Shinko AER-101-ORP Instruction Manual

Page 1
Digital Indicating ORP Meter
AER-101-ORP
Instruction Manual
Page 2
2
Preface
Thank you for purchasing our AER-101-ORP, Digital Indicating ORP Meter. This manual contains instructions for the mounting, functions, operations and notes when operating the AER-101-ORP. To ensure safe and correct use, thoroughly read and understand this manual before using this instrument. To prevent accidents arising from the misuse of this instrument, please ensure the operator receives this manual.
Characters Used in This Manual
Indication
Number, / -1 0 1 2 3 4 5 6 7 8 9
Indication
Alphabet A B C D E F G H I J K L M
Indication
Alphabet N O P Q R S T U V W X Y Z
Caution
• This instrument should be used in accordance with the specifications described in the manual. If it is not used according to the specifications, it may malfunction or cause a fire.
• Be sure to follow all of the warnings, cautions and notices. If they are not observed, serious injury or malfunction may occur.
• The contents of this instruction manual are subject to change without notice.
• Care has been taken to assure that the contents of this instruction manual are correct, but if there are any doubts, mistakes or questions, please inform our sales department.
• This instrument is designed to be installed through a control panel. If it is not, measures must be taken to ensure that the operator cannot touch power terminals or other high voltage sections.
• Any unauthorized transfer or copying of this document, in part or in whole, is prohibited.
a result of using this product, including any indirect damage.
Page 3
21
Safety Precautions (Be sure to read these precautions before using our products.)
The safety precautions are classified into 2 categories: “Warning” and “Caution”. Depending on the circumstances, procedures indicated by Caution may result in serious consequences, so be sure to follow the directions for usage.
Warning
Warning
• To prevent an electrical shock or fire, only Shinko or other qualified service personnel may handle the inner assembly.
• To prevent an electrical shock, fire or damage to the instrument, parts replacement may only be undertaken by Shinko or other qualified service personnel.
SAFETY PRECAUTIONS
• To ensure safe and correct use, thoroughly read and understand this manual before using this instrument.
• This instrument is intended to be used for industrial machinery, machine tools and measuring equipment. Verify correct usage after purpose-of-use consultation with our agency or main office. (Never use this instrument for medical purposes with which human lives are involved.)
• External protection devices such as protective equipment against excessive temperature rise, etc. must be installed, as malfunction of this product could result in serious damage to the system or injury to personnel. Also proper periodic maintenance is required.
• This instrument must be used under the conditions and environment described in this manual. Shinko Technos Co., Ltd. does not accept liability for any injury, loss of life or damage occurring due to the instrument being used under conditions not otherwise stated in this manual.
Caution with respect to Export Trade Control Ordinance
To avoid this instrument from being used as a component in, or as being utilized in the manufacture of weapons of mass destruction (i.e. military applications, military equipment, etc.), please investigate the end users and the final use of this instrument. In the case of resale, ensure that this instrument is not illegally exported.
Caution
Procedures which may lead to dangerous conditions and cause death or serious injury, if not carried out properly.
Procedures which may lead to dangerous conditions and cause superficial to medium injury or physical damage or may degrade or damage the product, if not carried out properly.
Page 4
4
1. Installation Precautions
Caution
This instrument is intended to be used under the following environmental conditions (IEC61010-1): Overvoltage category , Pollution degree 2
Ensure the mounting location corresponds to the following conditions:
• A minimum of dust, and an absence of corrosive gases
• No flammable, explosive gases
• No mechanical vibrations or shocks
• No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 ) that does not change rapidly, and no icing
• An ambient non-condensing humidity of 35 to 85%RH
• No large capacity electromagnetic switches or cables through which large current is flowing.
• No water, oil or chemicals or where the vapors of these substances can come into direct contact with the unit
• If the AER-101-ORP is mounted through the face of a control panel, the ambient temperature of the unit – not the ambient temperature of the control panel – must be kept to under 50 . Otherwise the life of electronic parts (especially electrolytic capacitors) of the unit will be shortened.
Note: Do not install this instrument on or near flammable material even though the case of
this instrument is made of flame-resistant resin.
2. Wiring Precautions
Caution
• Do not leave wire remnants in the instrument, as they could cause a fire or a malfunction.
• Use a solderless terminal with an insulation sleeve in which the M3 screw fits when wiring the AER-101-ORP.
• The terminal block of this instrument is designed to be wired from the left side. The lead wire must be inserted from the left side of the terminal, and fastened with the terminal screw.
