Siare AM5000 Service manual

AM5000 – AM5000/3 – AM5000/5
Service Manual
AM5000 - AM5000/3 - AM5000/5
AM5000
ANAESTESIA MODULE
SERVICE MANUAL
P.N.: DU5506130 Page 1 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual
GENERAL INDEX
1. USE OF SERVICE MANUAL.................................................................................................................3
1.1 Users of Manual................................................................................................................................3
1.2 Description of use.............................................................................................................................3
1.3 Safety Notes.....................................................................................................................................3
2. MAINTENANCE PROCEDURES...........................................................................................................5
2.1 General Information..........................................................................................................................5
3. TECHNICAL DESCRIPTION.................................................................................................................6
3.1AM5000 AM5000/3 - AM5000/5 anaesthesia module.......................................................................6
3.2 The Pneumatic Distributor................................................................................................................8
4. CALIBRATION AND TESTING............................................................................................................13
4.1 Flowmeter Box................................................................................................................................13
4.2 Final Test........................................................................................................................................24
5. PNEUMATICS......................................................................................................................................29
5.1 Pneumatic Diagram AM5000..........................................................................................................29
5.2 Components List FOR AM 5000/....................................................................................................30
5.3 Pneumatic Diagram AM5000/3.......................................................................................................31
5.4 Components List FOR AM 5000/3..................................................................................................32
5.5 Pneumatic Diagram AM5000/5.......................................................................................................33
5.6 Components List FOR AM 5000/5..................................................................................................34
SIARE is used throughout this manual as an abbreviation for
Via Giulio Pastore, 18 - 40056 Crespellano (BO) - ITALY
Manufacturer of the equipment described in this manual
P.N.: DU5506130 Page 2 of 34 Rev.: 3.0
Warning
SIARE HOSPITAL SUPPLIES S.r.l.
Tel.:+39- 051/969802 Fax: +39-051/969101
AM5000 – AM5000/3 – AM5000/5
Service Manual

1. USE OF SERVICE MANUAL

!

1.1 Users of Manual

This manual is strictly for use by SIARE technicians or qualified technicians authorized by SIARE: The technician authorized by SIARE has available the appropriate tools and spare parts and is trained with regards to the safety of the product. SIARE declines all responsibility with regards to technical assistance to the unit without formal authorization of SIARE. The correct and safe use for the patient and the operator of the unit, requires the knowledge of the instructions and advice written in this manual and the User’s manual.

1.2 Description of use

The symbol the safety of the patient and operator.
The manual describes the unit and its operation with the help of electrical and pneumatic diagrams. The User’s manual is an integral part of this manual, and the technician should have available a copy of the User’s manual and should known its contents before performing any of the operations described in this manual.
placed near an instruction, calls the reader’s attention to important information regarding

1.3 Safety Notes

WARNINGS
The unit has been designed and is manufactured in conditions that guarantee the quality of the product and its components, in order to ensure the maximum level of reliability and the safety of the patient and the operator. Therefore its safety is guaranteed only if it is used as per the instructions contained in this manual and the User’s manual which is an integral and unseparable part of the product documentation for the technical assistance.
For safety reasons, it is necessary to strictly follow the scheduled maintenance described in the User’s manual. The maintenance and the replacement of any parts has to be performed by authorized SIARE service dealers and only original SIARE parts or parts checked by SIARE should be used.
SIARE is not civically or criminally liable in the following cases:
1) Use in conditions and for reasons not stated or prescribed in this manual.
2) Lack or omission of scheduled maintenance as described in this manual.
3) Maintenance performed by personnel not authorized by SIARE.
4) Use of non original spare parts or spare parts not checked and approved by SIARE.
5) Connection with equipment that is not in compliance with relevant directives in effect.
P.N.: DU5506130 Page 3 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual
!
Do not use the equipment in the presence of flammable gases.
!
!
In order to avoid risk of explosion this unit should not be used in the in the presence of flammable anaesthetic gases such as ether of ciclopropane. This unit can operate only with anesthetics that are not flammable as per the directive relative to the electrical safety of the anaesthesia units.
Do not connect the unit to the patient using conductive antistatic tubes.
Due to the fact that this machine cannot be used with flammable anaesthetic agents, such as ether or ciclopropane, the use of antistatic patient tubes is not necessary. Since the use of anitstatic tubes can cause burns when using surgical equipment at high frequencies, their use is not allowed in any administration with this equipment.
The unit is not approved for use in areas with danger of explosion.
The unit cannot be used in the presence of explosive gases.
Before connecting the unit with other electrical equipment not described in this manual, check with the manufacturer.
Regarding electro-medical equipment’s general safety, it is important to follow all rules regarding the interaction between the equipment and the patient, the operator and the nearby environment.
In order to ensure proper and safe use of the equipment it is crucial to follow the instructions in this manual and the User’s manual and pay attention to the notes furnished in this User’s manual. In order to use this equipment, it is vital to know all the instructions in this manual.
The equipment should be inspected and the maintenance performed by SIARE authorized personnel every 6 (six) months. All maintenance performed by SIARE authorized personnel is recorded in the equipment’s maintenance log.
Every repair should be performed by SIARE authorized personnel. Siare is not liable for direct or indirect damage to people or things, due to technical assistance by personnel non-authorized by SIARE or improper use of the equipment, that is to say a use not described in the User’s manual or technical manual.
In order to repair equipment that is malfunctioning, has defects or is broken, the operator should contact SIARE or its authorized local service dealer. It is important, when requesting service, to specify the model and serial number of the equipment.
Use only the recommended accessories. The use of other accessories is authorized only by a written authorization from SIARE as per the safety directives in effect.
The equipment’s operation is authorized only in areas that conform to the safety directives in effect.
P.N.: DU5506130 Page 4 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual

