This work is protected by copyright. Any rights derived from the copyright shall be
reserved for SICK AG. Reproduction of this document or parts of this document is only
permissible within the limits of the legal determination of Copyright Law. Any modifica‐
tion, abridgment or translation of this document is prohibited without the express writ‐
ten permission of SICK AG.
The trademarks stated in this document are the property of their respective owner.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
5
1 ABOUT THIS DOCUMENT
1About this document
1.1Function of this document
This document describes:
Installation
•
Initial start-up
•
Operation via SOPAS ET
•
NOTICE
This technical documentation is only valid in combination with the MARSIC300 Operat‐
ing Instructions.
1.2Target group
This document is addressed to technicians (persons with technical understanding)
operating and maintaining the measuring system.
The technicians must have been trained on the device.
Requirements for the technician
The technician must be familiar with the exhaust gas technology on the ship (over‐
•
pressure, toxic and hot flue gases) and be able to avoid hazards when working on
gas ducts.
The technician must be familiar with handling compressed gas cylinders (test
•
gases).
The technician must be able to avoid hazards caused by noxious test gases.
•
Only allow an authorized electrician to work on the electrical system or electrical
•
subassemblies.
1.3Further information
SFU Gas Sampling System Operating Instructions
•
Sample Gas Line Operating Instructions
•
System documentation
•
Short instructions for MARSIC300
•
Optional: MPR (Meeting Point Router) Operating Instructions
•
Optional: Instrument Air Conditioning Operating Instructions
•
Optional: External Measuring Point Switchover Operating Instructions
•
Optional: Pressure Test Tool Operating Instructions
Optional: HOTSAMPLER (measuring point extension) Operating Instructions
•
1.4Symbols and document conventions
1.4.1Warning symbols
Table 1: Warning symbols
Symbol Significance
6
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Hazard (general)
Hazard by voltage
Subject to change without notice
Symbol Significance
Hazard in potentially explosive atmospheres
Hazard by explosive substances/mixtures
Hazard by toxic substances
Hazards by noxious substances
Hazard by high temperature
Hazard for the environment/nature/organic life
1.4.2Warning levels / Signal words
ABOUT THIS DOCUMENT 1
DANGER
Risk or hazardous situation which will result in severe personal injury or death.
WARNING
Risk or hazardous situation which could result in severe personal injury or death.
CAUTION
Hazard or unsafe practice which could result in less severe or minor injuries.
Notice
Hazard which could result in property damage.
Note
Hints
1.4.3Information symbols
Table 2: Information symbols
Symbol Significance
Important technical information for this product
Important information on electric or electronic functions
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
7
2 INSTALLATION
2Installation
2.1Gas supply terminology
Definition of utility gases:
Zero gas: Gas to adjust the zero point. Instrument air or nitrogen (N2)
•
Span gas: Gas to adjust the upper measuring range value
•
Test gas: Generic term for zero and span gas
•
Instrument air: Clean compressed air
•
Gas quality: see "Supply gases", page 87.
2.2Installation information
2.2.1Information on power supply
The operator is responsible for correct laying and connection of the electric lines.
WARNING
Danger to life by electric voltage
Only allow an authorized electrician to work on the electric system
b
ATTENTION
The analyzer power supply is configured for an individual power system
Check the configured power system against the system documentation provided.
If the analyzer power system does not match the power system on board: Please
b
contact SICK Customer Service.
When selecting and laying the electric lines for power supply, observe the applicable
local standards and guidelines.
2.2.2Notes on the gas supply
The operator is responsible for the correct laying of the sample gas lines.
ATTENTION
Risk of contamination of the analyzer by unclean instrument air.
Only use instrument air corresponding to the mandatory specification (see "Techni‐
b
cal data", page 78).
Install a suitable instrument air conditioning when necessary.
b
8
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
2.2.3Tube screw fitting
1
Swagelok screw fitting
Figure 1: Swagelok screw fitting
INSTALLATION 2
Push the tube up to the stop in the
b
tube screw fitting.
Turn the cap nut finger-tight.
Push-in fitting (pneumatic)
Figure 2: Push-in fitting with
retaining ring
Retaining ring
1
During initial assembly: Hold the fitting
b
bolt steady and tighten the cap nut
with 1 1/4 revolutions.
During refitting: Tighten the cap nut to
b
the previous position (the resistance
increases noticeably) and then slightly
tighten.
Figure 3: Using the pressure tool
Pressure tool
Connecting the tube: Push tube in.
b
Disconnecting the tube: Press the retaining ring in and pull the tube out.
b
It is easier to press the retaining ring in using the pressure tool provided.
2.3Scope of delivery
Please see the delivery documents for the scope of delivery.
2.4Provision by operator
To be provided by operator, especially:
Suitable flange on the exhaust duct (see “Operating Instructions SFU”)
•
Fixing accessories of enclosures (dowels, screws, etc.)
•
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
9
1
2
33
4
5 6
7
8
1
2
1
4
2 INSTALLATION
Fixing accessories for heated sample gas line and tube bundle cable
•
Power cable: see "Power supply", page 82
•
External power disconnection unit
•
Compressed air, instrument air as zero gas when necessary
•
Observe quality of operator's instrument air: see "Supply gases", page 87
Optional span gases
•
2.5Installation overview
Figure 4: Installation - overview
Gas sampling system
1
Heated sample gas line (with 2 measuring points: 2 sample gas lines)
2
Tube bundle cable (with 2 measuring points: 2 tube bundle cables) with pneumatic and elec‐
3
tric lines
Analyzer
4
Energy supply
5
Interfaces
6
Instrument air inlet
7
Option: Instrument air conditioning
Sample gas outlet
8
2.6Checklist for mechanical and electrical installation
NOTICE
Observe the sequence during installation. Connect the gas sampling units on the
exhaust duct as last task.
Incorrect assembly can create a risk of contaminating the gas extraction system. In this
case, exhaust gas can penetrate the unheated analyzer and possibly condensate.
1.First connect instrument air and power supply.
2.Then install the gas sampling system in the exhaust duct.
Observe laying information (Chapter see "Assembly information for sample gas lines
and tube bundle cable", page 11).
Table 3: Fitting and connecting system components
System componentReference
Install analyzer cabinetsee "Installing the analyzer cabinet", page 13
10
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
System componentReference
Connect sample gas line to analyzersee "Connect the sample gas line to the analyzer",
Connect tube bundle cable to analyzersee "Connecting the tube bundle cable to the ana‐
Connect signal lines to analyzersee "Connecting the signal lines to the analyzer",
Air and gas connections on analyzersee "Air and gas connections on analyzer",
Electrical connections on analyzersee "Electrical connections on the analyzer",
Install SFU gas sampling systemsee "Installing the gas sampling system", page 25
see "Setting up the Modbus-Profinet converter
(optional)", page 27
See Operating Instructions HOTSAMPLER
2.7Assembly information for sample gas lines and tube bundle cable
Installing the sample gas lines
INSTALLATION 2
WARNING
Risk of fire
Observe the laying instructions provided with the line.
b
Minimum clearance to other lines (for example, tube bundle cable): 10 cm.
b
Do not lay or roll up lines directly next to each other.
b
WARNING
Danger to life by electric voltage
Only allow an authorized electrician to work on the electric system
b
Start laying at the analyzer.
b
The end with the electric connection belongs on the analyzer.
°
Important: The screw fitting for the enclosure duct must be located at the
end of the electrical connection (analyzer side).
The end without electrical connection belongs on the gas sampling system.
°
Roll-up excess length on the gas sampling system.
Leave enough length for pulling the gas sampling system.
Protect the line from damage (chafing through vibration, mechanical load).
b
Observe minimum bend radius of 300 mm.
b
Installing the tube bundle cable
Start laying the tube bundle cable at the analyzer and roll-up excess length at the gas
sampling probe.
1.Assembly on the analyzer: see "Air and gas connections on analyzer", page 20
and see "Electrical connections on the analyzer", page 23.
2.Lay the tube bundle cable to the analyzer cabinet.
Attach excess length to the gas sampling probe.
°
Leave enough length for pulling the probe.
°
Protect the line from damage (chafing through vibration, mechanical load).
°
Minimum bend radius: 300 mm.
°
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
11
≥ 100 mm
250 mm
R
≥ 30
0 mm
2 INSTALLATION
3.Assembly on the gas sampling probe: see "Installing the gas sampling system",
page 25.
NOTE
Fit the sample gas lines and tube bundle cable on the cable strips with holes.
Observe the minimum distance and bend radius.
b
Figure 5: Sample gas line - distance and radius
12
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
2.8Installing the analyzer cabinet
Ø 12
1200
1254
1230
20
400
460
600
INSTALLATION 2
Figure 6: Analyzer cabinet - dimensional drawing
NOTICE
Observe clearances:
Top: 30 cm
•
Bottom: 20 cm
•
Observe the clearances for the heated sample gas line.
•
Install the analyzer in a well ventilated room at a location, when possible, with con‐
stant temperature conditions.
Observe the relevant ambient conditions: see "Ambient conditions", page 80.
•
Attach the enclosure with the mounting brackets provided on a suitable wall with
•
•
sufficient load-bearing capacity.
Secure the enclosure with 4 x M10 screw fittings (2 at bottom, 2 at top) with prop‐
•
erty class 8.8 (or higher).
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
13
1
5
4
7
6
32
1
6
7
8
5
2
3
4
2 INSTALLATION
Secure the screw fittings against loosening.
•
Drilling plan: See dimension drawings above.
•
Install the enclosure horizontal.
•
2.9Connect the sample gas line to the analyzer
WARNING
Danger to life by electric voltage
Only allow an authorized electrician to work on the electric system
b
Connect heated sample gas line to analyzer
The flexible wires of the heated sample
gas line are numbered.
Line assignment: see "Heated sample
gas lines", page 81
Figure 7: Heated sample gas line
Connection side without electric connec‐
1
tions on gas sampling system
Connection side with electric connec‐
2
tions on analyzer
Protective cap
3
2 x Pt100 connections (1 as reserve)
4
Power supply
5
Cable gland
6
Counter nut
7
1.Unscrew counter nut from the cable gland and pull off the sample gas line.
2.Lead sample gas line together with electrical connections from above through the
housing opening on the analyzer.
3.Push counter nut back over the sample gas line and electric connections.
14
Figure 8: Sample gas line - connection diagram
Sample gas line 1
1
Sample gas line 2 (option)
2
Cable gland
3
Counter nut
4
Clamping ring screw connection (cell)
5
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
TUBE 1
100
100
gnge
sw 2
sw 1
sw 3
sw 4
sw 5
sw 6
PT100 I
PT100 II
Heizkreis I
FILTER 1PROBE 1FILTER 2PROBE 2TUBE 2TUBE 3
INSTALLATION 2
Cable duct
6
Foam insulation
7
Finished assembly
8
4.Screw counter nut tight on the cable gland.
