This document corresponds to the version/revision level effective at the time of system delivery. Revisions to
hardcopy documentation are not automatically distributed.
The installation and service of equipment described herein is to be performed by qualified personnel who are
employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates
to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by
Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates
before attempting installation or service procedures.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danversi
MDS_sm.fm/04-02/kaupp
Page 4
Field Service ManualMedsi de Data Station
This page intentionally left blank.
ii Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 5
Medside Data StationField Service Manual
1IntroductionIn keeping with the Service Strategy for the Medside Data Station (MDS),
this Service Manual provides the necessary information required to
troubleshoot and service a Windows NT-based MDS. The MDS is powered
by an AC/DC power adapter, and can be placed on a desktop or attached
to a wall bracket. Control of all Medside Data Station functions is done by
means of a keyboard and a mouse. (It is recommended that the keyboard
and mouse be purchased through SIEMENS, to avoid any possible
incompatibility problems). The display screen has a 1280 x 1024 resolution
capabilities which enables the user to display and run applications such as
NFINITY
TM
Explorer and Webview.
I
For the purpose of clarification, special text in this Service Manual is
described below:
Bold Characters text that is to be typed in by the User.
Charactera space required between typed characters.
^
Italic Charactersa selection that is required by the User.
2TroubleshootingIf the Medside Data Station should fail to respond properly, use the
procedures below to aid in identifying and remedying the problem.
2.1Power Problem
2.1.1 No Response when power
On/Off switch toggled ON
Refer to Table 2-1 to troubleshoot Power-On problems.
Table 2-1Power-On Problem
ConditionsPossible Cause(s)Troubleshooting and Remedial Action
MDS connected
directly to Power
Adapter; Power
Adapter LED not
illuminated
MDS directly
connected to Power
Adapter; Power
Power source.
Power Adap te r
malfunction.
MDS Malfunction.
Power Adap te r
malfunction.
MDS Power Switch
Adapter LED On, MDS
charger LED not
MDS malfunction.
illuminated.
MDS directly
connected to Power
Adapter; MDS Charger
MDS Power Switch
malfunction.
MDS malfunction.
LED illuminated. No
Power On LED.
1. Assure Power Adapter is connected to an active
hospital power source.
2. If problem persists, disconnect power adapter from
MDS and measure voltage between Power Adapter
output pins.
•If voltage q 11.6 VDC or Q 13.8 VDC, replace
Power Adapter.
• If voltage = 11.6 to 13.8 VDC, contact TSS in
Solna or Danvers.
1. Disconnect power adapter from MDS and measure
voltage between Power Adapter output pins.
• If voltage q 11.6 VDC or Q 13.8 VDC, replace
power adapter.
• If voltage = 11.6 to 13.8 VDC, contact TSS in Solna
or Danvers.
1. Switch MDS Power to On.
2. If MDS fails to Power-up, contact TSS in Solna or
Danvers.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers1
MDS_sm.fm/04-02/kaupp
Page 6
Field Service ManualMedsi de Data Station
Table 2-1Power-On Problem (Continued)
ConditionsPossible Cause(s)Troubleshooting and Remedial Action
MDS not connected to
A/C Power Adapter;
Power switch On,
Internal UPS Battery
discharged.
Replace Internal battery.
Power LED not
illuminated.
MDS malfunction
2.1.2 Power On/Off Piezo Tone
Fails to Sound.
Table 2-2Power-off Alarm Malfunction
1. Connect MDS to A/C Power Adapter.
2. Switch MDS Power switch to On and verify that
battery charger LED illuminates.
Note: If Power On LED fails to illuminate, contact
TSS in Solna or Danvers.
3. If battery charger LED fails to illuminate, leave
power adapter connected to MDS for i 1hr.
4. After 1 hr. disconnect MDS from Power Adapter
and switch MDS Power On/Off switch Off, and then
On.
• If Power LED is green, reconnect MDS to Power
Adapter and leave MDS connected an additional
8 hours to charge internal battery.
• If Power LED is not green, replace Internal
batteries.
5. If problem still persists, contact TSS in Solna or
Danvers.
Symptom(s)Possible Cause(s)Troubleshooting and Remedial Action
Piezo tone fails to
sound when MDS
Speaker.
MDS malfunction.
Contact TSS in Solna or Danvers.
powered On, if MDS
loses power, or when
MDS is powered-Off.
2.1.3 Power-Up Sequence Fails to
Complete Properly
Table 2-3Power-up Process Malfunction
Symptom(s)Possible Cause(s)Troubleshooting and Remedial Action
Power inputs OK, but
MDS fails to complete
boot up.
Bios failure.
Software program
corrupted.
Hard Drive failure.
MDS malfunction.
1. If MDS displays error message proceed to
“Appendix B: Bios Messages” on page 43,and
“Appendix C: POST Error Codes” on page 47 to
identify failure.
2. Check BIOS configuration according to Section 2.2.
3. If BIOS configuration OK, reinstall Windows
according to Section 3 starting on page 7.
4. If problem still persists, contact TSS in Solna or
Danvers.
TM
NT4
2 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 7
Medside Data StationField Service Manual
SW 2
aaaa
Figure 2-1 MDS switch settings
2.1.4 No Video display
Table 2-4Video malfunction
Symptom(s)Possible Cause(s)Troubleshooting and Remedial Action
MDS power LED On.
No video on LCD/CRT
Display.
2.1.5 MDS Fails to boot properly
Table 2-5Power On/MDS Malfunction
Cable problem.
No power to Display.
Bad display.
Video switch set
incorrectly.
MDS malfunction.
ssss
1. Check both ends of video cable and ensure cable is
connected.
2. If problem persists, verify power source to display.
3. If problem persists, switch out display with a known
good display.
4. If problem persists, verify MDS video switch position
(see a in Figure 2-1) at rear of MDS is set to right.
5. If problem still persists, contact TSS Danvers/Solna.
SW 1
Symptom(s)Possible Cause(s)Troubleshooting and Remedial Action
MDS power LED On.
MDS Resets after
successful boot.
Watchdog timer switch
set incorrectly.
BIOS problem.
MDS malfunction
1. Verify Watchdog timer switch position (see s in
Figure 2-1) at rear of MDS is set to left.
2. If problem persists, check BIOS configuration as
described in Section 2.2.
3. If problem still persists, contact TSS Danvers/Solna.
2.2BIOS SetupThe MDS is configured at the factory for default settings that provide
proper operation. Use the following procedure to check MDS BIOS setup,
if MDS does not boot to WindowsNT boot loader screen.Changing BIOS
settings is not necessary on a new MDS, unless system failure occurs.
1. Switch MDS Power On/Off switch to On.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers3
MDS_sm.fm/04-02/kaupp
Page 8
Field Service ManualMedsi de Data Station
a
s
Figure 2-2 Main screen
2. Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
3. Press F9 key, then press <Enter> to load default BIOS configuration
settings.
4. Press F10 key, then press <Enter> to save configuration settings.
Note: After <Enter> key is pressed, MDS will reboot.
5. Verify that MDS boots to Windows NT OS loader screen. If MDS does
not boot to OS loader screen, proceed to step 6.
