Siemens Infinity MDS WIN NT Service Manual

s
IIIIN
NFFFFIIIINNNNIIIITTTTY
NN
Windows NT4 Version
Windows NT4 Version
Windows NT4 VersionWindows NT4 Version
Service Manual
Y Medside Data Station
Medside Data Station
YY
Medside Data Station Medside Data Station
ASK-T962-02-7600
EM Guidelines, 1997-04-02
ADVISORY
This document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Medside Data Station Field Service Manual
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
2.1 Power Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1.1 No Response when power On/Off switch toggled ON . . . . . . . . . . . . . . . . . . 1
2.1.2 Power On/Off Piezo Tone Fails to Sound. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1.3 Power-Up Sequence Fails to Complete Properly . . . . . . . . . . . . . . . . . . . . . . . 2
2.1.4 No Video display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1.5 MDS Fails to boot properly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 BIOS Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Reinstalling NT4 Operating System and Installing Optional Drivers . .7
3.1 Laptop Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.1 TCP/IP Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1.2 CDROM Share Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.3 Install MDS Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 MDS to Service Laptop Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Launch MDS Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 MDS Network Boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.5 Mapping MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.6 Install NT4 Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4 Phoenix BIOS Phlash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
4.1 Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Software Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.1 CDROM Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.2 Download Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.3 MDS Hard Drive Phlash Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
5.1 Opening MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.2 Replacing Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3 Replacing Hard Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.4 Replacing Memory / Daughterboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5 Closing MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6 Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
7 FLeakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Functional Verification Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Appendix A: Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Appendix B: Bios Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Appendix C: POST Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers i
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
This page intentionally left blank.
ii Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
1Introduction In keeping with the Service Strategy for the Medside Data Station (MDS),
this Service Manual provides the necessary information required to troubleshoot and service a Windows NT-based MDS. The MDS is powered by an AC/DC power adapter, and can be placed on a desktop or attached to a wall bracket. Control of all Medside Data Station functions is done by means of a keyboard and a mouse. (It is recommended that the keyboard and mouse be purchased through SIEMENS, to avoid any possible incompatibility problems). The display screen has a 1280 x 1024 resolution capabilities which enables the user to display and run applications such as
NFINITY
TM
Explorer and Webview.
I For the purpose of clarification, special text in this Service Manual is
described below:
Bold Characters text that is to be typed in by the User.
Character a space required between typed characters.
^
Italic Characters a selection that is required by the User.

2Troubleshooting If the Medside Data Station should fail to respond properly, use the

procedures below to aid in identifying and remedying the problem.

2.1 Power Problem

2.1.1 No Response when power On/Off switch toggled ON

Refer to Table 2-1 to troubleshoot Power-On problems.
Table 2-1 Power-On Problem
Conditions Possible Cause(s) Troubleshooting and Remedial Action
MDS connected directly to Power Adapter; Power Adapter LED not illuminated
MDS directly connected to Power Adapter; Power
Power source. Power Adap te r
malfunction. MDS Malfunction.
Power Adap te r malfunction.
MDS Power Switch
Adapter LED On, MDS charger LED not
MDS malfunction.
illuminated.
MDS directly connected to Power Adapter; MDS Charger
MDS Power Switch malfunction.
MDS malfunction.
LED illuminated. No Power On LED.
1. Assure Power Adapter is connected to an active hospital power source.
2. If problem persists, disconnect power adapter from MDS and measure voltage between Power Adapter output pins.
•If voltage q 11.6 VDC or Q 13.8 VDC, replace Power Adapter.
• If voltage = 11.6 to 13.8 VDC, contact TSS in Solna or Danvers.
1. Disconnect power adapter from MDS and measure voltage between Power Adapter output pins.
• If voltage q 11.6 VDC or Q 13.8 VDC, replace
power adapter.
• If voltage = 11.6 to 13.8 VDC, contact TSS in Solna
or Danvers.
1. Switch MDS Power to On.
2. If MDS fails to Power-up, contact TSS in Solna or
Danvers.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 1
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
Table 2-1 Power-On Problem (Continued)
Conditions Possible Cause(s) Troubleshooting and Remedial Action
MDS not connected to A/C Power Adapter; Power switch On,
Internal UPS Battery discharged.
Replace Internal battery.
Power LED not illuminated.
MDS malfunction