5.Unscrew cell cover and remove.
CAUTION
Hot parts in cell.
Let cell cool down before working on the cell.
b
6.Remove protective cap from sample gas line.
7.Insert sample gas line to stop in the clamping ring screw connection on the cell.
With 2 sample gas lines: Observe inlets:
Right inlet: Measuring point 1
°
Left inlet: Measuring point 2
°
8.Screw sample gas line tight on the clamping ring screw connection.
9.Attach red foam insulation on the clamping ring screw connection and bind
together with a cable clip. No thermal bridges may remain.
10. Close cell again.
11. Screw cable gland tight.
12. Push electric lines downwards through the cable duct.
13. Connect power supply of the sample gas line:
Figure 9: Power supply connection
Complete description of the interface on page 82.
Connect the optional second sample gas line (TUBE 2).
b
NOTICE
The connections on MARSIC300 must match the connections of the gas sampling sys‐
tem.
14. Connect Pt100 of the sample gas line:
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
15
PT100DIGITAL INPUTSDIGITAL OUTPUTSVALVE OUTPUTS
100
100
gnge
sw 2
sw 1
sw 3
sw 4
sw 5
sw 6
PT100 I
PT100 II
Heizkreis I
1
2
3
4
2 INSTALLATION
Figure 10: Pt100 connection
Complete description of the interface on page 86.
15. Connect Pt100 from the sample gas line 2 to pins 9 and 10.
2.10Connecting the tube bundle cable to the analyzer
16
Figure 11: Analyzer - overview
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Tube bundle cable 1
1
Tube bundle cable 2 (optional)
2
3
4
Valve block
Electronics unit
Subject to change without notice
1
4
2
3
6
5
Figure 12: Tube bundle cable - overview
Power supply
1
Pt100 lines
2
Grounding conductor
3
PTFE tube DN4/6
4
PA tube black DN6/8; Imprint “1”
5
Connections on analyzer and gas sampling probe must match
PA tube black DN6/8; Imprint “2”
6
Connections on analyzer and gas sampling probe must match
INSTALLATION 2
No. DesignationFunctionDimension
Power suppliesLines 1 and 2: Gas sampling filter
1
Lines 3 and 4: Probe tube (optional)
Signal lines (Pt100)Lines 1 and 2: Gas sampling filter
2
Lines 3 and 4: Probe tube (optional)
Grounding conductor (gnge)Ground1 x 4.0 mm
3
PTFE hose (white)Zero gasDN 4/6
4
PA hose (black)Control air main valveDN 6/8
5
PA hose (blue)Backflush airDN 6/8
6
4 x 1.5 mm
4 x 1.0 mm
WARNING
Danger to life by electric voltage
Only allow an authorized electrician to work on the electric system
b
CAUTION
The tube bundle cable is unusable when it is cut off too short.
The tube bundle cable is unusable when a lead inside is damaged when removing the
sheath.
If the tube bundle cable is long enough: Leave a piece as “reserve”.
b
Only remove the sheath from the tube bundle cable when you are technically capa‐
b
ble of doing this.
2
2
2
1.Remove 1900 mm of the tube bundle cable sheath.
2.Lead the tube bundle cable through the screw fitting on the enclosure top.
3.Lead tubes and lines of the tube bundle cable downwards in the cable duct.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
17
1 2 3 4 5 67
8
9
1
4
2
3
6
5
TUBE 1FILTER 1PROBE 1FILTER 2PROBE 2TUBE 2TUBE 3
!
$
§
"
gnge
2 INSTALLATION
4.Connect the 3 gas lines of one tube bundle cable to the valve block (connections
marked as “outlet” in the Figure below).
NOTICE
The gas connections on the valve block must match the gas connections of the
gas sampling system (see "SFU Gas Sampling System Operating Instructions").
Figure 13: Valve block
Outlet: Zero gas measuring point 1
1
Outlet: Zero gas measuring point 2 (option)
2
Outlet: Control air measuring point 1
3
Outlet: Backflush air measuring point 1
4
Outlet: Control air measuring point 2 (option)
5
Outlet: Backflush air measuring point 2 (option)
6
Inlet: Zero gas
7
Inlet: Control/backflush air
8
Inlet: Auxiliary control air
9
Red plug = dummy plug
-
5.Connect the heating of the gas sampling system (heated sample gas filter and
optional heated probe tube).
NOTICE
The connections on MARSIC300 must match the connections of the gas sampling
system.
Complete description of the interface, see "Tube bundle cable", page 81 Tube bundle
cable 2 optional to tube bundle cable 1.
18
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
1
4
2
3
6
5
PT100DIGITAL INPUTSDIGITAL OUTPUTSVALVE OUTPUTS
!
$
§
"
INSTALLATION 2
6.Connect the Pt100 of the gas sampling system (heated sample gas filter and
optional heated probe tube).
NOTICE
The connections on MARSIC300 must match the connections of the gas sampling
system.
Figure 15: Signal lines - tube bundle cable connection diagram
2.11Connecting the signal lines to the analyzer
4 digital inputs and outputs each are available as an option that must then be config‐
ured, see "Data interfaces / IO", page 55.
4 digital inputs
Fill level signal, condensate container
•
Condition, instrument air
•
Scrubber system on/off (can be linked with StBy MARSIC300)
•
Temperature error, weatherproof cover or other external alarm
•
4 digital outputs
Status (OK / Maintenance)
•
Status (OK / Failure)
•
Coefficient SO2/CO2 less than/greater than xy (definable)
•
The inputs and outputs are deactivated as standard. The inputs and outputs can be
activated and negated in SOPAS ET.
Digital inputs and outputs can also be redefined differently to the above description.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
19
PT100DIGITAL INPUTSDIGITAL OUTPUTSVALVE OUTPUTS
3
12
4
7
5
6
8
9
ß
2 INSTALLATION
Figure 16: Connection diagram - digital connections
2.12Air and gas connections on analyzer
WARNING
Hazard when pressure is too high
Hoses can burst when the pressure is too high.
Observe the maximum pressures of the gases provided by the operator: see "Sup‐
b
ply gases", page 87.
20
Figure 17: Overview - air and gas connections
Heated sample gas line for measuring point 1
1
Heated sample gas line for measuring point 2 (option)
2
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
INSTALLATION
Tube bundle cable 1
3
Tube bundle cable 2 (option)
4
Pressure reducer unit
5
Instrument air as zero gas/test gas (option)
6
Test gas inlet
7
Sample gas outlet
8
Sample gas outlet on cell
9
Fixing of line sample gas outlet on the cell
ß
Connect instrument air
Connect instrument air to the pressure control unit.
ATTENTION
Risk of contamination of the analyzer by unclean instrument air.
Only use instrument air corresponding to the mandatory specification (see "Techni‐
b
cal data", page 78).
Install a suitable instrument air conditioning when necessary.
b
2
Figure 18: Pressure reducer unit
Instrument air inlet with zero gas quality
1
Instrument air inlet for induction air ejector only
2
Manual valve for instrument air selection
3
3 pressure reducers (adjustable)
4
Manual valve - closed position
5
Manual valve - open position
6
The instrument air is used as both induction air for the ejector (cell) and zero/control
air.
Instrument air can be connected in 2 ways:
One common instrument air supply for ejector air and zero/control air (inlet 1).
b
Separate instrument air supply for:
b
Ejector air (inlet 2)
°
and zero/control air (inlet 1)
°
Instrument air quality
The requirements for instrument air quality are lower when the air is only used as
ejector air than when used as zero/control air (zero gas quality) (see "Supply
gases", page 87).
When connected just as instrument air supply with zero gas quality to be used as
b
common air for both ejector air and zero/control air (on inlet 1):
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
21
1
2
2 INSTALLATION
Set manual valve to position “open”.
w
When connected as one instrument air supply for the ejector (on inlet 2) and one
b
instrument air supply with zero gas quality (on inlet 1):
Set manual valve to position “closed”.
w
Connect span gas (option)
WARNING
Hazard when pressure is too high
Hoses can burst when the pressure is too high.
Observe the maximum pressures of the gases provided by the operator: see "Sup‐
b
ply gases", page 87.
Connect span gas to the span gas valve.
Figure 19: Test gas valve connection
Test gas valve
1
Test gas inlet
2
Connect sample gas outlet
CAUTION
Acidic condensate in sample gas outlet - risk of blockage
Acidic condensate forms in the sample gas outlet.
Always lay the sample gas outlet line running downwards so that no condensate
b
can accumulate.
Lay the line end in a suitable disposal device (drain or condensate collection con‐
b
tainer).
Do not kink the line and protect against frost.
b
The analyzer is fitted with an outlet hose (DIN 8/10) about 30 cm long when delivered.
Lay the sample gas outlet in a suitable exhaust duct.
•
The sample gas outlet must be open to the ambient pressure.
•
Recommendation for lengthening the outlet hose:
To prevent the outlet line clogging, it is recommended to lay a (1) new line starting at
the sample gas outlet of the cell to the disposal location.
22
To do this, see figure 17, page 20:
1.Open cell cover (loosen 4 screws on the side).
2.Unscrew sample gas outlet line on the sample gas outlet (bracket piece).
3.Loosen fixing of sample gas outlet line.
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
1
INSTALLATION 2
4.Remove the thermal insulation from the existing line and remove the line.
5.Lead a new line through the enclosure duct at the bottom on the enclosure floor
and connect it to the sample gas outlet of the cell.
6.Refit the fixing.
7.Refit the thermal insulation.
8.Close cell again.
To lengthen the existing hose piece: No cross-section narrowing may occur when con‐
necting the lengthening hose.
Sample gas outlet on housing:
Figure 20: Sample gas outlet - enclosure under‐
side
Sample gas outlet at bottom rear of the
1
housing
2.13Electrical connections on the analyzer
Connect power supply
ATTENTION
The analyzer power supply is configured for an individual power system
Check the configured power system against the system documentation provided.
If the analyzer power system does not match the power system on board: Please
b
contact SICK Customer Service.
The power supply is located on the left on the analyzer.
Figure 21: Power supply connections
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
23
1
ETHOE TH1USBCAN PROFIBUS
1 2 RS422/RS485 RS232 02DISP.
PT100DIGITAL INPUTSDIGITAL OUTPUTSVALVE OUTPUTS
I/0-MOD.
2 INSTALLATION
NOTICE
Install an external power disconnection unit which disconnects all connectors
°
and fuses near the analyzer.
The power disconnection unit must be marked clearly and be easily accessi‐
°
ble.
