Reading pulses via "Sensor 1" switching input (default setting)
NoteIn the default setting, the ICR does not read any bar codes.
Do not switch off the power supply during configuration.
1.Check the scope of delivery to ensure that it is complete.
2.Install the ICR in the reading station and align it with the stationary object with the 2-D
code (DataMatrix ECC 200) at a reading distance of 80 mm (3.15 in)(standard
device). Available reading area 40 mm x 32 mm (1.58 in x 1.26 in).
3.Install the CDB 420 or CDM 420 Connection Module.
4.Connect the ICR to the CDB 420 or CDM 420 Connection Module.
5.Install the reading pulse sensor.
6.Connect the sensor to the "Sensor 1" switching input in the CDB 420 or CDM 420.
7.Connect the host to the "host interface" in the CDB 420 or CDM 420.
8.Switch on the power supply for the CDB 420 or CDM 420.
Once the ICR has been started, the "Device Ready" LED lights up. The beeper beeps
twice to indicate that reading mode has been started.
9.Switch on the PC and start Windows
10. Install the CLV Setup operating software and the CLV Setup Help online help software,
which is supplied on the accompanying CD-ROM (“Manuals & Software“), on your PC.
11. Establish a physical connection between the PC and the Ethernet interface. To do so,
connect the PC directly to the ICR by means of a crossover cable or connect the PC and
ICR to the Ethernet by means of patch cables (OK = green "Ready" LED (Ethernet) lights
up on the ICR)
12. Start "CLV Setup". To initiate communication, select the Ethernet interface (O
INTERFACE). CLV Setup establishes communication with the ICR (standard IP address).
13. Upload the parameter set of the ICR (displayed on the tab pages).
14. Start the CLV Assistant and carry out the following steps consecutively: C
A
SSISTANT, ETHERNET ASSISTANT, SCANNER ADJUSTMENT, and DATAMATRIX AUTOSETUP.
15. To check the image online, launch the I
A
DJUSTMENT step.
16. Scanner Adjustment: Represent the 2-D code statically in the red illumination field
(pulsed) of the ICR. In I
MAGEFTP, make sure that the image is of a good quality in the
image memory.
17. DataMatrix AutoSetup: prepare AutoSetup and, when prompted, start the reading
trigger. Represent the 2-D code statically again in the red illumination field. When doing
so, retain the reading distance you adjusted previously.
18. Start DataMatrix AutoSetup. If successful, exit the Assistant after the final check. If not,
repeat AutoSetup.
19. Continue configuring the ICR using the setting options on the tabs in CLV Setup (reading
clock, data output). Download the modified parameter set to the ICR temporarily. Make
sure that data is transferred properly to the ICR.
20. If necessary, check and optimize the set parameter values. Download the parameter
set to the ICR permanently and save the parameter set as an "*.scl" configuration file
in CLV Setup.
The ICR can now be operated with the application-specific settings.
Fig. 4-1:Position of the securing threads on the ICR............................................................. 4-3
Fig. 4-2:ICR installation options using the mounting bracket no. 2 025 491............. 4-3
Fig. 4-3:Alignment of the ICR reading area with the code .................................................. 4-4
Fig. 4-4:Definition of the reading distance ................................................................................ 4-4
Fig. 4-5:Reading angles between the reading area and the code .................................. 4-5
Fig. 4-6:Installing the ICR parallel to the object surface....................................................... 4-6
Fig. 4-7:Count direction of the code position CP for bar codes along
the reading window............................................................................................................ 4-6
Fig. 4-8:Line scanner: Installation example for the external reading pulse sensor... 4-9
Fig. 5-1:Block diagram: Connection of the ICR to the CDB 420 or CDM 420
Fig. 6-19:CLV Assistant: Display of the performed steps and determined values.... 6-14
Fig. 6-20:CLV Setup: Status line with IP address/port of the Ethernet connection ..6-15
Fig. 6-21:CLV Setup: “Interface Options“ dialog box line with IP adress/port of
the Ethernet connection................................................................................................6-15
Fig. 6-22:ImageFTP: Image output, with the markings for CP limitations (violet),
the 2-D symbol (green) and the position of the DataMatrix decoder at
the end of reading pulse (blue) ..................................................................................6-16
Fig. 6-23:CLV Setup: Output of the reading result of the auxiliary interface for
2-D codes in the Terminal Emulator.........................................................................6-30
Fig. 6-24:Position of the symbol in the image field of the image memory ...................6-31
Fig. 6-25:CLV Setup: Output of the reading result of the auxiliary interface for
bar codes in the Terminal Emulator..........................................................................6-32
Fig. 6-26:CLV Setup: Display of the percentage evaluation of the auxiliary
interface for bar codes in the Terminal Emulator................................................6-35
Fig. 6-27:ImageFTP: Program window.........................................................................................6-37
Fig. 6-28:ImageFTP: Image output, with the marks for CP limitations (violet),
the 2-D symbol (green) and the position of the DataMatrix decoder
at the end of reading pulse (blue).............................................................................6-38
This document contains instructions for operating the 2-D code reader with fixed focus
•ICR 840 standard
The device is available
•with side reading window
•with front reading window
The Image Code Reader reads 2-D codes (DataMatrix ECC 200) using an image
recording and processing system. To do so, the device provides a rectangular reading
area in the reading plane. Currently, only images of stationary objects can be reliably
evaluated.