6. Toggle MDS On/Off power switch Off, and then On to reboot MDS.
7. Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
8. Enter correct date/time (a in Figure 2-2) for clinical site, using arrow/
number keys.
9. Verify correct “Primary Master” settings as shown in (s in Figure 2-
2).
Note: If Primary Master is incorrect, use up/down arrow keys to
select Primary Master, and then press <Enter> key.
4 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 9
Medside Data StationField Service Manual
a
Figure 2-3 Hard drive settings
10. Use +/- keys to select type to Auto (a in Figure 2-3).
11. Press ESC key to get back to “Main” tab.
a
s
Figure 2-4 Advanced settings
12. Use left/right arrow keys to select “Advanced” tab (a in Figure 2-4).
13. Verify correct “Advanced” settings as shown in Figure 2-4.
Note: If changes need to be made use up/down arrow keys.
14. Use up/down arrow keys to select I/O Configuration (s in Figure 2-4),
and then press <Enter> key.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers5
MDS_sm.fm/04-02/kaupp
Page 10
Field Service ManualMedsi de Data Station
Serial port A:
Figure 2-5 I/O Device configuration
15. Verify correct “I/O Device Configuration” settings as show n in Figure 2-
5.
Note: Use left/right arrow and +/- keys to make changes.
16. Press ESC key to return to main menu.
Boot
a
Figure 2-6 Boot settings
17. Use up/down arrow keys to select “Boot” tab (a in Figure 2-6).
18. Verify correct “Boot” order as shown in Figure 2-6.
Note: Use left/right arrow and +/- keys to make changes.
19. Press F10 key, and then press <Enter> to save configuration settings.
6 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 11
Medside Data StationField Service Manual
Note: After <Enter> key is pressed, MDS reboots.
20. Verify that Windows NT OS loader screen appears.
21. If MDS fails to boot to OS loader screen, proceed to Section 4 to
Phlash BIOS.
22. If Phlash BIOS setup procedure does not boot to Windows NT OS
loader screen, proceed to Section 3 and re-install Windows NT4. If
problem still exists, contact TSS in Danvers/Solna.
3 Reinstalling NT4
Operating System
and Installing
Optional Drivers
NT4 ReinstallationNT4 reinstallation requires making a network boot connection between an
Optional HardwareMDS Optional Hardware includes a Teac
This section explains how to reinstall Windows NT4 and also references
how to install Optional Hardware Drivers onto a Medside Data Station hard
drive.
Note: NT4 reinstallation should not be required on a new MDS.
The factory installs Windows
TM
NT4 on the MDS prior to shipment.
In addition to these instructions, the following Hardware and Software is
required:
• Siemens Service Laptop meeting minimum Hardware
requirements as specified by Med QM document ARTD-
001.719.06.04.02 (V01) M4 or greater.
• RJ45 crossover cable.
• MDS Software Recovery CDROM (shipped with MDS).
MDS RJ-45 network port, and a Windows 95/98 Laptop/PC RJ-45 network
port, and then transferring the Windows NT4 image from the Laptop/PC
CDROM drive to the MDS Hard Drive.
TM
CDROM Drive, BackpackTM
CDRW Drive, and Microsoft
Software” section in I
NFINITY
TM
Optical Mouse. Refer to “Loading
TM
Medside Data Station Reference Manual
for Teac CDROM drive and Backpack CDRW drive operation.
An Optical Mouse driver CDROM is shipped with each Optical Mouse.
This driver must be loaded through a network connection to the MDS.
Contact local IT department for network connection configuration.
3.1Laptop ConfigurationSetup the Service laptop as follows to allow network connectivity between
a Laptop CDROM drive and an MDS Hard drive.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers7
MDS_sm.fm/04-02/kaupp
Page 12
Field Service ManualMedsi de Data Station
Figure 3-1 Windows Main Menu
3.1.1 TCP/IP Setup1. Boot Service Laptop to Windows
2. Right-click on Network Neighborhood icon (see Figure 3-1) and click
on Properties.
TM
95/98 screen.
Figure 3-2 Network Window
3. At “Network” window (see Figure 3-2) click on Configuration Tab,
scroll down to Laptop TCP/IP Ethernet Adapter, and select Properties.
Note: TCP/IP Ethernet Adapter name is unique according to
specific adapter used on laptop. Refer to Service Laptop Ethernet
Adapter vendor document for specific name.
8 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 13
Medside Data StationField Service Manual
Laptop Settings
Figure 3-3 TCP/IP Window
4. Copy “IP Address” to the following
line:_____________________________
Note: IP Address is needed to reconfigure the service laptop back
to its original configuration, after completing NT4 reinstallation.
192. 168. 0. 1
255. 255. 255. 0
Figure 3-4 TCP/IP Window
5. Type new IP address 192.168.0.1 as shown in Figure 3-4, then click
on OK button.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers9
MDS_sm.fm/04-02/kaupp
Page 14
Field Service ManualMedsi de Data Station
Laptop Settings
Figure 3-5 Network window
6. At “Network” window (see Figure 3-5) click on Identification Tab.
7. Copy Computer Name to the following line:_______________________
3.1.2 CDROM Sh are
Configuration
The service laptop CDROM must be set up for file sharing. Complete the
following section to configure service laptop CDROM for file sharing.
1. Select Start and scroll to Programs, and then Windows Explorer.
Figure 3-6 Windows Explorer
10 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 15
Medside Data StationField Service Manual
2. At “Exploring” window, right-click on CDROM icon and select Sharing
from drop down menu (see Figure 3-6).
Note: Steps 2-4 can also be used to set up file sharing of laptop
hard drive (eg.“C”drive, “D” drive) and laptop floppy drive (“A”
drive) by selecting that drive in step 2 and providing a unique userprovided share name in step 3.
a
a
aa
Figure 3-7 Properties Window
3. At “Properties” window, click on Shared As button (a in Figure 3-7),
type CDROM in “Share Name:” box, click on Apply, then click OK.
a
Figure 3-8 Windows Explorer
4. Verify that there is a hand shown below the CDROM icon (see a in
Figure 3-8).
3.1.3 Install MDS Utility The MDS requires the use of a software utility to make a low level
connection between a MDS and a Service Laptop. This utility is also used
to transfer data between the two devices. Install the MDS software Utility
as follows:
1. Insert MDS Recovery CDROM (shipped with MDS) into Service
laptop CDROM Drive.
2. At Service laptop Windows screen, select Start and scroll to
Programs, then Windows Explorer.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers11
MDS_sm.fm/04-02/kaupp
Page 16
Field Service ManualMedsi de Data Station
a
Figure 3-9 Explorer Window
3. Click on CDROM icon (a in Figure 3-9) in left pane of “Explorer”
window.
4. Double click on Tftpboot directory in right pane of Explorer window.
5. Double-click on Remoteboot.exe file in right pane of Explorer window.
Figure 3-10Win Zip Extractor Window
6. Click on Unzip button.
Note: Files are extracted to tftpboot folder on the laptop “C” drive.
Once files have been extracted, the Win Zip Self-Extractor
windows appear, indicating 25 file(s) unzipped successfully (see
Figure 3-10).