• Tighten the terminal screw using the specified torque. If excessive force is applied to the screw when tightening, the terminal screw may be damaged.
• This instrument does not have a built-in power switch, circuit breaker and fuse. It is
necessary to install a power switch, circuit breaker and fuse near the instrument. (Recommended fuse: Time-lag fuse, rated voltage 250 V AC, rated current 2 A)
• For a 24V AC/DC power source, do not confuse polarity when using direct current (DC).
• Be sure to connect the ground terminal to earth for safety (D-class grounding). Keep the grounding of this unit separate from other electrical devices, such as motors.
• Do not apply a commercial power source to the sensor which is connected to the input terminal nor allow the power source to come into contact with the sensor.
• Use the ORP Combined Electrode Sensor in accordance with the sensor input specifications of the AER-101-ORP.
• Keep the input wires and power line separate.
Page 5
5
Note about the ORP Combined Electrode Sensor Cable
The ORP Combined Electrode Sensor cable is a highly-insulated (electrical) cable. Please handle it with utmost care as follows.
• Do not allow terminals and socket of the ORP Combined Electrode Sensor cable to come in contact with moisture or oil of any kind. Likewise, ensure fingers are clean, otherwise the insulation will deteriorate, resulting in unstable indication. Be sure to keep the cable dry and clean at all times. If the cable is stained, clean it with alcohol, and dry it completely.
• For calibration or electrode checking/replacement, the ORP Combined Electrode Sensor cable should be wired with sufficient length.
• Keep the ORP Combined Electrode Sensor cable and junction cable away from electrical devices, such as motors or their power lines from which inductive interference emanates.
Connection
The ORP Combined Electrode Sensor cable has the following terminals.
Symbol Terminal
M Metal electrode terminal R Reference electrode terminal
3. Operation and Maintenance Precautions
Caution
• Do not touch live terminals. This may cause an electrical shock or problems in
operation.
• Turn the power supply to the instrument OFF when retightening the terminal or
cleaning.
Working on or touching the terminal with the power switched ON may result in severe injury or death due to electrical shock.
• Use a soft, dry cloth when cleaning the instrument.
(Alcohol based substances may tarnish or deface the unit.)
• As the display section is vulnerable, do not strike or scratch it with a hard object or put
pressure on it.
Page 6
6
Contents
1. Model Page
1.1 Model ---------------------------------------------------------------------------------- 8
1.2 How to Read the Model Label -------------------------------------------------- 8
2. Names and Functions of Sections ----------------------------------------------- 9
3. Mounting to the Control Panel
3.1 Site Selection -------------------------------------------------------------------- 10
3.2 External Dimensions (Scale: mm) ------------------------------------------- 10
3.3 Panel Cutout (Scale: mm) ----------------------------------------------------- 11
3.4 Mounting and Removal -------------------------------------------------------- 12
4. Wiring ------------------------------------------------------------------------------------- 13
4.1 Lead Wire Solderless Terminal ------------------------------------------------ 14
4.2 Terminal Arrangement ----------------------------------------------------------- 15
5. Outline of Key Operation and Setting Groups
5.1 Outline of Key Operation -------------------------------------------------------- 16
5.2 Setting Groups --------------------------------------------------------------------- 16
6. Key Operation Flowchart ----------------------------------------------------------- 18
7. Setup
7.1 Turn the Power Supply to the AER-101-ORP ON ------------------------ 21
7.2 ORP Input Function Group ----------------------------------------------------- 22
7.3 EVT1 Action Group --------------------------------------------------------------- 23
7.4 EVT2 Action Group --------------------------------------------------------------- 29
7.5 EVT3 Action Group --------------------------------------------------------------- 35
7.6 EVT4 Action Group --------------------------------------------------------------- 41
7.7 Special Function Group ---------------------------------------------------------- 47
8. Calibration
8.1 Adjustment Mode ----------------------------------------------------------------- 52
8.2 Span Sensitivity Correction Mode -------------------------------------------- 53
8.3 Transmission Output Adjustment Mode ------------------------------------- 54
