Shinko WIL-101-ORP Instruction Manual

Plug-in Type
Digital Indicating ORP Meter
WIL-101-ORP
Instruction Manual
2
Preface
Characters Used in This Manual
Indication
Number, / -1 0 1 2 3 4 5 6 7 8 9
Indication
Alphabet A B C D E F G H I J K L M
Indication
Alphabet N O P Q R S T U V W X Y Z
Caution
• This instrument should be used in accordance with the specifications described in the manual. If it is not used according to the specifications, it may malfunction or cause a fire.
• Be sure to follow all of the warnings, cautions and notices. If they are not observed, serious injury or malfunction may occur.
• The contents of this instruction manual are subject to change without notice.
• Care has been taken to assure that the contents of this instruction manual are correct, but if there are any doubts, mistakes or questions, please inform our sales department.
• This instrument is designed to be installed on a DIN rail within a control panel. If it is not, measures must be taken to ensure that the operator cannot touch power terminals or other high voltage sections.
• Any unauthorized transfer or copying of this document, in part or in whole, is prohibited.
• Shinko Technos Co., Ltd. is not liable for any damage or secondary damage(s) incurred as a result of
using this product, including any indirect damage.
Safety Precautions (Be sure to read these precautions before using our products.)
The safety precautions are classified into 2 categories: “Warning” and “Caution”. Depending on the circumstances, procedures indicated by Caution may result in serious consequences, so be sure to follow the directions for usage.
Warning
Caution
Procedures which may lead to dangerous conditions and cause death or serious injury, if not carried out properly.
Procedures which may lead to dangerous conditions and cause superficial to medium injury or physical damage or may degrade or damage the product, if not carried out properly.
3
Warning
• To prevent an electrical shock or fire, only Shinko or other qualified service personnel may handle the inner assembly.
• To prevent an electrical shock, fire or damage to the instrument, parts replacement may only be undertaken by Shinko or other qualified service personnel.
SAFETY PRECAUTIONS
• To ensure safe and correct use, thoroughly read and understand this manual before using this
instrument.
• This instrument is intended to be used for industrial machinery, machine tools and measuring
equipment. Verify correct usage after purpose-of-use consultation with our agency or main office. (Never use this instrument for medical purposes with which human lives are involved.)
• External protection devices such as protective equipment against excessive temperature rise, etc.
must be installed, as malfunction of this product could result in serious damage to the system or injury to personnel. Also proper periodic maintenance is required.
This instrument must be used under the conditions and environment described in this manual.
Shinko Technos Co., Ltd. does not accept liability for any injury, loss of life or damage occurring due to the instrument being used under conditions not otherwise stated in this manual.
Caution with respect to Export Trade Control Ordinance
To avoid this instrument from being used as a component in, or as being utilized in the manufacture of weapons of mass destruction (i.e. military applications, military equipment, etc.), please investigate the end users and the final use of this instrument. In the case of resale, ensure that this instrument is not illegally exported.
PRECAUTIONS
1. Installation Precautions
Caution
This instrument is intended to be used under the following environmental conditions (IEC61010-1):
• Overvoltage category , Pollution degree 2
Ensure the mounting location corresponds to the following conditions:
• A minimum of dust, and an absence of corrosive gases
• No flammable, explosive gases
• No mechanical vibrations or shocks
• No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 ) that does not change rapidly, and no icing
• An ambient non-condensing humidity of 35 to 85 %RH
• No large capacity electromagnetic switches or cables through which large current is flowing.
• No water, oil or chemicals or where the vapors of these substances can come into direct contact
with the unit
• If the WIL-101-ORP is installed within a control panel, the ambient temperature of the unit – not the ambient temperature of the control panel – must be kept to under 50 . Otherwise the life of electronic parts (especially electrolytic capacitors) of the unit will be shortened.
Note: Do not install this instrument on or near flammable material even though the case of this
instrument is made of flame-resistant resin.
4
2. Wiring Precautions
Caution
• Do not leave wire remnants in the instrument, as they could cause a fire or a malfunction.
• Use a solderless terminal with an insulation sleeve in which the M3 screw fits when wiring the WIL-101-ORP.
• Tighten the terminal screw using the specified torque. If excessive force is applied to the screw when tightening, the terminal screw may be damaged.
• This instrument does not have a built-in power switch, circuit breaker and fuse.
It is necessary to install a power switch, circuit breaker and fuse near the instrument. (Recommended fuse: Time-lag fuse, rated voltage 250 V AC, rated current 2 A)
• For a 24 V AC/DC power source, do not confuse polarity when using direct current (DC).
• Do not apply a commercial power source to the sensor which is connected to the input terminal nor allow the power source to come into contact with the sensor.
• Use the ORP Combined Electrode Sensor in accordance with the sensor input specifications of the WIL-101-ORP.
• Keep the input wires and power line separate.
Note about the ORP Combined Electrode Sensor Cable
The ORP Combined Electrode Sensor cable is a highly-insulated (electrical) cable. Please handle it with utmost care as follows.
• Do not allow terminals and socket of the ORP Combined Electrode Sensor cable to come in contact with moisture or oil of any kind. Likewise, ensure fingers are clean, otherwise the insulation will deteriorate, resulting in unstable indication. Be sure to keep the cable dry and clean at all times. If the cable is stained, clean it with alcohol, and dry it completely.
• For calibration or electrode checking/replacement, the ORP Combined Electrode Sensor cable should be wired with sufficient length.
• Keep the ORP Combined Electrode Sensor cable and junction cable away from electrical devices, such as motors or their power lines from which inductive interference emanates.
Connection
The ORP Combined Electrode Sensor cable has the following terminals.
Symbol Terminal
M Metal electrode terminal R Reference electrode terminal
3. Operation and Maintenance Precautions
Caution
• Do not touch live terminals. This may cause an electrical shock or problems in operation.
• Turn the power supply to the instrument OFF when retightening the terminal or cleaning. Working on or touching the terminal with the power switched ON may result in severe injury or death due to electrical shock.
• Use a soft, dry cloth when cleaning the instrument.
(Alcohol based substances may tarnish or deface the unit.)
• As the display section is vulnerable, do not strike or scratch it with a hard object or put pressure on it.
5
Contents
1. Model Page
1.1 Model ---------------------------------------------------------------------------------- 7
1.2 How to Read the Model Label --------------------------------------------------- 7
2. Names and Functions of Sections ---------------------------------------------- 8
3. Mounting to the Control Panel
3.1 Site Selection ---------------------------------------------------------------------- 9
3.2 External Dimensions (Scale: mm) --------------------------------------------- 9
3.3 Mounting ---------------------------------------------------------------------------- 9
4. Removal -------------------------------------------------------------------------------- 11
5. Wiring ----------------------------------------------------------------------------------- 12
5.1 Lead Wire Solderless Terminal ------------------------------------------------ 13
5.2 Terminal Arrangement ------------------------------------------------------------ 13
5.3 Wire the Communication Line -------------------------------------------------- 14
6. Outline of Key Operation and Setting Groups
6.1 Outline of Key Operation -------------------------------------------------------- 16
6.2 Setting Groups --------------------------------------------------------------------- 16
7. Key Operation Flowchart --------------------------------------------------------- 18
8. Setup
8.1 Turn the Power Supply to the WIL-101-ORP ON ------------------------- 20
8.2 ORP Input Function Group ----------------------------------------------------- 20
8.3 Output Function Group ---------------------------------------------------------- 21
8.4 Special Function Group --------------------------------------------------------- 26
9. Calibration
9.1 Adjustment Mode ----------------------------------------------------------------- 33
9.2 Span Sensitivity Correction Mode -------------------------------------------- 34
9.3 Transmission Output Adjustment Mode ------------------------------------- 35
10. Measurement
10.1 Starting Measurement --------------------------------------------------------- 36
10.2 A Output --------------------------------------------------------------------- 36
10.3 ORP Input Error Alarm -------------------------------------------------------- 36
10.4 Cleansing Output --------------------------------------------------------------- 37
10.5 Manual Cleansing Mode ------------------------------------------------------ 38
10.6 Transmission Output ----------------------------------------------------------- 38
11. Communication
11.1 System Configuration Example ---------------------------------------------- 39
11.2 Setting Method of the ORP Meter ------------------------------------------- 39
11.3 Communication Procedure --------------------------------------------------- 40
11.4 Shinko Protocol
11.4.1 Transmission Mode ------------------------------------------------------ 41
11.4.2 Command Configuration ------------------------------------------------ 41
11.4.3 Checksum Calculation -------------------------------------------------- 42
11.5 Modbus Protocol
11.5.1 Transmission Mode ------------------------------------------------------ 43
11.5.2 Data Communication Interval ----------------------------------------- 43
11.5.3 Message Configuration ------------------------------------------------- 43
11.5.4 Message Example ------------------------------------------------------- 45
6
11.6 Communication Command Table
11.6.1 Notes about Setting/Reading Command --------------------------- 47
11.6.2 Setting/Reading Command -------------------------------------------- 47
11.6.3 Read Only Command --------------------------------------------------- 51
11.7 ORP Calibration and Transmission Output Adjustment via Communication Command --------------------------------------------- 52
11.7.1 Adjustment Mode --------------------------------------------------------- 52
11.7.2 Span Sensitivity Correction Mode ----------------------------------- 52
11.7.3 Transmission Output Adjustment ------------------------------------ 53
11.8 Notes on Programming Monitoring Software
11.8.1 How to Speed up the Scan Time ------------------------------------- 54
11.8.2 How to Read the Set Value Changes Made by Front Keypad Operation ------------------------------------ 54
11.8.3 Note when Sending All Set Values Simultaneously ------------- 54
12. Specifications
12.1 Standard Specifications ------------------------------------------------------- 55
12.2 Optional Specifications -------------------------------------------------------- 58
13. Troubleshooting
13.1 Indication -------------------------------------------------------------------------- 60
13.2 Key Operation -------------------------------------------------------------------- 60
13.3 Communication ------------------------------------------------------------------ 61
14 Character Tables ------------------------------------------------------------------- 62
7
1. Model
1.1 Model
W I L - 1 0 1 -ORP ,
Input Points 1 1 point
Input ORP ORP Combined Electrode Sensor
100 to 240 V AC (standard)
Supply Voltage
1 24 V AC/DC (*1)
EVT A output (A11, A12, A21, A22)
Option
TA Transmission output (*2)
(*1) Supply voltage 100 to 240 V AC is standard.
When ordering 24 V AC/DC, enter “1” after the input code.
(*2) If the TA option is ordered, the EVT option (A1 output only) will be automatically added.
1.2 How to Read the Model Label
The model label is attached to the left side of the case.
(Fig. 1.2-1)
Model, Option Input
Supply voltage, Power consumption Serial number Manufacturer Factory ID RoHS directive compliant
8
2. Names and Functions of Sections
(Fig. 2-1)
Displays
ORP Display: ORP value or characters in setting mode are indicated in red. Setting Display: Values in setting mode are indicated in red.
Indications differ depending on the selections in [Setting Display indication (p.27)].
Unit labels
Unit label: Attach the user’s unit of ORP Display and Setting Display from the included unit
labels if necessary.
Action Indicators
Power Indicator: When power supply to the instrument is turned ON, the yellow LED is lit. A1 Indicator: When A1 output (Contact output 1) is ON, the red LED is lit. A2 Indicator: When A2 output (Contact output 2) is ON, the yellow LED is lit. T/R Indicator: The yellow LED is lit during Serial communication TX output (transmitting).
Keys
UP key: Increases the numeric value. DOWN key: Decreases the numeric value. MODE key: Selects a setting group. SET/RESET key: Switches the setting modes, and registers the set value.
Light Sensor: Automatically measures and controls brightness of the ORP Display,
Setting Display and Action indicators.
Notice
When setting the specifications and functions of this instrument, connect mains power cable to terminals 13 and 14 first, then set them referring from “6. Outline of Key Operation and Setting Groups” to “8. Setup (pages 17 to 32)” before performing “3. Mounting to the Control Panel (p.9)” and “5. Wiring (p.12)”.
A2 Indicator
ORP Display
Setting Display
A1 Indicator
UP key
DOWN key
Power Indicator
T/R Indicator
MODE key
SET/RESET key
Light Sensor
Unit label
Unit label
9
3. Mounting to the Control Panel
3.1 Site Selection
Caution
Use within the following temperature and humidity ranges. Temperature: 0 to 50 (32 to 122 ) (No icing), Humidity: 35 to 85 %RH (Non-condensing) If the WIL-101-ORP is installed within a control panel, the ambient temperature of the unit – not the ambient temperature of the control panel – must be kept to under 50 . Otherwise the life of electronic parts (especially electrolytic capacitors) of the unit will be shortened.
This instrument is intended to be used under the following environmental conditions (IEC61010-1): Overvoltage category , Pollution degree 2
Ensure the mounting location corresponds to the following conditions:
• A minimum of dust, and an absence of corrosive gases
• No flammable, explosive gases
• No mechanical vibrations or shocks
• No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 ) that does not change rapidly
• An ambient non-condensing humidity of 35 to 85 %RH
• No large capacity electromagnetic switches or cables through which large current is flowing
• No water, oil or chemicals or where the vapors of these substances can come into direct contact with the unit
3.2 External Dimensions (Scale: mm)
(Fig. 3.2-1)
3.3 Mounting
(1) Hook the upper part of the socket on the DIN rail, and mount it (A clicking sound is heard).
(Fig. 3.3-1)
Hook the upper part of the socket on the DIN rail.
Socket (sold separately)
DIN rail
85
3
30
108
79
(29)
3.3
10
Caution
Before inserting the WIL-101-ORP into the socket, wire the unit while referring to Section “5. Wiring” (p.12).
(2) Check that the Lock Release has been lowered.
(Fig. 3.3-2)
(3) Insert the WIL-101-ORP into the socket.
(Fig. 3.3-3)
(4) Fix the WIL-101-ORP and the socket by pushing the Lock Release up.
(Fig. 3.3-4)
When inserting, be careful about
the position of pins and slots.
Lock Release
Lock Release
Check that the WIL-101-ORP and the socket are locked by pushing the Lock Release up.
11
4. Removal
(1) Turn the power supply to the unit OFF.
(2) Pull the Lock Release down, and release the WIL-101-ORP from the socket.
