Siemens Acuson X600, Acuson X700 User manual [gb]

Page 1
ACUSON X600 ACUSON X700 Diagnostic Ultrasound System Instructions for Use
Siemens Medical Solutions USA, Inc. 10789007-ABS-001-01-01
Page 2
www.siemens.com/healthcare
ACUSON X600, Product Version 1.0
ACUSON X700, Product Version 2.0
Software Version X700 VB20, X600 VA10
Cadence, Clarify, DIMAQ, Dynamic TCE, DTI, eSie Touch, Evolve Package, fourSight, microCase, MultiHertz, QuickSet, SuppleFlex, Siemens, SieClear, SieScape, SwiftLink, syngo, TGO, THI, Velocity Vector Imaging, X600, X700, and 3-Scape are trademarks or registered trademarks of Siemens AG or its subsidiaries and affiliates.
All other product names are trademarks or registered trademarks of their respective companies.
Siemens reserves the right to change system specifications at any time.
Global Business Unit and Legal Manufacturer
Siemens Medical Solutions USA, Inc. Ultrasound 685 East Middlefield Road Mountain View, CA 94043 U.S.A. Phone: +1-888-826-9702 www.siemens.com/healthcare
CE Declaration This product is provided with a CE marking in
accordance with the regulations stated in Council Directive 93/42/EEC of June 14, 1993 concerning Medical Devices. The CE marking only applies to medical devices that have been put on the market according to the above referenced Council Directive.
Unauthorized changes to this product are not covered by the CE marking and the related Declaration of Conformity.
EU Authorized Representative
Siemens AG Medical Solutions Henkestrasse 127 91052 Erlangen Germany
Siemens Global Headquarters
Siemens AG Wittelsbacherplatz 2 80333 Muenchen Germany
Global Siemens Healthcare Headquarters
Siemens AG Healthcare Sector Henkestrasse 127 91052 Erlangen Germany Phone: +49 9131 84-0
Page 3

Contents

Chapter 1
Introduction
Chapter 2
Safety and Care
Chapter 4
Examination Fundamentals
Chapter 6
Transesophageal Transducer
Chapter 7
Specialty Transducers
Chapter 8
Physiologic Function
General overview of the diagnostic ultrasound imaging system.
Detailed information on system safety and how to care for and maintain the system, transducers, and transducer accessories.
Chapter 3 System Setup
Detailed descriptions of how to transport, set up, and prepare the system for use, including transducer connection and system startup procedures.
Information on starting an examination, including instructions for entering and editing patient data and selecting an exam type, imaging mode, and transducer.
Chapter 5 Transducer Accessories and Biopsy
Attachment procedures for transducer accessories and an explanation of the biopsy (puncture) function, including a procedure for needle path verification.
Description of the transesophageal transducer, including cleaning and care information for the transducer.
Description of the following specialty transducers:
EV9F3
Explanation of the physiologic function.
Instructions for Use i
Page 4
Appendix A
Technical Description
Technical description of the ultrasound system.
Appendix B
Control Panel
Appendix E
Reserved for Future Use
Appendix F
Acoustic Output Reference
Acoustic output reporting tables.
Explanation of all controls and keys on the control panel, alphanumeric keyboard, and optional footswitch.
Appendix C On-screen Controls Explanation of all on-screen controls, including menus and soft key selections.
Appendix D Advanced Feature Controls Explanation of the on-screen controls for advanced features.
Note: Not all features and options described in this publication are available to all users. Please
check with your Siemens representative to determine the current availability of features and options.
ii Instructions for Use
Page 5

About the User and Reference Manuals

Publication
Includes
Explanation of the clinical software programs for use on the ultrasound system
Clinical references
The user and reference manuals contain descriptions for the following ultrasound systems:
ACUSON X600 diagnostic ultrasound system  ACUSON X700 diagnostic ultrasound system
Features and options unique to an ultrasound system are identified in Chapter 1 and Appendix A of the Instructions for Use.
The user and reference manuals consist of the following publications.
Instructions for Use Intended Audience
Technical description of the ultrasound system Safety and care information for the system and compatible transducers Descriptions of all system controls Procedures for system setup, examination fundamentals, and the biopsy
function
Acoustic output data
Features and Applications Reference*
System Reference* Description of customizable system settings
Electromagnetic Emissions and Immunity: Guidance and Manufacturer's Declaration*
*Languages supported by the user interface include a translation of this publication.
Descriptions of image acquisition and optimization, including optional imaging
features
General and exam-specific measurements and calculations Data management
Information about DICOM connectivity, network capabilities, and external
devices
  Information regarding the electromagnetic compatibility (EMC) testing of this
system
Instructions for Use iii
Page 6
Conventions
SET
Take a moment to familiarize yourself with these conventions.
Warnings, Cautions, and Notes
Cross-References Examples:
Customizable System Settings System settings available for customization are depicted as shown.
Keys and Controls
On-screen Objects
Selection of On-screen Objects
WARNING: Warnings are intended to alert you to the importance of
following the correct operating procedures where risk of injury to the patient or system user exists.
Caution: Cautions are intended to alert you to the importance of following
correct operating procedures to prevent the risk of damage to the system.
Note: Notes contain information concerning the proper use of the system and/or correct execution of a procedure.
See also: Biohazards, Safety and Care, Chapter 2, Instructions for Use
See also: Documentation Devices, Chapter 2, System Reference
See also: Alphanumeric Keyboard, p. 26
Example:
Default Settings > Automatic Freeze Response
Keys and controls located on the control panel are identified by uppercase, boldface type.
Example: Rotate the 2D control. Keys located on the keyboard are identified by boldface type. Example: Press the Exam key. On-screen objects such as menu selections, soft key selections, and buttons are
identified by boldface type. Example: The system displays the Patient Registration form. The SET key on the control panel functions as a point-and-select device (similar
to a computer mouse) when used with the trackball. "Select" or "click" describes this action: Roll the trackball to position the pointer (cursor) on an on-screen object and then
press the SET key. "Double-click" describes this action: Roll the trackball to position the pointer (cursor) on an on-screen object and then
press the SET key twice. "Drag" describes this action: Roll the trackball to position the pointer (cursor) on an on-screen object and then
press and hold the SET key. Roll the trackball to reposition the object and then release the
key.
iv Instructions for Use
Page 7
Intended Audience
User
Interaction with Ultrasound Equipment
Expected Experience and Other Characteristics
training in the field of cardiology
activities.
The intended audience for the user and reference manuals includes the following users.
Sonographer
Cardiologist
Maternal-fetal Medicine Obstetrician/ Perinatologist
Radiologist and Internist
System Administrator and Customer Service Engineer
Acquires diagnostic views of anatomy,
blood flow, and related pathology
Performs measurements and analysis of
the acquired images
Prepares exam data for review and
interpretation by a qualified physician
Performs invasive and non-invasive
ultrasound exams
Interprets exam data, including
echocardiography exam data
Writes and assembles exam findings in a
report
Performs ultrasound exams Interprets exam data Writes and assembles exam findings in a
report
Performs ultrasound exams Interprets exam data Writes and assembles exam findings in a
report
Configures the ultrasound system for use
in a networked environment
Ranges from novices (for example, students) to
advanced practitioners with certification in multiple subspecialties
Educated in anatomy, physiology, patient care,
and identification of pathology in ultrasound images
Many sonographers have a Bachelor's degree;
some have advanced degrees in related health care subjects
Medical doctor Expert in diagnostic imaging, including
computed tomography (CT), magnetic resonance imaging (MRI), X-ray, ultrasound, and nuclear medicine
Advanced training in imaging physics with
typically four to six years of post-doctoral
Medical doctor Manages high-risk obstetrical patients for the
safe and successful delivery of the fetus
Skilled in interpreting ultrasound exam data Medical doctors Expert in diagnostic imaging, including CT, MRI,
X-ray, ultrasound, and nuclear medicine
Advanced training in imaging physics with
typically two to six years of post-doctoral training in the field of radiology
A System Administrator is an individual within
your organization who is designated to set up system parameters to connect the ultrasound system or workstation to a picture archiving and communication system (PACS).
Customer Service Engineers are Siemens
representatives who configure the ultrasound system or workstation during software installation and support troubleshooting
Instructions for Use v
Page 8
vi Instructions for Use
Page 9