2.1.2 Power On/Off Piezo Tone Fails to Sound.

Table 2-2 Power-off Alarm Malfunction
1. Connect MDS to A/C Power Adapter.
2. Switch MDS Power switch to On and verify that battery charger LED illuminates.
Note: If Power On LED fails to illuminate, contact TSS in Solna or Danvers.
3. If battery charger LED fails to illuminate, leave power adapter connected to MDS for i 1hr.
4. After 1 hr. disconnect MDS from Power Adapter and switch MDS Power On/Off switch Off, and then On.
• If Power LED is green, reconnect MDS to Power Adapter and leave MDS connected an additional 8 hours to charge internal battery.
• If Power LED is not green, replace Internal batteries.
5. If problem still persists, contact TSS in Solna or Danvers.
Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action
Piezo tone fails to sound when MDS
Speaker. MDS malfunction.
Contact TSS in Solna or Danvers.
powered On, if MDS loses power, or when MDS is powered-Off.

2.1.3 Power-Up Sequence Fails to Complete Properly

Table 2-3 Power-up Process Malfunction
Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action
Power inputs OK, but MDS fails to complete boot up.
Bios failure. Software program
corrupted. Hard Drive failure. MDS malfunction.
1. If MDS displays error message proceed to
“Appendix B: Bios Messages” on page 43, and “Appendix C: POST Error Codes” on page 47 to
identify failure.
2. Check BIOS configuration according to Section 2.2.
3. If BIOS configuration OK, reinstall Windows according to Section 3 starting on page 7.
4. If problem still persists, contact TSS in Solna or Danvers.
TM
NT4
2 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
SW 2
aaaa
Figure 2-1 MDS switch settings

2.1.4 No Video display Table 2-4 Video malfunction

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action
MDS power LED On. No video on LCD/CRT Display.

2.1.5 MDS Fails to boot properly Table 2-5 Power On/MDS Malfunction

Cable problem. No power to Display. Bad display. Video switch set
incorrectly. MDS malfunction.
ssss
1. Check both ends of video cable and ensure cable is connected.
2. If problem persists, verify power source to display.
3. If problem persists, switch out display with a known good display.
4. If problem persists, verify MDS video switch position (see a in Figure 2-1) at rear of MDS is set to right.
5. If problem still persists, contact TSS Danvers/Solna.
SW 1
Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action
MDS power LED On. MDS Resets after successful boot.
Watchdog timer switch set incorrectly.
BIOS problem. MDS malfunction
1. Verify Watchdog timer switch position (see s in
Figure 2-1) at rear of MDS is set to left.
2. If problem persists, check BIOS configuration as described in Section 2.2.
3. If problem still persists, contact TSS Danvers/Solna.