(Fig. 4-1)
(3) Separate the WIL-101-ORP from the socket.
(Fig. 4-2)
(4) Remove the socket from the DIN rail by pulling the socket Lock Release
(at the bottom of the socket) down.
(Fig. 4-3)
Lock Release
Check that the WIL-101-ORP and the socket are unlocked by pulling the Lock Release down.
Lock Release
12
5. Wiring Warning
Turn the power supply to the instrument off before wiring or checking. Working on or touching the terminal with the power switched on may result in severe injury or death due to electrical shock.
Caution
• Do not leave wire remnants in the instrument, as they could cause a fire or a malfunction.
• Use a solderless terminal with an insulation sleeve in which the M3 screw fits when wiring the unit.
• Tighten the terminal screw using the specified torque. If excessive force is applied to the screw when tightening, the terminal screw may be damaged.
• This instrument does not have a built-in power switch, circuit breaker or fuse.
It is necessary to install a power switch, circuit breaker or fuse near the instrument. (Recommended fuse: Time-lag fuse, rated voltage 250 V AC, rated current 2 A)
• For a 24 V AC/DC power source, do not confuse polarity when using direct current (DC).
• Do not apply a commercial power source to the sensor which is connected to the input terminal nor allow the power source to come into contact with the sensor.
• Use the ORP Combined Electrode Sensor in accordance with the sensor input specifications of this unit.
• Keep the input wires and power line separate.
Note about the ORP Combined Electrode Sensor Cable
The ORP Combined Electrode Sensor cable is a highly-insulated (electrical) cable. Please handle it with utmost care as follows.
• Do not allow terminals and socket of the ORP Combined Electrode Sensor cable to come in contact with moisture or oil of any kind. Likewise, ensure fingers are clean, otherwise the insulation will deteriorate, resulting in unstable indication. Be sure to keep the cable dry and clean at all times.
If the cable is stained, clean it with alcohol, and dry it completely.
• For calibration or electrode checking/replacement, the ORP Combined Electrode Sensor cable should
be wired with sufficient length.
• Keep the ORP Combined Electrode Sensor cable and junction cable away from electrical devices,
such as motors or their power lines from which inductive interference emanates.
Connection
The ORP Combined Electrode Sensor cable has the following terminals.
Symbol Terminal
M Metal electrode terminal R Reference electrode terminal
13
5.1 Lead Wire Solderless Terminal
Use a solderless terminal with an insulation sleeve in which an M3 screw fits as follows. The tightening torque should be 0.63 N•m.
Solderless
Terminal
Manufacturer Model
Y-type Nichifu Terminal Industries CO.,LTD. TMEV1.25Y-3S
(Fig. 5.1-1)
5.2 Terminal Arrangement
M, R: Electrode sensor terminals (1-2)
ORP Combined Electrode Sensor
A1: A1 output (Contact output 1) terminals
(9-10) (When the EVT option or TA option is ordered)
A2: A2 output (Contact output 2) terminals
(11-12) (When the EVT option is ordered)
TRANSMIT OUTPUT:
Transmission output terminals (11-12) (When the TA option is ordered)
POWER SUPPLY: Power terminals (13-14)
24 V AC/DC (Enter 1 after the input code.)
RS-485: Serial Communication modular jack
Modular Jack Pin (WIL-101-ORP side arrangement)
(Fig. 5.2-1)
3.2mm
or more
5.
9
mm
or less
4 mm or less
4.8mm
or more
14
WIL-101-ORP (Max 31 units)
CDD or CPP (sold-separately)
CMC-001-3 (sold-separately)
Bottom of WIL-101-ORP
Host computer USB port
5.3 Wire the Communication Line
Connect to the modular jack at the bottom of the instrument, using CDD or CPP (sold separately).
Wiring Example Using a USB Communication cable CMC-001-3 (sold-separately)
(Fig. 5.3-1)
Wiring Example Using a Communication Converter IF-400 (sold-separately)
(Fig. 5.3-2)
Bottom of IF-400, WIL-101-ORP
Host Computer D-sub 9-pin Connector
Shield
Wire
FG
WIL-101-ORP (Max 31 units)
IF-400 (Sold Separately)
CDD or CPP (sold separately)
15
Shield Wire
Be sure to ground only one end of the shield wire so that current cannot flow to the shield wire. If both ends of the shield wire are grounded, the circuit will be closed, resulting in a ground loop.
This may cause noise. Be sure to ground the FG terminal.
Recommended cable: OTSC-VB 2PX0.5SQ (made by Onamba Co., Ltd.) or equivalent
(Use a twisted pair cable.)
Terminator (Terminal Resistor)
The terminator is mounted at the end of the wire when connecting a personal computer with multiple peripheral devices. The terminator prevents signal reflection and disturbance. Do not connect a terminator with the communication line because each WIL-101-ORP has built-in pull-up and pull-down resistors instead of a terminator. Communication converter IF-400 (sold separately) has a built-in terminal resistor.
16
6. Outline of Key Operation and Setting Groups
6.1 Outline of Key Operation
Setting items are divided into groups, and group selection has to be made with keypads. Press the key in ORP Display Mode or Cleansing Output Mode. The unit enters Group Selection Mode. Select a group with the key, and press the key. The unit enters each setting item. To set each setting item, use the or key, and register the set value with the key.
6.2 Setting Groups
Power ON
+ (3 sec) (*2) (*2)
ORP Display Mode or Manual Cleansing Mode
Cleansing Output Mode
(*1) + (3 sec) (*3)
Adjustment Mode
+ (3 sec) (*3)
Span Sens.Correc. Mode
+ (3 sec)
Transmission Output Transmission Output Zero Adjust. Mode Span Adjust. Mode
Group Selection Mode See pages 20-32 for details of each setting item.
ORP Input Function Moving average A output when Group data amount input errors occur
Output Function A11 type A22 OFF delay time Group
Special Function Set value lock Restore time after Group cleansing
(*1) Measurement starts, indicating ORP in the ORP Display, and the item selected in [Setting Display indication (p.27)]
of Special function group in the Setting Display. If power is turned ON again, the last mode (ORP Display Mode or Cleansing Output Mode) from when power was turned OFF will resume.
(*2) If (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)] in the Output Function
Group, the unit will enter Manual Cleansing Mode.
After cleansing action is complete, the unit automatically reverts to Cleansing Output Mode.
Abbreviation
s
:
Sens.: Sensitivity Correc.: Correction Adjust.: Adjustment
17
(*3) The unit cannot enter Adjustment Mode or Span Sensitivity Correction Mode in the following cases:
• When (Lock 1) ( Lock 2) or (Lock 3) is selected in [Set value lock (p.26)] in the Special Function Group
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)] in the Output Function Group, and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
: Available only when the TA option is ordered.
About Key Operation
+ (3 sec): Press and hold the key and key (in that order) together for 3 seconds. The unit will proceed to Manual Cleansing Mode.
+ (3 sec): Press and hold the key and key (in that order) together for 3 seconds.
The unit will proceed to Adjustment Mode.
+ (3 sec): Press and hold the key and key (in that order) together for 3 seconds.
The unit will proceed to Span Sensitivity Correction Mode.
+ (3 sec): Press and hold the key and key (in that order) together for 3 seconds.
The unit will proceed to Transmission Output Zero Adjustment Mode.
or : Press the or key. The unit will proceed to the next setting item,
illustrated by an arrow.
or : Press the or key until the desired setting mode appears.
• To revert to ORP Display Mode or Cleansing Output Mode, press and hold the key for 3
seconds while in any mode.
18
7. Key Operation Flowchart
Power ON
ORP Display Mode or + (3sec) (*1) Manual Cleansing (*1) Cleansing Output Mode mode
+ (3sec) (*2) Adjustment
mode
+ (3sec) (*2) Span Sensitivity
Correction mode
+ (3 sec) Transmission output Transmission output
In ORP Display Mode or Cleansing Output Mode, measurement starts, indicating ORP in the ORP Display and an item selected in [Setting Display indication(p.27)] of Special functiongroup in the Setting Display.
If power is turned ON again, the last mode
(ORP Display Mode or Cleansing Output
Mode) from when power was turned OFF
will resume.
Zero adjust. mode Span adjust. mode
ORP Input Function Output Function Group Group
Moving average A11 type A21 hysteresis type data amount
Input high limit A12 type A21 ON side
Input low limit A21 type A21 OFF side
ORP input A22 type A22 hysteresis type filter time constant
A output when input A11 value A22 ON side errors occur
A12 value A22 OFF side
A21 value A11 ON delay time
A22 value A12 ON delay time
A11 hysteresis type A21 ON delay time
A11 ON side A22 ON delay time
A11 OFF side A11 OFF delay time
A12 hysteresis type A12 OFF delay time
A12 ON side A21 OFF delay time
A12 OFF side A22 OFF delay time
Abbreviation:
Adjust.: Adjustment
19
About Setting Items
Moving average • Upper left: ORP Display: Indicates the setting item characters. data amount • Lower left: Setting Display: Indicates the factory default. • Right side: Indicates the setting item.
Transmission output Setting item in shaded section will be indicated only when the corresponding option is ordered. high limit A2 related items cannot be used if TA option is ordered.
About Key Operation
+ (3 sec): Press and hold , keys (in that order) for 3 sec. The unit enters Manual Cleansing Mode.
+ (3 sec): Press and hold , keys (in that order) for 3 sec. The unit enters Adjustment Mode.
+ (3 sec): Press and hold , keys (in that order) for 3 sec. The unit enters Span Sensitivity Correction Mode
+ (3 sec): Press and hold , keys (in that order) for 3 sec. The unit enters Transmission Output Zero Adjustment Mode.
or : Press the or key. The unit enters the next setting item, illustrated by an arrow.
• To revert to ORP Display Mode or Cleansing Output Mode, press and hold the key for 3 sec while in any mode.
Special Function Group
Set value lock A1 output allocation A2 ORP input error alarm
span when A output
ON Communication A2 output allocation protocol A2 ORP input error alarm
time when A output Instrument number Output ON time ON
when A1 output ON
A2 ORP input error alarm Communication Output OFF time span when A output speed when A1 output ON OFF
Data bit/Parity Output ON time A2 ORP input error alarm
when A2 output ON time when A output
OFF Stop bit Output OFF time
when A2 output ON ORP input error alarm
time unit Transmission output A1 ORP input error alarm high limit A type Number of cleansing
cycles Transmission output A2 ORP input error alarm low limit A type Cleansing interval
Transmission output status A1 ORP input error alarm in Adjustment Mode, Span span when A output Cleansing time Sensitivity Correction Mode ON
Set value HOLD A1 ORP input error alarm Restore time after
time when A output cleansing ON
Auto-light function
A1 ORP input error alarm
span when A output Setting Display OFF indication
A1 ORP input error alarm Indication time time when A output
OFF
(*1) When (Cleansing Output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)] in the Output Function Group, the unit will enter Manual
Cleansing Mode. After cleansing is complete, the unit automatically reverts to Cleansing Output Mode.
(*2) The unit cannot enter Adjustment Mode or Span Sensitivity Correction Mode in the following cases.
If (Lock 1) (Lock 2) or (Lock 3) is selected in [Set value lock (p.26)] in the Special Setting Group
When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)] in the Output Function Group, and cleansing action
is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
20
8. Setup
Setup should be done before using this instrument, according to the user’s conditions:
Setting the ORP input function (input high limit, input low limit), Output function (A11, A12, A21, A22
types), Special function (Communication, Cleansing action, etc.). Setup can be conducted in the ORP Input Function Group, Output Function Group and Special Function Group.
If the user’s specification is the same as the factory default of the WIL-101-ORP, or if setup has already been completed, it is not necessary to set up the instrument. Proceed to Section “9. Calibration (p.33)”.
8.1 Turn the Power Supply to the WIL-101-ORP ON.
For approx. 4 seconds after the power is switched ON, is indicated in the ORP Display. During this time, all outputs are in OFF status, and LED indicators except the Power Indicator go off. After that, measurement starts, indicating ORP in the ORP Display and an item selected in [Setting Display indication (p.27)] in the Setting Display. This status is called ORP Display Mode or Cleansing Output Mode.
8.2 ORP Input Function Group
To enter the ORP Input Function Group, follow the procedure below.
1
Press the Key once in ORP Display Mode or Cleansing Output Mode.
2
Press the Key once.
The unit proceeds to the ORP Input Function Group, and ’Moving average data amount’ appears.
Character Setting Item, Function, Setting Range Factory Default
Moving average data amount 3
• Sets the number of ORP inputs for calculating the moving average.
• Setting range: 1 to 20
Input high limit 1999 mV
• Sets the high limit value for ORP input indication.
• Setting range: Input low limit value to 1999 mV
Input low limit -1999 mV
• Sets the low limit value for ORP input indication.
• Setting range: -1999 mV to Input high limit value
ORP input filter time constant 0.0 sec.
• Sets ORP input filter time constant. If the value is set too large, it affects A output due to the delay of response.
• Setting range: 0.0 to 60.0 seconds
A output when input errors occur Disabled
• If input errors, such as ORP Combined Electrode Sensor burnout or short-circuited occur, A output Enabled/Disabled can be selected. If “Enabled” is selected, A output and A output status will be maintained when input errors occur. If “Disabled” is selected, A output and A output status will be turned OFF when input errors occur.
• If (Cleansing output) is selected in [A type], cleansing action takes priority regardless of this selection.
: Enabled
: Disabled
21
8.3 Output Function Group
To enter the Output Function Group, follow the procedure below.
1
Press the key twice in ORP Display Mode or Cleansing Output Mode.
2
Press the key.
The unit enters the Output Function Group, and ’A11 type‘ appears.
Character Setting Item, Function, Setting Range Factory Default
A11 type No action
• Selects an A11 type.
Note: If A11 type is changed, A11 value defaults to 0.
: No action
: ORP input low limit action : ORP input high limit action : Cleansing output
A11 Action
ORP input low limit action ORP input high limit action
If Medium Value is selected in [A11
hysteresis type]:
A11 ON sides
ON
OFF
A11 value
If Medium Value is selected in [A11
hysteresis type]:
A11 ON sides
ON
OFF
A11 value
If Reference Value is selected in [A11
hysteresis type]:
A11 ON side A11 OFF side
ON
OFF
A11 value
If Reference Value is selected in [A11
hysteresis type]:
A11 OFF side A11 ON side
ON
OFF
A11 value
(Fig. 8.3-1)
A12 type No action
• Selects an A12 type.