9. Measurement
9.1 Starting Measurement ----------------------------------------------------------- 55
9.2 EVT1 to EVT4 Outputs ---------------------------------------------------------- 56
9.3 Setting EVT1 to EVT4 Values ------------------------------------------------- 58
9.4 Cleansing Ouput ------------------------------------------------------------------ 59
9.5 Manual Cleansing Mode -------------------------------------------------------- 60
9.6 ORP Input Error Alarm ---------------------------------------------------------- 61
9.7 Cycle Automatic Variable Function ------------------------------------------- 62
9.8 Transmission Output ------------------------------------------------------------- 63
Page 7
7
10. Specifications
10.1 Standard Specifications ------------------------------------------------------- 64
10.2 Optional Specifications -------------------------------------------------------- 71
11. Troubleshooting
11.1 Indication -------------------------------------------------------------------------- 72
11.2 Key Operation -------------------------------------------------------------------- 73
12 Character Tables
12.1 Setting Group List --------------------------------------------------------------- 74
12.2 Adjustment Mode ---------------------------------------------------------------- 74
12.3 Span Sensitivity Correction Mode ------------------------------------------- 74
12.4 Transmission Output Adjustment Mode ----------------------------------- 74
12.5 Simple Setting Mode ----------------------------------------------------------- 74
12.6 ORP Input Function Group ---------------------------------------------------- 75
12.7 EVT1 Action Group ------------------------------------------------------------- 75
12.8 EVT2 Action Group ------------------------------------------------------------- 77
12.9 EVT3 Action Group ------------------------------------------------------------- 78
12.10 EVT4 Action Group ----------------------------------------------------------- 80
12.11 Special Function Group ------------------------------------------------------ 81
Page 8
8
1. Model
1.1 Model
A E R - 1 0 1 ORP
,
Input Points 1 1 point Input ORP ORP Combined Electrode Sensor
Supply Voltage
100 to 240 V AC (standard)
1 24 V AC/DC (*)
Option
C5 Serial communication RS-485 EVT3 EVT3, EVT4 Outputs (Contact outputs 3, 4)
(*) Supply voltage 100 to 240 V AC is standard.
When ordering 24 V AC/DC, enter “1” after the input code.
1.2 How to Read the Model Label
The model label is attached to the left side of the case.
(Fig. 1.2-1)
Model, Option Input indication
Supply voltage, Power consumption Serial number
Manufacturer
Factory ID
RoHS directive compliant
Page 9
9
2. Names and Functions of Sections
(Fig. 2-1)
Displays
ORP Display
ORP value or characters in setting mode are indicated in red/green/orange. Indications differ depending on the selections in [Backlight selection
(p.49)] and [ORP color (p.49)]. Setting Display
Values in setting modes are indicated in green.
Indications differ depending on the selections in [Backlight selection
(p.49)] and [Setting Display indication (p.50)]. Output Display
Backlight green
The bar graph is lit corresponding to the transmission output.
Indications differ depending on the selections in [Bar graph indication(p.50)].
Action Indicators: Backlight orange
EVT1 Lit when EVT1 output (Contact output 1) is ON. EVT2 Lit when EVT2 output (Contact output 2) is ON. EVT3 Lit when EVT3 output (Contact output 3) (EVT3 option) is ON. EVT4 Lit when EVT4 output (Contact output 4) (EVT3 option) is ON. T/R Lit during Serial communication (C5 option) TX output (transmitting). LOCK Lit when Lock 1, 2 or 3 is selected.
Unit label
UPPER Attach the user’s unit of ORP Display from the included unit labels if necessary. LOWER Attach the user’s unit of Setting Display from the included unit labels
if necessary.
Keys
UP key
Increases the numeric value.
DOWN key
Decreases the numeric value.
MODE key Selects a group.
SET key Switches the setting modes, and registers the set value.
ORP Display
Setting Display
Action Indicators
UP Key
DOWN Key
MODE Key
SET Key
Output Display
Unit label
Page 10
10
3. Mounting to the Control Panel
3.1 Site Selection
Caution
Use within the following temperature and humidity ranges:
Temperature: 0 to 50 (32 to 122 ) (No icing)
Humidity: 35 to 85 %RH (Non-condensing) If AER-101-ORP is mounted through the face of a control panel, the ambient temperature of the unit – not the ambient temperature of the control panel – must be kept to under 50 , otherwise the life of electronic parts (especially electrolytic capacitors) of the unit will be shortened.