2.2 BIOS Setup The MDS is configured at the factory for default settings that provide

proper operation. Use the following procedure to check MDS BIOS setup, if MDS does not boot to WindowsNT boot loader screen.Changing BIOS settings is not necessary on a new MDS, unless system failure occurs.
1. Switch MDS Power On/Off switch to On.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 3
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
a
s
Figure 2-2 Main screen
2. Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
3. Press F9 key, then press <Enter> to load default BIOS configuration settings.
4. Press F10 key, then press <Enter> to save configuration settings. Note: After <Enter> key is pressed, MDS will reboot.
5. Verify that MDS boots to Windows NT OS loader screen. If MDS does not boot to OS loader screen, proceed to step 6.
6. Toggle MDS On/Off power switch Off, and then On to reboot MDS.
7. Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
8. Enter correct date/time (a in Figure 2-2) for clinical site, using arrow/ number keys.
9. Verify correct “Primary Master” settings as shown in (s in Figure 2-
2).
Note: If Primary Master is incorrect, use up/down arrow keys to select Primary Master, and then press <Enter> key.
4 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
a
Figure 2-3 Hard drive settings
10. Use +/- keys to select type to Auto (a in Figure 2-3).
11. Press ESC key to get back to “Main” tab.
a
s
Figure 2-4 Advanced settings
12. Use left/right arrow keys to select “Advanced” tab (a in Figure 2-4).
13. Verify correct “Advanced” settings as shown in Figure 2-4. Note: If changes need to be made use up/down arrow keys.
14. Use up/down arrow keys to select I/O Configuration (s in Figure 2-4), and then press <Enter> key.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 5
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
Serial port A:
Figure 2-5 I/O Device configuration
15. Verify correct “I/O Device Configuration” settings as show n in Figure 2-
5.
Note: Use left/right arrow and +/- keys to make changes.
16. Press ESC key to return to main menu.
Boot
a
Figure 2-6 Boot settings
17. Use up/down arrow keys to select “Boot” tab (a in Figure 2-6).
18. Verify correct “Boot” order as shown in Figure 2-6. Note: Use left/right arrow and +/- keys to make changes.
19. Press F10 key, and then press <Enter> to save configuration settings.
6 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
Note: After <Enter> key is pressed, MDS reboots.
20. Verify that Windows NT OS loader screen appears.
21. If MDS fails to boot to OS loader screen, proceed to Section 4 to Phlash BIOS.
22. If Phlash BIOS setup procedure does not boot to Windows NT OS loader screen, proceed to Section 3 and re-install Windows NT4. If problem still exists, contact TSS in Danvers/Solna.
3 Reinstalling NT4
Operating System and Installing Optional Drivers
NT4 Reinstallation NT4 reinstallation requires making a network boot connection between an
Optional Hardware MDS Optional Hardware includes a Teac
This section explains how to reinstall Windows NT4 and also references how to install Optional Hardware Drivers onto a Medside Data Station hard drive.
Note: NT4 reinstallation should not be required on a new MDS. The factory installs Windows
TM
NT4 on the MDS prior to shipment.
In addition to these instructions, the following Hardware and Software is required:
• Siemens Service Laptop meeting minimum Hardware requirements as specified by Med QM document ARTD-
001.719.06.04.02 (V01) M4 or greater.
• RJ45 crossover cable.
• MDS Software Recovery CDROM (shipped with MDS).
MDS RJ-45 network port, and a Windows 95/98 Laptop/PC RJ-45 network port, and then transferring the Windows NT4 image from the Laptop/PC CDROM drive to the MDS Hard Drive.
TM
CDROM Drive, BackpackTM CDRW Drive, and Microsoft Software” section in I
NFINITY
TM
Optical Mouse. Refer to “Loading
TM
Medside Data Station Reference Manual
for Teac CDROM drive and Backpack CDRW drive operation. An Optical Mouse driver CDROM is shipped with each Optical Mouse.
This driver must be loaded through a network connection to the MDS. Contact local IT department for network connection configuration.

3.1 Laptop Configuration Setup the Service laptop as follows to allow network connectivity between

a Laptop CDROM drive and an MDS Hard drive.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 7
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
Figure 3-1 Windows Main Menu

3.1.1 TCP/IP Setup 1. Boot Service Laptop to Windows

2. Right-click on Network Neighborhood icon (see Figure 3-1) and click
on Properties.
TM
95/98 screen.
Figure 3-2 Network Window
3. At “Network” window (see Figure 3-2) click on Configuration Tab,
scroll down to Laptop TCP/IP Ethernet Adapter, and select Properties. Note: TCP/IP Ethernet Adapter name is unique according to
specific adapter used on laptop. Refer to Service Laptop Ethernet Adapter vendor document for specific name.
8 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
Laptop Settings
Figure 3-3 TCP/IP Window
4. Copy “IP Address” to the following
line:_____________________________ Note: IP Address is needed to reconfigure the service laptop back
to its original configuration, after completing NT4 reinstallation.
192. 168. 0. 1
255. 255. 255. 0
Figure 3-4 TCP/IP Window
5. Type new IP address 192.168.0.1 as shown in Figure 3-4, then click
on OK button.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 9
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
Laptop Settings
Figure 3-5 Network window
6. At “Network” window (see Figure 3-5) click on Identification Tab.
7. Copy Computer Name to the following line:_______________________

3.1.2 CDROM Sh are Configuration

The service laptop CDROM must be set up for file sharing. Complete the following section to configure service laptop CDROM for file sharing.
1. Select Start and scroll to Programs, and then Windows Explorer.
Figure 3-6 Windows Explorer
10 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
2. At “Exploring” window, right-click on CDROM icon and select Sharing from drop down menu (see Figure 3-6).
Note: Steps 2-4 can also be used to set up file sharing of laptop hard drive (eg.“C”drive, “D” drive) and laptop floppy drive (“A” drive) by selecting that drive in step 2 and providing a unique user­provided share name in step 3.
a
a
aa
Figure 3-7 Properties Window
3. At “Properties” window, click on Shared As button (a in Figure 3-7), type CDROM in “Share Name:” box, click on Apply, then click OK.
a
Figure 3-8 Windows Explorer
4. Verify that there is a hand shown below the CDROM icon (see a in
Figure 3-8).

