ChapterPageRevision
allall01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
POLYDOROS LX 30/50RX63-020.891.01Page 2 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
6 _______Changes to Previous Version_____________________________________ 6 - 1
Siemens AGRX63-020.891.01Page 3 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
0 - 4Contents
POLYDOROS LX 30/50RX63-020.891.01Page 4 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
General Notes1
1 - 1
General Notes1
- With distribution of these revision level, all preceding planning guides, Speed Infos (PGs) and drafts
lose their validity.
- All layouts issued by the Planning Departments must bear a note referring to the installation and
delivery conditions of Siemens Medical Engineering. The installation and delivery conditions must
be submitted with the layouts.
- Unless otherwise specified, all dimensions are indicated in "mm".
- The symbol indicates a change (see revision status).
- Orientation points
Points specific to system components to which reference is made when positioning system
components to each other or in the room.
The isocenter of a radiographic system is always illustrated as the orientation point.
-Fixpoints
Clearly marked points on system components, installation ceiling, walls or floor on which cable
outlets are located.
Illustration in the drawings: octagon with letter/number-combination.
The cable lengths specify the maximum fixpoint distances and thus the maximum distances
between the individual system components.
- Room height
The room height is the distance measured from the top surface of the floor to the bottom surface of
the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
- Room lighting
According to DIN 68 68-57 (international standard in preparation), the lighting in rooms in which
image playback devices (monitors) are used for diagnosis, the following requirements must be
met:
adjustable, no anti-glare screen, reproducible adjustment of the lighting (e.g. dimmer with
scale),
no glare or reflection from windows, lights and light boxes in the standard working position of
the monitors.
Hotline+ 49 (9191) 18 - 8080
Siemens AGRX63-020.891.01Page 1 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
1 - 2General Notes
Safety1
- The provisions of the relevant fire safety regulations must be observed for the premises.
- The system has been developed according to EN 60601 - 1.
- Minimum dimensions (e.g. room heights, safety distances) indicated in the planning guides are
marked "min."
- Basic strength against electromagnetic sources of interference.
Result of lightning discharges.
The protection targets of the different lightning protection areas up to the unit connection are also
specified in the IEC 1024, DIN 48810, VDE 0675 and in the DEMVT recommendations.
POLYDOROS LX 30/50RX63-020.891.01Page 2 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
General Notes1 - 3
System Configuration1
Siemens AGRX63-020.891.01Page 3 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
1 - 4General Notes
Power distributor
To be provided by the customer close to the system (recommendation)
External loads
Do not connect to the power supply source:
E-machines, air conditioners, elevators and general external loads
Cables belonging to other systems
Provide for shielding measures or disconnect from all generator cable groups
High voltage cables (max. 24 m)
Run them separately from the power cables, control cables and signal cables
X-ray tube assemblies
Max. 2 tube assemblies can be connected
(LX 30/50 Lite only 1 tube assemblies)
ZA (central power line connection)
Can be installed in an additional standard cabinet (option)
POLYDOROS LX 30/50RX63-020.891.01Page 4 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
Room Planning2
2 - 1
Dimensions of Generator Cabinet2
Louvres
Cable outlet on
the left and right
side for on-floor
duct or wall duct
POLYDOROS LX
Floor cable outlet
4 x ∅ 13 mm for floor fastening *1
1 : 20
*1Fastening materials are included in the shipment; 4 x B12/115 Liebig anchor bolts
The instructions from Liebig must be observed.
Siemens AGRX63-020.891.01Page 1 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
2 - 2Room Planning
Dimensions of Control Console2
Generator Control Console
Fold-out support
Touch-sensitive
keyboard
Generator control console with
console base
NOTE
Console base floor fastening
Floor mounting
4 x ∅ 9
1 : 20
The generator console can
be installed directly on the
generator.
The generator cover panel
has already been provided
with the fastening holes and
the cable outlet so that the
control console can be
installed centered on it.
Section A - A
Wall bracket for generator control console
4 x wall fastening
∅ 5 mm on-site
Cable outlet
NOTE
The generator control console is mounted in the wall bracket.
Cable outlet
1 : 10
POLYDOROS LX 30/50RX63-020.891.01Page 2 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
Room Planning2 - 3
Dimensions of KermaX Duo Master Display2
Installation in the Vicinity of the Generator Control Console
Can be pivoted
Opening, DIN 74-A 5
Cable lead-in
Bracket
e.g. wall installation
1 : 2
Obtain installation materials locally.
Siemens AGRX63-020.891.01Page 3 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
2 - 4Room Planning
This page intentionally left blank.
POLYDOROS LX 30/50RX63-020.891.01Page 4 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
Preparation for Installation3
3 - 1
On-site Electrical Installation3
Proposal for on-site line power distributor per DIN VDE 0100-710 or national regulations
Power cables for X-ray
Must be installed by customer
*1
On-Off switch with indicator light
Emerg. stop button with latching device
SIEMENS generator
Generator cabinet
SIEMENS
Must be installed by customer
*150 A/C with the PL LX 30/50 and LX 30/50 Lite, 63 A/C with the PL LX 80
To external conductive parts
1
Per DIN EN 50178 Classification VDE 0160,
2
following FI safety switches must be used exclusively:
At U(N) = 3 ∼ 400 V: I(N) = 63A, IdN = 30 mA for alternating and pulsed as well as smooth DC fault currents.
