marking
The device fulfils the statutory requirements of the following EC directives:
Pressure Equipment Directive 97/23/EC
Low Voltage Directive 73/23/EEC
EMC Directive 89/336/EC
as well as
EN 61010
EMC specification acc. to EN 61326/A1
NAMUR recommendations NE 21 and NE 43
SAFETY INSTRUCTIONS 1
The manufacturer certifies successful testing of the product by applying the CE marking.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
7
Page 8
1 SAFETY INSTRUCTIONS
1.3 Safety instructions from the manufacturer
1.3.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to German copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
SITRANS FX300
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.3.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect, incidental, punitive and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
8
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 9
SITRANS FX300
1.3.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.3.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
9
Page 10
1 SAFETY INSTRUCTIONS
1.3.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
SITRANS FX300
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.4 Safety instructions for the operator
10
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 11
SITRANS FX300
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-1: Scope of delivery
1 Measuring device in ordered version 2 Handbook 3 Certificates, calibration report and parameter datasheet 4 Bar magnet
2.2 Instrument versions
The devices come in the following versions:
Converter with display
Blind device
Sensor in flanged design
Sensor in sandwich design
The following designs are available as options:
with pressure sensor
with isolation valve for the pressure sensor
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
11
Page 12
2 DEVICE DESCRIPTION
2.2.1 Devices with connection flange
The measuring system consists of a measuring sensor and a signal converter. These elements form a permanent mechanical unit.
Figure 2-2: Flanged devices with display
1 Version with temperature sensor 2 Version with temperature sensor and optional pressure sensor 3 Version with temperature sensor, optional pressure sensor and shut-off valve
SITRANS FX300
2.2.2 Sandwich version
The sandwich version features 2 centring rings to aid with installation.
Figure 2-3: Sandwich versions with display
1 Version with temperature sensor 2 Version with temperature sensor and optional pressure sensor 3 Version with temperature sensor, optional pressure sensor and shut-off valve
12
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 13
SITRANS FX300
2.2.3 Device description
Figure 2-4: Device description
1 Signal converter 2 Cable feedthrough grey, standard version 3 Pressure sensor, optional 4 Shut-off valve, optional 5 Measuring sensor 6 Centering ring
DEVICE DESCRIPTION 2
2.2.4 Free air delivery measurement - FAD (optional)
A (air) compressor draws air from the ambient atmosphere, compresses it and delivers it at the required pressure. Since the ambient atmosphere also contains water vapour, what the compressor draws in is a mixture of air and water vapour. Free air delivery measurement is to be understood under this condition. Most manufacturers specify free air delivery only at standard intake conditions. What the user ultimately requires as process air must first be determined before measuring can take place with an accuracy of ±1%.
The vortex flowmeter with the optional FAD function can measure the free air delivery online, humidity and speed compensated, regardless of its function as standard flowmeter. The integrated software evaluates the free air delivery automatically online.
The menu-driven, user-friendly software prompts the operator to enter the pressure, relative humidity, the required as well as current discharge pressure.
The steam tables and compressibility tables are saved as standard. The measuring device is optionally available with a pressure sensor which measures the discharge pressure online, making manual input of the values unnecessary.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
13
Page 14
2 DEVICE DESCRIPTION
2.2.5 Gross heat meter
In almost all applications with saturated steam, the steam is used for heating. It is much more interesting to know how great the heat flow volume is that is available to the process, than to know how great the flow is in kg/h.
As the enthalpy of steam changes with the temperature, it cannot be assumed to be constant. The vortex flowmeter has a special feature which calculates the flow of steam and water as power. The enthalpy tables are permanently programmed into the memory of the device.
The online density-compensated mass flow is multiplied by the correct enthalpy to obtain the flow as power output.
SITRANS FX300
Power {Q
] = mass flow [Qm] x enthalpy [H]
H
If the gross heat meter is activated, both the totalizer for the absolute steam consumption as well as that for the energy run internally.
14
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 15
SITRANS FX300
2.2.6 Dual seal
To cover the requirements of ANSI/ISA-12.27.01-2003 Requirements for Process Sealing Between Electrical Systems and Flammable or Combustible Process Fluid Dual seal, a membrane vent is integrated in the neck of the instrument. This vent works in the improbable case of a leakage as an annunciation valve between the primary seal (Process) and secondary seal (electronic compartment).
DEVICE DESCRIPTION 2
1 Annunciation valve (Membrane vent)
The sealing between the pick-up and the Flow tube is considered as the primary seal. Its construction material is always the same as the Flow tube itself (e.g. Stainless steel 316L or Hastelloy C22). Chemicals used for the process, going through the Flow tube, must be compatible with the designated material. By using the membrane vent as annunciation valve, all requirements for a dual seal device in accordance to above mentioned standard, are fulfilled.
It protects the electronic compartment against access by process media, due to possible leakages.
The leakage of the primary seal is visible.
Regular maintenance should be performed by the customer to ensure this Dual Seal notification system is working properly or that there is no leak.
If a leakage is noticed:
Siemens customer service should be contacted for servicing or replacement of the meter.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
15
Page 16
2 DEVICE DESCRIPTION
2.3 Nameplate
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
SITRANS FX300
Figure 2-5: Example for nameplate
1 Device type 2 Manufacturer 3 Notified ATEX & DGRL bodies (only available if this option was ordered) 4 Connection data: nominal diameter and pressure rating 5 PED data 6 Ex data (only available if this option was ordered) 7 Electrical connection data 8 Manufacturer's website
Figure 2-6: Example for nameplate
1 Serial number 2 Order number 3 Type code 4 Item number 5 Fluid data 6 Software variant 7 Tag number
16
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 17
SITRANS FX300
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry and dust-free location.
Avoid extended direct exposure to the sun.
Store the device in its original packing.
The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
INSTALLATION 3
3.3 Transport
Use lifting straps wrapped around both process connections for transport.
Do not lift measuring devices by the signal converter housing for transport.
Never lift the measuring device by the pressure sensor.
Do not use lifting chains as they may damage the housing.
Figure 3-1: Transport instructions
CAUTION!
Non-secured devices can pose risk of injury. The centre of mass of the device is often higher than the point at which the lifting straps are attached. Prevent the measuring device from sliding or rotating accidentally.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
17
Page 18
3 INSTALLATION
3.4 Installation condtitions
INFORMATION!
For accurate volumetric flow measurement the measuring device needs a completely filled pipe and a fully developed flow profile. Please observe the instructions regarding inlet and outlet pipe runs as well as the installation position. In the event of vibrations their lowest in a transverse direction to the flowmeter.
CAUTION!
When installing the device in the piping, the following points must be observed:
Nominal diameter of connection pipe flange = nominal flange diameter of pipe!
Use flanges with smooth holes, e.g. welding neck flanges.
Align carefully the holes of the connecting flange and the flowmeter flange.
Check the compatibility of the gasket material with the process product.
Make sure that the gaskets are arranged concentrically. The flange gaskets must not project into the pipe cross-section.
The flanges have to be concentric.
There must not be any pipe bends, valves, flaps or other internals in the immediate inlet run.
Devices in sandwich version may only be installed using a centering ring.
Never install the device directly behind piston compressors or rotary piston meters.
Do not lay signal cables directly next to cables for the power supply.
vibrations on the piping, select the installation site so that the vibrations are at
vibrationsvibrations
SITRANS FX300
18
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 19
SITRANS FX300
3.4.1 Measurement of liquids
Prohibited installation
Prohibited installation
Prohibited installationProhibited installation
Figure 3-2: Upper pipe bend
CAUTION!
Prohibited: Installing the device in an upper pipe bend 1, because there is a risk of gas bubbles 2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement.
INSTALLATION 3
Figure 3-3: Downpipe and outlet
CAUTION!
Prohibited: Installing the device in a downstream pipe 3 or upstream pipe of a outlet 4. There is the risk of partially filled pipes.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
19
Page 20
3 INSTALLATION
Recommended installations for measurement of liquids
Recommended installations for measurement of liquids
Recommended installations for measurement of liquidsRecommended installations for measurement of liquids
CAUTION!
It is absolutely necessary to comply with the required inlet and outlet runs.
SITRANS FX300
1 If the device is installed in a downpipe, a standpipe must be installed immediately after it. 2 Installing the device in an inclined standpipe. 3 Installing the device in a vertical standpipe. 4 Installing the device in the lower pipe bend.
1 Above a horizontal pipe 2 On a vertical pipe
INFORMATION!
Depending on the installation position, you may have to rotate the display and/or the connection housing.
20
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 21
SITRANS FX300
3.4.2 Measurement of vapours and gases
Prohibited installation
1 Lower pipe bends 2 Condensate
DANGER!
Prohibited: Installing the device in a lower pipe bend 1, because there is a risk of condensate forming 2. Condensate can lead to cavitation and inaccurate measurement. Under certain circumstances the device can be destroyed and the measured product can leak.
INSTALLATION 3
Recommended installations
Recommended installations
Recommended installationsRecommended installations
CAUTION!
It is absolutely necessary to maintain the required inlet and outlet runs.
1 Inlet and outlet falling 2 Rising inlet - falling outlet
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
21
Page 22
3 INSTALLATION
3.4.3 Pipelines with control valve
INFORMATION!
To ensure smooth and correct measurement, the manufacturer recommends not installing the measuring device downstream from a control valve. This would run the risk of vortex formation, which would distort the measuring result.
Figure 3-4: Pipeline with control valve
1 Recommended: installing the device upstream 2 Not recommended: Installing the flowmeter directly downstream
upstream from the control valve at a distance of 5 DN
upstreamupstream
directly downstream of control valves, due to vortex formation.
directly downstreamdirectly downstream
SITRANS FX300
3.4.4 Preferred mounting position
Preferred mounting position
1 Above a horizontal pipe 2 underneath a horizontal pipe (not permitted with lines at risk for condensate ) 3 On a vertical pipe
INFORMATION!
Depending on the installation position, you may have to rotate the display and/or the connection housing. For further information refer to Turning the display on page 24
.
22
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 23
SITRANS FX300
3.4.5 Turning the connection housing
DANGER!
All work on the device electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed.
CAUTION!
Do not damage the electrical cable by overtwisting it. Do not remove the electrical connector.
INSTALLATION 3
Figure 3-5: Allen screws on connection housing
Disconnect the power supply from the measuring device.
Loosen the four screws 1 on the rear side of the connection housing.
Lift the connection housing and turn it to the required position in 90° steps.
Screw the connection housing back on.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
23
Page 24
3 INSTALLATION
3.4.6 Turning the display
DANGER!
All work on the device electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed.