1 Introduction

System Overview ................................................................................................ 3
System Review .............................................................................................. 4
Intended Use ....................................................................................................... 7
ACUSON X600 Ultrasound System ............................................................... 7
Indications for Use Statement ................................................................ 8
Transducers and Intended Applications ................................................. 9
ACUSON X700 Ultrasound System ............................................................. 12
Indications for Use Statement .............................................................. 13
Transducers and Intended Applications ............................................... 14
Image Screen Layout ........................................................................................ 18
Screen Saver ............................................................................................... 18
Sample Image Screen ................................................................................. 19
Status Bar ............................................................................................. 20
User-Defined System Settings ........................................................................ 20
QuickSets .......................................................................................................... 20
Instructions for Use 1 - 1
Page 10
1 Introduction
1 - 2 Instructions for Use
Page 11
1 Introduction
System Overview
The ACUSON X600 and ACUSON X700 systems are portable, digital diagnostic ultrasound imaging systems. The systems utilize advanced imaging processing and transducer technology. The operating systems are based on Windows technology.
The system software supports standard applications, exam-specific imaging presets, measurements, pictograms, annotations, reports, worksheets, and system diagnostics.
Operating modes for the system include:
2D-mode  Split mode  Dual-mode  4B-mode  2D/M-mode  Anatomical M-mode (for Cardiac imaging)  Pulsed Doppler  Color Doppler  Power Doppler  Steerable Continuous Wave Doppler  Auxiliary Continuous Wave Doppler
See also: Technical Description, Appendix A, Instructions for Use
Instructions for Use 1 - 3
Page 12
1 Introduction
System Review
1 User adjustable monitor 2 Storage shelf 3 Transducer cable hanger 4 Air filter 5 Black and white printer 6 Physio module (with ECG and
auxiliary connectors) 7 CD/DVD-R/W drive 8 Rear shelf 9 Power (partial) ON/OFF switch ( ) 10 Articulating arm
Example of the ultrasound system, left front view.
1 - 4 Instructions for Use
Page 13
1 Introduction
1 Alphanumeric keyboard 2 Transducer holder 3 Transducer cable hanger 4 Swivel wheel with brake 5 Transducer port for continuous wave
(pencil) transducers 6 Transducer ports 7 Integrated Gel warmer 8 Front handle 9 Back-lit control panel
Example of the ultrasound system, right front view.
Instructions for Use 1 - 5
Page 14
1 Introduction
1 Monitor 2 Cable hanger 3 Power panel with circuit breaker 4 Input/Output panel
Example of the ultrasound system, back view.
1 - 6 Instructions for Use
Page 15
1 Introduction
Intended Use
WARNING: The analysis of results from an ultrasound examination requires that you are trained
in the interpretation of diagnostic ultrasound studies and are qualified to make clinical diagnoses.
Caution: In the United States of America, federal law restricts this device to sale or use by, or
on the order of, a physician.
ACUSON X600 Ultrasound System
The ACUSON X600 ultrasound system supports the following applications:
Abdomen  Aorta  Breast  Cardiac  Cerebrovascular  Early Obstetrics  Emergency Medicine  Fetal Echo  Gynecology  Musculoskeletal  Obstetrics  Obstetrics (Japan)  Orthopedic  Pediatric Abdomen  Pediatric Echo  Pelvic Floor  Peripheral vascular  Renal  Small Parts  Testicle  Thyroid  Transcranial  Urology  Venous
Instructions for Use 1 - 7
Page 16
1 Introduction
Indications for Use Statement
Product Indications for Use Statement
ACUSON X600 Ultrasound System
The ACUSON X600 ultrasound imaging system is intended for the following applications: Cardiac Adult, Cerebrovascular, Peripheral Vascular, Transcranial, Fetal, Abdominal, Pediatric, Small Organ, Adult Cephalic, Musculo-skeletal Conventional, Obstetrical, Gynecological (including monitoring of ovarian follicle development), and Urological applications using different ultrasound transducers for different applications.
The recommended use of each transducer may vary depending on the transducer design, anatomical study, patient size and imaging approach. The system also provides the ability to measure anatomical structures and supports calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Caution: Ultrasound is used as an imaging aid, but may have further restrictions
specific to in-vitro fertilization (IVF), chorionic villus sampling (CVS), and percutaneous umbilical cord blood sampling (PUBS) procedures. Observe local laws and regulations.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
Note: This feature can be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging."
1 - 8 Instructions for Use
Page 17
1 Introduction
CURVED AND LINEAR ARRAY TRANSDUCERS
Transducers and Intended Applications
EMC Note: Operating the transducer in close proximity to sources of strong electromagnetic fields,
such as radio transmitter stations or similar installations may lead to temporary degradation or interference visible on the monitor screen. A lightening of image background may be noticed while visualizing hypoechoic structures, or color spectral interference, or jitter, or horizontal lines in the image screen may occur. The transducer and the system have been designed and tested to withstand such interference and will not be permanently damaged.
See Also: Electromagnetic Emissions and Immunity Guidance and Manufacturer's Declaration
Only the following transducers from Siemens are compatible with the Siemens ACUSON X600 ultrasound imaging system.
TRANSDUCER NAME
OPERATING FREQUENCY
1
MODES OF OPERATION2
INTENDED APPLICATIONS
CH5-2 2D-mode:
2.5 MHz–5.0 MHz Doppler:
2.0 MHz–2.7 MHz
C6-2 2D-mode:
3.1 MHz–5.0 MHz Doppler:
2.5 MHz–3.3 MHz
VF10-5 2D-mode:
6.2 MHz–10.0 MHz Doppler:
5.3 MHz–6.2 MHz
2D, C, D, M Abdomen
Early Obstetrics Emergency Medicine Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Peripheral Vascular Renal Urology Venous
2D, C, D, M Abdomen
Aorta Early Obstetrics Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Pediatric Abdomen Renal Venous
2D, C, D, M Breast
Cerebrovascular Emergency Medicine Musculoskeletal Orthopedics Peripheral Vascular Testicle Thyroid Venous
Instructions for Use 1 - 9
Page 18
1 Introduction
NAME
FREQUENCY1
OPERATION2
INTENDED APPLICATIONS
PHASED ARRAY TRANSDUCERS
ENDOCAVITY TRANSDUCERS
fourSight 4D TRANSDUCERS
Pelvic Floor
TRANSDUCER
OPERATING
VF12-4 2D-mode:
6.2 MHz–11.4 MHz Doppler:
4.0 MHz–6.7 MHz
4V1c 2D-mode:
1.7 MHz–4.0 MHz Doppler:
1.8 MHz–2.5 MHz
EC9-4w 2D-mode:
4.2 MHz–7.3 MHz Doppler:
4.0 MHz–5.3 MHz
MODES OF
2D, C, D, M Breast
Cerebrovascular Emergency Medicine Musculoskeletal Orthopedic Peripheral Vascular Small Parts Testicle Thyroid Venous
2D, C, D, M, SCW Abdomen
Cardiac Emergency Medicine Transcranial
2D, C, D, M Early Obstetrics
Gynecology Obstetrics Obstetrics (Japan) Urology
C6F2 2D-mode:
2.9 MHz–5.0 MHz Doppler:
2.5 MHz–3.1 MHz
C8F3 2D-mode:
3.3 MHz–5.7 MHz Doppler:
2.7 MHz–3.3 MHz
EV9F3 2D-mode:
4.7 MHz–8.0 MHz Doppler:
4.0 MHz–5.3 MHz
2D, C, D, M Abdomen
Early Obstetrics Fetal Echo Gynecology Obstetrics Obstetrics (Japan)
2D, C, D, M Abdomen
Early Obstetrics Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Pelvic Floor
2D, C, D, M Early Obstetrics
Gynecology Obstetrics Obstetrics (Japan) Pelvic Floor
1 - 10 Instructions for Use
Page 19
1 Introduction
NAME
FREQUENCY1
OPERATION2
INTENDED APPLICATIONS
1
Operating Frequency
Range of selectable operating frequencies for:
2D (brightness mode)
2D-mode, 2D-mode with Tissue Harmonic Imaging (THI)
transducers)
TRANSDUCER
OPERATING
MODES OF
CONTINUOUS WAVE TRANSDUCERS
CW2 2 MHz CW Cardiac
Pediatric Echo
CW5 5 MHz CW Cerebrovascular
Pediatric Echo
2D-mode Fundamental and harmonic imaging, including contrast Doppler Pulsed wave, continuous wave, and color imaging
2
Modes of Operation Includes one or more of the following system operating modes
C (color flow imaging) Color Doppler, Power Doppler D (Doppler)
Pulsed Wave Doppler, 2D/Doppler, 2D/Doppler with Color, 2D/Doppler with Power
M (motion mode)
M-mode, M-mode with THI, 2D/M-mode, 2D/M-mode with Color, 2D/M-mode with Power
SCW (steerable continuous wave
Steerable Continuous Wave Doppler (for phased array transducers)
Doppler) CW (continuous wave Doppler)
Auxiliary Continuous Wave Doppler (for continuous wave [pencil]
Instructions for Use 1 - 11
Page 20
1 Introduction
ACUSON X700 Ultrasound System
The ACUSON X700 ultrasound system supports the following applications:
Abdomen  Breast  Cardiac  Cerebrovascular  Early Obstetrics  Emergency Medicine  Fetal Echo  Gynecology  Musculoskeletal  Neonatal Head  Obstetrics  Obstetrics (Japan)  Orthopedic  Pediatric Abdomen  Pediatric Echo  Pelvic Floor  Penile  Peripheral vascular  Renal  Small Parts  Superficial Musculoskeletal  TEE (Adult transesophageal)  Testicle  Thyroid  Transcranial  Urology  Venous
1 - 12 Instructions for Use
Page 21
1 Introduction
Indications for Use Statement
Product Indications for Use Statement
ACUSON X700 Ultrasound System
The ACUSON X700 ultrasound imaging system, is intended for the following applications: Cardiac Adult, Cardiac Pediatric, Cardiac Transesophageal, Intracardiac Echo, Cerebrovascular, Peripheral Vascular, Transcranial, Fetal, Abdominal, Intra-operative (Specification), Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Musculo­skeletal Conventional, Musculo-skeletal Superficial, Obstetrical, Gynecological (including monitoring of ovarian follicle development), and Urological applications using different ultrasound transducers for different applications.
The recommended use of each transducer may vary depending on the transducer design, anatomical study, patient size and imaging approach. The system also provides the ability to measure anatomical structures and supports calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Caution: Ultrasound is used as an imaging aid, but may have further restrictions
specific to in-vitro fertilization (IVF), chorionic villus sampling (CVS), and percutaneous umbilical cord blood sampling (PUBS) procedures. Observe local laws and regulations.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
Note: This feature can be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging."
ACUSON AcuNav Ultrasound Catheter
The catheter is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
Instructions for Use 1 - 13
Page 22
1 Introduction
CURVED AND LINEAR ARRAY TRANSDUCERS
Transducers and Intended Applications
EMC Note: Operating the transducer in close proximity to sources of strong electromagnetic fields,
such as radio transmitter stations or similar installations may lead to temporary degradation or interference visible on the monitor screen. A lightening of image background may be noticed while visualizing hypoechoic structures, or color spectral interference, or jitter, or horizontal lines in the image screen may occur. The transducer and the system have been designed and tested to withstand such interference and will not be permanently damaged.
See Also: Electromagnetic Emissions and Immunity Guidance and Manufacturer's Declaration
Only the following transducers from Siemens are compatible with the Siemens ACUSON X700 ultrasound imaging system.
TRANSDUCER NAME
OPERATING FREQUENCY
1
MODES OF OPERATION2
INTENDED APPLICATIONS
4C1 2D-mode:
2.2 MHz-5.0 MHz Doppler:
2.2 MHz-3.1 MHz
6C2 2D-mode:
2.5 MHz-5.7 MHz Doppler:
2.7 MHz-3.6 MHz
C8-5 2D-mode:
4.2 MHz-7.6 MHz Doppler:
4.2 MHz-5.0 MHz
VF10-5 2D-mode:
6.2 MHz-10.0 MHz Doppler:
5.3 MHz-6.2 MHz
2D, C, D, M
Abdomen Early Obstetrics Emergency Medicine Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Peripheral Vascular Renal Urology Venous
2D, C, D, M Abdomen
Early Obstetrics Emergency Medicine Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Renal
2D, C, D, M Cerebrovascular
Neonatal Head Pediatric Abdomen Pediatric Echo
2D, C, D, M Breast
Cerebrovascular Emergency Medicine Musculoskeletal Orthopedics Peripheral Vascular Testicle Thyroid Venous
1 - 14 Instructions for Use
Page 23
1 Introduction
TRANSDUCER NAME
FREQUENCY1
OPERATION2
INTENDED APPLICATIONS
PHASED ARRAY TRANSDUCERS
4.0 MHz-5.0 MHz
OPERATING
VF12-4 2D-mode:
6.2 MHz-11.4 MHz Doppler:
4.0 MHz-6.7 MHz
VF16-5 2D-mode:
8.0 MHz-13.3 MHz Doppler:
5.3 MHz-6.2 MHz
4V1c 2D-mode:
1.7 MHz-4.0 MHz Doppler:
1.8 MHz-2.5 MHz
P8-4 2D-mode:
3.8 MHz-6.7 MHz Doppler:
3.6 MHz-4 MHz
V5Ms 2D-mode:
4.0 MHz-5.7 MHz Doppler:
3.6 MHz
AcuNav 8F/ SoundStar eco 8F
2D-mode:
5.0 MHz-8.9 MHz Doppler:
4.0 MHz-5.0 MHz
AcuNav 10F/ SoundStar 10F/ SoundStar eco 10F
2D-mode:
5.0 MHz-8.9 MHz Doppler:
MODES OF
2D, C, D, M Breast
Cerebrovascular Emergency Medicine Musculoskeletal Orthopedics Peripheral Vascular Small Parts Testicle Thyroid Venous
2D, C, D, M Breast
Cerebrovascular Emergency Medicine Penile Superficial Musculoskeletal Testicle Thyroid
2D, C, D, M, SCW
Abdomen Cardiac Emergency Medicine Transcranial
2D, C, D, M, SCW
Neonatal Head Pediatric Abdomen Pediatric Echo Renal
2D, C, D, M,
Transesophageal Echo
SCW
2D, C, D, M, SCW
Cardiac Adult Cardiac Pediatric Intra-Cardiac
2D, C, D, M, SCW
Cardiac Adult Cardiac Pediatric Intra-Cardiac
Instructions for Use 1 - 15
Page 24
1 Introduction
TRANSDUCER NAME
FREQUENCY1
OPERATION2
INTENDED APPLICATIONS
Urology
fourSight 4D TRANSDUCERS
OPERATING
MODES OF
ENDOCAVITY TRANSDUCERS
EC9-4w 2D-mode:
4.2 MHz-7.3 MHz Doppler:
4.0 MHz-5.3 MHz
2D, C, D, M Early Obstetrics
Gynecology Obstetrics Obstetrics (Japan) Urology
MC9-4 2D-mode:
4.2 MHz-7.3 MHz Doppler:
4.0 MHz-5.3 MHz
2D, C, D, M Early Obstetrics
Gynecology Obstetrics Obstetrics (Japan)
C6F2 2D-mode:
2.9 MHz-5.0 MHz Doppler:
2.5 MHz-3.1 MHz
2D, C, D, M Abdomen
Early Obstetrics Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Pelvic Floor
C8F3 2D-mode:
3.3 MHz-5.7 MHz Doppler:
2.7 MHz-3.3 MHz
2D, C, D, M Abdomen
Early Obstetrics Fetal Echo Gynecology Obstetrics Obstetrics (Japan) Pelvic Floor
EV9F3 2D-mode:
4.7 MHz-8.0 MHz Doppler:
4.0 MHz-5.3 MHz
2D, C, D, M Early Obstetrics
Gynecology Obstetrics Obstetrics (Japan) Pelvic Floor
CONTINUOUS WAVE TRANSDUCERS
CW2 2 MHz CW Cardiac
Pediatric Echo
CW5 5 MHz CW Pediatric Echo
Cerebrovascular
1 - 16 Instructions for Use
Page 25
1 Introduction
1
Operating Frequency
Range of selectable operating frequencies for:
2
Modes of Operation
Includes one or more of the following system operating modes
C (color flow imaging)
Color Doppler, Power Doppler
2D-mode Fundamental and harmonic imaging, including contrast Doppler Pulsed wave, continuous wave, and color imaging
2D (brightness mode) 2D-mode, 2D-mode with Tissue Harmonic Imaging (THI)
D (Doppler)
M (motion mode)
SCW (steerable continuous wave Doppler)
CW (continuous wave Doppler)
Pulsed Wave Doppler, 2D/Doppler, 2D/Doppler with Color, 2D/Doppler with Power
M-mode, M-mode with THI, 2D/M-mode, 2D/M-mode with Color, 2D/M­mode with Power
Steerable Continuous Wave Doppler (for phased array transducers)
Auxiliary Continuous Wave Doppler (for continuous wave [pencil] transducers)
Instructions for Use 1 - 17
Page 26
1 Introduction
Image Screen Layout
The monitor on the ultrasound system displays clinical images together with important operating parameters and patient data. There is a variety of on-screen overlays and graphical objects to aid in image evaluation.
Many fields or areas of data displayed on the screen are multi-functional. The image field can display a 2D-mode image, M-mode sweep, Doppler spectrum, and their combinations, sets of calipers, pictograms and annotation text, and CINE icons. An image can be inverted on a vertical axis and reversed on a horizontal axis to facilitate viewing and measurements.
The system displays reduced-size reference images (thumbnails) of images, clips, and volumes stored during the exam.
EMC Note: Operating the ultrasound imaging system in close proximity to sources of strong electromagnetic fields, such as radio transmitter stations or similar installations, may lead to interference visible on the monitor screen. However, the device has been designed and tested to withstand such interference and will not be permanently damaged.
Screen Saver
The screen saver feature automatically replaces the display with a blank screen after the system has been inactive for a specified number of minutes.
When the screen saver is activated, the system is automatically placed into freeze mode.
Restore the screen display by pressing any key or adjusting any control. The first key that you press will restore the view without performing a function. Press the key again to execute the command.
Use the system presets to enable the screen saver feature and to specify the delay period and the background for the blank screen.
Display > Monitor > Enable Screen Saver
Display > Monitor > Screen Saver Delay
Display > Monitor > Screen Saver Type
1 - 18 Instructions for Use
Page 27
1 Introduction
Sample Image Screen
Example of a typical image screen.
1 Information for identifying the patient, operator, and the hospital or clinic. 2 Time and Date 3 Panels including Thumbnails with controls for printing, deleting, and deselecting an image, clip, or
volume, Patient Info for displaying patient information and Results for displaying measurement results.
4 Soft key selections for a mode or function, including controls for image optimization, post-processing,
measurement types, annotations, and playback. Use the corresponding toggle key on the control panel to activate the selection. Rotate the PAGE control on the control panel to access additional selections for the mode or function. Press the PAG E control to access soft key selections for other modes or functions.
5 Status bar. Indicates the function currently assigned to the trackball, SELECT control, UPDATE key,
ESCAPE key.
Provides an indication of the following functions: video recording or paused recording, signal strength of
the wireless connection signal, and insert text or overwrite text for annotations. 6 Tab Indicator. Identifies the mode or function for the soft key selections. 7 Page Indicator for the soft key selections. The number of available "pages" for a specific mode or
function are identified by the number of sections in the "page indicator". The highlighted section
corresponds to the displayed "page" of soft key selections. 8 During the Measurement function, the system displays the Measurement Label menu for the priority
imaging mode and exam type. Use the CALIPER key to access the Measurement Menu. 9 Imaging Parameters. Displays settings for parameters adjusted by the soft key selections. 10 Transducer Frequency 11 Gain for the priority mode 12 Active Exam type 13 Active transducer name
Instructions for Use 1 - 19
Page 28
1 Introduction
if any.
Status Bar
Icon Description
Indicates the function currently assigned to the SELECT control.
Indicates the function currently assigned to the UPDATE key.
Indicates the function currently assigned to the ESCAPE key. Press the ESCAPE key to cycle through the functions for other active modes under control of the key,
Indicates the function currently assigned to the trackball. Press the SET key to cycle through other functions under control of the trackball, if any.
Indicates recording status during video recording using a DVR. A red icon displays during recording and a white icon displays during pause. The recording time also displays beside the icon.
When displayed, indicates recording function with the built-in microphone is activated.
Indicates strength of wireless connection signal.
User-Defined System Settings
You can customize many features of the ultrasound system by using the system presets to designate default settings, or presets. The values are stored in non-volatile memory and will remain intact when the system is powered off.
Each user of the system can determine settings for imaging preferences and default settings, then store them on a disk. Those user-defined settings can then be loaded along with new system software. The disk also serves as a backup.
QuickSets
The QuickSet feature allows you to capture an optimized configuration of imaging parameter settings for a combination of a specific transducer and exam.
1 - 20 Instructions for Use
Page 29

