SICK AG
Nimburger Str. 11 · D-79276 Reute · Germany
Original Documents
The English version 8010100/YK22 of this document is an original document from
SICK AG.
SICK AG assumes no liability for the correctness of an unauthorized translation.
In case of doubt, contact SICK AG or your local representative.
● Has been stored or operated under irregular conditions.
When safe operation is no longer possible:
▸ Put the GM700 out of operation, separate all connections from the power supply
WARNING: Hot surfaces when enclosure open
Pay attention to hot surfaces on the measuring cell when opening the GM700
enclosure.
1.1.1Electrical safety
and secure against unauthorized start-up.
WARNING: Hazard by voltage
▸ Pay careful attention to power supply connections.
▸ Do not interrupt protective conductor connections.
WARNING: Endangerment of electrical safety by not switching the power
supply off during installation and maintenance work.
An electrical accident can occur when the power supply to the equipment or
the lines is not switched off using a power isolating switch/circuit breaker for
installation and maintenance work.
▸ Before starting work on the equipment, ensure the power supply can be
switched off using a power isolating switch/circuit breaker in accordance
with DIN EN 61010.
▸ Make sure the power isolating switch is easily accessible.
▸ An additional disconnecting device is mandatory when the power isolating
switch is difficult to access or cannot be accessed when connecting the
equipment after installation.
▸ After completion of the work or for test purposes, the power supply may
only be activated again by authorized personnel complying with the safety
regulations.
WARNING: Endangerment of electrical safety through power cable with
incorrect rating
When a removable power cable is used, electrical accidents can occur when
the specifications are not fully observed.
▸ Always observe the exact specifications in the Operating Instructions
(Technical Data Section) when replacing a removable power cable.
1.1.2Grounding the device
10
CAUTION: Equipment damage through erroneous or missing grounding
Ensure the protective grounding of the equipment and/or lines involved during
installation and maintenance work in accordance with EN 61010-1.
▸ Wear suitable protective clothing and mask when using hot and/or aggressive
sample gases and/or with high dust loads.
▸ Never open the enclosure or switch off the purge air feed without taking appropriate
protective measures when the duct is pressurized.
WARNING: Toxic and caustic substances in parts in contact with sample gas
● Depending on the sample gas composition, parts of the GM700 in contact with
sample gas can contain toxic or caustic substances.
● When fitted in the measuring device, the reference cuvette contains the respective
gases to be measured. These gases must not escape.
Special care must be taken with the GME700-2 (HF) because this cell not only
contains gaseous HF but also a certain amount of liquid (max. 0.1 g HF for 8 cm
cell). HF / hydrofluoric acid is extremely toxic and caustic. It is imperative to avoid
inhaling or skin contact at all times. Particularly when there are indications of leaks
or defects, wear protective clothing (protective goggles, latex gloves) and carry out
work in a well ventilated room/area. Never open the cuvette under any circumstances. Check replacement cuvettes for any possible transport damage.
● Handle with care when test gases are used:
▸ When handling HF, keep an HF emergency set (including calcium gluconate gel)
available.
▸ Before working on the gas path, ask the operator which gases have been applied to
the GM700.
▸ Ask the operator whether, and how, the gas path of the GM700 has been cleaned.
▸ If necessary, clean the gas path with a suitable method.
▸ In case of doubt, take appropriate protective measures before working on the gas
path: Ensure adequate ventilation at the workplace or work under a vent. Wear
protective goggles or a safety mask, protective gloves and acid-proof protective
clothes.
SAFETY INFORMATION1
WARNING: Risk of fire through hot gas escaping in installations with
overpressure conditions
On installations with overpressure, the purge air hose can be severely
damaged by escaping hot gas and can catch fire depending on the
temperature.
On plants with overpressure as well as gas temperatures over 200°C:
▸ Ensure reverse flow is prevented by fitting a (trip) flap or a valve.
▸ Regularly check the functionality of the reverse flow safeguard.
WARNING: Eye injuries possible due to laser radiation
Observe all relevant information in these Operating Instructions.
NOTE: Deviations of EN60825-1:2014-08 “Laser safety” from the Directive
2006/25/EC
Due to the Implemented health and safety regulation for artificial optical radiation
concerning the Directive, deviations of EN60825-1:2014-08 “Laser safety” from
Directive 2006/25/EC have occurred.
▸ Always observe the user information in the Annex of these Operating Instructions!
“User information “Laser safety””, page 97.
The GM700 laser warning sign is located on the sender/receiver unit.
Fig. 1: Laser warning sign on the GM700 sender/receiver unit
The invisible laser beam within the SR-unit is not accessible when fitted. There is no risk for
the human eye when looking into the optic visor on the right side of the SR-unit.
CAUTION: Observe the following for the GM700 HF version:
As the limit values of laser class 1 of the IEC 60825-1:2014-5 are neither exceeded
in single case malfunctions nor normal operation under consideration of the service
and measuring state, the device is classified as laser class 1.
Requirement: Use of device software 9105060_YEK0 or a newer software version
with equal functionality.
CAUTION: Important information for safe handling of the GM700 SR-unit HF
version
For device software versions older than 9105060_YEK0, a higher laser radiation
might occur.
▸ Laser class 1 of the IEC 60825-1:2014-5: Only use device software
9105060_YEK0 or a newer software version.
CAUTION: Important information for safe handling of the GM700 SR-unit
(all versions)
▸ Never look directly into the output lens when opening the SR-unit or the reflector.
▸ Always wear laser goggles or switch the device off during maintenance work.
▸ Observe valid national limit values and respective standards for work safety.
Laser capacity
The wavelength of the laser varies for the respective measuring components (NH
HCl).
The maximum laser output capacity on the optical interface (front window) depends on the
respective measuring component and is shown on the laser warning sign.
–HF: “Laser warning sign on the GM700 sender/receiver unit”, page 12.
–NH
GM700 measuring system configurations demand immediate or short-term measures to
protect the measuring system should the purge air supply fail.
Measures for purge air feed failure
▸ Refer to Purge Air Unit Operating Instructions.
1.2Intended use
1.2.1Purpose of the device
The gas analyzer GM700 measures the sample gas component concentration, e.g. HF,
NH3 or HCl in a gas mixture (sample gas). For this purpose, the GM700 analyzer is
assembled at the measuring location and measures directly onsite (in-situ measurement).
!▸
Do not use the device to expose persons to radiation.
1.3Responsibility of user
Intended users
The GM700 may be operated by competent persons only who, based on their devicespecific training and knowledge of the device as well as knowledge of the relevant
regulations, can assess the tasks given and recognize the dangers involved.
Correct use
▸ Use the device only as described in these Operating Instructions. The manufacturer
bears no responsibility for any other use.
▸ Carry out the specified maintenance work.
▸ Do not remove, add or change any components in or on the device unless such changes
are officially allowed and specified by the manufacturer. Otherwise
– the device may become dangerous
– the manufacturer’s warranty becomes void
– the approval for usage in potentially explosive atmospheres becomes void.
Special local requirements
Follow local laws, regulations and company-internal operating directives applicable at the
installation location.
Responsibility for dangerous substances
WARNING: Mortal/health danger as a result of a gas path leakage
When the device is used to measure toxic gases: A leak, for example in the purge air
supply, can be an acute hazard for persons.
▸ Take sui tab l e sa fet y me asu res .
▸ Make sure these safety precautions are followed.
Safety precautions examples:
● Marking the device with warning signs
● Marking the operating area with warning signs
● Safety-related instruction of personnel who could be in the vicinity of the installation
site.
14
Retention of documents
▸ Keep these Operating Instructions available for reference.
▸ Pass on to a new owner/operator.
1.3.1Disposing of device parts harmful to the environment
DANGER: Possible materials harmful to the environment and health
GM700 components can have small shares of pollutants or hazardous substances
such as low lead content in printed circuit boards. The optional permanent cuvette
contains low concentrations of HF or HCl (according to device configuration, see scope
of delivery).
▸ Dispose of all GM700 components according to local laws, regulations and
company-internal operating directives applicable at the installation location such as:
▸ Printed circuit boards and similar electronic components
▸ Dispose of the permanent cuvette, when fitted, safely because it contain low
concentrations of HCl or HF. Therefore, do not simply destroy the cuvette.
The sender/receiver unit contains the optic and electronic subassemblies of the
measuring system. The gas concentration is captured here and the measured value
determined.
Measuring probe
Measuring probes are available in open design versions with integrated purge air guidance
system (GMP) as well as versions with a gas permeable sinter filter not requiring purge air
(GPP: G
detail”, page 19 onwards.
Purge air unit
To supply purge air to the SR-unit with open measuring probe (GMP) and thus to protect
against contamination and high gas temperatures. Use a certain purge air unit each for the
SR-unit depending on the application. The blower types for the SR-unit are designed
differently depending on the application.
as Permeable Probe). Both versions are described from “Measuring probe in
Further information on the purge air unit
▸ Refer to Purge Air Unit Operating Instructions.
Evaluation unit
The evaluation unit in the GM700 measuring system serves as user interface and is
responsible for measured value processing and output as well as control and monitoring
functions. The EvU can be located in the vicinity of the SR-unit. It can also be located up to
about 1000 meters from the sampling point, e.g. installed in the switch center or
monitoring center, and performs functions such as:
● Output of measured values, computed data and operating states
● Communication with the peripheral equipment
● Output of error messages and other status signals
● Control of automatic test functions and access during service (diagnosis)
Connection cables
Cable typePart No.