7. Click on OK, then Close in “Win Zip Self-Extractor” windows.
8. Close “Windows Explorer” window.
12 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 17
Medside Data StationField Service Manual
a
a
aa
Figure 3-11MDS, Laptop Hardware Connection
3.2MDS to Service
Laptop Interface
1. Connect Crossover Cable from Laptop Network Interface Card
Ethernet Port to MDS Main Ethernet Port (a in Figure 3-11).
Note: Ensure MDS is not connected to Hospital Network.
2. Connect MDS Power Adapter, Keyboard, Mouse, and Monitor
according to MDS Hardware Installation Instructions. Refer to Doc.
No. T951-01-7600 (shipped with MDS).
a
a
aa
d
d
dd
s
s
ss
Figure 3-12Windows Explorer
3.3Launch MDS Utility1. At Laptop Windows screen, select Start and scroll to Programs, and
then to Windows Explorer.
2. Double-click on C drive icon (a in Figure 3-12) in left pane of window.
3. Click on tftpboot folder (s in Figure 3-12) in left pane.
Note: Verify power On/Off switch on MDS is Off.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers13
MDS_sm.fm/04-02/kaupp
Page 18
Field Service ManualMedsi de Data Station
Finished- MDS Remote Boot
TFTPD- Trivial File Transfer Daemon
s
s
ss
d
d
dd
Figure 3-13Window TFTPD
4. Double-click on MDS Remote Boot icon (d in Figure 3-12) in right pane.
5. Verify “Finished - MDS Remote Boot” window (a in Figure 3-13)
opens, then “TFTPD - Trivial File Transfer Daemon” window (s in
Figure 3-13) opens, then DHCP/Bootp Server 1.6.5 window (d in
Figure 3-13) opens. If windows not displayed repeat steps 2-4 above.
DHCP/Bootp Server 1.6.5
a
a
aa
Note: DHCP/Boottp Server 1.6.5 window may open minimized. If
window is not displayed on Main screen,(d in Figure 3-13) check
to see if program is in Windows Task bar at bottom of screen.
aaaassssdddd
Figure 3-14Setup Utility Window
3.4MDS Network Boot1. Switch Power On/Off switch on MDS to On.
2. Press and hold F2 key to enter BIOS setup.
3. At PhoenixBIOS Setup Utility window, use MDS keyboard left/right
arrow keys to Select Boot Tab (a in Figure 3-14).
4. Use MDS keyboard Up/Down arrows to Select +Hard Drive (s in
Figure 3-14).
14 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 19
Medside Data StationField Service Manual
5. Hold down on Shift key, and momentarily press ! key.
6. Verify that an exclamation point ! appears to the left of +Hard Drive
(d in Figure 3-14).
7. Press MDS keyboard F10 key.
8. Press <Enter> to save and exit.
Figure 3-15MDS Service Menu
9. Verify MDS reboots to “MDS Service Menu” (see Figure 3-15).
10. Use MDS keyboard Up/Down arrows, if necessary, to Select Boot to
network login, and then press <Enter>.
3.5Mapping MDSAfter pressing <Enter>, the MDS Utility establishes a network link
between the Service Laptop and the MDS. Once this link has been
established, the A:\> prompt appears at bottom of screen.
Figure 3-16Map M.D.S. to CD ROM drive
The text after the A:\ prompt displays an example of how to map the MDS
to a Service Laptop CDROM Drive (see Figure 3-16).
11. After the A:\> prompt, type net
use^x:^^\\computername\shared
^
drivename, and then press <Enter>.
Note: Computer name = name noted in step 7 of Section 3.1.1,
andshared drive = name of drive typed in step 3 of Section 3.1.2.
12. At message “Type your user name, or press ENTER if it is
ADMINISTRATOR:”, press <Enter>.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers15
MDS_sm.fm/04-02/kaupp
Page 20
Field Service ManualMedsi de Data Station
13. At message “Type your password:”, press <Enter>.
14. At message “Please confirm your password, so that a password list
may be created:”, press <Enter>.
15. At A:\> prompt, type x: and then press <Enter>.
3.6Install NT4 Image
Caution:
Windows
where the MDS hard drive has been replaced with a new blank
hard drive, or if NT4 is corrupted and reinstallation is a final
troubleshooting procedure. All data (non-NT4) files must be
backed up before proceeding with reinstallation, as the MDS hard
drive is erased (formatted) during the reinstallation process.
TM
NT4 reinstallation must be performed only in cases
Figure 3-17Norton Ghost menu
1. At X:\> prompt type ghost, and then press <Enter>.
2. After Norton
TM
Ghost screen appears, click on OK to begin using
Norton Ghost Utility.
Figure 4-1 Utility Window
3. At Norton Ghost Utility window, select L
ocal Q Disk Q From Image.
4. Click on From Image, and then press <Enter>.
16 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 21
Medside Data StationField Service Manual
MDS_XXXX.
a
a
aa
Figure 4-2 Image Folder (Image file)
5. At “File name to load image from” window, click on MDS_xxxx.GHO
(see a in Figure 4-2).
Note: xxxx=language version.
Figure 4-3 Drive Number Window
6. At “Select local destination drive by clicking on the drive number”
window, click on OK button.
Figure 4-4 Destination Drive Window
7. At “Destination Drive Details” window, click on OK button.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers17
MDS_sm.fm/04-02/kaupp
Page 22
Field Service ManualMedsi de Data Station
Figure 4-5 Image Transfer Window
8. At “Question” window, click on Yes to proceed with disk load.
Note: Ghost image is transferred from the Service Laptop CDROM
drive to the MDS hard drive. Once Ghost image has been
transferred “Clone Complete” window appears. See Figure 4-6 on
page 18.
Continue
Figure 4-6 Clone Window
9. At “Clone Complete” window, click on Continue to return to Ghost
Main Utility page.
Quit
Figure 4-7 Main Window
18 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 23
Medside Data StationField Service Manual
10. At Main Utility window, click on Quit.
Yes
Figure 4-8 Quit Symantec Ghost Window
11. At “Quit Symantec Ghost” window, click on Yes to return to X:\>
prompt.
12. At MDS keyboard, press and hold Ctrl
+Alt +Delete keys to reboot
MDS.
13. Press and hold MDS F2 key to enter PhoenixBIOS Setup Utility.
14. Complete steps 2 and 3 in Section 3.4 (removing exclamation point
from left side of “+Hard Drive”) to enable MDS hard drive boot.
15. Press MDS keyboard F10 key, and then press <Enter> to save and
exit.
Figure 4-9 Windows Login Screen
16. Verify that MDS boots to Windows NT 0S loader screen (see Figure 4-
9), and then automatically continues to Hardware Profile/Configuration
Recovery Menu.
17. At Hardware Profile/Configuration Recovery Menu press <Enter> to
boot to Windows NT Workstation Setup (Sysprep) screen.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers19
MDS_sm.fm/04-02/kaupp
Page 24
Field Service ManualMedsi de Data Station
Figure 4-10 Setup screen
18. At “Windows NT Workstation Setup” screen (see Figure 4-10) click
Next.
Figure 4-11License Agreement screen
19. At “License Agreement” screen (see Figure 4-11) click on “I accept
this Agreement”, then click on Next.
