1) Can run on a computer with an operating system starting at Windows 2000
2) For installation and use, see the operating instructions "Visualization tool SICK LaserView Customer Edition“ (Part No.
8013787, English version) on the provided CD, "Manuals & Software LD-MRS" The visualization tool "SICK LaserView Customer Edition" will hencworth be called „LaserView“.
1)2)
FunctionVersion
Firmware (DSP)
Firmware (FPGA)
Firmware (DSP)
Firmware (FPGA)
Firmware (DSP)
Firmware (FPGA)
Firmware (DSP)
Firmware (FPGA)
as of V. 2.0.x
as of V. 9.0.x
as of V. 2.0.x
as of V. 9.0.x
as of V. 2.0.x
as of V. 9.0.x
as of V.3.0.x
as of V.9.6.x
Configuration software for field monitoring as of V. 2.22
Configuration software and visualization tool for mea-
as of V. 1.5.4
suring mode
TM
Copyright
Copyright 2009 - 2017
SICK AG Waldkirch
Auto Ident, Reute plant
Nimburger Strasse 11
79276 Reute
Germany
Trademark
Windows 2000
TM
, XPTM, VistaTM, Windows 7TM Windows 8TM, Windows 10
trademarks or trademarks of the Microsoft Corporation in the USA and other countries.
®
Adobe
Reader® is a registered trademark of Adobe Systems Incorporated.
Download via Internet: http://get.adobe.com/reader/
Version of these operating instructions
The latest version of these operating instructions can be obtained as PDF at www.sick.com.
Commissioning, configurationQualified personnel, e.g. technicians, engineers
Tab. 1-1: Target group
Qualified personnel, e.g. service technicians, factory
electricians
1.3Depth of information
This document contains all information required for mounting, electrical installation and
commissioning of the LD-MRS with factory settings.
All tasks are described step by step.
The configuration of the LD-MRS for user-specific situations in measuring mode as well as
measurement queries for displaying the scan data are done via the provided visualization
tool SICK LaserView Customer Edition or with messages. The LaserView software is also
used to configure the object data. Further Information is available in the corresponding document Article no. 8013787.
To configure field monitoring, use the SOPAS-ET configuration software. SOPAS-ET can also
display scanning data from measuring mode.
Further information about laser measurement technology is available from SICK AG, Division Auto Ident, and on the internet at www.sick.com.
1.4Symbology used
Some information in this documentation are highlighted to facilitate quick access to these
information:
Cross-referenceText in italics indicates a reference to more detailed information.
For your safety
Note!
A note indicates potential hazards that could involve damage or degradation of the functionality of the LD-MRS.
Warning!
A warning indicates an actual or potential hazard for the physical integrity of the user. They
are designed to help to prevent accidents.
The safety symbol beside the warning indicates the nature of the risk of accident, e. g. due
to electricity. The incremental warning category (CAUTION, WARNING, DANGER) indicates
the severity of the hazard.
Always read carefully and follow the warning notices.
ImportantThis important note informs about special issues.
ExplanationAn explanation provides background knowledge about technical relations.
RecommendationRecommendations are designed to assist in the decision-making process with respect to a
Basic settingsMarks a section listing the values of the factory settings.
AUDRATEThis typeface indicates a term in the configuration software SOPAS-ET.
B
Instructions for taking action are shown by an arrow. This symbol indicates an inst
2For your safety
certain function or a technical measure.
This symbol identifies a section that describes the operating steps with the SOPAS-ET configuration software.
This symbol refers to supplementary technical documentation.
ruction
hat only consists of one step, or a warning note with several steps that do not require any
t
particular order.
Multiple step instructions in a required sequence are identified by a sequential numbers.
8012948/ZMQ0/2017-06-09
This chapter deals with your own safety.
Read this chapter carefully before using the LD-MRS.
2.1Safety standards
The LD-MRS has been designed and manufactured under consideration of a risk analysis
and careful selection of the pertinent harmonized standards as well as other technical specifications. Hence, the LD-MRS matches state-of-the-art technology.
In design and production, established technological rules have been applied and observed.
Development and production in SICK AG ensure this quality standard by a certified quality
management system according to EN ISO 9001:2008.
If the user adheres to all safety notes in this operating instructions and uses the device as
intended, he is sufficiently protected.
2.2Authorized personnel
The LD-MRS must be mounted and operated by qualified personnel to ensure its proper and
safe function.
ImportantOnly trained and authorized service personnel of SICK AG may repair the LD-MRS.
The different tasks require the following qualification:
ActivitiesQualification
Mounting, maintenance– practical technical basic training
– knowledge of the established safety rules at work
Electrical installation,
device replacement
Commissioning, configuration
Processing the measured
values in EDP systems
Tab. 2-1: Required qualification for commissioning the LD-MRS
– practical electrical training
– knowledge of the established electric safety rules
– basic knowledge of Windows
– basic knowledge of data transmission
– Basic knowledge of how to establish and set up (address) Ethernet
connections when connecting the LD-MRS to the Ethernet
– Basic knowledge of how to establish and set up a CAN network
when using field monitoring
– Programming skills for the application to be created in measuring
The LD-MRS allows two applications with its radial field of view:
•The detection of objects around the location or installation site while continuously out-
putting measurements upon request, or
•Field monitoring of freely definable areas while signaling field violations via external
itch outputs
sw
The device is designed for application in the industrial sector, in particular outside.
ImportantEvery other use as well as modifications to the device, including those within the course of
mounting and electrical installation, will render void any warranty claim towards SICK AG.
Only operate the LD-MRS in the permitted ambient temperature range, see Chapter 9
Technical specifications, Page 94.
The user of the LD-MRS has to ensure that
•the device is only used in compliance with the listed specifications and environment
nditions, see Chapter 9.1 Data sheets, Page 94,
co
•country-specific standards and regulations are adhered to, depending on the kind of
ion,
operat
•the device is only used in proper, functional state,
•safety and warning labels attached to the device are not removed or covered and re-
main readable,
•the operating instructions are available on site in a readable form and that the autho-
rized personnel had read the operating instructions and is thus sufficiently qualified.
ImportantLoss of warranty!
The housing of the laser measurement sensor LD-MRS is closed with a seal. Breaking the
seal and opening the device causes a loss of warranty claims towards SICK AG. The housing
may only be opened by authorized service personnel of SICK AG.
Read the general safety notes carefully and strictly observe them in when working at or
ith the LD-MRS. Also observe the warning notes printed before
w
individual chapters
Safety notes
Observe the following points in order to ensure safe use as intended of the LD-MRS:
•The user must ensure that every person working on or with the LD-MRS has read and
understood these operating instructions.
•Official and legal requirements must be adhered to when operating the LD-MRS.
•For installation and usage of the LD-MRS as well as for commissioning and regular tech-
nical inspection, national/international legal requirements apply, in particular
– the accident prevention regulations/safety rules
– other pertinent safety rules
•Adhere to the following safety notes in order to prevent dangers for persons and/or
property:
– T
he operator must ensure by suitable instructions and inspections that the window
of the LD-MRS is always clean.
– Additionally, the local safety and accident prevention regulations apply for operating
the LD-MRS.
– A defect of the control functions can cause danger for human life or property dam-
age at the LD-MRS.
•Data integrity
In its products, SICK AG uses information technology, such as IO-Link or standard
te
chnology. The focus is on the availability of the products and their features.
nsistent assumption of SICK AG that the user will maintain the integrity and confiden-
co
tiality of data and rights that are affected by the use of the products.
The user must observe appropriate safety measures such as network separation, firewalls, viral protection and patch management according to the situation.
of this document.
the instructions in the
IP
It is the
2.4.1Range of application
Not permitted for personal protection!
The LD-MRS is NOT a device for ensuring personal protection as defined by applicable safety standards for machines!
This chapter provides information on design, special features and properties of the LD-MRS.
For information on assembly, electrical installation and startup of the LD-MRS, consul
the
section before starting.