2 Safety and Care

Operating Safety and Environment ................................................................... 3
System Symbols ............................................................................................ 3
Labels ............................................................................................................ 8
Biohazard Considerations............................................................................ 10
Note on Fetal Examinations ................................................................. 11
Acoustic Output — Mechanical and Thermal Indices .................................. 11
Mechanical and Thermal Indices .......................................................... 12
Transmit Power Control ........................................................................ 13
Transmit Power Display........................................................................ 14
Imaging Functions that Change Acoustic Output ................................. 15
Transducer Surface Temperature Limits ..................................................... 16
Electrical Safety ........................................................................................... 17
Level of Protection Against Electrical Shock — Transducers ............... 19
Defibrillators ......................................................................................... 19
Defibrillator Use and the ACUSON AcuNav Ultrasound Catheter ........ 19
Implantable Devices ............................................................................. 19
Possible Combinations with Other Equipment ...................................... 20
Maintaining Data Integrity ............................................................................ 21
Caring for the Ultrasound System ................................................................... 22
Daily Checklist ............................................................................................. 22
Maintenance ................................................................................................ 23
Repair ................................................................................................... 23
Siemens Authorized Care..................................................................... 23
Cleaning and Disinfecting ............................................................................ 24
Cleaning and Disinfecting the System .................................................. 24
Cleaning an Air Filter ............................................................................ 27
Caring for the Battery Pack .......................................................................... 29
Battery Pack Location .......................................................................... 29
Battery Pack Replacement and Disposal ............................................. 30
Caring for Documentation and Storage Devices ........................................... 31
Caring for Transducers .................................................................................... 31
Cleaning and Disinfecting Transducers ....................................................... 32
IPX8 Immersion Levels......................................................................... 33
Approved List of Pre-Cleaners and Low-Level Disinfectants ................ 35
Approved List of Disinfectants .............................................................. 36
Storage ........................................................................................................ 37
Repair .......................................................................................................... 37
Protective Case ........................................................................................... 37
Instructions for Use 2 - 1
Page 30
2 Safety and Care
Caring for Transducer Accessories ................................................................ 38
Transducer Sheaths .................................................................................... 38
Storage ................................................................................................. 38
Gel Pad........................................................................................................ 39
Storage ................................................................................................. 39
Needle Guide Bracket Kits........................................................................... 39
Storage and Transportation .................................................................. 39
Cleaning, Disinfecting, and Sterilizing Transducer Accessories .................. 40
Needle Guide Bracket Kits ................................................................... 40
Storage ................................................................................................. 42
Environmental Protection ................................................................................ 43
Product Recycling and Disposal .................................................................. 43
Hazardous Substances ........................................................................ 43
Caring for Batteries ...................................................................................... 44
Recycling Batteries ...................................................................................... 45
Disposing of the Packaging Materials .......................................................... 46
Disposing of Components and Accessories ................................................ 46
Energy Conservation ................................................................................... 46
2 - 2 Instructions for Use
Page 31
2 Safety and Care
Symbol
Explanation
Operating Safety and Environment
Do not operate the ultrasound imaging system until you fully understand the safety considerations and procedures presented in this manual.
System Symbols
Refer to this table to identify important symbols located on the ultrasound imaging system and transducers.
Alternating Current
V~ AC (alternating current) voltage source
Caution: Risk of electric shock
Type BF Defibrillator-proof Patient Connection
Type BF Applied Part
Type CF Patient Connection
Consult Instructions for Use
Operating Instruction – Mandatory Action (blue illustration)
Caution (black illustration)
Note: For systems and transducers shipped from the factory prior to 1 October 2010, the symbol means "Caution, consult accompanying documents."
General Warning (yellow and black illustration)
Monitor Menu Control
Monitor Menu Control, up (increase)
Monitor Menu Control, down (decrease)
Monitor Picture menu icon
Monitor Function menu icon
Instructions for Use 2 - 3
Page 32
2 Safety and Care
Symbol
Explanation
B/W
Monitor OSD menu icon
Monitor menu Exit icon
Protective Earth Ground
Signal Earth Ground
Signal Input
Signal Output
Video Connection (monochrome video signals)
Start (of action for equipment)
USB Connection
Ethernet 10/100BaseT Connection
Equipotential Connection
Printer Connection
Black and White Printer Connection VGA
Digital Video Interface (DVI)
Intentional transmitter of non-ionizing radiation.
Electronic Array Transducer Port
Continuous Wave Doppler Transducer Port
2 - 4 Instructions for Use
Page 33
2 Safety and Care
Symbol
Explanation
IPX8
EEC directive(s)
Unlocked position
Locked position
Audio
Do not lean against the monitor.
Shelf Weight Restriction
ON only for MAINS control
OFF only for MAINS control
Power "On" standby switch
ECG Socket
IVK
Protected against the effects of continuous immersion in water Do not open. Refer servicing to qualified service personnel.
Danger: Risk of explosion if used in the presence of flammable anesthetics
Manufacturer's declaration of product compliance with applicable EEC directive(s) and the European Notified Body
Manufacturer's declaration of product compliance with applicable
UL symbol for listing as recognized components for Canada and United States of America
Gost-R symbol indicates that this product is certified for conformity to the safety requirements of Russian state standards
(Installierte Volumen Komponente) Identifier of selected system components or parts for product
traceability UL classified symbol for Canada and United States of America
Bar Code
Instructions for Use 2 - 5
Page 34
2 Safety and Care
Symbol
Explanation
Environmentally friendly use period
Manufacturer
Date of Manufacture symbol with the date below
Do not install wet
Location of air filter
Storage temperature range (example)
Indicates this side up
Do not stack
Shipping weight (example)
Do not allow to get wet
Fragile. Handle with care.
Refer to the operator's manual for information about compatible needle guides.
Do not dispose of by dumping in garbage. Use a separate collection for electrical and electronic equipment.
Products bearing this symbol are subject to the European Community directive 2002/96/EC on waste electrical and electronic equipment (WEEE), amended by directive 2003/108/EC. For collection and disposal of the product, its components, or its accessories, contact your local Siemens representative.
2 - 6 Instructions for Use
Page 35
2 Safety and Care
Symbol
Explanation
WATCH YOUR HAND
Fuse
Pinch hazard WARNING: PINCH POINT
Control panel swivel
Control panel height adjustment
External input/output converter box
Instructions for Use 2 - 7
Page 36
2 Safety and Care
Labels
Location of labels.
1 Monitor label 2 System warning and Certification label 3 Identification label
Example of monitor label.
2 - 8 Instructions for Use
Page 37
2 Safety and Care
Example of system warning label with certification labels.
Example of identification label.
1 Product name 2 Product status number 3 Serial number 4 Serial number barcode 5 Model number 6 Model number barcode 7 Manufacturing date
Instructions for Use 2 - 9
Page 38
2 Safety and Care
Biohazard Considerations
WARNING: With the exception of systems licensed to use the ACUSON AcuNav catheter, this
equipment is not suitable for intracardiac use or direct cardiac contact.
See also: For additional information on the AcuNav catheter, refer to the user manual for the catheter.
WARNING: Siemens makes every effort to manufacture safe and effective transducers. You
must take all necessary precautions to eliminate the possibility of exposing patients, operators, or third parties to hazardous or infectious materials. These precautions should be considered in the use of any application that may indicate the need for such care, and during endocavity or intraoperative scanning; during biopsy or puncture procedures; or when scanning patients with open wounds.
WARNING: To eliminate the possibility of exposing patients, operators, or third parties to
hazardous or infectious materials, always dispose of hazardous or infectious materials according to local, state, and regional regulations.
WARNING: There have been reports of severe allergic reactions to medical devices containing
latex (natural rubber). Health care professionals are advised to identify latex-sensitive patients and to be prepared to treat allergic reactions promptly. For additional information in the U.S.A., refer to FDA Medical Alert MDA91-1.
WARNING: Ultrasound energy is transmitted more effectively through water than through tissue.
When using a standoff device of any kind, for example, a gel pad, the actual mechanical and thermal indices, MI and/or TI, may be higher than indicated in the output display on the system.
The assessment of the biological effects of diagnostic ultrasound on humans is a subject of ongoing scientific research. This system, and all diagnostic ultrasound procedures, should be used for valid reasons, for the shortest possible period of time, and at the lowest mechanical and thermal indices necessary to produce clinically acceptable images.
According to the ALARA (As Low As Reasonably Achievable) principles, acoustic output should be set to the lowest level required to satisfactorily perform the examination.
The ultrasound imaging system complies with the standards of the American Institute of Ultrasound in Medicine (AIUM), the National Electrical Manufacturer's Association (NEMA), the guidelines of the United States Food and Drug Administration (FDA), and the guidelines of the International Electrotechnical Commission (IEC) in terms of safety and acoustic output levels. The ultrasound output levels are stated to permit the user to critically evaluate the system settings in the event of new research findings being announced.
2 - 10 Instructions for Use
Page 39
2 Safety and Care
Note on Fetal Examinations
The following recommendation is excerpted from the National Institute of Health in the United States of America. Consensus Statement on the Use of Ultrasound Imaging During Pregnancy, Volume 5, No. 1, based on the recommendation issued at the Health Consensus Development Conference, February, 1984:
Ultrasound examination in pregnancy should be performed for a specific medical indication. The data on clinical efficacy and safety do not allow a recommendation for routine scanning at this time.
Ultrasound examination performed solely to satisfy the family's desire to know the fetal sex, to view the fetus, or to obtain a picture of the fetus should be discouraged. In addition, visualization of the fetus solely for educational or commercial demonstrations without medical benefit should not be performed.
In August 1994, the Food and Drug Administration (FDA) notified the medical community and the ultrasound industry regarding its concerns about the misuse of diagnostic ultrasound equipment for non-medical purposes, and to discourage patients from having sonograms for non-medical reasons.
The American Institute of Ultrasound in Medicine (AIUM) has also advocated the responsible use of diagnostic ultrasound for all fetal imaging (August 2005).
Acoustic Output — Mechanical and Thermal Indices
WARNING: Ultrasound procedures should be used for valid reasons, for the shortest period of
time, and at the lowest mechanical/thermal index setting necessary to produce clinically acceptable images.
The ultrasound system incorporates an output display of Mechanical and Thermal Indices to allow you to monitor, and to limit, the amount of ultrasound energy that is transferred to the patient.
Note: For systems distributed in the United States of America, refer to the Medical Ultrasound Safety ultrasound education program brochure produced by the AIUM that is shipped with the ultrasound system.
See also: Acoustic Output Reference, Appendix F, Instructions for Use
Instructions for Use 2 - 11
Page 40
2 Safety and Care
Mechanical and Thermal Indices
The system displays the Mechanical Index during real time imaging, in all imaging modes. The system displays the Thermal Indices during real-time imaging when the value is equal to or exceeds 0.4.
Note: During exams using Contrast Agent Imaging (CAI), the system always displays values for the Mechanical Index (MI) and the Maximum of Mechanical Indices measured at the active focal zones (MIF).
Location of mechanical and thermal indices on the image screen.
Indices display in the abbreviated form shown below:
MI: Mechanical Index  MIF: Maximum of the Mechanical Indices measured at the active focal zones
(displayed during CAI exams only)
TIB: Bone Thermal Index (fetal application)  TIS: Soft Tissue Thermal Index  TIC: Cranial Thermal Index
2 - 12 Instructions for Use
Page 41
2 Safety and Care
Transmit Power Control
Adjust the transmit power and the corresponding acoustic pressure delivered through the transducer to the patient by using the designated control on the system. It is the main system function that determines the transmitted intensity of ultrasound for all transducers and imaging modes during real-time imaging, though it is not the only function that affects the mechanical and thermal indices. The range and especially the maximum level of the mechanical and thermal indices differ depending on the transducers. In addition, each diagnostic exam type has preset values for mechanical and thermal indices.
See also: Imaging Functions that Change Acoustic Output, p. 2-15
Note: Maximum transmit acoustic intensity and the mechanical index for each exam type are limited
in accordance with the United States Food and Drug Administration's (FDA) recommendations and guidelines. System default transmit intensity and mechanical index values are always below the FDA recommendations for each exam type. Although some exam types may default to a condition of maximum allowable transmit power, there are other system controls or functions that could raise acoustic output levels.
To increase the transmit power: During real-time imaging, press the toggle key for P upward to increase transmit power.
To decrease the transmit power:
During real-time imaging, press the toggle key for P downward to decrease transmit power.
Instructions for Use 2 - 13
Page 42
2 Safety and Care
Percentage (%)
Decibels (dB)
Percentage (%)
Decibels (dB)
50%
-6 dB
2.0%
-34 dB
16%
-16 dB
0.63%
-44 dB
6.3%
-24 dB
0.25%
-52 dB
Transmit Power Display
The transmit power range is from 0.20% to 100%. Selecting 100%, in combination with other system controls or functions, generates the maximum acoustic intensity and mechanical index for each transducer, where:
I
: 720 mW/cm2 and MI 1.9
SPTA.3
100% 0 dB 4.0% -28 dB 79% -2 dB 3.2% -30 dB 63% -4 dB 2.5% -32 dB
40% -8 dB 1.6% -36 dB 32% -10 dB 1.3% -38 dB 25% -12 dB 1.0% -40 dB 20% -14 dB 0.79% -42 dB
13% -18 dB 0.40% -46 dB 10% -20 dB 0.50% -48 dB
7.9% -22 dB 0.32% -50 dB
5.0% -26 dB 0.20% -54 dB
Comparison of decibels to percentages.
When the power level is changed, the system briefly highlights the power value to indicate the change.
Location of power display on the image screen.
2 - 14 Instructions for Use
Page 43
2 Safety and Care
Imaging Functions that Change Acoustic Output
WARNING: Observe the real-time display of mechanical and thermal indices (MI/TI) at all times.
In addition to the adjustment of the transmit power, adjustment of the following imaging functions and/or controls may affect the acoustic output:
Automatic Time-out  M-mode cursor  Color and Power ROI Position and Size; Steering Angle for linear array transducers  Doppler Gate Position and Size; Steering Angle for linear array transducers  Doppler PRF  Color PRF  Exam Type  Field of View (Scan Angle); 2D Steering Angle for linear array transducers  Focus  Zoom  Frame Rate  Freeze  Image Depth  Imaging Mode  Multi-Frequency  Power On/Off  System Presets and QuickSets  THI  Resolution/Speed  Simultaneous/Update  Transducer  Gel pad use
Instructions for Use 2 - 15
Page 44
2 Safety and Care
Transducer
TMM
Still Air
<30.7° C
<44.3° C
<29.8° C
<36.4° C
<29.5° C
<36.5° C
Transducer Surface Temperature Limits
The following table provides the maximum surface temperature of the transducers compatible with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Note: The systemic uncertainty of the transducer surface temperature is estimated as 1.41%.
Maximum Temperature
4C1 CH5-2 6C2 C6-2 C8-5 VF10-5 VF12-4 VF16-5 C6F2 C8F3 EV9F3 EC9-4w MC9-4 4V1c P8-4 V5Ms AcuNav 8F AcuNav 10F SoundStar 10F SoundStar eco 8F SoundStar eco 10F CW2 CW5
TMM = Tissue Mimicking Material
<32.1° C <41.6° C <31.2° C <42.7° C <32.7° C <43.5° C <31.4° C <42.1° C <28.9° C <38.8° C <33.2° C <44.5° C <32.0° C <44.8° C <28.8° C <41.7° C <28.8° C <40.2° C <26.4° C <39.4° C <28.3° C <43.2° C <30.7° C <44.3° C
<32.8° C <45.5° C <29.4° C <39.3° C <30.3° C <43.2° C <29.5° C <36.5° C
<29.8° C <36.4° C
<29.8° C <36.4° C <30.6° C <40.2° C <29.4° C <38.1° C
2 - 16 Instructions for Use
Page 45
2 Safety and Care