20 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 25
Medside Data StationField Service Manual
Figure 4-12Name and Organization screen
20. At “Name and Organization” screen (see Figure 4-12), type in
Company or Hospital name in “Name” box, and Company or Hospital
Organization in “Organization” box, and then click Next.
21. Proceed to step 21 if using English version of Windows NT.
Otherwise at “Registration” screen type Product ID# in blank boxes,
and then click Next.
Note: Product ID# is located on Microsoft label at side panel of
MDS.
Figure 4-13Computer Name screen
22. At “Computer Name” screen (see Figure 4-13) type in a unique name
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers21
MDS_sm.fm/04-02/kaupp
Page 26
Field Service ManualMedsi de Data Station
(15 characters or less) for this MDS, then click on Next.
Note: If this MDS is used in a network, it is imperative that it has a
unique name. Network problems occur if 2 or more MDS’s share
an identical name.
Figure 4-14 Administrator screen
23. At “Administrator Account” screen (see Figure 4-14), type a
password (14 characters or less) in “Password” box, then re-type
same password in “Confirm Password” box, and then click on Next.
Note: If password has not been defined for this MDS, leave boxes
empty and click Next.
Figure 4-15Windows NT Setup screen
24. At “Windows NT Setup” screen (see Figure 4-15) click Finish.
25. Verify MDS reboots to OS loader V4.00 screen.
22 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 27
Medside Data StationField Service Manual
26. Pres s <Enter> and follow instructions on screen to complete NT4
Sysprep Utility.
Note: MDS reboots after NT4 Sysprep Utility is complete.
4 Phoenix BIOS Phlas hEach MDS is shipped with a default BIOS switch installed at the factory.
Phlashing BIOS is necessary only if current MDS BIOS is not operating
correctly or if an updated version is needed to correct MDS system
failures. If original BIOS is not operating correctly, complete Sect ion 2.2
before Phlashing BIOS. If Section 2.2 fails to correct problem, Phlash BIOS
as described below.
BIOS Phlash files are distributed in 4 formats:
• Existing Phlash files on MDS hard drive.
• Phlash files on MDS Software CDROM (shipped with each MDS).
• Upgrade Phlash files ordered from factory.
•Upgrade Phlash files downloaded from EM location
on MED-TD site (www-td.med.siemens.de).
Setup MDS to phlash the MDS BIOS as described in Section 4.1 and
Section 4.2 below.
a
a
aa
Figure 4-1 MDS (rear view)
4.1Hardware SetupVerify that BIOS switch (see a in insert in Figure 4-1) on rear of MDS is
set to right.
4.2Software SetupCopy Phlash software to MDS hard drive according to Section 4.2.1 for
CDROMor Section 4.2.2 below, if downloading from laptop.
4.2.1 CDROM Setup ProcedureUse this procedure to load Phlash software from MDS CDROM (shipped
with MDS), or Phlash Upgrades distributed on CDROM (ordered from
Factory).
1. Insert MDS CDROM into Service Laptop.
2. Configure service laptop for network boot according to Section 3.1
through Section 3.4.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers23
MDS_sm.fm/04-02/kaupp
Page 28
Field Service ManualMedsi de Data Station
a
a
aa
s
s
ss
d
d
dd
f
f
ff
Figure 4-2 DOS Window
3. At DOS A:> Phlash prompt, type C: (a in Figure 4-1) and press
<Enter>.
4. At DOS C:> prompt, type md
<Enter>.
Note: If message “A subdirectory or file phlash already exists”
appears, press <Enter>.
5. At DOS C:> prompt, type x: (d in Figure 4-1) and press <Enter>.
6. At DOS X:> prompt, type copy
and press <Enter>.
Note: If message “Overwrite C:\PHLASH\PHLASH.Exe (Yes/NO/
All) appears, type A and press <Enter>.
7. Verify that the following files scroll up the screen:
• Phlash\Phlash.exe
• Phlash\Zeus_XXX.rom
• Phlash\platform.bin
3 file(s) copied
8. At DOS X:> prompt, press Ctrl+Alt+Delete keys to reboot system.
9. Press F2 key and set BIOS for Hard drive boot. See step 2 and 3 of
Section 3.4.
10. Complete MDS Hard Drive Phlash Procedure. See Section 4.3.
4.2.2 Download Setup ProcedureFor Phlash upgrade using Electronic format go to TD Website (wwwtd.med.siemens.de).
phlash (s in Figure 4-1) and press
^
phlash^c:\phlash (f in Figure 4-1)
^
1. Select Product InformationzEM SystemzPCSz Softwar e PCSz
MDS. Download MDS Phlash directory from TD website and save
files to Service Laptop “C:\phlash” directory.
2. Configure laptop for network boot according to Section 3.1 through
step 7 of Section 3.4.
Note: In Section 3.1.2, configure file sharing for (C:) drive instead of
24 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 29
Medside Data StationField Service Manual
CDROM drive. At step 3 of Section 3.1.2 type C in “Share Name”
box.
3. At MDS DOS A:> Phlash prompt type C: (a in Figure 4-1) and press
<Enter>.
4At MDS DOS C:> prom pt, type md
phlash (s in Figure 4-1) and
^
press <Enter>.
Note: If message “a subdirectory or file phlash already exists”
appears, press <Enter>.
5. At MDS DOS C:> prompt, type x: (d in Figure 4-1) and press
<Enter>.
6. At MD S D O S X:> prompt, type copy
phlash^c:\phlash (f in Figure
^
4-1) and press <Enter>.
Note: If message “Overwrite C:\PHLASH\PHLASH.Exe (Yes/NO/
All) appears, type A and press <Enter>.
7. Verify that the following files scroll up the screen:
• Phlash\Phlash.exe
• Phlash\Zeus_XXX.rom
• Phlash\platform.bin
1 file(s) copied
8. At DOS X:> prompt, Ctrl+Alt+Delete keys to reboot system.
9. Press F2 key and set BIOS for Hard drive boot (see step 2 and step 3
of Section 3.4).
10. Complete MDS Hard Drive Phlash Procedure. See Section 4.3.
4.3MDS Hard Drive
Phlash Procedure
MS_DOS
Figure 4-3 NT Boot loader window
1. Switch MDS Power On/Off switch to On, to boot MDS to Boot
Loader screen.
Note: Do not attempt to complete this operation by entering DOS
through the Windows environment.
2. Using keyboard up/down arrows, select MS-DOS (as shown in Figure
4-3).
3. Press <Enter>, then immediately press and hold F5 key.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers25
MDS_sm.fm/04-02/kaupp
Page 30
Field Service ManualMedsi de Data Station
Figure 4-4 Boot window
4. Verify that message “MS-DOS is bypassing your config.sys and
autoexec.bat files” is displayed. If message is not displayed repeat
step 1 and 2 until message is displayed.
5. At DOS C:> prompt, type cd
phlash and press <Enter>.
^
6. At DOS C:> PHLASH prompt, type dir and press <Enter>.
7. Verify that the following files appear in the phlash directory:
• Phlash.exe
•Platform.bin
• Zeus_XXX.ROM
Note: XXX indicates this ROM update version (e.g xxx=020 for
ROM).