2. MAINTENANCE PROCEDURES

!

2.1 General Information

In order to ensure the safety of the patient and the operator, the unit should undergo an inspection
and test after 800 hours of use or every 6 months, whichever comes first. The inspection and test require a specific knowledge of the unit, and therefore have to be made by specially trained SIARE authorized personnel.
The anesthetist or doctor is responsible for the ordinary maintenance of the unit, as described in this chapter. Cleaning, disinfecting, sterilization and replacement of parts should be done as per the instructions in the user’s manual in order to avoid damage to the equipment that could also endanger the safety of the patient and operator.
The components used have been selected after technical and comparative tests in the designing phase of the machine. Furthermore, the same components are always tested during the manufacturing cycle in order to obtain the maximum level of safety and reliability for the operator and patient.
Therefore whenever a part needs to be replaced, it must be an original spare part, which has been checked and tested by SIARE.
SIARE assumes responsibility for all provisions of the law, if the unit is used and maintained as per the instructions in this manual and the technical manual. The Technical Assistance Report, signed by the authorized SIARE technician, is proof of the completion of the scheduled maintenance.
P.N.: DU5506130 Page 5 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual

3. TECHNICAL DESCRIPTION

ILOST
PQ V
AM5000
ILMNOST
U
P
Q
R
V
I
L
M
N
OST
U
P
QR V
AM5000/3
AM5000/5
W
X

3.1AM5000 AM5000/3 - AM5000/5 anaesthesia module

AM5000 - AM5000/3 – AM5000/5 regulates the supply and concentration of the gas mixture (Air, O2, N2O), and also supplies it to the anaesthetic gas vaporizer.
It also allows the selection of the gas mixture to be supplied (Air-O2, or N2O-O2) and the enrichment (with oxygen) of the supplied gas mixture for emergencies. The AM5000 - AM5000/3 – AM5000/5 is also equipped with the MIX-LIFE device that guarantees a minimum of 25% oxygen in all open-tap conditions.
By using the three gauges located on the front panel, the medical gases supply pressure coming from the central gas installation can be continuously controlled (precision ± 10%). The flowmeters permit the measurement of the flow the of the corresponding gases with a precision of ± 10% of the displayed value or ± 1% of the bottom of the scale, by choosing the highest of the two values.
I OXYGEN supply gauge. L NITROUS OXIDE supply gauge. M AIR supply gauge. N Selector switch for AIR / NITROUS OXIDE. This safety device:
avoids the contemporaneous supply of air and nitrous oxide only the selected gas will be available on the flowmeters.
O OXYGEN BY-PASS key. By pressing this key pure oxygen is released into the anaesthesia circuit with a
flow of approximately 60 l/min. P OXYGEN flow regulator for fresh gases. It opens counter-clockwise. Q NITROUS OXIDE flow regulator for fresh gases. It opens counter-clockwise.
P.N.: DU5506130 Page 6 of 34 Rev.: 3.0
Front view
AM5000 – AM5000/3 – AM5000/5
Service Manual
The opening of this regulator automatically supplies a flow of oxygen of about 25% of the total mixture.
A
B
D
E
H
G
F
I
AM5000
A
B
C
D
E
H
G
F
I
AM5000/3 AM5000/5
The flow of oxygen can be viewed on flowmeter (S). This safety device (MIX LIFE) avoid the incorrect
administration of hypoxic mixtures.
R AIR flow regulator for fresh gases. It opens counter-clockwise. S OXYGEN flowmeter (Max flow 12 l/min). In the model for low flows, there is a second flowmeter (W)
with a lower scale of 1 liter. T NITROUS OXIDE flowmeter (Max flow 12 l/min).. In the model for low flows, there is a second
flowmeter (X) with a lower scale of 1 liter. U AIR flowmeter (Max flow 12 l/min). V Selector switch for the exit of fresh gases (BREATHING SYSTEM or TO AND FRO).
Back View
A OXYGEN inlet from the main medical gas installation. B NITROUS OXIDE inlet from the main medical gas installation. C COMPRESSED AIR inlet from the main medical gas installation. D OXYGEN inlet from the cylinder’s pressure reducer. E NITROUS OXIDE inlet from the cylinder’s pressure reducer. F AIR/OXYGEN outlet for the supply of the lung ventilator, of the active scavenging system and the
tracheal aspiration unit.
In optimal conditions these connector supply COMPRESSED AIR.
If there is no COMPRESSED AIR or there is not enough pressure, the unit automatically switches the
distribution to OXYGEN.
These connectors are supplied standard with the machine if the relative accessories are in use G Emergency OXYGEN outlet. H Exit of fresh gases connector for TO and FRO. I FRESH GAS outlet for the Breathing System.
P.N.: DU5506130 Page 7 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual

3.2 The Pneumatic Distributor

3 3
1 2
1 2
2 4
1 2
1 2
2 4
3 2
1 2
3 2
8
1 1
1 2
1 0
6
2 8
9
2
5
1
3
2
1 5
2
1 0
8
3 0
1 2
2 3
2 8
9
2 8
1 0
1 3
1 3
1 4
1 5
1 5
1 6
1 6
1 7
1 7
1 7
1 8
1 8 1 8
1 8
1 9
1 9
2 0 2 0
2 1
2 2
2 2
2 9
4 5
2 5
2 6
2 7
O 2
N 2 O
A I R
T O F L O W M E T E R
F R O M V A P O R I Z E R
A
B
C
D
1
3
4
132
4 4
4 4
4 4
4 4
4 4
NOTE: THE HERE BELOW INDICATED SCHEMES AND THE PROCEDURES, HAVE BEEN UNIFIED FOR THE THREE MODULES AM5000 – AM5000/3 – AM5000/5. REMEMBER: DO NOT CONSIDER THE AIR LINE FOR ANAESTHESIA MODULE A 2 FLOWETERS.
Below is the pneumatic diagram which describes the unit:
The main purpose of the pneumatic distributor is to:
1. take the medical gases form the pipeline connections or cylinder connections.
2. monitor the operating pressure (pressure gauges)
3. select the type of mix that should be administered to the patient (gas selector AIR-N20)
4. take the selected gases to the flowmeter box
5. allow the mixed gases to be taken to the patient ventilation system (breathing system)
6. enrich the oxygen mixture in case of need (O2 bypass)
All the supply inlets are protected by a non-return valve, in order to avoid a leak of gas in case there is no supply.
P.N.: DU5506130 Page 8 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual
As secondary functions, it allows: supplies an auxiliary outlet of only oxygen which is automatically closed for connections to reanimation bags
for emergencies,
supplies auxiliary outlet for O2 and Air (Aux outlet Air/O2) for devices such as the lung ventilator, ventury
aspirator, etc...
The auxiliary outlets for Air/O2 supply medical air if it has a pressure higher than that of the oxygen minus
0.5 bar. The non-return valves 18 and 26 are necessary in order to avoid that air and O2 mix due to the pressure difference between the two gases’ supply.
Each gas cylinder or pipeline supply conforms to the DISS standard in order to avoid mistakes in identifying the gases.
The gauges have a colored band indicating the range at which there is the correct supply pressure.
The gases supplied to the AM5000 - AM5000/3-AM5000/5 are connected to the flowmeter box in order to regulate the flows and to mix the gases that are then sent to the anaesthetic gas vaporizers. From these they are taken back to the AM5000 - AM5000/3-AM5000/5 in to later be sent to the connector for the fresh gas supply in the patient ventilation circuit. The gases mixed with the anaesthetic can be later mixed with pure oxygen for enrichment of the mix that can be done by the operator by using the bypass button.
The selector enables using two different types of breathing systems without disconnection between the tubes: the first is the patient circuit of the breathing system and the second, is the To and Fro, which is used for manual ventilation of the patient’s breathing apparatus.
P.N.: DU5506130 Page 9 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual
The following diagram illustrates the connections between the parts described above, in this case a
F
LOW
M
ETERO
AIR2N O
2FL
OWMETERFLOWMETE
R
flowmeter box with three flowmeters.
43
40
32
34
42
36
39
41
40
37
35
38
34
40
VAPORIZER
P.N.: DU5506130 Page 10 of 34 Rev.: 3.0
AM5000 – AM5000/3 – AM5000/5
Service Manual
FLOWMETER BOX
FLO
WMETERO
SAFETY
VALVE
TO
VAPORIZER
2
N O
REG.
PRESS.
REG.
PRESS.
2FLOW
METER1234FLOW
M
ETERO
AIR
SAFETY
VALVE
TO
VAPORIZER
2
N O
REG.
PRESS.
REG.
PRESS.
2FL
O
WME
T
ERFLOWM
E
TER
11234
Can be used in the two, three and five flowmeter models.
P.N.: DU5506130 Page 11 of 34 Rev.: 3.0
Loading...
+ 23 hidden pages