3.1Design of the device
The laser measurement sensor LD-MRS consists of a laser measurement system and a mirror construction. The components - laser measurement system and mirror construction - are
mounted in a rugged housing.
When looking from the front at the LD-MRS, the right side of the device contains three connections ("Ethernet", "Data interfaces/synchronization" and "Power"), the left side contains
a ventilation unit.
At the rear side of the device, four protruding lugs have a hole each to attach the device at
the point of installation.
The LD-MRS delivery includes the following components:
QuantityComponentComment
1Laser measurement sensor LD-MRSType depending on order
1Notes on Device with information and elec-
trical circuit diagram for getting started
1CD-ROM "Manuals & Software LD-MRS"
Tab. 3-1: Delivery of LD-MRS
The product information for the LD-MRS laser measurement sensor (Part No. 8012945)
contains all of the ordering information for the product and accessories.
Chapter 10.3 Ordering information, Page 105 provides an overview of consumables for
care and maintenance of the LD-MRS.
3.2.1Contents of the CD, "Manuals & Software LD-MRS" (Part No. 2050264)
ComponentComment
SoftwareSICK LaserView Customer EditionVisualization tool for measurements; can run on
standard computers (Windows
Software SOPAS-ET Configuration software for field monitoring
including online help (Java), can run on standard
computers (Windows
Operating instructions for the "laser measurement sensor LD-MRS"
Operating instructions for the
"SICK LaserView Customer Edition visualization tool"
Ethernet Data ProtocolPDF in German and English
CAN Data ProtocolPDF in German and English
Tab. 3-2: Contents of the CD-ROM "Manuals & Software LD-MRS"
PDF version in German and English
PDF version in German and English
Included in the LD-MRS packaging
TM
and Linux)
TM
)
ImportantThe publications and programs on the CD-ROM are also available for download at
www.sick.com.
Download address for PDF visualization software in the Internet:
http://get.adobe.com/reader/
For general system requirements refer to the data sheet for the LD-MRS (see Chapter 9
Technical specifications, Page 94).
The requirements for Mounting, Electrical installation and Commissioning and configura-
tion are summarized in the respective chapters.
3.3.1Product features and functions (overview)
Performance featureCharacteristic
General advantages• reliable detection of objects in the field of view of the LD-MRS
• high scanning range and compact housing
• simultaneous measurement on 8 planes
• LD-MRS HD: expanded penetration through dust by special receiver unit and extended
dust filter software
Safety and comfort for the user• Rugged, compact aluminum housing also available as lightweight housing
• laser class 1, laser switches off in case of faults
• enclosure rating IP 69K in mounted condition
• upgradable by firmware update via data interface Ethernet
•low power consumption
• wide range of supply voltage
Simple operation / parametrization• Configuration of measuring mode using the SICK LaserView Customer Edition visual-
ization tool with Windows and Linux
• And configuration of field monitoring using the SOPAS-ET the configuration software
with Windows
• alternatively using commands in messages
Operating states• wait mode: Configuration of the LD-MRS
• Operating mode measurement: Scanning the field of view
• Operating mode "Object tracking". Scanning the field of view and tracking of up to 64
objects in field of view.
• Field monitoring mode: scanning the field of view and monitoring specific fields for
changes (field intrusion)
Output of results• Measurements: via the Ethernet data interface
• Object tracking: pre-processed object data allows for object tracking of up to 64 objects simultaneously, as well as the output of their dynamic properties (e.g.. speed, direction, size, etc.)
• Field violations: via a maximum of 10 externals switch outputs with the help of two optional CAN modules (accessory) with eight configurable I/O ports each, optional arming of specific, freely-definable evaluation instances through 2 external switching
inputs
Its applications are wide-ranging. In particular, the following can be stated:
•Container loading / handling
•Traffic / Transport
•Robots / UAV
•Collision protection
•Autonomous industrial vehicles
•Security monitoring
•Topography and urban surveying
In measuring mode, the measurements can be processed externally to identify and analyse
objects.
In field monitoring, external switch outputs signal changes in the monitored areas (fields).
Evaluation instances can be defined with an application-specific evaluation strategy per
evaluation field and assigned to switch outputs and combined. Combining two switch outputs activates a specific evaluation instance only when needed.
The object tracking provides pre-processed object data. Up to 64 objects can be processed
at the same time. Dynamic parameters such as direction, speed and even acceleration are
available for every object. This data is available when vehicle data (ego motion data) is provided.
3.4Operating principle of the device
The LD-MRS serves for contact-free and directional detection of the sensor surroundings, or
rather of the objects located within the radial field of view.
The object detection is done with laser beams that the LD-MRS emits in four stacked
planes. The device measures the distance and the direction (the angle to the LD-MRS) of
the object. From the measured data, the LD-MRS calculates the position of the object in the
sensor co-ordinate system.
The resulting profiles of the different planes are called scan, see Chapter 3.5 Measuring
mode, Page 36.
The LD-MRS issues the measured data in reference to its sensor co-ordinates. The data contain information about the measuring plane, distance, echo number, echo pulse width and
class. It also contains angle information (horizontal and vertical).
Due to the detailed presentation and thus the extensive amount of data, the LD-MRS issues
the measured data via the Ethernet interface only.
The factory defaults of the LD-MRS are designed to facilitate immediate commissioning of
the LD-MRS. The user can modify some parameters to optimize the LD-MRS in regard to the
application at hand.
8012948/ZMQ0/2017-06-09
3.4.1Measuring process and measuring properties
The LD-MRS is a measurement instrument basing on the Time-of-Flight (ToF) technology, i.
e.
the LD-MRS uses laser beams to detect distance and angle of objects.
It radially scans the surroundings with laser beams deflected by a rotating mirror, receives
the echos with a photo diode receiver and outputs the data processed based on a runtime
calculation via the Ethernet interface.
By the permanent rotation of the mirror in connection with the laser beam, it is possible to
build a complete profile of the surroundings within the field of view of the LD-MRS. The scan
data of the LD-MRS consist of values for distance, angle, and echo pulse width.
The measurement properties base on
•time-of-flight measurement,
•multi-echo capability,
•multi-layer technology,
•the scanning range and the relation of angle to range,
•the angular resolution and the scanning frequency
The multi-layer technology of the LD-MRS allows for a pitch angle compensation by means
of four scan planes with different vertical angles, e.g. if the device is mounted to a vehicle.
Thus, the LD-MRS detects the object reliably even when accelerating or braking the vehicle.
Fig. 3-5: Principle of the scan planes, color-coded
Caption:
1vertical aperture angle
2one of the four scan planes
3LD-MRS
The photo diode receiver of the LD-MRS consists of four independent receivers arranged in
a line.
These four receivers enable the implementation of the multi-layer technology.
One receiver is assigned to each plane, thus dividing the vertical aperture angle into four
scan planes.
These four scan planes are scanned interlaced. This means that the combination of two
planes is always scanned simultaneously (first e. g. the yellow and the green plane, then the
blue and the red plane), see Chapter 3.4.8 Angular resolution and scanning frequency,
Page 29.
PlaneEcho 1Echo 2Echo 3
yellowscan plane 4scan plane 4scan plane 4
greenscan plane 3scan plane 3scan plane 3
bluescan plane 2scan plane 2scan plane 2
redscan plane 1scan plane 1scan plane 1
Tab. 3-5: Naming conventions (colour code see Fig. 3-5)
Colour hues visualize the planes and colour saturation of the echoes. Tab. 3-5 illustrates the
naming convention for the planes and their predefined colors that are used for the visualisation. The colour saturation in the visualization decreases from echo to echo.
Example for a case with three echoes, see Fig. 3-3
If a laser beam hits a glass pane, for example, a part of the light is reflected and triggers a
measurement (echo 1).
Most of the light passes the window pane and might hit a rain drop which then again reflects
a part of the light (echo 2).
The remaining light is then reflected by an object, which then results in the third measured
value (echo 3).
