The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: RXR2-130.101.02.03.02 Doc. Gen. Date:08.97
Replaces: RXR2-130.101.02.01.02
0 - 2Revision
ChapterPageRevision
01 to 403
11 to 201
21 to 402
31 to 201
41 to 1002
51 to 202
61 to 201
71 to 202
SIREMOBIL CompactRXR2-130.101.02Page 2 of 4Siemens AG
Rev. 0308.97TD SP 1Medical Engineering
Contents0 - 3
Page
1 _______General information and overview_________________________________1 - 1
7 ______ Changes to previous version_____________________________________7 - 1
SIREMOBIL CompactRXR2-130.101.02Page 4 of 4Siemens AG
Rev. 0308.97TD SP 1Medical Engineering
General information and overview1
To ensure compliance with the applicable provisio ns of the Federal Performance Standard, the user of the equipment has to ensure that the maintenance procedure described
below is performed at intervals of si x months or less. In case the user does not comply
with these requirements, the manufactu rer can no longer be held liable.
If the results of any of the tests described in this document exceed the tolerances given,
the required adjustments must be performed.
Contact Siemens service.
After performing any adjustment, all appl icable tests must be repeated.
Customer:Customer No.:Room:
BZ-No.:Type No.:Serial No.:
1 - 1
After completion of maintenance:1
I hereby certify t hat all proce dures described i n these main tenance instruc tions have been
completed as specified and that the tolerance values are within the accepted range.
_______________________________
Date Signature
Siemens AGRXR2-130.101.02Page 1 of 2SIREMOBIL Compact
Medical Eng ineeringRev. 0111.96TD SP 1
1 - 2General information and overview
CAUTION
WARNING
1 Required reference documents1
- DHHS Test certificate 1a+ and Acceptance Test Prot ocol
(in Register 3 of the Logbook)
2 Required test equipment1
- PTW-NomexSerial-No.: ..........Date of calibration ..........
- FlukeSerial-No.: ..........Date of calibration ..........
-Calculator
3 Safety Notes1
All local safety regulations must be observed.
4 Notes for the Maintenance Instructions1
- Compliance with the applicable s ections of the Federal Performance Standards is
assured when all procedures descri bed in these instructions have been perfo rmed
and tolerance values are within the sp ecified range.
- The unit movements required for thes e procedures must be performed cautio usly to
prevent the unit from being damaged in the event of a malfunction.
Checks or adjustments to be performed with radiation ON are
marked with the radiation warning symbol .
- When performing safety checks or adjus tments with radiation ON, radiati on protection
clothing must be worn.
SIREMOBIL CompactRXR2-130.101.02Page 2 of 2Siemens AG
Rev. 0111.96TD SP 1Medical Engineering
Visual checks2
1 Radiation protection2
Make sure that there is no mechanical damage on the SIREPHOS tube housing assembly
and image intensifier which could impair the radiation protection.
SIREPHOS and image intensifier
have been checked for mechanical
damage
Damage present (brief description):
YES
NO
2 - 1
2 Checking the DHHS and Identification Labels2
Confirm that the certificat ion labels as well as serial number and model number labels are
legible. The illustrati on shows the location of the labels.
If labels are required, contact the Regulatory Affairs department in Iselin, N.J.