This instrument is intended to be used under the following environmental conditions (IEC61010-1): Overvoltage category , Pollution degree 2
Ensure the mounting location corresponds to the following conditions:
• A minimum of dust, and an absence of corrosive gases
• No flammable, explosive gases
• No mechanical vibrations or shocks
• No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 )
that does not change rapidly
• An ambient non-condensing humidity of 35 to 85 %RH
• No large capacity electromagnetic switches or cables through which large current
is flowing
• No water, oil or chemicals or where the vapors of these substances can come into
direct contact with the unit
3.2 External Dimensions (Scale: mm)
(Fig. 3.2-1)
11.5
104.5 (when terminal cover is used)
98.5
106.2
91
Gasket
Screw type mounting bracket
Terminal cover (sold separately)
48
96
Page 11
11
3.3 Panel Cutout (Scale: mm)
Caution
If lateral close mounting is used for the unit, IP66 specification (Drip-proof/Dust-proof) may be compromised, and all warranties will be invalidated.
Lateral close mounting n: Number of units mounted
(Fig. 3.3-1)
130
92
0.8
0
45
0.5
0
n×48-3
0.5
0
92
0.8
0
Page 12
12
3.4 Mounting and Removal
Caution
As the case is made of resin, do not use excessive force while screwing in the mounting bracket, or the case or mounting brackets could be damaged. The tightening torque should be 0.12 N•m.
How to mount the unit
Mount the unit vertically to the flat, rigid panel to ensure it adheres to the Drip-proof/ Dust-proof specification (IP66). Mountable panel thickness: 1 to 8 mm (1) Insert the unit from the front side of the panel. (2) Attach the mounting brackets by the holes at the top and bottom of the case, and
secure the unit in place with the screws.
How to remove the unit
(1) Turn the power to the unit OFF, and disconnect all wires before removing the unit. (2) Loosen the screws of the mounting brackets, and remove the mounting brackets. (3) Pull the unit out from the front of the panel.
(Fig. 3.4-1)
UP
Page 13
13
4. Wiring
Warning
Turn the power supply to the instrument off before wiring or checking. Working on or touching the terminal with the power switched on may result in severe injury or death due to electrical shock.
Caution
• Do not leave wire remnants in the instrument, as they could cause a fire or a malfunction.
• Use a solderless terminal with an insulation sleeve in which the M3 screw fits when wiring the AER-101-ORP.
• The terminal block of this instrument is designed to be wired from the left side. The lead wire must be inserted from the left side of the terminal, and fastened with the terminal screw.
• Tighten the terminal screw using the specified torque. If excessive force is applied to the screw when tightening, the terminal screw may be damaged.
• This instrument does not have a built-in power switch, circuit breaker and fuse. It is
necessary to install a power switch, circuit breaker and fuse near the instrument. (Recommended fuse: Time-lag fuse, rated voltage 250 V AC, rated current 2 A)
• For a 24 V AC/DC power source, do not confuse polarity when using direct current (DC).
• Be sure to connect the ground terminal to earth for safety (D-class grounding). Keep the grounding of this unit separate from other electrical devices, such as motors.
• Do not apply a commercial power source to the sensor which is connected to the input terminal nor allow the power source to come into contact with the sensor.
• Use the ORP Combined Electrode Sensor in accordance with the sensor input specifications of the AER-101-ORP.
• Keep the input wires and power line separate.
Page 14
14
Note about the ORP Combined Electrode Sensor Cable
The ORP Combined Electrode Sensor cable is a highly-insulated (electrical) cable. Please handle it with utmost care as follows.
• Do not allow terminals and socket of the ORP Combined Electrode Sensor cable to
come in contact with moisture or oil of any kind. Likewise, ensure fingers are clean, otherwise the insulation will deteriorate, resulting in unstable indication.
Be sure to keep the cable dry and clean at all times. If the cable is stained, clean it with alcohol, and dry it completely.
• For calibration or electrode checking/replacement, the ORP Combined Electrode Sensor cable should be wired with sufficient length.
• Keep the ORP Combined Electrode Sensor cable and junction cable away from
electrical devices, such as motors or their power lines from which inductive interference emanates.
Connection
The ORP Combined Electrode Sensor cable has the following terminals.
Symbol Terminal M Metal electrode terminal R Reference electrode terminal
4.1 Lead Wire Solderless Terminal
Use a solderless terminal with an insulation sleeve in which an M3 screw fits as follows. The tightening torque should be 0.63 N•m.
Solderless
Terminal
Manufacturer Model
Tightening
Torque
Y-type
Nichifu Terminal Industries CO.,LTD. TMEV1.25Y-3
0.63 N•m
Japan Solderless Terminal MFG CO.,LTD. VD1.25-B3A
Ring-type Nichifu Terminal Industries CO.,LTD. TMEV1.25-3
Japan Solderless Terminal MFG CO.,LTD. V1.25-3
(Fig. 4.1-1)
5.8 mm
max.