3.4.5Scanning range and relation of angle to range
Scanning range
1
Fig. 3-6: Scanning range
2
3
Caption:
1central working range (green)
2lateral scanning range (light green)
3LD-MRS
The LD-MRS has been designed with a central scanning range of 85° for four scan planes.
The scanning range can be extended between +35° and +50° or –50° and –60° to a total
range of 110°. The lateral scanning ranges only provide two instead of four scan planes.
The following table Fig. 3-7 of the LD-MRS and LD-MRS HAD shows the sensing range in relation to the remission.
Fig. 3-8: Sensing range in relation to remission
3.4.7Filter
Noise filter
The noise filter reduces the effects of atmospheric interference such as rain, spray or dust
with a clutter filter.This filter classifies the scanning data and distinguishes normal measurements from noise.
The filter affects the field evaluation and the output measurements.
The LD-MRS has for scan planes (layers).Every second measurement within a layer takes
place at close range up to approximately 15 m (49.2 ft) with reduced sensitivity. For greater
distances, the measurements are taken at normal sensitivity. In addition, the measurements are evaluated. Criteria such as proximity relationships between measurements with
normal sensitivity as well as the measuring characteristic of the sensor are taken into account.
When the filter is set, measurements classified as clutter are not transmitted to the SOPASET interface and are therefore not included in the field evaluation.
ImportantThe close range sensitivity is reduced by the clutter filter. Consequently, targets with a lower
remission are more difficult for the LD-MRS to identify.
Only the last echo
This filter causes only the last echo of a laser pulse to be measured. If for example part of
the energy is reflected as a first or second echo by a pane of glass or raindrop, only the last
echo of the actual object is measured. The filter affects the field evaluation and the output
measurements.
Fig. 3-9: Example of three echoes of a laser pulse
Caption:
1.Echo of a pane of glass
2.Echo of a raindrop
3.Threshold voltage 1
4.Threshold voltage 2
5.Echo of an object
3.4.8Angular resolution and scanning frequency
The LD-MRS can be operated with three different scanning frequencies (12.5 Hz, 25 Hz and
50 Hz), which allow four different settings of the angular resolution.
The operator can configure these parameters.
Angular resolution
Fig. 3-10: Angular resolution
Two scan planes each are measured and analysed simultaneously. If an angular resolution
of 0.125° is specified for a certain range, then 0.125° is the angular step between two
scan planes (e. g. red-blue) and their partners (e. g. yellow-green). The angle for the next
measurement on the same plane is twice as high, in this example 0.25°.
Fig. 3-11: Example: Angular resolution of 0.25° at distance d = 25 m (82 ft)
Caption:
Y_G1distance between measured points in one measurement plane (here: Y_G1 = 0.1833 m)
Y_G2distance between measured points between two laser pulses (here: Y_G2 = 0.0742 m)
Y_Swidth of measured point (here: Y_S = 0.0349 m)
X_Layer height of individual measured point (here: X_Layer = 0.3491 m)
The following tables illustrate some values as examples:
Distance d [m]Y_S [m]Y_G1 [m]Y_G2 [m]X_Layer [m]
100.01400.02970.00790.1396
250.03490.07420.01960.3491
500.06980.14840.03930.6981
1000.13960.29670.07851.3963
Tab. 3-6: Angular resolution 0.125°
Distance d [m]Y_S [m]Y_G1 [m]Y_G2 [m]X_Layer [m]
100.01400.07330.02970.1396
250.03490.18330.07420.3491
500.06980.36650.14840.6981
1000.13960.73300.29671.3963
Tab. 3-7: Angular resolution 0.25°
Distance d [m]Y_S [m]Y_G1 [m]Y_G2 [m]X_Layer [m]
100.01400.16060.07330.1396
250.03490.40140,018330.3491
500.06980.80290,36650.6981
1000.13961.60570,73301.3963
Tab. 3-8: Angular resolution 0.5°
In measuring mode for a scanning frequency of 12.5 Hz, you can choose between an angle
resolution that is constant or changes according to sector (settable with the SICK LaserView
Customer Edition). This function is not possible for field monitoring.
The scanning frequencies of 25 Hz and 50 Hz only allow for constant angular resolution.
3.4.9Impact of the object surface on the measurement
The signal received from a perfectly diffuse reflecting white surface corresponds to the definition of a remission of 100 %. As a result of this definition, the remissions for surfaces that
reflect the light bundled (mirrored surfaces, reflectors), are more than 100 %.
Fig. 3-16: Reflection of the laser beam at the surface of an object
Most surfaces reflect the laser beam diffusely in all directions.
The reflection of the laser beam will vary as a function of the surface structure and colour.
Light surfaces reflect the laser beam better than dark surfaces and can be detected by the
LD-MRS over larger distances. Brilliant white plaster reflects approx. 100% of the incident
light, black foam rubber approx. 2.4 %. On very rough surfaces, part of the energy is lost due
to shading. This reduces the scanning range of the LD-MRS.
Fig. 3-17: Reflection angle
The reflection angle is the same as the angle of incidence. If the laser beam is incident perpendicularly on a surface, the energy is optimally reflected, see Fig. 3-17. If the beam is incident at an angle, a corresponding energy and scanning range loss is incurred.
Fig. 3-18: Degree of reflection
If the reflected energy returned is over 100% (basis: Kodak standard) the incident beam is
not reflected diffusely in all directions, but is reflected in a specific direction. As a result a
large portion of the energy emitted can be received by the laser distance measurement device. Plastic reflectors ("cats’ eyes”), reflective tape and triple prisms have these properties.
Fig. 3-19: Mirror surfaces
At mirror surfaces the laser beam is almost entirely deflected, see Fig. 3-19.
Instead of the surface of the mirror, it is possible that the object on which the deflected laser
beam is incident may be detected.
Fig. 3-20: Object smaller than diameter of the laser beam
Objects that are smaller than the diameter of the laser beam cannot reflect all the energy
of the laser light, see Fig. 3-20. The energy in the portion of the laser light that is not reflected is lost. This means that the scanning range is less than would be possible theoretically
based on the surface of the object.
The LD-MRS has an Ethernet interface for configuration and data transmission for measuring mode.
The Ethernet interface is designed as TCP/IP interface; it supports full duplex.
By factory default, the Ethernet interface is configured as follows:
•IP address: 192.168.0.1
•Subnet mask: 255.255.255.0
•TCP port: 12002
Depending on the configuration of the LD-MRS, the Ethernet interface provides different
data types:
•Scan data (see Chapter 3.5 Measuring mode, Page 36)
– LD-MRS: data type 0x2202
•Warnings and fault messages
– LD-MRS: data type 0x2030
•Commands and responses
– data types 0x2010 and 0x2020
For a description of the electrical connections and cables of the data interfaces, refer to
Chapter 5.3 Electrical connections and cables, Page 66.
See Ethernet Data Protocol for further information.
3.5Measuring mode
3.5.1Scan data
Scanning the sensor surroundings within the scanning range of the LD-MRS is called the
scan process.
The total set of measured data (= scan data) of a scan process, consisting of individual scan
points, are called scan.
Upon request, the LD-MRS issues the scan data via the Ethernet interface.
In the scan process, the transmitted laser pulses are reflected by objects within the measurement range. These echo pulses are received and analysed by the LD-MRS. Every detected echo pulse is represented by a scan point with the following main properties:
•position of the point
•width of the echo pulse
•scan plane and echo number
The LD-MRS generates a two-dimensional profile of the surroundings, with additional height
information (three-dimensional information) resulting from the multi-layer technology, see
Fig. 3-4, page 24.
Typical presentation of the scan data is a bird's view, i.e. the view from above onto the measurement plane.
For the measured value output, the LD-MRS supplies measured values to the Ethernet interface.
It is prerequisite for this data output that the LD-MRS is in the measurement mode. Normally, the LD-MRS starts measurement mode via the automatic power system. If the LD-MRS
stops during measured value output, there are two ways to start the measurement mode:
•Start using the SICK LaserView Customer Edition visualization tool
•Start using a telegramm, see Telegram Listing Ethernet Data Protocol, art. no.