INFORMATION!
If the measuring device is installed in a vertical pipe, you will have to turn the display by 90 installed below a pipe, turn 180
SITRANS FX300
°
; if
°
.
Figure 3-6: Turning the display
Turn the display as follows:
Disconnect the power supply from the measuring device.
Unscrew the cover in front of the display 1 from the connection housing.
Pull the display 2 carefully a few centimetres out of the anchor fitting and turn it to the
required position 3.
Press the display onto the spacer pins 4, until it clicks.
Turn the cover with gasket 5 back onto the housing and tighten it by hand.
24
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 25
SITRANS FX300
3.4.7 Heat insulation
CAUTION!
The area above the converter support must not be heat-insulated. The heat insulation 3 may only extend to the maximum height 1 shown below up to the connecting screws of the measuring sensor.
INSTALLATION 3
Figure 3-7: Heat insulation on connection piece and signal cable
1 Max. height of insulation up to intermediate piece between measuring sensor and signal converter 2 Max. thickness of the insulation up to the bend of the pressure pipe 3 Insulation
CAUTION!
The heat insulation 3 may only extend as far as the bend of the pressure sensing line 2.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
25
Page 26
3 INSTALLATION
3.5 Inlet and outlet runs
3.5.1 Minimum inlet runs
SITRANS FX300
1 General inlet run without disturbing flow 20 DN 2 Behind a control valve 50 DN 3 After a pipe diameter reduction 20 DN
26
1 After a single bend 90° 20 DN 2 After a double bend 2x90° 30 DN 3 After a double three-dimensional bend 2x90° 40 DN
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 27
SITRANS FX300
3.5.2 Minimum outlet runs
INSTALLATION 3
Figure 3-8: Minimum outlet runs
1 Upstream of pipe expanders, pipe bends, control valves, etc. 5 DN 2 Upstream of metering points 56 DN
INFORMATION!
The interior of the pipe at the metering points must be free of burrs and other flow impediments. The measuring device has an internal temperature sensor. External temperature sensor must be located 5 DN behind the flow sensor outlet. Use sensors that are as short as possible to avoid disturbing the flow profile.
3.5.3 Flow straightener
If, due to the type of installation, the required inlet runs are not available, the manufacturer recommends using flow straighteners. Flow straighteners are installed between two flanges upstream of the device and shorten the required inlet run.
Figure 3-9: Flow straightener
1 Straight inlet run upstream of straightener 2 DN 2 Flow straightener 3 Straight pipe run between flow straightener and device 8 DN 4 Minimum straight outlet run 5 DN
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
27
Page 28
3 INSTALLATION
3.6 Installation
3.6.1 General installation notes
CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. The regional occupational health and safety directives must always be observed.
The following procedures have to be carried out before installing the device:
Remove all transportation safety devices and protective coverings from the device.
Ensure that the gaskets have the same diameter as the pipelines.
Note the correct flow direction for the device. This is indicated by an arrow on the housing of
the measuring sensor.
On metering points with varying thermal loads, the flowmeters have to be mounted with stress
bolts (DIN 2510).
Stress bolts or bolts and nuts are not included in the scope of delivery.
Ensure that the measuring flange is concentrically fitted.
Note the exact installation length of the measuring device when preparing the metering point.
SITRANS FX300
Figure 3-10: Preparing the metering point
1 Installation length of measuring device + thickness of gaskets.
CAUTION!
The internal diameter of the pipelines, the measuring sensor and the gaskets must match. The gaskets may not protrude into the flow.
Figure 3-11: Internal diameter
1 Internal diameter connection pipe 2 Internal diameter flange and gasket 3 Internal diameter measuring sensor
28
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 29
SITRANS FX300
3.6.2 Installing devices in flange design
INSTALLATION 3
Figure 3-12: Installing devices in flange design
1 Gasket 2 Bolts with fixing nuts
Use bolts and fastening nuts 2 to attach the measuring device to one side of the flange.
While doing so, insert the gaskets 1 between measuring sensor and flange and align them.
Check that the gasket is concentric and that it is not protruding into the pipe cross-section.
Install the gasket, bolts and fastening nuts on the other side of the flange.
Align the measuring device and the gaskets so they are concentric.
Now tighten all nuts bit by bit alternately across the diagonal.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
29
Page 30
3 INSTALLATION
3.6.3 Installing devices in sandwich design
SITRANS FX300
Figure 3-13: Installation using centering ring
1 Measuring sensor 2 Centering ring 3 Bolts with fixing nuts 4 Hole 5 Hole 6 Gasket
Push the first bolt 3 through the hole 5 of both flanges.
Screw on the nuts and washers to both ends of the bolt 3 but do not tighten them.
Install the second bolt through the holes 4.
Place the measuring sensor 1 between the two flanges.
Insert the gaskets 6 between measuring sensor 1 and flanges and align them.
Check that the flange is concentric.
Install the remaining bolts, washers and nuts. Do not yet tighten the nuts.
Turn the centring ring 2 in a counter-clockwise direction and align the device.
Check that the gaskets 6 are concentric; they must not protrude into the pipe cross-section.
Now tighten all nuts bit by bit alternately across the diagonal.
30
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 31
SITRANS FX300
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
ELECTRICAL CONNECTIONS 4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
31
Page 32
4 ELECTRICAL CONNECTIONS
4.2 Connecting the signal converter
Figure 4-1: Signal converter housing with housing cover
1 Housing cover of the electrical terminal compartment 2 Electrical connection terminals with the housing cover open 3 Terminal A current output - 4 Terminal A+ current output + 5 Terminal B pulse output - 6 Terminal B+ Pulse output + 7 PE terminal in housing 8 PE terminal on connection piece between measuring sensor and signal converter.
SITRANS FX300
Both grounding terminals 7 and 8 are equally effective from a technical point of view.
The following procedures are to be performed:
Unscrew the housing cover 1 of the electrical terminal compartment.
Thread the connection cable through the leadthrough in the housing.
Connect the cable for the current output and the cable for the optional pulse output as shown
in the cable terminal diagrams below. To facilitate installation the connection plug can be removed from the device. The plug is configured in such a way as to prevent reverse polarity.
Connect the grounding to terminal 7 alternatively use the PE terminal on the connection
piece between measuring sensor and signal converter 8.
Tighten the cable glands.
Hand-tight the housing cover with gasket.
32
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 33
SITRANS FX300
ELECTRICAL CONNECTIONS 4
4.3 Electrical connection of current and pulse output
Current output:
Current output:
Current output:Current output: In some cases, a shielded or twisted cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit).
Pulse output:
Pulse output:
Pulse output:Pulse output: When using the pulse output, two separate signal circuits are necessary if the pulse output is utilized together with analogue signals. Each signal circuit requires its own power supply. The total resistance must be adapted so that the total current I
Connection current output on terminals A, A+ Connection pulse output on terminals B, B+
does not exceed 100 mA.
tot
Figure 4-2: Electrical connection of current and pulse output
1 Power supply per signal circuit 2 Optional display unit
3 Load for HART 4 e.g. counter
®
250 Ω
The maximum load resistance is calculated as follows:
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
33
Page 34
4 ELECTRICAL CONNECTIONS
4.3.1 Power supply
INFORMATION!
The supply voltage has to be between 14 VDC and 36 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the measuring device).
The required supply voltage can be calculated using the formula below:
U
= RL * 22mA + 14V
ext.
where
= the minimum supply voltage and
U
ext.
= the total measuring loop resistance is.
R
L
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
SITRANS FX300
4.3.2 Totalizer / pulse output
The basic units for the totalizer and the pulse output are mmmm
3333
norm.
norm. for standard volume and kg
mmmm
norm.norm. The unit and the conversion factor can be changed in menu item 3.2.8 "Conf.Tot". User specific units (User Def.) can also be entered but the conversion factor must always be entered based on the base unit. For a sample calculation see chapter 6.2.5 "Change units".
The maximum frequency of the pulse output is 0.5 Hz.
Figure 4-3: Pulse output
0.5 Hz
1 f
max
2 closed 3 open 4 Pulse 250 ms
kg for mass.
kgkg
3333
for volume,
34
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 35
SITRANS FX300
The pulse output is a passive "open collector" output which is electrically separated from the current interface and the measuring sensor. It can be configured as a high current output or NAMUR output using a jumper on the amplifier board.
Jumper in NAMUR setting
Jumper in NAMUR setting
Jumper in NAMUR settingJumper in NAMUR setting
Ri = 900 Ω U
= 36 VDC
max
ELECTRICAL CONNECTIONS 4
The maximum frequency of the pulse output is f
Jumper in high current setting
Jumper in high current setting
Jumper in high current settingJumper in high current setting
Open: Maximum voltage U
Closed: Maximum current I
= 36 VDC Closed current IR < 1 mA
max
= 100 mA Voltage U < 2 VDC
max
The maximum frequency of the pulse output is f
max
max
= 0.5 Hz
= 0.5 Hz
Figure 4-4: Jumper settings pulse output
1 Electronic insert 2 Jumper on board 3 Jumper in high current setting 4 Jumper in NAMUR setting
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
35
Page 36
4 ELECTRICAL CONNECTIONS
4.4 Grounding connections
The grounding can be done either by connecting the PE terminal in the housing or the PE terminal on the connection piece between measuring sensor and signal converter. Both of these electrical connections are equally effective from a technical point of view.
SITRANS FX300
Figure 4-5: Grounding connection
1 Electrical grounding connection on connection piece between measuring sensor and signal converter. 2 Electrical grounding connection on housing
DANGER!
The measuring device has to be grounded properly to achieve accurate measurement. The grounding wire may not transfer any interference voltage. Do not use this grounding wire to ground any other items of electrical equipment.
36
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 37
SITRANS FX300
4.5 Protection category
The measuring device meets all requirements of protection category IP66/67.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection category must be ensured again.
Figure 4-6: Cable feedthrough
ELECTRICAL CONNECTIONS 4
Therefore it is essential to observe the following points:
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
The electrical cables used must be undamaged and must comply with regulations.
The cables must be laid with a loop 1 upstream of the measuring device to prevent water
from getting into the housing.
The cable feedthroughs 2 must be tightened.
Align the measuring device so that the cable feedthrough is never facing up 3.
Close the unused cable feedthroughs using blind plugs 4.
Do not remove the required cable bushing from the cable feedthrough.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
37
Page 38
5 START-UP
5.1 Start
After the device is switched on, the display shows the following in sequence
1. Testing...
2. Device type
Sofware Version - Revision
The device performs a self-test and switches to measurement mode. Here, all of the parameters preset for the customer are analysed and checked for plausibility, and the current measured value is displayed.