3.2 mm
3.2 mm
5.8 mm
max.
Page 15
15
4.2 Terminal Arrangement
(Fig. 4.2-1) GND Ground
POWER SUPPLY
100 to 240 V AC or 24 V AC/DC (when 1 is added after the model)
For 24 V DC, ensure polarity is correct.
EVT1 EVT1 output (Contact output 1)
EVT2 EVT2 output (Contact output 2)
TRANSMIT OUTPUT Transmission output
M, R Electrode sensor
RS-485 Serial communication RS-485 (C5 option)
2 connectors are wired internally. Use the included wire harnesses C5J and C0J.
EVT3 EVT3 output (Contact output 3) (EVT3 option)
Use the included wire harness HBJ.
EVT4 EVT4 output (Contact output 4) (EVT3 option)
Use the included wire harness HBJ.
Page 16
16
5. Outline of Key Operation and SettingGroups
5.1 Outline of Key Operation
There are 2 setting modes: Simple Setting mode, and Group Selection mode in which setting items are divided into groups. To enter Simple Setting mode, press the key in ORP Display mode or Cleansing Output mode. To enter Group Selection mode, press the key in ORP Display mode or Cleansing Output mode.
Select a group with the key, and press the key. The unit enters each setting item.
To set each item, use the or key, and register the set value with the key.
5.2 Setting Groups
POWER ON Simple Setting Mode
EVT1 value EVT4 value (*4)
ORP Display Mode or Manual Cleansing Mode Cleansing Output Mode (*1) Adjustment Mode
Span Sens. Correct. mode
+ (3 sec)
Trans. output Trans. output Zero adjust mode Span adjust mode
Group Selection Mode Setting Item selection
ORP Input Group Moving average ORP input filter
data amount time constant
EVT1 Action EVT1 type EVT1 cycle Group extended time
EVT2 Action EVT2 type EVT2 cycle Group extended time
EVT3 Action EVT3 type (*4) EVT3 cycle Group (*4) extended time (*4)
EVT4 Action EVT4 type (*4) EVT4 cycle Group (*4) extended time (*4)
Special Function Set value lock ORP input error Group alarm time unit
Abbreviations: Sens.: Sensitivity Correct: Correction
Trans.: Transmission Adjust: Adjustment
+ (3 sec) (*2)
+ (3 sec) (*3)
+ (3 sec) (*3)
(*2)
Page 17
17
[About each mode and setting items]
(*1) In ORP Display mode or Cleansing Output mode, measurement starts, indicating
the item selected in [Backlight selection (p.49), ORP color (p.49), Bar graph indication (p.50) and Setting Display indication (p.50)] in Special Function group. If power is turned ON again, the last mode (ORP Display mode or Cleansing Output mode) from when power was turned OFF will resume.
(*2) If (Cleansing output) is selected in [EVT1 type to EVT4 type] in the EVT1
to EVT4 Action Groups, the unit will enter Manual Cleansing mode. After cleansing action is complete, the unit automatically reverts to Cleansing Output mode.
(*3) The unit cannot enter Adjustment mode or Span sensitivity correction mode in the
following cases:
(Lock 1) ( Lock 2) or (Lock 3) is selected in [Set value lock (p.47)].
• When (Cleansing output) is selected in any of [EVT1 type to EVT4 type (pp. 23, 29, 35, 41)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
(*4) Setting groups and items with dotted lines are indicated only when the EVT3 option
is ordered.
[Key Operation]
+ (3 sec): Press and hold the key and key (in that order) together for 3 seconds. The unit will proceed to Manual Cleansing mode.
+ (3 sec): Press and hold the key and key (in that order) together
for 3 seconds. The unit will proceed to Adjustment mode.
+ (3 sec): Press and hold the key and key (in that order) together
for 3 seconds. The unit will proceed to Span sensitivity correction mode.
+ (3 sec): Press and hold the key and key (in that order) together
for 3 seconds. The unit will proceed to Transmission output Zero adjustment mode.
, : Press the or key. The unit will proceed to the next setting
item, illustrated by an arrow.
, : Press the or key until the desired setting mode appears.
• To revert to ORP Display mode or Cleansing Output mode, press and hold the
key for 3 seconds while in any mode
Page 18
18
6. Key Operation Flowchart
POWER ON
ORP Display M
ode
+ (3 sec)(*2) Manual (*2)
or Cleansing Cleansing mode
Output Mode (*1)
+ (3sec)(*3) Adjustment
mode
+ (3 sec) Span sensitivity
correction mode
+ (3 sec) Trans. output Trans. output
Zero adjust. mode Span adjust. mode
ORP Input Group
Simple Setting Mode Moving average
EVT1 value data amount
Input high limit
EVT2 value
Input low limit
EVT3 value
ORP input filter time
EVT4 value constant
[About Setting Items]
EVT1 value
Upper left
:
ORP Display:
Indicates
the
setting item characters.