Cable (CAN bus) SR-unit – purge air fixture measuring probe, length 0.8 m 2 023 704
● Cable (CAN bus), SR-unit – evaluation unit, length 4 m
● Extension up to 1000 m, with terminal box ; with 24 V supply
[1]
Cable
(only for version Cross-Duct and GMP probe)
2 cables
extension (only for GMP probe)
Cable SR-unit purge air fixture – filter monitor of purge air unit, length 3 m
extension (only for Cross-Duct and GMP probe version)
Tab le 1 : Connection cable Part Nos.
[1] One length included in scope of delivery
[2] Included in scope of delivery
SR-unit purge air fixture – filter monitor of purge air unit, 5 m
[2]
SR-unit purge air fixture - filter monitor of purge air unit, length 2 m
To install the purge air fixtures of the SR-unit and reflector on
the gas duct. The purge air fixture is fitted on the flange and
then the SR-unit is fitted on the purge air fixture. ANSI or DIN
flanges provided by the customer can be used alternatively to
the flanges supplied.
Integrated sensors
All probe versions have a built-in pressure sensor as well as an integrated temperature
sensor PT 1000 that continually measures the medium temperature of the probe in the
active measuring path. The measured data are transferred via the CAN bus interface of the
measuring probe and can then be displayed on the evaluation unit.
EPA conformity
When using a GPP probe, an audit measurement conforming to EPA CFR 40 Part 60 and/
or Part 75 can be carried out with the device fitted as far as usable for the application.
Shortest reaction times and high temperature stability characterize the GMP series
measuring probes. Continuous purge air feed is required for operation. In the current GMP
probe series, the air is outlet into the duct at 90° to the gas flow (directed purge air). The
GMP probe has a locking device on the opening for sample gas that is activated with a
lever on the probe flange.
Fig. 6: GMP measuring probe (with open measuring gap)
2.2.1.2GPP gas diffusion probe in dry or wet version
Temperature sensor PT 1000
Gas permeable ceramic filter
This version is more suitable for higher dust contents because, on GPP probes, dust
particles are separated on the filter element and therefore kept away from the measuring
path. The GPP (G
as Permeable Probe) should also be chosen to allow an EPA-conform
audit measurement as well as for low flow speeds or irregular flow profiles.
Fig. 7: Measuring gap of the GPP measuring probe
PRODUCT OVERVIEW2
Both variants of the GPP differ in the respective filter which provides suitability for different
application areas in accordance with Table “GM700 measuring probes in comparison”,
page 22.
Advantages of GPP probes
GPP probes do not require a purge air supply and are easy to maintain. They are fitted with
an automatically controlled heater to reliably prevent condensate on the optical interfaces.
The electronics for heating control, temperature and pressure measurement are located
reliably protected in a sturdy cast-metal enclosure acting as measuring probe section
between duct flange and SR-unit. Both the electric connections for CAN bus and power
supply as well as the test gas connection with which the audit measurement according to
EPA Guideline CFR 40, Part 60 or Part 75 can be carried out are located on the enclosure
as shown.
This Table shows an overview of the features of the different measuring probes. All
measuring probe versions are compatible with all SR-units. The SR-unit is calibrated to the
respective probe length before delivery.
Measuring probeGMP (open probe)GPP (dry)
Design
Max. gas temperature. 430°C
Gas check according to EPA
specification possible
Purge air supply requiredYes–
Heating on optical interfaces in
the probe
Flow rate of sample gas1...40 m/s< 40 m/s
Suitable for wet sample gasYes–
Maximum duct pressure
Measurable componentsNH
System-response time (T90)≥ 5s≥ 120 s
Duct diameter
Dust concentration<2g/m
Probe lengths available [m]1.0/1.5/ 2.01.0/1.5 / 2.0
Active measuring paths available
[mm]
Tab le 2 : Measuring probes in comparison
[1] The maximum temperature used for the measurement depends on the application.
[2] As from gas temperature 300 °C
[3] As from a gas temperature of minimum 130 °C
[4] Probes with shorter active measuring paths on request (possibly required for higher concentrations or
narrower duct diameters)
[4]
Measuring path open in flow
direction; purge air guidance
with outlet aligned 90° to gas
flow
[1]
Gas diffusion probe with
ceramic filter, for dry sample
gas
–Yes
–Yes, with integrated control
±120 hPa, depending on purge
air supply
, HF, HClNH
3
±120 hPa
[2]
, HCl
3
[3]
> 360 mm> 300 mm
3
act.<30g/m3 act.
250/500/750/1000/1250250/500/750/1000
2.2.1.4Special versions
Apart from the standard probes shown, other versions made from stainless steel type
1.4539 especially resistant to acid are available. Special versions can be manufactured
according to customer demands on request.
2.3GM700 options and accessories
● Terminal b ox for CAN bus with 24 V power supply
The optional terminal box is available for distances between SR-unit and EvU larger than
4 meters via a CAN bus cable provided by the customer. The total length of all the CAN
bus connections in the GM700 measuring system can be up to 1000 m in this case.
● Cover plate for purge air fixture of the sender/receiver unit
To maintain purge air supply in cases where this must remain in operation at the
measuring location when the SR-unit is dismounted.
● Weatherproof cover for SR-unit and purge air unit
Required for outdoor installation - dimension drawings, see “Dimensions of weatherproof
● Chart recorder, single or multi-channel, for measured value recording. Protocol(s) can of
course be produced using customer systems.
● Air heater for purge air supply
For special application conditions to prevent condensate. An air heater is required when
the difference between gas temperature and dew point temperature is too small. The
following practical rule of thumb serves as guideline:
An air heater is recommended when
Gas temperature [°C] – dew point temperature [°C] < abs. humidity [%].
Values are compared without considering the units of measure.
2.4Measuring principle
The light of the laser diode radiates through the sample gas and is then detected by a
photo diode. The laser diode wavelength is set to a single absorption line of the sample gas
component. This absorption line is scanned by modulating the wavelength and then the
transmission signal (relation between signal sent and received) captured by a photo diode.
An appropriate signal evaluation returns the magnitude of the absorption line from which
the gas concentration is then calculated. This method is called Tunable Diode Laser Spectroscopy (TDLS) or Tunable Diode Laser Absorption Spectroscopy (TDLAS).
A laser specially developed for gas analysis is used as light source in the GM700. This laser
diode radiates a wavelength with a narrow line width so that an absorption line can be
reliably scanned. A Peltier element and a temperature sensor fitted in the laser diode
enclosure ensure a precise temperature and therefore exact wavelength stabilization of the
system.
PRODUCT OVERVIEW2
The laser beam from the sender/receiver unit travels through the active measuring path
and then impinges on the reflector on the other side of the gas duct. The beam is then
reflected back to the sender/receiver unit. After the beam has passed through the
measuring path twice, during which laser light absorption specific to the gas occurs, the
light is focused on a receiver optics.
2.4.1Reference cuvette for wavelength stabilization
The GM700 is equipped with:
● A permanently filled cuvette for the adjustment of the analyzer laser diode to the
reference position of the gas absorption line (Line-Locking).
2.4.2Signal evaluation
The optimized algorithms of the GM700 evaluation electronics process the measurement
signal of the receiver element together with the corresponding parameters based on the
TDLS measuring method. This is based on the physical characteristics of the gas
molecules that absorb light energy in certain wavelengths. These optimized algorithms
ensure the concentrations of sample gas components are determined without crosssensitivities to other gases. The differential absorption measurement also eliminate dust
influences.
If you make changes to the specifications ordered for your application or if a device is to be
used for an application different to the one originally planned, please pass this information
onto your local sales representative so that we can determine the application options
under changed conditions and whether new adjustment/parameter settings are required.
The following steps are normally taken before starting-up the measuring system
● Project planning
“Project Planning Checklist”, page 24.
● Initial onsite installation
▸ The following preparatory work normally carried out by the customer is described in
“Initial onsite installation”, page 27:
▸ Flange installation, “Installing the flanges with tube”, page 28.
▸ Preparation for installing the purge air unit, “Installation preparation for the purge air unit
(for GMP probe)”, page 30.
▸ Laying the signal and power supply cables to the measuring point, “Signal and power
supply cables”, page 31.
● Installation preparation for the EvU, “Electrical connection in the evaluation unit EvU”, page 40
▸ Possibly preparation of signal cables for the interface to the peripheral equipment,
“Signal and power supply cables”, page 31.
● Device installation
To allow speedy start-up, the following components are normally installed ready for
operation before the start-up date; see also “Installation”, page 34.
– Purge air unit
– Evaluation unit(s).
● Start-up
The actual start-up is carried out by trained personnel or Customer Service. This work is
described in “Start-up”, page 46. The main activities are adjustment tasks on the
GM700 system related to the application.
3.2Project Planning Checklist
Project planning, step by step
The following Checklist simplifies performing and controlling project planning measures
required before start-up in the correct sequence. Technical data and dimensional drawings
of system components, see “Dimension drawing”, page 85 and following.