5.2 Start-up and control
INFORMATION!
The flowmeter is largely maintenance-free. Observe the application limits in respect of temperature and medium. Additional information refer to Technical data on page 67
SITRANS FX300
.
38
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 39
SITRANS FX300
6.1 Display and operating elements
With the cover open the device is operated by using the mechanical keys on the front; with the cover closed a bar magnet is used.
CAUTION!
The switching point of the magnetic sensors is directly under the glass panel above the appropriate symbol. Only touch the symbol vertically and from the front using the bar magnet. Touching it from the side may cause a malfunction.
OPERATION 6
Figure 6-1: Display and operating elements
1 Enter button (bar magnet) 2 Right button (bar magnet) 3 Up button (bar magnet) 4 Up button (mechanical) 5 Enter button (mechanical) 6 Right button (mechanical) 7 Display
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
39
Page 40
6 OPERATION
The mechanical keys and keys for the bar magnet have the same functionality. In this documentation the keys are represented as symbols to describe the operating functions:
Mechanical Bar magnet Symbol
Table 6-1: Description of keys
6.2 Operating principles
SITRANS FX300
^
6.2.1 Functional description of the keys
Switch from measuring mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
In change mode:
In change mode: Move the input cursor one position to the right; after the last digit the
In change mode:In change mode: input cursor jumps back to the beginning.
In measuring mode:
^^^^ Switch to one menu level higher
Table 6-2: Description of the operating keys
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
Change between the menu items within a menu level
In change mode:
In change mode: Changing parameters and settings; running through the available
In change mode:In change mode: characters; shifting the decimal point to the right.
Return to measuring mode with a query whether the data should be accepted
6.2.2 Switch from measuring mode to menu mode
Measuring mode Operation Menu mode
156.3
156.3
156.3156.3 kg/h
kg/h
kg/hkg/h
1.1.1
1.1.1
1.1.11.1.1 Language
Language
LanguageLanguage
40
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 41
SITRANS FX300
6.2.3 Navigation within the menu structure
Navigation within the menu is by means of the and ^ buttons. Pressing button takes you one menu level lower, ^ takes you one menu level higher. If you are already located at the lowest level (function level), you can use the button to go in the change mode, which can be used to set data and values. If you are located at the first level (main menu), you can use the ^ key to exit the menu mode and return to the measuring mode.
OPERATION 6
Measuring
mode
Table 6-3: Navigation menu structure
Main menu↑→ Sub-menu↑→ Function
^^^^ ^^^^ ^^^^ ^^^^
6.2.4 Changing the settings in the menu
Use the and keys to navigate in the menu in which you want to change a setting or value.
Use the key to activate the change mode in the selected menu.
i The current values or settings are displayed.
Use the and keys to change the value or setting.
Save the new value or setting using the ^ key. Some menu items contain several setting options. They are displayed in sequence by pressing the ^ key.
i Takes you back to the main menu.
Press the ^ key to save the settings.
i The query "Save Yes" appears. Switch between "Yes" and "No" by pressing the key.
Switch between "YES" and "NO" by pressing the key.
Edit
^^^^
Save Yes ^^^^ Changes saved.
Save No ^^^^ Changes not saved.
An update is carried out and the display returns to measuring mode.
The display returns to measuring mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal plausibility check. If implausible inputs have been made, the display remains in the current menu, and the changes are not saved.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
41
Page 42
6 OPERATION
Example: changing the default parameter from m3/h to l/min
Procedure Display Procedure Display
SITRANS FX300
1 3 x
2 3 x
3
4 3 x
3 x 1.1.1111
3 x 3 x
3 x 1.1.4444
3 x 3 x
3 x L/min
3 x 3 x
6.2.5 Changing units
When entering numbers and values in floating point format, the maximum possible accuracy is
0.003%. The accuracy depends on both the position of the decimal point and the length of the number entered. Numerical values and factors are displayed in the first line of the 10-digit display. Numerical values are either displayed in floating point format (123.4567890) or are expressed in exponential format (123456E002). The conversion factor of the totalizer and the pulse output is, however, expressed in whole numbers.
Input values exponential format
107.2
3
/h
m
Language
Max. Flow
m3/h
Unit
Unit
5 ^^^^ 0000600.00
L/min
6 ^^^^ Display
Unit
7 ^^^^ 1111.1.4
Max. Flow
8 ^^^^ Save Yes
9 ^^^^ 1787.0000
L/min
Display position 1 2 3 4 5 6 7 8 9 10
Decimal point - - - - - -
Input values 0 ...
9
0 ...90 ...90 ...
9
0 ...90 ...9E -
or 0
0 ...30 ...
8
To select the exponential function, the decimal point must be located between the 2nd and 5th decimal position.
Flow units
The base units are mmmm measurement and kg/h
3333
/h for volume flow measurement, mmmm
kg/h for mass flow measurement.
kg/hkg/h The flow units can be changed in menu item 1.1.4 "Max.Flow". User defined units can be entered using "User Def." The unit (text) as well as the conversion factor (number) can be entered here. The conversion factor must always be entered based on the base unit.
3333
/h stand.
/h stand. for standard volumetric flow
/h stand./h stand.
Conversion
Formula New unit (User Def.)
Example: 1 litre / h = 0.001 *
(User Def.) ==== A1 Coeff.
(User Def.)(User Def.)
A1 Coeff. **** Base unit
A1 Coeff.A1 Coeff.
Base unit
Base unitBase unit
m3/h
Menu items
1.1.3 Maes.Inst Volume / Standard volume / Mass
1.1.4 Max. Flow Unit (User Def.) / Text / A1 Coeff.
Unit (User Def.) / Text / A1 Coeff. / Max. Flow / Flow Displ
Unit (User Def.) / Text / A1 Coeff.Unit (User Def.) / Text / A1 Coeff.
42
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 43
SITRANS FX300
Totalizer / pulse output
The base units for the totalizer and the pulse output are mmmm
3333
mmmm The unit and the conversion factor can be changed in menu item 3.2.8 "Tot.Conf.". User defined units (User Def.) can be selected and entered. The conversion factor of the unit must always be entered based on the base unit.
The maximum pulse output frequency is 0.5 Hz. To ensure that the pulse output does not exceed
0.5 Hz, the conversion factor of the totalizer must be selected and set accordingly. The pulse output represents an exact copy of the whole number value of the internal totalizer.
Max. pulses per hour
norm. for standard volume and kg
kg for mass.
kgkg
3333
for volume,
OPERATION 6
fmax. 0.5 Hz 1 pulse-break 2 seconds Max. pulses / hour = 1800
1800 pulses
18001800
Example
Flow Qmax. Factor A1 Coef. Pulses / litre Pulses / h Comment
5.6 m3/h
0.001 1 pulse / 1 litre 5600 not possible
0.01 1 pulse / 10 litres 560 560 < 1800 = ok.
Menu items
1.1.3 Maes.Inst Volume / Standard volume / Mass
1.1.4 Max. Flow Unit (User Def.) / Text / A1 Coeff. / Max. Flow
3.2.5 Function P Yes
3.2.7 Totalizer Tot. on
3.2.8 Tot. Conf. User def. / Unit Text / A1 Coeff.
Yes / No
YesYes
Tot. on / Tot. off
Tot. onTot. on
User def. / Unit Text / A1 Coeff. / Preset Value / Reset / Display
User def. / Unit Text / A1 Coeff.User def. / Unit Text / A1 Coeff.
Max. Flow / Flow Displ
Max. FlowMax. Flow
6.2.6 Measures in the event of faulty indications
If the indications on the display or the responses to keypad commands are faulty, you have to do a hardware reset. Switch the power supply OFF and ON again.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
43
Page 44
6 OPERATION
6.3 Overview of the most important functions and units
INFORMATION!
A complete list of all functions and short descriptions is provided in the appendix. All default parameters and settings are adapted for the specific customer.
Level Designation Explanation
1.1.1 Language Select the menu language
1.1.4 Max. Flow Maximum flow rate
1.1.5 Min. Flow Minimum flow rate
1.1.6 Timeconst. Time constant, damping value [s]
2.1.1 Test I Check current output
2.1.2 Test P Test pulse output
3.1.1 Error Msg. Error indicator
Table 6-4: The most important functions
The value set is represented by a 20 mA analogue current output. If the current value exceeds the preset value, an alarm is indicated.
The set value does not
not represent the 4 mA value of the current output.
notnot
Yes: error messages are displayed No: error messages are suppressed. A flashing cursor at the top left indicates that error messages are present.
SITRANS FX300
Volume mass units
Volume mass units
Volume mass unitsVolume mass units
Volume Standard volume Mass
Liquids, steams, gases Air Gas Liquids, steams, gases
m3/h
3
/min
m
3
/s
m L/h L/min L/s
3
/h
ft
3
/min
ft
3
/s
ft gal/h gal/min gal/s ImpGal/h ImpGal/min ImpGal/s
3
/h
cm
3
/min
cm
3
/s
cm
3
/h
dm
3
/s
dm bbl/h bbl/d User Def.
Table 6-5: Flow measurements and units
FAD m3/h
3
FAD m FAD m
/min
3
/s FAD L/h FAD L/min FAD L/s
3
/h
FAD ft
3
/min
FAD ft
3
/s
FAD ft User Def.
m3/h (norm)
3
/min (norm)
m
3
/s (norm)
m L/h (norm) L/min (norm) L/s (norm)
3
/d (std.)
ft
3
/h (std.)
ft
3
/min (std.)
ft
3
/s (std.)
ft User Def.
kg/h kg/min kg/s t/h t/min t/s lb/h lb/min lb/s g/min g/s User Def.
44
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 45
SITRANS FX300
Unit totalizer
Unit totalizer
Unit totalizerUnit totalizer
Volume Standard volume Mass
Liquids, steams, gases Gas Liquids, steams, gases
m L
3
ft
3
ft gal ImpGal
cm dm
bbl User Def.
Table 6-6: Totalizer units
Temperature - Pressure - Power - Energy - Density units
Temperature - Pressure - Power - Energy - Density units
Temperature - Pressure - Power - Energy - Density unitsTemperature - Pressure - Power - Energy - Density units
OPERATION 6
3
/h
m3 norm L norm
3
std.
ft User Def.
3
3
kg t lb g User Def.
Temp. Absolute
pressure (over)
°C °F
K User Def.
Table 6-7: Additional units
Pa(g) kPa(g) MPa(g)
2
kg/ms
2
kp/cm atm(g) torr(g) bar(g) mbar(g) psi(g)
2
(g)
lbf/ft kgf/cm
inHg(g) mmHg(g) mmH2O(g)
User Def.