4492.
801
3.5.3Data communication using messages
In measuring mode, the LD-MRS communicates with a connected computer using messages via the Ethernet interface.
The following functions can be run using messages:
•request for measured values by the PC and subsequent output of the measured values
by the LD-MRS to the PC
•Setting the parameter using a computer to configure the LD-MRS
•parameters and status log querying by the PC
A detailed list of all possible telegrams is included the Telegram Listing Ethernet Data Protocol, art. no. 8014492.
By means of object tracking, the LD-MRS outputs preprocessed data that can then be used
for advanced applications. No additional hardware or software is required for object tracking.
The LD-MRS outputs the following dynamic object parameters in relation to the data for its
own movement:
•Position of the object
•Speed of the object
•Size (length and width) of the object
•Direction of movement of the object
•Age (scans, time) of the object
The areas of application include:
•Navigation
•Collision prevention
•Monitoring
Object tracking allows up to 128 objects to be tracked at the same time. The object data is
then processed in real time. This real-time data is output via an Ethernet interface.
When the "Show Contour" option is activated, the LD-MRS uses the raw data to generate an
object contour. The scan data is divided to form groups consisting of related scan points.
This division is dependent on the vehicle's own movement.
Fig. 3-23: Bounding box
When the "Show Bounding Box" option is activated, the LD-MRS uses the raw data to generate a rectangular border around the object contour.
The vehicle's own movement and the historical scan data will affect the object contour.
When the "Show Object Boxes" and "Show Direction" options are activated, the LD-MRS
uses the raw data to generate a rectangular box around the object and also outputs the following data:
•Position of the object
•Speed of the object
•Size (length and width) of the object
•Direction of movement of the object
•Age (scans, time) of the object
•Contour of the object
Ego motion data:
It is possible to transfer information on the proper motion data of the vehicle to the LD-MRS.
The scanner mounted on the vehicle passes on the data to the LD-MRS via Ethernet or CAN
bus. This data is taken into account during object tracking and when calculating speed and
direction properties of the detected objects. Without this information, the scanner assumes
that its mounting/vehicle is not moving
The "Suppressor" filter option can be activated during object tracking to improve the measurement results under difficult conditions. This filter reduces the sensitivity of the four lower levels to ensure more accurate measurement.
3.6.3Reduced sensitivity
If the sensitivity is reduced, the "NoiseControl" sensitivity settingwill be deactivated because the reduction in sensitivity also results in excellent ambient light immunity.
If the "Lower 4 layers reduced range" option has been selected under the "LayerRangeReduction" setting, the measurements in the near range up to 15m will not be used during
object tracking.
Depending on the application, reducing the sensitivity can filter out erroneous measurements and prevent them from being taken in the first place. For use on a vehicle in a dusty
environment, it is possible to reduce the sensitivity of the lower layers yet retain the full
scanning range of the upper layers to make it possible to detect obstacles or objects that
are further away.
3Reduced sensitivityReduced sensitivityProfile measurement with
sun in background
Tab. 3-9: Setting the layer sensitivity
If the LD-MRS has been mounted upside down, make sure that the "Upper layers" and
"Lower layers" columns are swapped over and that the "Upside down active" option has
been activated.
The sensitivity is generally reduced to 60% of the original scanning range.
Fig. 3-27: Example for measurement result with „Clutter“
For every echo that reflects off an object, the signal amplitude is assessed twice – with a
lower and higher threshold. If both thresholds are exceeded, it can be assumed that the
measurement target is solid and fixed. If only the lower threshold is exceeded, the target is
scattered:
For example, dust.
Within the output of measured values, the measured values are marked as "Clutter" and are
not included in the object calculation
This option should only be used in cases when it is known that there is a chance of disruptive measurements occurring in the measuring situation. If not, this may result in inaccurate measurement results.
The filter can only be used with the full range of layers ("Layer Reduction" > "Full Range").
In extreme cases of ambient light (e.g. sunlight), erroneous measurements may be taken,
which can be evaluated and filtered by performing a plausibility check on a downstream external computer.
The "Noise control active" function activates an automatic sensitivity setting that monitors
erroneous measurements at >100m and uses them to update the sensitivity of the receivers. The automatic sensitivity adjustment means that the scanning range will be between
the maximum and the threshold for reduced sensitivity.
If the LD-MRS has been mounted upside down, it is necessary to activate the "Upside down
active" option. The installation position is selected from the Laserview/Telegram listing to
ensure that the scan data output, coordinates and object data can be output correctly.
If two LD-MRS 8Ls are used, it is possible to extend the scanning range by mounting the
LD-MRS 8L on the left-hand side, upside down in the direction of travel.
With the aid of integrated field monitoring, the LD-MRS evaluates up to 16 evaluation fields
within its scan range. It signals field violations (events) through a maximum of 10 logical outputs. The logical outputs can be assigned via the CAN bus to external, physical ports (that
can each be configured as inputs or outputs) of a maximum of two CAN modules (optional
accessory). A CAN module has 8 I/O ports.
Systems for collision protection, object protection or access monitoring can be created using field monitoring.
8012948/ZMQ0/2017-06-09
Fig. 3-30: Principle of field monitoring
The LD-MRS uses up to 16 evaluation cases to adapt to the evaluated situation. For each
evaluation instance, and, the following are selected: one of 16 configurable evaluation
fields, one evaluation strategy per field, a logical output and in certain circumstances a combination of the logical inputs that activate the evaluation instance. For each logical output,
a link can be selected that determines the result of the output if more than one evaluation
instance affects the output.
In the example inFig. 3-30, evaluation field 1 is used in evaluation instance 1, and evaluation field 2 is used in evaluation instance 2. Both evaluation instances affect logical output
OUT1. If the results of the evaluation instances are AND-linked, the output only switches
when both evaluation instances report results.
An evaluation instance determines which evaluation field is evaluated in which way (evaluation strategy), and which logical output it affects. Up to 16 evaluation instances can be configured, and all configured and evaluation instances are simultaneously active.
The following is specified with SOPAS-ET for each evaluation instance:
•Logical inputs that (specifically) activate the evaluation instance
•Associated evaluation field to be monitored
•Evaluation strategy per evaluation field
•Logical output that is affected by the evaluation instance
•Response time of the logical output
•Manipulation protection per evaluation field
•Reset behavior of an activated output
Logical inputs
If the evaluation instance is not to be continuously active, an input combination can be configured that activates the evaluation instance.
Input 1Input 2Evaluation instance
Active highActive highEvaluation instance 1
Active highActive lowEvaluation instance 2
Active lowActive highEvaluation instance 3
Active lowActive lowEvaluation instance 4
Tab. 3-10: Example of a combination of inputs
ImportantA combination of inputs can also be defined for several evaluation instances. In this case
(for example), two evaluation instances are active simultaneously.
Evaluation strategy for evaluation field
The LD-MRS offers three different evaluation strategies:
•Pixel evaluation
The LD-MRS evaluates the entire area of the field; each beam is
an
object enters the field, this event is transmitted to the appropriate output
•Hiding
The LD-MRS evaluates the entire area of the field. Objects up to a certain size can
hidden.
•Contour
An object is only identified when it is larger than the configured hidden size.
used for evaluation. If
.
be
The LD-MRS evaluates the presence of a contour that is supposed to be continuous
and
completely in the evaluation field. This allows the LD-MRS to determine that for example a door is opening outward, or that the position of the LD-MRS is changing. In addition, the entrance of a vertical evaluation field or the deflection of the laser beam with
a mirror can be identified.
By using the hiding function, the absence of a part the contour can be hidden up to
certain si
A response time is defined for the above evaluation strategies. In order for the LD-MRS to
identify an object, it must be detected at one location for the duration of the response time
when the "pixel evaluation" and "hiding" evaluation strategies are used. With the "contour"
evaluation strategy, the contour violation must be detected for at least the duration of the
response time at one location.