▸ For round duct cross-sections: 3 times the duct diameter
▸ For rectangular cross-sections: Hydraulic diameter
▸ If these specifications cannot be met: Inlet section > outlet section,
2
/3 : 1/3; uniform concentration spread whenever possible
e.g.
Determine sampling point
Observe national regulations
such as VDI 3950
Emission sampling
point
Application conditions
Pressure conditions at
the sampling point
▸ Obtain official approval for emission sampling point.
▸ Provide for calibration openings at an easily accessible place.
▸ Ensure the GM700 and calibration probe do not influence each
other; the calibration gland should be located at a minimum
distance of 0.5 m away from the measuring device, upstream
▸ Observe Technical Data for duct/ambient conditions!
▸ Gas temperature above/below dew point (dry/wet)
▸ A fitting location with partial vacuum in the duct is ideal.
▸ For duct pressures > 10 mbar, contact SICK on correct selection of
the purge air blower.
▸ Flange fitting is planned in steel ducts as standard; corresponding
flanges with tube are normally included in the GM700 scope of
delivery.
▸ Stone stacks or ducts with thick walls demand an onsite retaining
plate and, possibly, a longer version of the flange with tube; “Install-
ing the flanges with tube”, page 28.
● “Dimension drawings of flange with tube”, page 89.
▸ Provide an opening of suitable size for the flange tube.
▸ Plan adequate clearance for installation and maintenance activities
for the duct insulation cutout.
▸ Plan clearances for handling the SR-unit
▸ Ensure the ambient temperature for the SR-unit is within the
specified range.
SR-unit
→
p. 25, §3.3.1
Select suitable flange
with tube
Duct openings
selection
▸ For installation outside, plan a weatherproof cover
Tools for start-up and
maintenance
▸ When working on the zero path; prerequisites: Clean ambient
atmosphere free from sample gas; weatherproof: Plan the zero
path or order from SICK.
▸ Plan installation location on the duct in immediate vicinity (5 m) of
the GM700 SR-unit.
▸ Keep purge air hoses to the purge air fixture (SR-unit) as short as
possible (pressure loss approx. 1.2 mbar per meter).
Purge air unit
→
p. 28, §3.3.4
Fitting location
selection
▸ Ensure secure cable laying.
▸ Ensure dry and, whenever possible, dust-free intake air on the
purge air unit, use a preliminary filter as necessary.
▸ The intake air temperature should be between 0 and 55 °C. Heat
the purge air for T < 0 °C; see air heater option ( “GM700 options
and accessories”, page 22).
▸ For installation outside, plan a weatherproof cover
Evaluation unit
→
p. 28, §3.3.6
Determine the fitting
location
Tab le 3 : Project Planning Checklist
▸ Plan assembling the unit at an easily accessible location
▸ Ambient temperature within the specifications, “Technical data
▴ : Marking for fitting position,
pointing in flow direction
Steel pipe
50 x 5
DIN 2391
Purge air unit fastening
Platform
50 mm protrusion for circular duct cross-section
The work described in the following can be carried out by the customer's installers.
Prerequisite is that all the required specifications have been made based on the Project
Planning Checklist.
3.3.1Installation preparation at the sampling point
This Section describes the welding work on the duct including making fixing elements
onsite.
Fig. 8: Installation recommendation: Mounting flange and purge air unit (duct diameter not
representative)
CAUTION: Protective measures at the sampling point
▸ Always shut down the installation before any work on the duct!
▸ Secure parts to be separated, e.g. with wire binding, to prevent damage caused
by falling par ts.
▸ Take appropriate protective measures against hot, explosive gases or toxic gases
that could possibly escape from the duct.
▸ Take all the necessary safety measures during welding work, among others
against the risk of explosion or fire in the duct atmosphere and on the duct
insulation.
▸ If necessary, seal off the mounting flange with a cover securely until device
installation (e,g, for overpressure in duct).
3.3.2Uncovering the duct
▸ Remove any duct insulation to uncover an area of approx. 800 x 1500 mm (W x H) to
prepare the duct for the following work.
▸ Keep the insulation material removed for later refitting and/or provide new suitable
insulation material.
3.3.3Installing the flanges with tube
SICK delivers two flanges with tube with 240 mm total length and 125 mm inner diameter
as standard. A version with 500 mm total length is available for installation locations with
thicker insulation or for stone stacks. Special versions can be manufactured on request.
Onsite flanges, including ANSI flanges, can also be used.
Fig. 9: Standard flange with tube
Reinforcement with junction plates recommended
Due to the weight of the sender/receiver unit, we recommend reinforcing the flange tube
fixture onsite with junction plates.
3.3.4Installation preparation for the purge air unit (for GMP probe)
● Maximum length of cable to measuring device complies with project planning.
Installation of purge air unit , see Operating Instructions of purge air unit.
NOTE: Adequate purge air pressure
▸ During project planning, make sure the purge air pressure is sufficient to force the
purge air into the gas duct. Please contact SICK Customer Service or your local
representative when necessary.
3.3.5Duct insulation
▸ Refit the thermal duct insulation; reinforce the insulation when necessary.
WARNING: Observe the ambient temperature!
The SR-unit of the GM700 is designed for an ambient temperature of max. +50 °C.
Radiant heat on the enclosure surface can, among others, create temperatures higher
than the measured air temperature. Therefore, design insulation and radiation
shielding so that temperature limits are reliably maintained.
3.3.6Installation preparation for the evaluation unit
The installation location for the evaluation unit was defined within the project planning
framework ( “Project Planning Checklist”, page 24). The maximum total cable length of all
CAN bus connections in the GM700 measuring system of 1 000 m was taken into account,
bearing in mind that the closer the device is to the measuring point, the easier it is to use.
Installation location preparation
The evaluation unit has fastening openings in the enclosure for easy fixing.
▸ Based on the EvU dimensional drawing according to “Dimension drawing evaluation
unit GM700: Sheet steel enclosure (dimensions in mm)”, page 88 and/or “Dimension
drawing evaluation unit GM700: Cast metal enclosure (dimensions in mm)”, page 88
ensure enough space is available at the planned installation location to attach these, for
wiring as well as opening the enclosure door.
▸ Drill suitable openings as installation points as required.
The onsite supply and signal cables are laid beforehand to facilitate subsequent
installation and start-up of the GM700 system components. Suitable cable ducts and/or
empty conduits are installed for cables already prefabricated and delivered with the
GM700 system. Suitably qualified personnel or Customer Service connect the prepared
cables to the device during installation and/or start-up.
3.4.1Signal and power supply cables
WARNING: Observe safety information!
● Always observe safety information and relevant safety regulations.
● During all work on electrical equipment, disconnect such equipment from the mains,
check that the equipment is potential free and make sure that no third person can
switch the equipment back on again without authorization. Leave the power supply
switched off during the following device installation.
● The power connection must be made via a circuit-breaker switch.
An installation location in the vicinity of the sampling point is generally selected for the EvU
so that the 4 m CAN bus cable in the scope of delivery is sufficient for cabling without
additional installation effort.
Installation away from the evaluation unit
A terminal box with a 24 V power supply unit can be delivered when the EvU is to be located
at a greater distance from the SR-unit. This is then connected to the SR-unit using the 4 m
CAN bus cable delivered with the measuring system. A customer cable suitable for CAN
bus applications, 6-pole cable (twisted pair wires and shielded), then leads to the EvU. The
total length of the CAN bus connections, including the one to the reflector, may be up to
1000 meters. When performing maintenance or service, it must be possible to deinstall
the EvU temporarily and connect it directly to the SR-unit at the sampling point.
Laying the cables
▸ Provide adequate cable lengths at the connection points.
▸ Whenever possible, do not lay power supply cables immediately next to signal cables.
▸ Protect open ends of preinstalled cables against weather effects until device
installation.
▸ Install separate power supply cables and circuit breakers for:
– Purge air units; additional motor circuit breakers and optional protective phase failure
switches.
– Evaluation unit
WARNING: Take precautions to prevent accidental switching off of the purge
air supply.
▸ Attach a clearly visible warning against accidental switching off the separation
equipment for the purge air unit.
▸ Install easily accessible cable ducts or empty conduits for the prefabricated cables to
those delivered with the system ( “Circuit diagram (probe version)”, page 32) marked
with one or two plug-in connectors . Approx. 2 m cable lengths each should be
available at the sampling point for later maintenance work on the measuring system
when dismounted from the duct.
▸ Lay onsite cables (shown without plugs) according to “Circuit diagram (probe version)”,
page 32.
– Wire cross-section specifications are recommendations from which cables for analog
and binary signals can slightly deviate (not however for the CAN bus connections
and/or power supply cables).
– Start with the system internal connections of the GM700.
Status and signal cables from the EvU to the connection terminals of the customer's
status/message devices can be added later as required.
This Section describes the installation and installation work for the GM700 measuring
system before the actual start-up. Completion of the onsite preinstallation in accordance
with “Preparations”, page 34 is assumed.
4.1.1Checking the scope of delivery
▸ Check the delivery against the belonging delivery note and make sure the complete
measuring system has been delivered as ordered.
▸ Check the specifications on mains voltage and frequency on the type plates of the
GM700 components match the installation conditions, delivery note and the order.