(g) (g)
2
(g)
Power Energy Density
kJ/h MJ/h GJ/h Btu/h th kcal/h User Def.
kW MW TR kJ MJ GJ Btu th kcal kWh MWh User Def.
3
kg/m kg/L g/L h/mL
3
g/cm
3
g/ft g/ImpGal g/gal kg/gal Lb/gal Lb/ImpGal User Def.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
45
Page 46
6 OPERATION
6.4 Error messages
Error message Cause Measure
SITRANS FX300
No Signal No signal from vortex amplifier Check connector
In the event of measuring sensor problems contact service.
Low freq Sample frequency too low Contact service.
High Freq Sample frequency too high Contact service.
Low Flow Flow rate less than the set minimum
flow rate q
min
Q too high Flow rate higher than the set maximum
flow rate q
max
Signal converter continues to display the current flow rate but the accuracy of the measurement may be affected.
Corrective measures depend on the application! If the flow rate exceeds the maximum value, the measuring sensor may be physically damaged.
Inv. Config. Invalid configuration data in FRAM
(permanent memory)
Check entire configuration! Contact service if error message persists.
Amp. Fail Error in the pre-amplifier stage Contact service.
Chk. Instal. Quality of the vortex signal too poor 1. Check q
min
2. Excessive pipe vibration and faulty flow profile
Low Signal Amplitude of vortex signal is too low 1. q
2. If q
ok?
min
OK, contact service
min
Hi. Signal Amplitude of vortex signal is too high Occurs with media of high density
L.Temp.Phy Operating temperature lower than
specified
H.Temp.Phy Operating temperature higher than
specified
Hi.P.Phys Pressure higher than specified
1. Check q
2.If q
Take corrective measures within process.
Take corrective measures within process as quickly as possible, otherwise both the measuring sensor and the signal converter may be
max
OK, contact service
max
damaged.
Tsens Shrt. Temperature sensor short circuit Shows error at temperature sensor!
Tsens Open Open temperature sensor
Contact service.
P. Sen. Fail Faulty pressure sensor Displays error at pressure sensor!
Contact service.
Table 6-8: Error messages
46
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 47
SITRANS FX300
6.5 Menu structure
6.5.1 Overview of firmware versions
There are three firmware versions, each of which is tailored to a different use of the measuring device:
Basic:
Basic: liquids and gases without compensation, saturated steam with density compensation
Basic:Basic: using the temperature
Steam:
Steam: saturated and superheated steam with density compensation using pressure and
Steam:Steam: temperature, gross heat meter
Gas:
Gas: gas, gas mixture and wet gas with density compensation using pressure and
Gas:Gas: temperature, FAD (free air delivery measurement)
The menu structures differ from one another depending on the firmware version. The following table contains an overview of all menu items in the first menu level. For a complete description of the menu, note the firmware version of your measuring device and follow the relevant instructions in the table.
OPERATION 6
Menu items Firmware version
Basic
1. Quick setup refer to
2. Tests refer to
3. Setup refer to
Setup (firmware version
- basic)
5. Service The service menu is only accessible to service personnel and is not described.
Table 6-9: Overview of menu structure
Menu item Quick Setup Menu item Tests Menu item
on page 51
Firmware version Steam
on page 49
on page 50
refer to
Setup (firmware version
- steam)
Menu item
on page 54
Firmware version Gas
refer to
Setup (firmware version
- gas)
Menu item
on page 58
INFORMATION!
The measuring device has been preset at the factory in accordance with the customer order. Therefore, subsequent configuration via the menu is only necessary if the intended use of the measuring device changes.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
47
Page 48
6 OPERATION
6.5.2 Entering values in change mode
Moves the insertion point one position to the right; after the last position, the insertion point returns to the beginning.
Cycles through available values and characters; moves the decimal point to the right.
^^^^
Accepting the entry.
6.5.3 Character selection in change mode
Depending on the menu function, you can choose from the following characters:
SITRANS FX300
Numbers
0 1 2 3 4 5 6 7 8 9
Lowercase letters
a b c d e f g h i j
k l m n o p q r s t
u v w x y z
Uppercase letters
A B C D E F G H I J
K L M N O P Q R S T
U V W X
Y
Z
Special characters
°
- . /
1 "blank"
2 3
" % & : < = >
1
48
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 49
SITRANS FX300
6.5.4 Menu item Quick Setup
Level Designation Selection / entry field Explanation
1.1.1 Language
German ↑... German language French French language English ↑...^ English language
1.1.2 Location
0000000000 Location......^
1.1.3 Maes.Inst
Volume Volume measurement Norm. Vol. 1 Standard volume flow measurement Mass ↑...^ Mass flow measurement
1.1.4 Max. Flow
2
m3/h Unit ↑...^
888888888 Upp.Limit
00600.0000 or other value
3
/h ↑...↑...^
m
Display Unit / Display %Max Flow ↑...^
3
m3/h norm Unit ↑...^
00600.0000 or other value
3
/h norm ↑...↑...^
m
Display Unit / Display %Max Flow ↑...^
4
kg/h Unit
00600.0000 or other value kg/h ......^
Display Unit / Display %Max Flow ↑...^
1.1.5 Min. Flow
1.1.6 Timeconst.
1 only with gas 2 available if Maes.Inst = Volume (see menu item 1.1.3) 3 available if Maes.Inst = Norm. Vol. (see menu item 1.1.3) 4 available if Maes.Inst = Mass (see menu item 1.1.3)
001111111
00300.0000
3
/h ^
m
00002.0000 s ......^
Select menu language
Enter the name of the location (max. 10 characters)
Select type of flow measurement
Set maximum flow rate
Select unit for volume flow measurement
Enter value for the maximum volume flow
Display flow rate in units / in % of maximum flow rate
Select unit for standard volume flow measurement
Enter value for the maximum standard volume flow rate
Display flow rate in units / in % of maximum flow rate
Select unit for mass flow measurement
Enter value for the maximum mass flow rate
Display flow rate in units / in % of maximum flow rate
The minimum flow rate value is factory-set.
Enter the time constant for the measured value output in seconds (0–20s) 0: disabled
OPERATION 6
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
49
Page 50
6 OPERATION
6.5.5 Menu item Tests
Level Designation Selection / entry Explanation
SITRANS FX300
2.1.1 Test I
2.1.2 Test P
Table 6-10: Menu item Test
4 mA ^ Test current output
8 mA ^
12 mA ^
16 mA ^
20 mA ^
0.5003 Hz ^ Test pulse output
Every current value has to be confirmed with ^. Only after this is the current output set to the respective value. The following is then displayed:
"Contin. Yes":
"Contin. Yes": continue with the test of current output function
"Contin. Yes":"Contin. Yes":
"Contin. No":
"Contin. No": back to menu
"Contin. No":"Contin. No":
The test current is maintained until the menu has been exited.
50
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 51
SITRANS FX300
6.5.6 Menu item Setup (firmware version - basic)
Level Designation Selection / entry Explanation
3.1 Set display
3.1.1 Error Msg.
3.1.2 Cycle.Disp.
3.1.3 Code 1
Yes ↑... Shows error messages in measuring mode in plain text
No ↑...^ Flashing cursor in the top left corner of the display
Yes ↑... Cyclically display measured values in measuring mode
No ↑...^ Do not display measured values cyclically
Yes ↑... Activate access code:
No ↑...^ Deactivate access code
Display error
alternating with the measured values
indicates errors.
Display measured values
(6 s intervals)
Set security query 1 (menu)
→→→↑↑↑^^^
OPERATION 6
3.2 I/O settings
3.2.1 4-20mA off
3.2.5 Function P
3.2.7 Totalizer
3.2.8 Tot. Conf.
(available if Maes.Inst = Volume, see menu item
1.1.3)
(available if Maes.Inst = Norm. Vol., see menu item 1.1.3)
4-20mA ↑... Current output 4-20 mA 4-20/22E ↑... Current output 4-20 mA or error current 22 mA 4-20/3.55E ↑...^ Current output 4-20 mA or error current 3.55 mA
Yes ↑... Activate pulse output No ↑...^ Deactivate pulse output
Tot. on ↑... Start totalizer Tot. off ↑...^ Stop totalizer (value is held)
3
m Unit ↑...^
0000000000
3
......^
m
Reset yes / Reset no ↑...^
Disp. off / Disp. on ↑...^
m3 norm Unit ↑...^
0000000000
3
norm ↑...↑...^
m
Reset yes / Reset no ↑...^
Disp. off / Disp. on ↑...^
Current output 4-20 mA and error current
Pulse output
Totalizer
Configure totalizer
Select totalizer unit for volume flow measurement
Enter preset totalizer value
Reset totalizer / Do not reset totalizer/
Display totalizer / Do not display totalizer
Select totalizer unit for standard volume flow measurement
Enter preset totalizer value
Reset totalizer / Do not reset totalizer/
Display totalizer / Do not display totalizer
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
51
Page 52
6 OPERATION
Level Designation Selection / entry Explanation
SITRANS FX300
(available if Maes.Inst = Mass, see menu item
1.1.3)
3.3 HART settings
3.3.1 Poll. Adr.
3.3.2 HART SV
3.3.3 HART TV
3.3.4 HART 4V
3.4 Set fluid and medium
3.4.1 Fluid 1
3.4.2 Medium
(available if Fluid = Gas or Liquid, see menu item
3.4.1)
(available if Fluid = Steam, see menu item 3.4.1)
1 Density check
3.5 Set pressure, temperature and density
3.5.1 T-sensor
3.5.3 Sat.P/T
(verfügbar wenn FLUID = DAMPF, siehe Menüpunkt
3.4.1)
3.5.4 Temp. Opr.
kg Unit ↑...^
0000000000 kg ......^
Reset yes / Reset no ↑...^
Disp. off / Disp. on ↑...^
000 0 – 15 →↑...→↑...^
Total Flow ^ HART secondary variable
Temp. ^ HART tertiary variable
Density ^
Density ^ HART quaternary variable
Temp. ^
Gas ... Gas Liquid. ↑... Liquid Steam ...^ Steam
Custom ... Customer medium
Sat.Steam ...^ Saturated steam
Custom
No ... No temperature sensor available Yes ...^ Temperature sensor available
Sat. T ^ Density calculation for saturated steam with the
°C Unit ...^
0000000.0 °C ^
Disp. on / Disp. off ↑...^
Select totalizer unit for mass flow measurement
Enter preset totalizer value
Reset totalizer / Do not reset totalizer/
Display totalizer / Do not display totalizer
Enter HART polling address for "Multidrop operation"
Set fluid type
Set medium
Internal temperature sensor
Density calculation for saturated steam
saturated temperature
Operating temperature
Set temperature unit
Operating temperature
Display temperature / Do not display temperature
52
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 53
SITRANS FX300
Level Designation Selection / entry Explanation
OPERATION 6
3.5.6 Dens. Opr.
3.5.9 Dens. Norm 2
3.6
3.6.1 Remote 0.000
3.6.2 Cable Coef 14.074
1 Disp. off Disp. on 2 available if Maes.Inst = Norm.Vol.
kg/m Unit ...^
00000.0000 kg/m
00000.0000 kg/m
length/m
Cable Coef
3
3
^ 1
3
......^
Density at operating pressure and operating temperature
Set density unit
Operating density
Enter density for reference conditions (pressure and temperature)
(max. = 30m)
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
53
Page 54
6 OPERATION
6.5.7 Menu item Setup (firmware version - steam)
Level Designation Selection / entry Explanation
3.1 Set display
3.1.1 Error Msg.
3.1.2 Cycle.Disp.
3.1.3 Code 1
Yes ↑... Shows error messages in measuring mode in plain text
No ↑...^ Flashing cursor in the top left corner of the display
Yes ↑... Cyclically display measured values in measuring mode
No ↑...^ Do not display measured values cyclically
Yes ↑... Activate access code:
No ↑...^ Deactivate access code
Display error
alternating with the measured values
indicates errors.
