• This Service Manual is intended to give service
personnel sufficient knowledge about the KION
MultiGas 2000 (KMG 2000) to be able to diagnose
any possible problem or failure and to identify
which part or component that causes the problem
or failure. The description is limited to a level,
which allows the service personnel to track
problems down to parts that are found in the
spare part list.
• This Service Manual is also intended to give
information how to perform preventive
maintenance and to make functional checks and
calibrations.
• Documentation for the KMG 2000 consists of:
– SC 7000 / SC 9000XL – User’s guide
– KION – Operating Manual
– KION – Service Manual
– KION MultiGas 2000 – Service Manual
– KION MultiGas 2000 – Installation Instructions
– KION MultiGas 2000 – Spare Parts List
• The SC 7000 / SC 9000XL – User’s guide and the
KION – Operating Manual are indispensable
complements to the Service Manual for proper
servicing.
Symbols used in this manual
• ESD sensitive components – Make
sure to take precautions to avoid
damaging ESD sensitive components.
• Special waste – Make sure to discard disposables and replaced parts
according to local regulations and in
an environmentally acceptable way.
• Technical training – Refers to the
Technical training supplied by
Siemens.
• Service contract – Refers to the
Service contract supplied by
Siemens.
Text inside a box is used to highlight important
information.
• In addition to the Important information given
here and in the related documents, always pay
attention to applicable local and national
regulations.
• Responsibility for the safe functioning of the
equipment reverts to the owner or user in all
cases in which service or repair has been done by
a non-professional or by persons who are not
employed by or authorized by Siemens, and when
the equipment is used for other than its intended
purpose.
• Data on internal pressures are given in Pa (bar).
Airway pressures are given in cm H
• KION Display – Siemens Patient Care Monitors,
e. g. SC 7000 or SC 9000XL.
Installation
• Only personnel trained and authorized
by Siemens shall be permitted to install
the KMG 2000. The installation and
handing-over procedures are described
in the KION MultiGas 2000 – Installation
Instructions.
Functional check
• After any installation, maintenance or service
intervention in the KMG 2000, perform a
”Functional verification”. Refer to chapter Service
procedures, section Functional verification in this
Service Manual.
4
Siemens-Elema AB
E382 E392E 061 10 02 02
Service ManualKION MultiGas 2000
1. Important
Hazard notices
• Before disassembling or assembling the KION
and/or the KMG 2000, make sure that the:
– Gas supply is disconnected.
– Mains power cable is disconnected.
– KION power switch is set to Off. If the power
switch is set to Standby or On, the internal
battery will supply power to the PC boards.
– Internal battery is disconnected when working
inside the Power & Communication interface.
– The system is cleaned according to instructions
in the KION – Operating Manual, chapter
Routine cleaning.
• With power supply connected to the KION
System, there are energized electrical components
inside the unit. All personnel must exercise
extreme caution if fault tracing or adjustments are
performed with power supply connected.
Warnings
• The KMG 2000 must not be used with flammable
anesthetic agents.
• Water emptied from the water trap
should be disposed of in accordance
with local regulations for contaminated
and biologically hazardous fluids.
• The water trap should be disposed of in
accordance with local regulations for
contaminated and biologically hazardous
items.
• The evacuated gas must be handled in a suitable
manner to prevent hazardous levels of anesthetic
gases to be built-up in the ambient air.
• Analyzer performance is likely to be degraded in
the presence of operating electrosurgical or
diathermy equipment.
Service
• The KMG 2000 must be serviced at
regular intervals by personnel trained
and authorized by Siemens.
Any maintenance or service must be
noted in a log book.
• Preventive maintenance must be performed at
least once every year as long as the unit is not
used more than normal. Normal operation is
estimated to correspond to approx. 5.000 hours
of operation. Details are found in this Service
Manual, chapter ”Preventive maintenance”.
• It is recommended that maintenance
and service is done as a part of a service
contract with Siemens.
• All disposable parts shall be discarded
according to local regulations and in an
environmentally acceptable way.
• When working with ESD sensitive
components, always use a grounded
wrist band and a grounded work
surface. Adequate service tools must
always be used.
To the responsible service personnel
• The contents of this document are not binding.
If any significant difference is found between the
product and this document, please contact
Siemens for further information.
• We reserve the right to modify products without
amending this document or advising the user.
• Only personnel trained and authorized
by Siemens shall be permitted to
perform installation, service or
maintenance of the KION and/or the
KMG 2000. Only Siemens genuine
spare parts must be used. PC boards (spare parts)
must always be kept in a package for sensitive
electronic devices. Siemens will not otherwise
assume responsibility for the materials used, the
work performed or any possible consequences of
same.
E382 E392E 061 10 02 02
Siemens-Elema AB
5
KION MultiGas 2000Service Manual
2. Introduction
KION
Display
MGA-001X
2.1 General description
The KION MultiGas 2000 (KMG 2000) is intended for
integration in the KION Anesthesia System.