Observe the maximum power input of the complete system: see "Power sup‐
ply", page 82.
The onsite wiring system to the power source of the system must be installed
°
and fused according to the relevant regulations.
Always connect a protective ground to PE.
°
Route the electric lines through the screw fittings of the enclosure.
b
Connect the electric lines.
b
Connect signal line (optional)
1
b
b
Connect Ethernet (optional)
Figure 22: Connections overview
2 x signal line ducts
Lead cable through the enclosure duct.
Attach shield as shown in the Figure above.
24
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
1
INSTALLATION 2
Table 4: Data interfaces - overview
PlugConnection for
ETH0Ethernet (e.g. SOPAS ET), MPR (remote maintenance), communication via
ETH1Internal
USBInternal
SD cardSD card (on the right, next to USB)
CAN1Internal
CAN2Internal
RS422, RS485Internal
RS232 (top plug)Internal
O2 (bottom plug)O2 sensor
DISP (top plug)Display
I/O-MOD (bottom plug)Internal
Modbus TCP
Ethernet cable duct
1
Lead cable through the enclosure duct.
b
Attach shield as shown in the corresponding Figure above.
b
Connect Ethernet to ETH0 (network or computer with SOPAS ET).
b
Plug type: RJ 45.
Connect MPR (remote maintenance via SICK Meeting-Point Router) to ETH0.
b
Further information, see “MPR Operating Instructions”.
2.14Installing the gas sampling system
NOTICE
Observe the ambient conditions of the gas sampling system: See “SFU Gas Sam‐
b
pling System Operating Instructions”.
Leave the sample gas line and tube bundle cable long enough to be able to pull
b
the sample gas system out of the exhaust duct.
All connections must match the connections in the analyzer (see "Connecting the
b
tube bundle cable to the analyzer", page 16 and see "Connect the sample gas line
to the analyzer", page 14).
Fitting the flange
Install the flange of the gas sampling system as described in the “SFU Gas Sam‐
b
pling System Operating Instructions”.
Observe the 10° tilt of the probe tube during installation.
°
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
25
min.70
10°
~
~30
1
2
34
2 INSTALLATION
Figure 23: Flange assembly
Probe tube
1
Stack wall
2
Welding neck flange
3
Gas sampling filter
4
Gas connections
Connect the following gas connections on the gas sampling system:
b
Heated sample gas line
°
Tube bundle cable
°
Electrical connections
Connect the following electric lines of the tube bundle cable on the gas sampling
b
system:
°
°
°
°
Black line 1: Main valve
•
Blue line 2: Backflush
•
White PTFE tube: Instrument air/test gas
•
Power supply, gas sampling system
Power supply, probe tube (when heated)
Filter Pt100 line
Pt100 line, probe tube (optional, when heated)
26
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Tube bundle cable
1
4
2
3
6
5
Figure 24: Tube bundle cable - overview
Power supply
1
Pt100 lines
2
Grounding conductor
3
PTFE tube DN4/6
4
PA tube black DN6/8; Imprint “1”
5
Connections on analyzer and gas sampling probe must match
PA tube black DN6/8; Imprint “2”
6
Connections on analyzer and gas sampling probe must match
INSTALLATION 2
2.15
No. NameFunctionDimension
Power suppliesLines 1 and 2: Gas sampling filter
1
Lines 3 and 4: Probe tube (optional)
Signal lines (Pt100)Lines 1 and 2: Gas sampling filter
2
Lines 3 and 4: Probe tube (optional)
Grounding conductor (gnge)Ground1 x 4.0 mm
3
PTFE hose (white)Zero gasDN 4/6
4
PA hose (black)Control air, main valveDN 6/8
5
PA hose (blue)Backflush airDN 6/8
6
Fit the gas sampling system on the flange
NOTICE
Risk of contamination of gas sampling system
First install the gas sampling system on the exhaust duct just before the analyzer
b
is switched on.
Install the gas sampling system: See “SFU Gas Sampling System Operating
b
Instructions”.
Setting up the Modbus-Profinet converter (optional)
4 x 1.5 mm
4 x 1.0 mm
2
2
2
As an option to Modbus, the MARSIC300 can also be configured with a Profinet or Profi‐
bus protocol. This requires an external Modbus-Profinet converter that may require
some adjustments.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
27
INSTALLATION
2
Configuration:
1.Install the converter onsite on a DIN rail and create a connection to the MAR‐
SIC300 (Modbus-TCP).
2.Connect the converter to 24 V DC.
3.Connect a PC via one of the Modbus-TCP connections of the converter with a net‐
work cable (crossover not required).
Use the “IP config” software on the Product-CD to find the IP address of the con‐
verter in the network using its MAC address (the MAC address is on a label on the
converter (next to the TCP interface).
Figure 25: Menu window Anybus IPconfig
4.Start a web browser (IE7.0 or 8.0), enter the IP address and a connection to xGateway to the web interface.
5.Define your specific IP address incl. subnet of the converter that matches your net‐
work and safety settings.
Figure 26: Menu Configuration/Modbus Client
6.Enter the name, IP address and port of the MARSIC300.
28
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
INSTALLATION
Figure 27: Menu Configuration/Modbus Servers
7.Enter address from PROFINET and save the settings. The address must be the
same as in the operator-side SPS.
2
Figure 28: Menu Configuration/PROFINET IO
8.Click “Apply” to activate all changes.
Figure 29: Menu Tools/X-gateway management
9.Setup the hardware configuration on the operator-side (operator SPS) according to
the following Table:
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
29
2 INSTALLATION
Figure 30: Byte sequence
10. Set the byte sequence. The byte sequence can be set in MARSIC300 when
desired.
A connection to the PC with SOPAS ET is then necessary.
30
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
INSTALLATION 2
Figure 31: Menu Tree Parameters/Modbus in SOPAS ET
Set the register swap in SOPAS ET to CD_AB for Siemens controls.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
31
INSTALLATION
2
Figure 32: Menu Modbus in SOPAS ET
2.16Setting up the Modbus-Profinet converter (optional)
As an option to Modbus, the MARSIC300 can also be configured with a Profinet or Profi‐
bus protocol. This requires an external Modbus-Profibus converter that may have to be
adjusted.
How to configure the external converter is described in the following:
1.Install the converter onsite on a DIN rail and create a connection to the MAR‐
SIC300 (Modbus-TCP).
2.Connect the converter to 24 V DC.
3.Connect a PC via one of the Modbus-TCP connections of the converter with a net‐
work cable (crossover not required).
Use the “IP config” software on the Product-CD to find the IP address of the con‐
verter in the network using its MAC address (the MAC address is on a label on the
converter (next to the TCP interface).
32
Figure 33: Menu window Anybus IPconfig
4.Start a web browser (IE7.0 or 8.0), enter the IP address and a connection to xGateway to the web interface.
5.Define your specific IP address incl. subnet of the converter that matches your net‐
work and safety settings.
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Figure 34: Menu Configuration/Modbus Client
6.Enter the name, IP address and port of the MARSIC300.
INSTALLATION
2
Figure 35: Menu Configuration/Modbus Servers
7.Enter the address of PROFIBUS and save the settings. The address must be the
same as in the operator-side SPS.
Figure 36: Menu Configuration/PROFIBUS DP-V1
8.Click “Apply” to activate all changes.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
33
2 INSTALLATION
Figure 37: Menu Tools/X-gateway management
9.Setup the hardware configuration on the operator-side (operator SPS) according to
the following Table:
Figure 38: Byte sequence
10. Set the byte sequence. The byte sequence can be set in MARSIC300 when
desired.
A connection to the PC with SOPAS ET is then necessary.
34
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
INSTALLATION 2
Figure 39: Menu Tree Parameters/Modbus in SOPAS ET
Set the register swap in SOPAS ET to CD_AB for Siemens controls.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
35
2 INSTALLATION
Figure 40: Menu Modbus in SOPAS ET
36
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
3Initial start-up
NOTE
Prerequisite: The system is completely installed and connected.
Before switching on
1.Check connections are connected properly as described in the “Installation” Chap‐
2.Relieve the pressure on all pressure regulators in the analyzer (see "Air and gas
3.Adjust the manual valve in the analyzer as described in see "Air and gas connec‐
4.Open the external instrument air supply.
5.Set pressures: see "Air and gas connections on analyzer", page 20.
6.Check the power supply of the analyzer matches the setting on the ship network:
7.Check the voltage connection on the gas sampling system.
8.Install the gas sampling system on the flange of the exhaust duct: See “SFU Gas
INITIAL START-UP 3
ters.
connections on analyzer", page 20): Turn the regulators counterclockwise.
tions on analyzer", page 20 .
see "Electrical connections on the analyzer", page 23.
Sampling System Operating Instructions”.
Switching on
1.Switch on the external power disconnection unit.
2.The green "Power" LED on the control panel goes on:
Energy supply is available.
3.The yellow and red LEDs go on sporadically.
4.Booting appears several times on the screen.
5.The Measuring screen appears.
6.The system heats up:
Only the green LED is on.
°
Display: Init/Heating up.
°
A downwards counter displays the maximum duration of the process.
7.Display: Conditioning.
8.Only the green LED is on and Measuring is shown in the status line.
The system is ready for operation.
If a measured values blinks: The measured value is outside the calibration range.
When the yellow or red LED is on: Press the Diag button and clear the error: Error
list see "Error messages and possible causes", page 93.
Final tests and parameter settings
1.Perform a leak tightness check: see "Leak tightness check during initial startup",
page 75.
2.Perform an internal adjustment: Display menu Adjustment/internal adjustment.
3.Connect computer with SOPAS ET on ETH0 (see "Connections in analyzer",
page 83) (see "Software SOPAS ET", page 39).
If the customer network is already connected: Disconnect the customer network.
4.If required: Set the times for automatic adjustments.
To change in SOPAS ET: Menu Adjustment/Parameter/Start times.
w
5.If required: Configure the measuring point switchover.
In SOPAS ET: Menu Parameterization/Sequence Controls/Sampling point pro‐
w
gram.
6.Set the IP address of the customer network for the ETH0.
°
In the SOPAS ET user interface: Click the pen symbol in the device tile of the
MARSIC300.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
37
3 INITIAL START-UP
Figure 41: SOPAS ET device
tile
7.Disconnect the computer from ETH0 and connect the customer network to ETH0.
Initial start-up is completed.
38
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
4Configuration software
4.1Software SOPAS ET
SOPAS ET can also be used to the configure MARSIC300 and to access the MAR‐
SIC300 logbook.
SOPAS ET runs on an external PC connected to the MARSIC300 via the Ethernet inter‐
face.
Connect MARSIC300 with software SOPAS ET
1.Make sure all electric and pneumatic connections are connected properly, start-up
has been carried out correctly and the system is running without error messages.