Electrical Safety
WARNING: To avoid electric shock, use a protective earth connection to connect the ultrasound
system to the mains power supply. The protective earth connection ensures that the mains circuit breaker will disconnect the power supply in the event of a short circuit.
WARNING: For systems used in the U.S.A.: To ensure grounding reliability, only connect the
system to a hospital-grade power outlet.
WARNING: The AC power connector plug for the ultrasound system is a three-prong grounded
plug (in the U.S.A.) and should never be adapted to any two-prong (non-grounded) outlet, either by modifying the plug or by using an adapter. In the U.S.A., proper grounding requires the AC power connector plug to be plugged into a hospital-grade power outlet.
WARNING: To avoid electrical shock, never modify the ultrasound system's AC power
connector plug, as doing so may overload your facility's power circuits. To ensure grounding reliability, connect the system only to an equivalent outlet.
WARNING: To avoid electrical shock, never use equipment or a MAINS power cord that shows
signs of wear or tampering, or whose ground plug has been bypassed using an adapter.
WARNING: Equipment connected to the ultrasound system and in the patient zone must be
powered from a medically-isolated power source or must be a medically-isolated device. Equipment powered from a non-isolated source can result in chassis leakage currents exceeding safe levels. Chassis leakage current created by an accessory or device connected to a non-isolated outlet may add to the chassis leakage current of the ultrasound system.
WARNING: Using an extension cord or multi-socket outlet setup to provide power to the imaging
system, or to the system's peripheral devices may compromise the system grounding and cause your system to exceed leakage current limits.
WARNING: To avoid electrical shock and damage to the ultrasound system, power off and
unplug the equipment from the AC power outlet before cleaning and disinfecting.
WARNING: To avoid electrical shock and damage to the control panel resulting from ingress of
liquid, place the gel and gel warmer on the side of the system closest to the patient.
WARNING: Do not pour any fluid onto the ultrasound system surfaces, as fluid seepage into the
electrical circuitry may cause excessive leakage current or system failure.
WARNING: To ensure proper grounding and leakage current levels, it is the policy of Siemens
to have an authorized Siemens representative or Siemens-approved third party perform all on-board connections of documentation and storage devices to the ultrasound system.
Instructions for Use 2 - 17
Page 46
2 Safety and Care
WARNING: To prevent excessive leakage current from contacting the patient, do not touch a
user-accessible connector on the system while touching or scanning the patient. User­accessible connectors include the ECG connector, a USB connector, and any other audio, video, or data transmission connectors.
WARNING: Connecting peripheral devices to accessory outlets on the ultrasound system
effectively creates a medical electrical system, resulting in a reduced level of safety.
WARNING: Do not modify this equipment without authorization from Siemens.
Caution: To maintain the safety and functionality of the ultrasound system, maintenance must
be performed every 24 months. Electrical safety tests must also be performed at regular intervals as specified by local safety regulations, or as needed.
Caution: To avoid the possibility of static shock and damage to the system, avoid the use of
aerosol spray cleaners on the monitor screens.
Caution: Do not use spray cleaners on the ultrasound system, as this may force cleaning fluid
into the system and damage electronic components. It is also possible for the solvent fumes to build up and form flammable gases or damage internal components.
Caution: To reduce the risk of fire and subsequent equipment damage, use only 26 gauge
(0.14 mm
2
) or heavier wire for the cable connecting to the Ethernet port located on the
ultrasound system.
EMC Note: Operating the ultrasound system in close proximity to sources of strong electromagnetic fields, such as radio transmitter stations or similar installations, may lead to interference visible on the monitor screen. However, the device has been designed and tested to withstand such interference and will not be permanently damaged.
2 - 18 Instructions for Use
Page 47
2 Safety and Care
Level of Protection Against Electrical Shock — Transducers
WARNING: Only use Type BF transducers with the ultrasound system to maintain a level of
protection against electrical shock.
According to EN 60601-1 and IEC 60601-1, the assemblies for the linear, curved, and phased array transducers provide a "Level of Protection Against Electrical Shock" of "Type BF."
The Type BF icon is located on the transducer label.
Defibrillators
WARNING: The ECG function is designed to withstand the effects of defibrillation. However,
when possible, disconnect the ECG leads during defibrillation since a malfunction of the safety controls could otherwise result in electrical burns for the patient.
For patient safety, be sure to use defibrillators that do not have grounded patient circuits.
Defibrillator Use and the ACUSON AcuNav Ultrasound Catheter
WARNING: The catheter is designed to withstand the effects of defibrillation. However, when
possible, disconnect the connector from the ultrasound system during defibrillation because a malfunction of the safety controls could otherwise result in electrical burns for the patient.
The catheter is designed to withstand the effects of defibrillation. There are no exposed conductive surfaces distal to the handle. Within the flexible shaft, a chassis ground shield covers all active circuits and conductors.
Recovery Time After Defibrillation During a Catheter Procedure
If you disconnect the SwiftLink catheter connector from the ultrasound system before defibrillation, the recovery time after defibrillation is equal to the time required to reconnect the connector to the ultrasound system.
The recovery time after defibrillation for the ultrasound system: Three (3) minutes if defibrillation is performed with the ultrasound system on.
Implantable Devices
WARNING: Ultrasound systems, like other medical equipment, use high-frequency electrical
signals that can interfere with implantable devices such as pacemakers and implantable cardioverter-defibrillators (ICDs). If the patient has such an implantable device, you should be aware of any interference in its operation and immediately power off the ultrasound system.
Instructions for Use 2 - 19
Page 48
2 Safety and Care
Possible Combinations with Other Equipment
WARNING: Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective EN and IEC standards (for example, EN 60950 and IEC 60950 for data processing equipment and EN 60601-1 and IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standards EN 60601-1-1 and IEC 60601-1-1. Anyone who connects additional equipment to any of the signal input or signal output ports configures a medical system and is therefore responsible that the system complies with the requirements of the system standards EN 60601-1-1 and IEC 60601-1-1. Siemens can only guarantee the performance and safety of the devices listed in the Instructions for Use. If in doubt, consult the Siemens service department or your local Siemens representative.
The ultrasound system can support two on-board documentation devices at one time.
On-board peripheral devices must be installed by an authorized Siemens representative or by a Siemens-approved third party. Devices installed by other people will be at the user's risk and may void the system warranty.
Barcode Reader
WARNING: To avoid electrical shock, do not connect the barcode reader to the system when
more than two on-board documentation devices are connected to the system.
WARNING: Do not stare into the beam of the barcode reader. This action can create a risk for
injury to the user.
You can connect a Siemens-approved barcode reader directly to a USB port on the ultrasound system. The barcode reader provides direct entry of patient information in the registration form, for example, patient name, patient identifier, or physician identifier.
2 - 20 Instructions for Use
Page 49
2 Safety and Care
Maintaining Data Integrity
Important Information
To ensure data integrity:
To prevent the loss of data that results from power failures and other system "down"
occurrences, you must archive important data, such as patient records, onto an external recording media, such as a CD or a network.
Loss of data is to be expected and its retrieval is not normally possible under the following
conditions: loss of power to the ultrasound system, hard disk failure, CPU failure, system lockup, and other similar causes.
Should an abnormal system shutdown occur, retrieval of data not saved to the hard disk or
not archived to an external recording media is not normally possible. An abnormal system shutdown occurs if you do not power off the ultrasound system using
the partial power on/off switch ( ) located on the control panel. Other examples of abnormal system shutdown include: equipment malfunction, loss of power, or pressing and holding the partial power on/off switch longer than 4 seconds.
Should an abnormal system shutdown occur, the system may initially require additional
time to reboot or to respond to user input. This is due to the operating system performing a background scan of the hard disk to detect and segregate any truncated or corrupted files.
Instructions for Use 2 - 21
Page 50
2 Safety and Care
Caring for the Ultrasound System
It is the responsibility of the user to verify that the ultrasound system is safe for diagnostic operation on a daily basis. Each day, prior to using the system, perform each of the steps in the Daily Checklist.
All exterior parts of the system, including the control panel, keyboard, and transducers, should be cleaned and/or disinfected as necessary or between uses. Clean each component to remove any surface particles. Disinfect components to kill vegetative organisms and viruses.
The air filter on the ultrasound system must be cleaned regularly to maintain proper system cooling. Remove and check the air filter weekly, and clean as needed.
Daily Checklist
WARNING: To minimize the risk of cross-contamination and infectious diseases, a sterile,
non-pyrogenic transducer sheath must be in place during procedures requiring sterility.
WARNING: To avoid electrical shock, you must visually inspect a transducer prior to use. Do not
use a transducer that has a cracked, punctured, or discolored casing or a frayed cable.
Discoloration Exception: The use of the approved disinfectants, Cidex OPA and Gigasept FF, may cause discoloration of transducer housings, including the face of the transducer. You can continue to use a transducer if it is discolored due to the use of these specific disinfectants only.
See also: Approved List of Disinfectants, p. 2-36
Perform the following each day before using the ultrasound system:
Visually inspect all transducers. Do not use a transducer which has a cracked, punctured,
or discolored casing or a frayed cable.
Visually inspect all power cords. Do not turn on the power if a cord is frayed or split, or
shows signs of wear. If your system's power cord is frayed or split, or shows signs of wear, contact your Siemens
service representative for power cord replacement.
Visually inspect the ECG connector and the cable. Do not use the ECG function if the
connector or cable is damaged or broken.
Verify that the trackball, DGC slide controls, and other controls on the control panel are
clean and free from gel or other contaminants.
Once the system is powered on:
Visually check the on-screen displays and lighting.  Verify that the monitor displays the current date and time.  Verify that the transducer identification and indicated frequency are correct for the active
transducer.
2 - 22 Instructions for Use
Page 51
2 Safety and Care
Maintenance
Caution: To maintain the safety and functionality of the ultrasound system, maintenance must
be performed every 24 months. Electrical safety tests must also be performed at regular intervals as specified by local safety regulations, or as needed.
Repair
WARNING: Do not modify this equipment without authorization from Siemens.
For questions regarding repair or replacement of any equipment parts on your system, contact your Siemens service representative.
Siemens Authorized Care
Installers and operators must observe any statutory regulations that govern the installation, operation, inspection, and maintenance of this equipment.
To ensure the safety of patients, operators, and third parties, the equipment must be inspected every 24 months, and the replacement of parts is performed as necessary. This maintenance must be performed by a qualified Siemens authorized representative. It is important to inspect the equipment more frequently if it is operated under extraordinary conditions.
Perform inspections and maintenance at the prescribed intervals to avoid worn and hazardous parts due to wear. Contact the Siemens service department for information regarding the required maintenance. As manufacturers and installers of ultrasound equipment, Siemens cannot assume responsibility for the safety properties, reliability, and/or performance of the equipment, if:
Installations, extensions, readjustments, modifications, additions, or repairs are carried out
by persons not specifically authorized by Siemens.
Components that affect the safe operation of the system are replaced by parts not
authorized by Siemens.
The electrical installation of the room where the equipment is located does not meet the
power and environmental requirements stated in this manual.
The equipment is not used in accordance with the operating instructions.  The system is operated by personnel not adequately educated or trained.
Siemens suggests that you request any person who performs maintenance, or repairs, to provide you with a certificate showing:
The nature and extent of the work performed  Changes in rated performance  Changes in working ranges  Date of service  Name of person or firm performing the service  Signature of person performing the service
Technical documentation pertinent to the ultrasound system is available at an additional charge. However, this does not in any way constitute an authorization to conduct repairs or maintenance. Siemens refuses all responsibility whatsoever for repairs that are performed without the express written consent of the Siemens service department.
Instructions for Use 2 - 23
Page 52
2 Safety and Care
Cleaning and Disinfecting
You must take all necessary precautions to eliminate the possibility of exposing patients, operators, or third parties to hazardous or infectious materials. Use universal precautions when cleaning and disinfecting. You should treat all portions of the imaging system that come in contact with human blood or other body fluids as if they were known to be infectious.
All exterior parts of the system, including the control panel, and transducers, should be cleaned and/or disinfected as necessary or between uses. Clean each component to remove any surface particles. Disinfect the components to kill vegetative organisms and viruses.
Cleaning and Disinfecting the System
WARNING: To avoid electric shock and damage to the system, always power off and
disconnect the equipment from the AC power source before cleaning and disinfecting.
WARNING: Contents of some disinfecting agents are known to be health hazards. Their
concentration in the air must not exceed an applicable specified limit. Comply with the manufacturer's instructions when using these agents.
WARNING: Disinfectants and cleaning methods listed are recommended by Siemens for
compatibility with product materials, not for biological effectiveness. Refer to disinfectant label instructions for guidance on disinfection efficacy and appropriate clinical uses.
WARNING: The use of any disinfectants other than those specified in the instructions for use
may damage the ultrasound system and accessory surfaces and, as a result, may create electrical hazards for the patients and/or users.
Caution: To avoid the possibility of static shock and damage to the ultrasound system, avoid the
use of aerosol spray cleaners on the monitor screens.
Caution: Do not clean the system with chlorinated or aromatic solvents, acidic or basic
solutions, isopropyl alcohol or strong cleaners such as ammoniated products, as these can damage the surface of the system. Isopropyl alcohol exception: It is acceptable to use isopropyl alcohol only when recommended for cleaning the trackball assembly. Use the recommended cleaning procedure.
See also: Trackball Cleaning, p. 2-25
Caution: Do not use spray cleaners on the ultrasound system, or pour fluid onto the system
surfaces, as fluid can seep into the system and damage electronic components. It is also possible for the solvent fumes to build up and form flammable gases or damage internal components.
Caution: Do not pour any fluid onto the ultrasound system surfaces, as fluid seepage into the
electrical circuitry may cause excessive leakage current or system failure.
2 - 24 Instructions for Use
Page 53
2 Safety and Care
System Surfaces
The following instructions describe cleaning the surface of the ultrasound system, including the trackball and transducer holder.
After cleaning system surfaces, including the trackball and transducer holder, you can disinfect the surface with an approved disinfectant wipe.
To clean the surface of the ultrasound system:
1. Power off ( ) the ultrasound system and unplug the power cord from the power outlet.
2. Use a clean gauze pad or lint-free cloth, lightly moistened with a mild detergent, to wipe the surface of the ultrasound system.
Take particular care to clean the areas near the trackball and the slide controls. Ensure these areas are free of gel and any other visible residue.
Ensure that cleaning solution does not seep into the control panel, keyboard, or any other openings.
3. After cleaning, use a clean, lint-free cloth to dry the surface.
4. After cleaning, reconnect the ultrasound system power cord into the power outlet.
To clean the liners of the transducer holders:
1. Remove the liner from the transducer holder.
2. Clean the liner under running water, using a mild detergent and dry with a lint-free cloth.
3. Reinsert the liner into the transducer holder.
See also: System Setup, Chapter 3, Instructions for Use
To clean the trackball:
Caution: Do not drop or place foreign objects inside the trackball assembly. This may affect the
trackball's operation and damage the system.
Caution: Do not submerge the front bezel and Teflon seal in isopropyl alcohol - this may
damage the adhesive bond between the two components.
1. Remove the trackball front panel bezel by rotating the bezel counterclockwise.
2. Remove the ball.
3. Clean the ball with tissue and isopropyl alcohol.
4. Clean the Teflon seal (located in the front panel bezel) with a tissue and isopropyl alcohol.
5. Clean the inside of the trackball assembly with a cotton swab and isopropyl alcohol.
6. Allow the assembly parts to completely dry before reassembly.
7. Replace the ball and front panel bezel.
Instructions for Use 2 - 25
Page 54
2 Safety and Care
Ultrasound System
Transducer Liners
Trackball Assembly
Approved Disinfectant Wipes for the Ultrasound System Surfaces
The following matrix provides a list of approved disinfectant wipes for use on the ultrasound system and surfaces of the listed components.
Sani-Cloth
AF
   