This document provides information on
•Installation and electrical installation
•Startup
•Operation and configuration (parameterizing)
•Maintenance
•Exchanging the device while retaining the parameter set
•Special applications and procedures
The ICR 84x Image Code Reader with all its variants is simply referred to as “ICR“ in the
document, except where a distinction is necessary.
1.2Target audience
This document is intended for persons who are responsible for the following activities:
1.2.1Installation, electrical installation, maintenance and replacement
Electricians and service technicians
1.2.2Startup, operation and configuration
Technicians and engineers
1.3Information content
This document contains all of the information necessary for the installation, electrical
installation and startup of the ICR with the factory default settings.
A series of step-by-step instructions is provided for each of these activities.
The ICR is configured for specific applications using the Windows-based “CLV Setup“
program. Further assistance is also available in the form of the online help system “CLV
Setup Help“. The procedure for installing and operating the software is described in the
Appendix.
Additional information on the structure of the Image Code Reader and 2-D code/bar code
technology is available from the Auto Ident division of SICK AG.
Internet address: www.2d-code.com.
For the ICR to function properly and safely, it must be installed and setup by sufficiently
qualified personnel.
The end user must be supplied with the operating instructions.
The end user must be provided with expert tuition and is advised to read the operating
instructions.
The following qualifications are required for the various tasks involved:
2.1.1Installation and maintenance
•General technical training
•Knowledge of the standard guidelines relating to safety in the workplace
2.1.2Electrical installation and replacement
•Practical electrical training
•Knowledge of the common electrical safety guidelines
•Knowledge regarding the operation of the devices in the relevant application
(e.g. conveyor belt)
2.1.3Startup, operation and configuration
•Knowledge regarding the operation of the devices in the relevant application
(e.g. conveyor belt)
•Knowledge of the hardware and software environment of the relevant application
(e.g. conveyor belt)
•Basic knowledge of Windows 95
Windows XP
•Basic knowledge of an HTML browser (e.g. Internet ExplorerTM)
•Basic knowledge of data transmission
•Basic knowledge of Ethernet connections
•Basic knowledge of 2-D code/bar code technology
TM
TM
/98TM, Windows NT4.0TM, Windows 2000TM or
2.2Intended use
The ICR automatically scans and decodes 2-D codes and bar codes. It is installed in a
reading station and reads these codes on objects which are not moved.
The ICR transfers the data content of the decoded codes via its data interface (Host, CAN
or Ethernet) to a host for further processing.
Any warranty claims vis-à-vis SICK AG will be rendered invalid if the device is used for any
other purpose or if changes are made to the device, also as part of the installation and
electrical installation procedures.
NoteDon’t open the device. The producer warranty will be forfeited if the device is opened.
2.3General safety precautions and protection measures
¾ Read the general safety precautions carefully and observe them at all times. This also
applies to the warnings and operating instructions in the individual chapters of this
document.
2.3.1RF interferences
The ICR 84x is exclusively intended for use in an industrial environment.
In case of use in residential areas, RF interference may occur.
2.3.2Electrical installation
Risk of injury by electrical current
In the CDM 420 Connection Module, the CMP 400 Power Supply Module is connected to a
mains voltage of 100 to 250 V AC/50 to 60 Hz.
¾ When working with electrical equipment, always follow the relevant safety regulations.
2.3.3LED illumination for reading area
The ICR 84x is classified in LED class 1.