Display measured values
(6 s intervals)
Set security query 1 (menu)
→→→↑↑↑^^^
SITRANS FX300
3.2 I/O settings
3.2.1 4-20mA off
3.2.2 Variable I
(available if Meter Type = Gross Heat, see menu item 5.4.1, service menu)
3.2.3 Power Unit
(available if Meter Type = Gross Heat, see menu item 5.4.1, service menu)
3.2.4 FS power
(available if Variable I = Power, see menu item
3.2.2)
3.2.5 Function P
3.2.6 Variable P
(available if Meter Type = Gross Heat, see menu item 5.4.1, service menu)
3.2.7 Totalizer
3.2.8 Tot. Conf.
4-20mA ↑... Current output 4-20 mA 4-20/22E ... Current output 4-20 mA or error current 22 mA 4-20/3.55E ↑...^ Current output 4-20 mA or error current 3.55 mA
Maes.Inst ... Flow rate Power ...^ Power
kJ/h Unit ↑...^
Disp. on / Disp. off ↑...^
7000000.00 kJ/h ^
Yes ↑... Activate pulse output No ↑...^ Deactivate pulse output
Total Flow ↑... Total flow Energy ...^ Energy
Tot. on ↑... Start totalizer Tot. off ↑...^ Stop totalizer (value is held)
Current output 4-20 mA and error current
Set output variable for current output
Set unit for heat output
Select unit for measuring heat output
Display measured value/ Do not display measured value
End value for power measuring range
Power value at 20 mA current output
Function P
Set output variable for the pulse output
Totalizer
Configure totalizer
54
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 55
SITRANS FX300
Level Designation Selection / entry Explanation
(available if Maes.Inst = Volume, see menu item
1.1.3)
(available if Maes.Inst = Norm. Vol., see menu item 1.1.3)
(available if Maes.Inst = Mass, see menu item
1.1.3)
3.2.9 E.tot. on
(available if Meter Type = Gross Heat, see menu item 5.4.1, service menu)
3.2.10 Energ.Unit
(available if Meter Type = Gross Heat, see menu item 5.4.1, service menu)
3
m Unit ↑...^
0000000000
3
......^
m
Reset Yes / Reset No ↑...^
Disp. off / Disp. on ↑...^
m3 norm Unit ↑...^
0000000000
3
norm ↑...↑...^
m
Reset Yes / Reset No ↑...^
Disp. off / Disp. on ↑...^
kg Unit ↑...^
0000000000 kg ......^
Reset Yes / Reset No ↑...^
Disp. off / Disp. on ↑...^
Tot. on ↑... Start totalizer Tot. off ↑...^ Stop totalizer (value is held)
kJ Unit ↑...^
0000000000 kJ ↑...↑...^
Reset No / Reset Yes ↑...^
Disp. on / Disp. off ↑...^
Select totalizer unit for volume flow measurement
Enter preset totalizer value
Do not reset counter/ reset counter
Do not display totalizer Display totalizer /
Select totalizer unit for standard volume flow measurement
Enter preset totalizer value
Do not reset totalizer/ Reset totalizer
Do not display totalizer Display totalizer /
Select totalizer unit for mass flow measurement
Enter preset totalizer value
Do not reset totalizer/ Reset totalizer
Do not display totalizer Display totalizer /
Energy totalizer
Set unit for heat energy
Set unit for heat energy
Enter preset totalizer value
Do not reset heat energy/ Reset heat energy
Display heat energy / Do not display heat energy
OPERATION 6
3.3 HART settings
3.3.1 Poll. Adr.
3.3.2 HART SV
000 0 – 15 →↑...→↑...^
Total Flow ↑... Energy ↑...^
#available if Meter Type = Gross Heat (menu item 5.3.1, service menü)
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Enter HART polling address for "Multidrop operation"
HART secondary variable
55
Page 56
6 OPERATION
Level Designation Selection / entry Explanation
SITRANS FX300
3.3.3 HART TV
3.3.4 HART 4V
3.4 Set fluid and medium
3.4.1 Fluid 1
3.4.2 Medium
3.4.5 Dry. Fact.
1 Density check
3.5 Pressure, temperature and density
3.5.1 T-sensor
3.5.2 P-sensor
3.5.3 Sat.P/T
3.5.4 Temp. Opr.
3.5.5 Pres. Opr.
(available if Medium = Sat.Steam, see menu item
3.4.2)
(available if Medium = Sat.Steam, see menu item
3.4.2)
(for saturated steam with Sat. Temp. or Sup.Steam, see menu items 3.4.2 and
3.5.3)
(available for Sat.Steam with Sat. P or Sup.Steam, see menu items 3.4.2 and
3.5.3)
Temp. ... Pressure ... Density ↑...^
Temp. ... Pressure ... Density ↑...^
Steam ^ Steam
Sat.Steam ... Saturated steam Sup.Steam ...^ Overheated steam
0000001.00
0.85 TO 1 ↑... ...^
No ... No temperature sensor available Yes ↑...^ Temperature sensor available
Intern ... Internal pressure sensor
--- ↑...^ No pressure sensor available
Sat. Temp. ↑... Density calculation for saturated steam from saturated
Sat. Pres. ↑...^ Density calculation for saturated steam from saturated
°C Unit ↑...^
0000000.0 °C ^
Disp. on / Disp. off ↑...^
Pa Unit ↑...^
0000000.0 Pa ^
Disp. on / Disp. off ↑...^
HART tertiary variable
HART quaternary variable
Fluid type
Process medium
Enter portion of dryness in saturated steam
Factor = 1 - %water mass
Internal temperature sensor
Pressure sensor
Density calculation for saturated steam
temperature
pressure
Operating temperature
Set temperature unit
Operating temperature
Display temperature / Do not display temperature
Operating pressure
Set pressure unit
Operating pressure
Display pressure / Do not display pressure
56
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 57
SITRANS FX300
Level Designation Selection / entry Explanation
OPERATION 6
3.5.6 Dens. Opr.
kg/m Unit ...^
00000.0000 kg/m
3.5.10 P-Excit. V
0005.00000
V ^ (available if P-Sensor = Internal, see menu item
3.5.2)
3.5.11 P-Sen.P1V1
(available if P-Sensor = Internal, see menu item
3.5.2)
0001.00000
P1 kg/cm
...^
0002.00000
V1 mV ↑...↑...^
3.5.12 P-Sen.P2V2
(available if P-Sensor = Internal, see menu item
3.5.2)
0005.00000
P2 kg/cm
...^
0048.00048
V2 mV ↑...↑...^
3.6
3.6.1 Remote 0.000 length/m
3.6.2 Cable Coef 14,074 Cable Coef
3
3
^ 1
2
g ↑...
2
g ↑...
Density at operating pressure and operating temperature
Set density unit
Operating density
Exciting voltage of pressure sensor
Pressure sensor: Enter 1st calibration point
Pressure sensor: Enter 2nd calibration point
(max. = 30m)
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
57
Page 58
6 OPERATION
6.5.8 Menu item Setup (firmware version - gas)
Level Designation Selection / entry Explanation
3.1 Set display
3.1.1 Error message
3.1.2 Cycle.Disp.
3.1.3 Code 1
Yes ↑... Shows error messages in measuring mode in plain text
No ↑...^ Flashing cursor in the top left corner of the display
Yes ↑... Cyclically display measured values in measuring mode
No ↑...^ Do not display measured values cyclically
Yes ↑... Activate access code:
No ↑...^ Deactivate access code
Display error
alternating with the measured values
indicates errors.
Display measured values
(6 s intervals)
Set security query 1 (menu)
→→→↑↑↑^^^
SITRANS FX300
3.2 I/O settings
3.2.1 4-20mA off
3.2.5 Function P
3.2.7 Totalizer
3.2.8 Tot. Conf.
(available if Maes.Inst = Volume, see menu item
1.1.3)
(available if Maes.Inst = Norm. Vol., see menu item 1.1.3
4-20mA ↑... Current output 4-20 mA 4-20/22E ... Current output 4-20 mA or error current 22 mA 4-20/3.55E ↑...^ Current output 4-20 mA or error current 3.55 mA
Yes ↑... Activate pulse output No ↑...^ Deactivate pulse output
Tot. on ↑... Start totalizer Tot. off ↑...^ Stop totalizer (value is held)
3
m Unit ↑...^
0000000000
3
......^
m
Reset Yes / Reset No ↑...^
Disp. off / Disp. on ↑...^
m3 norm Unit ↑...^
0000000000
3
norm ↑...↑...^
m
Reset yes / Reset no ↑...^
Disp. off / Disp. on ↑...^
Current output 4-20 mA and error current
Pulse output
Totalizer
Configure totalizer
Select totalizer unit for volume flow measurement
Enter preset totalizer value
Reset totalizer Do not reset totalizer/
Do not display totalizer Display totalizer /
Select totalizer unit for standard volume flow measurement
Enter preset totalizer value
Reset totalizer Do not reset totalizer/
Do not display totalizer Display totalizer /
58
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 59
SITRANS FX300
Level Designation Selection / entry Explanation
OPERATION 6
(available if Maes.Inst = Mass, see menu item
1.1.3)
3.3 HART settings
3.3.1 Poll. Addr..
3.3.2 HART SV
3.3.3 HART TV
3.3.4 HART 4V
kg Unit ↑...^
0000000000 kg ......^
Reset Yes / Reset No ↑...^
Disp. off / Disp. on ↑...^
000 0 – 15 →↑...→↑...^
Total Flow ↑... FAD ...^
#available if Meter Type = FAD meter (see menu item 5.3.1, service personnel only)
Temp. ... Pressure ... Density ↑... FAD ...^
#available if Meter Type = FAD meter (see menu item 5.3.1, service personnel only)
Temp. .. Pressure ... Density ↑... FAD ...^
#available if Meter Type = FAD meter (see menu item 5.3.1, service personnel only)
Select totalizer unit for mass flow measurement
Enter preset totalizer value
Reset totalizer Do not reset totalizer/
Do not display totalizer Display totalizer /
Enter HART polling address for "Multidrop mode"
HART secondary variable
HART tertiary variable
HART quaternary variable
3.4 Set fluid and medium
3.4.1 Fluid 1
1 When changing fluid: check density
Gas ... Gas Gas Mix. ↑... Gas mixture Wet Gas ↑...^ Wet gas
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Set fluid type
59
Page 60
6 OPERATION
Level Designation Selection / entry Explanation
SITRANS FX300
3.4.2 Medium
(available if Fluid = Gas / Wet Gas, see menu item
3.4.1)
3.4.3 % Gas
(available if Fluid = Gas Mix., see menu item 3.4.1)
3.4.4 %Rel.Hum .