8. At DOS C:> PHLASH prompt, type phlash
zeus^XXX.rom (where
^
xxx=ROM version displayed in step 6) and press <Enter>.
Figure 4-5 Phlash program
9. Verify that Phlash program begins loading new BIOS version (see
Figure 4-5).
26 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 31
Medside Data StationField Service Manual
Figure 4-6 Phlash Window
10. Once Phlash program is completed, verify message “Phlash memory
has been successfully programmed” is displayed in PhoenixPhlash
Status box (see Figure 4-6).
Note: If message is not displayed, repeat steps 1-6.
11. Power-down MDS.
12. Power-up MDS and press and hold F2 key until BIOS screen appears.
13 At “Main” tab of PhoenixBIOS Setup Utility screen, verify that new
BIOS version is displayed at right side of “BIOS Version:”.
14. Press F9 key, and then press <Enter> to install BIOS default settings.
15. Press F10 key, and then press <Enter> to save and exit BIOS.
5Replacement
Procedures
5.1Opening MDS1. Remove all cables attached to MDS.
Caution:
The MDS contains PC boards that can be affected by static
discharge. Work in a static-protected environment.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers27
MDS_sm.fm/04-02/kaupp
Page 32
Field Service ManualMedsi de Data Station
a
Figure 5-1 MDS top view
2. Set MDS upright on clean surface.
3. Remove and save 6 Phillips-head screws (a in Figure 5-1) that hold
top cover to MDS.
4. Remove top cover and set aside.
5. Set MDS upside down on clean surface.
28 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 33
Medside Data StationField Service Manual
s
s
ss
a
a
aa
Figure 5-2 MDS (bottom view)
s
s
ss
a
a
aa
Figure 5-3 MDS exploded bottom view front panel
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers29
MDS_sm.fm/04-02/kaupp
Page 34
Field Service ManualMedsi de Data Station
6. Insert small blade screwdriver between front panel and chassis of
MDS (as shown at a in Figure 5-2 and in Figure 5-3 “exploded
view”), close to each of three panel locking tabs indicated by s in
Figure 5-2, and carefully lift up on screwdriver to release front panel
from each locking tab (s in Figure 5-2 and in Figure 5-3 “exploded
view”), and then pull front panel out so that tabs cannot reset.
7. Remove and set front panel aside.
ssss
ssss
ssss
dddd
aaaa
Figure 5-4 MDS rear view
8. Set MDS upright on clean surface.
9. Remove and save 2 Phillips-head screws (a in Figure 5-4), ten posts
(s in Figure 5-4) and three nuts on auxiliary jacks (d in Figure 5-4)
that secure rear panel to MDS.
10. Remove rear panel and set aside.
ssss
aaaa
Figure 5-5 Front Panel battery
5.2Replacing Battery11. Lift up and then pull battery (a in Figure 5-5) out of front panel
housing to gain access to battery connector (s in Figure 5-5).
12. Pull out battery connector, and then remove and set battery aside.
Note: Note polarity of battery cable for reference when
reassembling MDS.
30 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 35
Medside Data StationField Service Manual
5.3Replacing Hard Drive
a
a
aa
Figure 5-6 MDS (top cover removed)
1. Set MDS upright on clean surface.
2. Unplug ribbon cable connector (a in Figure 5-6) from motherboard,
and fold back.
d
d
dd
d
d
dd
Figure 5-7 MDS (front view)
s
s
ss
a
a
aa
3. Remove and save 4 Phillips-head screws and sleeves (d in Figure 5-
7) that secure primary hard drive to top and bottom of front housing.
4. Remove hard drive from front housing, disconnect ribbon cable, and
set hard drive on flat clean surface.
5. Align pins on hard drive to ribbon cable connector, and carefully press
into place.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers31
MDS_sm.fm/04-02/kaupp
Page 36
Field Service ManualMedsi de Data Station
Note: Ribbon cable connector is keyed and can only be inserted
on the hard drive in one orientation.
6. Insert hard drive into front housing and secure with screws removed
in step 5 above.
7. Follow procedure of Section 5.3 in reverse order to reassemble MDS,
and then proceed to Section 5.5.
5.4Replacing Memory /
Daughterboard
a
a
aa
a
a
aa
d
d
dd
s
s
ss
s
s
ss
Figure 5-8 MDS (top cover removed)
1. Unplug ribbon cable connector (a in Figure 5-8) from mother board,
and fold back.
2. Remove and save plastic ethernet cover (s in Figure 5-8).
3. Remove and save 6 Phillips-head screws (d in Figure 5-8) that secure
daughter board to mother board.
4. Lift daughter board up to separate from mother board, slide board
slightly to right to separate from PCMCIA guide slot, and then pull
board toward rear of MDS to remove board.
5. Do either a, b, or c, as appropriate.
a)If replacing daughter board, locate replacement board in position on
MDS and perform steps 4 - 1 above in reverse to reassemble unit.
Then proceed to Section 5.5.
b)If replacing memory module, go to step 6.
32 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 37
Medside Data StationField Service Manual
.
s
sa
ss
a
aa
Figure 5-9 Memory slots
6. Lift defective memory module a or s (not installed in Figure 5-9) out
of memory slot and remove module.
7. Align pins on replacement memory module to connector on
motherboard (a or s in Figure 5-9).
8. Carefully seat memory module into connector, and then press down
to lock into place.
Note: Memory module is slotted and can only be inserted in only
one orientation, and snaps into place when properly installed.
9. Locate daughter board in position on MDS and perform steps 4 - 1
above in reverse to reassemble unit.
10. Proceed to Section 5.5.
5.5Closing MDS1. Set MDS upright on clean surface.
2. Align screw holes on rear panel to screw holes on back of MDS.
3. Insert and tighten 2 Phillips-head screws removed in step 8 of Section
5.1.
4. Insert and tighten 10 post removed in step 9 of Section 5.1.
5. Insert and tighten 3 nuts removed in step 10 of Section 5.1.
6. Extend PCMCIA eject button out, so that front panel can be installed.
7. Align tab slots on front panel to tabs on bottom of MDS, and carefully
press front panel into place.
Note: Tabs snap into place when properly installed.
8. Align screw holes on top cover to screw holes on MDS.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers33
MDS_sm.fm/04-02/kaupp
Page 38
Field Service ManualMedsi de Data Station
9. Insert and tighten 6 Phillips-head screws removed in step 3 of Section
5.1.
10. Proceed to Section 6,
6 Functional CheckThe following procedures check the MDS’s hard drive, memory, power
circuits, power-up sequence, power indicator, and software. Begin the
procedure with the MDS powered off. Record all values in “Functional
Verification Checklist” on page 37. Retain a copy of test results with your
records.
aaaa
ssss
Figure 6-1 MDS (front view)
1. Connect cables removed from MDS during step 1 of Section 2.1
2. Power up MDS, and press and hold down F2 key to enter Bios setup.
a
Figure 6-2 Boot-Up Self-Test Screen
3. During boot up, verify memory test passes (a in Figure 6-2).
34 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 39
Medside Data StationField Service Manual
a
s
Figure 6-3 PhoenixBios Utility Screen
4. At “MAIN” tab of PhoenixBIOS setup utility screen verify the
following:
4.1) Cache Ram, System Memory and Extended memory are as
shown by a in Figure 6-3.
