Siemens LX 30 50 User manual

POLYDOROS LX 30/50
Planning Guide
POLYDOROS LX 30/50
AX
Also for: POLYDOROS LX 30/50 Lite POLYDOROS LX 80
English Print No.: RX63-020.891.01.01.02 Doc. Gen. Date: 03.04
Replaces: RX63-020.021.01.07.02
© Siemens AG 2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
0 - 2 Revision
Chapter Page Revision
all all 01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions
Contents 0 - 3
Page
1 _______General Notes _________________________________________________ 1 - 1
General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
2 _______Room Planning ________________________________________________ 2 - 1
Dimensions of Generator Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Dimensions of Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Dimensions of KermaX Duo Master Display . . . . . . . . . . . . . . . . . . . . . . 2 - 3
3 _______Preparation for Installation_______________________________________ 3 - 1
On-site Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
4 _______System Connections____________________________________________ 4 - 1
Notes on Laying Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 2
Layout of Fixpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
List of Fixpoints, POLYDOROS LX 30/50 . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
List of Fixpoints Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
5 _______Technical Data _________________________________________________ 5 - 1
Electrical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Line Resistance R
Weights and Heat Dissipation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Packing and Transport Routes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Paint Colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 2
for Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
i
6 _______Changes to Previous Version_____________________________________ 6 - 1
Siemens AG RX63-020.891.01 Page 3 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
0 - 4 Contents
POLYDOROS LX 30/50 RX63-020.891.01 Page 4 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

General Notes 1

1 - 1

General Notes 1

- With distribution of these revision level, all preceding planning guides, Speed Infos (PGs) and drafts lose their validity.
- All layouts issued by the Planning Departments must bear a note referring to the installation and delivery conditions of Siemens Medical Engineering. The installation and delivery conditions must be submitted with the layouts.
- Unless otherwise specified, all dimensions are indicated in "mm".
- The symbol indicates a change (see revision status).
- Orientation points Points specific to system components to which reference is made when positioning system components to each other or in the room. The isocenter of a radiographic system is always illustrated as the orientation point.
-Fixpoints Clearly marked points on system components, installation ceiling, walls or floor on which cable outlets are located. Illustration in the drawings: octagon with letter/number-combination. The cable lengths specify the maximum fixpoint distances and thus the maximum distances between the individual system components.
- Room height The room height is the distance measured from the top surface of the floor to the bottom surface of the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
- Room lighting According to DIN 68 68-57 (international standard in preparation), the lighting in rooms in which image playback devices (monitors) are used for diagnosis, the following requirements must be met:
adjustable, no anti-glare screen, reproducible adjustment of the lighting (e.g. dimmer with scale),
no glare or reflection from windows, lights and light boxes in the standard working position of the monitors.
Hotline + 49 (9191) 18 - 8080
Siemens AG RX63-020.891.01 Page 1 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
1 - 2 General Notes

Safety 1

- The provisions of the relevant fire safety regulations must be observed for the premises.
- The system has been developed according to EN 60601 - 1.
- Minimum dimensions (e.g. room heights, safety distances) indicated in the planning guides are marked "min."
- Basic strength against electromagnetic sources of interference. Result of lightning discharges. The protection targets of the different lightning protection areas up to the unit connection are also specified in the IEC 1024, DIN 48810, VDE 0675 and in the DEMVT recommendations.
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions
General Notes 1 - 3

System Configuration 1

Siemens AG RX63-020.891.01 Page 3 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
1 - 4 General Notes
Power distributor
To be provided by the customer close to the system (recommendation)
External loads
Do not connect to the power supply source:
E-machines, air conditioners, elevators and general external loads
Cables belonging to other systems
Provide for shielding measures or disconnect from all generator cable groups
High voltage cables (max. 24 m)
Run them separately from the power cables, control cables and signal cables
X-ray tube assemblies
Max. 2 tube assemblies can be connected
(LX 30/50 Lite only 1 tube assemblies)
ZA (central power line connection)
Can be installed in an additional standard cabinet (option)
POLYDOROS LX 30/50 RX63-020.891.01 Page 4 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

Room Planning 2

2 - 1

Dimensions of Generator Cabinet 2

Louvres
Cable outlet on the left and right side for on-floor duct or wall duct
POLYDOROS LX
Floor cable outlet
4 x 13 mm for floor fastening *1
1 : 20
*1 Fastening materials are included in the shipment; 4 x B12/115 Liebig anchor bolts
The instructions from Liebig must be observed.
Siemens AG RX63-020.891.01 Page 1 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
2 - 2 Room Planning