3.1.3 Install MDS Utility The MDS requires the use of a software utility to make a low level connection between a MDS and a Service Laptop. This utility is also used to transfer data between the two devices. Install the MDS software Utility as follows:

1. Insert MDS Recovery CDROM (shipped with MDS) into Service
laptop CDROM Drive.
2. At Service laptop Windows screen, select Start and scroll to
Programs, then Windows Explorer.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 11
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
a
Figure 3-9 Explorer Window
3. Click on CDROM icon (a in Figure 3-9) in left pane of “Explorer”
window.
4. Double click on Tftpboot directory in right pane of Explorer window.
5. Double-click on Remoteboot.exe file in right pane of Explorer window.
Figure 3-10Win Zip Extractor Window
6. Click on Unzip button.
Note: Files are extracted to tftpboot folder on the laptop “C” drive. Once files have been extracted, the Win Zip Self-Extractor windows appear, indicating 25 file(s) unzipped successfully (see
Figure 3-10).
7. Click on OK, then Close in “Win Zip Self-Extractor” windows.
8. Close “Windows Explorer” window.
12 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Medside Data Station Field Service Manual
a
a
aa
Figure 3-11MDS, Laptop Hardware Connection
3.2 MDS to Service
Laptop Interface
1. Connect Crossover Cable from Laptop Network Interface Card
Ethernet Port to MDS Main Ethernet Port (a in Figure 3-11).
Note: Ensure MDS is not connected to Hospital Network.
2. Connect MDS Power Adapter, Keyboard, Mouse, and Monitor
according to MDS Hardware Installation Instructions. Refer to Doc. No. T951-01-7600 (shipped with MDS).
a
a
aa
d
d
dd
s
s
ss
Figure 3-12Windows Explorer

3.3 Launch MDS Utility 1. At Laptop Windows screen, select Start and scroll to Programs, and

then to Windows Explorer.
2. Double-click on C drive icon (a in Figure 3-12) in left pane of window.
3. Click on tftpboot folder (s in Figure 3-12) in left pane.
Note: Verify power On/Off switch on MDS is Off.
ASK-T962-02-7600 Siemens Medical Solutions, EM-PCS Danvers 13
MDS_sm.fm/04-02/kaupp
Field Service Manual Medsi de Data Station
Finished- MDS Remote Boot
TFTPD- Trivial File Transfer Daemon
s
s
ss
d
d
dd
Figure 3-13Window TFTPD
4. Double-click on MDS Remote Boot icon (d in Figure 3-12) in right pane.
5. Verify “Finished - MDS Remote Boot” window (a in Figure 3-13)
opens, then “TFTPD - Trivial File Transfer Daemon” window (s in
Figure 3-13) opens, then DHCP/Bootp Server 1.6.5 window (d in Figure 3-13) opens. If windows not displayed repeat steps 2-4 above.
DHCP/Bootp Server 1.6.5
a
a
aa
Note: DHCP/Boottp Server 1.6.5 window may open minimized. If window is not displayed on Main screen,(d in Figure 3-13) check to see if program is in Windows Task bar at bottom of screen.
aaaassssdddd
Figure 3-14Setup Utility Window

3.4 MDS Network Boot 1. Switch Power On/Off switch on MDS to On.

2. Press and hold F2 key to enter BIOS setup.
3. At PhoenixBIOS Setup Utility window, use MDS keyboard left/right
arrow keys to Select Boot Tab (a in Figure 3-14).
4. Use MDS keyboard Up/Down arrows to Select +Hard Drive (s in
Figure 3-14).
14 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Loading...
+ 40 hidden pages