Connect your computer with SICK software SOPAS ET installed (on the Product CD
or free in internet under www.sick.com):
CONFIGURATION SOFTWARE
4
Figure 42: MARSIC300 network connection
NOTE
SOPAS ET connects automatically the first time MARSIC300 is started and
inquires whether the driver files (SDD files) for the connected device should be
downloaded. Downloading is not necessary because SOPAS ET loads the driver
files direct from the MARSIC300.
2.Start SOPAS ET on the PC and search for devices. When the MARSIC300 appears
in the right area in SOPAS ET, drag & drop it to the left Project area.
To change the IP setting of the MARSIC300, click on the pen icon (marked red in
the Figure) and confirm the message with “Yes”.
Figure 43: Menu dialog TCP/IP settings
3.If the MARSIC300 was not found, login on the MARSIC300 on the control panel as
“Authorized Operator” (password is “1234”). You can change the network settings
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
39
BackEnter
Measure
1 Digital outputs
2 Digital inputs
3 Ethernet
I/O config.3.3
/Diag/I/O
4 CONFIGURATION SOFTWARE
for the MARSIC300 here. Make sure the PC and the MARSIC300 are in the same
subnet and no firewall is active.
Set the device parameters matching your network in this menu and confirm with
“OK”. A warning appears to confirm that the device is set offline. Confirm with
“Yes”. The network configuration has been changed, however the old configuration
can still be seen. The IP configuration is updated on the operator panel and in
SOPAS ET after a restart (switching the MARSIC300 off and on).
4.2Saving parameters
Figure 44: Menu settings IP configuration
After configuration, the MARSIC300 must be adjusted to the network settings of
the customer network.
SOPAS ET can be used to backup and restore the MARSIC300 parameters:
40
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
CONFIGURATION SOFTWARE 4
1.Click the left mouse button to select the desired device in the main window.
2.Select “Project / Project save as” to save the SOPAS ET-Project file in a local direc‐
tory on the PC.
Figure 45: Menu dialog - Save Project file
4.3Backing up the logbook with SOPAS ET
It is useful for diagnostic purposes when the MARSIC300 logbook is made available to
Service. It can be downloaded easily and, e.g. sent per e-mail.
1.Login as “Authorized User”.
2.Select path: MARSIC300 / Diagnosis / Logbook.
3.Select “All” in the scroll down menu.
4.Select “Export” to save the log file in a local directory on the PC.
Figure 46: Menu Logbook in SOPAS ET
4.4Passwords
There are 2 passwords:
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Real constant (RCi)
Logbookchapter 7.4
System statuschapter 7.5.1
System infochapter 7.5
Timemeterchapter 7.5
Parameterchapter 6
Measuring componentschapter 6.1
Global definitions
Measuring screenchapter 6.2
Measuring Screen x
I/Ochapter 6.5
Hardware plan
Data
–External data
–OPC outputs (OPCOi)
–Modbus values (MBVi)
–Modbus input values (MBIVi)
–Modbus input flags (MBIFi)
–Modbus reference flags (MBIRFi)
Sequence control programschapter 6.4
Sampling point program (MPP)
Temperature controlchapter 6.6.1
Pressure controlchapter 6.6.2
Flow ratechapter 6.6.3
O2 sensorchapter 6.6.4
Logbookchapter 6.6.5
Instrument displaychapter 6.6.6
Devicechapter 6.6.7
Emitterchapter 6.6.8
Modbuschapter 6.6.7
Adjustmentchapter 6.5.4
Parameterchapter 6
Concentrations
Adjustment factors
Start times
Manual adjustmentchapter 6.3
Automatic adjustmentchapter 6.3
Maintenancechapter 8
Testschapter 8.1
Digital inputs
Digital outputs
Operating stateschapter 8.2
System maintenancechapter 8.3
Service logchapter 8.5
Restart systemchapter 8.4
Confirm active messageschapter 8.6
Replacing the analyzerchapter 8.8
Replacing the electronicschapter 8.9
Loading/saving the parameterschapter 8.7
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
43
5 MEASURING SCREENS AND DATA STORAGE
5Measuring screens and data storage
Only the configured Measuring screens are shown.
b
Figure 48: Menu Measuring
Screens
Configuring Measuring screens: see "Measuring screens", page 47.
•
Scaling Measuring screens.
•
5.1Measuring screen
The Measuring screen is shown as measured value box, bar graph or line writer
depending on the configuration.
Measured value box
Here, the measured values are shown in numeric format.
Double-click the desired Measuring screen
Component
1
Unit
2
Measured value
3
44
Figure 49: Menu Measuring Screen/Measuring Screen 1 ... x
Bar graph
Here, the measured values are shown as a bar graph.
Component
1
Unit
2
Measured value
3
Figure 50: Bar graph
Line Writer and data storage
Here, the measured values are shown in numeric format as measured value box and in
a time chart.
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Figure 51: Line Writer
MEASURING SCREENS AND DATA STORAGE 5
Component
1
Line Writer - shows the
2
activated measured val‐
ues (see under “Scaling
Measuring screens”).
The Line Writer first
starts the display when
it is called up for the
first time.
Changing the user level
deletes the history of
the line display.
5.2Data storage function
Start:
b
Start data storage.
Data that have been configured are stored (independent of the graphic represen‐
tation).
Maximum 65536 entries per curve are stored in a file and then a new file created
automatically (with an incremental index).
A dialog field appears above the line diagram to enter a target file name (.txt):
w
Figure 52: Storing the recording
Stop:
b
Stop recording.
View:
b
View recorded data.
A dialog field appears to select the file with the stored values.
The following then appears above the line diagram:
Figure 53: Loading the recording
Reset:
b
Delete all line displayed (no effect on data storage).
Scaling Measuring screens
Tap on a display to call up a screen for scaling:
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
45
5 MEASURING SCREENS AND DATA STORAGE
Figure 54: Menu Scaling
Font color (Line Writer colors are prede‐
1
fined)
Measuring screen precision
2
Example:
-2: 123.45
-1: 1234.5
0 : 12345
1 : 123450
Scale start value
3
Scale end value
4
(for Line Writer: for y-axis)
Line display active / not active
5
(for Line Writer)
46
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
6Parameterization
6.1Measuring components
Global definitions
This menu displays the global device information (e.g.: number of filter wheels, number
of activated components).
Figure 55: Menu Parameterization/Measuring component/Global Definitions
Active measuring components. Checkmark: Active
1
Internal use: Error threshold of reference energy as from which a message is generated.
2
Definition of measuring components
This menu displays the setting for the respective measuring component.
This menu only serves as information.
PARAMETERIZATION 6
Figure 56: Menu Parameterization/Measuring components/Definition of
components
6.2Measuring screens
Up to 8 Measuring screens can be configured and displayed:
Figure 57: Menu Tree Parame‐
ters/Measuring Screens in
SOPAS ET
Select the desired Measuring screen
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
47
6 PARAMETERIZATION
Double-clicking a Measuring screen opens a screen in which the Measuring screen can
be configured:
Selected Measuring screen
1
Do not display the Measuring screen
2
Line Writer (see example below)
3
1 measured value box
4
4 measured value boxes
5
16 measured value boxes (see example
6
below)
2 large, 8 small measured value boxes
7
6 bar graphs (see below)
8
3 bar graphs + 2 measured value boxes
9
3 bar graphs + 8 measured value boxes
ß
(hidden)
Click “Save” to store
à
Figure 58: Menu Measuring Screens - Lay‐
out selection
Measured value box
Example: Measuring screen 1 with layout III
(16 measured value boxes)
1 b
Enter desired names (tags). MAR‐
SIC300 tags: see "Tags (variable
names)", page 100
Scaling the measured value box: see
"Measuring screen", page 44
Figure 59: Menu Measuring Screens Example layout
The Measuring screen now looks like this.
Figure 60: Menu Measuring Screens - Com‐
ponent names
Bar graph
The component names and units can be, for
example, from the factory settings (e.g. RVi) the
source of a filter (e.g. FVi) or from the analyzer.
48
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Figure 61: Menu Measuring Screens - Bar
graph
The Measuring screen then looks like this:
PARAMETERIZATION 6
Example: Measuring screen 1 with bar
graph layout I (6 bars)
1 b
Enter desired names (tags). MAR‐
SIC300 tags: see "Tags (variable
names)", page 100
Scaling the measured value box: see
"Measuring screen", page 44
Figure 62: Bar graph - example
Line Writer
Figure 63: Menu Measuring Screens - Tags
The Measuring screen then looks like this:
Enter desired names (tags).
b
1
Scaling the measured value box: see
"Measuring screen", page 44
2
Scaling the time axis [sec, min or hrs],
scaling the y-axis: see "Measuring
screen", page 44
Measured value box
1
Line Writer
2
Figure 64: Menu Measuring Screens - Line
Writer
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
49
6 PARAMETERIZATION
6.3Adjustment functions
Start times
Menu: Adjustment/Parameter/Start times
This menu displays the start times of the “Cyclic triggers (CT1 .. CT16)”.
Figure 65: Menu Start times
Manual adjustment
Menu: Adjustment/Manual adjustment
Further information, see “MARSIC300 Operating Instructions”
Cyclic trigger name
1
Next start time
2
Deactivation via operator
panel or menu: Parameter/
Variables and functions/
Cyclic trigger (CTi)
Measured value of this
1
component is set to
zero
Measured value of this
2
component is set to the
nominal concentration
of the test medium
Measured values of all
3
components listed
above are set to zero
Figure 66: Menu Manual adjustment
Automatic adjustment
Menu: Adjustment/Automatic adjustment
Further information, see “MARSIC300 Operating Instructions”
50
Figure 67: Menu Automatic adjustment
Start adjustment of program displayed
1
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
The available programs are system-specific: see System documentation.
Parameter
6.3.1.1Concentrations of test gases
Menu: Adjustment/Parameter/Concentrations
This menu displays the concentrations of the span gases and internal adjustment filter.
The concentrations of the test gases can be set.
Figure 68: Menu Concentrations
PARAMETERIZATION 6
Enter the concen‐
1
tration of the test
medium
Only internal signif‐
2
icance
6.3.1.2Adjustment factors
Menu: Adjustment/Parameter/Adjustment factors
This menu serves to set the correction factors (adjustment filter wheel optional).
•
•
•
Figure 69: Menu Adjustment factors
1
2
3
6.3.1.3Start times
Menu: Adjustment/Parameter/Start times
This menu displays the start times of the "Cyclic triggers (CT1 .. CT16)“.
The measured value is computed with both adjustment factors.
Factor "Internal adjust." is set automatically to "1.000" when factor "Span gas" is
changed.