Note: If A12 type is changed, A12 value defaults to 0.
: No action
: ORP input low limit action : ORP input high limit action : Cleansing output
• A12 action: See (Fig. 8.3-1).
A21 type No action
• Selects an A21 type.
Note: If A21 type is changed, A21 value defaults to 0.
: No action
: ORP input low limit action : ORP input high limit action : Cleansing output
• A21 action: See (Fig. 8.3-1).
22
Character Setting Item, Function, Setting Range Factory Default
A22 type No action
• Selects an A22 type.
Note: If A22 type is changed, A22 value defaults to 0.
: No action
: ORP input low limit action : ORP input high limit action : Cleansing output
• A22 action: See (Fig. 8.3-1, p.21).
A11 value 0 mV
• Selects an A11 value. See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A11 type].
• Setting range: -1999 to 1999 mV
A12 value 0 mV
• Selects an A12 value. See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A12 type].
• Setting range: -1999 to 1999 mV
A21 value 0 mV
• Selects an A21 value. See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A21 type].
• Setting range: -1999 to 1999 mV
A22 value 0 mV
• Selects an A22 value. See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A22 type].
• Setting range: -1999 to 1999 mV
A11 hysteresis type Reference Value
• Selects A11 output hysteresis type (Medium or Reference Value). See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A11
type].
: Medium Value
Sets the same value for both ON and OFF sides in relation to A11 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to A11 value. Both ON and OFF sides need to be set individually.
A11 ON side 10 mV
• Sets the span of A11 ON side. See (Fig. 8.3-1, p.21). If (Medium Value) is selected in [A11 hysteresis type], the span of ON/OFF side will be the same value.
• Not available if (No action) or (Cleansing output) is selected in [A11 type].
• Setting range: 0 to 200 mV
A11 OFF side 10 mV
• Sets the span of A11 OFF side. See (Fig. 8.3-1, p.21).
• Not available if (Medium Value) is selected in [A11 hysteresis type]. Not available if (No action) or (Cleansing output) is selected in [A11 type].
• Setting range: 0 to 200 mV
23
Character Setting Item, Function, Setting Range Factory Default
A12 hysteresis type Reference Value
• Selects A12 output hysteresis type (Medium or Reference Value). See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A12
type].
: Medium Value
Sets the same value for both ON and OFF sides in relation to A12 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to A12 value. Both ON and OFF sides need to be set individually.
A12 ON side 10 mV
• Sets the span of A12 ON side. See (Fig. 8.3-1, p.21). If (Medium Value) is selected in [A12 hysteresis type], the span of ON/OFF side will be the same value.
• Not available if (No action) or (Cleansing output) is selected in [A12 type].
• Setting range: 0 to 200 mV
A12 OFF side 10 mV
• Sets the span of A12 OFF side. See (Fig. 8.3-1, p.21).
• Not available if (Medium Value) is selected in [A12 hysteresis type]. Not available if (No action) or (Cleansing output) is selected in [A12 type].
• Setting range: 0 to 200 mV A21 hysteresis type Reference Value
• Selects A21 output hysteresis type (Medium or Reference Value). See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A21
type].
: Medium Value
Sets the same value for both ON and OFF sides in relation to A21 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to A21 value. Both ON and OFF sides need to be set individually.
A21 ON side 10 mV
• Sets the span of A21 ON side. See (Fig. 8.3-1, p.21).
If (Medium Value) is selected in [A21 hysteresis type], the span of ON/OFF side will be the same value.
• Not available if (No action) or (Cleansing output) is selected in [A21
type].
• Setting range: 0 to 200 mV
A21 OFF side 10 mV
• Sets the span of A21 OFF side. See (Fig. 8.3-1, p.21).
• Not available if (Medium Value) is selected in [A21 hysteresis type].
Not available if (No action) or (Cleansing output) is selected in [A21 type].
• Setting range: 0 to 200 mV
24
Character Setting Item, Function, Setting Range Factory Default
A22 hysteresis type Reference Value
• Selects A22 output hysteresis type (Medium or Reference Value). See (Fig. 8.3-1, p.21).
• Not available if (No action) or (Cleansing output) is selected in [A22
type].
: Medium Value
Sets the same value for both ON and OFF sides in relation to A22 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to A22 value. Both ON and OFF sides need to be set individually.
A22 ON side 10 mV
• Sets the span of A22 ON side. See (Fig. 8.3-1, p.21).
If (Medium Value) is selected in [A22 hysteresis type], the span of ON/OFF side will be the same value.
• Not available if (No action) or (Cleansing output) is selected in [A22
type].
• Setting range: 0 to 200 mV
A22 OFF side 10 mV
• Sets the span of A22 OFF side. See (Fig. 8.3-1, p.21).
• Not available if (Medium Value) is selected in [A22 hysteresis type].
Not available if (No action) or (Cleansing output) is selected in [A22
type].
• Setting range: 0 to 200 mV A11 ON delay time 0 sec.
• Sets A11 ON delay time.
The A11 output does not turn ON (under the conditions of turning ON) until the time set in [A11 ON delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A11
type].
• Setting range: 0 to 9999 seconds A12 ON delay time 0 sec.
• Sets A12 ON delay time.
The A12 output does not turn ON (under the conditions of turning ON) until the time set in [A12 ON delay time] elapses.
• Not available if (No action) or (Cleansing Output) is selected in [A12
type].
• Setting range: 0 to 9999 seconds A21 ON delay time 0 sec.
• Sets A21 ON delay time.
The A21 output does not turn ON (under the conditions of turning ON) until the time set in [A21 ON delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A21
type].
• Setting range: 0 to 9999 seconds A22 ON delay time 0 sec.
• Sets A22 ON delay time.
The A22 output does not turn ON (under the conditions of turning ON) until the time set in [A22 ON delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A22
type].
• Setting range: 0 to 9999 seconds
25
Character Setting Item, Function, Setting Range Factory Default
A11 OFF delay time 0 sec.
• Sets A11 OFF delay time.
The A11 output does not turn OFF (under the conditions of turning OFF) until the time set in [A11 OFF delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A11
type].
• Setting range: 0 to 9999 seconds A12 OFF delay time 0 sec.
• Sets A12 OFF delay time.
The A12 output does not turn OFF (under the conditions of turning OFF) until the time set in [A12 OFF delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A12
type].
• Setting range: 0 to 9999 seconds A21 OFF delay time 0 sec.
• Sets A21 OFF delay time.
The A21 output does not turn OFF (under the conditions of turning OFF) until the time set in [A21 OFF delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A21
type].
• Setting range: 0 to 9999 seconds A22 OFF delay time 0 sec.
• Sets A22 OFF delay time.
The A22 output does not turn OFF (under the conditions of turning OFF) until the time set in [A22 OFF delay time] elapses.
• Not available if (No action) or (Cleansing output) is selected in [A22
type].
• Setting range: 0 to 9999 seconds
26
8.4 Special Function Group
To enter the Special Function Group, follow the procedure below.
1
Press the key 3 times in ORP Display Mode or Cleansing Output Mode.
2
Press the key.
The unit enters the Special Function Group, and the ’Set value lock‘ appears.
Character Setting Item, Function, Setting Range Factory Default
Set value lock Unlock
• Locks the set values to prevent setting errors.
(Unlock): All set values can be changed.
(Lock 1): None of the set values can be changed. (Lock 2): Only A11, A12, A21, A22 values can be changed. (Lock 3): All set values – except Adjustment value, Span sensitivity correction
value, Transmission output Zero adjustment value, Transmission output Span adjustment value – can be temporarily changed. However, they revert to their previous value after the power is turned off because they are not saved in the non-volatile IC memory. Do not change setting items (A11, A12, A21, A22 types). If they are changed, they will affect other setting items. Be sure to select Lock 3 when changing the set value frequently via software communication. (If a value set via software communication is the same as the value before the setting, the value will not be written in the non-volatile IC memory.)
Communication protocol Shinko protocol
• Selects communication protocol.
: Shinko protocol
: Modbus ASCII mode : Modbus RTU mode
Instrument number 0
• Sets the instrument number. (The instrument numbers should be set one by one when
multiple instruments are connected.)
• Setting range: 0 to 95 Communication speed 9600 bps
• Selects a communication speed equal to that of the host computer.
: 9600 bps
: 19200 bps : 38400 bps
Data bit/Parity 7 bits/Even
• Selects data bit and parity.
: 8 bits/No parity
: 7 bits/No parity : 8 bits/Even : 7 bits/Even : 8 bits/Odd : 7 bits/Odd
Stop bit 1 bit
• Selects the stop bit.
: 1 bit
: 2 bits
27
Character Setting Item, Function, Setting Range Factory Default
Transmission output high limit 1999 mV
• Sets the Transmission output high limit value.
(This value correponds to 20 mA DC output.)
If Transmission output high limit and low limit are set to the same value, the transmission output will be fixed at 4 mA DC.
• Available only when Transmission output (TA option) is ordered.
• Setting range: Transmission output low limit to 1999 mV Transmission output low limit -1999 mV
• Sets the Transmission output low limit value.
(This value correponds to 4 mA DC output.)
If Transmission output high limit and low limit are set to the same value, the transmission output will be fixed at 4 mA DC.
• Available only when Transmission output (TA option) is ordered.
• Setting range: -1999 mV to Transmission output high limit
Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode
Last value HOLD
• Selects Transmission output status in Adjustment Mode or Span Sensitivity Correction
Mode.
• Available only when Transmission output (TA option) is ordered.
• Selection item
: Last value HOLD (Retains and outputs the last value before adjustment or
span sensitivity correction.) : Set value HOLD (Outputs the value set in [Set value HOLD].) : Measured value (Outputs the value measured in Adjustment Mode or Span
Sensitivity Correction Mode.)
Set value HOLD 0 mV
• Sets the Set value HOLD.
• Available only when (Set value HOLD) is selected in [Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode].
• Setting range: -1999 to 1999 mV
Auto-light function Disabled
• Selects Auto-light Enabled/Disabled.
: Disabled : Enabled
Setting Display indication No indication
• In ORP Display Mode or Cleansing Output Mode, selects an item to be indicated in the
Setting Display.
: No indication
: A11 value : A12 value : A21 value : A22 value
Indication time 00.00
• In ORP Display Mode or Cleansing Output Mode, sets the indication time of the
displays from no key operation until displays go off. Displays remain lit when set to 00.00. Displays become lit when any key is pressed while in unlit status.
• Not available if (No indication) is selected in [Setting Display indication].
• Setting range: 00.00 (Remains lit)
00.01 to 60.00 (Minutes.Seconds)
28
Character Setting Item, Function, Setting Range Factory Default
A1 output allocation A11 type
• Selects A1 output allocation.
For A1 output, A11 type, A12 type, A21 type and/or A22 type are allocated. Output is OR output. However, if (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], the Cleansing output will be given priority.
• Available when A output (EVT option) or Transmission output (TA option) is
ordered.
: A11 type
: A12 type : A21 type : A22 type : A11, A12 types : A21, A22 types : A11, A21 types : A12, A22 types : A11, A12, A21, A22 types
A2 output allocation A21 type
• Selects A2 output allocation.
For A2 output, A11 type, A12 type, A21 type and/or A22 type are allocated. Output is OR output. However, if (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], the Cleansing output will be given priority.
• Available when A output (EVT option) is ordered.
Not available if Transmission output (TA option) is ordered.
: A11 type : A12 type : A21 type : A22 type : A11, A12 types : A21, A22 types : A11, A21 types : A12, A22 types : A11, A12, A21, A22 types
29
Character Setting Item, Function, Setting Range Factory Default
Output ON time when A1 output ON 0 sec.
• Sets Output ON time when A1 output is ON.
If ON time and OFF time are set, A1 output can be turned ON/OFF in a configured cycle when A1 output is ON. (Fig. 8.4-1)
• Available when A output (EVT option) or Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds
Timing chart of Output ON time and OFF time while in A1 output ON
(Fig. 8.4-1)
Output OFF time when A1 output ON 0 sec.
• Sets Output OFF time when A1 output is ON.
If ON time and OFF time are set, A1 output can be turned ON/OFF in a configured cycle when A1 output is ON. See (Fig. 8.4-1).
• Available when A output (EVT option) or Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds
Output ON time when A2 output ON 0 sec.
• Sets Output ON time when A2 output is ON.
If ON time and OFF time are set, A2 output can be turned ON/OFF in a configured cycle when A2 output is ON. See (Fig. 8.4-1).
• Available when A output (EVT option) is ordered.
Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds
Output OFF time when A2 output ON 0 sec.
• Sets Output OFF time when A2 output is ON.
If ON time and OFF time are set, A2 output can be turned ON/OFF in a configured cycle when A2 output is ON. See (Fig. 8.4-1).
• Available when A output (EVT option) is ordered.
Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds
A1 ORP input error alarm A type No action
• Selects A type in order to assess A1 ORP input error alarm.
: No action : A11 type : A12 type : A21 type : A22 type
OFF
ON
OFF
ON
A1 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual A1 output turning OFF.
Actual A1 output
30
Character Setting Item, Function, Setting Range Factory Default
A2 ORP input error alarm A type No action
• Selects A type in order to assess A2 ORP input error alarm.
• Not available if Transmission output (TA option) is ordered.
• Selection item : No action : A11 type : A12 type : A21 type
: A22 type
A1 ORP input error alarm span when A output ON
0 mV
• Sets span to assess A1 ORP input error alarm when A output (selected in [A1 ORP input error alarm A type]) is ON. Refer to ‘ORP Input Error Alarm’ on p.31.
• Setting range: 0 to 1999 mV
When set to 0, ORP input error alarm is disabled.
A1 ORP input error alarm time
when A output ON
0 sec.
• Sets time to assess A1 ORP input error alarm when A output (selected in [A1 ORP input error alarm A type]) is ON. Refer to ‘ORP Input Error Alarm’ on p.31.
• Setting range: 0 to 9999 seconds or minutes (Time unit follows the selection in [ORP input error alarm time unit].)
When set to 0, ORP input error alarm is disabled.