*or any bleach wipe with <1% sodium hypochlorite and no other active ingredients = compatible NC = not compatible NA = not applicable (not tested)
Sani-Cloth
AF3
Sani-Cloth
Bleach Wipe*
Sani-Cloth
HB
Sani-Cloth
Plus
Super
Sani-Cloth
2 - 26 Instructions for Use
Page 55
2 Safety and Care
Cleaning an Air Filter
The air filter on the ultrasound system must be cleaned regularly to maintain proper system cooling. Remove and check the air filter weekly, and clean as needed.
The filter location is marked with the air filter symbol.
Location of air filter.
Instructions for Use 2 - 27
Page 56
2 Safety and Care
To remove and clean the air filter:
Caution: Do not scrub, stretch, or bend the filter, or apply heat to the filter, as doing so could
damage the filter.
1. Power off and unplug the power cord from the power outlet.
2. Grasp the air filter tray and pull it out of the system.
1 Air filter 2 Air filter tray
Removing the air filter tray.
3. Rinse the air filter tray with running water and allow the filter to completely dry. To hasten drying, you may gently shake the air filter tray, or blot the filter with a clean,
lint-free cloth.
Caution: Do not insert the air filter tray into the ultrasound system with a wet filter as this can
damage the system.
4. Slide the air filter tray back into the ultrasound system.
5. Plug the power cord into the power outlet.
2 - 28 Instructions for Use
Page 57
2 Safety and Care
Caring for the Battery Pack
WARNING: Do not strike or drop the battery pack because this can cause heat generation,
bursting, or fire. Compromising the structural integrity of the battery pack can result in leakage or explosion and the potential for personal injury.
WARNING: Do not use the battery pack if it leaks fluid or has changed shape. If skin or clothing
comes in contact with fluid from the battery pack, thoroughly wash the area immediately with clean water. If any fluid comes in contact with a user's eyes, immediately flush their eyes with water and seek medical attention.
WARNING: Do not allow the battery pack to contact water. Compromising the structural integrity
of the battery pack can result in leakage or explosion and the potential for personal injury.
WARNING: Do not disassemble the battery pack. Compromising the structural integrity of the
battery pack can result in leakage or explosion and the potential for personal injury.
The battery pack is designed to maintain system memory for a maximum of twenty minutes.
The length of the charge time is three hours.
Battery Pack Location
The battery pack is located on the front side of the ultrasound system, where the power is located. The battery pack cover must be removed to access the battery pack.
Battery pack location.
1 Battery pack location
Instructions for Use 2 - 29
Page 58
2 Safety and Care
+ +
+ +
2
1
Battery Pack Replacement and Disposal
Replace the battery pack when it is no longer able to hold a charge. When the system is used with the Mobile QuikStart option and frequently disconnected
from the AC power (for a duration of greater than 20 minutes), replace the battery every 6 months.
When the system is used with the Mobile QuikStart option and only occasionally
disconnected from the AC power (for a duration of approximately 20 minutes), replace the battery every year.
1 AC OK indicator LED 2 Battery pack charge indicator LED
Example of an AC Tray panel with the battery pack charge indicator LED.
The battery pack charge indicator LED is located at the rear of the system, on the AC Tray panel. A green blinking LED indicates that the battery is actively charging. A solid green LED indicates that the battery pack is fully charged. If the LED is not illuminated, then there is a problem with the battery pack, or the battery pack may be missing.
2 - 30 Instructions for Use
Page 59
2 Safety and Care
Caring for Documentation and Storage Devices
Caution: Computer viruses on a USB-compatible device can infect the ultrasound system.
Check for viruses before connecting the device to the system.
Note: Study data stored directly to a USB device can be lost. Do not use a USB device for permanent data storage.
For information on the care of an optional documentation or storage device, please refer to the manufacturer's operating instructions that accompanied the device.
Caring for Transducers
WARNING: To minimize the risk of cross-contamination and infectious diseases, endocavity and
intraoperative transducers must be cleaned and high-level disinfected after each use.
WARNING: Prior to each use, inspect the endocavity or intraoperative transducer for signs of
mechanical damage such as cracks, cuts, tears, perforations, or protrusions. Do not use the transducer if the transducer appears damaged in any way. Any damage could cut the patient and compromise the electrical safety of the transducer, causing possible patient or user injury. Contact your local Siemens representative.
WARNING: During neurosurgical procedures, if a transducer becomes contaminated with tissue
or fluids of a patient known to have Creutzfeld-Jacob disease, the transducer should be destroyed, as it cannot be sterilized.
WARNING: When using an endocavity or intraoperative transducer with a CF type applied part,
the patient leakage currents may be additive.
Caution: Transducers are sensitive instruments – irreparable damage may occur if they are
dropped, knocked against other objects, cut, or punctured. Do not attempt to repair or alter any part of a transducer.
Caution: To avoid cable damage, do not roll the system over transducer cables.
Caution: To avoid damage to the transducer, do not use transducer sheaths containing an
oil-based coating or petroleum- or mineral oil-based ultrasound coupling agents. Use only a water-based ultrasound coupling agent.
Caution: Follow all instructions provided by manufacturers of sterile goods (transducer sheaths)
to ensure proper handling, storage, and cycling of all sterile goods.
Take extreme care when handling or storing transducers. They must not be dropped, jarred, or knocked against other objects. Do not allow transducers to come into contact with any sharp-edged or pointed object.
Instructions for Use 2 - 31
Page 60
2 Safety and Care
Cleaning and Disinfecting Transducers
WARNING: To avoid electrical shock and damage to the system, disconnect the transducer
prior to cleaning or disinfecting.
WARNING: Disinfectants and cleaning methods listed are recommended by Siemens for
compatibility with product materials, not for biological effectiveness. Refer to disinfectant label instructions for guidance on disinfection efficacy and appropriate clinical uses.
Caution: Do not sterilize transducers using hot steam, cold gas, or Ethylene Oxide (EO)
methods. Before applying any other methods which might be recommended by manufacturers of sterilization equipment, please contact your Siemens representative.
Caution: To avoid damage to the transducer, observe the immersion levels indicated for each
transducer type. Do not immerse or allow the cable or connector of a transducer to become wet.
Caution: The transducers have been designed and tested to be able to withstand high-level
disinfection as recommended by the manufacturer of the disinfectant product. Carefully follow the disinfectant manufacturer's instructions. Do not immerse for more than one hour.
Caution: Do not use abrasive cleaning agents, organic solvents such as benzene, isopropyl
alcohol, or phenol-based substances, or cleaning agents containing organic solvents to clean or disinfect transducers. These substances can damage the transducers.
Caution: Do not use a spray cleaner on a transducer, as this may force cleaning fluid inside the
housing and damage the transducer.
All transducers should be cleaned and disinfected prior to their use on each patient. Endocavity and intraoperative transducers require high-level disinfection prior to use.
See also: Transesophageal Transducer, Chapter 6, Instructions for Use
2 - 32 Instructions for Use
Page 61
2 Safety and Care
To clean a transducer:
1. Disconnect the transducer from the system.
2. Moisten a clean gauze pad with water and wipe the transducer to remove any gel or particles remaining on the transducer. If water is not effective, then you can use an approved pre-cleaner or low-level disinfectant.
3. Carefully wipe the entire transducer, including the cable and connector.
4. After cleaning or disinfecting, use a clean cloth to dry the transducer.
To disinfect or high-level disinfect a transducer:
1. Disconnect the transducer from the system.
2. Thoroughly clean, rinse, and dry the transducer.
3. Take care to keep the cable strain relief and connector of the transducer dry while immersing the transducer in an approved disinfectant to the level indicated in the following illustration.
4. Carefully follow the disinfectant manufacturer's instructions for disinfection or high-level disinfection.
5. After disinfecting or high-level disinfecting, use a clean cloth to dry the transducer.
IPX8 Immersion Levels
Caution: To avoid damage to the transducer, observe the immersion levels indicated for each
transducer type. Transducers with the protection level IPX8 are indicated by the presence of the "IPX8" symbol on the connector of the transducer.
Transducers meet Ingress Protection level IPX8 of EN 60539 and IEC 60539 to the depth of the immersion line shown in the illustration only for transducers with the "IPX8" symbol on the connector of the transducer.
Example of transducer labels with IPX8 symbol.
Instructions for Use 2 - 33
Page 62
2 Safety and Care
IPX8 Immersion Levels.
1 Endocavity 2 Linear 3 Curved 4 Phased 5 fourSight 4D 6 fourSight 4D (endocavity) 7 Continuous Wave (CW) 8 V5Ms
Caution: Do not immerse the label located on the cable of the CW transducer.
Note: Intraoperative transducers are immersible up to the strain relief on the connector.
2 - 34 Instructions for Use
Page 63
2 Safety and Care
4C1
CH5-2
    