Checkmark “Active”: This measuring component is computed.
Display: Current factor
Input: New factor
Confirm: Use new factor
Cyclic trigger name
1
Next start time
2
Deactivation via operator
panel or menu: Parameter/
Variables and functions/
Cyclic trigger (CTi)
Figure 70: Menu Start times
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
51
6 PARAMETERIZATION
6.3.1Manual adjustment
Menu: Adjustment/Manual adjustment
Further information, see "MARSIC300 Operating Instructions”
Figure 71: Menu Manual adjustment
6.3.2Automatic adjustment
Menu: Adjustment/Automatic adjustment
Further information, see "MARSIC300 Operating Instructions”
Measured value of this
1
component is set to
zero
Measured value of this
2
component is set to the
nominal concentration
of the span gas
Measured values of all
3
components listed
above are set to zero
Figure 72: Menu Automatic adjustment
Start adjustment of program displayed
1
The available programs are system-specific: see System documentation.
6.4Measuring point switchover / Sequence control program
Sampling point program
Menu: Parameterization/Sequence control program/Sampling point program
This menu serves to configure "measuring at several sampling points".
52
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
12
344
55
6
3
PARAMETERIZATION 6
CAUTION
The programs are automatically controlled.
The device no longer measures correctly when a program is interrupted with Start/
Pause/Stop/Continue.
Only use these functions when you are sure you can assess the consequences.
b
Figure 73: Measuring point switchover - diagram
Measuring point 1
1
Measuring point 2
2
Hold time
3
Active
4
Duration
5
Cycle time
6
Cycle time
The complete defined measuring cycle including measuring point switchover is com‐
pleted within the cycle time.
Duration
The time during which a measuring point is active and measuring (including run-in time
after measuring point switchover).
Hold time
After measuring point switchover: The time in which the last valid measured value of
the measuring point is held until the new measuring point has run-in (purge processes
etc.).
The hold time can be entered individually for each measuring point and serves, for
example, to control the release of the measured value to Modbus or similar.
Active
Valid measured values of the active measuring point are available after the run-in time.
Figure 74: Menu Sampling point program (MPP)
Display: Internal tag to enable the sampling point program
1
Display: Run-time for the current sampling point
2
Start the sampling point program
3
Interrupt the sampling point program (pause)
4
Display: Active sampling point
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
5
Display: Index of active sampling point
6
53
6 PARAMETERIZATION
Cancel the sampling point program
7
Continue the sampling point program after "Pause"
8
Input: Number of sampling points
9
Minimum measuring time per sampling point
ß
Checkmark: Fixed cycle time 285 s (“DeSOx time”)
à
Checkmark: Fixed cycle time from "Cycle time"
á
Checkmark: Fixed sampling point time
â
Input: Cycle time
ã
DeSOx-Time (à)
The DeSOx-Time is based on Guideline MEPC.259(68). This requires that each measur‐
ing point must be measured once within 4:45 minutes (285 seconds).
To enter in the menu below: Subtract the hold times from the 4:45 minutes and divide
the remaining time by the number of measuring points. For example: For 2 measuring
points with the same hold time, each measuring point is active for 142 seconds.
This can be used, for example, to control the release of the measured value on Modbus.
Fix.Cycle-Time (á)
The“Fix.Cycle-Time” corresponds to the “DeSOx-Time”, however the cycle time is not
specified fixed but can be entered explicitly (see menu above).
Fix.Msp-Time (â)
Measuring on each measuring point continues for the time specified under “Duration”
in the menu.
This means different measuring times can be defined per measuring point. If the dura‐
tion set is shorter than the “hold time + minimum measuring time”, the duration is
automatically set to this time. The following is also valid - the total of “hold times + the
minimum measuring time” per measuring point may not be longer than the cycle time.
Otherwise the cycle time is increased accordingly.
54
Figure 75: Measuring point program - menu description
Sampling point index
1
Sampling point name
2
Sampling point sequence
3
Measuring time at the sampling point including run-in times
4
Run-in time
5
Checkmark active: Sampling point active
6
Figure 76: Measuring point program - setting options
Display: Sampling point index
1
Checkmark: Sampling point active
2
Sampling point name
3
Overall duration at the sampling point
4
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Run-in time
5
Position in the sequence of the sampling point program
6
NOTICE
The green arrow (load all parameters) must be pressed after making changes in order
to see the effect of the changes for both measuring points. This then updates the data
in the Table accordingly.
Figure 77: Load parameters - button
6.5Data interfaces / IO
Menu: Parameterization/I/O
This menu displays the data interfaces.
Hardware Plan
Menu: Parameterization/I/O/Hardware plan
PARAMETERIZATION 6
6.5.1Digital inputs
CAN bus address x
Displays the I/O modules present in the selected CAN bus gateway.
NOTE
The sequence of the specified modules must match the sequence of the plugged in
modules (starting at the gateway).
Serves to assign a name, unit and data type to the measured values stored by the Mas‐
ter in the Holding register, whereby i is the index. Two registers incrementing from the
offset are used per measured value. A status is not considered.
Edit window:
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
(default value: 130 l/h)
Input: Maximum flow rate warning thresh‐
4
old (default value: 600 l/h)
Input: Maximum flow rate error message
5
(default value: 1000 l/h)
Figure 85: Menu Flow monitoring
Menu: Parameterization/O2 Sensor
This menu serves to view the configuration of the O2 sensor.
6.6.5Logbook
Menu: Parameterization/Logbook
This menu serves to configure the logbook.
Changing the setting deletes all entries.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
59
6 PARAMETERIZATION
Figure 86: Menu Logbook
6.6.6Instrument display
Menu: Parameterization/Instrument display
This menu serves to configure how the instrument display shows information:
•
•
•
If an error occurs:
An error counter is incremented.
1
The error message is stored.
2
When the logbook is full:
Message "Logbook error" is output and no fur‐
3
ther entries are stored.
The current entries overwrite the oldest entries.
4
A relevant message is not output.
The Measuring screen.
The password for the operator panel on the display.
If parameters have been changed and the Measuring screen is being shown on
the instrument display: The display on the instrument display must be reinitialized
so that the changes can be shown on the instrument display.
1Press the menu-dependent button "Menu“ on the device.
2Then press "MEAS“.
6.6.7System
60
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Figure 87: Menu Device display
Password, comprising 4 digits. (Information: Is first used after a hardware reset (Restart)).
1
Validity duration of the password-protected level. (Information: Is first used after a hardware
2
reset (Restart)).
Configuration (see "Tags (variable names)", page 100) of the Measuring screen.
3
Start and end values of the display range of the bar and line charts. Start value must be
4
lower than the end value (no plausibility check during input)..
Counter format of the display: Number of decimal places.
5
Menu: Parameterization/System parameters
This menu serves to display and change system parameters.
Subject to change without notice
PARAMETERIZATION 6
Figure 88: Menu Device parameters
System name
1
Checkmark: The IP configuration of the MARSIC300 can be changed in SOPAS ET.
2
Device language.
3
Accept with “Initialize” (see "Setting operating states", page 69)
Use PC time on MARSIC300.
4
6.6.8Emitter
Menu: Parameterization/Emitter
This menu serves to view the parameters for the emitter.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
61
7 DIAGNOSIS
7Diagnosis
7.1Control values adjustment, sensors and signals
Zero drift
Menu: Diagnosis/Control values/Zero drift
This menu displays the drift "with zero gas" (since the last drift "reset") and can be
reset.
This drift is recalculated as from the relevant adjustment.
Figure 89: Menu Zero drift
Reset: Reset zero drift
b
Both “Reset” have the same effect.
Span gas drift
Menu: Diagnosis/Control values/Drift span gas
This menu displays the drift “with test gas” (since the last drift “reset”) and can be
reset.
This drift is recalculated as from the relevant adjustment.
This menu displays the current state of
the digital signals (DIi, DOi, limit val‐
ues)
Digital values:
. = off (0)
I = on (1)
Figure 94: Menu Digital signals
Measuring signals
Menu: Diagnosis/Signals/Measuring signals
This menu displays the measuring sig‐
nals.
Let SICK Customer Service evaluate
these signals.
Figure 95: Menu Measuring signals
Internal signals
Menu: Diagnosis/Signals/Diagnosis internal
Figure 96: Menu Internal signals
This menu displays the signals.
Let SICK Customer Service evaluate
these signals.
64
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Boolean values
Menu: Diagnosis/Signals/Boolean values
Figure 97: Menu Boolean values
Real values
Menu: Diagnosis/Signals/Real values
Figure 98: Menu Real values
Modbus values
DIAGNOSIS 7
This menu displays
the Boolean values
(BVi) and limit val‐
ues (LIi).
This menu displays
the current real val‐
ues (RVi).
This menu shows the current Modbus values and Modbus flags
Modbus values
•
Modbus input values
•
Modbus input flags
•
Modbus reference flags
•
Further information on Modbus: see "Modbus", page 56 and see "Data interfaces / IO",
page 55
Filtered values
Menu: Diagnosis/Signals/Filtered values
This menu displays
the current filtered
values (FVi).
Figure 99: Menu Filtered values
Integer values
Menu: Diagnosis/Signals/Integer values
This menu displays
the current integer
values (IVi).
Figure 100: Menu Integer values
Real constants
Menu: Diagnosis/Signals/Real constants
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
65
7 DIAGNOSIS
7.4Logbook
This menu displays
the current real
constants (RCi).
Figure 101: Menu Real constants
The logbook records warnings, value overflows and underflows as well as errors and
serves to document and reconstruct occurring events. The logbook can be saved with
SOPAS ET (see "Saving parameters", page 40) and, for example, sent to SICK Customer
Service per e-mail.
Maximum number of entries: 6000.
(Representation: uncompressed data storage)
Figure 102: Menu Diagnosis/Logbook
Table 10: Logbook - Menu legends
NameRemark
Logbook fill level in %.
If the font is red: The logbook is full.
Warning mode: No further entries are made.
Ring buffer mode: The oldest entries are overwritten.
Data storage:
Symbol not crossed out: Compressed.
Symbol crossed out: Uncompressed.
Significance and default setting: see "Logbook", page 59
Ring buffer mode warning mode, significance and default setting: see "Logbook", page 59
1
2
3
4
Number of entries of selected filter.
Only filtered messages are shown.
Failure (active)
•
Failure (all)
•
Maintenance request (active)
•
Maintenance request (all)
•
Uncertain (active)
•
Uncertain (all)
•
Other (active)
•
Other (all)
•
All active
•
All
•
Attention: All logbook entries in MARSIC300 are deleted
All entries selected with the filter (see in the Table further above) are stored on the PC as log file in C:\Own files.