Manipulation projection
When evaluation fields are defined at a distance from the LD-MRS, a small, close-range object (between the LD-MRS in the field) may cast a large shadow in the observed fields. Without manipulation protection, shadowed objects are not detected in the field. When
manipulation protection is activated, the LD-MRS signals a field violation when the shadowed area is larger than the configured minimum object size, and the configured "expanded
response time" has passed.
ImportantThe "Negation of the results" function does not correspond to the setting option "active
8012948/ZMQ0/2017-06-09
Fig. 3-31: Manipulation protection against shadow
Evaluation field
A field is selected from the configuration evaluation fields for the evaluation instance. Is
shape must corresponds to the evaluation strategy (see Chapter 3.8.2 Evaluation fields,
Page 50).
Logical output
One of the logical outputs is selected for the evaluation instance. If several evaluation instances act on one logical output, specify how the results of the evaluation instances are
linked (see Chapter 3.8.3 Linking the evaluation instances to the logical output, Page 50).
Negation of the results
By negating the results, the field evaluation is conversely transmitted to the logical output.
The output that is used is hence switched (for example) when the evaluation field is free, or
when the contour has not been violated.
high/active low" (level) for the external, physical outputs (see Chapter 3.8.3 Linking the
evaluation instances to the logical output, Page 50).
Up to 16 evaluation fields can be configured using field monitoring. There is almost no restriction to the configuration of the size and shape of these the 16 evaluation fields.
1
2
3
4
5
6
Fig. 3-32: Examples of evaluation field shapes
Caption:
1Evaluation field for contour monitoring
2Rectangular evaluation field at a distance from the LD-MRS
3Polygonal evaluation field
4Rotated evaluation field
5LD-MRS scanning range
6LD-MRS
The evaluation fields can be marked directly with the SOPAS-ET corresponding to the requirements of the application. Evaluation fields can have the following shapes and positions:
•Polygon
•Rectangle
•Extending up to the LD-MRS
•At a distance from the LD-MRS
When the range to be monitored changes, the LD-MRS can be reconfigured using software
without additional installation.
3.8.3Linking the evaluation instances to the logical output
If several evaluation instances act on one logical output, specify how the results of the evaluation instances are linked.The results can be AND-linked or OR-linked.
The external, physical outputs are configured by default as "active high".
Resetting a logical output
Logical outputs are reset immediately by default. Howev er, a del ay up to 10 second s can be
set (for example to activate a horn, or to send the output signal to a PLC).
Alternately, the output can be reset by and input. The output is only reset when the assigned
input assumes the configured status.
The LD-MRS features two sender elements and four receiver elements. The elements
are mounted such that they each cover a particular viewing angle and thus measure different approach angles.
The rotating mirror that deflects the lasers is mounted at a slight angle to the axis of rotation. This design means that the front and back of the mirror each have a different field of
vision, thus extending the vertical range covered.
The lower scan field is generated by the back of the mirror. The upper scan field is generated
by the front of the mirror.
Fig. 3-33: Function of the 8-layer sensors – upper scan field
Fig. 3-34: Function of the 8-layer sensors – lower scan field
Fig. 3-35: Function of the 8 layer sensor – overlapping scan fields
The alignment of the mirror provides a continuous vertical field of vision. The upper and lower scan fields start to overlap at a scanning angle of -50°.
3.10Physical switching inputs and outputs
3.10.1External switching inputs (digital)
The LD-MRS does not have a direct, physical switching input.
A maximum of two external switch inputs can be implemented for the LD-MRS using a CAN
module (optional accessory).
These inputs can turn specific evaluation instances on and off (see Chapter 3.8.1 Evalua-
tion instances, Page 48). The external outputs of the LD-MRS can be reset by using inputs
(Chapter Resetting a logical output, Page 50).
3.10.2External switching outputs (digital)
The LD-MRS does not have a direct, physical switching output.
A maximum of 10 external switching outputs can be implemented for the LD-MRS using a
maximum of two optional CAN modules (together with the optional switching inputs).
If an output is switched by the field monitoring, the LD-MRS reports field violations or contour violations as the results (results). Use SOPAS-ET to specify which evaluation instance
acts on which output.
3.10.3CAN module (optional accessory)
The aforementioned AND module is available with two types of protection:
•IP 20: type μCAN.8.dio-SNAP (Part No. 6038825), plastic housing for top hat rail instal-
Only operate the CAN module when the housing is closed!
Product description
Laser measurement sensor LD-MRS
•IP 67: type μCAN.8.dio-BOX (Part No. 6041328), small metal housing with seven
eaded holes for cable glands (4 x M16, 3 x M20) and 36 terminals.
thr
An additional set (Part No. 6043917) consisting of 5 cable glands/2 blind covers
(in
metal) is available.
Fig. 3-37: μCAN.8.dio-BOX module
The addre
ss range of 1 to 127 is available in the CAN bus. By default, the LD-MRS has the
address 1. The transmission rate of 500 kBit/s is fixed by the LD-MRS and needs to be set
for the two CAN modules.
When a cable is used to connect to the LD-MRS (Part No. 2050647 (2 m (6.56 ft))/Part No.
2050648 (8 m (26.2 ft)) in maximum configuration, both CAN modules are connected to
the end of the bus via the 9-pin D-sub connector. The required terminals are created using
the internal switches of each module. If only one module is connected to the LD-MRS, the
9-pin D-sub connector (Part No. 6042511) offers an integrated termination
end o
f the bus as an adapter.
resistor for one
3.11Status indicators and controls
The LD-MRS does not have any status indicators like LEDs, and it does not have any controls
like switches. The device is configured for measurement mode by means of commands in
messages or using the "SICK LaserView Customer Edition" visualization tool via the Ethernet
interface and is switched to measuring mode. Configuration for field monitoring is performed using the SOPAS-ET.
After mounting the LD-MRS, adjust the matching angles and the distance to the reference
plane.
4.3.1Mounting the LD-MRS
For mounting the LD-MRS safely at the mounting location, 4 M6 screws with washers and
lock washers are required. The supply voltage of the LD-MRS must be switched off.
4.3.2Distance between LD-MRS and object/surface to be monitored
a
cc
b
s
Fig. 4-1: Beam expansion
Caption:
avertical beam expansion for four layers
bvertical beam expansion for two layers (one laser pulse)
cvertical beam expansion for one layer
sdistance in mm
The reference plane for the required distance to the floor/wall is the sensor point of origin,
see Chapter 9.2.2 LD-MRS laser measurement sensor: Dimensions, Page 98.
The distance-dependant beam expansion can be calculated with this formula:
–vertical beam expansion for one layer = distance × 0.014
–horizontal beam expansion for one layer = distance × 0.0014
For some example values see Tab. 3-6, Page 30, Tab. 3-7 and Tab. 3-8.
1LD-MRS
2Connection of the CAN module to the LD-MRS
3Power supply line
4Ethernet data cable
5CAN module(s) with digital I/O ports
6Power supply line for the CAN module
7Power supply
8Desktop/laptop (temporary, configuration)
The following list provides an overview of the required installation steps:
•Connect the LD-MRS to the supply voltage
•Connect the PC temporarily (configuration)
•Measuring mode: Turn on the data interface for operation (depending on the use)
- or -
•Field monitoring: Wiring of the external switching inputs and outputs
Connect the module to the powe
r supply
After finishing
Chapter 6 Commissioning and configuration, Page 81.
the electrical installation, commission and configure the LD-MRS, see
5.2Planning the electrical installation
In general, observe the following requirements for electrical installation:
•A supply voltage of 9 to 27 V DC, typical 12/24 V DC (function extra-low voltage in ac-
cordance with IEC 60364-4-41).
The voltage supply must provide at least 36 W power output. The device can be connected to a supply by solar energy.