A1 ORP input error alarm span when A output OFF
0 mV
• Sets span to assess A1 ORP input error alarm when A output (selected in [A1 ORP input error alarm A type]) is OFF. Refer to ‘ORP Input Error Alarm’ on p.31.
• Setting range: 0 to 1999 mV
When set to 0, ORP input error alarm is disabled.
A1 ORP input error alarm time
when A output OFF
0 sec.
• Sets time to assess A1 ORP input error alarm when A output (selected in [A1 ORP input error alarm A type]) is OFF. Refer to ‘ORP Input Error Alarm’ on p.31.
• Setting range: 0 to 9999 seconds or minutes (Time unit follows the selection in [ORP input error alarm time unit].)
When set to 0, ORP input error alarm is disabled.
A2 ORP input error alarm span when A output ON
0 mV
• Sets span to assess A2 ORP input error alarm when A output (selected in [A2 ORP input error alarm A type]) is ON. Refer to ‘ORP Input Error Alarm’ on p.31.
• Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 1999 mV
When set to 0, ORP input error alarm is disabled.
A2 ORP input error alarm time
when A output ON
0 sec.
• Sets time to assess A2 ORP input error alarm when A output (selected in [A2 ORP input error alarm A type]) is ON. Refer to ‘ORP Input Error Alarm’ on p.31.
• Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds or minutes (Time unit follows the selection in [ORP input error alarm time unit].)
When set to 0, ORP input error alarm is disabled.
31
Character Setting Item, Function, Setting Range Factory Default
A2 ORP input error alarm span when A output OFF
0 mV
• Sets span to assess A2 ORP input error alarm when A output (selected in [A2 ORP input error alarm A type]) is OFF. Refer to ‘ORP Input Error Alarm’.
• Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 1999 mV
When set to 0, ORP input error alarm is disabled.
A2 ORP input error alarm time
when A output OFF
0 sec.
• Sets time to assess A2 ORP input error alarm when A output (selected in [A2 ORP input error alarm A type]) is OFF. Refer to ‘ORP Input Error Alarm’.
• Not available if Transmission output (TA option) is ordered.
• Setting range: 0 to 9999 seconds or minutes (Time unit follows the selection in [ORP input error alarm time unit].) When set to 0, ORP input error alarm is disabled.
ORP input error alarm time unit Second(s)
• Selects ORP input error alarm time unit.
• Selection item : Second(s)
: Minute(s)
Number of cleansing cycles 0 (Continuous cleansing)
• Sets the number of cleansing outputs. (Fig. 8.4-2)
• Available for this setting item and all sebsequent items if (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp. 21, 22)].
• Setting range: 0 to 10 (0: Continuous cleansing)
Cleansing interval 360 minutes
• Sets an interval between cleansings. (Fig. 8.4-2)
• Setting range: 60 to 3000 minutes
Cleansing time 600 sec.
• Sets the cleansing output time during the cleansing output interval. (Fig. 8.4-2)
• Setting range: 1 to 1800 seconds
Restore time after cleansing 600 sec.
• Sets the time to restore units to normal operation after cleansing output. (Fig. 8.4-2)
• Setting range: 1 to 1800 seconds
ORP Input Error Alarm ORP input error alarm is used for detecting actuator trouble. Even if ORP input error alarm time has elapsed, and if ORP input does not become higher than ORP input error alarm span, the unit assumes that actuator trouble has occurred, and writes Status flag 2 (A1, A2 ORP input error alarm output flags). In Serial communication, status can be read by reading Status flag 2 (A1, A2 ORP input error alarm output flags).
ORP input error alarm is disabled in the following cases.
• During Adjustment Mode, Span Sensitivity Correction Mode.
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
• When ORP input error alarm time is set to 0 seconds, or ORP input error alarm span is set to 0 mV
32
Cleansing Output Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing
(Fig. 8.4-2)
Programmed
action
Programmed
action
Programmed
action
33
9. Calibration
Adjustment Mode, Span Sensitivity Correction Mode and Transmission Output Adjustment Mode are described below.
9.1 Adjustment Mode
When using a brand-new sensor, please calibrate in Adjustment Mode.
By setting the adjustment value, calibrates ORP value indicated on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The unit cannot enter Adjustment Mode in the following cases:
• When (Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.26)]
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines the procedure for calibration.
(1) When selecting (Last value HOLD) in [Transmission output status in Adjustment Mode,
Span Sensitivity Correction Mode (p.27)], select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
(2) Press and hold the and keys (in that order) together for 3 seconds in ORP Display Mode
or Cleansing Output Mode.
The unit enters Adjustment Mode, and indicates the following.
Display Indication
ORP Display and ORP value are indicated alternately. Setting Display The adjustment value is indicated.
(3) Immerse the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential
difference 260 mV).
(4) Set an adjustment value with the or key so that ORP value is approximately 260 mV (at 20 ).
For other temperature and electrical potentials, refer to the temperature characteristics of your standard solution. Adjustment range: -200 to 200 mV
(5) Press the key.
Adjustment Mode is complete, and the unit reverts to ORP Display Mode or Cleansing Output Mode.
34
9.2 Span Sensitivity Correction Mode
When calibrating periodically, please calibrate in Span Sensitivity Correction Mode.
By setting the Span sensitivity correction value in percentage, calibrates ORP value indicated on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The unit cannot enter Span Sensitivity Correction Mode in the following cases:
• When (Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.26)]
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines the procedure for calibration.
(1) When selecting (Last value HOLD) in [Transmission output status in Adjustment Mode,
Span Sensitivity Correction Mode (p.27)], select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
(2) Press and hold the and keys (in that order) together for 3 seconds in ORP Display Mode
or Cleansing Output Mode. The unit enters Span Sensitivity Correction Mode, and indicates the following.
Display Indication
ORP Display and ORP value are indicated alternately. Setting Display Span sensitivity correction value is indicated.
(3) Immerse the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential
difference 260 mV).
(4) Set a span sensitivity correction value with the or key so that ORP value is approximately
260 mV (at 20 ). For other temperature and electrical potentials, refer to the temperature characteristics of your standard solution.
Setting range: 50 to 150%
(5) Press the key.
Span Sensitivity Correction Mode is complete, and the unit reverts to ORP Display Mode or Cleansing Output Mode.
35
9.3 Transmission Output Adjustment Mode
Fine adjustment of Transmission output is performed. This ORP meter is adjusted at the factory, however, differences may occur between the indication value of the connected equipment (recorders, etc.) and output value of this unit. In this case, perform Transmission output Zero adjustment and Span adjustment. Transmission Output Adjustment Mode is available only when the TA option is ordered.
The unit cannot enter Transmission Output Zero Adjustment Mode in the following case.
• During Adjustment Mode or Span Sensitivity Correction Mode
• When (Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.26)]
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines the procedure for Transmission output adjustment.
1
Press and hold the and key (in that order) together for approx. 3 seconds in ORP Display Mode or Cleansing Output Mode. The unit enters Transmission Output Zero Adjustment Mode, and indicates the following.
Display Indication
ORP Display
Setting Display
Transmission output Zero adjustment value
2
Set Transmission output Zero adjustment value with the or keys, while viewing the
value indicated on the connected equipment (recorders, etc.).
Setting range: 5.00% of Transmission output span
3
Press the key. The unit enters Transmission Output Span Adjustment Mode, and indicates the following.
Display Indication
ORP Display
Setting Display
Transmission output Span adjustment value
4
Set Transmission output Span adjustment value with the or keys, while viewing the
value indicated on the connected equipment (recorders, etc.).
Setting range: 5.00% of Transmission output span
5
Press the key. The unit reverts to the Transmission Output Zero Adjustment Mode. Repeat steps
2
to
5
if necessary.
6
To finish the Transmission output adjustment, press the key in Transmission
Output Span Adjustment Mode.
The unit reverts to ORP Display Mode or Cleansing Output Mode.
36
10. Measurement
10.1 Starting Measurement
After mounting to the control panel, and wiring, setup and calibration are complete, turn the power to the instrument ON. For approx. 4 seconds after the power is switched ON, characters are indicated in the ORP Display. During this time, all outputs are in OFF status, and action indicators except the Power Indicator go off. After that, measurement starts, indicating ORP value in the ORP Display and the item selected in [Setting Display indication (p.27)] in the Setting Display. This status is called ORP Display Mode or Cleansing Output Mode.
10.2 A Output
When (ORP input low limit action), (ORP input high limit action) is selected in [A11, A12, A21, A22 type (pp.21, 22)], the A output is turned ON if ORP value exceeds the A value. A1 or A2 output is turned ON depending on the settings in [A1, A2 output allocation (p.28)] and [Output ON time/OFF time when A1/A2 output ON (p.29)]. However, if (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], the Cleansing output will be given priority.
If ON time and OFF time are set, A1 or A2 output can be turned ON/OFF in a configured cycle when A1 or A2 output is ON. (Fig. 10.2-1)
Timing chart of Output ON time and OFF time while in A1 output ON
(Fig. 10.2-1)
A output status can be read by the status flag (A11, A12, A21, A22 output bit) in Serial communication. A output status, when input errors occur, differs depending on the selection in [A output when input errors occur (p.20)].
• If (Disabled) is selected, A output and A output status will be turned OFF when input errors occur.
• If (Enabled) is selected,A output and A output status will be maintained when input errors occur.
10.3 ORP Input Error Alarm
ORP input error alarm is used for detecting actuator trouble. Even if ORP input error alarm time has elapsed, and if ORP input does not become higher than ORP input error alarm span, the unit assumes that actuator trouble has occurred, and writes Status flag 2. In Serial communication, status can be read by reading Status flag 2 (A1, A2 ORP input error alarm output flags).
ORP input error alarm is disabled in the following cases.
• During Adjustment Mode, Span Sensitivity Correction Mode.
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and
cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
• When ORP input error alarm time is set to 0 seconds, or ORP input error alarm span is set to 0 mV
OFF
ON
OFF
ON
Actual A1 output
A1 output to which ON time and OFF time are set.
ON
time
ON
time
ON
time
O
FF
time
O
FF
time
ON time is turned OFF, caused by the actual A1 output turning OFF.
37
10.4 Cleansing Output
If (Cleansing output) is selected in any one of [A11, A12, A21, A22 type (pp.21, 22)], the unit will enter Cleansing Output mode. An A output (for which the cleansing output is selected) will turn ON during the configured cleansing time. When the cleansing interval finishes after restore time has passed, this is counted as one cleansing cycle, and the configured number of cleansing cycles will be repeated. While cleansing is being performed using the ‘Cleansing Time’ and ‘Restore Time after Cleansing’ settings, other outputs are in OFF status. ORP value is retained.
When cleansing is not being performed, programmed operation continues. When power is turned ON again, the unit starts from the first cleansing cycle. After the configured number of cleansing cycles are finished, the A output (for which the cleansing output is selected) is turned OFF, and other outputs perform their programmed operations, however, they are in Cleansing Output mode.
Cleansing Output Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing
(Fig. 10.4-1)
• While A1 is performing cleansing action, and if A type (to which Cleansing output is set) is selected for A2, A2 cleansing action will be the same as the (currently performing) A1 cleansing action.
While A2 is performing cleansing action, and if A type (to which Cleansing output is set) is selected for A1, A1 cleansing action will be the same as the (currently performing) A2 cleansing action.
• During Adjustment Mode, Span Sensitivity Correction Mode or Transmission Output Mode, if cleansing action initiates after restore time has passed, the cleansing action will not be performed in the current session.
• If the number of cleansing cycles is changed in [Number of cleansing cycles] during cleansing action, the new number will be valid from the next cleansing cycle.
If any output other than (Cleansing output) is selected in [A11, A12, A21, A22 type (pp.21,
22)], the unit will revert to ORP Display Mode.
Programmed
action
Programmed
action
Programmed
action
38
10.5 Manual Cleansing Mode
By pressing the and keys simultaneously for 3 seconds, the unit enters Manual Cleansing Mode. In Manual Cleansing Mode, cleansing action is performed using the ‘Cleansing Time’ and ‘Restore Time after Cleansing’ settings. After cleansing is completed, the unit automatically reverts to Cleansing Output Mode.
During cleansing action, manual cleansing via key operation is invalidated, so the unit cannot move to Manual Cleansing Mode. During Manual Cleansing Mode, if Programmed cleansing action initiates after Restore time has passed, the cleansing action will not be performed in the current session. Cleansing action by Manual Cleansing Mode is counted as 1 cleansing cycle.
Manual Cleansing Mode Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing
(Fig. 10.5-1)
10.6 Transmission Output
Converting ORP value to analog signal every input sampling period, outputs in current. If Transmission output high limit and low limit are set to the same value, Transmission output will be fixed at 4 mA DC.
Resolution 12000
Current 4 to 20 mA DC (Load resistance: Max 550 )
Output accuracy Within 0.3% of Transmission output span
During cleansing action, Manual Cleansing Mode via key operation is invalidated Programmed cleansing mode takes priority.
During Manual Cleansing Mode, the Programmed cleansing action will be invalidated. Cleansing action by Manual Cleansing Mode is counted as one cleansing cycle
Programmed
action
Programmed
action
Programmed
action
Key operation
39
11. Communication
11.1 System Configuration Example
When Using USB Communication Cable CMC-001-3 (sold separately)
(Fig. 11.1-1)
When Using Communication Converter IF-400 (sold separately)
(Fig. 11.1-2)
11.2 Setting Method of the ORP Meter
Communication parameters can be set in the Special Function Group. To enter the Special Function Group, follow the procedure below.
1
Press the key 3 times in ORP Display Mode or Cleansing Output Mode.
2
Press the key twice. ’Communication Protocol‘ appears.
3
Set each items. (Use the or key for settings, and register the value with the key.)
Character Setting Item, Function, Setting Range Factory Default
Communication protocol Shinko protocol
• Selects communication protocol.
: Shinko protocol : Modbus ASCII mode : Modbus RTU mode
Instrument number 0
• Sets the instrument number.
The instrument numbers should be set one by one when multiple instruments are connected in Serial communication, otherwise communication is impossible.
• Setting range: 0 to 95
Communication Converter
IF-400 (Sold separately)
No.0
No.2
No.30
RS-232C RS-485
No.1
No. 0
No. 2
No. 30
No. 1
WIL-101-
ORP
Host computer
Host computer
WIL-101-
ORP
USB communication cable
CMC-001-3 (Sold separately)
40
Character Setting Item, Function, Setting Range Factory Default
Communication speed 9600 bps
• Selects a communication speed equal to that of the host computer.