Format: CSV (comma-separated list). Viewable in, e.g., EXCEL.
66
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
NameRemark
5
6
7
To update the display: Click the display.
Scroll to older entries.
Scroll to current entries.
Consecutive number of the message.
Red LED: Message still pending.
Green LED: Message no longer pending.
DIAGNOSIS 7
8
9
ß
à
á
â
ã
ä
Triggering unit: System, measured value name (sample gas component), subassembly, Evaluation module.
Number of times the error has occurred.
Significance and default setting: see "Logbook", page 59
Only with "compressed data storage".
Logbook message (error messages see MARSIC300 Operating Instructions).
F = Failure
M = Maintenance request
C = Check (function control / maintenance)
U = Uncertain
X = Extended
Format: yy-mm-dd
For "uncompressed": When message occurred.
For "compressed": Last time message occurred
Format: hh:mm:ss
For "uncompressed": When message occurred.
For "compressed": Last time message occurred
Format: yy-mm-dd
For "uncompressed": When message was deleted
For "compressed": Last time message was deleted.
Format: hh:mm:ss
For "uncompressed":When message was deleted
For "compressed":Last time message was deleted.
7.5System info
System info
Menu: Diagnosis/System info
This menu displays system status (see “MARSIC300 Operating Instructions”) and sys‐
tem information.
Serial number
•
IP address
•
Software version
•
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
67
7 DIAGNOSIS
7.5.1System status
7.5.2System info
LEDs in accordance with the
1
operator panel
With measuring point switch‐
2
over: Current measuring
point
With measuring point switch‐
3
over: Remaining time until
measuring point switchover
Figure 103: Menu Diagnosis/System status
Name of current sequence
4
control program
Remaining time of current
5
sequence control program
System info
Menu: Diagnosis/System info
This menu displays system status (see “MARSIC300 Operating Instructions”) and sys‐
tem information.
Serial number
•
IP address
•
Software version
•
7.5.3Timemeter
This menu serves to view diverse operating hours counters.
Operating hours
1
This counter shows the total operating
time (“Power on”) of the Analyzer mod‐
ule.
Measuring duration
2
This counter shows the total time sam‐
ple gas was fed.
Operating time of the light source
3
Operating time of the filter of the gas
4
sampling system
Operating time of the cell inlet filter
5
Figure 104: Menu Diagnosis/Timemeter
68
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
8Maintenance
8.1Test digital I/O
Menu: Maintenance/Tests
This menu serves to test the digital interfaces.
b
b
b
b
Test digital inputs
MAINTENANCE 8
Click the desired interface (mark).
Perform with „Test“.
A menu to set the parameters appears.
(Field “Save” has no significance).
Figure 105: Menu Digital inputs
Test digital outputs
Figure 106: Menu Digital outputs
8.2Setting operating states
Menu: Maintenance/Operating states
This menu serves to switch the operating states of the MARSIC300 on.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
LED on:
Status signal “Mainte‐
nance” is switched on.
System Stop.
2
Switch to measuring
3
operation.
Switch to measuring
4
operation (after
changes in menu:
Parameterization/Meas‐
uring components/...).
This menu serves to start various maintenance procedures.
Figure 108: Menu System maintenance
Standby
1
Switches the system to Standby to put it out of operation for some time.
Exit Standby
2
Switches the system back to regular measuring operation.
(After it was switched to Standby using item "1" of this menu).
Cancel system maintenance
3
Abort a program started in this menu.
70
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
MAINTENANCE 8
Blow back probe
4
Triggers a backflush of the probe tube (as well as the inlet filter when fitted on the
probe tube) with instrument air.
Press "Start" to start the backflush.
•
Operating state: "Maintenance".
•
The measuring screen appears with a downwards counter to the end of the
•
adjustment.
The system switches back to operating state "Measure" when the adjustment
•
has completed (when "Maintenance" was set manually beforehand: Back in
"Maintenance").
Leakage test
5
Start a leak tightness check.
Test pressure sensors
6
Checks the pressure sensors.
Perform this check when you have the impression that a pressure sensor is defec‐
tive.
“Test OK” means: The pressure sensors are OK.
•
“Test failed” means: One of the pressure sensors is defective. Replace pres‐
•
sure control module.
Adjust pressure sensors
7
After replacing the pressure control module: Perform this menu item.
8.4Restart
8.5Service log
NOTE
Further information, see “MARSIC300 Operating Instructions”.
Menu: Maintenance/Restart System
This menu serves to restart the system.
Start "Restart system"
1
Figure 109: Menu Restart
system
Menu: Maintenance/Service Log
Completed maintenance work can be entered in this Table as clear text.
Figure 110: Menu Service Log
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
71
8 MAINTENANCE
Date [dd.mm.yyyy]
1
Service engineer name
2
Completed maintenance work in clear text
3
8.6Acknowledging messages
Menu: Maintenance/Confirm active messages
This menu resets all pending active messages.
LED "MAINTENANCE REQUEST“ goes off.
Figure 111: Menu Confirm
active messages
8.7Loading/saving the parameters
Parameters can be loaded/saved as parameter sets in different versions:
Table 11: Parameter sets
Factory setting/opt/analyser/backup/
Parameter 1/pccard/backup1/
Parameter 2/pccard/backup2/
Parameter 3/pccard/backup3/
Reset active messages
1
72
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
MAINTENANCE 8
Figure 112: Menu Tree Loading/saving parameters in SOPAS ET
8.8
Loading/saving the parameters after replacing the analyzer
Load the configuration of the old analyzer module to the new analyzer module after
replacement. This can be done easily using SOPAS ET.
1.Insert the SD card of the previous electronics in the new electronics.
2.Switch the device on.
3.Login as “Authorized User”.
4.Select menu Maintenance/Replace analyzer.
5.Click <Load customer-specific parameters> (This loads just the customer-specific
data from the SD card).
6.Wait until the device restarts.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
73
8 MAINTENANCE
Figure 113: Menu Tree Analyzer replacement in SOPAS ET
8.9
Loading/saving the parameters after replacing the electronics
Load the configuration of the old electronics unit to the new electronics unit after
replacement. This can be done easily using SOPAS ET.
1.Insert the SD card of the previous electronics in the new electronics.
2.Switch the device on.
3.Login as “Authorized user”.
4.Select menu Maintenance/Electronics replacement.
5.Click <Load all parameters> (this load all data from the SD card).
6.Wait until the device restarts.
Figure 114: Menu Tree Electronics replacement in SOPAS ET
74
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
9Maintenance
1
9.1Leak tightness check during initial startup
1.Perform the leak tightness check when the device is running.
WARNING
Risk of burns on hot cell
The cell is very hot (approx. 200 °C).
For the leak tightness check, the line on the sample gas outlet must be unscrewed
when the cell is hot.
Use heat-resistant gloves.
b
Use heat-resistant tool.
b
1.Start program Maintenance/Maintenance Sys./Leakage Test .
2.Wait until message “Close outlet - discon. purge” appears.
MAINTENANCE
9
Figure 115: Sample gas outlet (inside
view)
Sample gas outlet
1
Figure 117: Sample gas outlet - cell
Sample gas outlet on cell
1
Figure 116: Sample gas outlet (out‐
side view)
Sample gas outlet at bottom
1
rear of the housing
Figure 118: Analyzer module - Con‐
nections
Rear (thin) flush air line
1
Instrument air valve (shown
2
"open")
3.On Analyzer module: Disconnect rear purge air line (press ring to do so).
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
75
9 MAINTENANCE
4.Close off sample gas outlet gas-tight:
Either at the end of the sample gas outlet line (the line does not end at the
°
enclosure duct but runs through to the cell in the enclosure).
The leak test tool kit has a suitable plug to close off the sample gas outlet
line.
Or on the cell at the sample gas outlet (10 mm clamping ring sealing plug,
°
the plug is also in the leak test tool kit).
To do this, open the cell: Loosen 4 side screws and remove cover.
5.The pressure in the system slowly rises.
Message “close air valve” appears when the pressure is ≥ 1200 hPa (after about
30 seconds) (the current pressure is shown in the “Measured value display”).
6.Close instrument air valve.
The pressure no longer rises and measurement starts automatically after
°
about 20 seconds: Measurement duration approx. 5 minutes.
Pressure loss during this time must not exceed 20 hPa. A message is shown:
°
7.Open instrument air valve again.
8.Wait until message “Reopen outlet - connect purge” appears.
9.Reconnect rear purge air line.
10. Restore sample gas outlet to its original state.
“Test OK - open air valve”: Test successful.
•
“Test failed - open air valve”: Test unsuccessful: Analyzer switches to
•
“Maintenance request” state.
9.2Leak tightness check with pressure test tool
This Section describes the leak tightness check with the SICK “Leak test tool kit”.
Table 12: Screw fittings
CapSize
Probe tube
Sample gas inlets2 x clamping ring screw fittings 8/10
Span gas inlet1 x clamping ring screw fitting 4/6
Ejector block3 x clamping ring screw fitting 4/6
Procedure
1.Switch analyzer to “Standby”: see "System maintenance (Stand-by, Leakage test,
etc.)", page 70.
2.Flush system for 10 minutes in this state.
Start removing the probe during the flush duration: See “SFU Gas Sampling Sys‐
tem Operating Instructions”.
3.Close off external instrument air supply.
4.Close probe tube at gas inlet.
5.Open cell enclosure.
6.Unscrew exhaust gas line 1 from sample gas outlet of ejector.
76
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
1
Figure 119: Sample gas outlet (inside view)
Ejector sample gas outlet
1
MAINTENANCE
9
7.Connect PTFE connecting hose of the test tool to sample gas output of ejector.
8.Unscrew all further lines marked in the Figure and close connections on the cell
gas-tight.
9.Perform a leak tightness check: See “Pressure Test Tool Operating Instructions”.
Test duration: 5 minutes
°
The "leak" pressure rise must be < 10 mbar (< 0.14 psi). If the pressure rise
°
is higher, search for and clear the leak in the gas path.
10. Connect all lines again after successful leak tightness check.
11. Refit the probe.
12. Switch standby off again.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
77
Ø 12
1200
1254
1230
20
400
460
600
10 TECHNICAL DATA
10Technical data
NOTE
The Technical data depend to some extent on the individual equipment of your analyzer.
See the system documentation provided for the configuration of your analyzer.
Observe the specified quality of the instrument air.
If required, provide for instrument air conditioning.
GasQualityInlet pressureFlow rate
Instrument air (zero
gas quality)
Instrument air solely
as induction air for
ejector
External span gasPrecision: ± 2 %
Particle size max. 1 μm
Oil content max. 0.1 mg/m3
Pressure dew point max. –
40 °C
Purity class 2 (ISO 8573)
Particle size max. 5 μm
Oil content max. 1 mg/m3
Pressure dew point max.