4.1.2Installation prerequisites
The following prerequisites are applicable for the work described in the following:
▸ Plan safe usage/application within the limits defined in “Technical data GM700 probe
system”, page 80.
▸ Compliance with the specifications made during project planning (according to “Project
Planning Checklist”, page 24) and correct performance of onsite preinstallation
according to “Mechanical preparation”, page 47.
WARNING: Power supply OFF!
During the following work, it must be ensured that the power supply to the devices and
cables involved is switched off and can only be activated by the personnel carrying out
the work when the work is completed and/or for test purposes and then under
consideration of valid safety regulations.
4.2.1Information on the SR-unit and measuring probe
The GM700 SR-unit and measuring probe are first fitted on the duct during start-up ( “Fit-
ting the SR-unit on the measuring probe”, page 48) because these components first
require an adjustment away from the gas-carrying duct. To avoid problems during start-up,
the SR-unit and measuring probe must be stored in a dry place free from dust, preferably
at room temperature, until start-up.
CAUTION: Do not fit the SR-unit and measuring probe before start-up
Unfavorable ambient conditions or atmosphere in the measuring channel can
damage the measuring system which prevents start-up.
Apart from that, there is a health risk when opening the duct depending on the pressure,
gas temperature and composition in the sample gas duct.
4.3Installing the purge air units (for GMP probe)
Installation of purge air unit
→ Operating Instructions of purge air unit.
INSTALLATION4
4.3.1Terminal box with 24 V power supply unit (option)
▸ Install the terminal box with 24 V power supply unit in the vicinity of the sampling point.
▸ Secure the enclosure using both mounting holes (∅ 5 mm)
▸ The cable length available from the terminal box to the SR-unit is 4 m.Take the empty
conduits laid for the prefabricated cables during onsite preinstallation into account.
Fig. 12: Fitting the terminal box with power supply unit
The fitting location for the evaluation unit was defined during project planning ( “Project
Planning Checklist”, page 24) and prepared during onsite preinstallation as required.
▸ Make sure the CAN bus connection to the SR-unit selected during project planning is
usable at the planned installation location. The CAN bus connection cable delivered as
standard is 4 m long and serves to connect the evaluation unit directly at the sampling
point.
▸ Ensure easy access without problems. In particular, make sure the swivel door of the
evaluation unit can be opened without hindrance after fitting.
4.4.1Installing the evaluation unit – sheet metal enclosure version
▸ Drill mounting holes ∅7.2 mm (for M8) at the fitting location according to the Drilling
plan.
▸ Attach the evaluation unit at the installation location using the 4 planned fastening
brackets with suitable screws.
Fig. 13: Installing the evaluation unit (sheet metal enclosure version)
4.4.2Installing the evaluation unit – cast-metal enclosure version
117.5
330
235
Swivel door
Mounting holes ∅ 7.2 mm
Mounting surface
The mounting holes are
accessible when the swivel
door is open.
Mounting holes
∅ 7.2 mm
1.
2.
▸ Drill mounting holes ∅7.2 mm (for M8) at the fitting location according to the Drilling plan.
Fig. 14: Mounting holes layout (Drilling plan) to fit the EvU (cast-metal enclosure)
INSTALLATION4
▸ Open the enclosure cover with a control cabinet key and swivel open.
▸ Attach the evaluation unit at the installation location using the 3 planned mounting
holes with suitable screws (M8 x 20).
Fig. 15: Fitting the evaluation unit (cast enclosure)
Onsite preparation for electrical installation has been described in “Preparations for elec-
trical installation”, page 31. The cables laid as described there are now connected to the
system components.
CAUTION: Observe safety information as well as relevant safety regulations!
During all work on electrical equipment, disconnect such equipment from the mains,
check that the equipment is potential free and make sure that no third person can
switch the equipment back on again without authorization.
Electrical connections for purge air unit → Purge air unit Operating Instructions.
4.5.1CAN bus wiring options
As already described in the project planning on “Preparations for electrical installation”,
page 31, the following options are available for wiring the CAN bus connection between SR-
unit and evaluation unit:
● Standard cable, 4 m, prefabricated
● Terminal box with prefabricated 4 m long cable to SR-unit; a cable provided by the
customer is used to connect to the evaluation unit.
Information on selecting a suitable type of wiring can be found under “Circuit diagram
Prefabricated cable with plug,
4 m long, to SR-unit
Cable to evaluation unit
(provided by customer)
Cable to power supply
(provided by customer) (see
safety information)
CAN cable signals
ColorSignal
Pink+24 V
GreyGND
Yel lowC AN- H
GreenCAN-L
BrownCAN-GND
Whiten.c.
WARNING: Endangerment of electrical safety
● The power cables must be adequately dimensioned.
● The power connection must be made via a circuit-breaker switch.
● The grounding cable must be connected.
Wiring in terminal box
Connections in the terminal box are wired as follows:
Fig. 16: Terminal box for CAN bus connection between SR-unit and evaluation unit
▸ Connect bridge (jumper) according to suitable voltage supply (ST2).
▸ Lead CAN cable (provided by customer) through the right PG screw fitting to terminal
strip.
▸ Connect shielding on the PG screw fittings on the enclosure.
▸ Connect wires to terminal strip ST5 as shown in “Evaluation unit connections”, page 40;
check that a twisted pair cable is used for CAN-H and CAN-L. Connect the respective
signals in the EvU and terminal box.
WARNING: Observe connection values for power supply!
The evaluation unit is configured to 230 V AC on delivery.
▸ Plug the respective bridges for 115 and/or 120 V AC as shown on the connection
plate of the evaluation unit.
▸ Ensure the power supply has been installed (see evaluation unit connections) according
to the specifications (observe national requirements), but with the power switched off.
▸ Connect protective conductor (PE) to the terminal on the enclosure floor
▸ Lead the signal cable for inputs and outputs through the PG screw fittings on the EvU
enclosure floor and wire according to “Evaluation unit connections”, page 40.
▸ When using the CAN cable provided by the customer, connect the wires to the “Sensor”
terminal strip. Do not connect +24 V and GND (ground).
Fig. 18: See steps 1 to 4 below Connecting the CAN bus cable to the evaluation unit
WARNING: Burns by touching the cable glands
The temperature on the cable glands can be >60 °C.
Graphic display for measured value
display and navigation
Status LEDs to indicate operating and
malfunction states
Key pad for navigation and
entering data
Key pad for menu selection
5Handling the Evaluation Unit
5.1User qualifications
This Section describes how to operate the GM700 measuring system with the evaluation
unit (EvU). The evaluation unit is available with either a sheet metal enclosure (protection
class IP 65) or a cast-metal enclosure (protection class IP 67). The work described in this
Section can be carried out by qualified customer operating personnel. Setting parameters
does however demand comprehensive knowledge of the measuring system, measuring
technology and specific measuring task.
5.2Operating elements
The evaluation unit of the analysis system serves to display, enter and set parameters and
control functions on the system. The operator panel with the display, status indicators and
key pad is accessible when the enclosure door is opened.
Fig. 19: Evaluation unit display and operating elements (shown with sheet metal enclosure)
42
Arrow keys
Enter
Display in measuring
mode
Navigate, select, scroll or edit menu items, variables, units or digits.
Execute the selected menu contents or commands.
Displays all current measured values (temperature values resp, gas
concentration);
Displays computed values
This Section describes the standard start-up at the end of which GM700 starts measuring
operation.
6.1.1Required qualifications and further prerequisites
Previous training by SICK or a qualified sales partner is recommended for technicians or
engineers planning on performing a start-up. This training provides knowledge so that
participants can recognize and handle situations that demand measures going beyond the
standard procedures described here. Apart from the start-up itself, employees of SICK or
trained sales partners are also able to make recommendations for the actual measuring
operation as well as to define the maintenance interval based on the specific plant
conditions.
Standard start-up
Each individual measuring system is already configured to the individual usage conditions
at the factory and therefore the standard start-up procedure described in this Section can
normally be carried by qualified engineers or measurement technicians without the special
training. Prerequisites are:
● Exact adherence to the application requirements specified in the order
● The possibility, during start-up, to ask the advice of a trained specialist from SICK or the
respective sales partner should special questions arise that go beyond the normal
scope of the standard procedures described
Start-up procedure overview
After the general preparation which mainly deals with checking the work already carried
out, it is then necessary to carry out a zero adjust with the measuring system on a
measuring path free from sample gas, the zero path. This can be done either at the
sampling point or another location, e.g. in a closed room. At the sampling point itself, the
purge air unit and the SR-unit are then put into operation in succession. Finally, the
evaluation unit is switched on and checked; the parameters can then be set for the
individual demands.
6.2Start-up steps overview
Prerequisites for successful start-up:
● Installation conditions match the requirements for the measuring system (temperature,
pressure).
● Sampling point must be accessible without danger or problems.
● All power supply and signal cables installed and connected.
● The system, apart from the SR-unit and purge air fixtures which are attached to the duct
flange later, must be fully installed and wired.
● Purge air supply, when fitted, must be ready for use.