The KMG 2000 delivers gas measurement data to be
shown on the KION Display.
The KMG 2000 is a side-stream multigas analyzer
intended to monitor respiratory and anesthetic gases
sampled from the breathing circuits of adult, pediatric
and neonatal patients. The gases measured are O
CO
, N2O, Halothane, Enflurane, Isoflurane,
2
Sevoflurane and Desflurane in any combination.
,
2
KION
as 2000
ltiG
u
M
All gases in the gas sample, except O
and measured by an infrared sensor. O
and measured by a paramagnetic O
, are identified
2
is identified
2
sensor.
2
The information created by the KMG 2000 is
delivered to the KION Display where the information
is calculated and shown.
6
Siemens-Elema AB
E382 E392E 061 10 02 02
Service ManualKION MultiGas 2000
2.2 Mechanical overview
2.2.1 Water trap
The DRYLINE™ water trap consists of two parts, a
filter housing with hydrophobic anti-bacterial filter and
a container for separated waste.
2.2.2 Water trap receptacle
The water trap is connected to the water trap
receptacle. Switches on the water trap receptacle
detects that a water trap is connected. The sample
gas flow is routed from the water trap receptacle to
the KMG 2000 via a Nafion
2.2.3 AION
The AION™ is the gas measuring unit. It contains the
following modules:
• Pneumatic Module: Interconnection between the
different parts of the gas sampling system. Includes
two solenoid valves; the zero calibration valve (ZCV)
and the purge flow valve (PFV).
• Gas Measurement Bench (GMB): Identifies and
measures the gases CO
Enflurane, Isoflurane, Desflurane and Sevoflurane.
• Pump Module: Creates and controls the gas
sampling flow.
TM
tube.
, N2O, Halothane,
2
• AION PC board: Integrated in the AION base
module. Contains CPU, EEPROM and software.
Controls the gas sampling system and performs
calculations from the measured parameters.
2.2.4 O
The paramagnetic O
the O
sensor
2
sensor identifies and measures
concentration in the gas sample.
2
2
2.2.5 SCM300 PC board
The SCM300 PC board is an interconnection between
modules within the KMG 2000 and includes also all
external electrical connectors. The RS232 protocol
converters are located on this PC board.
A metal shield covering the external connectors
mounted on the SCM300 PC board protects the
interior from electromagnetic interference from the
outside.
2.2.6 Rear panel
The evacuation outlet connector and all external
electrical connectors are mounted on the rear panel.
E382 E392E 061 10 02 02
Siemens-Elema AB
7
KION MultiGas 2000Service Manual
2.3 Functional principles
This description refers to the functional block
diagram, see chapter Diagrams.
2.3.1 Gas sampling system
The KMG 2000 samples a continuous gas flow from
the patient breathing circuit to measure concentrations and to identify gases in the breathing gas. A gas
pump creates and controls the gas flow through the
sampling system.
From the gas sampling line, the gas enters the water
trap where water droplets and secretions are
separated from the flow and collected in the water
trap container.
In the water trap, the gas flow is divided in two parts:
• The sampling gas flow, approx. 90% of the total gas
flow.
• The purge gas flow, approx. 10% of the total flow.
Both gas flows passes a filter inside the water trap.
This filter protects the sampling system from
condensed water, secretions, bacteria and other kind
of contamination.
The sampling gas flow passes:
• The Nafion tube where the moisture content and
the temperature in the sampled gas are equalized
to that of the ambient air.
• The Zero Calibration Valve (ZCV). Zero calibrations
are performed automatically at regular intervals.
• The Gas Measurement Bench (GMB). The CO
and anesthetic agents are identified and measured
in the Gas Measurement Bench (GMB).
• The O2 Sensor where the O2 content is identified
and measured.
The purge gas flow passes:
• The Purge Flow Valve (PFV). The PFV automatically
flushes the sampling system if the sampling gas
flow is obstructed.
Both gas flows are now mixed and connected to the
pump. After passing the pump, the gas flow is routed
back to the patient breathing circuit via the evacuation
outlet.
, N2O
2
2.3.2 Data communication
Analog signals from the GMB and the O
converted to digital signals by the AION CPU. This
sensor are
2
digital information is delivered to the KION Display via
the Protocol Converter (PrC).
2.3.3 Control system
The pump and thus the gas sampling flow rate is
controlled by information from the water trap
receptacle. Switches in the water trap receptacle
identifies if a water trap is connected.
Calibration information is stored separately in
EEPROMs in the GMB, in the pump and in the AION
PC board.
Software upgrades/updates of the AION-SW and the
PrC-SW, can be performed via a PC connected to P1.
The switches SW1 and SW2 selects the flash memory
to be upgraded/updated.