: 9600 bps : 19200 bps : 38400 bps
Data bit/Parity 7 bits/Even
• Selects data bit and parity.
: 8 bits/No parity : 7 bits/No parity : 8 bits/Even : 7 bits/Even : 8 bits/Odd : 7 bits/Odd
Stop bit 1 bit
• Selects the stop bit.
: 1 bit : 2 bits
4
Press the key multiple times. The unit reverts to ORP Display Mode or Cleansing Output
Mode.
11.3 Communication Procedure
Communication starts with command transmission from the host computer (hereafter Master) and
ends with the response of the WIL-101-ORP (hereafter Slave).
• Response with Data
When the master sends the reading command, the slave responds with the corresponding set value or current status.
• Acknowledgement
When the master sends the setting command, the slave responds by sending acknowledgement after processing is terminated.
• Negative Acknowledgement
When the master sends a non-existent command or value out of the setting range, the slave returns a negative acknowledgement.
• No Response
The slave will not respond to the master in the following cases:
• Global address (Shinko protocol) is set.
• Broadcast address (Modbus protocol) is set.
• Communication error (framing error, parity error)
• Checksum error (Shinko protocol), LRC discrepancy (Modbus
(Fig.11.3-1) ASCII mode), CRC-16 discrepancy (Modbus RTU mode)
Communication Timing of the RS-485
Master Side (Take note while programming)
When the master starts transmission through the RS-485 communication line, the master is arranged so as to provide an idle status (mark status) transmission period of 1 or more characters before sending the command to ensure synchronization on the receiving side. Set the program so that the master can disconnect the transmitter from the communication line within a 1 character transmission period after sending the command in preparation for reception of the response from the slave. To avoid collision of transmissions between the master and the slave, send the next command after carefully checking that the master has received the response. If a response to the command is not returned due to communication errors, set the Retry Processing to send the command again. (It is recommended to execute Retry twice or more.)
Slave Side
When the slave starts transmission through the RS-485 communication line, the slave is arranged so as to provide an idle status (mark status) transmission period of 1 or more characters before sending the response to ensure synchronization on the receiving side. The slave is arranged so as to disconnect the transmitter from the communication line within a 1 character transmission period after sending the response.
Command
Data
Command
Acknowledgement
Command
Negative Acknowledgement
Command
No response
Master
Slave
41
11.4 Shinko Protocol
11.4.1 Transmission Mode
Shinko protocol is composed of ASCII. Hexadecimal (0 to 9, A to F), which is divided into high order (4-bit) and low order (4-bit) out of 8-bit binary data in command is transmitted as ASCII characters. Data format Start bit: 1 bit
Data bit: 7 bits Parity: Even Stop bit: 1 bit
Error detection: Checksum
11.4.2 Command Configuration
All commands are composed of ASCII. The data (set value, decimal number) is represented by hexadecimal numbers. The negative numbers are represented in 2's complement. Numerals written below the command represent the number of characters.
(1) Setting Command
Header
(02H)
Address
Sub
address
(20H)
Command type (50H)
Data
item
Data Checksum
Delimiter
(03H)
1 1 1 1 4 4 2 1
(Fig. 11.4.2-1)
(2) Reading Command
Header
(02H)
Address
Sub
address
(20H)
Command type (20H)
Data
item
Checksum
Delimiter
(03H)
1 1 1 1 4 2 1
(Fig. 11.4.2-2)
(3) Response with Data
Header
(06H)
Address
Sub
address
(20H)
Command type (20H)
Data
item
Data Checksum
Delimiter
(03H)
1 1 1 1 4 4 2 1
(Fig. 11.4.2-3)
(4) Acknowledgement
Header
(06H)
Address Checksum
Delimiter
(03H)
1 1 2 1
(Fig. 11.4.2-4)
(5) Negative Acknowledgement
Header
(15H)
Address
Error code
Checksum
Delimiter
(03H)
1 1 1 2 1
(Fig. 11.4.2-5)
Header: Control code to represent the beginning of the command or the response.
ASCII codes are used. Setting command, Reading command: STX (02H) fixed Response with data, Acknowledgement: ACK (06H) fixed Negative acknowledgement: NAK (15H) fixed
Instrument Number (Address): Numbers by which the master discerns each slave.
Instrument numbers 0 to 94 and Global address 95. ASCII codes (20H to 7FH) are used by adding 20H to instrument numbers 0 to 95 (00H to 5FH). 95 (7FH) is called Global address, which is used when the same command is sent to all the slaves connected. However, the response is not returned.
Sub Address: 20H fixed
42
Command Type: Code to discern Setting command (50H) and Reading command (20H) Data Item: Classification of the command object.
Composed of 4-digit hexadecimal numbers, using ASCII. [Refer to “11.6. Communication Command Table”. (pp. 47 to 51)]
Data: The contents of data (set value) differ depending on the setting command.
Composed of 4-digit hexadecimal numbers, using ASCII. [Refer to “11.6. Communication Command Table”. (pp. 47 to 51)]
Checksum: 2-character data to detect communication errors.
Refer to “11.4.3 Checksum Calculation”.
Delimiter: Control code to represent the end of command.
ASCII code ETX (03H) fixed
Error Code: Represents an error type using ASCII.
1 (31H)-----Non-existent command 2 (32H)-----Not used 3 (33H)-----Setting outside the setting range 4 (34H)-----Status unable to be set (e.g. During Adjustment Mode, Span Sensitivity
Correction Mode or Transmission output adjustment, etc.)
5 (35H)-----During setting mode by keypad operation
11.4.3 Checksum Calculation
Checksum is used to detect receiving errors in the command or data. Set the program for the master side as well to calculate the checksum of the response data from the slaves so that communication errors can be checked. The ASCII code (hexadecimal) corresponding to the characters which range from the address to that before the checksum is converted to binary notation, and the total value is calculated. The lower one byte of the total value is converted to 2’s complement, and then to hexadecimal numbers, that is, ASCII code for the checksum.
Checksum Calculation Example
Data item 0008H (Moving average data amount): 1 (0001H) Address (instrument number): 0 (20H)
• 1’s complement: Reverse each binary bit. 0 will become 1 and vice versa.
• 2’s complement: Add 1 to 1’s complement.
(Fig. 11.4.3-1)
STX ETX
P 0 0 0 8 0 0 0 1
02H 20H 20H 50H 30H 30H 30H 38H 30H 30H 30H 31H 03H
[Characters above are represented in ASCII.]
Checksum
Checksum calculation range
[e.g.]
E
45H 37H
7
20H 20H 50H 30H 30H 30H 38H 30H 30H 30H 31H
0010 0000 0010 0000 0101 0000 0011 0000 0011 0000 0011 0000 0011 1000 0011 0000 0011 0000 0011 0000 0011 0001
+
10 0001 1001
[Hexadecimal] [Binary]
1110 0110
1
+
1110 0111
E 7
45H
37H
[1's complement]
[2's complement]
[Hexadecimal]
[ASCII]
Checksum
43
11.5 Modbus Protocol
11.5.1 Transmission Mode
There are 2 transmission modes (ASCII and RTU) in Modbus protocol.
ASCII Mode
Hexadecimal (0 to 9, A to F), which is divided into high order (4-bit) and low order (4-bit) out of 8-bit binary data in command is transmitted as ASCII characters. Data format Start bit: 1 bit
Data bit: 7 bits Parity: Even (No parity, Odd), Selectable Stop bit: 1 bit (2 bits). Selectable
Error detection : LRC (Longitudinal Redundancy Check)
RTU Mode
8-bit binary data in command is transmitted as it is. Data format Start bit: 1 bit
Data bit: 8 bits Parity: No parity (Even, Odd), Selectable Stop bit: 1 bit (2 bits). Selectable
Error detection: CRC-16 (Cyclic Redundancy Check)
11.5.2 Data Communication Interval ASCII Mode
1 second or less Max.1 second of interval between ASCII mode characters
RTU Mode
3.5 character transmission times or less To transmit continuously, an interval between characters which consist of one message, must be within 3.5 character transmission times. If an interval lasts longer than 3.5 character transmission times, WIL-101-ORP assumes that transmission from the master is finished, which results in a communication error, and will not return a response.
11.5.3 Message Configuration
ASCII Mode
ASCII mode message is configured to start by Header [: (colon)(3AH)] and end by Delimiter [CR (carriage return) (0DH) + LF (Line feed)(0AH)].
Header
(:)
Slave
address
Function
Code
Data
Error check
LRC
Delimiter
(CR)
Delimiter
(LF)
RTU Mode
RTU mode is configured to start after idle time is processed for more than 3.5 character transmissions, and end after idle time is processed for more than 3.5 character transmissions.
3.5 idle
characters
Slave
address
Function
code
Data
Error check
CRC-16
3.5 idle
characters
(1) Slave Address
Slave address is an individual instrument number on the slave side, and is set within the range 0 to 95 (00H to 5FH). The master identifies slaves by the slave address of the requested message. The slave informs the master which slave is responding to the master by placing its own address in the response message. Slave address 00H (Broadcast address) can identify all the slaves connected. However slaves do not respond.
44
(2) Function Code
The function code is the command code for the slave to undertake the following action types.
(Table 11.5.3-1)
Function Code
Contents
03 (03H) Reading the set value and information from slaves 06 (06H) Setting to slaves
Function code is used to discern whether the response is normal (acknowledgement) or if any error (negative acknowledgement) has occurred when the slave returns the response message to the master. When acknowledgement is returned, the slave simply returns the original function code. When negative acknowledgement is returned, the MSB of the original function code is set as 1 for the response. (For example, when the master sends request message setting 10H to the function code by mistake, slave returns 90H by setting the MSB to 1, because the former is an illegal function.) For negative acknowledgement, the exception codes below (Table 11.5.3-2) are set to the data of the response message, and returned to the master in order to inform it of what kind of error has occurred.
(Table 11.5.3-2)
Exception Code Contents
1 (01H) Illegal function (Non-existent function) 2 (02H) Illegal data address (Non-existent data address) 3 (03H) Illegal data value (Value out of the setting range)
17 (11H)
Shinko protocol error code 4 [Status unable to be set (e.g.) During
Adjustment Mode, Span Sensitivity Correction Mode or Transmission output adjustment, etc.]
18 (12H) Shinko protocol error code 5 (During setting mode by keypad operation)
(3) Data
Data differs depending on the function code. A request message from the master is composed of data item, amount of data and setting data. A response message from the slave is composed of a number of bytes, data and exception codes in negative acknowledgements, corresponding to the request message.
Effective range of data is -32768 to 32767 (8000H to 7FFFH).
(4) Error Check
ASCII Mode
After calculating LRC (Longitudinal Redundancy Check) from the slave address to the end of data, the calculated 8-bit data is converted to two ASCII characters, and are appended to the end of message.
How to Calculate LRC
1
Create a message in RTU mode.
2
Add all the values from the slave address to the end of data. This is assumed as X.
3
Make a complement for X (bit reverse). This is assumed as X.
4
Add a value of 1 to X. This is assumed as X.
5
Set X as an LRC to the end of the message.
6
Convert the whole message to ASCII characters.
RTU Mode
After calculating CRC-16 (Cyclic Redundancy Check) from the slave address to the end of the data, the calculated 16-bit data is appended to the end of message in sequence from low order to high order.
How to calculate CRC-16
In the CRC-16 system, the information is divided by the polynomial series. The remainder is added
to the end of the information and transmitted. The generation of a polynomial series is as follows. (Generation of polynomial series: X16+ X15+ X2+ 1)
1
Initialize the CRC-16 data (assumed as X) (FFFFH).
2
Calculate exclusive OR (XOR) with the 1st data and X. This is assumed as X.
3
Shift X one bit to the right. This is assumed as X.
45
4
When a carry is generated as a result of the shift, XOR is calculated by X of3and the fixed value (A001H). This is assumed as X. If a carry is not generated, go to step5.
5
Repeat steps3and4until shifting 8 times.
6
XOR is calculated with the next data and X. This is assumed as X.
7
Repeat steps3to5.
8
Repeat steps3to5up to the final data.
9
Set X as CRC-16 to the end of message in sequence from low order to high order.
11.5.4 Message Example ASCII Mode
Numerals written below the command represent the number of characters.
1
Reading [Slave address 1, Data item 0080H (ORP value)]
• A request message from the master Amount of data means how many data items are to be read. It is fixed as (30H 30H 30H 31H).
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(30H 33H)
Data Item
[0080H]
(30H 30H 38H 30H)
Amount of Data
[0001H]
(30H 30H 30H 31H)
Error Check
LRC
(37H 42H)
Delimiter
(0DH 0AH)
1 2 2 4 4 2 2
• Response message from the slave in normal status [100 mV (0064H)] The number of response bytes means the number of bytes of the data which has been read. It is fixed as (30H 32H).
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(30H 33H)
Number of
Response Bytes
[02H]
(30H 32H)
Data
[0064H]
(30H 30H 36H 34H)
Error Check
LRC
(39H 36H)
Delimiter
(0DH 0AH)
1 2 2 2 4 2 2
• Response message from the slave in exception (error) status (When a data item is incorrect) The function code MSB is set to 1 for the response message in exception (error) status (83H).
The exception code 02H (Non-existent data address) is returned (error).
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(38H 33H)
Exception Code
[02H]
(30H 32H)
Error Check
LRC
(37H 41H)
Delimiter
(0DH 0AH)
1 2 2 2 2 2
2
Setting [Slave address 1, Data item 0008H (Moving average data amount) 1 (0001H)]
• A request message from the master
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(30H 36H)
Data Item
[0008H]
(30H 30H 30H 38H)
Amount of data
[0001H]
(30H 30H 30H 31H)
Error Check
LRC
(46H 30H)
Delimiter
(0DH 0AH)
1 2 2 4 4 2 2
• Response message from the slave in normal status
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(30H 36H)
Data Item
[0008H]
(30H 30H 30H 38H)
Amount of data
[0001H]
(30H 30H 30H 31H)
Error Check
LRC
(46H 30H)
Delimiter
(0DH 0AH)
1 2 2 4 4 2 2
• Response message from the slave in exception (error) status (When a value out of the setting range is set) The function code MSB is set to 1 for the response message in exception (error) status (86H). The exception code 03H (Value out of the setting range) is returned (error).