– Measuring range
– Limit values
– Inputs and outputs to be used
Start-up runs in two main steps:
st
● 1
main step: Zero adjust on the measuring path free from sample gas ( “Zero adjust”,
page 52)
nd
● 2
main step: Installation and start-up at the sampling point
( “Evaluation unit start-up”, page 56)
Tools and equipment
▸ Provide the following tools and equipment:
– Personal protective equipment as required, e.g. for hot and/or aggressive sample
gases
– 1 19 mm open-end wrench
– Allen key set
– Insulated screwdriver set for electrical connection work etc.
● Fastening parts included in the GM700 measuring system scope of delivery:
– 4 each M 16 x 60 screws with washers and self-locking nuts to fasten the purge air
fixtures on the duct-side flange with tube
– For each purge air fixture:
3 nuts with washers and each with 10 cup springs to fasten the purge air fixtures on
the SR-unit and/or reflector
– Sealing ring to seal the SR-unit, reflector and purge air fixture connection
● Optical cleaning cloth without detergents, e.g. SICK Part No. 4003353
6.3Mechanical preparation
WARNING: Avoid hazards through sample gases!
To avoid health hazards, the following work step may not be carried out during the preparation described in this Section but first within the scope of the respective descriptions
in the following Sections:
▸ Connect the power supply to the SR-unit
6.3.1Checking the scope of delivery
▸ Check the exterior of the SR-unit and measuring probe to ensure they are not damaged.
▸ Make sure the supply voltages on the type plates of the GM700 device components
comply with the installation conditions.
The supply voltage of the GM700 components can be changed between 115 V and 230 V
on site by the SICK service personnel when necessary.
6.3.1.1Transport safety devices
▸ Remove the transport safety devices shown below as well as any protective stickers,
marked as such, depending on the device version.
The front cover of the SR-unit is clamped between the flange fixture and enclosure. To
remove it, open the lock and swivel the flange fixture up (see Fig.). Keep the transport
safety devices as required.
Cup springs:
10 per attachment, facing each
other (= 5 pairs)
Centering disc
Sealing ring
Device flange of the
measuring probe
Lock nut
Threaded pin of the flange
fixture (fixed)
Measuring probeSR-unit/ flange fixture
For GMP probe: 3±0.5
For GPP probe: 2±0.5
Connection CAN cable
(SR-unit and probe)
SR-unit
Low-pressure monitor
SLV
Temperature sensor
PT 1000
Connection CAN cable to evaluation
unit (via terminal box when necessary)
Fig. 22: Cross-section of the connection fitted between the SR-unit and measuring probe
6.3.3Electrical connections on the sender/receiver unit
▸ If the flange fixture was removed from the SR-unit, attach it again. Insert the hinge
bolt again and close the quick-release fasteners.
Set the lever at the probe flange to “open” to open the closing device of the probe in
sample gas direction. See GMP measuring probe (with open measuring gap)
▸ Connect cables accordingly:
– Connect and secure CAN bus cable between SR-unit and probe.
– Connect and secure CAN bus cable to EvU.
– When necessary, connect the temperature sensor and the low-pressure monitor (to
SLV, when fitted).
▸ Switch the power supply on.
Fig. 23: Connections on the SR-unit with GMP measuring probe
Maintenance Mode
Test Analog Out
Test Analog In
Test Relay
Maintenance
ß back
Mode: On
Maintenance Mode
ß back
Alignment tool lever
1.
2.
6.3.4Optical alignment
6.3.4.1Optical alignment for device version for NH
Alignment of the optical axis is checked using the visor on the right hand side of the SR-unit
enclosure and adjusted using the screws on the mounting flange according to the Ladjustment. “Using the alignment tool”, page 51. The power supply for the evaluation unit
must be switched on.
1 Switch to Maintenance mode
On the evaluation unit:
▸ Press button “maint”
▸ Select “Maintenance Mode” and activate with
▸ Activate Maintenance mode
Mode: On (maintenance mode active)
2 Align the sender/receiver unit
▸ First turn (1.) and then pull down (2.) the lever of the built-
in alignment tool to bring the alignment tool into adjustment position “Using the alignment tool”, page 50.
▸ Look through the visor on the right enclosure side diago-
nally from above at the mirror with the target.
▸ As shown in “Using the alignment tool”, page 51, align
the position of the light spot (red on the HF device version,
green on the NH
the screws of the purge air fixture on the flange so that the light spot appears on the
target in position “Circle”.
device version) of the laser beam from the reflector. To do this, adjust
3
and HF measurement
3
Enter
Fig. 24: Using the alignment tool
▸ Only adjust the screws as shown in “Using the alignment tool”, page 51.
▸ First turn (1.) and then pull down (2.) the lever of the built-in alignment tool to bring
the alignment tool into adjustment position “Using the alignment tool”, page 51.
Adjust the screws of the purge air fixture on the flange so that the measuring channel
brightness reaches the maximum value.The adjustment, including the warming up time,
takes about 2.5 to 5 hours whereby the actual work is completed in less than 30 minutes.
Devices with automatic check cycle perform the first check cycle after the warming up
phase.
The LED display of the device side serves as additional aid for rough setting.
After the evaluation unit has been connected to the power supply, a warm-up time of
approx. 2.5 - 5hours (depending on ambient conditions) is required before zero adjust can
be carried out.
▸ After the warm-up phase, align the optical axis as described in “Optical alignment”,
page 50.
Select menu cal:
● Zero Adjust – manual zero adjust (measuring path free from sample gas)
▸ Activate Calibration mode (button “cal”), execute menu item Zero Adjust.
Confirm the prompt. Zero point adjustment runs and
6.3.5.1Types of zero point checks
also see “Gas connections on the GPP measuring probe for zero gas and/or test gas”,
page 57.
● When removed:
SR-unit installed and aligned on probe in ambient air suitable for zero conditions.
● Devices with GPP type probes can be checked when fitted by feeding suitable zero gas
(nitrogen, clean compressed air in instrument air quality)
the check with test gas. This function can also be executed in intervals, controlled by the
device. During this procedure, the probe can cool down and swivel, resulting in a zero
point error. This effect can be minimized by setting the gas flow to the required amount
and limiting the duration needed for the check. Especially for higher gas temperatures
(from 200°C), the zero gas has to be pre-heated to minimize swiveling. This can be
achieved by feeding the zero gas via a heated sample gas line or passively with a ¼“
metal tube, which is pushed beneath the channel insulation for contact with the hot
duct wall. Here, the pre-heating spiral 2065326 can be used or an installation fitted to
the construction.
Zero Measuring is displayed.
[1]
. The sequence is equal to
[1] ISO8573 Class 2 for particle size, amount of contaminator and oil, H2O dew point
The transition from the outlet of the channel insulation to the connection at the zero gas
inlet or the probe has to be insulated, has to reach at least 100°C, but may not exceed
120 °C on the probe connection.
6.3.6Fitting the weatherproof cover for the SR-unit
The weatherproof cover is used when the measuring system is operated outdoors. It is
available as an accessory.
Fig. 27: Weatherproof cover for the GM700 SR-unit
The weatherproof cover is installed in two steps:
Fig. 28: Weatherproof cover for GM700 SR-unit and reflector unit
1 Installing the installation plate on the flange of the purge air fixture.
▸ Lay the weatherproof cover upside down on the floor.
▸ Open and unhinge the locks on both sides.
▸ Pull the installation plate upwards and remove it from the cover.
▸ Place the cover from above on the installation plate.
▸ Click the locks in and close.
6.3.7Evaluation unit start-up
The evaluation unit is configured with a standard parameter set at the factory and,
therefore, is ready to start measuring operation immediately. With the exception of the
following steps, therefore, no additional measures are required for start-up for standard
applications:
▸ Based on the operator information in “Handling the Evaluation Unit”, page 42, check
whether the measured values are output correctly on the LCD.
If error or warning messages are displayed:
▸ Use the operator information in “Handling the Evaluation Unit”, page 42 as well as the
display of error and warning messages in “Error messages”, page 76 to locate and
rectify the error cause.
If the error cause cannot be rectified with this information, contact the Service department
at SICK or your local sales partner for further advice.
▸ If necessary, parameterize the evaluation unit in accordance with the requirements of
the measuring task and the installation environment “Handling the Evaluation Unit”,
6.3.8.1Gas connections on the GPP measuring probe for zero gas and/or test gas
Depending on the variant, the GM700 analyzer performs independent zero point
measurements in certain intervals to ensure high precision with very small measuring
ranges. A zero gas/test gas connection and a second gas connection for nitrogen purging
or dynamic zeroing directly on the measuring probe are available.
An overpressure is created in the measuring path of the probe during zero or test gas feed.
If this pressure is high enough, the complete sample gas is drained out of the measuring
path because a gas flow runs in the reverse direction through the probe filter. The gas flow
in the filter must be larger than the diffusion velocity of the gases in order to drain out the
sample gas from the measuring path. The analyzer performs the zero point measurements
automatically during measuring operation. They can also be triggered manually with the cal
button on the evaluation unit. “Menu overview”, page 43 and following.
START-UP6
Fig. 31: Gas inlets on the measuring probe
Equipment
● Gas hoses:
– Teflon hose (PTFE), type 650 P 0707,
– Steel pipe 1.4571,
/4" with cutting-clamping ring connection (Swagelok)
1
/4"
1
/4"
57
6START-UP
Clamping rings
Gas pipe
(open end)
Cap nut
Inner
bushings
1 Connecting the measuring probe
▸ Connect zero gas (instrument air) to 1st gas inlet “Zero
gas”.