The 24 V DC power is supplied via connector P2 from
the KION power supply. Both the SCM300 PC board
and the AION PC board has separate DC/DC
converters and 5 V stabilizers.
8
Siemens-Elema AB
E382 E392E 061 10 02 02
Service ManualKION MultiGas 2000
3. Description of functions
This description in this chapter also refers to the
functional block diagram, see chapter Diagrams.
3.1 Water trap
90%
Gas sample
MGA-005X
Purge inlet Sample inlet
MGA-006X
Upper
switch
Lower
switch
10%
Flat cable
PC board
The DRYLINE™ water trap protects the gas analyzer
system from condensed water, secretions, bacterial
contamination and dust. It consists of two parts, a
filter housing with hydrophobic anti-bacterial filter and
a container for separated waste.
The filter housing has one gas sample inlet to be
connected to the sampling tube.
In the water trap, the sampling flow is divided in two
parts:
• Sampling gas flow (approx. 90%).
• Purge gas flow (approx. 10%).
There are two outlets in the filter housing, one for
each gas flow.
The DRYLINE water trap to be used with the
KMG 2000 is designed for a sampling flow of 120–
200 ml/min. For further information, see Water trap
receptacle below.
3.2 Water trap receptacle
The water trap is connected to the water trap
receptacle. There are two gas connections on the
receptacle; one for the sampling gas flow and one for
the purge gas flow.
The two electrical switches on the water trap
receptacle detects if a water trap is present and
identifies the version of the connected water trap.
The water trap receptacle is designed to identify two
different water trap versions:
• The high-flow water trap. This version of the water
trap is to be used with the KMG 2000. The water
trap activates both switches. The sampling flow is
by default set to 200 ml/min, but can be changed to
120 ml/min via the KION Display, refer to the User's
manual.
• The low-flow water trap. This version of the water
trap is not used in this application. It activates only
the lower switch. This low-flow water trap is intended to be used in applications where the sampling
gas is not routed back to the patient circuit.
If no water trap is connected or if a water trap
without container is connected, both switches are
inactive. The pump, and thus the gas flow through
the KMG 2000, is automatically stopped to prevent
contamination.
The control signal cable from the water trap
receptacle PC board is connected to the SCM300 PC
board via connector P4a.
E382 E392E 061 10 02 02
Siemens-Elema AB
9
KION MultiGas 2000Service Manual
3.3 Nafion tube
The NafionTM tube equalizes the moisture content and
the temperature in the sampled gas to that of the
ambient air. This is to reduce the influence on the
measurements from humidity and to prevent from
water condensation inside the sampling system.
3.4 AION
Pneumatic
Module
Gas Measurement
Art.Nr.
Ser.Nr.
ARTEMA
Bench
Pump
Module
Base Module
including
AION PC board
The AION is the gas measuring unit. It is a complete
unit and must not be disassembled. The following
modules are included in the AION:
• Pneumatic Module
• Gas Measurement Bench
• Pump Module
• Base Module including AION PC board
3.4.1 Pneumatic Module
The pneumatic module interconnects the pump, the
gas measurement bench and the O
are two solenoid valves in the pneumatic module; the
sensor. There
2
zero calibration valve and the purge flow valve.
MGA-007X
The KMG 2000 Service Software is required to:
• Perform an O
sensor and/or the AION has been replaced.
sensor calibration if the O
2
2
• Update the configuration stored in the AION
EEPROM if the AION unit has been replaced.
This is done to secure traceability and to allow
correct revision number to be shown on the
KION Display.
Zero Calibration Valve (ZCV): To establish a gas
zero concentration reference, ambient air is routed
through the Gas Measurement Bench via the Zero
Calibration Valve.
The automatic zero calibration of the AION will be
performed:
• During warm-up
• At a temperature change >1 °C (in the GMB)
• Every 4 hours
• At gas span calibration.
Purge Flow Valve (PFV): During normal sampling
condition there is always a small purge flow, approx.
10% of the total flow, through the KMG 2000. This
purge flow is created by the purge flow restrictor.
If there is an obstruction in the sampling line, this will
be detected by the flow sensor in the pump as a
decreased flow.
If the total flow from the two water trap outlets drops
below 40 ml/min for more than one second, the
forced purge flow-function is activated during 12
seconds. This forced purge flow is created by the
Purge Flow Valve connected in parallell with the
purge flow restrictor. The forced purge flow forces
water and secretions in the sampling line into the
water trap container. The forced purge flow is routed
from the water trap directly to the pump and out
through the evacuation outlet. It will not pass the Gas
Measurement Bench or the O
Sensor.
2
If the total flow remains below the limit value, the
forced purge flow-function will be repeated up to four
times until further forced purge cycles are inhibited.
The purge function is automatic and is not possible to
start manually.
10
Siemens-Elema AB
E382 E392E 061 10 02 02
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.