Under normal and sensible conditions, the accessible beam of the LED illumination is not
hazardous. Blinding, impairment of ability to see color, or other irritations, however, cannot
be excluded.
¾ The entire glass window acts as a LED outlet aperture.
¾ Caution – use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous radiation exposure.
¾ Do not open the housing.
(Opening the housing does not deactivate the laser diodes).
¾ Observe the laser protection specifications (latest version).
Radiation power
The LEDs of the illumination operate at a wavelength of λ =617nm ± 15 nm (visible red
light) with a pulse duration of max. 5 ms. The energy in the human eye is < 1,7 J/m
The product is classified in LED class 1 in accordance with EN 60825-1 and IEC 60825-1
(for publication date, see the warning sign on the device).
Warning labels
The warning label (Fig. 2-1, Page 2-3) in three languages is located on the wide side of the
housing (see Fig. 3-1, Page 3-5).
Fig. 2-1: Black-yellow signed warning labels found on the ICR
If the ICR is installed in a machine/panel with the result that the warning labels are no
longer visible, additional warnings (not included in the scope of delivery) must be
provided on the machine beside the emergence aperture of the LED radiation.
Activation and deactivation of the LEDs when reading is controlled by the reading pulse
(trigger source).
A timer (illumination timeout) automatically deactivates the LEDs 10 minutes (default
setting) after a continuous reading pulse is initiated in Reading mode with switching input
pulse modes “Sensor Input“ and “Serial Interface“. However, it does not end the reading
interval. In this case, the ICR outputs the following message to the auxiliary interface:
“Illumination safety timeout“
The reading interval must be terminated by resetting the trigger signal. The LEDs activated
again by the next reading pulse.
The illumination timeout can be set in the range of 1 min to 25 h or deactivated (see
Table 6-3, Page 6-25).
The Illumination LEDs are periodically activated in the operating mode “Percentage
Evaluation“ and are always activated in the pulse mode “Free Running“ in Reading
mode.
2.4Quick stop and quick restart
2.4.1Switching the ICR off
¾ Switch off the power supply or remove the ICR cable plug from the connection module.
This can result in loss of the following (at the most):
•The application-specific parameter set, if it was only stored temporarily in the ICR
•The last reading result
•Daily operating data
(operating hours counter, reading interval count, good read count, no read count,
maximum duration reading interval, minimum duration reading interval, number of
matches with match code 1, number of matches with match code 2, number of No
Matches.)
2.4.2Switching the ICR on again
¾ Switch on the supply voltage or reattach cable plug of the ICR to the connection module.
The ICR resumes operation with the last permanently stored parameter set and resets
the daily operating data.
The ICR is designed to cause minimum impact on the environment. It does not contain any
silicone-based materials on the housing surface and, therefore, does not represent any
problems for paint sprayers in paint shops, for example.
2.5.1Power requirements
The power requirements are particularly low: The ICR has a max. power consumption of
typically 7 W (max. 10 W).
The value is given for devices with disconnected switching outputs.
2.5.2Disposal after final removal from service
Dispose of unusable or irreparable devices in accordance with the respective state
regulations on waste disposal in a manner compatible with the environment. The ICR can be
separated into recyclable secondary raw materials and special-category waste (electronic
scrap) (see Chapter 7.3 “Disposal“, Page 7-2).
At present SICK AG does not take back devices which have become unusable or
irreparable.
•An information sheet (note on device) with terminal diagram and Quick Start instructions
Depending on the number of devices ordered, one or more copies of the following:
•CD ROM (no. 2 029 112) with
– "CLV Setup" program for Windows
(HTML files)
– "CLV-Connect" PC software (HTML files showing terminal diagrams)
– ICR 84x Operating Instructions in English and German as PDF edition as well as
additional publications (connections module, other SICK bar code scanners)
– freely available "Acrobat Reader" PC software for reading PDF files
TM
and the "CLV Setup Help" online help system
NoteThe latest versions of all the current publications/programs on the CD ROM can also be
downloaded from www.sick.com.
Depending on the number of copies ordered, the delivery includes (optional):
•ICR 84x Operating Instructions in English and/or German (printed edition)
An overview of the available installation accessories, connection modules, incremental
encoder, cables, plug connections, and sensors for reading pulse generation is contained in
Chapter 10.9 “Accessories“, Page 10-40.