Header
(3AH)
Slave
Address
(30H 31H)
Function
Code
(38H 36H)
Exception Code
[03H]
(30H 33H)
Error Check
LRC
(37H 36H)
Delimiter
(0DH 0AH)
1 2 2 2 2 2
46
RTU Mode
Numerals written below the command represent the number of characters.
1
Reading [Slave address 1, Data item 0080H (ORP value)]
• A request message from the master Amount of data means how many data items are to be read. It is fixed as (0001H).
3.5 idle
Characters
Slave
Address
(01H)
Function
Code
(03H)
Data Item
(0080H)
Amount of
data
(0001H)
Error Check
CRC-16 (85E2H)
3.5 idle
characters
1 1 2 2 2
• Response message from the slave in normal status [100 mV (0064H)] The number of response bytes means the number of bytes of the data which has been read. It is fixed as (02H).
3.5 idle
Characters
Slave
Address
(01H)
Function
Code (03H)
Number of
Response Bytes
(02H)
Data
(0064H)
Error Check
CRC-16
(B9AFH)
3.5 idle
characters
1 1 1 2 2
• Response message from the slave in exception (error) status (When data item is incorrect). The function code MSB is set to 1 for the response message in exception (error) status (83H). The exception code (02H: Non-existent data address) is returned (error).
3.5 idle
Characters
Slave
Address
(01H)
Function
Code (83H)
Exception Code
(02H)
Error Check
CRC-16
(C0F1H)
3.5 idle
characters
1 1 1 2
2
Setting [Slave address 1, Data item 0008H (Moving average data amount) 1 (0001H)]
• A request message from the master
3.5 idle
Characters
Slave
Address
(01H)
Function
Code (06H)
Data Item
(0008H)
Data
(0001H)
Error Check
CRC-16
(C9C8H)
3.5 idle
characters
1 1 2 2 2
• Response message from the slave in normal status
3.5 idle
Characters
Slave
Address
(01H)
Function
Code (06H)
Data Item
(0008H)
Data
(0001H)
Error Check
CRC-16
(C9C8H)
3.5 idle
characters
1 1 2 2 2
• Response message from the slave in exception (error) status (When a value out of the setting range
is set) The function code MSB is set to 1 for the response message in exception (error) status (86H). The exception code (03H: Value out of the setting range) is returned (error).
3.5 idle
Characters
Slave
Address
(01H)
Function
Code (86H)
Exception
Code
(03H)
Error Check
CRC-16 (0261H)
3.5 idle
characters
1 1 1 2
47
11.6 Communication Command Table
11.6.1 Notes about Setting/Reading Command
• The data (set value, decimal) is converted to hexadecimal numbers. A negative number is represented in 2's complement.
• When connecting multiple slaves, the address (instrument number) must not be duplicated.
• Data item 0200H to 0209H (User save area 1 to 10) can be read or set in 1 word units.
Effective range of data is -32768 to 32767 (8000H to 7FFFH).
• Modbus protocol uses Holding Register addresses. The Holding Register addresses are created
as follows. A Shinko command data item is converted to decimal number, and the offset of 40001 is added. The result is the Holding Register address. Using Data item 0001H (Input high limit) as an example:
Data item in the sending message is 0001H, however, Modbus protocol Holding Register address is 40002 (1 + 40001).
• Even if options are not ordered, setting or reading via software communication will be possible.
Command contents of the A11, A12, A21, A22 will function, however, Transmission output command contents will not function.
(1) Setting Command
• Up to 1,000,000 (one million) entries can be stored in non-volatile IC memory. If the number of settings exceeds the limit, the data will not be saved. So, ensure the set values are not frequently changed via software communication. (If a value set via software communication is the same as the value before the setting, the value will not be written in non-volatile IC memory.)
• Be sure to select Lock 3 when changing the set value frequently via software communication. If Lock 3 is selected, all set values – except Adjustment value, Span sensitivity correction value, Transmission output Zero adjustment value, Transmission output Span adjustment value – can be temporarily changed. However, they revert to their previous value after the power is turned off because they are not saved in the non-volatile IC memory. Do not change setting items A11, A12, A21 and A22 types. If they are changed, they will affect other setting items.
• Setting range of each item is the same as that of keypad operation.
• When the data (set value) has a decimal point, a whole number (hexadecimal) without a decimal point is used.
• If A type is changed in [A11, A12, A21, A22 type] (A11: 0003H, A12: 0050H, A21: 0051H, A22: 0052H), the A11, A12, A21 or A22 value (A11: 0004H, A12: 0053H, A21: 0054H, A22: 0055H) will default to 0 (zero). Output status of A11, A12, A21 or A22 will also be initialized.
• Settings via software communication are possible while in Set value lock status.
• The Instrument Numbers and Communication Speed of the slave cannot be set by the software communication function. They can only be set via the keypad.
• When sending a command by Global address [95 (7FH), Shinko protocol] or Broadcast address [0 (00H), Modbus protocol], the same command is sent to all the slaves connected. However, the response is not returned.
(2) Reading Command
• When the data (set value) has a decimal point, a whole number (hexadecimal) without a decimal point is used for a response.
11.6.2 Setting/Reading Command Shinko
Command
Type
Modbus
Function
Code
Data Item Data
50H/20H 06H/03H 0001H Input high limit Set value 50H/20H 06H/03H 0002H Input low limit Set value 50H/20H 06H/03H 0003H A11 type
0000H: No action 0001H: ORP input low limit action 0002H: ORP input high limit action 0003H: Cleansing output
50H/20H 06H/03H 0004H A11 value Set value
48
Shinko
Command
Type
Modbus
Function
Code
Data Item Data
50H/20H 06H/03H 0005H A11 ON side Set value 50H/20H 06H/03H 0006H A11 ON delay time Set value 50H/20H 06H/03H 0007H A11 OFF delay time Set value
50H/20H 06H/03H 0008H Moving average data amount Set value 50H/20H 06H/03H 0030H Set value lock 0000H: Unlock
0001H: Lock 1 0002H: Lock 2
0003H: Lock 3 50H/20H 06H/03H 0032H Transmission output high limit Set value 50H/20H 06H/03H 0033H Transmission output low limit Set value
50H/20H 06H/03H 0035H Auto-light function 0000H: Disabled
0001H: Enabled 50H/20H 06H/03H 0036H Setting Display indication
0000H: No indication
0001H: A11 value
0002H: A12 value
0003H: A21 value
0004H: A22 value 50H/20H 06H/03H 0037H Indication time Set value
50H/20H 06H/03H 0040H ORP input filter time constant Set value 50H/20H 06H/03H 0041H A output when input
errors occur
0000H: Enabled
0001H: Disabled 50H/20H 06H/03H 0044H Adjustment Mode 0000H: ORP Display Mode or
Cleansing Output Mode
0001H: Adjustment Mode 50H/20H 06H/03H 0045H Adjustment value Set value 50H/20H 06H/03H 0046H Span Sensitivity Correction
Mode
0000H: ORP Display Mode or
Cleansing Output Mode
0001H: Span Sensitivity
Correction Mode
50H/20H 06H/03H 0047H Span sensitivity correction
value
Set value
50H/20H 06H/03H 0048H Output ON time when A1
output ON
Set value
50H/20H 06H/03H 0049H Output OFF time when A1
output ON
Set value
50H/20H 06H/03H 004AH Output ON time when A2
output ON
Set value
50H/20H 06H/03H 004BH Output OFF time when A2
output ON
Set value
50H/20H 06H/03H 0050H A12 type
0000H: No action
0001H: ORP input low limit action
0002H: ORP input high limit action
0003H: Cleansing output 50H/20H 06H/03H 0051H A21 type
0000H: No action
0001H: ORP input low limit action
0002H: ORP input high limit action
0003H: Cleansing output 50H/20H 06H/03H 0052H A22 type
0000H: No action
0001H: ORP input low limit action
0002H: ORP input high limit action
0003H: Cleansing output 50H/20H 06H/03H 0053H A12 value Set value 50H/20H 06H/03H 0054H A21 value Set value 50H/20H 06H/03H 0055H A22 value Set value 50H/20H 06H/03H 0056H A12 ON side Set value
49
Shinko
Command
Type
Modbus
Function
Code
Data Item Data
50H/20H 06H/03H 0057H A21 ON side Set value 50H/20H 06H/03H 0058H A22 ON side Set value
50H/20H 06H/03H 0059H A12 ON delay time Set value
50H/20H 06H/03H 005AH A21 ON delay time Set value
50H/20H 06H/03H 005BH A22 ON delay time Set value
50H/20H 06H/03H 005CH A12 OFF delay time Set value
50H/20H 06H/03H 005DH A21 OFF delay time Set value
50H/20H 06H/03H 005EH A22 OFF delay time Set value 50H/20H 06H/03H 006AH A1 output allocation 0000H: A11 type
0001H: A12 type
0002H: A21 type
0003H: A22 type
0004H: A11, A12 types
0005H: A21, A22 types
0006H: A11, A21 types
0007H: A12, A22 types
0008H: A11, A12, A21, A22 types 50H/20H 06H/03H 006BH A2 output allocation
0000H: A11 type
0001H: A12 type
0002H: A21 type
0003H: A22 type
0004H: A11, A12 types
0005H: A21, A22 types
0006H: A11, A21 types
0007H: A12, A22 types
0008H: A11, A12, A21, A22 types
50H 06H 007FH
Key operation change flag clearing
0001H: Clear change flag
50H/20H 06H/03H 0100H
A11 hysteresis type 0000H: Medium Value
0001H: Reference Value 50H/20H 06H/03H 0101H
A12 hysteresis type 0000H: Medium Value
0001H: Reference Value 50H/20H 06H/03H 0102H
A21 hysteresis type 0000H: Medium Value
0001H: Reference Value 50H/20H 06H/03H 0103H
A22 hysteresis type 0000H: Medium Value
0001H: Reference Value 50H/20H 06H/03H 0104H A11 OFF side Set value
50H/20H 06H/03H 0105H A12 OFF side Set value
50H/20H 06H/03H 0106H A21 OFF side Set value
50H/20H 06H/03H 0107H A22 OFF side Set value
50H/20H 06H/03H 0108H Number of cleansing cycles Set value
50H/20H 06H/03H 0109H Cleansing interval Set value
50H/20H 06H/03H 010AH Cleansing time Set value
50H/20H 06H/03H 010BH Restore time after cleansing Set value
50H 06H 010CH Manual Cleansing Mode 0001H: Manual Cleansing Mode
50H/20H 06H/03H 010FH
Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode
0000H: Last value HOLD
0001H: Set value HOLD
0002H: Measured value 50H/20H 06H/03H 0110H Set value HOLD Set value
50
Shinko
Command
Type
Modbus
Function
Code
Data Item Data
50H/20H 06H/03H 0111H
A1 ORP input error alarm A type
0000H: No action
0001H: A11 type
0002H: A12 type
0003H: A21 type
0004H: A22 type
50H / 20H 06H / 03H 0112H A2 ORP input error alarm
A type
0000H: No action
0001H: A11 type
0002H: A12 type
0003H: A21 type
0004H: A22 type
50H / 20H 06H / 03H 0115H A1 ORP input error alarm
span when A output ON
Set value
50H / 20H 06H / 03H 0116H A1 ORP input error alarm
time when A output ON
Set value
50H / 20H 06H / 03H 0117H A1 ORP input error alarm
span when A output OFF
Set value
50H / 20H 06H / 03H 0118H A1 ORP input error alarm
time when A output OFF
Set value
50H / 20H 06H / 03H 0119H A2 ORP input error alarm
span when A output ON
Set value
50H / 20H 06H / 03H 011AH A2 ORP input error alarm
time when A output ON
Set value
50H / 20H 06H / 03H 011BH A2 ORP input error alarm
span when A output OFF
Set value
50H / 20H 06H / 03H 011CH A2 ORP input error alarm
time when A output OFF
Set value
50H / 20H 06H / 03H 0125H ORP input error alarm
time unit
0000H: Second(s)
0001H: Minute(s)
50H / 20H 06H / 03H 0126H
Transmission Output Adjustment Mode
0000H: ORP Display Mode or
Cleansing Output Mode
0001H: Transmission Output Zero
Adjustment Mode
0002H: Transmission Output Span
Adjustment Mode
50H / 20H 06H / 03H 0127H
Transmission output Zero adjustment value
Set value
50H / 20H 06H / 03H 0128H
Transmission output Span adjustment value
Set value
50H / 20H 06H / 03H 0200H User save area 1 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0201H User save area 2 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0202H User save area 3 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0203H User save area 4 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0204H User save area 5 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0205H User save area 6 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0206H User save area 7 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0207H User save area 8 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0208H User save area 9 -32768 to 32767 (8000H to 7FFFH) 50H / 20H 06H / 03H 0209H User save area 10 -32768 to 32767 (8000H to 7FFFH)
51
11.6.3 Read Only Command
Shinko
Command
Type
Modbus
Function
Code
Data Item Data
20H 03H 0080H ORP value ORP value 20H 03H 0081H Status flag 1
0000 0000 0000 0000 2
15
to 2
0
20to 28digit: Not used (Always 0) 29digit: ORP value has exceeded 1999 mV
0: Normal 1: Exceeding 1999 mV
210digit: ORP value is less than -1999 mV
0: Normal 1: Less than -1999 mV
211digit: Unit Status
0: ORP Display Mode or Cleansing Output Mode 1: Setting mode
212digit: Adjustment status flag
0: ORP Display Mode or Cleansing Output Mode 1: Adjustment Mode
213digit: Span sensitivity correction status flag
0: ORP Display Mode or Cleansing Output Mode
1: Span Sensitivity Correction Mode 214digit: A1 output 0: OFF 1: ON 215digit: Change in key operation 0: No 1: Yes
20H 03H 0091H Status flag 2
0000 0000 0000 0000 2
15
to 2
0
20digit: Not used (Always 0) 21digit: A2 output 0: OFF 1:ON 22digit: Not used (Always 0) 23digit: A11 output flag (*) 0: OFF 1: ON 24digit: A12 output flag (*) 0: OFF 1: ON 25digit: A21 output flag (*) 0: OFF 1: ON 26digit: A22 output flag (*) 0: OFF 1: ON 27digit: Cleansing action (Cleansing time)
0: Programmed action 1: During cleansing time
28digit: Cleansing action (Restore time after cleansing)
0: Programmed action 1: During Restore time after cleansing
29digit: Manual cleansing action status flag
0: No Manual cleansing action
1: During Manual cleansing action 210digit: Not used (Always 0) 211, 212digits: Transmission output adjustment status flag
2
12
2
11
Status
0 0 ORP Display Mode or Cleansing Output Mode 0 1 During transmission output Zero adjustment in
Transmission Output Adjustment Mode
1 0 During transmission output Span adjustment in
Transmission Output Adjustment Mode 213digit: A1 ORP input error alarm output flag 0: OFF 1: ON 214digit: A2 ORP input error alarm output flag 0: OFF 1: ON 215digit: Not used (Always 0)
(*) A output flag:
When (Cleansing output) is selected in [A type]: The A output flag changes to 1 (ON). If any item other than Cleansing output is selected:
When A output is ON: The A output flag changes to 1 (ON). When A output is OFF: The A output flag changes to 0 (OFF).