•Electrical isolation of RS232 and sync:
•If cables are long or the sensor and evaluation computer are far apart from one anoth-
er, potential differences can occur. To prevent damage, please use an RS232
on the sync inputs and
•Measuring mode:
omputer (such as a vehicle computer) with an Ethernet interface for pr
Host c
the
measurements from the LD-MRS
•Field monitoring:
Externa
– processing the output switching signals from the LD-MRS
– optional arming of specific evaluation instances using the switching inputs of th
•PC/laptop with Ethernet interface for temporary connection for configuration/visualiza-
tion of the measured data
•Connection lines: see the order information in the product information "Product family
of the laser measurement sensor LD-MRS" or Tab. 5-4, Page 67.
l control required for
-MRS
LD
outputs for electrical isolation.
isolator
ocessing
e
8012948/ZMQ0/2017-06-09
Damage hazard!
The sensor must not be operated without supply voltage ground or a potential differ-
ence between data ground and supply voltage ground.
•Electrical connection may only be established or disconnected when not under voltage.
•Specify and implement the wire diameter and the correct fusing according to valid stan-
dards.
•For data cables only use shielded cables (twisted pair).
•In order to prevent noise, do not lay data cables parallel to power supply or motor ca-
bles, e. g. in cable ducts.
•Lay all cables such that there is no risk of tripping and all cables are protected against
damage.
•Do not switch on the supply voltage for the LD-MRS before all connection works are fin-
ished and the wiring has been checked thoroughly
•In applications with high solar radiation, please lay the cables in protective tubes.
5.4.1Connect the supply voltage
Supply voltage requirements
The LD-MRS requires for operation a supply voltage of 9 to 27 V DC (function extra-low voltage in accordance with IEC 60364-4-41). The function extra-low voltage can be generated
by a power supply with a safety transformer in accordance with IEC 742.
ImportantThe output circuit of the power supply must be safely electrically isolated from the input
circuit, this feature is normally provided by a safety transformer in accordance with IEC
742.
The LD-MRS consumes the following power:
ProcessPower consumption
Switch-ontypically 36 W (1.5 A at 24 V DC)
permanent operationtypically 8 W (0.34 A at 24 V DC)
permanent operationmax. 10 W (0.4 A at 24 V DC)
Tab. 5-5: Power consumption of the LD-MRS
The optional CAN module consumes approximately 1.5 W at DC 8 to 50 V without a load.
ImportantIn order to ensure protection against short circuit / overload of the supply cable, dimension
and fuse the wire cross-sections accordingly. The valid standards must be observed.
Make sure that the voltage supply to which the LD-MRS will be connected, is switched
off.
1.Connect the supply cable, e. g. Part no. 2049823 (2 m (6.56 ft)), to the LD-MRS
, see
Fig. 5-3, page 70.
2.Connect the supply cable voltage-free to the voltage supply.
In order to bridge distances greater than 20 m (65.6 ft) (cable no. 2049825), use
extensio
n cable at the consumer side between the supply cable and the voltage supply.
an
ImportantIf you remove the bunch pin plugs of the supply cable and connect the cores directly to
the terminal block, make sure to connect the shielding of the cable also at a suitable
ground point.
Fig. 5-5: Design of the CAN module μCAN.8.dio-SNAP
Caption:
1Terminal for the power supply V+
2Terminal V
P
3Terminal for the power supply GND
4Switch for termination
5DIP switch for the transmission rate
6DIP switch for the address
7Terminal for the CAN bus CAN_H and CAN bus CAN_L
8LED
9Terminal strip of digital switching inputs/outputs 1 to 4
blTerminal strip of digital switching inputs/outputs 5 to 8
CAN bus
Since the fixed transmission rate of the LD-MRS is 500 kBit/s, the maximum bus line length
is 100 m (328 ft).
ISO 11898 recommends the following wire cross section: 0.25 mm
40 m (0 to 131.2 ft)) and 0.34 mm
2
(22 AWG) (for 40 to 100 m (131.2 to 328 ft)).
2
(24 AWG) (for 0 to
The bus must be connected to each end of the cable with a 120 Ohm resistor (terminated).
It is recommendable to use a shielded (braided), twisted two-wire cable with a surge impedance of 120 Ohm.
Fig. 5-6: Connect two CAN modules to the CAN interface of the LD-MRS
Caption:
1LD-MRS
2CAN module
When a cable (Part No. 2050647 (2 m (6.56 ft))/ 2050648 (8 m (26.2 ft))) is used to connect to the LD-MRS in maximum configuration, both CAN modules are connected to the end
of the bus via the 9-pin D-sub connector, see Fig. 5-6
CAN_H1 (Pin 7) (black)
CAN_H2 (Pin8) (black)
120Ω
CAN_L2
(Pin4) (red)
Int. termination resistance (green)
CAN_L1 (Pin 2) (brown)
1
3
8012948/ZMQ0/2017-06-09
Fig. 5-7: Connect one CAN module to the CAN interface of the LD-MRS
Caption:
1LD-MRS
2CAN module
If only one CAN module is connected to the LD-MRS, the CAN terminating resistance (D-Sub
connector) Part no. 6042511 is integrated in the termination resistance for the other end
of the bus, see Fig. 5-7, page 73.
The required termination is provided by a CAN module via the internal termination switch;
see Fig. 5-4, page 71 and Fig. 5-5, page 72.
Termination is active when the termination switch is in the ON position.
Digital switching inputs
The digital switching inputs of the CAN modules are connected to the positive supply voltage
V
The digital switching output of the CAN modules are connected to the positive supply voltage V
(high side driver).
+PWR
Fig. 5-9: A circuit diagram of digital I/O terminal (high side driver)
Caption:
1logic
2impedance R
3V
+PWR
5.4.3Configuring the CAN module
Setting the CAN address
Fig. 5-10: DIP switch for address (example: address 9)
For addressing in the CAN bus, the address range 1 to 127 is available. By default, the address of theLD-MRS is 1. Each address assigned to a device may only occur once (such as
10, 11, 12).
ImportantSwitch 8 must always be in the 0 position (OFF).
The switches may never all be in the 0 position (OFF) at the same time.
In this configuration, the CAN module will not start.
To configure using the visualization tool SICK LaserView Customer Edition (measuring
mode) or the configuration software program SOPAS-ET (field monitoring), use a crossover
cable such as Part No. 2049826 to connect the LD-MRS to the ethernet port of the computer.
RS-232 interface
In order to be able to access the RS-232 interface as auxiliary data interface, connect the
LD-MRS via a data cable, e. g. Part no. 2049831, to the COM port of the PC, see Fig. 5-12.
*) Twisted wire pair:
Sync-in (3) with Data GND (4); Sync-out (2) with Data GND (5); TxD (1) with Data GND (6); RxD (7)
with Data GND (8); CAN_L1 (9) with CAN_H1 (10); CAN_L2 (11) with CAN_H2 (12)
8012948/ZMQ0/2017-06-09
Tab. 5-8: Pin assignment of the cable splitter 1:3
Do not start up without testing by a qualified technician!
Before a system equipped with the LD-MRS is started up for the first time, it needs to be
checked and released by a technician.
Follow the instructions in Chapter 2 For your safety, Page 9.
Start up and configuration of the LD-MRS is independent of the intended use with the following tools:
•SICK LaserView Customer Edition visualization tool for measuring mode.
The installation and use of SICK LaserView Customer Edition as well as the configuration of the LD-MRS are described in the operating instructions for the SICK LaserV
C
ustomer Edition visualization tool (Part No. 8013787, English version).
This document is also on the CD, "Manuals & Software LD-MRS".
•SOPAS-ET configuration program for field monitoring.
The installation and use of the SOPAS-ET are described in this section as well
device
-related online help in SOPAS-ET.