3.1.2Variants
The ICR is available in the following variants:
Type (red light) Part No.CategoryHost interface typeReading window
The following are required to start up and operate the ICR:
Electrical connecting:
1.A SICK connection module for power supply of the ICR (15 to 30 V DC) and connection
of the data and functional interfaces.
Available types:
– CDB 420-001 (no. 1 023 885) for 10 to 30 V DC, enclosure rating max. IP 65
– CDB 420-101 (no. 1 024 305) for 10 to 30 V DC, enclosure rating max. IP 65
– CDM 420-0001 (no. 1 025 362) for 10 to 30 V DC, enclosure rating max. IP 65
– or –
Alternatively, a non-SICK power supply unit with a voltage output of 15 to 30 V DC
(functional extra-low voltage pursuant to IEC 364-4-41) and at least 12 W power
output.
Use cable no. 6 010 137 with 15-pin D-Sub HD socket and one open end for
connecting the ICR to the power supply unit.
2.The following operating voltages/power output values
– CDB 420-001: 10 to 30 V DC, pursuant to IEC 364-4-41, at least 10 W
– CDM 420-0001: 10 to 30 V DC, pursuant to IEC 364-4-41, at least 10 W.
100 to 250 V AC, 50 to 60 Hz when using the CMP 400 Power Supply Module
– If the following modules are additionally built-in in the CDB 420 module:
CMC 400 Connection Module Cloning: 10 to 30 V DC, additionally 0.5 W
– If the following modules are additionally built-in in the CDM 420 module:
CMC 400 Connection Module Cloning: 10 to 30 V DC, additionally 0.5 W
CMD 400 Connection Module Display: 18 to 30 V DC, additionally 1 W
CMF 400 Connection Module Fieldbus: 18 to 30 V DC, additionally 2 W
3.With external clock pulse (start/stop of reading interval) supply via the “Sensor 1“
switching input: a suitable reading pulse sensor for signaling the presence of an object
with a bar code, e. g. a photoelectric reflex switch.
4.With optional external clock pulse (stop of reading interval) supply via the “Sensor 2“
switching input: a suitable reading pulse sensor for signaling the end of reading intervall,
e. g. a photoelectric reflex switch.
5.If an external field illumination is necessary: a suitable light source to be connected to
the “Result 1“ switching output.
6.To delay the external reading pulse using a track controll: a suitable incremental
encoder.
7.A higher-level computer (host) with a data interface of type RS 422/485 or RS 232.
8.To connect the ICR to the Profibus DP and the DeviceNet: the corresponding CMF 400
Connection Module Fieldbus (Operating instructions see Chapter 10.11
“Supplementary documentation“, Page 10-44).
9.For connection of the ICR to the CAN bus: the operating instructions “Using the CAN
interface“ (no. 8 009 180, English).
Starting-up and configuring:
10. A PC (at least 80486, 66 MHz, 16 MB RAM, CD ROM drive, serial interface, mouse
(recommended) with Windows 95
Windows XP
11. For configuring the ICR via the auxiliary interface a 3-core RS 232 data cable (null
modem cable) with two 9-pin D-Sub sockets for connecting the PC to the auxiliary
interface of the ICR in the CDB 420 or CDM 420 Connection Module,
e. g. no. 2 014 054. Pin 2 (RxD) and pin 3 (TxD) are crossed.
12. For configuring the ICR and outputting images via the Ethernet interface:
For the connection of the ICR to the network: a standard Ethernet data cable (patch
cable), e.g. no. 6 026 083.
For the connection of the ICR to the PC (network card): an Ethernet crossover line,
e. g. no. 6 026 084.
13. To use the online help system “CLV Setup Help“, an HTML browser is required,
e. g. “Internet Explorer
TM
“ browser.
3.1.4Product features and functions (overview)
High-performance scanner:
•CMOS sensor of newest technology (1.3 Mega pixel)
•Fixed focus
•Front reading window, variant: side reading window
•Reading range (DOF) 58 to 106 mm (2.29 to 4.18 in), standard type
•Reading area (dependent of the reading distance), 40 mm x 32 mm (1.58 in x 1.26 in),
at reading distance 80 mm (3.15 in), standard type
•Resolution 0.15 to 2.0 mm (5.9 to 78.7 mil), standard type
•Image refresh rate/decoding frequency 25 Hz at highest resolution
•Function to adjust on the code print quality
•Variable active evaluation range of the image buffer, thereby higher image refresh rate
Safety and user-friendly features:
•Robust, compact metal housing, max. IP 65, CE certification
•LED class 1, illumination diodes switches off if reading interval is active for too long and
if the output power is exceeded
•Automatic self-test on startup. Can also be triggered at any time
•Diagnosis tools for installing and monitoring the system
•Parameterized output of reading diagnosis data in the read result
•Operating data query, and error messages on request
•Test string function for signaling readiness for operation
•Future proof thanks to firmware update via serial data interface (flash PROM)
•Low power consumption, other voltage range
Easy operation/configuration:
•With "CLV Setup" PC software for Windows (online), Help system and four assistants
(Connection Assistant, Ethernet Assistant, Scanner Adjustment, and DataMatrixAutoSetup).
