Siemens QPA228.FWNC Series, QPA228.FWSC Series, QAA2280.FWNC, QAA2280.FWSC, QFA3280.FWNC Operator Interface Manual

...
Operator Interface Guide
Room Unit for TEC, and ATEC
Overview
These devices are used with BACnet Terminal Equipment Controllers (TEC), BACnet Actuat or TEC (ATEC) Controll ers, and BACnet Programmable Controllers (PTEC).
Product Numbers
Docume nt No. 125- 703
September 19, 2017
QAA2280.FWNC QPA228x.FWSC QAA2280.FWSC QFA3280.FWNC QPA228x.FWNC QFA3280.FWSC
Accessories
Description Product Number
Passkey 544-643A 25-foot (7.6 m) cable with
connections 50-foot (15.2 m) cable with
connections 100-foot (30.5 m) cable with
connections Replacement Relative
Humidity Sensing Element for QFA and QP A
Replacement Housing Base (QAA, QF A types) Replacement Housing Base
(QPA types) Room Unit Backplate,
10-pack Power Supply (QPA types) Wall Gasket, 10-pack
588-100A
588-100B 588-100C
AQF3060
563-102-01
563-120 AQM2200-INTL
AQM2200 563-102 GSKT KIT
Operation
Operation modes
Figure 1. Operation Modes.
Normal Mode
In Normal Mode, the displ ay is updated with temperature, humi dity, carbon dioxide, and/or occ upied status (all where appl icable) on a set time cycle.
The display shows tem perature in the desired units (as set by the cli pable j umper on the PCA). Depending on the model, the humidity and/or carbon dioxide value(s) will be displayed as well. When multiple variables need to be displayed (temperature, relative humidity, and/or carbon dioxide) the display will cycle through the required values at a rate of one change every three to five seconds.
Page 1 of 4
Document No. 125-703 Operator Interface Guide September 19, 2017
Operation modes, Continued
Occupied Mode
If the TEC is currently in Unoccupied Mode (display shows man outside of the house), pressing the override button (button designated by an image of a man inside a house) results in the display showing the word OCCUPIED and the request is sent to the TEC to override the unoccupied mode. If the TEC does not get overridden, and stays in unoccupied mode, the display on the sensor reverts back to the unoccupied symbol as soon as possible.
If the TEC is currently in occupied mode, pressing the override button flashes the word OCCUPIED, and does nothing else. This functionality is intended solely to let the user know that the butt on press was recognized.
Occupied Mode Unoccupied Mode
Display Display
Setpoint Adjustment Mode
The temperature setpoint is adjusted by using the plus and minus buttons. The resulting changes in setpoint are displayed on t he di spl ay in 1.0°F or 0.5°C increments.
The setpoint adjustment will display for three seconds. If, during those three seconds, a setpoint button is pressed again, the setpoint will be adjusted accordingly and be displayed, and t he three- second count down will restart. If there is no user input for more than three seconds, the room unit will return to Normal Mode.
Press the plus/minus buttons t o scroll through the various options available for each parameter to display the desired option.
Press the occupancy override button again to move to the next user adjustable parameter.
When fi nished, remove the passkey f rom t he HMI port. User setti ng changes will be saved when the passkey is removed.
Figure 2. Inserting the Passkey.
NOTE: When the passkey is plugged i n,
communication between the room unit and the TEC is disabled. See note in Laptop Pass-thru Mode section for more information.
Settings
Set Pt Disp - determines how the user views the
temperature setpoint adjustment. The default setting is NUMERIC, and displ ays in degrees. The user can select between a NUMERIC and a GRAPHIC setpoint displ ay.
Set Pt Min - determines the minimum temperature
setpoint value. The default setting is 55°F (12.5°C). The user selects the minimum setpoint the room unit should request. The setpoint limit is ultimately defined in the controller. If the setpoint is displayed graphically as a sliding bar, this sets the left end of the slider bar. The limits are 55°F to the Set Pt Max value. The adj ustm ent step size is 1°F (0.5°C).
Passkey Mode
Insert the passkey in the sensor HMI (Human-Machine Interface) port to setup all of the user functionality listed below. For each parameter, use the following sequence:
Press the occupancy override button to edit the parameter.
Page 2 of 4 Siemens Industry, Inc.
Set Pt Max - determines the maximum temperature setpoint value. The default setting is 95°F (35°C). The user selects the max imum setpoint the sensor should request. The setpoint limit is ultimately defined in the controller. If the setpoint is displayed graphically as a sliding bar, this sets the right end of the slider bar. The limits are Set Pt Min to 95°F (35°C). The adjustment step size is 1°F (0.5°C).
Document No. 125-703
Operator Interface Guide
September 19, 2017
Disp Temp? – determines whether the temperature value should be displayed. The default value is YES.
Disp RH? – determines whether the humidity val ue should be displayed. The default value is YES, if the device is a combination temperature/humidity or temperature/ humi dity, CO2 room unit. (This option is conditional based on the presence of humidity measurement capability, and only offered with QFA3x80.F or QPA2x84. F types.)
Disp CO2? – determines whether the carbon dioxide value should be displayed. The default value is YES, if the device is a CO2 Room Unit. (This option is conditional based on the presence of CO measurement capability, and only offered with QPA2x84.F types.)
Disp Occ? – determines whether the occupancy status should be displayed. The default value is YES if the device offers occupanc y override capability. NOTE: This does not enable/disable the Occupancy override button functionality, just the display of the occupancy status.
T Cal – allows the user to field calibrate the temperature display and output through a bias adjustment. The default value is + 0°F. Adjustments can be made between -5°F and + 5°F, in 0.5°F increments.