C8-5
NA
VF10-5
VF12-4
C8F3
EV9F3
4V1c
P8-4
V5Ms
CW5
Approved List of Pre-Cleaners and Low-Level Disinfectants
The following matrix provides a list of approved pre-cleaners and low-level disinfectants for all transducers.
Enzol Transeptic Hi-Tor Plus Theracide Plus T-Spray II Super SaniCloth
 
6C2
C6-2
VF16-5 C6F2 NA NA NA NA
EC9-4w
MC9-4
CW2
= compatible NC = not compatible NA = not applicable (not tested)
 
NA
 
 
NC NA NA
NA NA
 
NA
   
NA NA NA NA NA
NA NA NA
 
NA
     
NA NA
   
NC
NA NA
Instructions for Use 2 - 35
Page 64
2 Safety and Care
Cidex
Cidex Plus
Cidex OPA
Milton
Virkon
Gigasept FF
Resert XL HLD
4C1
CH5-2
C8-5
NA
VF10-5
VF12-4
C6F2    NA
NA
NA
NA
C8F3
EV9F3
4V1c
P8-4
V5Ms
CW5      
NA
Approved List of Disinfectants
The following matrix provides a list of approved disinfectants for all transducers.
Note: The approved disinfectants, Cidex OPA and Gigasept FF, may discolor transducer housings, including the face of the transducer. There is no associated degradation of imaging performance or transducer reliability.
 
6C2 C6-2 NA
   
VF16-5 NA NA
NA NA NA NA NA
NA NA NC
NA NA NA NA
NA NA
EC9-4w
NA
MC9-4 NA
   
NA
CW2 NA
= compatible
NC = not compatible NA = not applicable (not tested)
NA
NA
2 - 36 Instructions for Use
Page 65
2 Safety and Care
Storage
Store transducers in a clean and dry environment. Extreme temperatures or humidity may damage a transducer.
See also: Technical Description, Appendix A, Instructions for Use
Repair
Do not attempt to repair or alter any part of the transducer. Contact your service representative at Siemens immediately if a transducer appears to be damaged or malfunctions in any way.
Protective Case
Due to the mechanical sensitivity of transducers, Siemens recommends that you always use the transducer case when you ship a transducer or transport it from one place of examination to another. The case is specially designed to protect the sensitive parts of the transducer. Be sure that all parts of the transducer are properly placed inside the case before you close the lid.
Instructions for Use 2 - 37
Page 66
2 Safety and Care
Caring for Transducer Accessories
Instructions are provided for the following accessories for transducers:
Transducer Sheaths  Gel Pads  Needle Guide Bracket Kits
See also: Transducer Accessories and Biopsy, Chapter 5, Instructions for Use
Transducer Sheaths
WARNING: There have been reports of severe allergic reactions to medical devices containing
latex (natural rubber). Health care professionals are advised to identify latex-sensitive patients and to be prepared to treat allergic reactions promptly. For additional information in the U.S.A., refer to FDA Medical Alert MDA91-1.
WARNING: To minimize the risk of cross-contamination and infectious diseases, a sterile,
non-pyrogenic transducer sheath must be in place during procedures requiring sterility.
WARNING: Only a sterile transducer sheath provides the sterile barrier required for surgical
procedures. To ensure sterility of a procedure, always place a sterile sheath on a transducer, as transducers cannot be sterilized using hot steam, cold gas, or Ethylene Oxide (EO) methods.
Caution: Siemens recommends that you follow all instructions provided by manufacturers of
sterile goods (transducer sheaths) to ensure proper handling, storage, and cycling of all sterile goods.
Transducer sheaths are single-use items used to ensure proper acoustic coupling and provide a prophylactic barrier for the intended ultrasound application. Sheaths are available for all transducers. Siemens recommends the use of market-cleared transducer sheaths.
Using a disposable latex transducer sheath on a transducer reduces the possibility of cross-contamination. Always use a protective transducer sheath for endocavity exams, and when scanning an open wound or an area where the skin is not intact.
Storage
WARNING: Before use, examine sterile goods, such as sheaths, for any material flaws. Some
packaging may list an expiration date. Any product showing flaws, or whose expiration date has passed, should not be used.
Caution: Do not store transducer sheaths in direct sunlight, as ultraviolet damage can result.
Latex products have a limited shelf life, and should be stored in a cool, dry, dark place with an ambient temperature between -5°C and +40°C and up to 80% relative humidity at +40°C.
2 - 38 Instructions for Use
Page 67
2 Safety and Care
Gel Pad
Before use, examine the gel pad for any material flaws. Thinning, bulging, or brittleness of the material indicates damage. Any product showing flaws should not be used.
See also: Transducer Accessories and Biopsy, Chapter 5, Instructions for Use
Storage
Do not store gel pads below 5°C nor above 57°C. Gel pads have a limited shelf life. Before use, examine these products for any material flaws. Some packaging may list an expiration date. Any product showing flaws, or whose expiration date has passed, should not be used.
Needle Guide Bracket Kits
WARNING: If a needle guide becomes contaminated with tissue or fluids of a patient known to
have Creutzfeld-Jacob disease, then the needle guide should be destroyed. Sterilization is not effective against Creutzfeld-Jacob contamination.
Needle guide bracket kits are available for biopsy and puncture procedures for specific transducers.
See also: Transducer Accessories and Biopsy, Chapter 5, Instructions for Use
Storage and Transportation
Always clean and sterilize or high-level disinfect components used in a needle puncture or biopsy procedure after each use.
Refer to the in-box instructions for storage and transportation information.
Instructions for Use 2 - 39
Page 68
2 Safety and Care
Cleaning, Disinfecting, and Sterilizing Transducer Accessories
WARNING: Ensure the accessories for a transducer are properly cleaned, sterilized, or
disinfected as appropriate before each use to avoid possible patient contamination.
Needle Guide Bracket Kits
Needle guide bracket kits are available for specific transducers. Instructions follow for the cleaning, disinfecting, and sterilization of each kit. Bracket assemblies should be cleaned and sterilized or high-level disinfected after each use.
See also: For a list of transducers compatible with the needle guide accessories, refer to Chapter 5 in this manual.
CH4-1, SG-3, SG-5, and Infiniti Plus Needle Guide Bracket Kits
Disposable Needle Guide
WARNING: The needle guide is packaged sterile and is a single-use item. Do not use if the
packaging indicates signs of tampering or if the expiration date has passed.
The needle guide for use with the CH4-1, SG-3, SG-5, and Infiniti Plus bracket kit is a single­use item. Refer to the in-box instructions for disposal instructions.
Reusable Bracket
WARNING: Needle Guide Bracket kits are packaged non-sterile. High-level disinfect these
products prior to their first use.
EV9F3 Reusable Needle Guide Bracket Kit (Stainless Steel)
WARNING: Needle Guide Bracket kits are packaged non-sterile. Sterilize these products prior to
their first use.
The stainless steel endocavity needle guide is a reusable item. Refer to the in-box instructions for attachment and care procedures, including cleaning and sterilization.
2 - 40 Instructions for Use
Page 69
2 Safety and Care
EC9-4w Disposable Needle Guide Bracket Kit
WARNING: The needle guide is packaged sterile and is a single-use item. Do not use if the
packaging indicates signs of tampering or if the expiration date has passed.
The disposable endocavity needle guide is a single-use item. Refer to the in-box instructions for disposal instructions.
EC9-4w Reusable Needle Guide Bracket Kit (Stainless Steel)
WARNING: The needle guide is packaged non-sterile. Sterilize this product prior to its first use.
The stainless steel endocavity needle guide is a reusable item. Refer to the in-box instructions for attachment and care procedures, including cleaning and sterilization.
VF12-4 Reusable Needle Guide Bracket Kit
WARNING: The needle guide is packaged non-sterile. Sterilize this product prior to its first use.
Refer to the in-box instructions for attachment and care procedures, including cleaning and sterilization.
Instructions for Use 2 - 41
Page 70
2 Safety and Care
Universal Reusable Needle Guide Bracket Kit S (Stainless Steel)
WARNING: Needle Guide Bracket kits are packaged non-sterile. Sterilize these products prior to
their first use. Refer to the sterilization procedures for the Reusable Needle Guide Bracket Kit S (Stainless Steel) in the following pages.
Prior to sterilization, clean the Universal Reusable Needle Guide Bracket and insert(s) using an enzymatic cleaner.
To clean:
1. Rinse the bracket and insert(s) with water.
2. Soak the bracket and insert(s) in an enzymatic cleaner. Carefully follow the manufacturer's instructions.
3. Rinse the bracket and insert(s) with water to remove any debris and remaining cleaner.
To sterilize:
Sterilize the bracket and needle guide insert(s) using a wrapped, gravity-displacement
steam sterilization at a temperature of 121°C to 123°C for an exposure time of 15 to 30 minutes.
Storage
Always clean and sterilize components used in a needle puncture or biopsy procedure after each use.
2 - 42 Instructions for Use
Page 71
2 Safety and Care
Environmental Protection
Product Recycling and Disposal
Dispose of this product according to local, state, and regional regulations.
Batteries and electrical and electronic equipment can contain hazardous substances. If released, the hazardous substances can harm people and the environment.
Siemens provides disassembly instructions to treatment facilities for the safe and proper removal and recycling of electrical and electronic components in this product. For more information, contact your local Siemens representative.
To the extent required by local laws and regulations, Siemens has programs for the return of used products. For more information, contact your local Siemens representative.
Hazardous Substances
WARNING: This product consists of devices that may contain mercury, which must be recycled
or disposed of in accordance with local, state, or federal laws.
Hazardous Substance Included in the Product? Location
Mercury Yes
A small amount of mercury (maximum
5.0 mg/lamp) is included in the backlight of the flat panel monitor.
Instructions for Use 2 - 43
Page 72
2 Safety and Care
Caring for Batteries
WARNING: Do not strike or drop batteries, allow batteries to contact water or other fluids,
disassemble batteries, allow conductive object to contact a battery's terminals, cause a battery to become short circuited, heat batteries, or expose batteries to fire. Any of these actions can compromise the structural integrity of a battery. Compromising the structural integrity of a battery can result in battery leakage, heat generation, fire, or explosion, causing possible personal injury.
WARNING: Do not use a battery if it leaks fluid or has changed shape. If skin or clothing comes
in contact with fluid from the battery, thoroughly wash the area immediately with clean water. If any fluid comes in contact with a user's eyes, immediately flush their eyes with water and seek medical attention.
WARNING: Replace batteries with the same or equivalent type. Use of incompatible batteries
can result in battery leakage, heat generation, fire, or explosion, causing possible personal injury.
WARNING: Do not attempt to recharge non-rechargeable batteries, such as the batteries
included on printed circuit boards. Recycle non-rechargeable batteries according to local, state, and regional regulations.
For maximum battery life, observe all of the following steps when the system is not in use:
Keep the system plugged into the power outlet.  Ensure the mains circuit breaker is in the on position (I = ON). Store the system in an environment with low humidity and a temperature range between
0°C and 21°C, where possible.
2 - 44 Instructions for Use
Page 73
2 Safety and Care
Recycling Batteries
WARNING: Never dispose of batteries by burning or by flushing into any waste water system,
for example, a lavatory. Compromising the structural integrity of a battery can result in leakage or explosion and the potential for personal injury.
WARNING: Do not throw batteries into the trash. Collect and recycle used batteries separate
from other waste.
Item Estimated Use Period Handling Instructions
Lithium batteries 1.5 years
Battery packs (lithium-ion battery)
6 months
Recycle batteries according to local, state, and regional regulations. Use a battery collection program available in your country to recycle batteries.
To the extent required by local laws and regulations, Siemens will collect and recycle batteries for this product at no charge. Contact your local Siemens representative for battery shipment instructions.
See also: Caring for Batteries, p. 2-44
The battery is expected to hold a sufficient charge for
1.5 years from the start of service if:
The system is never plugged into the power outlet. The system is plugged into the power outlet, but the mains
circuit breaker is in the off position (
1.5 years.
If the system clock no longer keeps time, it could be time to replace the battery. Contact your local Siemens representative.
See also: Caring for the Battery Pack, p. 2-29
The battery is rechargeable.
= OFF) for the entire
Instructions for Use 2 - 45
Page 74
2 Safety and Care
Disposing of the Packaging Materials
Dispose of or recycle the packaging materials according to local, state, and regional regulations.
To the extent required by local laws and regulations, Siemens will collect and dispose of packaging materials for this product. For more information, contact your local Siemens representative.
Disposing of Components and Accessories
WARNING: Observe local, state, and regional regulations for the disposal of the ultrasound
system components and accessories.
Component or Accessory Handling Instructions
air filter tray
See also: Cleaning an Air Filter, p. 2-27
The air filter tray is reusable. Replace the air filter tray if the air filter is damaged in any
way. For example, the air filter has a hole, or the air filter tray will not fit in the air filter tray slot.
Energy Conservation
See also: Supplying Power to the System, Chapter 3, Instructions for Use
For moderate energy conservation when the system is not in use, place the system in standby status.
For improved energy conservation when the system is not in use, power off the system. Keep the system plugged into the power outlet. Ensure the mains circuit breaker is in the on position (I = ON).
For maximum energy conservation when the system is in storage, power off and unplug the system from the power outlet or switch the mains circuit breaker to the off position ( = OFF).
Note: Unplugging the system from the power outlet or switching the mains circuit breaker to the off position (
= OFF) for long periods of time can shorten the life of the system batteries.
2 - 46 Instructions for Use
Page 75