52
11.7 ORP Calibration and Transmission Output Adjustment via Communication Command
Like a keypad operation, there are also 2 methods in calibration via communication command:
Adjustment Mode and Span Sensitivity Correction Mode
11.7.1 Adjustment Mode
When using a brand-new sensor, please calibrate in Adjustment Mode.
By setting the adjustment value, calibrates ORP value indicated on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The following outlines the procedure for calibration.
1
When selecting Last value HOLD (0000H) at Data item 010FH (Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode), select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
2
Set Data item 0044H (Adjustment Mode) to 0001H. The unit enters Adjustment Mode.
3
Immerse the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential
difference 260 mV).
4
Set the adjustment value at Data item 0045H (Adjustment value) so that ORP value is 260 mV (at 20 ).
For other temperature and electrical potentials, refer to the temperature characteristics of your standard solution. If 212digit is read at Data item 0081H (Status flag 1), 1 (Adjustment Mode) will be returned.
5
Set Data item 0044H (Adjustment Mode) to 0000H.
Adjustment Mode is complete, and the unit will revert to ORP Display Mode or Cleansing Output Mode. If 212digit is read at Data item 0081H (Status flag 1), 0 (ORP Display Mode or Cleansing Output Mode) will be returned.
If errors (e.g. ORP value is outside the measurement range) occur in Adjustment Mode, and if adjustment cannot be carried out, Error code 1 (Exceeding 1999 mV, Less than -1999 mV) will be returned when 29and 210digits are read at Data item 0081H (Status flag 1). To cancel the error code, set Data item 0044H (Adjustment Mode) to 0000H.
The unit will revert to ORP Display Mode or Cleansing Output Mode.
11.7.2 Span Sensitivity Correction Mode
When calibrating periodically, please calibrate in Span Sensitivity Correction Mode.
By setting the Span sensitivity correction value in percentage, calibrates ORP value indicated on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The following outlines the procedure for calibration.
1
When selecting Last value HOLD (0000H) at Data item 010FH (Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode), select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
2
Set Data item 0046H (Span Sensitivity Correction Mode) to 0001H. The unit enters Span Sensitivity Correction Mode.
3
Immerse the ORP Combined Electrode Sensor in the Standard solution (Quinhydrone potential
difference 260 mV).
53
4
Set the span sensitivity correction value at Data item 0047H (Span sensitivity correction
value) so that ORP value is 260 mV (at 20 ). For other temperature and electrical potentials, refer to the temperature characteristics of your standard solution. If 213digit is read at Data item 0081H (Status flag 1), 1 (Span Sensitivity Correction Mode) will be returned.
5
Set Data item 0046H (Span Sensitivity Correction Mode) to 0000H.
Span Sensitivity Correction Mode is complete, and the unit will revert to ORP Display Mode or Cleansing Output Mode. If 213digit is read at Data item 0081H (Status flag 1), 0 (ORP Display Mode or Cleansing Output Mode) will be returned.
If errors (e.g. ORP value is outside the measurement range) occur in Span Sensitivity Correction Mode, and if span sensitivity correction cannot be carried out, Error code 1 (Exceeding 1999 mV, Less than -1999 mV) will be returned when 29and 210digits are read at Data item 0081H (Status flag 1). To cancel the error code, set Data item 0046H (Span Sensitivity Correction Mode) to 0000H. The unit will revert to ORP Display Mode or Cleansing Output Mode.
11.7.3 Transmission Output Adjustment
Fine adjustment of Transmission output is performed. This ORP meter is adjusted at the factory, however, differences may occur between the indication value of the connected equipment (recorders, etc.) and output value of this instrument. In this case, perform Transmission output Zero adjustment and Span adjustment.
The following outlines the procedure for adjustment.
1
Set Data item 0126H (Transmission Output Adjustment Mode) to 0001H.
The unit moves to Transmission Output Zero Adjustment Mode. If 212, 211digits are read at Data item 0091H (Status flag 2), 01 (During transmission output Zero adjustment in Transmission Output Adjustment Mode) will be returned.
2
Set the Transmission output Zero adjustment value at Data item 0127H (Transmission output
Zero adjustment value), while viewing the value displayed on the connected equipment (recorders, etc.). Setting range: 5.00% of Transmission output span
3
Set Data item 0126H (Transmission Output Adjustment Mode) to 0002H.
The unit moves to Transmission Output Span Adjustment Mode. If 212, 211digits are read at Data item 0091H (Status flag 2), 10 (During transmission output Span adjustment in Transmission Output Adjustment Mode) will be returned.
4
Set the Transmission output Span adjustment value at Data item 0128H (Transmission output
Span adjustment value), while viewing the value displayed on the connected equipment (recorders, etc.). Setting range: 5.00% of Transmission output span
5
Repeat steps
1
to
4
if necessary.
6
To finish the Transmission output adjustment, set Data item 0126H (Transmission Output
Adjustment Mode) to 0000H. The unit reverts to ORP Display Mode or Cleansing Output Mode.
54
11.8 Notes on Programming Monitoring Software
11.8.1 How to Speed up the Scan Time
When monitoring multiple units of the WIL-101-ORP, set the program so that the requisite minimum pieces of data such as Data item 0080H (ORP value), Data item 0081H (Status flag 1), Data item 0091H (Status flag 2) can be read. For other data, set the program so that they can be read only when their set value has been changed. This will speed up the scan time.
11.8.2 How to Read the Set Value Changes Made by Front Keypad Operation
If any set value is changed by keypad operation, the instrument sets [0081H (Status flag 1) 215: Change in key operation] to 1 (Yes). There are 2 methods of reading the set value changes made by the front keypad:
(1) Reading Method 1
1
On the monitoring software side, check that [0081H (Status flag 1) 215: Change in key
operation] has been set to 1 (Yes), then read all set values.
2
Clear [0081H (Status flag 1) 215: Change in key operation], by setting Data item 007FH (Key operation change flag clearing) to 0001H (Clear change flag). If 007FH (Key operation change flag clearing) is set to 0001H (Clear change flag) during the setting mode of the instrument, Error code 5 (35H, Shinko protocol) or Exception Code 18 (12H, Modbus protocol) will be returned as a negative acknowledgement. And [0081H (Status flag 1) 215: Change in key operation] cannot be cleared. Set a program so that all set values can be read when a negative acknowledgement is returned.
3
Read all set values again after acknowledgement is returned.
(2) Reading Method 2
1
On the monitoring software side, check that [0081H (Status flag 1) 215: Change in key
operation] has been set to 1 (Yes), then set 007FH (Key operation change flag clearing) to 0001H (Clear change flag).
2
Set the program depending on the acknowledgement or negative acknowledgement as follows.
When acknowledgement is returned:
Consider it as settings completed, and read all set values.
When Error code 5 (35H, Shinko protocol) or Exception code 18 (12H, Modbus protocol) is returned as a negative acknowledgement:
Consider it as still in setting mode, and read the requisite minimum pieces of data such as 0080H (ORP value), 0081H (Status flag 1), 0091H (Status flag 2), then return to step1.
Thus, programs which do not affect the scan time can be created using the methods described above, even if set values on the monitoring software will not be updated until settings are complete.
11.8.3 Note when Sending All Set Values Simultaneously
• If A11, A12, A21 or A22 type is changed at Data item 0003H (A11 type), 0050H (A12 type), 0051H (A21 type) or 0052H (A22 type), the A11, A12, A21 or A22 value will default to 0 (zero). Output status of A11, A12, A21 or A22 will also be initialized.
First, send the A11, A12, A21, A22 type, then send the A11, A12, A21, A22 value set at Data item 0004H (A11 value), 0053H (A12 value), 0054H (A21 value) and 0055H (A22 value).
55
12. Specifications
12.1 Standard Specifications Rating
Rated scale Input Input Range Resolution
ORP Combined Electrode Sensor
-1999 to 1999 mV 1 mV
Input ORP Combined Electrode Sensor Supply voltage Model WIL-101-ORP WIL-101-ORP 1
Supply voltage 100 to 240 V AC 50/60 Hz 24 V AC/DC 50/60 Hz Allowable voltage fluctuation range
85 to 264 V AC 20 to 28 V AC/DC
General Structure
External dimensions 30 x 88 x 108 mm (W x H x D, including socket) Mounting DIN rail Case Material: Flame-resistant resin, Color: Light gray Panel Membrane sheet Indication structure Display
ORP Display
Red LED 4-digits,character size 10 x 4.6 mm (H x W)
Setting Display
Red LED 4-digits,character size 10 x 4.6 mm (H x W)
Action indicator PWR (Yellow) Lit when power supply is ON A1 (Red) Lit when A1 output (Contact output 1) is ON. A2 (Yellow) Lit when A2 output (Contact output 2) is ON. T/R (Yellow) Lit while in Serial communication TX output
(transmitting)
Setting structure Setting method: Input system using membrane sheet key
Indication Performance
Repeatability
Within 5 mV (at equivalent input)
Linearity
Within 5 mV (at equivalent input) Input sampling period 125 ms Time accuracy Within 1% of setting time
Standard Functions
Adjustment
For successful measurement of ORP, ORP value in the sensor location, electrode performance and standard solution accuracy respectively play an
important role for obtaining reliable data. By setting the adjustment value, calibrates ORP value displayed on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
Span Sensitivity Correction
By setting the span sensitivity correction value in percentage, calibrates ORP value displayed on the WIL-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
56
Serial communication
The following operations can be carried out from an external computer. (1) Reading and setting of various set values (2) Reading of the ORP value and status (3) Function change, adjustment
(4) Reading and setting of user save area
Cable length 1.2 km (Max.), Cable resistance: Within 50 (Terminators are not
necessary, but if used, use 120 or more on one side.)
Communication line EIA RS-485 Communication method
Half-duplex communication
Communication speed
9600, 19200, 38400 bps (Selectable by keypad)
Synchronization method
Start-stop synchronization
Code form ASCII, binary Communication protocol
Shinko protocol, Modbus ASCII, Modbus RTU (Selectable by keypad)
Data bit/Parity 8-bits/No parity, 7-bits/No parity, 8-bits/Even, 7-bits/Even, 8-bits/Odd,
7-bits/Odd (Selectable by keypad)
Stop bit 1 bit, 2 bits (Selectable by keypad) Error correction Command request repeat system Error detection Parity check, Checksum (Shinko protocol), LRC (Modbus protocol ASCII),
CRC-16 (Modbus protocol RTU)
Data format
Communication
Protocol
Shinko
Protocol
Modbus ASCII Modbus RTU
Start bit 1 1 1 Data bit 7 7 8
Parity Even
Even (No parity, Odd) Selectable
No parity (Even, Odd) Selectable
Stop bit
1 1 (2)
Selectable
1 (2) Selectable
Insulation/Dielectric Strength
Circuit Insulation Configuration
Insulation resistance 10 M or more, at 500 V DC
Dielectric strength
Between power terminal - ground: 1.5 kV AC for 1 minute Between input terminal - ground: 1.5 kV AC for 1 minute Between input terminal - power terminal: 1.5 kV AC for 1 minute
Input
A1 output
A2 output or
Transmission
output
Serial
communication
Power supply
Electrically
insulated
57
Attached Functions
Set value lock Lock 1: None of the set values can be changed.
Lock 2: Only A11, A12, A21, A22 values can be changed. Lock 3: All set values – except Adjustment value, Span sensitivity correction
value, Transmission output Zero adjustment value, Transmission output Span adjustment value – can be temporarily changed.
However, they revert to their previous value after the power is
turned off because they are not saved in the non-volatile IC memory. Outside measurement range
ORP value is outside the measurement range: If the value is less than
-1999 mV or exceeds 1999 mV, the following will be indicated.
ORP Display Setting Display
Less than -1999 mV: -1999 is flashing. Exceeding 1999 mV: 1999 is flashing. However, when ORP value is outside the measurement range, and if the unit proceeds to Adjustment Mode or Span Sensitivity Correction Mode, the
ORP Display will be unlit, and the Setting Display will flash . Power failure countermeasure
The setting data is backed up in the non-volatile IC memory.
Self-diagnosis The CPU is monitored by a watchdog timer, and if an abnormal status is
found on the CPU, the WIL-101-ORP is switched to warm-up status. Warm-up Indication For approx. 4 seconds after the power is switched ON, is indicated
in the ORP Display. Display sleep function In ORP Display Mode or Cleansing Output Mode, selects the item to be
indicated in the Setting Display.
’No indication’ ‘A11, A12, A21 or A22 value’ can be selected in [Setting
Display indication (p.27)].
If any item other than ’No indication’ is selected, and if indication time is
set, the display (no operation status) becomes unlit after the indication time
has passed.
By pressing any key, the display re-lights.
If the indication time is set to 0, the display remains lit, and this function
does not work. Auto-light function Automatically measures and controls brightness of the ORP Display,
Setting Display and action indicators.