(available if Fluid = Wet Gas, see menu item 3.4.1)
3.4.6 FAD Unit.
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.7 Suct.Temp.
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.8 Atm.Press.
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.9 Fil. P. Drop.
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.10 Inlet RH
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.11 Actual Rpm
(available if Meter Type = FAD meter, see menu item 5.3.1, service personnel only)
3.4.12 Rated Rpm.
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
3.4.13 Outlet RH
(available if Meter Type = FAD meter, see menu item 5.4.1, service personnel only)
Set process medium
Air ... Air Ammonia ... Ammonia ... Argon ... Argon etc. ↑...^ other gases not listed here
Determine composition of gas mixture
050.000000 Air ......^
100.00 Total % ^
0000000.00 % Rel.Hum ↑... ...^
FAD m3/h Unit ↑...^
Disp. on / Disp. off ↑...^
°C Unit ...^
0000200.00 °C ↑......^
Pa Unit ↑...^
00001.0000 Pa ......^
Pa Unit ...^
00000.0000 Pa ......^
0000060.00 % Rel.Hum ↑... ...^
0001500.00 RPM ^
0001500.00 RPM ^
0000100.00 % Rel.Hum ↑... ...^
Select gas and enter percentage
Total percentage
Determine relative humidity
Determine unit for FAD
Set unit for FAD
Display measured value / Do not display measured value
Enter temperature in suction area of compressor
Select temperature unit
Enter temperature value
Atmospheric pressure
Select pressure unit
Enter pressure value
Pressure drop via filter at compressor inlet
Select pressure unit
Enter pressure value
Enter relative humidity in suction area of compressor
Current speed of compressor motor in revolutions per minute
Rated speed of compressor motor in revolutions per minute
Enter relative humidity at measuring device (compressor outlet)
60
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 61
SITRANS FX300
OPERATION 6
Level Designation Selection / entry Explanation
3.5 Pressure, temperature and density
3.5.1 T-sensor
No ... No temperature sensor available Yes ↑...^ Temperature sensor available
3.5.2 P-sensor
Intern ... Internal pressure sensor
--- ↑...^
3.5.4 Temp. Opr.
°C Unit ↑...^
0000000.0 °C ^
Disp. on / Disp. off ↑...^
3.5.5 Pres. Opr.
Pa Unit ...^
0000000.0 Pa ^
Disp. on / Disp. off ↑...^
3.5.6 Dens. Opr. Density at operating pressure and operating
3
kg/m Unit ↑...^
00011.0000
3
^
kg/m
3.5.7 Temp.Norm .
(available if Maes.Inst =
00000020.0 °C ↑......^
Norm. Vol., see menu item 1.1.3)
3.5.8 Pres.Norm
(available if Maes.Inst =
00000000.0 Pa ......^
Norm. Vol., see menu item 1.1.3)
3.5.9 Dens. norm
(for unknown gases using standard volume flow
00001.2900
3
^
kg/m
measurement, see menu items 1.1.3, 3.4.1 and
3.4.2)
3.5.10 P-Excit. V
(available if P-Sensor =
0005.00000 V ^
Internal, see menu item
3.5.2)
3.5.11 P-Sen.P1V1
(available if P-Sensor = Internal, see menu item
3.5.2)
0001.00000 P1 kg/cm
2
g ↑...
...^
0002.00000 V1 mV ↑...↑...^
Internal temperature sensor
Pressure sensor
No pressure sensor available
Operating temperature
Set temperature unit
Operating temperature
Display temperature / Do not display temperature
Operating pressure
Set pressure unit
Operating pressure
Display pressure / Do not display pressure
temperature
Set density unit
Operating density
Enter normal temperature Unit as 3.5.4
Enter normal pressure Unit as 3.5.5
Enter density for reference conditions (pressure and temperature)
Exciting voltage of pressure sensor
Pressure sensor: 1st calibration point
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
61
Page 62
6 OPERATION
Level Designation Selection / entry Explanation
SITRANS FX300
3.5.12 P-Sen.P2V2
3.6
3.6.1 Remote 0.000
3.6.2 Cable coeff 14.074
(available if P-Sensor = Internal, see menu item
3.5.2)
0005.00000 P2 kg/cm
...^
0048.00048 V2 mV ↑...↑...^
length/m
Cable Coef
2
g ↑...
Pressure sensor: 2nd calibration point
(max. = 30m)
62
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 63
SITRANS FX300
7.1 Replacing signal converter / LCD display
The signal converter has to be replaced by a converter of the same type. The following parameters have to be observed for it:
The item number must match: 2.143670.xxx
The software version 2 must match.
The basic version has no software identification
The gas version is marked: "gas"
The steam version is marked: "steam"
SERVICE 7
Figure 7-1: Board, example
1 Connector for display 2 Version number, software identification 3 Connector for measuring sensor 4 Spacer pins 5 Fixing screw (Phillips)
The following steps are to be performed:
Switch off the power supply.
Unscrew the front cover.
Detach the display from the spacer pins.
Detach the display cable 1.
Detach the measuring sensor cable 3.
Loosen the three fixing screws 5 (Phillips).
Pull out signal converter.
Insert new signal converter.
Tighten the three screws.
Attach measuring sensor cable 3 (Caution: ensure that the measuring sensor cables are not
on top of the spacer pins 4 May be damaged).
Attach display cable 1.
Attach display in desired position, apply even pressure to the entire surface.
Screw on cover by hand.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
63
Page 64
7 SERVICE
7.2 Spare parts availability
The manufacturer adheres to the basic principle that operational spare parts for each device or each important accessory part will be kept available for a period of 10 (ten) years after delivery of the last production run for that device.
Operational spare parts are defined as parts that are subject to faults in normal operation.
7.3 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, technical support and training.
INFORMATION!
For more precise information, please contact your local representative.
7.4 Returning the device to the manufacturer
SITRANS FX300
7.4.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
64
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 65
SITRANS FX300
7.4.2 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
SERVICE 7
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
Returning adress: Siemens Flow Instruments A/S Nordborgvej 81 DK - 6430 Nordborg Att. Quality department E1-Ø21
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
65
Page 66
8 TECHNICAL DATA
8.1 Functional principle
Vortex flowmeters are used to measure the volumetric flow of gases, vapours and liquids at completly filled pipes. The measuring principle is based on the Karman vortex street. The measuring tube contains a bluff body, behind which vortex shedding occurs. The frequency ffff of the vortex shedding is proportional to the flow rate vvvv. The nondimensional Strouhal number SSSS describes the relationship between vortex frequency ffff, width bbbb of bluff body and the mean flow velocity vvvv:
S.v
f
=
b
The vortex frequency is recorded at the sensor and evaluated at the converter.
SITRANS FX300
Figure 8-1: Functional principle
66
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 67
SITRANS FX300
8.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website .
Measuring system
Application range Flow measurement of liquids, gases and vapours
Operating method / measuring principle
Measured value
Measured value
Measured valueMeasured value
Primary measured value Number of separated vortices
Secondary measured value Operating and standard volumetric flow, mass flow
Signal converter
Signal converter
Signal converterSignal converter
Versions Compact
Measuring sensor
Measuring sensor
Measuring sensorMeasuring sensor
Standard Basic device in flange version (with integrated temperature
Optional Basic device with additional pressure measurement
Display and user interface
Display and user interface
Display and user interfaceDisplay and user interface
Local display 2 lines, 10 characters per line
Operating and display languages German, English, French
TECHNICAL DATA 8
Karman vortex street
Option: Ex version
measurement)
Basic device in sandwich version (with integrated temperature measurement)
Basic device with additional pressure measurement and shut-off valve for pressure sensor
Dual measuring device in both flange and sandwich version (redundant measurement)
Dual measuring device in flange version with additional pressure measurement
Measuring accuracy
Reference condition Water at 20ºC
Air at 20°C and 1.013 bar abs.
Accuracy Based on volume flow Liquids ±0.75% of measured value (Re 20000)
±2.0% of measured value (10000 < Re < 20000)
Gases and steams ±1.0% of measured value (Re 20000)
±2.0% of measured value (10000 < Re < 20000)
Pressure and temperature compensation: ±20000% of measured value (Re 20000); ±10000% of measured value (20000 < Re < 20000)
Repeatability ±0.1% of measured value
Long-term stability ±0.01% of measured value
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
67
Page 68
8 TECHNICAL DATA
Operating conditions
Temperature
Temperature
TemperatureTemperature
Product -40+240°C / -40+465°F
Ambient Non-Ex: -40+85°C / -40+185°F
Storage -50+85°C / -58+185°F
Pressure
Pressure
PressurePressure
Product Max. 100 bar / 1450 psi; Information on higher pressures on request.
Ambient Atmosphere
Chemical properties
Chemical properties
Chemical propertiesChemical properties
Density Taken into consideration when sizing.
Viscosity < 10 cP
Reynold's number 10000...2300000
Recommended flow velocities
Recommended flow velocities
Recommended flow velocitiesRecommended flow velocities
Liquids 0.37m/s / 0.98…23 ft/s (optional up to 10 m/s / 32.8 ft/s taking
Gases and steams 2.080 m/s / 6.6262.5 ft/s
Other conditions
Other conditions
Other conditionsOther conditions
Protection category IP 66/67
SITRANS FX300
Ex: -40…+60°C / -40…+140°F
cavitation into account)
DN15: 3.045 m/s / 9.8148 ft/s; DN25: 2.070 m/s / 6.6…230 ft/s
For detailed information, see chapter "Flow tables".
