The reproduction, transmission or
use of th is do cu ment or its con ten ts
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
3 _______Starting up RAI I.I./TV adapter N93 ________________________________3 - 1
4 _______RAS-programming for the generator_______________________________4 - 1
Siemens AGTDRX1RX99-051.032.02Register 04RAS
Medical EngineeringRev. 0103.95
0 - 4Contents
Page
RASRegister 04RX99-051.032.02TDRX1Siemens AG
Rev. 0103.95Medical Engineering
NOTICE
NOTICE
General1
Tools and measurement devices required1
Except for the standard installation kit , the various tools and measurement devices are listed as well as specified in part 3 of the
ARTD.
Service PCaccording to spec’sARTD 3.1.0
PT Emulation96 60 960 RE999
ST 32097 04 586 Y 4364
Connection cable99 00 440 RE999
Adapter99 00 218 RE999
Meas. device for fiber optic
cables
Documentation required1
99 60 812 RE999
1 - 1
Depending on the system configuration, you need the documentat ion for the generator,
unit, or DFR system.
Safety instructions1
When performing the steps and tests described please adhere to:
- the product-specific safety instructions contained in this document
- the safety instructions RA0-000.012.29.. . in Register 2 of the
generator or system manual, and
- the general safety instructions contained in Register 2 of the TIfolder.
Abbreviations used1
RASX-ray system interface
RAI I.I./TV AdapterX-ray system interface for TV system, monitor, Bucky wall
stand and connection panel for a number of the cables in
the corrugated hose coming from the image intensifier.
RWGBucky wall stand
DFRDigital Fluoro Radiography
LWLFiber optic cable
Siemens AGTDRX1RX99-051.032.02Register 04RAS
Medical EngineeringRev. 0103.95
1 - 2General
Inserting the RAS ring1
Control station
POLYSPOT/
Primary loop
FLUOROSPOT H
RAS
Fig. 1
Sub-station
GENERATOR
Sub-station
UNIT
Methodology1
Secondary loop 2
RAS
Sub-station
RASRASSEP
Sec. loop 1
R
A
S
kk-cable
R
A
S
Sub-station
KK1B
I.I./TV
Adapter
TV
SYSTEM
RWG
The RAS is a standardized interface for the components of an X-ray installation.
The RAS replaces the kk-connections. Microprocessor-controlled components exchange
information directly, r endering the kk -inte rface technology s uperfl uous. Only ti me-relevant
signals, e.g. analog signals from the Iontomat or multiplier will be forwarded via kk-connections.
The RAS network is configured as a ring (refer to the example shown in fig. 1).
Each ring is stabilized via a loop feedback. In the event of interference on the ring, the
system components linked by a secondary loop are automatically bypassed to maintain
the ring functions.
Data transfer between stations in the ring is serial via optic fiber cables.
Each RAS station consists of an RAS board or a combination of several RAS boards
depending on the system components (generator, diagnostic device ..) and their location
in the RAS network.
TV systems are connected to the network via a special interface, the I.I./TV adapter.
A kk1B connector for the I.I./TV adapter is avai lable to connect a Bucky wall stand.
RASRegister 04RX99-051.032.02TDRX1Siemens AG
Rev. 0103.95Medical Engineering
RAS programming for DFR systems2
POLYSPOT/POLYTRON S PLUS2
Setting the RAS group index:
• Connect the service PC to POLYSPOT/POLYTRON S PLUS (M80 X30).
• Start the service program ST 320.
Refer to the setting instructions POLYSPOT R41-030.032.01...
and setting instructions POLYTRON S PLUS R42-025.032.01...
• Set the RAS level index to 1 under XRAYSYSTEM/ROOM/COMMONDATA in the
Configurations menu (enter 0 Index 1).
2 - 1
FLUOROSPOT H2
Setting the RAS group index:
• Connect the service PC to the FLUOROSPOT (Transiti on Panel J27).
• Start the service program ST 320.
Refer to the start-up instructions FLUOROSPOT RX41-020.034.01....
• Enter 1 under RAS group index on page 2 in the System Configur ation menu
Siemens AGTDRX1RX99-051.032.02Register 04RAS
Medical EngineeringRev. 0103.95
2 - 2RAS programming for DFR systems
This page intentionally left blank.
RASRegister 04RX99-051.032.02TDRX1Siemens AG
Rev. 0103.95Medical Engineering
NOTICE
Starting up RAI I.I./TV adapter N933
• Connect the pocket terminal to board RAS 10. Use the same adapter board (D851) as
for POLYDOROS 50S/80S, or connect the service PC to th e pocket terminal emulation.
• Set switch S2 into position 2 (Confi g) on board RAS 10.
• Switch on the system
The label PT510 or PC on the board is without importance. Set a
baud rate of 9600 Bd when you connect a PC.
*2
3 - 1
RAS
BVFS
CR
MODE:
N
Y
MODE:
Y
CR
TV-STATION
N
Y
TV-STATION
Y
VB3
VA0
*1
JUSTAGE
Y/N
JUSTAGE
Y/N
Y/N
Y/N
*1
If errors occur during intial st art-up of the
adapter, press "CR" again.
*2
Or the actual software status
When you connect a PC, the <ENTER>key
ccorresponds to FNC, CR
Enter "Y".
(for TV configured)
CR
1
Siemens AGTDRX1RX99-051.032.02Register 04RAS
Medical EngineeringRev. 0103.95
3 - 2Starting up RAI I.I./TV adapter N93
If required move cursor with "←" or"→" to the
1
GROUP-INDEX
0000000000000000
1
GROUP-INDEX
0000000000000001
digit to be changed, enter "0" or"1".
Always program
Group index 1
0000.0000.0000.0001
CR
ASSIGNMENT UNIT 1
N
Y/N
Y
ASSIGNMENT UNIT 1
YY/N
CR
2
At this point, you are programming the connection for the TV system which is always connected to unit G1 of the generator:
UNIT 1 = g1
Enter "Y" for UNIT 1.
(When you enter "Y" for "UNIT 1", the system
stops the queries for additional units).
RASRegister 04RX99-051.032.02TDRX1Siemens AG
Rev. 0103.95Medical Engineering
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.