3 System Setup

Initial Setup ......................................................................................................... 3
Moving the System ............................................................................................. 3
Swivel Locking Brake .................................................................................... 4
Front Swivel Locking Brake .................................................................... 4
Back Swivel Locking Brake .................................................................... 5
Prior to the Move ........................................................................................... 6
During the Move ............................................................................................ 8
Shipping the System .................................................................................... 10
After the Move ............................................................................................. 10
System Startup ................................................................................................. 11
Plugging in the System ................................................................................ 11
Supplying Power to the System ................................................................... 12
QuikStart Feature (Battery-Powered Standby Mode) .................................. 14
Adjusting Controls on the Monitor ................................................................ 15
Controls on the Monitor ........................................................................ 15
Adjusting Brightness ............................................................................. 16
Adjusting the Monitor with the Monitor Menu Selections ...................... 17
Audio and Microphone Volume ............................................................ 18
Connecting and Disconnecting Transducers ............................................... 19
ACUSON X600 System ........................................................................ 19
ACUSON X700 System ........................................................................ 20
Array Transducers ................................................................................ 21
Continuous Wave Transducers ............................................................ 22
Protective Transducer Holder ............................................................... 23
Transducer Cable Management ........................................................... 23
Connecting System Accessories ................................................................. 23
Footswitch ............................................................................................ 23
ECG (EKG) Cables .............................................................................. 24
Auxiliary Inputs ..................................................................................... 24
Wireless Network Connections .................................................................... 25
Approved Wireless Adapters ................................................................ 25
Connecting the Wireless Adapter to the System .................................. 26
Informational Icons for the Wireless Adapter ........................................ 26
Instructions for Use 3 - 1
Page 76
3 System Setup
Input/Output Panel Connections ..................................................................... 27
Input/Output Converter Box (Video Converter) ............................................ 29
Connecting Peripheral Equipment ............................................................... 30
System Ergonomics ......................................................................................... 32
General System Settings ................................................................................. 34
Modifying Hospital or Clinic Name ............................................................... 34
Setting System Date and Time .................................................................... 34
Configuring the Documentation Controls ..................................................... 36
Software Installation ......................................................................................... 37
Software-Based Option Installation ................................................................ 39
3 - 2 Instructions for Use
Page 77
3 System Setup
Initial Setup
The ultrasound system is initially unpacked and installed by a Siemens representative. Your Siemens representative will verify the operation of the system. Any transducers, documentation and storage devices, accessories, and options delivered with your system are also connected and installed for you.
Each day before you use the ultrasound system, perform the Daily Checklist.
See also: Daily Checklist, Safety and Care, Chapter 2, Instructions for Use
Moving the System
WARNING: Preparations before moving the system are important to minimize potential damage
to sensitive components and to avoid safety hazards. Review the moving instructions before moving the system.
WARNING: To avoid damage to the monitor and the potential for personal injury to the user,
ensure the monitor does not swivel during transport. Lock the vertical position of the flat panel monitor prior to moving the ultrasound system according to the directions provided in this section of the instructions.
Caution: Do not push the flat panel monitor to move the system. Pushing on the flat panel
monitor can cause loss of control over the system and damage to the moving parts of the flat panel monitor.
Caution: Do not lean on the flat panel monitor. Subjecting the flat panel monitor to heavy loads
or extreme pressure can damage the ultrasound system.
Caution: Do not park, or leave unattended, on a slope. Even when the rear brakes are
engaged, the system may slide down a ramp.
Caution: To prevent damage to the system during a move, retract or close any doors and trays
on the documentation devices or components on the ultrasound system. Ensure components do not protrude from the system.
The ultrasound system is designed to be a mobile unit. Before moving the system to another location, you must prepare for the move by powering off and securing the system.
Instructions for Use 3 - 3
Page 78
3 System Setup
Lock the brake
Press the locking lever down with your foot into the locked position.
ON (on)
into place.
Swivel Locking Brake
The ultrasound system has four swivel locking brakes. The two front brakes each have a locking and locking release lever and a swivel and swivel release lever. The two back brakes each have a locking lever and a release lever. Set the swivel locking brakes with the levers on each wheel.
Front Swivel Locking Brake
Example of wheel with locking brake (unlocked).
1 Locking release lever 2 Locking lever 3 Swivel lock lever 4 Swivel release lever
To Do This Symbol
Release the brake
Lock the swivel
Release the swivel
Press the locking release lever down with your foot to release
the brake.
Press the swivel lock lever with your foot until the swivel brake locks
Press the swivel release lever with your foot until the brake unlocks
and the lock swivel lever releases.
OFF (off)
3 - 4 Instructions for Use
Page 79
3 System Setup
Example of wheel with locking brake (locked).
Example of wheel with locking brake (unlocked).
To
Do This
Symbol
position
Back Swivel Locking Brake
Lock the brake
Release the brake
Press the locking lever down with your foot into the lowest (locked)
Press the locking release lever down with your foot to release
the brake.
ON (on)
OFF (off)
Instructions for Use 3 - 5
Page 80
3 System Setup
Prior to the Move
1. Power off the ultrasound system. The power (partial) on/off ( ) switch is located on the upper left of the control panel.
2. Unplug the power cord from the wall outlet. Pull on the plug, not the cord.
3. Secure the power cord to avoid rolling the system wheels over the cord.
4. To ensure that the transducers are transported safely, remove each transducer and place it in its protective carrying case.
5. Retract or close any doors or trays of the documentation devices or components on the ultrasound system.
Retracting open doors or trays.
6. Disconnect off-board documentation devices from the system.
7. Transport gel and CD/DVD disks separately.
8. Disconnect the optional footswitch.
3 - 6 Instructions for Use
Page 81
3 System Setup
Example of transport lock in the unlocked position.
Example of transport lock in the locked position. Example of monitor in vertical position for transport.
9. Lock the position of the flat panel monitor for transport: a. Align the flat panel monitor to the front, center of the ultrasound system. b. Push and rotate the transport lock into the locked position. Ensure the locking pin
engages with the hole on the articulating arm.
1 Pin hole 2 Transport lock
1 Pin 2 Pin hole 3 Transport lock
10. Release both the front and rear brakes.
Instructions for Use 3 - 7
Page 82
3 System Setup
During the Move
Caution: When moving the ultrasound system, protect it from environmental changes including:
moisture, winds, dirt and dust, and extreme heat or cold exposure.
Caution: Avoid moving the ultrasound system on outside surfaces with loose dirt, contaminates,
or standing liquids.
Caution: Care should be taken to minimize shock and vibration of the ultrasound system. Avoid
uneven surfaces that contain an abrupt height change or jarring surface irregularities.
Caution: For systems installed with an on-board documentation device located on the rear shelf
of the ultrasound system, avoid damage to cables and connectors protruding from the documentation device, particularly when moving the system around corners or through doorways.
See also: Supplying Power to the System, p. 3-12
See also: Transducers — Care, Safety and Care, Chapter 2, Instructions for Use
3 - 8 Instructions for Use
Page 83
3 System Setup
Protruding cables and connectors.
You can move the ultrasound system from room to room within a facility and easily reposition the system during an examination. Be careful on inclines and uneven surfaces. The ultrasound system can be moved across pavement and other hardened parking lot surfaces.
Note: The wheels of the ultrasound system must be locked when transporting by vehicle. The ultrasound system must be sufficiently anchored to the vehicle floor or walls such that it does not shift or move during transport.
Instructions for Use 3 - 9
Page 84
3 System Setup
Shipping the System
When shipping the system, perform the following tasks, as appropriate.
To prepare the system for shipment over long distances or rough terrain:
1. Repack the system in the factory packaging and crate.
2. Load the system into a vehicle using a lift gate. To prevent lateral movement of the system, secure the system with cargo straps.
To prevent sudden jarring of the system during transport, provide shock cushions beneath the system.
After the Move
Caution: Make sure the ultrasound system has proper ventilation during operation. Do not
position the system against walls or hard surfaces that would impede free ventilation around the system.
Caution: Do not allow linens, bedding, and/or hanging curtain partitions to block the ultrasound
system's ventilation.
Caution: Obstructed fans can cause potential system overheating, system performance
degradation, or failure.
Caution: Brakes are most effective on a level surface. Never park the system on an incline
greater than five degrees.
See also: Swivel Locking Brake, p. 3-4
See also: Daily Checklist, Safety and Care, Chapter 2, Instructions for Use
1. Position the system: Make sure the system is not placed against walls or fabrics that obstruct perimeter air flow to the system cooling fans.
2. Set the brakes: Set the front and rear brakes.
3. Plug in the cord: Plug the power cord into a hospital-grade or local equivalent wall outlet.
4. Power on: Power on ( ) the ultrasound system.
5. Check the display: After the boot-up sequence is complete, verify that the image display is stable, that you can select a transducer, and that the system responds to selections made on the control panel.
3 - 10 Instructions for Use
Page 85
3 System Setup
+
+
+
+
1 2 3 4
5
6
System Startup
The first step to operating the ultrasound system is to connect the system to a power source.
Plugging in the System
WARNING: Before connecting the ultrasound system to a power supply, you must read and
understand the Electrical Safety section of Chapter 2, Instructions for Use.
The ultrasound system has a non-detachable power cord.
1 Equipotential connector 2 MAINS circuit breaker 3 MAINS Voltage Selector 4 Accessory Outlet Output Voltage Selector 5 Status Indicator (for Service diagnostics) 6 Connector cover for power cord
Example of power panel.
To plug in the system:
Connect the power cord to the MAINS supply:
200V~ to 240V~ systems: connect the power cord to a standard MAINS receptacle.
For example, a "Schuko" receptacle (CEE 7-7 standard).
100V~ to 120V~ systems used in the U.S.A.: connect the power cord to a hospital-
grade MAINS receptacle.
100V~ to 120V~ systems outside the U.S.A.: connect the power cord to a standard
MAINS receptacle.
Instructions for Use 3 - 11
Page 86
3 System Setup
Amber
Stand-by mode
Supplying Power to the System
The ultrasound system is powered on and off using the partial power on/off switch ( ) located on the upper left of the keyboard.
Note: This switch does not completely shut down or disconnect the system from the power mains. This switch only powers on, or off, a portion of the ultrasound system. To completely disconnect the system from the power mains, the circuit breaker located on the back panel must be switched from On to Off.
Caution: Wait approximately 20 seconds between powering the system off and then on again.
This allows the system to complete its shutdown sequence.
Power on/off button display System status:
Blue Powered on
Off (not lit) Powered off
To power on the system:
1. Before using the system, perform the Daily Checklist.
See also: Daily Checklist, Safety and Care, Chapter 2, Instructions for Use
2. Verify the power cord is plugged into the system and then into the power supply.
3. Power on ( ) the ultrasound system. When powered on, the system runs through a series of self-diagnostic and calibration
tests. The tests last a few minutes, after which the system is ready for use.
Note: The system will not run through the complete power-on routine if a problem occurs. Instead, an error code or message appears on the screen to indicate the problem. Please note the message and call your local Siemens service representative.
The system is factory-configured to display initially in 2D-mode.
4. Visually check the on-screen displays and lighting indicated in the Daily Checklist.
To power off the system:
Note: To cancel the power off procedure, select the Cancel button in the Shut Down System
dialog box.
1. Briefly press the partial power on/off ( ) button on the upper left of the ultrasound system. The system displays the Shut Down System dialog box.
2. Select the Shut Down button. Wait approximately 30 seconds before powering on the ultrasound system.
Note: In case the system is down, use the compulsory shutdown by pressing and holding the power on/off ( ) button for at least five seconds.