RH Cal – allows the user to field calibrate the humidity display and output through a bias adjustment. The default value is + 0%. Adjustments can be made between -5% and + 5%, in 0.5% increments. (This option is conditional based on t he presence of humidity measurement capability, and only offered wit h QFA3x80.F or QPA2x84.F types.)
Scr Saver – allows the user to activat e or deactivat e the display screen saver. Available options are: NONE, OFF, and FADING. The default value is NONE.
NONE – The display operates at the programmed brightness level without any interrupti on.
Off – The display will turn off aft er 30 seconds. Any button press will wake it up (without doing anything else), and then the device will operate in Normal Mode for 30 seconds until the display turns off again.
2
Fading – Within 30 seconds of no operator interacti on, the entire display will fade to a display brightness of 1. The display will operate at this brightness until the user presses a button. At that time, the programmed display brightness will be restored.
Factory Defaults - allows the user to reset all parameters to fac tory defaults. The default is NO. Available options for this parameter are YES and
NO.
Parameter Default Value
Set Pt Disp NUMERIC Set Pt Min 55°F (13°C) Set Pt Max 95°F (35°C) DISP Temp? YES Disp RH? YES Disp CO2? YES Disp Occ? YES T Cal + 0°F (0°C) RH Cal + 0% CO2 Cal +0 PPM Brightness 5 Scr Saver NONE Factory Defaults NO
CO2 Cal – allows the user to field cali brate t he carbon dioxi de displ ay and output through a bias adjustment. The default value is +0 parts per million (PPM). Adjustments can be made between -175 and + 50 PPM, in 5 PPM increments. (This option is conditional based on the presence of CO
2
measurement capability, and only offered with QPA2x84.F types.)
Laptop Pass-thru Mode
To communicat e with the TEC using Datamate Base, Datamate Advanced, or Microsoft Windows
TM
Controll er Interf ace Software (WCIS), connect a PC­compatible cable (see Accessories) from a lapto p to the sensor HMI port at the bottom of the room unit. This puts the room unit in laptop mode, which deactivat es all buttons and di spl ays LAPTOP on the
Brightness – allows the user to adjust the display brightness. Valid values are 1, 2, 3, 4, 5, 6, 7, 8, 9 and 10, with 1 being dimmest and 10 bei ng brightest. The default setting is 5.
display. In this mode, the controller communicates only with the laptop and still allows the temperature, humidity, or CO2 values (where applicable) to be updated in the controller. The room unit controls communication, and updates values every 10 seconds. The room unit returns to Normal Mode once the cable is disconnected.
Siemens Industry, Inc. Page 3 of 4
Document No. 125-703 Operator Interface Guide September 19, 2017
NOTE: The Room Unit (Firmware Revi si on 016 or
lower) will stop updating ROOM TEMP, RM STPT, and OVRD if a passkey or laptop cable is connected to the Room Unit. This issue will affect the normal commissioning process, because the Specialist cannot leave the HMI cable connected to the Room Unit to monitor the TEC’s control. This affects all Series 2200 and 3200 models.
Furthermore, if the TEC is initialized, power is cycled (which has the same effect as initializing the TEC) or the Application (subpoint 2) i s commanded to another value, the following will occur:
The RM TEMP will read and control to 57°F (12°C).
The RM STPT with read and control to 98°F (37°C).
Once the passkey or HMI cable or passkey is disconnected, t he ROOM TEMP will update within 10 seconds. The RM STPT will not update until t he setpoint setting is changed on the Room Unit or until the Room Unit cycles power.
The Room Unit (Firmware Revi si on 029 or higher) in combinat ion with new release TEC products ROOM TEMP, RM STPT, and OVRD will be updated if a passkey or laptop cable is connected to the Room Unit, but the buttons will not work. New release TEC P roducts Use a SENSOR SEL point that allows the TEC to see all inputs from the Digital Room Sensor, the default setting does not allow the TEC to see the points.
Set-up of Displayed Temperature Units
The factory default for displayed temperature units is °F. To change the display to °C mode, snip the wire jumper on the back of the PCB (the visible side when the unit is taken off the wall).
Figure 3. Jumper Location.
Error Messages
Message Meaning
NO TEC
NO SENSOR
0xx
No communication with TEC. This error will display on the lower right (icon) section of the display, and the upper portion of the display should continue to display temperature v alues (and/or humidity and CO2 values where applicable).
Or a bad crimp on the RJ11 cable connection.
NOTE: This error will display whenever TEC subpoints are not available or when the TEC (or its subpoints) are not responding to commands. The default HMI baud rate on the TEC is 1200, but the room unit baud rat e i s self-adjusting.
No temperature (or humidity or CO2 where applicable) sensor detected on the PCA. This error will display until the error is corrected. This message might also display if the wrong controller-to-room unit­cable is used or if the replacement sensor element is incorrectly installed on the PCA.
Room unit firmware revision, where X is a number between 0 and 9. For Room Unit s (Revision 016 and lower) this is visible upon power-up for a f ew seconds until communication is established with the TEC or ATEC. For Revision 029 and higher, this is visible for a few seconds upon power-up and when the laptop cable is disconnected.
Information in this publication is based on c urrent specifications. Th e company reserves the right to make changes in specifications and models as desig n improvements ar e i ntroduced. Produc t or company names mentioned herein may be the tr ademarks of their respecti ve owners. © 2017 Siemens Industry, Inc.
Siemens Industry, Inc.
Building Technologies Division 1000 Deerfield Parkway Buffalo Grove, IL 6 0089-4513 USA +1-847-215-1000
Your feedback is important to us. If you have
comments about this document, please send them
toSBT_technical.editor.us.sbt@siemens.com
Document No. 125-703
Printed in the USA
Page 4 of 4
Loading...