+3 °C
Purity class 3 (ISO 8573)
Concentration: 80% ... 100%
of measuring range
The span gas must comply
with the specifications of the
standards to be applied (e.g.,
MARPOL Annex VI)
600 ... 700 kPa
(6.0 ... 7.0 bar)
500 ... 700 kPa
(5.0 ... 7.0 bar)
Max. 400 kPa
(4.0 bar)
Approx. 350 l/h
Approx. 1300 l/h
Approx. 350 l/h
10.13Tube connections
Table 36: Tube connections
ConnectionDimension
Sample gas inletClamping ring screw connection 6 mm
Ejector induction airDN 6/8
Span gas inletClamping ring screw connection 6 mm
Gas outletDN 8/10
10.14Torques
Tighten all screw connections, for which no tightening torque or no pretension force is
specified in drawings or Assembly Instructions, according to VDI 2230.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
87
10 TECHNICAL DATA
Exceptions to this rule are all connections with screws that are not screw connections in
the real sense. This includes hose clips, cable glands, screw fittings, gas connections,
screws for circuit boards etc. Tighten these screw fittings as evenly as possible with a
much lower torque (hose clips 1 Nm, other screw fittings according to manufacturer
specifications).
Select the next lowest torque valid for the screw for mixed materials and special screws
such as relieved screws.
The friction coefficient serving as basis is (screw fitting without lubrication)
µk=µG=0.14. The calculated values are valid for room temperature (T=20°C).
Table 37: Torques
Size MPitch PTightening torque Ma (Nm)
3.64.65.68.8, A2 u.
A4-80
1.60.40.050.050.170.28
20.450.10.110.350.6
2.50.450.210.230.731.23
30.50.5411.31.72
3.50.60.851.31.92.63.2
40.71.0222.54.45.1
50.822.758.710
613.54.6101518
81.258.411253643
101.51722497284
121.75293985125145
1424662135200235
1627195210310365
182.597130300430500
202.5138184425610710
222.5186250580830970
24323531573010501220
273350470110015501800
303.5475635145021002450
333.5645865200028003400
36410801440260037004300
39413301780340048005600
10.912.9
88
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
11Annex
+ALLG/7100
Test gas
Instrument air
Quality 2
Instrument air
Quality 1
Sample gas outlet
P
A
-KK0-KK6
-EB41
/112.1
sample gas line
-EB51
/112.4
sample gas line
ZrO2
Cell
Measuring
head
DRM
123
2
3
M20
M63
M20
M63
M20
> 2 bar
M40
M40
M20
Measuring Point 2 option
Zero gas
measuring Point 1
Zero gas
measuring Point 2 (option)
Control air
measuring Point 1
Back purging air
measuring Point 1
Control air
measuring Point 2 (option)
Back purging air
measuring Point 2 (option)
P
A
-KK1.1
P
R
B
-KK2.1
P
R
B
-KK3.1
P
A
-KK1.2
P
R
B
-KK2.2
P
R
B
-KK3.2
-RN2
-RN1
123
4
5
6
7
8
9
-KK0
/107.1
25psi
25psi
Bundle of cables and pipes
1
BK
2
BU
1
WH
3
/110.1
-EB42-WD
-EB42
/110.1
230V
Heated probe tube
Sample point
Stack
Flange size
See Device Pass
PT100
Power
N / WH
R / BU
ST / BK
25psi
25psi
Bundle of cables and pipes
1
BK
2
BU
1
WH
3
/111.1
-EB52-WD
-EB52
/111.1
230V
Heated probe tube
Sample point
Stack
Flange size
See Device Pass
PT100
Power
N / WH
R / BU
ST / BK
Optional. If present, differing
requirements of I-air to I-air
Quality1
Quality2:
Particle size max. 5 μm,
oil content max. 1 mg/m3,
pressure dew point max
3°C (26,6°F).
Cleanliness class 3 (ISO 8537)
Instrument air
Particle size max. 1 µm,
oil content max. 0.1 mg/m3,
pressure dew point max.
–40 °C (-40 °F)
Cleanliness class 2 (ISO 8537)
-RN3
BC
A
D
F
zero gas
3 bar
backpurge / control
5-7 bar
Auxiliary control air
5-7 bar
Heated probe tube
OPTION
Heated probe tube
OPTION
-BQ1
Instrument air green
6 +/- 0,2 bar
zero gas blue
3 bar
10
0
BAR
10
0
BAR
10
0
BAR
11.1Gas flow plan
ANNEX
11
Figure 129: Gas flow plan
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
89
11 ANNEX
11.2Modbus register
11.2.1Device status (Discrete Inputs [1xxxx], Function Code 02)
Table 38: Discrete Inputs
Discrete Input No.Date/Time
1000Device group status malfunction
1001Device group status maintenance request
1002Device group status maintenance
1003Device group status outside specification
1004Measured data are live/current (0 = held, 1 = alive)
1008Gas feed via measuring point_01
1009Gas feed via measuring point_02
1010Gas feed via measuring point_03
1011Gas feed via measuring point_04
1012Gas feed via measuring point_05
1013Gas feed via measuring point_06
1014Gas feed via measuring point_07
1015Gas feed via measuring point_08
1016Gas feed via measuring point_09
1017Gas feed via measuring point_10
1018Gas feed via measuring point_11
1019Gas feed via measuring point_12
1024Zeroize (Val./Adj.)
1025Span check (Val./Adj.)
1026Zeroize/span check (Val./Adj.), total signal
11.2.2Measured variables (Input Register [3xxxx], Function Code 04)
If not measuring: Register = “0”
Table 39: Input Register - measured variables
Input Register NoMeasured variable
1000 [Float]SO2
1002 [Float]CO2
1004 [Float]Ratio SO2/CO2
1006 [Float]H2O
1008 [Float]NO
1010 [Float]NO2
1012 [Float]NOx
1014 [Float]NH3
1016 [Float]CO
1018 [Float]CH4
1020 [Float]O2
1022 [Float]VOC
90
Active Sampling Point = 0 means maintenance or value not active.
1050 [DInt]
1052 [DInt]
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
≠ 0 measured value comes from measuring point x and is active.
Shown Sampling Point = 0 means maintenance.
≠ 0 measured value comes from measuring point x.
Subject to change without notice
Input Register NoMeasured variable
1
0
0
0,5
1
1,5
50
100
150
200
250
300
350
510152025303540455055606570
2
2,5
-0,5
-1
-1,5
-2
-2,5
2,5
3
3,5
4
4,5
5
2
1,5
1
0,5
0
1
2
4
3
Figure 130: Time diagram
Measured value course (exemplary)
1
Signal course (exemplary)
2
Feed via measuring point 2
3
Measured variable comes from measuring point 2 (in Hold mode)
4
Measured variable
Measured variable in Hold mode
Active Sampling Point (Input Reg. 1050)
Shown Sampling Point (Input Reg. 1052)
Gas feed via measuring point 1 (DI 1008)
Gas feed via measuring point 2 (DI 1009)
1054 [Float]Optional: [pH washing water]
1056 [Float]Optional: [PAH washing water (oil content)]
1058 [Float]Optional: [Clouding washing water]
ANNEX 11
11.2.3Device-internal monitoring values (Input Register [3xxxx], Function Code 04)
Table 40: Input Register - monitoring value
Input Register NoInternal variable
2000 [Float] Sample gas flow
2002 [Float] T_Cell (heating measuring cell)
2004 [Float] T_Optics (heating optics)
2006 [Float] T_Ext1 (external heating 1)
2008 [Float] T_Ext2 (external heating 2)
... ...
2018 [Float] T_Ext7 (external heating 7)
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
2020 [Float] T (ambient)
2022 [Float] T (LPMS01)
2024 [Float] T (LPMS02)
2026 [Float] T (LPMS03)
2028 [Float] p(0)
2030 [Float] p(2)
91
11 ANNEX
Input Register NoInternal variable
2032 [Float] p (ambient)
2034 [Float] Δp
3000 [Float] SO2 low [0..250]
3002 [Float] SO2 high [0..2000]
11.2.4Triggering the MARSIC 300 (Coils [0xxxx], Function Code 15, write multiple Coils)
Table 41: Activation - coils
Coil No.Action
2000 Select measuring point_01
2001 Select measuring point_02
2002 Select measuring point_03
2003 Select measuring point_04
2004 Select measuring point_05
2005 Select measuring point_06
2006 Select measuring point_07
2007 Select measuring point_08
2008 Select measuring point_09
2009 Select measuring point_10
2010 Select measuring point_11
2011 Select measuring point_12
2012 Trigger Standby
2013 Trigger BlowBack
2014 Trigger Zeroize
2015 Trigger Span Set (internal)
2016 Trigger zero validation
2017 Trigger span validation (internal)
2018 Trigger adjustment O2
Remarks:
The external selection is activated as soon as release signal '1' is set. If this is not
•
done, the measuring points are automatically triggered sequentially.
Standby: 1 = Standby, 0 = Measuring operation.
•
Zero and span set: “1” The adjustment/validation process starts for 5 seconds;
•
the trigger signal must then be disabled.
11.2.5VDI 4301 conform range (Holding Register [4xxxx], Function Code 03)
Table 42: VDI - Holding Register
Holding Register No.Measured variable
4000 [32 Bit Float, scaled according to VDI]SO2
4002 [32 Bit DInt]Device status
4004 [32 Bit Float, scaled according to VDI]CO2
4006 [32 Bit DInt]Device status
4008 [32 Bit Float, scaled according to VDI]Ratio SO2/CO2
4010 [32 Bit DInt]Device status
4012 [32 Bit Float, scaled according to VDI]H2O
4014 [32 Bit DInt]Device status
4016 [32 Bit Float, scaled according to VDI]NO
92
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
Holding Register No.Measured variable
4018 [32 Bit DInt]Device status
4020 [32 Bit Float, scaled according to VDI]NO2
4022 [32 Bit DInt]Device status
4024 [32 Bit Float, scaled according to VDI]NOx
4026 [32 Bit DInt]Device status
4028 [32 Bit Float, scaled according to VDI]NH3
4030 [32 Bit DInt]Device status
4032 [32 Bit Float, scaled according to VDI]CO
4034 [32 Bit DInt]Device status
4036 [32 Bit Float, scaled according to VDI]CH4
4038 [32 Bit DInt]Device status
4040 [32 Bit Float, scaled according to VDI]O2
4042 [32 Bit DInt]Device status
4044 [32 Bit Float, scaled according to VDI]VOC
4046 [32 Bit DInt]Device status
4064 [32 Bit Float, scaled according to VDI]Number of active measuring point
4066 [32 Bit DInt]Device status
4068 [32 Bit Float, scaled according to VDI]Number of triggered measuring point
4070 [32 Bit DInt]Device status
4072 [32 Bit Float, scaled according to VDI]pH washing water
4074 [32 Bit DInt]Device status
4076 [32 Bit Float, scaled according to VDI]PAH washing water (oil content)
4078 [32 Bit DInt]Device status
4080 [32 Bit Float, scaled according to VDI]Clouding washing water
4082 [32 Bit DInt]Device status
ANNEX 11
Remark: The device status is added for each measured value for conformity reasons.