3 - 12 Instructions for Use
Page 87
3 System Setup
To restart the system:
1. Briefly press the partial power on/off ( ) button on the upper left of the ultrasound system. The system displays the Shut Down System dialog box.
2. Select the Restart button. The system will power off and restart.
Entering and Exiting Standby Mode
The standby function is not available during video playback or use of an application. Examples of applications include: Stress Echo, Axius EF, and SieScape.
To enter Standby mode:
Use the system presets to automatically close the current study when standby mode is activated.
General 1 > Close Study on System Standby
Note: The system preserves patient data when in standby mode. To avoid the possibility of data loss,
Siemens recommends that you close the current patient study, activate 2D-mode, and wait until the system finishes writing data to external storage media before entering standby mode.
1. Briefly press the partial power on/off ( ) button on the upper left of the ultrasound system.
2. Select Stand by from the displayed dialog box. The system enters standby mode.
3. To use the QuikStart battery-powered feature in standby mode, wait approximately 20 seconds before unplugging the ultrasound system.
To exit Standby mode:
Note: You must wait approximately 20 seconds after the system powers off before powering on ( )
the ultrasound system.
1. If you are using the QuikStart feature in standby mode, plug the power cord into a power outlet before exiting standby mode.
2. Press the partial power on/off ( ) button on the upper left of the ultrasound system. The system is ready for use in less than 15 seconds.
3. To re-enter standby mode, wait approximately 20 seconds after powering on the system.
Instructions for Use 3 - 13
Page 88
3 System Setup
Power on/off button display
Battery status
Flashing amber (fast)
Battery power is low.
QuikStart Feature (Battery-Powered Standby Mode)
The QuikStart feature for portable studies decreases the time required to power the system on or off by using the installed battery to place the ultrasound system in a standby status.
The system can maintain the standby status for approximately 30 minutes when the system's power cord is not plugged into the power supply.
When the system is in standby status running on battery power, the power on/off ( ) button functions as a battery power indicator.
Steady amber More than 60% charge remaining. Flashing amber (slow) More than 30% charge remaining.
Flashing amber (fast) with audible beep
Battery power is very low. If the system is not plugged in, it will shut down automatically within 5 minutes.
3 - 14 Instructions for Use
Page 89
3 System Setup
Example of monitor, front view.
Monitor button
Description
to the right or up.
Adjusting Controls on the Monitor
Use the controls located on the underside of the monitor to adjust the brightness settings of the monitor. Additional settings are available as menu selections from the monitor.
See also: Monitor Menu Selections, p. 3-17
Note: The language setting for the monitor menu is separate from the language setting of the
ultrasound system.
1 Monitor On/Off LED 2 Microphone 3 Microphone On/Off LED
Controls on the Monitor
Menu
Up (+)
Down (-)
Press Menu to activate and deactivate the brightness control. Press and hold Menu and Down (for at least 8 seconds) to display
the monitor menu with additional controls. Press Up to increase the settings for a control or to move the selection
Press Down to decrease the settings for a control or to move the selection to the left or down.
Example of monitor, side view.
1 Menu 2 Up (+) 3 Down (-)
Instructions for Use 3 - 15
Page 90
3 System Setup
Adjusting Brightness
Note: Factory-defined imaging presets use a default setting for the brightness of the monitor.
Adjusting the brightness setting on the monitor may affect the image optimization intended by the factory-defined imaging presets.
For consistency in image reproduction, adjustments to the brightness of the viewing monitor should be made prior to adjusting the print quality of installed documentation devices.
These processes assure consistent quality in the image display and reduces the potential for image quality issues.
To adjust the brightness of the monitor:
Note: Always begin adjusting your monitor with the controls set to the factory defaults, and then
adjust each control separately to suit your preference and the lighting conditions in the room.
1. Press the control on the monitor to display the setting for Brightness.
2. Press the control to decrease the setting or press the control to increase the setting. Decrease the setting for darker images; increase the setting for lighter images.
To restore and lock the factory default monitor settings for brightness:
1. Press the control on the monitor to display the setting for Brightness.
2. Press and hold the  control or the control (for at least 3 seconds). The system displays the OSD Main Menu is locked message.
3. To unlock the monitor settings, repeat step 2.
3 - 16 Instructions for Use
Page 91
3 System Setup
Exit
Message
Action
NO SIGNAL
OUT OF FREQUENCY
Adjusting the Monitor with the Monitor Menu Selections
You can adjust the monitor controls through the monitor menu.
Monitor Menu Selections
Menu Icon Menu Selection
Picture Brightness Sharpness Color Mode
Function Scale Information Memory Recall SBC Exit OSD Language OSD H-Position OSD V-Position Half Tone Exit Exit
To adjust monitor controls from the monitor menu:
1. Press and hold and  simultaneously on the monitor to display the monitor menu.
2. Press or to highlight a selection.
3. Press to select a Menu Icon.
4. Press or to select a menu selection and press to adjust the setting.
5. Press or to change the setting.
6. Select the EXIT button or press to confirm the setting and return to the main Selection screen.
7. To exit the monitor menu, allow the menu to time out (for 10 seconds), or press  or  to select EXIT, and then press again.
Monitor Error Messages
Check the signal cable Check if the operating frequency is out of range
Instructions for Use 3 - 17
Page 92
3 System Setup
Audio and Microphone Volume
You can adjust the volume of the speakers. Use the system presets to set a default volume level.
General 1 > Audio
You can also adjust the volume using the volume control on the control panel located above the DGC controls.
Example of the volume control.
To adjust the volume during Doppler:
Rotate the volume control on the control panel.
To activate the microphone during recording:
Use the system presets to adjust the volume during video playback (DVR).
General 1 > Audio > Line-in Volume
1. Press the volume control to activate the microphone. The system illuminates the LED located on the front of the monitor.
2. To adjust the volume, rotate the volume control.
3 - 18 Instructions for Use
Page 93
3 System Setup
Connecting and Disconnecting Transducers
Caution: Ensure that the system is in freeze before connecting and disconnecting transducers.
If a transducer is disconnected before the image is frozen, the system will display an error message, and it will be necessary to reset the system before continued use.
You can connect multiple transducers to the ultrasound system, with one transducer being the active transducer.
Note: When three transducers are connected to the system, the names of the transducers display as soft key selections.
ACUSON X600 System
Example of transducer ports.
1 Three 260-pin ports for standard array transducers 2 Continuous Wave (CW) Doppler port for CW Doppler pencil transducers
Instructions for Use 3 - 19
Page 94
3 System Setup
ACUSON X700 System
Example of transducer ports.
1 One 192-element port for micro-pinless transducers 2 Two 260-pin ports for standard array transducers 3 Continuous Wave (CW) Doppler port for CW Doppler pencil transducers
3 - 20 Instructions for Use
Page 95
3 System Setup
Array Transducers
Caution: You must freeze the system before connecting or disconnecting a transducer.
Note: When transducer connectors are being attached to or disconnected from the system,
resistance may be encountered due to the special shielding material inside the connectors. This is normal for these transducers.
1 Locked 2 Unlocked
Example of locked and unlocked positions of the connector.
To connect an array transducer:
1. Hold the transducer connector with the cable extending upward from the connector.
2. (260-pin array transducers) Insert the connector pins into the system port. (Micro-pinless transducers) Insert the connector into the system port.
3. Adjust the connector until you can turn the lever on the transducer connector clockwise to lock it in position.
This secures the connector in position and ensures the best possible contact.
4. Place the transducer in the transducer holder and drape the cable through the cable hangers.
To disconnect an array transducer:
Caution: To avoid damaging the transducer cable, do not pull on the cable to disconnect the
transducer. Use the following instructions.
1. Turn the lock on the connector housing counterclockwise until it unlocks.
2. Firmly grasp the transducer connector and carefully remove it from the system port.
3. Store each transducer in its protective carrying case.
Instructions for Use 3 - 21
Page 96
3 System Setup
Continuous Wave Transducers
Caution: You must freeze the system before connecting or disconnecting a CW transducer.
Connect a continuous wave transducer to the round sector port located to the left of the array ports.
Continuous wave transducer port.
1 Connector 2 Retractable connector housing 3 Transducer cable
Example of CW transducer connector.
To connect a continuous wave transducer:
1. Align the connector key until it fits smoothly into the port.
2. Insert the connector into the system port until it locks into position.
To disconnect a continuous wave transducer:
Caution: To avoid damaging the transducer cable, do not pull on the cable to disconnect the
transducer. Use the following instructions.
1. Pull on the connector housing ring to disengage the locking mechanism and remove the connector from the system port.
2. Store each transducer in its protective carrying case.
3 - 22 Instructions for Use
Page 97
3 System Setup
Protective Transducer Holder
Caution: Transducer holders have variable sizes both in depth and diameter. To avoid
transducer damage, you must use the holder and/or insert provided for transducers that have small or large diameter handles or for specialty transducers.
After connecting a transducer to the system, place the transducer in the protective holder attached to the control panel platform. A holder can also be used for the coupling agent (gel).
The liners of the transducer holders on the sides of the control panel are interchangeable and replaceable.
Transducer Cable Management
WARNING: To avoid injury to the patient or operator, use the cable hooks located on the
ultrasound system to manage one or more transducer cables. Entanglement with the transducer cables could result in injury.
After you have connected and secured a transducer, drape the transducer cable through one of the cable hangers located on the back of the control panel.
These hangers provide support for the transducer cables, keep cables off the floor, and help to prevent tangling of the cables when more than one transducer is connected to the system.
Connecting System Accessories
Footswitch
The ultrasound system has an optional dual-pedal, watertight footswitch.
Ingress Protection
With the exception of the connector on the footswitch, the footswitch meets Ingress Protection level IPX8 of EN 60539 and IEC 60539.
To connect the footswitch to the system:
Insert the footswitch connector into the USB port located on the back of the system.
Instructions for Use 3 - 23
Page 98
3 System Setup
ECG (EKG) Cables
The ECG feature allows the system to display a scrolling ECG waveform on the image screen.
ECG label identifying the connector socket.
To connect the ECG cables:
Insert the six-pin ECG connector into the socket labeled ECG on the left side of the
system.
Auxiliary Inputs
WARNING: The Aux 1 input is not intended for direct connection to the patient. To reduce the
risk of electrical shock while using this connection, ensure that the external source (peripheral equipment) is properly set up and is designed for direct patient connection.
Aux 1
Aux 1 connector socket.
To connect an external auxiliary cable:
1. Connect the one-pin external auxiliary cable to the socket labeled "Aux 1" on the front of
2. To connect the external auxiliary cable to a device (external source), see the device
the system.
manufacturer's operating instructions.
3 - 24 Instructions for Use
Page 99
3 System Setup
Wireless Network Connections
(Requires the wireless option license)
The ultrasound system can send data, such as studies, images, and clips, to a network location over a wireless network.
Use the system presets to configure wireless network connections.
Wireless
When the ultrasound system is connected to a wireless network, the system displays an icon indicating the connection to the wireless network and the signal strength on the status bar.
Note: Siemens recommends connecting the ultrasound system to wireless networks that use only the 5 GHz frequency bandwidth to reduce potential radio interference from the 2.4 GHz frequency bandwidth.
Approved Wireless Adapters
Caution: Use of wireless adapters not tested for use and approved by Siemens can damage the
ultrasound system. Do not connect any wireless adapters to the system except the wireless adapters listed below.
Note: Use of an unapproved wireless adapter may cause loss of data transferred over a wireless network.
The following wireless adapters have been tested for use and approved by Siemens.
Linksys AE1000  ASUS USB-N53
For information on the transmission and reception frequencies used by the wireless adapter, refer to the manufacturer's operating instructions included with the wireless adapter.
You must abide by the regulatory requirements of your country to use a wireless adapter. Refer to the manufacturer's operating instructions for information regarding certified use of the wireless adapter in your country.
Instructions for Use 3 - 25
Page 100
3 System Setup
Connecting the Wireless Adapter to the System
Caution: Maintain a distance of at least 25 cm of radius between the wireless adapter and
possible sources of interference. Possible sources of interference can be other medical devices and systems, including devices and systems in compliance with CISPR emission standards.
Caution: Ultrasound systems operate in the range of radio frequencies (RF) and are susceptible
to electromagnetic interference generated by other RF energy sources. To prevent this interference, increase the distance between the ultrasound system and the interfering RF energy source.
Siemens recommends connecting the wireless adapter to the leftmost USB port on the Input/Output panel of the ultrasound system.
Informational Icons for the Wireless Adapter
Icon Explanation
Connected to a wireless network, strong signal strength
Connected to a wireless network, good signal strength
Connected to a wireless network, weak signal strength
Connected to a wireless network, poor signal strength
Cannot connect to a wireless network, no signal
3 - 26 Instructions for Use
Loading...