- Order from a wholesale electrical supplier, Order No. FI 5SZ3 466 0KG05 all-current sensitive
- Order from SPH2 (Med Department, previously INAK: Part No. 49 54 470 Y7933)
- (Width of the FI safety switch is 144 mm = 8TE, installation on standard rails)
For countries in which this norm does not apply, the following FI safety switches can be used with voltages > 400 V:
At U(N) > 3 ∼ 400 V; I(N) = 125 A, IdN = 30 mA
- Doepke-Norden (order from SPH2, Part No. 51 41 168 Y7933)
System breaker
3
Option: Radiation warning lamp
4
Numbers are free cable lengths in m
[ ]
"Equipping of High Voltage Systems with Electronic Control Elements"
, the
Siemens AGRX63-020.891.01Page 1 of 2POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
3 - 2Preparation for Installation
This page intentionally left blank.
POLYDOROS LX 30/50RX63-020.891.01Page 2 of 2Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
System Connections4
4 - 1
Notes on Laying Cables4
Suggested cable run
Cable duct depth 60 mm
if required, deeper at cable crossovers.
Lay high voltage and power cables separately from control and video cables
(use shielding if possible).
Make absolutely sure to:
Avoid cable looping and, as far as possible, crossovers.
• Lay cables in separate conduits or closed cable ducts.
• If open cable ducts are used, lay the cables separately using metal dividers or similar devices.
Metal dividers
Cable duct
Siemens AGRX63-020.891.01Page 1 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
4 - 2System Connections
Overview 4
Unit side
for the connection of:
- max. 4 unit
- max. 4 Iontomat
measures
- max. 1 fluoroscopy *2
unit
+ 1 tomography unit
Generator cabinet
*1
*2 I.I. connection
only for tomogr. or
*1
fluoroscopy units
*2 Sirecon compact
Iontomat measuring
chamber
X-ray tube
assembly
*1For systems with XCS Communication system, the KK interface and thus the cables to be con-
nected are not applicable
*2Not applicable with the LX 80 and LX 30/50 Lite (no fluoroscopy)
NOTE
POLYDOROS LX 30/50RX63-020.891.01Page 2 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
The relevant PGs must be observed for system shipments!
System Connections4 - 3
Layout of Fixpoints4
Generator cabinet
KermaX Duo
KermaX Duo
Master
Display
Slave
Display
(Option)
*1 Power cable
Control cable 8 m per option 20 m
Option
Generator
control console
internal
length
1.4 m
Wall or table
installation
internal
length
after
attachment
*1Calculate the cross section for the main power cable of the generator accordingly.
2
Per IEC 601-1, the minimum cross section for the LX 30/50 is 10 mm
2
it is 10 mm
For LX 80, 16 mm
.
2
minimum cross section, not per IEC 601-1
, for the LX 30/50 Lite,
Siemens AGRX63-020.891.01Page 3 of 4POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21
4 - 4System Connections
List of Fixpoints, POLYDOROS LX 30/50 4
Cable
har-
ness
No.
1PU1P11100246 x 24
2aPU1CR1550246 x 128Option: 20m
2bPU1CR1550246 x 126.6 or 18.6 Option: Console base
2cPU1CR1550246 x 12
from
fix-
point
to
fix-
point
Cable duct
cross-
section in
2
mm
Conduit,
diameter
inside in
inches
Minimum
opening in
mm
Maximum
fixpoint
distance in
m
approx. 8 or
20
Remark
P1 = Unit connection
Option:
Wall or table
installation, max. fixpoint distance depending on attachment
length
3PU1P122003∅ 42
4PU1
Kermax
5
Master
Kermax
6
Master
Kermax
7
Master
Kermax
8
Master
PU1
P1
2801
2801
*12801
Kermax
Slave
2801
∅ 1522.5Patch cable
∅ 15
∅ 15
∅ 1524Patch cable (option)
High-voltage cable
On-site power cable
Patch cable
shipped length: 24 m
Patch cable (option)
shipped length: 24 m
*1to 2nd work station (3D stand or Floor-Ceiling stand)
List of Fixpoints Used4
FixpointSubsystemRemark
P1UnitFloor fixpoints
CR1Generator control consoleFloor fixpoints
PU1GeneratorFloor fixpoints
POLYDOROS LX 30/50RX63-020.891.01Page 4 of 4Siemens AG
Rev. 0103.04CS SD 21Medical Solutions
Technical Data5
5 - 1
Electrical Data 5
Power supplyInput powerInternal
fuse
FluoroscopyRadiography
3/N/Gnd,
POLYDOROS LX 30
3/N/Gnd,
POLYDOROS LX 30 Lite
3/N/Gnd,
POLYDOROS LX 50
3/N/Gnd,
POLYDOROS LX 50 Lite
3/N/Gnd,
POLYDOROS LX 80
*1A pre-transformer is required for 440/480 V (LX 30/50 and LX 30/50 Lite, Part No. 97 51 652
X1269, for LX 80, Part No. 48 19 756) that is installed in the generator at the factory.
*2The tolerances apply to the connection terminals in the generator.
∼ 400/440/480 V
+10% *1, *2
-15
50/60 Hz,
± 1 Hz
∼ 400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
∼ 400/440/480 V
-15 +10% *1, *2
50/60 Hz
± 1 Hz
∼ 400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
∼ 400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
1.2 kVA56 kVA
n.a.56 kVA
1.2 kVA94 kVA
n.a.80 kVA
n.a.125 kVA
35 A
slow-blow
35 A
slow-blow
35 A
slow-blow
35 A
slow-blow
50 A
slow-blow
Line Resistance Ri for Generator5
(per EN 60601-2-7, max. values in Ohm at UN - 10%)
UN (V) P (kW)30 kW50 kW80 kW
4000.440.170.11
4400.500.200.14
4800.640.240.16
Siemens AGRX63-020.891.01Page 1 of 2POLYDOROS LX 30/50
Medical SolutionsRev. 0103.04CS SD 21