• Lower left: Setting Display: Indicates the factory default.
Right side: Indicates the
s
etting
item.
EVT3 value
Setting item in shaded section will be indicated only when
the corresponding option is ordered.
[Each Mode and Setting Items]
(*1) In ORP Display mode or Cleansing Output mode, measurement starts, indicating the item selected in [Backlight selection (p.49), ORP
color (p.49), Bar graph indication (p.50) and Setting Display indication (p.50)] in Special Function group. If power is turned ON again, the
last mode (ORP Display mode or Cleansing Output mode) from when power was turned OFF will resume.
(*2) If (Cleansing output) is selected in [EVT1 type to EVT4 type] in the EVT1 to EVT4 Action Groups, the unit will enter Manual
cleansing mode. After cleansing action is complete, the unit automatically reverts to Cleansing Output mode.
(*3) The unit cannot enter Adjustment mode or Span sensitivity correction mode in the following cases:
(Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.47)] in the Special Function group.
• When (Cleansing output) is selected in any of [EVT1 type to EVT4 type (pp. 23, 29, 35, 41)], and cleansing action is
performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
[About Key Operation]
+ (3 sec): Press and hold the and keys (in that order) together for 3 sec. The unit will enter Manual Cleansing mode.
+ (3 sec): Pressand hold the and keys (in that order) together for 3 sec.The unit will enter Adjustment mode.
+ (3 sec): Press and hold the and keys (in that order) together for3 sec. The unit will enter Span sensitivity correction mode.
+ (3 sec): Press and hold the and keys (in that order) together for 3 sec. The unit will enter Transmission outputZero adjustment mode.
, : Press the or key. The unit will proceed to the next setting item, illustrated by an arrow.
To revert to ORP Display mode or Cleansing Output mode, press and hold the key for 3 sec. while in any mode.
Abbreviations: Sens.: Sensitivity, Correct.: Correction
Trans: Transmission Adjust: Adjustment
Page 19
19
EVT1 Action EVT2 Action EVT3 Action Group Group Group
EVT1 type Setting items: Same as item Setting items: Same as item
settings for EVT1 Action Group. settings for EVT1 Action Group.
For EVT2, read EVT1 column For EVT3, read EVT1 column
EVT1 value and substitute EVT1 with EVT2. and substitute EVT1 with EVT3.
(e.g.) (e.g.)
EVT1 proportional band
EVT1 reset
EVT1 hysteresis type
Output OFF time when
EVT1 ON side EVT1 output ON
EVT1 ORP input error
EVT1 OFF side alarm EVT type
EVT1 ORP input error
EVT1 ON delay alarm span when time EVT output ON
EVT1 OFF delay EVT1 ORP input error time alarm time when
EVT output ON
EVT1 proportional
cycle EVT1 ORP input error
alarm span when
EVT1 output high EVT output OFF
limit
EVT1 ORP input error
EVT1 output low alarm time when limit EVT output OFF
Output ON time when EVT1 cycle variable
EVT1 output ON range
EVT1 cycle extended time
Page 20
20
EVT4 Action Special Function Group Group
Setting items: Same as item Set value lock settings for EVT1 Action Group.
For EVT4, read EVT1 column
and substitute EVT1 with EVT4. Communication (e.g.) protocol
Instrument number
Communication ORP color speed reference value
Data bit/Parity ORP color range
Stop bit Backlight time
Transmmision Bar graph indication output type
Transmmision
EVT output when
output high limit
input errors occur
Transmmision Setting Display output low limit indication
Trans. output status Number of cleansing
in Adjust. mode, Span cycles
sensitivity correct.mode
Cleansing interval
Set value HOLD
Cleansing time
Backlight selection
Restore time after
ORP color cleansing
ORP input error alarm time unit
Abbreviations: Trans.: Transmission Adjust: Adjustment
Correct: Correction
Page 21
21
7. Setup
Setup should be done before using this instrument according to the user’s conditions:
Setting the ORP input function (input high limit, low limit), EVT1, EVT2, EVT3 (EVT3 option) and EVT4 (EVT3 option) types, Communication (C5 option), Transmission output, Indication settings (Backlight selection, ORP color, etc.) and Cleansing action.