4.2) Primary Master drive capacity is displayed (s in Figure 6-3).
5. Press F10 key then press <Enter> to save and exit PhoenixBIOS
Utility menu.
6. Verify two power LED’s on front panel illuminate green (s in Figure 6-
1), Medside Data Station emits a brief tone, and monitor display
begins boot sequence.
7. Verify MDS boots to OS loader screen.
8. Using Up/Down arrows select Windows NT WorkstationVersion4.00
and press <Enter>.
Note: If no selection is made, this screen times out in 25 seconds
and then boots to Hardware Profile/Configuration Recovery Menu.
9. At Hardware Profile/Configuration Recovery Menu press <Enter> to
boot to NT4 Login Window.
Note: If NT Login Window does not appear, reboot MDS.
10. At OS Loader scre en selec t Windows NT Workstation 4.00 [VGA
mode] and press <Enter>.
11. At NT4 Login Window, press Ctrl+Alt+Delete to Login.
12. Click on OK at Login information windows to boot to “MAIN” screen.
Note: Do not enter name or password.
13. Verify that correct language appears on Windows NT Workstation
main screen.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers35
MDS_sm.fm/04-02/kaupp
Page 40
Field Service ManualMedsi de Data Station
14. Perform leakage current test and functionally verify proper operation
of reassembled MDS before returning MDS to clinical service.
Proceed to Section 7.
7 FLeakage Current
Test
Leakage current tests assure that under both normal and fault conditions,
any leakage current does not exceed values given in Table 7-1.
1. Perform leakage test with MDS power supply plugged into leakage
tester. See Figure 7-1.
LEAKAGE
TESTER
AC/DC
Power
Adapter
Medside
MONITOR
Data
Station
Figure 7-1 MDS Earth leakage current test setup
2. Follow leakage tester manufacture’s instructions to measure each of
leakage currents given in Table 7-1.
• Earth leakage
• Enclosure leakage (case)
Table 7-1Leakage Current Test
TESTMax. Current
Earth Leakage.5ma@240VAC
.250ma@120VA
Enclosure leakage (case).1ma@240VAC
.05ma@120VAC
3. Verify that current does not exceed values given in Table 7-1.
4. Record all va lues in “Functional Verification Checklist” on page 37
36 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
Location: _____________________________MDS Serial Number: ___________ _Installed SW Version: _________
File a copy of this report with site documentation, and retain a copy for your records. The Siemens RG may also
require a copy of these test results. For MDS’s serviced in U.S.A., also forward copy of completed Functional
Verification Checklist per applicable SSG installation procedure.
r = Test Passed
Power Circuits and
Startup
• Power LEDs_______
• Brief tone_______
• OS loader screen_______
• Correct language_______
Hard Drive _______
256MB Memory• Boot up_______
• BIOS Utility screen_______
Leakage Current Test
• Earth leakage_______
• Enclosure leakage (case)_______
MDS has passed all required tests.
__________________________________ ___________________________________ ________________
Name Printed Signature Date
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers37
MDS_sm.fm/04-02/kaupp
Page 42
Field Service ManualMedsi de Data Station
This page intentionally left blank.
38 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 43
Medside Data StationField Service Manual
Appendix A: Spare Parts
Table A-1Spare Parts
Part Art. No.Part NameDwg. Ref.
72 59 257 E553UDrive Cable Figure A-1 on page 40
72 65 627 E553U128M Memory Module Figure A-2 on page 40
72 59 869 E533UE/M ASY CBL FRNT PANEL MDS Figure A-3 on page 40
72 62 046 E553UPCB ASY Daughterboard Figure A-4 on page 41
72 59 307 E553UBattery Module Figure A-5 on page 41
72 58 812 E553UMEC PRT CVR Front MDS Figure A-6 on page 42
72 59 851 E553UE/M ASY OFF/ON SWITCH Figure A-7 on page 42
72 65 619 E553UEM SPR MDS Hard DriveFigure A-8 on page 42
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers39
MDS_sm.fm/04-02/kaupp
Page 44
Field Service ManualMedsi de Data Station
Figure A-1 Drive Cable
Figure A-2 128M Memory Module
Figure A-3E/M ASY CBL FRNT PANEL MDS
40 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 45
Medside Data StationField Service Manual
Figure A-4 PCB ASY DAUGHTERBOARD
Figure A-5 Battery Mod ule
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers41
MDS_sm.fm/04-02/kaupp
Page 46
Field Service ManualMedsi de Data Station
Figure A-6 MEC PRT CVR FRONT MDS
Figure A-7 E/M ASY OFF/ON SWITCH
Figure A-8 EM SPR MDS Hard Drive
42 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 47
Medside Data StationField Service Manual
Appendix B: Bios Messages
The following is a list of the messages that the BIOS displays. Most error
messages occur during POST test. See “Appendix C: POST Error Codes”
on page 47. Some messages display information about a hardware device,
e.g., the amount of memory installed. Other messages may indicate a
problem with a device, such as the way it has been configured. The
following list of messages includes explanations of error messages and
possible remedies for reported problems.
*If your system displays one of the messages marked below with an
asterisk (*), write down the message and contact TSS Danvers or TSS
Solna. If the MDS fails after making changes in the BIOS Setup menus,
reset the computer, enter BIOS Setup and verify Setup (see Section 3) to
correct the error.
0200 Failure Fixed Disk
Fixed disk is not working or not configured properly. Check to see if fixed
disk is attached properly. Run BIOS Setup. Find out if the fixed-disk type
is correctly identified (see Section 3.1).
0210 Stuck key
Stuck key on keyboard.
0211 Keyboard error
Keyboard not working.
*0212 Keyboard Controller Failed
Keyboard controller failed test. May require replacing keyboard
controller.
0213 Keyboard locked - Unlock key switch
Unlock the system to proceed.
0220 Monitor type does not match CMOS - Run SETUP
Monitor type not correctly identified in Setup
*0230 Shadow Ram Failed at offset: nnnn
Shadow RAM failed at offset
was detected.
*0231 System RAM Failed at offset: nnnn
System RAM failed at offset
was detected.
*0232 Extended RAM Failed at offset: nnnn
nnnn of the 64k block at which the error
nnnn of in the 64k block at which the error
Extended memory not working or not configured properly at offset
*0250 System battery is dead - Replace and run SETUP
The CMOS clock battery indicator shows the battery is dead.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers43
MDS_sm.fm/04-02/kaupp
nnnn.
Page 48
Field Service ManualMedsi de Data Station
0251 System CMOS checksum bad - Default configuration used
System CMOS has been corrupted or modified incorrectly, perhaps by
an application program that changes data stored in CMOS. The BIOS
installed Default Setup Values. If you do not want these values, enter
Setup and enter correct values (see Section 3.1). If the error persists,
contact TSS Danvers/Solna.
*0260 System timer error
The timer test failed. Requires repair of system board.
*0270 Real time clock error
Real-Time Clock fails BIOS hardware test. May require board repair.