▸ Take off the cap nut from the gas inlet.
▸ Place the cap nut and clamping rings on the open end
of the gas pipe.
▸ Insert the pipe in the open screw fitting.
▸ Tighten the nut with an open end wrench.
▸ Connect zero gas (instrument air) “N
zeroing to 3rd gas inlet as described above.
2 Connecting the instrument air
▸ Connect the pressure reducer to the instrument air supply.
▸ Connect the gas hose from the probe to the pressure regulator outlet.
▸ Set pressure reducer to 250 kPa.
6.3.9Starting operating mode
On the evaluation unit:
▸ Switch Maintenance mode off:
Mode: Off
Start Measuring mode by pressing the meas button.
The analyzer now starts in Measuring mode.
purging or dyn.
2
When using a gas hose, an inner bushing (Part No. 5309138) must be used to support the
hose and clamping ring.
▸ Check for error or warning messages, especially on the
status LEDs.
▸ If error messages are displayed, clear according to “Error
messages”, page 76.
▸ Set parameters, special times and measuring range, “Handling the Evaluation Unit”, page 42.
The device performs temperature stabilization after a change of the temperature range.
This can take approx. half an hour depending on the environment.
● Warning message “DEV TEMP” is output
▸ Wait for half an hour before performing a zero adjust.
On the evaluation unit:
▸ Select a suitable ambient temperature so that the heater of the SR-unit can stabilize the
optics temperatures.
▸ Activate Parameter mode.
▸ Select menu Settings, then Ambient Temp.
▸ Select the temperature range according the Table below that matches the ambient
conditions at the measuring location.
Possible ambient temperature ranges (test cuvette)
The GM700Version with Measuring Probe requires very little maintenance. This Section
describes regular maintenance work to be carried out on the GM700 measuring system.
Qualifications
Inspection and maintenance tasks described in this Section can be carried out by Service
technicians familiar with the device based on the information in these Operating
Instructions and having an in-depth knowledge of the relevant safety regulations.
Maintenance intervals depend on individual application conditions and should be clarified
with SICK Service and/or with a trained engineer or technician at the local sales partner.
If no other specifications have been made, the following recommendation is applicable:
● Maintenance interval – 4 weeks
7.1.1Maintenance protocol
Keep a log of maintenance work done. A simple notebook recording the maintenance
dates, work done, special observations, and required consumables and spare parts is
adequate.
DANGER: Important safety information for all service work
Always observe the following information when carrying out service work to avoid injury
or damage to the measuring system:
▸ Wear suitable protective clothing and a protective mask when the sample gas is hot
and/or aggressive or has a high dust load, or when the sample gas duct is
pressurized. Never open the enclosure or disengage the quick-release locks without
taking suitable protective measures.
▸ If the conditions in the sample gas duct are particularly problematic and hinder or
prevent work on the open duct, despite the use of protective equipment, the
maintenance work must be carried out when the sample gas duct is out of service or
after it has been flooded with ambient air.
▸ The purge air supply must run continuously.
▸ If visual inspection of the power supply cable reveals damage to the insulation or
strain-relief clamp, switch off the power supply to the cable in question immediately.
7.2Preparation and general preparatory work
▸ Have the following equipment available for service work:
– At least the following tools: 2 x 24 mm open-end or box wrenches, 1 x 19 mm open-
end or box wrench, Allen key set, insulated screwdriver set for electrical connection
work
– Optical cleaning cloth without detergents, e.g. Part No. 4 003 353
– Distilled water, clean cloths and, if necessary, a dusting brush
– If the duct is pressurized, a suitable protective cover is required for the duct-side
flange with tube.
60
Maintenance tasks for purge air unit→ Purge air unit Operating Instructions.
▸ Check the enclosure of the SR-unit for damage such as cracks.
▸ If a weatherproof cover is used, check its condition.
▸ Clean contaminated GM700 system components.
▸ Inspect cables thoroughly for damage and pay particular attention to any signs of
abrasion or bending at cable ducts. Prefabricated cables are available as spare parts.
7.3.2Cleaning the front window on the SR-unit
WARNING: Take care when the duct is pressurized!
▸ If the duct is pressurized, you must take the appropriate protective measures, in
particular wear a protective mask.
▸ Prepare a suitable cover for the opening on the device flange of the SR-unit and keep
it available.
▸ Fit the cover immediately on the device flange opening after swiveling the SR-unit
open.
CAUTION: Important information for safe handling of the GM700 SR-unit HF
version
For device software versions older than 9105060_YEK0, a higher laser radiation
might occur.
▸ Laser class 1 of the IEC 60825-1:2014-5: Only use device software
9105060_YEK0 or a newer software version.
▸ Never look directly into the exit lens when opening the SR-unit or the reflector.
▸ Always wear laser goggles or switch the device off during maintenance work.
▸ Observe valid national limit values and respective standards for work safety.
MAINTENANCE7
NOTE: Deviations of EN60825-1:2014-08 “Laser safety” from the Directive
2006/25/EC
Due to the Implemented health and safety regulation for artificial optical radiation
concerning the Directive, deviations of EN60825-1:2014-08 “Laser safety” from
Directive 2006/25/EC have occurred.
▸ Always observe the user information in the Annex of these Operating Instructions!
“User information “Laser safety””, page 97.
▸ Switch the GM700 off.
▸ Open the SR-unit enclosure with a control cabinet key and swivel open.
▸ Check the optical interface (front window) on the SR-unit for contamination and, if
necessary, clean using an optical cleaning cloth. Do not use any detergents because
these leave invisible residues that could falsify the measuring result. The cleaning cloths
can be moistened with distilled water if necessary.
▸ Close and lock the enclosure again to protect the cleaned optical interface against
The evaluation unit is designed for maintenance-free operation over the entire service life
of the measuring system. If the evaluation unit is mounted outdoors, the following simple
checks should be carried out regularly due to the load resulting from changing weather
conditions:
▸ Visual inspection
– Is the enclosure undamaged and the fitting intact?
– Does the enclosure front door open and close easily?
– Is the enclosure window free from moisture?
▸ Does the illuminated LC-Display on the evaluation unit function correctly?
▸ Open the evaluation unit door and check the following:
– Are the cable connections OK?
– Is the enclosure dry inside?
▸ If any one of these points is negative, clarify the cause when possible.
▸ Carry out the necessary repairs.
If the evaluation unit is damaged (e.g. LC-Display failure):
▸ Contact SICK Service or the local sales representative.
Exposed electrical connections
The electrical connections are exposed when the evaluation unit door is open.
Follow the relevant safety regulations
7.5Controlling and tracking the laser working point during measurement
The working point of the laser needs to be tracked when the concentration in the sample
gas is not sufficient. This is done automatically with the help of the fitted reference cell.
7.5.0.1Selecting the ambient temperature range
NOTE: Change of temperature range
The device performs temperature stabilization after a change of the temperature range.
This can take approx. half an hour depending on the environment.
● Warning message “DEV TEMP“ is output
▸ Wait 30 min. before performing filter box measurement.
On the evaluation unit:
▸ Select the appropriate ambient temperature to allow the heater of the SR-unit to
stabilize the optics temperature.
–Activate Parameter mode
– Select Settings and then the Ambient Temp. menu.
– Select the temperature range according the Table below that matches the ambient
conditions at the measuring location.
Possible ambient temperature ranges (test cuvette)
▸ Position the filter box with adapter plate on the SR-unit:
▸ Insert the filter box in the SR-unit hinge
▸ Insert and fasten the bolt
▸ Swivel the filter box on the SR-unit in and lock with the locking device
▸ Fit the reflector for GM700 on the filter box.
Fig. 36: Installing the filter box on the measuring device
7.5.4Carry out filter box measurement
▸ Switch the filter box on; the warm-up phase takes about 2.5 hours.
▸ Switch on the zero gas pump or connect a different zero gas.
▸ Switch all chamber valves to “Zero gas” and the Measure/Purge valve to “Purge”.
▸ Connect test gas; set the primary pressure to approx. 1000 hPa (1 bar).
▸ Note the diameters of individual chambers and/or chamber combinations of the filter
box and the concentration values; see Section 9.6.1, page 96.
▸ At the same time, set the respective filter chambers to “Test gas”.
▸ Switch the Measuring/Purging valve to “Purging” for 2 to 3 minutes (until the
measured value has stabilized) and then to “Measuring”.
▸ The overpressure from the purge phase now dissipates.
▸ When the measured value has stabilized again, read off and note the value.
Restart measuring mode
▸ Purge the filter box with zero gas.
On the EvU:
▸ Exit operating mode “Box measuring” by pressing back.
▸ Disassemble the filter box with plate and reflector from the SR-unit and store safely.
▸ Refit the SR-unit back onto the sampling point in the correct position.
Special characteristics of a measuring component or also the availability of test gases with
the precision required for reference materials require the generation of the span gas from
aqueous solutions with known content. This is the standard case for GM700-2 (HF).
A test gas system with hose pump, scales and evaporator is required in addition to the
heated test cell GMK10 (example: Hovacal).
Requirements
WARNING: Aggressive and corrosive gases
All materials used must be suitable for the measuring components, e.g. HCl, HF.