Displaying the image memory content via the ImageFTP program.
•Alternatively with simple command strings, also for use with special devices
•Four status LEDs
•Beeper to confirm device functions or operating steps (can be switches off)
•Percentage evaluation - for assessing the quality of the reads (bar codes only)
•Special functions for system installation
2-D code/Bar code evaluation:
•DataMatrix ECC 200/all standard bar code types
•Max. 50 codes per reading pulse (max. 4,000 characters)
•Separation of identical bar codes of the same code type
•Code comparison (max. 2 matchcodes), can also be used as filter or sort criterion for
the reading result
•Sort sequences: code position, FIFO, LIFO, code length list
•Manipulation of the data output string via filter or format mask
Data communication:
•Main data interface: Host interface (with variable telegram structure), communication
can be routed via the Ethernet interface
•Auxiliary data interface: Terminal interface (with fixed telegram structure) with special
diagnosis functions, communication can be routed via the Ethernet interface
Reading pulse:
•External reading pulse, via switching input(s) or serial data interface
•Free running with timeout
Electrical interfaces:
•Serial host interface (RS 232, RS 422/485) with variable transfer rate and protocol
•Serial terminal interface (RS 232) as auxiliary data interface with fixed transfer rate and
fixed protocol
•CAN interface for integration in the SICK CAN scanner network or a CANopen network
•Ethernet interface with TCP/IP or FTP
•2 switching inputs for external reading pulse and special function (e. g. teach-in of match
code or encoder increment)
•2 switching outputs for signaling defined events in reading mode as well as for
triggering or direct powering an external field illumination
Connections:
•All interfaces excepting the Ethernet interface are connected via one 15-pin D Sub HD
plug, Ethernet interface: RJ 45 10 base T socket
•CDB 420 or CDM 420 Connection Module for connection to host (stand-alone) and for
integration in SICK CAN scanner network
•CMF 400 Connection Module Fieldbus in the CDM 420 Connection Module for
connection to field bus systems
The ICR uses an image recording and processing system (CMOS matrix sensor with 1280
x 1024 pixels) to scan 2-D codes and bar codes with a rectangular reading area. To do so,
LEDs integrated in the housing illuminate the area to be read. The LEDs generate a red
illumination field (pulsed) with a configurable flash duration. To start a reading cycle, the ICR
switches on the illumination. It absorbs the light reflected from the reading area (Fig. 3-2)
through the lens as gray values of an image area, which are stored in an electronic image
memory (image refresh rate: 25 Hz at full memory utilization).
To prevent distorted images in the image buffer memory, the object with the code must
remain stationary while the image is being recorded.
All dimensions in mm (in)
Fig. 3-2: Reading area of the ICR in the focus position (standard type)
The reading area (surface), which is mapped in sharp focus by the ICR in its image memory,
depends on the reading distance and, depending on the resolution, must be within the
reading area (DOF). Fig. 9-4, Page 9-5 shows the available reading area.
If the appropriate configuration settings are made to reduce the image buffer memory area
(image geometry) used for the evaluation, this increases the potential image refresh rate.
The refresh rate with the VGA resolution (640 x 480 pixels) is 60 Hz, for example. The ICR
automatically adjusts the flash frequency. With a maximum of two working areas, a
maximum of just two separate strips in the image field can be evaluated.
As 2-D-codes or bar codes are detected in the image data, the ICR's decoding algorithms
determine the code(s) contents. At the end of the reading cycle or immediately after
successful decoding, the ICR outputs the code data information via its data interface (main
data interface: Host, CAN or Ethernet interface) to a host/PC for further processing.
PC
Operation
Parameterizing etc.