In normal mode, the LD-MRS is fully-automatic and does not require any operator intervention.
iew
as the
6.1Overview for the startup steps for field monitoring
•Install the SOPAS-ET configuration program on the computer
•Connect PC to LD-MRS
•Take the LD-MRS into operation with the factory defaults
•To optimize the LD-MRS functionality, adjust the LD-MRS and adapt the configuration
he LD-MRS
of t
•Use SOPAS-ET to create an application-specific set of parameters for field monitoring,
and save
•Test LD-MRS for correct functionality
them permanently in LD-MRS and on the computer
6.2SOPAS-ET configuration program
SOPAS-ET is used for interactive configuration. With this configuration program, the parameters can be set and tested for the evaluation behavior and output characteristics of the LDMRS for field monitoring is needed. The configuration data can also be saved and archived
as a parameter set (project file) on the computer.
You can find help for using the program interface and individual options in SOPAS-ET:
•HELP menu, HELP F1: detailed online help on the program interface and individual op-
tions
•HELP window (bottom left in the user interface): Context-sensitive help for the most re-
cently appeared dialog
•ToolTips: place the mouse pointer over an entry field. A small reference text (tool tip)
6.3.2Start the SOPAS-ET and open the scanning assistant
1. Start the SOPAS-ET.
SOPAS-ET opens the program window with the English user interface by default.
2. To change the language setting, click C
interface
language to an other language in the menu by going to T
ANCEL in the start dialogue and change the user
OOLS, OPTIONS.
3. Once the language has been changed, close SOPAS-ET and restart it.
4. In the dialog window, select the option C
5. In the main window under S
e S
Th
CANASSISTANT dialog appears.
CANASSISTANT, click the button CONFIGURATION.
REATENEWPROJECT and confirm with OK.
6.3.3Configure the ethernet connection
ImportantDeactivate all the programs on your computer that access the ethernet or TCP/IP.
1. In the S
bo
2.
Confirm the settings in the S
Th
CANASSISTANT dialogue, go to INTERNETPROTOCOL, IP COMMUNICATION and check the
x A
CTIVATE IP COMMUNICATION and USE AUTOIP.
CANASSISTANT dialog by clicking OK.
e S
CANASSISTANT dialog closes.
6.3.4Run a scan
1. In the S
2. Select the listed devices and confirm with A
The connected devices are searched over the
SOPAS
CANASSISTANT dialogue, click the button STARTSCAN.
DDDEVICE.
connection.
-ET adds the listed devices to the project tree.
6.4Initial startup
The LD-MRS is adapted to the local field monitoring by SOPAS-ET. A user-specific set of parameters is created with SOPAS-ET. The parameters set can either be loaded from the device (upload), or it can be created independently.
The parameter set is then loaded onto the LD-MRS (download). This is either done immediately (the SOPAS-ET option D
DOWNLOADALLPARAMETERSTOTHEDEVICE).
ImportantAfter configuration, permanently save the parameter set on the LD-MRS.
OWNLOADIMMEDIATELY) or manually (the SOPAS-ET command
Saved project file with
archived parameter set
(*.spr)
Hard disk
Upload
Download
LD-MRSComputer with SOPAS-ET
Commissioning and configuration
Laser measurement sensor LD-MRS
RecommendationIn addition, the parameter set can be saved as a project file (sprfile with configuration data)
and archived on the computer.
Fig. 6-1: Principle of data storage
6.4.1Configure the LD-MRS
You can configure the LD-MRS interactively using SOPAS-ET.
All of the configurable parameters for the LD-MRS are compiled in a device description (jar
or sdd file) for SOPAS-ET. You can access the parameters through the project tree for the
device description.
The functions of the parameters are explained depending on the context in an online help
(press [F1]). The valid value range and default are listed n the display window P
INFO (right-click when the pointer is over the parameter.
ImportantThe progra m a cce ss to the LD-MR S i s pa ssword protected. After configuration is successful,
change the password to ensure protection.
User levelPassword
Authorized clientClient
Tab. 6-2: Password
Use the structure of the project tree in SOPAS-ET to configure the required parameters for
your application.
Do not turn off the power during configuration.
If the power is turned off during configuration, all of the configured parameters are lost.
1. In the O
PTIONS menu, select the command LOGONTOTHEDEVICE, and log on with the pass-
word "client" as an A
2. Configure the LD-MRS using the parameters in the SOPAS-ET for the desired application.
You can find help on using the program interface and individual options in SOPAS-ET:
ARAMETER
UTHORIZED CLIENT.
RecommendationTo reset the LD-MRS back to its original settings, first export the data of a device with the
Reset configuration
original settings to a file. This device data can later be loaded to a configured device to reset
the configuration to the original settings.
Use the graphic scan display in SOPAS-ET to verify the generated measurements and measuring range online.
1. In the project tree, select LD-MRS /
2. To start measurement, click the button D
3. Compare the measured line with the desired results.
Important–The S
does not
4. After the test measurement is over, permanently save the configuration on the LD-MRS:
In the menu, go to LD-MRS/P
CANDISPLAY on the MONITOR depends on the available power of the computer and
occur in real time. Consequently, not every measurement is visualized.
MONITOR/EVALUATIONMONITOR.
ISPLAYSCANLINE.
ARAMETERS/SAVEPERMANENTLY.
6.6Synchronization
Fig. 6-2: Scheme of the data flow in a system with one LD-MRS and up to two external devices
Caption:
1LD-MRS
2"Data interfaces/synchronization" connecting line
3External device (e.g. camera; optional)
4External synchronization pulse signal generator (optional)
In a fusion system with an LD-MRS and an external device (e. g. a camera), the LD-MRS
must be synchronized with the external component to minimize time shift during data collection of both components.
This reduces the necessity of scan data corrections during processing.
During the synchronization process, the LD-MRS adjusts its scanning frequency (in a small
range) so that the measurement in a selected angle happens at the time the external device
presents a synchronization pulse.
Another option is to define the LD-MRS as master which presents its synchronization pulse
to the external device, which can adjust its processing frequency of measurement (e. g. taking pictures) to this synchronization signal.
ImportantThe synchronization frequency must be between 8.5 Hz and 50 Hz.
Frequencies above and beyond these values are not accepted by the LD-MRS.
The precision of the synchronization (i. e. the time difference between synchronization
pulse and the moment of crossing the angle that was specified as SyncAngle) is approx.
±1,000 μs for a scanning frequency of 12.5 Hz if no external forces act on the LD-MRS (especially no angle acceleration).
This is the equivalent of an angular precision of ± 4.5° around the true 0° direction.
For the scanning frequency of 25 Hz and 50 Hz the precision is ± 500 μs, equivalent to an
angular precision of ± 4.5° or ± 9°, respectively.
The external synchronization frequency f
50 Hz and should have a relative precision better than 0.1 % (the better f
can be selected in a range of 8.5 Hz ≤ f
sync
sync
≤
sync
, the better the
laser measurement sensor).
Fig. 6-3: Details synchronisation
Therefore we recommend not to send the synchronization signal from a PC because they
generally have less precise timers. Instead use a designated micro-controller to generate
the synchronization signal.
The external synchronization frequency and the scanning frequency do not need to match.
The external synchronization frequency must be within ± 4 Hz of the scanning frequency f
(f
= f
can
sync
± 4 Hz). The LD-MRS adjusts the scanning frequency to the external syn-
scan
chronization frequency.
When the external synchronization frequency fails, the LD-MRS adjusts its scanning frequency again to the initially specified value.
The voltage level to detect an external trigger signal must have level E/A-232 (formerly RS-
232).
The LD-MRS detects a logical positive flank as the moment of synchronization, see T
SyncTime
in Fig. 6-4.
6.6.2Synchronization systems
There are three possible synchronization systems:
•LD-MRS and external device synchronized by the LD-MRS
•LD-MRS and external device synchronizing the LD-MRS
•LD-MRS and two external units, one of which synchronizes the LD-MRS, the other being
synchronized by the LD-MRS
The external unit can also be a different LD-MRS.
Each LD-MRS starts as a master unit, sending trigger points to the next unit. The LD-MRS
always sets the output signal to t
Pulse_out
= 20 μs; see the red output signal inFig. 6-4,
page 87.
When the trigger input receives a transition from low to high, the LD-MRS switches to the
salve mode. This stops the further active sending of the Sync Pulse. The device then only
copies and transmits the master trigger.