The manufacturer recommends to exclusively use the evaporator of the test gas
system for this check.
Components required:
● Zero path; Part No. 46483
● Test cell GMK10 with heated sample gas line
● Settings of the test gas system (Hovacal)
Consider the following when entering the nominal value of the concentration (mg/m
− Test cell length of 150 mm
− Test gas temperature depending on setting (120 ...
140 °C)
− Measured value output in mg/m
state relative to an active measuring path length of
1 m (1000 mm)
Input nominal value (in standard state; 0 °C, 1013 hPa) in test gas system (Hovacal):
Example: Test value = 5 mg/m
3
3
in the operating
3
):
The values are displayed in mg/m3 (operation):
Comparison of actual display with entered nominal value:
● Aqueous solution of test component (e.g. 0.01 M for HF)
● Distilled water
● Nitrogen 4.6 or better (compressed gas cylinder)
● Length of hose for safe discharge of the test gas
– The CAN line from the SR-unit to the purge air fixture may not be connected, the
system switches to using the default values for pressure and temperature.
If necessary, disconnect this CAN line. Connect the remaining connections of the
measuring system, “Electrical connections on the sender/receiver unit”, page 49.
Connect the cable for power supply to the GMK10.
Connect the sample gas line from the test gas system.
If necessary, connect the power supply for all GM700 components
Step 2: Settings:
As standard, the temperature of the test cell and the heated sample gas line is
120 °C. Temperatures up to 140°C are possible depending on the application.
Check the temperature on the GMK10.
– Check the temperature indicator on the GMK10 cover.
– Set the desired temperature accordingly when the indicator does not show 120 °C:
– Remove the cover of the GMK10 and set the temperature on the temperature
controller. If required, see separate Operating Instructions of the temperature
It is possible to perform the check of the GM700 with a separate reflector on a zero
tube or a similar construction.
Independent therefrom, the optical alignment is always performed to the probe
position (marking "O, circle" on the target).
WARNING: Hazard by dangerous laser radiation
When the device is operated without probe, dangerous exposure to radiation may
occur.
▸ Always close the reflector when the device is operated without measuring probe.
On the EvU: Enter the temperature chosen for the test cell, e.g. 120 °C, as default
value and enter 1013 hPa as default value for pressure.
Step 3: Set the evaporator temperature on the test gas generator to 180 °C. The test cell is
purged with nitrogen during the warming up phase.
Warming up time: Approx. one hour
Measurement can begin after a warming up time of at least one hour. However, we
recommend to first purge the system by evaporating pure water.
NOTE: Software version UD81 or higher
Software version UD81 or higher must be installed in the GM700 measuring system for
this procedure!
See Section “Menu overview”, page 43 →p. 45, 5.2.1
Step 4: Perform measurement:
When stable conditions exist − to be seen by the trend of the measured values −
activate boxmeasuring mode on the GM700 EvU
– On the EvU: Call up menu cal with the cal button, activate
with
Enter.
– Enter password
“1 2 3 4”.
“Boxmeasuring” and start
The measuring device now starts a zero adjust and then switches to operating mode
“Box measuring”. As for each filter box measurement, the system first performs a
temporary zero adjust and then changes to boxmeasuring mode. The output of the
measured values is performed analog to standard filter box measurement in mg/m
3
(during operation, relative to 1 m active measuring path).
Since the setpoint setting at the test gas system (Hovacal) is "standard", correct the
value comparison (ambient) pressure and (test cell) temperature.
Recommendation
● To prevent pressure build-up in the measuring cell, do not use higher flow
rates than
● Pay attention to the unhindered discharge of the sample gas at the gas
4 l/min. for the measurements.
outlet.
● The standard solution used for the test must be chosen so that the H
concentration of the generated test gas does not exceed 30% by volume
at any test point. Typical value for HF: 0.01M.
Step 5: Terminate the measurement
After termination of boxmeasuring mode (“
back“ button), the conditions for normal
measuring mode are reestablished, e.g. indication of measured value as parameterized.
Terminate the measurement with the “back” button.
Reinstall the GM700 SR and reflector unit to the purge air fixtures at the sampling
point and continue Measuring mode.
NOTE: Purge the measuring system with evaporated distilled water
After the end of the measurements, the system must be purged with vaporized
distilled water until the zero point is reached again!
Before switching off the heaters, the system must be purged with dry nitrogen for a
minimum of 5 minutes.
This Section shows how malfunctions of the GM700 can be recognized, diagnosed and
cleared. It is primarily aimed at the operating personnel responsible for the current
operation of the measuring system as well as maintenance technicians responsible for
clearing malfunctions.
GM700 malfunctions are categorized according to the anticipated effects.
Damage to the measuring system itself
Depending on the installation conditions and measuring system version, a purge air failure
could cause damage to the GM700 .
8.2Purge air failure
Failure of the purge air supply demands measures to be taken immediately or within a
short time, depending on the installation conditions, to protect the measuring system. A
purge air failure, however, rarely occurs in practice. It is still however wise to be prepared
for such an occurrence to prevent damage to the measuring system.
Indications of a potential purge air failure
● Error message on systems that are equipped with a pressure difference sensor
● Purge air unit power supply failure
● Increase in enclosure temperature of the GM700 SR-unit
● Rapid increase in contamination on optical interfaces of the SR-unit
● Hose for the purge air supply to the purge air fixture is loose or visibly damaged
Tools for troubleshooting
● Suitable protective equipment (protective clothing, gloves, etc.) that enable the gas duct
to be opened safely and the SR-unit as well as the reflector removed under the given
installation conditions (hot/aggressive/noxious/dust-laden sample gases, overpressure
in the duct).
● 2 wrenches to remove the SR-unit and reflector and, possibly, other tools required to
restore the power supply.
● Flange cover to seal off the flange opening when the measuring system is removed.
8.3Integrated monitoring and diagnosis system
The GM700 is equipped with an integrated system that
constantly monitors the operating state of the SR-unit and
evaluation unit. Appropriate messages are generated and
logged in the devices for subsequent evaluation should any
deviations from normal operating conditions occur.
Depending on the anticipated effects, the messages of the
system components are categorized into error messages and
warning messages: Warning messages are generated if the
measurement results are not (yet) directly affected by the
change in the system state. Nevertheless, it is important that
the cause(s) be investigated and corrected, e.g. by means of
maintenance measures, to avoid further malfunctions and
damage to the device in particular.
Error messages are generated when measuring operation is no longer possible or no
longer reliable. These warning and error messages are stored in the integrated message
memory of the evaluation unit.
8.3.1Display and retrieval of messages on the evaluation unit
Component/ToolSignalsNote
Operation
Service
Warning
Malfunction
Diagnosis mode
Functional impairment on system
Warning LED goes on
Front panel
Evaluation unit
Malfunction LED blinks
Error Storage TableCall menu Malfunction
Warning Storage TableCall menu Warning
Output
for serious problems
(malfunction, error
message)
[1]The relay is active during normal operation (no malfunctions), i.e. the contact is closed.
Relay 1 inactive
[1]
that will not directly lead to corrupt
measured values.
Functional impairment on system
that can lead to system failure or
restricted function.
Plain-text message(s) for errors that
have occurred to localize and clear
the problem.
See “Troubleshooting Table”.
Plain-text message(s) for pending
warnings
Group malfunction
Troubleshooting Table
If a warning or malfunction is signaled, first call up pending error messages in the Error
menu. Then localize the possible cause and clear the malfunction; see Troubleshooting
Tab le.
Error indicationPossible causesClearance
● LED Malfunction blinks;
(
LED Warning could go on)
● Relay 1:
Group malfunction
74
Plain-text error messages
indicate possible causes
Tab le 1 1: Troubleshooting Table
▸ Start Diagnosis mode (diag):
–Call menu
– Check and clear the specified malfunction.
MEAS: REF CONC
(only for devices with
closed cuvette)
Sensor communication
SIG: DARK VALUE
SIG: K HIGH
SIG: K LOW
SIG: M HIGH
SIG: M LOW
SIG: R HIGH
SIG: R LOW
Z MEAS: RANGE
(only for devices with NH3
measurement)
Table 13: Error messages
Incompatible software (SR-unit)
No absorption line found
Temperature measurement on Peltier
element defective:
● Possible hardware defect
Laser wavelength adjustment outside
allowed range
● Absorption line “lost”
● Laser diode defective
Measuring results from measuring
channel not plausible:
● Incorrect parameter values
● Strong signal interference
● Hardware defect
Measuring results from measuring
channel not plausible:
● Incorrect parameter values
● Strong signal interference
● Hardware defect
Gas concentration in reference cuvette
too low
SR-unit not connected correctly▸ Check CAN connection and repair if necessary
Dark values of receiver element too
high:
● Possible hardware defect
Monitoring channel signal too high▸ Contact Service
Monitoring channel signal too low▸ Contact Service
Measuring channel signal too high▸ Contact Service
Measuring channel signal too low:
● Optical interfaces (front window) of
SR-unit or reflector contaminated
● Optical axes alignment too
inaccurate
● Hardware defect
Reference channel signal too high▸ Contact Service
Reference channel signal too low▸ Contact Service
Unallowed results during zero gas
measurement:
● Zero gas missing
● Optical interfaces (front window) of
SR-unit or reflector contaminated
● Optical axes alignment too
inaccurate
▸ Check software version
▸ Contact Service
▸ Restart device, see above
▸ If not successful, contact Service
▸ Restart device, see above
▸ If not successful, contact Service
▸ Restart device, see above
▸ If not successful, contact Service
▸ Check parameters measuring path, substitute
for
temperature and pressure
▸ Increase parameter Average
Attention: Only by trained personnel!