Device status format. The individual status has the same significance as the device sta‐
tus of Discrete Inputs 1000 et seq.
Bit No.
0Malfunction
1Maintenance
2Maintenance request
3Outside the specification
4Test operation according to VDI
11.3Error messages and possible causes
Current pending messages are shown on the device display.
A combined list of messages is included in SOPAS ET (see "MARSIC300 Technical
Data").
NOTE
The following Table only includes those messages with classification "X" that are impor‐
tant for information.
Messages not included in the following Table have no further significance for operation.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
93
11 ANNEX
NOTE
Display of current device state data: Menu Diagnosis/System param.
C = Classification
F = Failure
M = Maintenance request
U = Uncertain
E = Extended
Trigger: System
Table 43: Error codes - System
CodeError textC DescriptionPossible clearance
S001Temperature too highFMeasuring cell temperature too highWhen T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check plug-in connector
on electronics unit.
When plug OK: Replace cell.
Optic head temperature too highWhen T < 356 °C:
When housing temperature >= 55 °C: Check
housing fan.
When housing temperature < 55 °C: Replace
electronics unit.
When T >= 356 °C: Check plug-in connector
on electronics unit.
When plug OK: Replace Analyzer module.
Temperature of heating for a subassembly
too high
LPMS01 (1/2 control) temperature too highWhen housing temperature < 55°C: Electron‐
LPMS02 (power electronics) temperature too
high
S002Temperature too lowFAfter x minutesCheck system documentation to clarify which
Check device documentation to clarify which
subassembly is affected.
When T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check subassembly plugin connector.
When plug OK: Replace subassembly
ics unit fan functioning?
Yes: Replace electronics unit.
No: Replace Analyzer module.
When housing temperature >= 55 °C: Check
housing fan.
When housing temperature < 55 °C: Replace
electronics unit.
When housing temperature >= 55 °C: Check
housing fan.
subassembly is affected (heating circuit 1 ..7).
Temperature displayed < -30 °C: Pt100
•
short circuit: Replace subassembly
For heated sample gas line: Connect
reserve Pt100
For analyzer: Replace Analyzer module
For optic head: Replace Analyzer mod‐
ule
Reset on circuit breaker under electron‐
•
ics unit possible: Check all cables
involved for damage (see "Circuit break‐
ers", page 86)
Check all plugs are plugged correctly.
Reset not possible: Replace subassem‐
•
bly affected
94
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
ANNEX 11
CodeError textC DescriptionPossible clearance
S004Flow too lowFFlow too lowSample gas flow and instrument air flow too
S005Pressure too highFPressure too highOnly instrument air pressure too high:
S006Pressure too lowFPressure too lowReplace pressure control module.
S008ChopperF24 V power supply malfunction: Replace elec‐
S009Motor filter wheel 1FFilter wheel motor does not detect reference
S010Motor filter wheel 2
S011Motor filter wheel 3
S012EmitterFEmitter voltage incorrect: Replace electronics
S0135 Volt powerFReplace electronics unit.
S01424 Volt powerFReplace electronics unit.
S015Detector signalFReplace Analyzer module.
S016Ref. energy too lowFIf further error messages are pending: Clear
S024No active componentFWhen all "active" checkmarks of all compo‐
S025Evaluation module failure FEvaluation module could not be started"Load backup": Menu Maintenance/Save
S026Evaluation mod. file errorFFiles for Evaluation module not created
position
nents are inactive
(espec, config, condition, measval)
low: Replace cell
Sample gas flow too low and instrument air
flow OK: Gas sampling system defective
Instrument air flow too low and sample gas
flow OK: Check all hose connections.
When all hose connections OK: Replace valve
block.
Check and set connected instrument air.
•
Set correct pressure on pressure
•
reducer unit.
Only sample gas pressure too high:
Set sample gas pressure within device
•
specification
Instrument air and sample gas pressure too
high:
Exhaust gas hose crimped/blocked
•
Counter-pressure in exhaust duct too
•
high
Check all hose connections
•
If this does not work:
Replace the pressure control module
•
Otherwise: Replace Analyzer module
•
tronics unit.
24 V power supply OK: Replace Analyzer mod‐
ule.
24 V power supply malfunction: Replace elec‐
tronics unit.
24 V power supply OK: Replace Analyzer mod‐
ule.
unit.
Emitter voltage OK and emitter power incor‐
rect: Replace emitter.
corresponding error.
If no further error messages are pending:
Replace cell.
Check in SOPAS ET.
parameters.
If error remains: "Load default setting" .
If error remains: Replace electronics unit.
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
95
11 ANNEX
CodeError textC DescriptionPossible clearance
S033Dev. zero point too highM Configured for measuring componentCheck zero gas for pressure and cleanness.
Perform maintenance on compressed air con‐
ditioning unit
Perform manual zero point adjustment (menu:
Adjustment/Zero point).
If deviation still too high: Replace Analyzer
module
S034Configuration I/O moduleM CONF (I/O) configuration error, module found
S035Ref. energy too lowMIf further error messages are pending: Clear
S036O2 sensor failureM Error bit O2 Error=1Please contact SICK Customer Service
S038Channel 1 errorM OV0 (I/O) signals that the desired current is
S039Channel 2 errorM
S040Flow too highM Flow too highTest pressure sensors with program "Mainte‐
S041Flow too lowM Flow too lowPlease contact SICK Customer Service
S042Controller IO or HC busyM BSY (I/O and HC3X) signals that the module
S043Emitter weakMReplace emitter.
S045dev. span gas adjust too
high
S046dev. Int. adjust too highM When F_Medium calculation is refused
S047dev. O2 adjust too highM When F_Medium calculation is refused
S048alarm O2 measure valueM
S049FlashCard not recognizedM FlashCard not recognizedPlease contact SICK Customer Service
S050adjust factor is ZeroM If one of factors F_Medium or F_Filter in
S057Energy too highU When at least one energy value > 5*Energie‐
S058Energy too lowU Energy too lowIf a further emitter error is pending: Replace
S072Module not foundE I/O (EXIST)Please contact SICK Customer Service
S089ZeroENew zero recordedExtinction value set to zero is displayed
S090AFE F_Filter has been recalculatedNew factor determined and measured value
S091Communication problemEInternal communication problemPlease contact SICK Customer Service
S092Adjustment canceledE Adjustment canceledRestart adjustment. If this does not work:
S093Dark measurementE
S094System startE
S095Adjust. zero canceledE
S096Backup doneE
does not correspond to nominal configura‐
tion
not reached on Analog module connection
(node y, module z).
microcontroller is still executing the previous
command
M When F_Medium calculation is refused
because outside tolerable range; Configured
for measuring component
because outside tolerable range; Configured
for measuring component
because outside tolerable range; Configured
for measuring component
range -0,000001 < x < 0,000001
MAX
Check I/O module, check configuration: IO
hardware plan
corresponding error.
If no further error messages are pending:
Replace cell.
Check I/O module, cable damage
nance/Maint Sys./Test pressure sensors".
If this does not work: Replace pressure control
module.
Perform adjustment with zero gas and span
gas.
If error remains: Replace Analyzer module.
Perform "Internal reference" adjustment.
If error remains: Replace Analyzer module.
Check span gas, check entry for span gas con‐
centration.
Check span gas, check entry for span gas con‐
centration
Please contact SICK Customer Service
emitter.
Otherwise replace Analyzer module
from adjustment displayed
Please contact SICK Customer Service
96
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
Subject to change without notice
CodeError textC DescriptionPossible clearance
S097Backup deniedE
S098AME
S112failure IO-NodeE
S113Check sum errorFBCK (I/O) shows the transfer made from
Master to Slave (controller) had an incorrect
checksum and the Slave did not save the
data.
S114Communication errorFCOM (I/O) communication error with an /I/O
module.
S115High/low voltageFPF0 (I/O) signals the internal voltage monitor‐
ing of the 5 V and 24 V supply voltages
exceed or underflow a range.
S116Output without currentFTOO (HC3X)
Check I/O module, cable damage
Please contact SICK Customer Service
This Table contains solution proposals that can only be processed by specially trained
personnel.
ANNEX 11
8017585/YXD9/V3-1/2017-05 | SICKT EC H NI C AL IN F OR M AT I ON | MARSIC300
Subject to change without notice
97
11 ANNEX
Trigger: Evaluation process
Table 44: Error codes - Evaluation process
CodeError textC Possible clearance
E001Operating system errorU Please contact SICK Customer Service
E002Temp. too low
E003Incorrect configuration
E004Incorrect configuration
E005Internal file error
E006Incorrect configuration
E007
...
E009
E010
...
E012
E013
...
E021
E022Resolution too high/low
E023Numeric error
E024Incorrect configuration
E025Internal file error
E026Numeric error
E027Incorrect configuration
E028Incorrect configuration
E029Unknown error
E030Operating system error
E031Operating system error
E032
...
E034
E035Numeric error
E036Syntax error
E037Error during processing
E038Extinction too high
E039Internal file error
E040Internal file error
E097Evaluation uncertain
E098Medium temp. too high/low
E099Medium pressure too high/low
E100Medium flow too high/low
E101Measured value too high/low
E102Evaluation uncertain
E103Evaluation uncertain
Internal file error
Incorrect configuration
Internal file error
Internal file error
98
T EC H NI C AL IN F OR M AT I ON | MARSIC3008017585/YXD9/V3-1/2017-05 | SICK
check
M063M Test failedRepeat test and observe messages dis‐
M064Pressure not
M065Air valve not closed
M066Leakage > Limit
M067X Deviation="xx"Serves as information
C MessageClearance
Clearance see above: S001 and S002
Clearance see above: S001 and S002
No action required
by user
M Flow alarm (meas‐
uring)
Clearance see above: S004
Next measuring point activation after mes‐
sage acknowledged
C MessageClearance
Clearance see above: S004
uring)
Next measuring point activation after mes‐
sage acknowledged
No action necessary
started
X Test passed
played;
reached
Check seating of connections, disconnect
heating hose from cell and close sample
gas inlet off with dummy plug (from leak
tightness check set)
If leaky: Exchange pressure control module
and cell otherwise exchange gas sampling
system