The external synchronization signal should remain in high state for at least t
Pulse,min
=
10 μs, see blue input signal in Fig. 6-4, page 87.
If no other external trigger is present, the LD-MRS switches back to master mode after approx. 300 ms.
8012948/ZMQ0/2017-06-09
6.7Switching off the LD-MRS
Switch off the supply voltage of the LD-MRS.
The parameter set most recently stored in the LD-MRS remains valid.
The LD-MRS is classified as enclosure rating IP 69k and is only protected against permeating liquids in that regard.
Do not use aggressive chemicals!
If necessary, clean the housing with a soft, damp cloth (lent-
window deterg
If mounted, the LD-MRS can also be cleaned with a high pressure cleaner as long as the
maximum values regarding pressure, duration and distance between high pressure cleaner
and LD-MRS comply with DIN 40050. In addition, protect the plugs of the LD-MRS with suitable measures when cleaning with a high-pressure cleaning device because the plugs are
classified as enclosure rating IP 68.
Else water can permeate the seals or damage parts of the LD-MRS housing.
ent.
free). Use a mild antistatic
7.3Exchanging an LD-MRS
ImportantOnly trained and authorized service personnel of SICK AG may repair the LD-MRS.
A faulty or damaged LD-MRS must be dismounted and replaced by a new or repaired LDMRS.
7.3.1Exchanging the device
1.Switch off the supply voltage of the LD-MRS
2.
Unplug all of the device's plugs and remove the cables.
3.If necessary, appropriately mark the alignment of the LD-MRS.
4.Remove the LD-MRS from the holder.
5.Mount a new device in reverse sequence and connect it.
6.Switch on the supply voltage of the LD-MRS.
.
7.3.2Configure the new device.
Measuring mode:
The replaced LD-MRS must be reconfigured to the application with the SICK LaserView Customer Edition configuration program.
See the operating instructions, "SICK LaserView Customer Edition visualization tool“ (Part
no. 8013787), section "Changing and saving the configuration of the LD-MRS".
Field monitoring:
1.Use SOPAS-ET to load the parameter set for the application saved on the co
the
If a malfunction arises in the LD-MRS, it emits warning and error messages via the ethernet
interface.
The operator can extract the warnings and fault messages from the Ethernet data stream.
If the SICK LaserView Customer Edition is used to visualize the scan data, the errors are displayed there.
See the operating instructions, "SICK LaserView Customer Edition visualization tool (Part
No. 8013786), section "Trace log".
The warnings and fault messages contain Bit fields, each Bit represents one warning or one
fault. Available are a maximum of twice 16 Bit each for warnings and faults.
8.2Overview of possible errors and faults
8.2.1Mounting error
•LD-MRS not mounted conveniently (e. g. limited field of vision)
•LD-MRS adjusted incorrectly (angle of rotation)
8.2.2Error in the electrical installation
•Interfaces of the LD-MRS switched incorrectly
8.2.3
•Device limits (scanning range) not considered
•The parameters are not adapted to local conditions.
8.2.4Fault during operation
•Device error (hardware/software)
Parametrization error
8.3Monitor error and fault indications
The LD-MRS monitors itself during operation:
•After switching on the supply voltage, the LD-MRS first runs a self-test checking import-
ant hardware components before initialization (loading the parameter set and initializing the device
•In case of an error, the LD-MRS sends warning and error messages via the Internet in-
terface; see Chapter 8.1 Warnings and fault messages, Page 91.
The following tools are required for troubleshooting:
•A digital measurement device (measuring current/voltage)
•A computer with the SICK LaserView Customer Edition visualization tool and SOPAS-ET
configuration software
•Tools
FaultPossible causeRectification
1. SICK LaserView Customer Edition:
Configuration.
PC with SICK LaserView Customer Edition cannot communicate with the LDMRS.
2. Measurement mode: no scan data.• LD-MRS inconveniently mounted or ad-
3. Field monitoring:
No signals from the field violations
4. Field monitoring:
The CAN modules are not working
(correctly) (optional accessory)
1) For additional assistance, see the chapter "Diagnosis" in the manual for the CAN module on the CD, "Manuals & Software LD-MRS“ (Part No. 2050246)
2) 1 = ON, 0 = OFF
Tab. 8-1: Troubleshooting the LD-MRS
• Supply voltage for LD-MRS not switched
on.
• PC not connected to LD-MRS.
justed.
• LD-MRS parametrized incorrectly.
• Field not assigned to an evaluation instance
• Switching output not activated or incorrectly configured
• The supply voltage V+ is not connected to
the CAN electronics, the poles are
switched or it is not turned on
• The supply voltage V
for the power driver
P
(switching outputs) is not connected
• Switch 8 for the CAN address dip switch
was accidentally set to 1
2)
(for program-
ming via an RS-232)
• Incorrect transmission rate
•Wrong termination
• CAN addressed several times
• The logical inputs/outputs in the LD-MRS
incorrectly assigned to the physical I/O
ports in the CAN modules
• The switching logic of the inputs/outputs
is inverted, active high, active low
• Check supply voltage, switch on.
• Connect PC via Ethernet interface of the
LD-MRS.
• Mount LD-MRS at more suitable place or
readjust.
• Set correct parameters for the LD-MRS.
• In SOPAS-ET: check the evaluation instance and provide the correct parameters if necessary.
• In SOPAS-ET: check the CAN parameters.
Check the configuration of the external
outputs.
Check the wiring between the LD-MRS
and the CAN module.
• Connect the supply voltage (8 to 50 V DC)
to V+ and GND and turn on if necessary
• For example, bridge the supply voltage to
V+ to V
P
Set switch 8 for the CAN address dip
•
switch to 0
2)
• In each CAM module, set a transmission
rate of 500 kBit/s (set the dip switch
transmission rate from left to right:
2)
1110)
• Activate the termination in both modules
with the Termination switch.
If there is only one CAN module, terminate
the open bus end with a 120 Ohm resistance.
• Configure a unique CAN address for each
bus user
• Check the assignment of the logical inputs/outputs in SOPAS-ET
• Check the assignment of the switching
logic in SOPAS-ET
If an error can not be rectified by the listed measures, the LD-MRS might be defective. The
operator can not repair the LD-MRS and thus restore its functionality after a fault. However,
the LD-MRS can be rapidly replaced by the user. See Chapter 7.3 Exchanging an LD-MRS,
Page 90.
If an error occurs which cannot be eliminated, please contact SICK Service:
•International: Competent SICK branch office or SICK subsidiary
– For telephone numbers and e-mail addresses please see the back page of these op-
erating instructio
For the postal address please visit www.sick.com.
–
Only return device after consultation with the SICK Service.
ImportantOnly trained and authorized service personnel of SICK AG may repair the LD-MRS.
Dimensions (H x W x D)99 mm x 22.5 mm x 114.5 mm125 mm x 80 mm x 57 mm without
Threaded holes for cable glands –4 x M16 x 1.5
Weightapprox. 150 g (5.3 oz.)Approximately 515 g (18.2 oz.) without
Enclosure ratingIP 20IP 67
EMC testEmitted interference: according to EN 50081-2, EN 55022, class A; EN 61000-6-4 (2001)
Noise immunity: according to EN 50082-2, EN 61000-6-2 (2001)
Ambient operating temperature–40 to +85 °C (–40 to +185 °F)
MountingSnap-in on TS35 top hat rail in the control
cabinet
1) Set of cable glands (metal), Part no. 6043917 deliverable as optional accessory, consisting of:
2 cable glands M16 x 1.5 for a line diameter of 5 to 9 mm (0.2 to.35 in), 3 cable glands M20 for a line diameter of 10 to 14 mm (0.39 to 0.55 in) including filler plug,
2 dummy covers M16, 3 dummy covers M20
Tab. 9-2: Technical specifications for the CAN modules (from the manufacturer's data sheets and manuals)
= 500 mA per output, sustained short-circuit proof, total current of all