Other
Power consumption Approx. 8 VA Ambient temperature 0 to 50 (32 to 122 ) Ambient humidity 35 to 85 %RH (non-condensing) Weight Approx. 200 g (including the socket) Accessories included Instruction manual: 1 copy
Unit label: 1 sheet Accessories sold separately
Socket ASK-001-1 (Terminal screw fall prevention, finger-safe structure)
Environmental specification
RoHS directive compliant
58
12.2 Optional Specifications A Output (Option Code: EVT or TA)
A output If ORP value exceeds the A value, the A output will be turned ON
for each A output type. Regardless of options being ordered, A output status can be read via the Status flag (A11, A12, A21, A22 output bit) in Serial communication.
A output status, when input errors occur, differs depending on the selections in [A output when input errors occur (p.20)] as follows:
• If (Disabled) is selected, the A output and A output status will be turned OFF if input errors occur.
• If (Enabled) is selected, the A output and A output
status will be maintained if input errors occur. Setting range -1999 to 1999 mV Setting accuracy Same as indication accuracy Action ON/OFF action A ON side A OFF side
0 to 200 mV
A type
One type can be selected from the following with the keypad.
• No action
• ORP input low limit action
• ORP input high limit action
• Cleansing output
Output Relay contact, 1a
Control capacity 3 A 250 V AC (Resistive load)
1 A 250 V AC (Inductive load cos =0.4)
Electrical life 100,000 cycles A ON delay time 0 to 9999 seconds A OFF delay time 0 to 9999 seconds
A1, A2 output allocation
For A1 or A2 output, A11 type, A12 type, A21 type and/or A22 type are
allocated.
Output is OR output.
However, if (Cleansing output) is selected in any of [A11, A12,
A21, A22 type (pp.21, 22)], Cleansing output will be given priority. Output ON time/ OFF time when A1/A2 output ON
If Output ON time and OFF time are set, A1 or A2 output can be turned
ON/OFF in a configured cycle when A1 or A2 output is ON.
ORP input error alarm
Detects actuator trouble. Even if ORP input error alarm time has elapsed, and if ORP input does not become higher than ORP input error alarm span, the unit assumes that actuator trouble has occurred, and writes Status flag 2 (A1, A2 ORP input error alarm output flags). In Serial communication, status can be read by reading Status flag 2 (A1, A2 ORP input error alarm output flags).
ORP input error alarm is disabled in the following cases.
• During Adjustment Mode, Span Sensitivity Correction Mode.
• When (Cleansing output) is selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
• When ORP input error alarm time is set to 0 seconds, or ORP input
error alarm span is set to 0 mV
59
Transmission Output (Option Code: TA)
Transmission output
Converting ORP value to analog signal every input sampling period, and outputs the value in current. If Transmission output high limit and low limit are set to the same value, the transmission output will be fixed at 4 mA DC.
Resolution 12000 Current 4 to 20 mA DC (Load resistance: Max 550 )
Output accuracy Within 0.3% of Transmission output span Transmission output adjustment
Fine adjustment of Transmission output is performed via Transmission
output Zero adjustment and Span adjustment. Transmission output status in Adjustment Mode, Span Sensitivity Correction Mode
Transmission output status can be selected in Adjustment Mode or Span
Sensitivity Correction Mode.
Last value HOLD: Retains and outputs the last value before Adjustment or
Span Sensitivity Correction. Set value HOLD: Outputs the value set in [Set value HOLD]. Measured value: Outputs the measured value in Adjustment Mode or Span
Sensitivity Correction Mode.
60
13. Troubleshooting
If any malfunction occurs, refer to the following items after checking that power is being supplied to the WIL-101-ORP.
13.1 Indication Problem Possible Cause Solution
The ORP Display is unlit. The time set in [Indication time
(p.27)] has passed.
If any key is pressed while displays are unlit, it will re-light. Set the indication time to a suitable time-frame.
The ORP Display is dark.
(Enabled) is selected in
[Auto-light function (p.27)].
Select (Disabled).
Calibration may not have finished. Perform calibration.Indication of the ORP Display is unstable or irregular.
Specification of ORP Combined
Electrode Sensor may not be
suitable.
Replace the sensor with a suitable one.
Electrode sensor terminal screws
have become loose.
Tighten the screws securely.
Electrical insulation of electrode
sensor terminals has deteriorated.
Clean the terminals with alcohol, and dry completely.
The electrode is not clean. Clean the electrode.
Air bubbles are attached to the
electrode.
Make sure there are no bubbles in the measurement solution.
The electrode has not been
placed in the measurement
solution.
Install the electrode in the measurement solution, maintaining a consistent volume.
There may be equipment that
interferes with or makes noise
near the WIL-101-ORP.
Keep WIL-101-ORP clear of any potentially disruptive equipment.
Setting Display is unlit. (No indication) is selected
in [Setting Display indication
(p.27)].
Select any other item except
(No indication).
Select A11, A12, A21 or A22 value.
[ ] is flashing in the Setting Display.
This indicates that the ORP value
is outside the measurement range
(less than -1999 mV or exceeding
1999 mV).
Check the measuring environment.
[ ] is indicating in the ORP Display.
Internal memory is defective. Contact our agency or us.
13.2 Key Operation Problem Possible Cause Solution
None of the set values can be changed The values do not change by the or key.
(Lock 1) is selected in
[Set value lock (p.26)].
Select (Unlock).
Only A value can be set. Other settings are
impossible The values do not change
by or key.
(Lock 2) is selected in
[Set value lock (p.26)].
Select (Unlock).
61
Problem Possible Cause Solution
Unable to enter Manual Cleansing Mode.
(Cleansing output) is not selected in any of [A11, A12, A21 or A22 type (pp.21, 22)].
Select (Cleansing output) in any of [A11, A12, A21 or A22
type (pp.21, 22)]. Cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
Execute Manual cleansing after
Cleansing action is completed.
(Lock 1), (Lock 2) or (Lock 3) has been selected in [Set value lock (p.26)].
Select (Unlock).
Unable to enter calibration mode (Adjustment Mode or Span Sensitivity Correction Mode).
(Cleansing output) has been selected in any of [A11, A12, A21, A22 type (pp.21, 22)], and cleansing action is performing using the ‘Cleansing Time’ and ‘Restore Time after Cleansing’ settings.
Perform calibration after cleansing action is complete.
13.3 Communication
Check that power is being supplied to the master and slave that customers use. If communication failure still occurs, check the following.
Problem Possible Cause Solution
Communication failure Communication cable is not
securely connected, or is disconnected/defective.
Check the communication cable and connector.
Incorrect wiring of the communi­cation cable and/or connector
Check the communication cable
and connector. Imperfect contact between the communication cable and the connector, or between the communication connector and instrument port
Check the communication cable
and connector.
Communication speed of the slave does not match that of the master.
Check the communication
speed of the slave and master. The data bit, parity and stop bit of the master do not correspond to those of the slave.
Check the data bit, parity and
stop bit of the master and the
slave. The instrument number (address) of the slave does not correspond to that of the command.
Check the instrument number
(address) of the slave and
command. The instrument numbers (addresses) are duplicated in multiple slaves.
Check the instrument numbers
(addresses) of the slave.
Make sure that the program is appropriate for the transmission timing.
Check the program.
A non-existent command code has been sent.
Check the command code.Although communication is
occurring, the response is negative acknowledgement.
The setting command data exceeds the setting range of the slave.
Check the setting range of the
slave. The WIL-101-ORP cannot be set while in Adjustment Mode or Span Sensitivity Correction Mode.
Check the slave status.
The WIL-101-ORP is in the front keypad operation setting mode.
Return the unit to ORP Display
Mode or Cleansing Output Mode.
62
14. Character Tables
The following shows our character tables. Use data column for your reference.
Setting Groups
Character Setting Group
ORP Input Function Group Output Function Group Special Function Group
Adjustment Mode
Character Setting Item, Setting Range Factory Default Data
Adjustment value 0 mV(*)
-200 to 200 mV
(*) and ORP value are displayed alternately.
Span Sensitivity Correction Mode
Character Setting Item, Setting Range Factory Default Data
Span sensitivity correction value 100%(*)
50 to 150%
(*) and ORP value are displayed alternately.
Transmission Output Adjustment Mode
Character Setting Item, Setting Range Factory Default Data
Transmission output Zero
adjustment value
0.00%
5.00% of Transmission output span
Transmission output Span
adjustment value
0.00%
5.00% of Transmission output span
ORP Input Function Group
Character Setting Item, Setting Range Factory Default Data
Moving average data amount 3
1 to 20
Input high limit 1999 mV
Input low limit to 1999 mV
Input low limit -1999 mV
-1999 mV to Input high limit
ORP input filter time constant 0.0 sec.
0.0 to 60.0 seconds
A output when input errors occur Disabled
: Enabled : Disabled
63
Output Function Group
Character Setting Item, Setting Range Factory Default Data
A11 type No action
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
A12 type No action
Selection item: Same as those of A11 type
A21 type No action
Selection item: Same as those of A11 type
A22 type No action
Selection item: Same as those of A11 type
A11 value 0 mV
-1999 to 1999 mV
A12 value 0 mV
-1999 to 1999 mV
A21 value 0 mV
-1999 to 1999 mV
A22 value 0 mV
-1999 to 1999 mV
A11 hysteresis type Reference Value
: Medium Value : Reference Value
A11 ON side 10 mV
0 to 200 mV
A11 OFF side 10 mV
0 to 200 mV
A12 hysteresis type Reference Value
: Medium Value : Reference Value
A12 ON side 10 mV
0 to 200 mV
A12 OFF side 10 mV
0 to 200 mV
A21 hysteresis type Reference Value
: Medium Value : Reference Value
A21 ON side 10 mV
0 to 200 mV
A21 OFF side 10 mV
0 to 200 mV
A22 hysteresis type Reference Value
: Medium Value : Reference Value
A22 ON side 10 mV
0 to 200 mV
A22 OFF side 10 mV
0 to 200 mV
A11 ON delay time 0 sec.
0 to 9999 seconds
64
Character Setting Item, Setting Range Factory Default Data
A12 ON delay time 0 sec.
0 to 9999 seconds
A21 ON delay time 0 sec.
0 to 9999 seconds
A22 ON delay time 0 sec.
0 to 9999 seconds
A11 OFF delay time 0 sec.
0 to 9999 seconds
A12 OFF delay time 0 sec.
0 to 9999 seconds
A21 OFF delay time 0 sec.
0 to 9999 seconds
A22 OFF delay time 0 sec.
0 to 9999 seconds
65
Special Function Group
Character Setting Item, Setting Range Factory Default Data
Set value lock Unlock
: Unlock : Lock 1 : Lock 2 : Lock 3
Communication protocol Shinko protocol
: Shinko protocol : Modbus ASCII mode : Modbus RTU mode
Instrument number 0
0 to 95
Communication speed 9600 bps
: 9600 bps : 19200 bps : 38400 bps
Data bit/Parity 7 bits/Even
: 8 bits/No parity : 7 bits/No parity : 8 bits/Even : 7 bits/Even : 8 bits/Odd : 7 bits/Odd
Stop bit 1 bit
: 1 bit : 2 bits
Transmission output high limit 1999 mV
Transmission output low limit to 1999 mV
Transmission output low limit -1999 mV
-1999 mV to Transmission output high limit
Transmission output status in Last value HOLD Adjustment Mode, Span Sensitivity Correction Mode
: Last value HOLD : Set value HOLD : Measured value
Set value HOLD 0 mV
-1999 to 1999 mV
Auto-light function Disabled
: Disabled : Enabled
Setting Display indication No indication
: No indication : A11 value : A12 value : A21 value : A22 value
Indication time 00.00
00.00 (Remains lit))
00.01 to 60.00 (Minutes.Seconds)
66
Character Setting Item, Setting Range Factory Default Data
A1 output allocation A11 type
: A11 type : A12 type : A21 type : A22 type : A11, A12 types : A21, A22 types : A11, A21 types : A12, A22 types : A11, A12, A21, A22 types
A2 output allocation A21 type
: A11 type : A12 type : A21 type : A22 type : A11, A12 types : A21, A22 types : A11, A21 types : A12, A22 types : A11, A12, A21, A22 types
Output ON time when A1 output ON 0 sec.
0 to 9999 seconds
Output OFF time when A1 output ON
0 sec.
0 to 9999 seconds
Output ON time when A2 output ON 0 sec.
0 to 9999 seconds
Output OFF time when A2 output ON 0 sec.
0 to 9999 seconds
A1 ORP input error alarm A type No action
: No action : A11 type : A12 type : A21 type : A22 type
A2 ORP input error alarm A type No action
: No action : A11 type : A12 type : A21 type : A22 type
A1 ORP input error alarm span when A output ON
0 mV
0 to 1999 mV
A1 ORP input error alarm time when A output ON
0 sec.
0 to 9999 seconds or minutes
67
Character Setting Item, Setting Range Factory Default Data
A1 ORP input error alarm span when A output OFF
0 mV
0 to 1999 mV
A1 ORP input error alarm time when A output OFF
0 sec.
0 to 9999 seconds or minutes
A2 ORP input error alarm span when A output ON
0 mV
0 to 1999 mV
A2 ORP input error alarm time when A output ON
0 sec.
0 to 9999 seconds or minutes
A2 ORP input error alarm span when A output OFF
0 mV
0 to 1999 mV
A2 ORP input error alarm time when A output OFF
0 sec.
0 to 9999 seconds or minutes
ORP input error alarm time unit Second(s)
: Second(s) : Minute(s)
Number of cleansing cycles 0 (Continuous cleansing)
0 to 10 (0: Continuous cleansing)
Cleansing interval 360 minutes
60 to 3000 minutes
Cleansing time 600 sec.
1 to 1800 seconds
Restore time after cleansing 600 sec.
1 to 1800 seconds
***** Inquiries *****
For any inquiries about this unit, please contact our agency or the
vendor where you purchased the unit after checking the following.
[Example]
• Model --------------------------- WIL-101-ORP
• Serial number ----------------- No. 155F05000
In addition to the above, please let us know the details of the
malfunction, or discrepancy, and the operating conditions.
No. WIL11CORPE2 2016.04
SHINKO TECHNOS CO., LTD.
OVERSEAS DIVISION
:Head Office
URL: E-mail:
2-5-1, Senbahigashi, Minoo, Osaka, Japan http://www.shinko-technos.co.jp overseas@shinko-technos.co.jp
Tel : Fax:
+81-72-727-6100 +81-72-727-7006
Loading...