0271 Check date and time settings
BIOS found date or time out of range and reset the Real-Time Clock. May
require setting legal date (1991- 2099).
0280 Previous boot incomplete - Default configuration used
Previous POST did not complete successfully. POST loads default values
and offers to run BIOS Setup. If the failure was caused by incorrect
values and they are not corrected, the next boot will likely fail. On
systems with control of wait states, improper Setup settings can also
terminate POST and cause this error on the next boot. Run Setup and
verify that the wait-state configuration is correct. This error is cleared the
next time the system is booted.
*0281 Memory Size found by POST differed from CMOS
Memory size found by POST differed from CMOS.
*02B2 Incorrect Drive A type - run SETUP
Type of floppy drive A: not correctly identified in Setup. Contact TSS
Danvers/Solna.
*02B3 Incorrect Drive B type - run SETUP
Type of floppy drive B: not correctly identified in Setup. Contact TSS
Danvers/Solna.
02D0 System cache error - Cache disabled
RAM cache failed and BIOS disabled the cache. A disabled cache slows
system performance considerably. Contact TSS Danvers/Solna.
*02F0: CPU ID:
CPU socket number for Multi-Processor error.
*02F4: EISA CMOS not writeable
ServerBIOS2 test error: Cannot write to EISA CMOS.
*02F5: DMA Test Failed
ServerBIOS2 test error: Cannot write to extended DMA (Direct Memory
Access) registers.
*02F6: Software NMI Failed
ServerBIOS2 test error: Cannot generate software NMI (Non-Maskable
Interrupt).
44 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 49
Medside Data StationField Service Manual
*02F7: Fail-Safe Timer NMI Failed
ServerBIOS2 test error: Fail-Safe Timer takes too long.
Device Address Conflict
Address conflict for specified
device.
Allocation Error for: device
Run ISA or EISA Configuration Utility to resolve resource conflict for the
specified device.
*CD ROM Drive
CD ROM Drive identified.
Entering SETUP...
Starting Setup program
*Failing Bits: nnnn
The hex number nnnn is a map of the bits at the RAM address which
failed the memory test. Each 1 (one) in the map indicates a failed bit. See
errors 230, 231, or 232 above for offset address of the failure in System,
Extended, or Shadow memory.
Fixed Disk n
Fixed disk
n (0-3) identified.
Invalid System Configuration Data
Problem wi th NVRAM (CMOS) data.
I/O device IRQ conflict
I/O device IRQ conflict error.
PS/2 Mouse Boot Summary Screen:
PS/2 Mouse installed.
nnnn kB Extended RAM Passed
Where
nnnn is the amount of RAM in kilobytes successfully tested.
nnnn Cache SRAM Passed
Where
nnnn is the amount of system cache in kilobytes successfully
tested.
nnnn kB Shadow RAM Passed
Where
nnnn is the amount of shadow RAM in kilobytes successfully
tested.
nnnn kB System RAM Passed
Where
nnnn is amount of system RAM in kilobytes successfully tested.
One or more I2O Block Storage Devices were excluded from the Setup
Boot Menu
There was not enough room in the IPL table to display all installed I2O
block-storage devices.
Operating system not found
Operating system cannot be located on drive C:. Enter Setup and see if
fixed disk properly identified.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers45
MDS_sm.fm/04-02/kaupp
Page 50
Field Service ManualMedsi de Data Station
*Parity Check 1 nnnn
Parity error found in the system bus. BIOS attempts to locate the
address and display it on the screen. If it cannot locate the address, it
displays ????. Parity is a method for checking errors in binary data. A
parity error indicates that some data has been corrupted.
Parity Check 2 nnnn
Parity error found in the I/O bus. BIOS attempts to locate the address and
display it on the screen. If it cannot locate the address, it displays
???.
Press <F1> to resume, <F2> to Setup, <F3> for previous
Displayed after any recoverabl e erro r messa ge. Pres s <F1> to start the
boot process or <F2> to enter Setup and change the settings. Press
<F3> to display the previous screen (usually an initialization error of an
Option ROM, i.e., an add-on card). Write down and follow the
information shown on the screen.
Press <F2> to enter Setup
Optional message displayed during POST.
PS/2 Mouse:
PS/2 mouse identified.
Run the I2O Configuration Utility
One or more unclaimed block storage devices have the Configuration
Request bit set in the LCT. Run an I2O Configuration Utility (e.g. the SAC
utility).
System BIOS shadowed
System BIOS copied to shadow RAM.
UMB upper limit segment address: nnnn
Displays the address
nnnn of the up per lim it of Upper Memory Blocks,
indicating released segments of the BIOS which can be reclaimed by a
virtual memory manager.
Video BIOS shadowed
Video BIOS successfully copied to shadow RAM.
46 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 51
Medside Data StationField Service Manual
Appendix C: POST Error Codes
Recoverable POST ErrorsWhenever a recoverable error occurs during POST, PhoenixBIOS displays
an error message describing the problem.
Terminal POST ErrorsThere are several POST routines that issue a POST Terminal Error
message and shut down the system if the routines fails. Before shutting
down the system, the terminal-error handler issues a beep code signifying
the test point error, writes the error to port 80h, attempts to initialize the
video, and writes the error in the upper left corner of the screen (using both
mono and color adapters).
Test Point Error CodeAt the beginning of each POST routine, the BIOS outputs the test point
error code to I/O address 80h. Use this code during trouble shooting to
establish at what point the system failed and what routine was being
performed. If external hardware error is displayed, (eg.mouse, keyboard,
etc.) Check external hardware and hardware connections, then reboot
MDS. If other errors are displayed, write down error code and contact TSS
Danvers/Solna. If the BIOS detects a terminal error condition, it halts POST
and attempts to display the error code on upper left corner of the screen.
If the system hangs before the BIOS can process the error, the value
displayed at the port 80h is the last test performed. In this case, the screen
does not display the remaining error code.
The following is a list of the checkpoint codes displayed and written at the
start of each test, and the beep codes issued for terminal errors. Unless
otherwise noted, these codes are valid for PhoenixBIOS 4.0 Release 6.x.