Installation conditions
Inlet run 20 x DN (without disturbing flow, after pipe narrowing, after a single
Outlet run 5 x DN
Dimensions and weights For detailed information refer to chapter "Dimensions and weights".
90° bend)
≥ 30 x DN (after a double bend 2x90°) ≥ 40 x DN (after a double three-dimensional bend 2x90°) ≥ 50 x DN (after control valves) ≥ 2 DN before flow straightener; ≥ 8 DN after flow straightener
(specified values apply only to original 20 DN inlet run)
Materials
Measuring sensor and process connections
Converter housing Die-cast Aluminium
Pressure sensor gasket Standard: FPM
Measuring tube gasket Standard: 1.4435/316L
Standard: 1.4404/316L
Option: Hastelloy® C-22 on request
Option: FFKM
Option: Hastelloy® C-276
Selection depends on measuring sensor material/medium.
68
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 69
SITRANS FX300
Process connections
Flange version
Flange version
Flange versionFlange version
DIN EN 1092-1 DN15...300 in PN16…100
ASME B16.5 ½...12" in 150…600 lbs
JIS B 2220 DN15...300 in JIS 10…20 K
Sandwich version
Sandwich version
Sandwich versionSandwich version
DIN DN15...100 in PN100 (higher pressures on request)
ASME ½...4" in 600 lbs (higher pressures on request)
JIS DN15...100 in 1020 K (higher pressures on request)
Electrical connections
Power supply Non-Ex: 14…36 VDC
Current output
Current output
Current outputCurrent output
Description of the used abbreviations
Measuring range
Load Minimum 0 Ω; maximum RL = ((Uext. - 14 VDC) / 22 mA)
Error signal Acc. to NAMUR NE43
TECHNICAL DATA 8
For detailed information on combination flange/pressure rating, refer to chapter "Dimensions and weights".
Ex: 14…30 VDC
U
= external voltage; RL = load + resistance
ext
4...20 mA (max. 20.8 mA) + HART® protocol
Upper value: 21.0 mA
Lower value: 3.6 mA (not with HART® protocol)
Pulse output
Pulse output
Pulse outputPulse output
Pulse rate Max. 0.5 pulse/s (corresponds to 1800 pulses/hour)
Power supply Non-Ex: 24 VDC as NAMUR or open < 1 mA, maximum 36 V, closed
®
HART
HART
HARTHART
Device revision 1
Physical layer FSK
Device category Transmitter, galvanically isolated System requirements Load min. 250 Ω
Multidrop operation 4mA
100 mA, U < 2 V
Ex: 24 VDC as NAMUR or open < 1 mA, maximum 30 V, closed 100 mA, U < 2 V
HART® protocol via current output
Approvals and certifications
ATEX ATEX II 2G EEx d ia [ia] IIC T6
FM Class I, II, III
Div 1/2, groups A-G
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
69
Page 70
8 TECHNICAL DATA
8.3 Dimensions and weights
8.3.1 Flange versions
SITRANS FX300
Option: version with 2 transmitters
a = 133 mm / 5.24" b = 105 mm / 4.13"
c = 179 mm / 7.05"
Dimension H x 2 Specified weight + 2.80 kg
Flange version EN 1092-1
Size Pressure
rating
DN PN d D L H l With Without
15 40 17.3 95 200 265 174.3 6.1 5.5
15 100 17.3 105 200 265 174.3 7.1 6.5
25 40 28.5 115 200 265 174.3 7.9 7.3
25 100 28.5 140 200 265 174.3 9.9 9.3
40 40 43.1 150 200 270 174.5 10.8 10.2
40 100 42.5 170 200 270 174.5 14.8 14.2
50 16 54.5 165 200 275 174.5 12.7 12.1
50 40 54.5 165 200 275 174.5 12.9 12.3
50 63 54.5 180 200 275 174.5 16.9 16.3
50 100 53.9 195 200 275 174.5 18.4 17.8
80 16 82.5 200 200 290 174.25 17.4 16.8
80 40 82.5 200 200 290 174.25 19.4 18.8
80 63 81.7 215 200 290 174.25 23.4 22.8
80 100 80.9 230 200 290 174.25 27.4 26.8
100 16 107.1 220 250 310 174.5 22 21.4
100 40 107.1 235 250 310 174.5 25 24.4
100 63 106.3 250 250 310 174.5 30 29.4
Dimensions [mm] Weight [kg]
pressure sensor
70
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 71
SITRANS FX300
TECHNICAL DATA 8
Size Pressure
Dimensions [mm] Weight [kg]
rating
DN PN d D L H l With Without
pressure sensor
100 100 104.3 265 250 310 174.5 36 35.4
150 16 159.3 285 300 325 196.5 35.8 35.2
150 40 159.3 300 300 325 196.5 41.8 41.2
150 63 157.1 345 300 325 196.5 59.8 59.2
150 100 154.1 355 300 325 196.5 67.8 67.2
200 10 206.5 340 300 350 208.5 38.4 37.8
200 16 206.5 340 300 350 208.5 38.4 37.8
200 25 206.5 360 300 350 208.5 47.4 46.8
200 40 206.5 375 300 350 208.5 55.4 54.8
250 10 260.4 395 380 370 236.5 58.0 57.4
250 16 260.4 405 380 370 236.5 59.0 58.4
250 25 258.8 425 380 370 236.5 75.0 74.4
250 40 258.8 450 380 370 236.5 93.0 92.4
300 10 309.7 445 450 395 261 76.3 75.7
300 16 309.7 460 450 395 261 82.8 82.2
300 25 307.9 485 450 395 261 99.3 98.7
300 40 307.9 515 450 395 261 128.1 127.5
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
71
Page 72
8 TECHNICAL DATA
a = 133 mm / 5.24" b = 105 mm / 4.13"
Flange version ASME B16.5
SITRANS FX300
c = 179 mm / 7.05"
Size Pressure
Dimensions [mm / inches] Weight [kg / lbs]
rating
NPS Class d D L H l With Without
pressure sensor
½ 150 15.8 / 0.6 90 / 3.5 200 / 7.9 265 / 10.4 174.3 / 6.9 5.1 / 11.2 4.5 / 9.9
½ 300 15.8 / 0.6 95 / 3.7 200 / 7.9 265 / 10.4 174.3 / 6.9 5.5 / 12.1 4.9 / 10.8
½ 600 13.9 / 0.5 95 / 3.7 200 / 7.9 265 / 10.4 174.3 / 6.9 5.7 / 12.6 5.1 / 11.2
1 150 26.6 / 1.1 110 / 4.3 200 / 7.9 265 / 10.4 174.3 / 6.9 6.8 / 15.0 6.2 / 13.7
1 300 26.6 / 1.1 125 / 4.9 200 / 7.9 265 / 10.4 174.3 / 6.9 7.8 / 17.2 7.2 / 15.9
1 600 24.3 / 1.0 125 / 4.9 200 / 7.9 265 / 10.4 174.3 / 6.9 8.1 / 17.9 7.5 / 16.5
1½ 150 40.9 / 1.6 125 / 4.9 200 / 7.9 270 / 10.6 174.3 / 6.9 8.9 / 19.9 8.3 / 18.3
1½ 300 40.9 / 1.6 155 / 6.1 200 / 7.9 270 / 10.6 174.3 / 6.9 11 / 24.3 10.4 / 22.9
1½ 600 38.1 / 1.5 155 / 6.1 200 / 7.9 270 / 10.6 174.3 / 6.9 12 / 26.5 11.4 / 25.1
2 150 52.6 / 2.1 150 / 5.9 200 / 7.9 275 / 10.8 174.3 / 6.9 11.6 / 25.6 11 / 24.3
2 300 52.6 / 2.1 165 / 6.5 200 / 7.9 275 / 10.8 174.3 / 6.9 13 / 28.7 12.4 / 27.4
2 600 49.3 / 1.9 165 / 6.5 200 / 7.9 275 / 10.8 174.3 / 6.9 14.5 / 32.0 13.9 / 30.6
3 150 78 / 3.1 190 / 7.5 200 / 7.9 290 / 11.4 174.5 / 6.9 20.4 / 45.0 19.8 / 43.7
3 300 78 / 3.1 210 / 8.3 200 / 7.9 290 / 11.4 174.5 / 6.9 23.4 / 51.6 22.8 / 50.2
3 600 73.7 / 2.9 210 / 8.3 200 / 7.9 290 / 11.4 174.5 / 6.9 24.4 / 52.8 23.8 / 52.5
4 150 102.4 / 4.0 230 / 9.1 250 / 9.8 310 / 12.2 176.5 / 7.0 24 / 52.9 23.4 / 51.6
4 300 102.4 / 4.0 255 / 10.0 250 / 9.8 310 / 12.2 176.5 / 7.0 32 / 70.6 31.4 / 69.2
4 600 97.2 / 3.8 275 / 10.8 250 / 9.8 310 / 12.2 176.5 / 7.0 41 / 90.4 40.4 / 89.1
6 150 154.2 / 6.1 280 / 11.0 300 / 11.8 325 / 12.8 196.5 / 7.7 36.8 / 81.1 36.2 / 79.8
6 300 154.2 / 6.1 320 / 12.6 300 / 11.8 325 / 12.8 196.5 / 7.7 51.8 / 114.2 51.2 / 112.9
6 600 146.3 / 5.8 355 / 14.0 300 / 11.8 325 / 12.8 196.5 / 7.7 76.8 / 169.3 46.2 / 101.9
8 150 202.7 / 8.0 345 / 13.6 300 / 11.8 350 / 13.8 208.5 / 8.2 50.6 / 146.4 50.0 / 145.7
8 300 202.7 / 8.0 380 / 15.0 300 / 11.8 350 / 13.8 208.5 / 8.2 75.4 / 190.4 74.8 / 189.7
10 150 254.5 / 10.0 405 / 15.5 380 / 15.0 370 / 14.6 236.5 / 9.3 75.0 / 197.1 74.4 / 195.8
10 300 254.5 / 10.0 455 / 17.9 380 / 15.0 370 / 14.6 236.5 / 9.3 107.0 / 252.2 106.4 / 239.9
12 150 304.8 / 1 485 / 19.1 450 / 17.7 395 / 15.6 261 / 10.3 106.9 / 318.3 106.3 / 317.0
12 300 304.8 / 1 520 / 20.5 450 / 17.7 395 / 15.6 261 / 10.3 151.9 / 415.4 151.3 / 414.0
72
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 73