Dimensions of Control Console 2

Generator Control Console
Fold-out support
Touch-sensitive keyboard
Generator control console with console base
NOTE
Console base floor fastening
Floor mounting 4 x ∅ 9
1 : 20
The generator console can be installed directly on the generator. The generator cover panel has already been provided with the fastening holes and the cable outlet so that the control console can be installed centered on it.
Section A - A
Wall bracket for generator control console
4 x wall fastening
5 mm on-site
Cable outlet
NOTE
The generator control console is mounted in the wall bracket.
Cable outlet
1 : 10
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions
Room Planning 2 - 3

Dimensions of KermaX Duo Master Display 2

Installation in the Vicinity of the Generator Control Console
Can be pivoted
Opening, DIN 74-A 5
Cable lead-in
Bracket
e.g. wall installation
1 : 2
Obtain installation materials locally.
Siemens AG RX63-020.891.01 Page 3 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
2 - 4 Room Planning
This page intentionally left blank.
POLYDOROS LX 30/50 RX63-020.891.01 Page 4 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

Preparation for Installation 3

3 - 1

On-site Electrical Installation 3

Proposal for on-site line power distributor per DIN VDE 0100-710 or national regulations
Power cables for X-ray
Must be installed by customer
*1
On-Off switch with indicator light
Emerg. stop button with latching device
SIEMENS generator
Generator cabinet
SIEMENS
Must be installed by customer
*1 50 A/C with the PL LX 30/50 and LX 30/50 Lite, 63 A/C with the PL LX 80
To external conductive parts
1
Per DIN EN 50178 Classification VDE 0160,
2
following FI safety switches must be used exclusively: At U(N) = 3 ∼ 400 V: I(N) = 63A, IdN = 30 mA for alternating and pulsed as well as smooth DC fault currents.
- Order from a wholesale electrical supplier, Order No. FI 5SZ3 466 0KG05 all-current sensitive
- Order from SPH2 (Med Department, previously INAK: Part No. 49 54 470 Y7933)
- (Width of the FI safety switch is 144 mm = 8TE, installation on standard rails) For countries in which this norm does not apply, the following FI safety switches can be used with voltages > 400 V: At U(N) > 3 400 V; I(N) = 125 A, IdN = 30 mA
- Doepke-Norden (order from SPH2, Part No. 51 41 168 Y7933)
System breaker
3
Option: Radiation warning lamp
4
Numbers are free cable lengths in m
[ ]
"Equipping of High Voltage Systems with Electronic Control Elements"
, the
Siemens AG RX63-020.891.01 Page 1 of 2 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
3 - 2 Preparation for Installation
This page intentionally left blank.
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 2 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

System Connections 4

4 - 1

Notes on Laying Cables 4

Suggested cable run
Cable duct depth 60 mm
if required, deeper at cable crossovers.
Lay high voltage and power cables separately from control and video cables (use shielding if possible).
Make absolutely sure to:
Avoid cable looping and, as far as possible, crossovers.
Lay cables in separate conduits or closed cable ducts.
If open cable ducts are used, lay the cables separately using metal dividers or similar devices.
Metal dividers
Cable duct
Siemens AG RX63-020.891.01 Page 1 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
4 - 2 System Connections

Overview 4

Unit side
for the connection of:
- max. 4 unit
- max. 4 Iontomat measures
- max. 1 fluoroscopy *2 unit + 1 tomography unit
Generator cabinet
*1
*2 I.I. connection
only for tomogr. or
*1
fluoroscopy units
*2 Sirecon compact
Iontomat measuring
chamber
X-ray tube assembly
*1 For systems with XCS Communication system, the KK interface and thus the cables to be con-
nected are not applicable
*2 Not applicable with the LX 80 and LX 30/50 Lite (no fluoroscopy)
NOTE
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions
The relevant PGs must be observed for system shipments!
System Connections 4 - 3

Layout of Fixpoints 4

Generator cabinet
KermaX Duo
KermaX Duo Master Display
Slave Display (Option)
*1 Power cable
Control cable 8 m per option 20 m
Option
Generator control console
internal length
1.4 m
Wall or table installation
internal length after attach­ment
*1 Calculate the cross section for the main power cable of the generator accordingly.
2
Per IEC 601-1, the minimum cross section for the LX 30/50 is 10 mm
2
it is 10 mm For LX 80, 16 mm
.
2
minimum cross section, not per IEC 601-1
, for the LX 30/50 Lite,
Siemens AG RX63-020.891.01 Page 3 of 4 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
4 - 4 System Connections

List of Fixpoints, POLYDOROS LX 30/50 4

Cable
har-
ness
No.
1 PU1 P1 1100 2 46 x 24
2a PU1 CR1 550 2 46 x 12 8 Option: 20m
2b PU1 CR1 550 2 46 x 12 6.6 or 18.6 Option: Console base
2c PU1 CR1 550 2 46 x 12
from
fix-
point
to
fix-
point
Cable duct
cross-
section in
2
mm
Conduit, diameter inside in
inches
Minimum
opening in
mm
Maximum
fixpoint
distance in
m
approx. 8 or
20
Remark
P1 = Unit connection
Option:
Wall or table installation, max. fix­point distance depen­ding on attachment length
3 PU1 P1 2200 3 42
4PU1
Kermax
5
Master
Kermax
6
Master
Kermax
7
Master
Kermax
8
Master