Table C-1Checkpoint codes
CodePOST Routine Description
02hVerify Real Mode
03HDisable Non-Maskable Interrupt (NMI)
04hGet CPU type
06hInitialize sys tem har dwa re
07hDisable shadow and execute code from ROM
08hInitialize chipset with initial POST values
09hSet IN POST flag
0AhInitia lize CPU registers
0BhEnable CPU cache
0ChInitialize cache to initial POST values
0EhInitialize I/O component
0FhInitialize the local bus IDE
10hInitialize Power Management
11hLoad alternate registers with values POST values
12hRestore CPU control word during warm boot
13hInitialize PCI Bus Mastering devices
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers47
MDS_sm.fm/04-02/kaupp
Page 52
Field Service ManualMedsi de Data Station
Table C-1Checkpoint codes (Continued)
14hInitialize keyboar d contr oller
16hBIOS ROM checksum
17hInitialize cache before memory Auto size
18h8254 timer initiali zat ion
1Ah8237 DMA controller initialization
1ChRese t Programmable Interrupt Controller
20hTest DRAM refresh
22hTest 8742 keyboard controller
24hSet ES segment register to 4 GB
28hAuto size DRAM
29hInitialize POST Memory Manager
2AhClear 512kb base RAM
2ChRAM failure on address line xxxx
2EHRAM failure on address line xxxx* of low byte of
memory bus
2FhEnable cache before system BIOS shadow
32hTest CPU bus-clock frequency
33hInitialize Phoenix Dispatch Manager
36hWarm start shut down
38hShadow system BIOS ROM
3AhAuto size cache
3ChAdva nce d configuration of chipset registers
3DhLoad alternative registers with CMOS values
41hInitialize extended memory for RomPilot
42hInitialize inter ru pt vec to rs
45hPOST device initialization
46hCheck ROM copyright notice
47hInitialize I20 sup po rt
48hCheck video configuration against CMOS
49hInitialize PCI bus devices and devices
4AhInitialize all video adapters in system
4BhQuietBoot start (optional)
4ChShadow video BIOS ROM
4EhDisplay BIOS copyright notice
4FhInitialize MultiBoot
50hDisplay CPU type and speed
51hInitialize EISA bo ard
52hTest Keyboard
54hSet key click if enabled
55hEnable USB devices
58hTest for unexpected interrupts
59hInitialize POST dis play
5AhDisplay prompt “Press F2 to enter SETUP”
5BhDisplay CPU cache
5ChTest RAM between 512 and 640 kb
48 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 53
Medside Data StationField Service Manual
Table C-1Checkpoint codes (Continued)
60hTest extended memory
62hTest extended memory address line
64hJump to User Patch1
66hConfigure advanced cache register
67hInitialize Mul ti Proces s or AP IC
68hEnable external and CPU caches
69hSetup System Management Mode (SSM) area
6AhDisplay external L2 cache size
6BhLoad custom defaults (optional)
6ChDisplay shadow-area message
6EhDisplay possible high address for UMB recovery
70hDisplay error messages
72hCheck for configuration errors
76hCheck for keyboard errors
7ChSetup hardware interrupt vectors
7DhInitialize Intelligent System Monitoring
7EhInitialize coprocessor if present
80hDisable on board Super I/O ports and IRQ’s
81hLate POST device initialization
82hDetect and install external RS232 ports
83hConfigure non-MCD IDE controllers
84hDetect and install external parallel ports
85hInitialize PC compatible PnP ISA devices
86hRe-initialize on board I/O ports
87hConfigure Motherboard Devices
88hInitialize BIOS Data Area
89hEnable Non-Maskable Interrupts (NMI’s)
8AhInitialize Extended BIOS Data Area
8BhTest and initialize PS/2 mouse
8ChI nitialize floppy controll er
8FhDetermine number of ATA drives (optional)
90hInitialize hard-d isk cont ro ller s
91hInitialize local-bus hard-disk controllers
92hJump to UserPatch2
93hBuild MPTABLE for multi-processor boards
95hInstall CD ROM for bo ot
96hClear huge ES segment register
97hFix up Multi Processor table
98hSearch for optional ROM’s. One long, two short
beeps on checksum failure
99hCheck for SMART drive (optional)
9AhShadow option ROM”s
9ChSet up Power Management
9DhInitialize security engine (optional)
9EhEnable hardware interrupts
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers49
MDS_sm.fm/04-02/kaupp
Page 54
Field Service ManualMedsi de Data Station
Table C-1Checkpoint codes (Continued)
9FhDetermine number of ATA and SCSI’s drives
A0hSet time of day
A2hCheck key lock
A4hInitialize typematic rate
A8hErase F2 prompt
AAhScan for F2 key stroke
AChEnter Setup
AEhClear Boot flag
B0hCheck for errors
B1hInform RomPilot about the end of POST.
B2hPOST done - prepare to boot operating system
B4hOne short beep before boot
B5hTerminate QuietBoot (optional)
B6hCheck password (optional)
B7hI nitialize AC PI BIOS
B9hPrepare Boot
BAhInitialize SMBIOS
BBhInitialize PnP Option ROM”s
BChClear parity checkers
BDhDisplay MultiBoot menu
BEhClear screen (optional)
BFhCheck virus and backup reminders
C0hTry to boot with INT 19
C1hInitia lize POST Error Manager (PEM)
C2hInitia lize error logging
C3hInitialize error display function
C4hInitia lize sys tem error handler
C5hPnPnd dual CMOS (optional)
C6hInitialize note dock (optional)
C7hInitialize note dock late
C8hForce check (optional)
C9hExtended checksum (optional)
CAhRedirect Int 15h to enable remote keyboard
CBhRedirect Int 13h to Memory Technology Devices such as
ROM,RAM, PCMCIA, and serial disk
CChRedirect Int 10h to enable remote serial video
CDhRe-map I/O and memory for PCMCIA
CEhInitialize digitizer and display message
D2hUnknown interrupt
The following are for boot block in Flash ROM
E0hInitialize the chipset
E1hInitialize the bridge
E2hInitialize the CPU
E3hInitialize system timer
E4hInitialize system I/O
50 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 55
Medside Data StationField Service Manual
Table C-1Checkpoint codes (Continued)
E5hCheck force recovery boot
E6hChecksum BIOS ROM
E7hGo to BIOS
E8hSet Huge Segment
E9hInitialize Multi Processor
EAhInitialize OEM special code
EBhInitialize PIC and DMA
EChInitialize Memory type
EDhInitialize Memory size
EEhShadow Boot Block
EFhSystem memory test
F0hInitialize interrupt vec to rs
F1hInitialize Run Time Clock
F2hInitialize video
F3hInitialize System Management Manager
F4hOutput One beep
F5hClear Huge Segment
F6hBoot to Mini DOS
F7hBoot to Full DOS
* If BIOS detects error code 2C, 2E, or 30 (base 512K RAM error), it
displays an additio nal word-bitmap (xxxx) address line after the error code.
For example, "2C 0002" means address line 1 (bit one set) has failed. "2E
1020" means data bits 12 and 5 (bits 12 and 5 set) have failed in lower 16
bits.
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers51
MDS_sm.fm/04-02/kaupp
Page 56
Field Service ManualMedsi de Data Station
This page intentionally left blank.
52 Siemens Medical Solutions, EM-PCS, Danvers ASK-T962-02-7600
MDS_sm.fm/04-02/kaupp
Page 57
Medside Data StationField Service Manual
For additional support, Siemens customers can contact their local Siemens Service Representatives. Siemens
Customer Support Engineers can contact the following as required:
In North and South America:In Europe, Asia, Africa, Australia and New Zealand
Siemens Medical Systems, Inc.Siemens-Elema AB
EM-PCSEM
Technical Service and SupportTechnical Service and Support
16 Electronics Avenue171 95 Solna, Sweden
Danvers, MA 01923 USA
Tel: (978) 907-7500Tel: Int+46-8-730-7641
FAX: (978) 907-7546FAX: Int+46-8-986 662
ASK-T962-02-7600Siemens Medical Solutions, EM-PCS Danvers53
MDS_sm.fm/04-02/kaupp
Page 58
Medside Data Station - Field Ser vice Manual
Order No. 74 94 144 E553U