Photoelectric switch
Reading pulse
Signal
Path increment
Teach-in match code 1
End of reading interval
•The SMART decoder (SICK Modular Advanced Recognition Technology) for decoding
bar codes with a small aspect ratio (ratio of the code height to the code length), bar
codes that are dirty or damaged, as well as bar codes that are tilted excessively
(azimuth angle)
•The tried-and-tested standard decoder of the CLV series for bar codes
The ICR derives useful diagnosis data from the reading process and transfers it to the host.
It also records operating data that can be requested at any time. The reading quality
of bar codes can be checked in the “Percentage Evaluation“ operating mode. The reading
quality of 2-D codes can be checked using the reading diagnosis data in normal Reading
mode. On request, the ICR outputs the image buffer memory content as binary or grey scale
bitmap via the Ethernet interface.
To start the reading process when an object is located in the reading area, the ICR requires
a suitable trigger. This opens a time window (“reading interval“) in the ICR for the reading
procedure. In the default setting, this trigger is supplied by an external reading pulse sensor.
Alternative trigger sources include Free Running mode and a command sent via the host
interface.
The current operating status is indicated by four LEDs. A beeper indicates the status of the
reading result. In the default setting, the “Good Read“ function is selected for this.
If the trigger is supplied externally by a sensor, the “Sensor 1“ switching input signals the
start of the reading procedure to the ICR. The “Sensor 2“ switching input is used to signal
the end of reading procedure or to enter a path increment. Alternatively, it can be used
e. g. to teach in a match code. The “Result 1“ and “Result 2“ switching outputs can be
assigned various functions and trigger external devices, such as a PLC. Using the “Result 1“
output, an external field illumination can additionally be triggered or directly powered (max.
100 mA) by the ICR.
The ICR is operated and configured via the serial auxiliary interface (auxiliary data interface)
with the user interface of the “CLV Setup“ software or via the host/auxiliary interface and
command strings. Alternatively, the Ethernet interface is available for both types of
operation.
The ICR outputs system and error messages as error codes, which can be requested from
the error memory using command strings.
The ICR is operated and configured as described above. A variety of parameter options
allow you to adapt the device to a wide range of applications.
The following can be defined:
•The configuration of the code types to be read
•The reading, evaluation and output characteristics
•The communication parameters of the host interface, CAN or Ethernet interface
•The structure of the data output string for “Good Read“ and “No Read“ (host interface)
•The function of the auxiliary interface
Chapter 10.3 “Installation and operating instructions for the PC-based “CLV Setup“
program“, Page 10-3 describes the procedure for installing the “CLV Setup“ program and
explains how to use it. Configuration (parameterizing) is explained in Chapter 6.3.2
“Configuring the ICR for the application with the Setup Assistant“, Page 6-4.
3.3.2Function of the LEDs
General LEDs
Four LEDs indicate the operating status, activity of the laser diode, the reading result status
and data transfer on the host interface. The LEDs (Fig. 3-4) are located on the rear of the
device toward the bottom. Table 3-2, Page 3-9 shows the meaning of the LEDs in the
different operating modes/functions.
Orange• Lights up after power-up if the self-test was successful and the wait time
for reading the profile bar codes (not suitable for ICR) has expired.
Orange• Lights up while the laser diodes for reading the profile bar codes (not
suitable for ICR) are active.
Orange• Lights up constantly
• Extinguishes with new operating mode/function
• Lights up when the laser diodes are active
(The laser diodes are activated/deactivated by reading pulse).
• Lights up constantly in the pulse mode Free Running, as the laser diodes
are always active.
OrangeLED is connected to the “Result 2“ switching output. It indicates the
selected result status for the set pulse duration of the output.
• Lights up after a good read
(default setting: Good Read)
• Lights up (match code comparison active) if the read code matches the
predefined match code(s) and the corresponding result status output is
selected for the “Result 2“ output.
Orange• Flickers while the ICR transfers data to the host on the host
interface.
Orange• Lights up constantly (Free Running mode)
OrangeBehavior depends on the reading quality:
• Extinguishes if reading rate < 30 %
• Flashes twice a second if reading rate is 30 % to 70 %
• Flashes five times a second if reading rate is 70 % to 90 %
• Lights up constantly if reading rate > 90 %
Table 3-2: Meaning of the general LEDs
LEDs of the Ethernet interface
Two LEDs, integrated in the RJ 45 socket 10baseT (Fig. 3-1, Page 3-5), indicate the status
of the Ethernet connection. Table 3-3 lists the meaning of the LEDs.