▸ Restart device, see above
▸ If not successful, contact Service
▸ Restart device, see above
▸ If not successful, contact Service
▸ Exchange reference cuvette
▸ Contact Service
▸ Contact Service
▸ Clean front window, “Cleaning the front window on the
SR-unit”, page 61.
▸ Check optical alignment and correct if necessary,
→
“Optical alignment”, page 50.
▸ If not successful, contact Service
▸ Check zero gas supply and repair if necessary
▸ Clean front window, “Cleaning the front window on the
SR-unit”, page 61
▸ Check optical alignment and correct if necessary,
No detectable sample gas
concentration (see Interval)
Deviation of absorption line
position:
● Line not adjusted perfectly dur-
ing start process
Monitoring measurability of H2O:
● This warning appears when the
gas temperature drops below
the switch threshold set
Gas concentration in reference
cuvette too low; measurement still
possible
Optic of SR-unit out of allowed
temperature range:
● Warm-up phase after switching
on
● Ambient temperature too high
Tab le 1 4: Warning messages
▸ Feed test gas
▸ Wait until the warning message disappears after a few
minutes
▸ Restart device
– Activate Maintenance mode with maint button
– Trigger menu item Reset System or
– switch device off and on again
– If not successful, contact Service
“---” appears on the EvU display. This signals that no value
is available. The warning message disappears again as
soon as the temperature exceeds the switch threshold
▸ Plan and prepare reference cuvette replacement
▸ Wait until the temperature has stabilized
▸ Select different ambient temperature range, “Selecting
Operating voltage:
230 V AC – no jumper
115 V AC – jumper
5 V DC
LEDs light up:
Internal voltage supply
present
24 V DC
LEDs light up when inputs
are connected, i.e. a signal
is pending
Digital inputs,
2 and 3
Analog
output
Relay
outputs
1...4
Fig. 41: Troubleshooting on the evaluation unit: LED displays, signals and fuses
TROUBLESHOOTING AND CLEARING MALFUNCTIONS8
Evaluation unit not responding
▸ Check the power supply on the GM700, check the operating voltage set
▸ Check the fuse in the evaluation unit; check the 24 V/5 V supply indicator in the
evaluation unit, when doing so, remove the plug-in terminal on the cable to the receiver.
▸ If these indicators only light up when the plug-in connector has been removed, check the
cabling first.
Communication fault between evaluation unit and GM700 SR-unit
Error message:
Check following connections:
Sensor Communication???
▸ Connection between evaluation unit and SR-unit.
▸ Cable connection on the plug-in terminal in the evaluation unit.
▸ Cable to SR-unit.
▸ Outer plug-in connectors on SR-unit.
▸ Inner plug-in connectors in SR-unit.
Minimum and maximum measuring range end values are only applicable for single
components. Extremely small or large end values for different components cannot be
combined in all cases.
Concerns the deviation of EN 60825-1:2014-08 “Laser safety” from Directive 2006/25/
EC and/or the health and safety regulation, issued based on the Directive, regarding
artificial optical radiation (“OstrV”)
This product has been classified and manufactured according to the harmonized standard
EN 60825-1:2014-08.
However, this manufacturer classification does not in every case result in the compliance/
non-compliance of the exposure limit values with the European Directive 2006/25/EC.
The limit values for the exposure of workers to artificial optical radiation are determined in
“Directive 2006/25/EC of the European Parliament and of the Council of 5 April 2006 on
the minimum health and safety requirements regarding the exposure of workers to risks
arising from physical agents (artificial optical radiation)” of the European Union. This
Directive has been implemented by the health and safety regulations concerning artificial
optical radiation (OStrV) into national legislation.
The values for maximum permissible exposure (MPE) stated in EN 60825-1:2014-08,
which are used as basis for the risk assessment, are based on the exposure limit values of
an internationally accepted scientific board (ICNIRP), which represents international stateof-the-art technology with its guidelines. These MPE values of the EN 60825-1:2014-08,
however, do not comply with the exposure limit values of the European Directive 2006/25/
EC, as they are higher or lower in certain areas. Both MPE observations are based on the
ICNIRP, the EN 60825-1:2014-08, however, takes the more recent guidelines of the
ICNIRP into account.
ANNEX10
SICK assumes that the limit value concept of Directive 2006/25/EC will be adapted
according to EN 60825-1:2014-08, however, SICK is not aware of the time horizon.
Our product has been manufactured and classified according to EN 60825-1:2014-08; it is
therefore possible that it deviates from the European minimum requirements for protection
against risks due to laser radiation at the workplace.
The following overview shows applications of the product where the exposure limit values
of the European Directive 2006/25/EC can be exceeded under the predictable conditions.
The GM700-02 emits maximum 15 mW in correct operation, when a current software
version is used (version 9105060_YEK0 or a newer version with same functionality),
parameters are set correctly and measured values are valid. This is below the limit values
of laser class 1, according to IEC 60825-1:2014-5 (or EN 60825-1:2014-8) and below the
limit values of EN 60825-1:2007, as well as Directive 2006/25/EC and the limit values of
the health and safety regulations concerning artificial optical radiation (“OstrV”).
When the enclosures of the sender unit and the reflector unit are closed (normal
operation), no optical radiation emerges.
The GM700-02 can emit maximum 25 mW, provided software version 9105060_YEK0
(or a newer version with same functionality) is used, even in case of an unfavorable
parameter choice or a reasonably predictable error case.
This is below the accessible emission limit (AEL = accessible emission limit) of laser
class 1, according to IEC 60825-1:2014-5 (30.3 mW for GM700-02 laser radiation), but
factor 1.6 above the limit value of the European health and safety regulations for
artificial optical radiation or EN 60825-1:2007.
As general precautionary measure for all GM700, the user is requested to avoid looking
into the beam and to switch off the device and thus the laser for all maintenance and
service work.
The following overview shows applications of product GM700-02 where the exposure limit
values of the European Directive 2006/25/EC can be exceeded under the predictable
conditions
Limit values for accessible radiation for the radiation of the GM700-02
EN 60825-1:2007, Directive 2006/25/EC,
IEC 60825-1:2014-5
Limit value laser class 1
λ
worst case
λ
typical
=1272 nm
= 1278 nm
30.3 mW
41.3 mW
Limit value laser class 3R
λ
worst case
λ
typical
=1272 nm
= 1278 nm
155 mW
211.8 mW
Tab le 3 0: Limit values for accessible radiation for the radiation of the GM700-02
Power output of the GM700 in planned operating state (normal operation/maintenance)
Power output GM700
Normal operation, with correct
parameter choice and valid
measurement
15 mW
Correct operation with SW as
from version 9105060_YEK0
Power output GM700
in reasonably predictable single
error case or with incorrect
25 mW
parameter settings with SW as
from version 9105060_YEK0
Remark 1: 10% safety value additionally considered for the measurement uncertainties and
rounded up to 1 mW.
Remark 2: When old software versions are used (prior to 9105060_YEK0), dangerous laser
radiation can be emitted especially during start-up with low ambient temperatures (<0°C).
Tab le 3 1: Power output of the GM700 in planned operating state
Health and safety regulations concerning
artificial optical radiation (“OstrV”)
15.6 mW
15.6 mW
80 mW
80 mW
< AEL of laser class 1 (IEC 60825-1:2014-05)
< AEL of laser class 1 (IEC 60825-1:2007-03)
< AEL laser class 1 (IEC 60825-1:2014-05)
< AEL laser class 3R (IEC 60825-1:2007-03)
(factor approx. 1.6 above LK1 )
10.2Positioning the probe reflector when the probe must be realigned
Fitting the probe:
Vertical measuring gap
Fitting the probe:
Horizontal measuring gap
Only necessary when the measuring probe must be realigned in flow direction (e.g. from
vertical to horizontal direction) but the SR-unit must sit vertically.
Fig. 48: Aligning the measuring probe
ANNEX10
The reflector in the probe end must, in accordance with the probe rotation (e.g. at 90°), be
turned in the opposite direction to this rotation (e.g. –90°) so that the alignment of the
optical beam path to SR-unit is maintained.
▸ Loosen the 3 screws on the probe end (screws with nuts remain on the reflector
enclosure); turn the cover 30° to the right and pull the reflector unit out.
Socket and slotted screw position (view
rotated 180°)
Slotted screw
Screw connection for
reflector holder
Screws
Fig. 49: Reflector position and reflector position marking (socket screw)
100
▸ Loosen the screw connections in the reflector holder and remove it.
▸ Take both holders and the wedge with the graphite seals out.
▸ Take the hollow triple with the triple holder out, position it in the new probe alignment in
the appropriate position (e.g. –90°).
Swap the socket and slotted screws: The socket screw must be rotated in the triple holder
groove.
The socket screw shows the position of the hollow triple. It must always point towards
the connector on the SR-unit so that the alignment is correct.