SITRANS FX300
8.3.2 Sandwich version
TECHNICAL DATA 8
a = 133 mm / 5.24" b = 105 mm / 4.13"
c = 179 mm / 7.05"
Dimension H x 2 Specified weight + 2.80 kg
Sandwich version EN
Size Pressure
rating
DN PN d D L H l With Without
15 100 16 45 65 265 174.25 4.1 3.5
25 100 24 65 65 265 174.25 4.9 4.3
40 100 38 82 65 270 174.5 5.5 4.9
50 100 50 102 65 275 174.5 6.6 6
80 100 74 135 65 290 174.25 8.8 8.2
100 100 97 158 65 310 176.5 10.1 9.5
Dimensions [mm] Weight [kg]
pressure sensor
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
73
Page 74
8 TECHNICAL DATA
a = 133 mm / 5.24" b = 105 mm / 4.13"
c = 179 mm / 7.05"
Sandwich version ASME
SITRANS FX300
Size Pressure
Dimensions [inches] Weight [lbs]
rating
NPS Class d D L H l With Without
pressure sensor
½ 150 0.63 1.77 2.56 10.43 6.82 9.04 7.72
½ 300 0.63 1.77 2.56 10.43 6.82 9.04 7.72
½ 600 0.55 1.77 2.56 10.43 6.82 9.04 7.72
1 150 0.94 2.56 2.56 10.43 6.82 10.8 9.48
1 300 0.94 2.56 2.56 10.43 6.82 10.8 9.48
1 600 0.94 2.56 2.56 10.43 6.82 10.8 9.48
1½ 150 1.5 3.23 2.56 10.63 6.87 12.13 10.8
1½ 300 1.5 3.23 2.56 10.63 6.87 12.13 10.8
1½ 600 1.5 3.23 2.56 10.63 6.87 12.13 10.8
2 150 1.97 4.02 2.56 10.83 6.87 14.55 13.23
2 300 1.97 4.02 2.56 10.83 6.87 14.55 13.23
2 600 1.97 4.02 2.56 10.83 6.87 14.55 13.23
3 150 2.91 5.31 2.56 11.42 6.82 19.4 18.08
3 300 2.91 5.31 2.56 11.42 6.82 19.4 18.08
3 600 2.91 5.31 2.56 11.42 6.82 19.4 18.08
4 150 3.82 6.22 2.56 12.21 6.95 22.27 20.94
4 300 3.82 6.22 2.56 12.21 6.95 22.27 20.94
4 600 3.82 6.22 2.56 12.21 6.95 22.27 20.94
74
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 75
SITRANS FX300
8.4 Flow tables
Measuring ranges
TECHNICAL DATA 8
EN 1092-1
Water
Air
DN
Size Q
NPS
min
[m3/h]
Q
max
Q
min
[gph]
ASME B16.5
15 ½ 0.45 5.0 119 1321
25 1 0.81 11.40 214 3012
40 1½ 2.04 28.57 539 7547
50 2 3.53 49.47 933 13069
80 3 7.74 108.37 2045 28629
100 4 13.30 186.21 3514 49192
150 6 30.13 421.86 7960 111445
200 8 52.66 792.42 13911 209335
250 10 90.5 1266.8 23908 334653
300 12 113.41 1839.8 29960 486024
Values based on water at 20°C / 68°F
15 ½ 6.79 32.56 1794 8602
25 1 10.20 114.0 2695 30116
40 1½ 25.35 326.63 6697 86287
50 2 43.89 565.49 11595 149388
80 3 96.14 1238.60 25398 327207
100 4 165.19 2128.27 43639 562236
150 6 374.23 4821.57 98862 1273738
200 8 702.95 9056.8 185700 2392553
250 10 1123.7 14478 29685 3824683
300 12 1632.1 21028 431155 55550104
Values based on air at 20°C / 68°F and 1.013 bar abs / 14.891 psig
Q
max
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
75
Page 76
8 TECHNICAL DATA
SITRANS FX300
Flow rate limits
Product Nominal sizes Minimum flow rates Maximum flow rates
EN ASME [m/s] [ft/s] [m/s] [ft/s]
Liquids DN15300 ½…12"
Gas, steam DN15300 ½…12"
1 Minimum flow rate 0.3 m/s - maximum flow rate 7 m/s 2 Minimum flow rate 0.984 ft/s - maximum flow rate 23 ft/s 3 Minimum flow rate 2 m/s - maximum flow rate 80 m/s; DN15 up to 45 m/s and DN25 up to 70 m/s 4 Minimum flow rate 6.6 ft/s - maximum flow rate 262 ft/s; DN15 up to 148 ft/s and DN25 up to 230 ft/s
0.5x(998/ρ)
6x(1.29/ρ)
0.5
1.64x(998/ρ)0.5 2
1
0.5
3
16.4x(1.29/ρ)0.5
4
7x(998/ρ)
7x(998/ρ)
0.47
0.47
23x(998/ρ)0.47 2
1
23x(998/ρ)0.47 4
3
ρ = operating density [kg/m3]
76
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 77
SITRANS FX300
Measuring range saturated steam: 1...7 bar
TECHNICAL DATA 8
Overpressure
1 3.5 5.2 7
[bar]
Density [kg/m³] 1.13498 2.4258 3.27653 4.16732
Temperature [°C] 120.6 148.2 160.4 170.6
Flow rate min. max. min. max. min. max. min. max.
DN EN
1092-1
100 4 193.14 2415.5 282.36 5162.7 328.16 6973.3 370.09 8869.2
150 6 437.56 5472.4 639.69 11696 743.45 15798 838.44 20093
200 8 821.91 10279 1201.6 21970 1396.5 29675 1574.9 37743
250 10 1313.9 16433 1920.9 35122 2232.5 47439 2517.7 60337
300 12 1908.3 23866 2789.8 51010 3242.4 68899 3656.6 87630
NPS
[kg/h] [kg/h] [kg/h] [kg/h] ASME B16.5
15 ½ 5.87 36.97 7.68 79.0 8.93 106.68 10.06 135.69
25 1 11.82 129.39 17.28 276.40 20.09 373.53 22.66 474.82
40 1½ 29.64 370.71 43.33 792.33 50.63 1070.2 56.8 1361.2
50 2 51.31 641.82 75.02 1371.8 87.19 1852.8 98.33 2356.6
80 3 112.41 1405.8 164.33 3004.7 191 4058.4 215.39 5161.8
Measuring range saturated steam: 10.5...20 bar
Overpressure
[bar]
Density [kg/m³] 5.88803 7.60297 9.31702 10.5442
Temperature [°C] 186.2 198.5 208.5 215
Flow rate min. max. min. max. min. max. min. max.
DN EN
1092-1
100 4 439.91 12238 499.9 14013 553.38 15608 588.69 16666
150 6 996.62 27725 1132.5 31747 1253.7 35359 1333.7 37756
200 8 1872.1 52079 2127.3 59634 2354.9 66419 2505.2 70921
250 10 2992.7 83254 3400.71 95333 3764.6 106180 4004.9 113380
300 12 4346.5 120920 4939.1 138460 5467.5 154210 5816.5 164660
NPS ASME B16.5
15 ½ 12.78 191.71 16.51 247.55 20.23 303.36 22.89 343.32
25 1 26.93 670.88 30.6 857.88 33.87 955.48 36.04 1201.41
40 1½ 67.51 1878.2 76.72 2150.7 84.93 2395.3 90.35 2557.7
50 2 116.89 3251.7 132.82 3723.4 147.03 4147 156.42 4428.1
80 3 256.03 7122.4 290.93 8155.8 322.06 9083.7 342.62 9699.3
10.5 14 17.5 20
[kg/h] [kg/h] [kg/h] [kg/h] [kg/h]
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
77
Page 78
8 TECHNICAL DATA
Measuring range saturated steam: 15...100 psig
SITRANS FX300
Overpressure
15 50 75 100
[psig]
Density [lbs/ft³] 0.0719 0.1497 0.2036 0.2569
Temperature [°F] 249.98 297.86 320.36 338.184
Flow rate min. max. min. max. min. max. min. max.
DN EN
1092-1
100 4 428.81 5405.7 618.51 11246 721.21 15291 810.06 19291
150 6 971.47 12246 1401.2 25478 1633.9 34642 1835.2 43703
200 8 1824.8 23004 2632.1 47859 3069.1 65072 3447.2 82092
250 10 2917.2 36774 4207.7 76508 4906.4 104030 5510.8 131230
300 12 4236.8 53410 6111.1 111120 7125.8 151080 8003.6 190600
NPS
[lbs/h] [lbs/h] [lbs/h] [lbs/h] [lbs/h] ASME B16.5
15 ½ 12.9 82.70 16.83 1720.12 19.62 234.0 22.04 295.23
25 1 26.25 289.40 37.86 602.09 44.15 818.63 49.59 1032.76
40 1½ 65.81 829.61 94.92 1726 110.68 2346.7 124.32 2960.5
50 2 113.94 1436.3 164.34 2988 191.63 4062.9 215.23 5125.6
80 3 249.57 3146.1 360 6545.3 419.74 8899.4 471.45 11227
Measuring range saturated steam: 150...300 psig
Overpressure
[psig]
Density [lbs/ft³] 0.3627 0.4681 0.5735 0.6792
Temperature [°F] 366.08 388.04 406.22 422.06
Flow rate min. max. min. max. min. max. min. max.
DN EN
1092-1
100 4 962.54 26766 1093.5 30643 1210.4 34124 1317.2 37324
150 6 2180.6 60639 2477.4 69421 2742.1 77307 2984 84556
200 8 4096.1 113900 4653.6 130400 5150.7 145210 5605.2 158830
250 10 6548.1 182090 7439.3 208460 8234.1 232140 8960.6 253910
300 12 9510.2 264460 10805 302760 11959 337150 13014 368770
NPS ASME B16.5
15 ½ 27.79 416.68 35.86 573.83 43.94 659.14 52.04 780.29
25 1 58.93 1459.16 66.94 1875.90 74.1 2089.00 80.63 2284.90
40 1½ 147.72 4107.2 167.83 4702.8 185.76 5237 202.15 5728
50 2 255.75 7111.9 290.56 8141.9 321.6 9066.8 350 9917
80 3 560.19 15578 636.44 17834 704.43 19860 766.6 21722
150 200 250 300
[lbs/h] [lbs/h] [lbs/h] [lbs/h] [lbs/h]
78
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Page 79
SITRANS FX300
NOTES 9
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
79
Page 80
Loading...