LEDIndicationFunction
ReadyGreenIndicates the physical Ethernet connection
DataYellowLights up when the ICR receives or sends addressed data via
Table 3-3: Meaning of the LEDs of the Ethernet interface
The beeper uses different tone sequences and lengths (Table 3-4) to signal whether
functions have been executed successfully and whether any malfunctions have occurred.
For information on troubleshooting, seeChapter 8.5 “Troubleshooting“, Page 8-5.
In the default setting, the beeper is activated and indicates the result status “Good Read“
for the Reading mode. It is assumed in these Operating Instructions, that the beeper is
operated with the default setting in Reading mode.
The sound opening of the beeper is located on the rear, narrow side of the device and below
the LED labeling.
Operating modeTone sequenceBeeper function
StartBeep• Signals that the self-test after power-up was successful
Reading modeBeep Beep• Confirms that the device has assumed Reading mode after power
up and that the wait time of 5 s for reading the profile bar codes
(not suitable for ICR) has elapsed
Beep• Confirms a successful read (good read; default setting) and the
reading result output
Exceeding the
illumination timeout
Table 3-4: Beeper function
Tip
Beep Beep Beep• Signals that the illumination LEDs have been deactivated after the
illumination timeout of 10 min (default setting) was exceeded in
Reading mode. The reading interval is still active.
The behavior of the beeper in Reading mode can be changed on the DEVICE CONFIGURATION
tab of the CLV Setup user interface.
ON/OFF:
¾ To switch off the beeper, click on the O
FF radio button in the BEEPER group.
Output function for the result status:
1.Click the R
The R
2.Click the B
ESULT OUTPUT PARAMETERS button.
ESULT OUTPUT PARAMETERS dialog box is displayed.
EEPER list field in the RESULT FUNCTIONS section.
The list containing the available result status functions appears.
3.Click the desired function and choose “OK“.
4.Perform a download to the ICR. This is done by clicking in the toolbar.
The D
OWNLOAD PARAMETER dialog box is displayed.
5.Confirm the dialog box by selecting the P
ERMANENT save option.
The ICR operates the beeper with the values selected for the function of the result status.
•Aligning the ICR with the code (adjusting the reading distance)
•Mounting the ICR
•Installing the CDB 420 or CDM 420 Connection Module
•Connecting the IRC to the CDB 420 or CDM 420 Connection Module
•Adjusting the ICR
•Installing the reading pulse sensor for triggering the reading pulse
NoteDon’t open the device. The producer warranty will be forfeited if the device is opened.
4.2Installation preparations
4.2.1Laying out the components to be installed
•ICR 84x Image Code Reader
4.2.2Laying out accessories
•Mounting bracket no. 2 025 491 with 2 M5 x 16 mm (0.63 in) screws (not included in
the ICR scope of delivery)
– or –
Alternatively, if the bracket is supplied by the user:
Stable installation device that allows the alignment of the ICR to be varied in the x and
y axes. The ICR weighs approx. 900 g (31.75 oz) with the connection cable.
Two M5 screws for the ICR. The screw length depends on the wall thickness of the
bracket used. Depth of engagement in ICR max. 8 mm(0.31 in) from housing surface
•CDB 420 or CDM 420 Connection Module (not included in the scope of delivery of the
ICR)
•Reading pulse sensor(s) for external reading pulse trigger, e. g. photoelectric reflex
switch(es)/photoelectric proximity switch(es) (not included in the scope of delivery of
the ICR)
4.2.3Laying out the required tools
•Two M5 screws for fastening the SICK mounting bracket (strength: 8 mm (0.31 in)) to
the installation base. The screw length depends on the wall thickness of the base
When you select the mounting location, the distance between the ICR and the host and
between the ICR and the code are important.
Distance between ICR and host
The ICR can be installed in a bus connection max. 1,200 m (3,936.96 ft) from the host. The
distance which can be achieved depends on the selected model of the host interface and
the set data transfer rate, however (seeTable 5-5, Page 5-5). Via the Ethernet interface the
ICR can connected to the Ethernet network.
Distance between the ICR and the CDB 420 or CDM 420 Connection Module
The CDB 420 or CDM 420 Connection Module must not be located further than 10 m
(32.8 ft) from the ICR, since the “CLV Setup“ program on the PC accesses the auxiliary
interface of the ICR via this module (RS 232) by default.