Setup can be conducted in the ORP Input group, EVT1, EVT2, EVT3, EVT4 Action groups and Special Function group.
If user’s specification is the same as the factory default of the AER-101-ORP, or if setup has already been completed, it is not necessary to set up the instrument. Proceed to Section “8. Calibration (p.52)”.
7.1 Turn the Power Supply to the AER-101-ORP ON.
For approx. 4 seconds after the power is switched ON, the following characters are indicated on the ORP Display and Setting Display.
ORP Display Setting Display
Unlit
During this time, all outputs are in OFF status, and action indicators go off. After that, measurement starts, indicating the item selected in [Backlight selection (p.49), ORP color (p.49), Bar graph indication (p.50), Setting Display indication (p.50)] in the Special Function group. This status is called ORP Display mode or Cleansing Output mode.
Page 22
22
7.2 ORP Input Group
To enter the ORP Input group, follow the procedure below.
1
Press the key in ORP Display mode or Cleansing Output mode.
2
Press the key. The unit proceeds to the ORP Input group, and “Moving average data amount” will appear.
Character Setting Item, Function, Setting Range Factory Default
Moving average data amount 3
• Sets the number of ORP inputs for calculating the moving average.
• Setting range: 1 to 20
Input high limit 2000 mV
• Sets the high limit value for ORP input indication.
• Setting range: Input low limit to 2000 mV
Input low limit -2000 mV
• Sets the low limit value for ORP input indication.
• Setting range: -2000 mV to Input high limit
ORP input filter time constant 0.0 sec.
• Sets ORP input filter time constant. Even when ORP value before filter process changes as shown in (Fig. 7.2-1), if the filter time constant “T” is set, the ORP value changes as shown in (Fig. 7.2-2) so that ORP value after finishing filter process can reach 63% (of the desired ORP value) after T seconds have passed. If the filter time constant is set too large, it affects EVT action due to the delay of response. (e.g.) If the LSD (least significant digit) of the ORP value prior to
filter process is fluctuating, it can be suppressed by using the filter time constant.
(Fig. 7.2-1) (Fig. 7.2-2)
• Setting range: 0.0 to 60.0 sec.
ORP value
Time (sec.)
ORP value
Time (sec.)
T
Page 23
23
7.3 EVT1 Action Group
To enter the EVT1 Action group, follow the procedure below.
1
Press the key twice in ORP Display mode or Cleansing Output mode.
2
Press the key.
The unit proceeds to the EVT1 Action group, and ”EVT1 type“ will appear.
Character Setting Item, Function, Setting Range Factory Default
EVT1 type No action
• Selects an EVT1 output (Contact output 1) type. (Fig 7.3-1, p.28)
Note: If EVT1 type is changed, EVT1 value defaults to 0.
: No action : ORP input low limit action : ORP input high limit action : Cleansing output : ORP input error alarm output
EVT1 value 0 mV
• Sets EVT1 value. (Fig 7.3-1, p.28)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Setting range: Input low limit to Input high limit
EVT1 proportional band 0 mV
• Sets EVT1 proportional band. (Fig 7.3-1, p.28) ON/OFF control action when set to 0.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Setting range: 0 to 4000 mV
EVT1 reset 0 mV
• Sets EVT1 reset value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: 200 mV
Page 24
24
Character Setting Item, Function, Setting Range Factory Default
EVT1 hysteresis type Reference Value
• Selects EVT1 output hysteresis type (Medium or Reference Value). (Fig. 7.3-1, p.28)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the P control action.
: Medium Value Sets the same value for both ON and OFF sides in relation to EVT1 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to EVT1 value.
Both ON and OFF sides need to be set individually.
EVT1 ON side 10 mV
• Sets the span of EVT1 ON side. (Fig. 7.3-1, p.28) If (Medium Value) is selected in [EVT1 hysteresis type], the span of ON/OFF side will be the same value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the P control action.
• Setting range: 0 to 200 mV
EVT1 OFF side 10 mV
• Sets the span of EVT1 OFF side. (Fig. 7.3-1, p.28)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the P control action, or if (Medium Value) is selected in [EVT1 hysteresis type].
• Setting range: 0 to 200 mV
EVT1 ON delay time 0 sec.
• Sets EVT1 delay time. The EVT1 output does not turn ON after the input value exceeds the EVT1 value until the time set in [EVT1 ON delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
Page 25
25
Character Setting Item, Function, Setting Range Factory Default
EVT1 OFF delay time 0 sec.