PU1
P1
280 1
280 1
*1 280 1
Kermax
Slave
280 1
15 22.5 Patch cable
15
15
15 24 Patch cable (option)
High-voltage cable
On-site power cable
Patch cable shipped length: 24 m
Patch cable (option) shipped length: 24 m
*1 to 2nd work station (3D stand or Floor-Ceiling stand)

List of Fixpoints Used 4

Fixpoint Subsystem Remark
P1 Unit Floor fixpoints
CR1 Generator control console Floor fixpoints
PU1 Generator Floor fixpoints
POLYDOROS LX 30/50 RX63-020.891.01 Page 4 of 4 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

Technical Data 5

5 - 1

Electrical Data 5

Power supply Input power Internal
fuse
Fluoroscopy Radiography
3/N/Gnd,
POLYDOROS LX 30
3/N/Gnd,
POLYDOROS LX 30 Lite
3/N/Gnd,
POLYDOROS LX 50
3/N/Gnd,
POLYDOROS LX 50 Lite
3/N/Gnd,
POLYDOROS LX 80
*1 A pre-transformer is required for 440/480 V (LX 30/50 and LX 30/50 Lite, Part No. 97 51 652
X1269, for LX 80, Part No. 48 19 756) that is installed in the generator at the factory.
*2 The tolerances apply to the connection terminals in the generator.
400/440/480 V +10% *1, *2
-15
50/60 Hz,
± 1 Hz
400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
400/440/480 V
-15 +10% *1, *2
50/60 Hz
± 1 Hz
400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
400/440/480 V
-15 +10% *1, *2
50/60 Hz,
± 1 Hz
1.2 kVA 56 kVA
n.a. 56 kVA
1.2 kVA 94 kVA
n.a. 80 kVA
n.a. 125 kVA
35 A
slow-blow
35 A
slow-blow
35 A
slow-blow
35 A
slow-blow
50 A
slow-blow

Line Resistance Ri for Generator 5

(per EN 60601-2-7, max. values in Ohm at UN - 10%)
UN (V) P (kW) 30 kW 50 kW 80 kW
400 0.44 0.17 0.11
440 0.50 0.20 0.14
480 0.64 0.24 0.16
Siemens AG RX63-020.891.01 Page 1 of 2 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
5 - 2 Technical Data

Weights and Heat Dissipation 5

Weight [kg] Heat dissipation [W]
Generator cabinet
(maximum configuration)
Control console
Console base
Wall bracket
LX 30/50 and LX 80
LX 30/50 Lite
approx. 230
approx. 208
approx. 2.5 max. 20
approx. 27 n.a.
approx. 1.0 n.a.
approx. 300

Environmental Conditions 5

Operation Transport Storage
Admissible ambient temp. + 10
Admissible relative air humi­dity
Admissible air pressure 700 hPa ... 1060 hPa 500 hPa ... 1060 hPa 500 hPa ... 1060 hPa
*1 Limitation with oil-filled containers (high voltage transformer/tube unit) only up to - 20° C.
° ... + 40° C- 40° ... + 70° C *1 - 40° ... + 70° C *1
20% ... 75% 10% ... 95% 10% ... 95%

Packing and Transport Routes 5

Largest crate L 1160 x W 560 x H 1240 mm
Heaviest single part approx. 240 kg
Minimum door width and floor width for transport
min. approx.
815 mm 940 mm
Door width Floor width

Paint Colors 5

Main color Medical white C610
Combination color Neutral gray C612
Info on spray cans and paint sticks can be found in the Project Manager Handbook (PMH) Chapter 1
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 2 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions

Changes to Previous Version 6

Chapter Page Change
6 - 1
0 - 6
3 3-1 On-site Electrical Installation updated.
4 4-2 Fixpoint updated.
4 4-3 Fixpoints in the Overview and *1 text updated.
4 4-4 Fixpoints in the Fixpoint List and list of Fixpoints Used updated.
5 5-1 Technical data updated and *2 text added.
5 5-2 Paint colors updated.
66-1
The entire PG changed from 021 to 891 (Rev. level set to 01)
Chapter, Transport Conditions, eliminated and replaced by Changes to Previous Version.
Siemens AG RX63-020.891.01 Page 1 of 2 POLYDOROS LX 30/50 Medical Solutions Rev. 01 03.04 CS SD 21
6 - 2 Changes to Previous Version
This page intentionally left blank.
POLYDOROS LX 30/50 RX63-020.891.01 Page 2 of 2 Siemens AG Rev. 01 03.04 CS SD 21 Medical Solutions
Loading...