• Sets EVT1 delay time. The EVT1 output does not turn OFF after the input value exceeds the EVT1 value until the time set in [EVT1 OFF delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
EVT1 proportional cycle 30 sec.
• Sets EVT1 proportional cycle.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: 1 to 300 sec.
EVT1 output high limit 100%
• Sets EVT1 output high limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: EVT1 output low limit to 100%
EVT1 output low limit 0%
• Sets EVT1 output low limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: 0% to EVT1 output high limit
Output ON time when EVT1 output ON 0 sec
• Sets Output ON time when EVT1 output is ON. If ON time and OFF time are set, EVT1 output can be turned ON/OFF in a configured cycle when EVT1 output is turned ON.(Fig. 7.3-2, p.28)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
Page 26
26
Character Setting Item, Function, Setting Range Factory Default
Output OFF time when EVT1 output ON 0 sec.
• Sets Output OFF time when EVT1 output is ON. If ON time and OFF time are set, EVT1 output can be turned ON/OFF in a configured cycle when EVT1 output is turned ON.(Fig. 7.3-2, p.28)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
EVT1 ORP input error alarm EVT type No action
• Selects EVT output type (except EVT1 type) in order to assess
EVT1 ORP input error alarm.
• Available only when (ORP input error alarm output) is selected in [EVT1 type]
: No action : EVT2 action : EVT3 action : EVT4 action
EVT1 ORP input error alarm span when EVT output ON
0 mV
• Sets span to assess EVT1 ORP input error alarm when EVT output (selected in [EVT1 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT1 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT1 ORP input error alarm time when EVT output ON
0 sec.
• Sets time to assess EVT1 ORP input error alarm when EVT output (selected in [EVT1 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT1 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 27
27
Character Setting Item, Function, Setting Range Factory Default
EVT1 ORP input error alarm span
when EVT output OFF
0 mV
• Sets span to assess EVT1 ORP input error alarm when EVT output (selected in [EVT1 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT1 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT1 ORP input error alarm time when EVT output OFF
0 sec.
• Sets time to assess EVT1 ORP input error alarm when EVT output (selected in [EVT1 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT1 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
EVT1 cycle variable range 50.0%
• Sets EVT1 cycle variable range.
• Available when (ORP input low limit action) or
(ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: 1.0 to 100.0%
EVT1 cycle extended time 0 sec.
• Sets EVT1 cycle extended time.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type].
• Not available for the ON/OFF control action.
• Setting range: 0 to 300 sec.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 28
28
EVT1 Action
EVT1 Type P Control Action ON/OFF Control Action
ORP input low limit action
EVT1 proportional band
ON
OFF
EVT1 value
If Medium Value is selected in [EVT1 hysteresis type]:
EVT1 ON sides
ON
OFF
EVT1 value If Reference Value is selected in [EVT1 hysteresis type]:
EVT1 ON side EVT1 OFF side
ON
OFF
EVT1 value
ORP input high limit action
EVT1 proportional band
ON
OFF
EVT1 value
If Medium Value is selected in [EVT1 hysteresis type]:
EVT1 ON sides
ON
OFF
EVT1 value
If Reference Value is selected in [EVT1 hysteresis type]:
ON
OFF
EVT1 value
(Fig 7.3-1)
Timing chart of EVT1 output ON time and OFF time while in EVT1 output ON
(Fig. 7.3-2)
OFF
ON
OFF
ON
Actual EVT1 output
EVT1 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual EVT1 output turning OFF.
EVT1 ON side
EVT1 OFF side
Page 29
29
7.4 EVT2 Action Group
To enter the EVT2 Action group, follow the procedures below.
1
Press the key 3 times in ORP Display mode or Cleansing Output mode.
2
Press the key.
The unit proceeds to the EVT2 Action group, and ”EVT2 type“ will appear.
Character Setting Item, Function, Setting Range Factory Default
EVT2 type No action
• Selects an EVT2 output (Contact output 2) type. (Fig 7.4-1, p.34)
Note: If EVT2 type is changed, EVT2 value defaults to 0.
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT2 value 0 mV
• Sets EVT2 value. (Fig 7.4-1, p.34)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Setting range: Input low limit to Input high limit
EVT2 proportional band 0 mV
• Sets EVT2 proportional band. (Fig 7.4-1, p.34) ON/OFF control action when set to 0.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Setting range: 0 to 4000 mV
EVT2 reset 0 mV
• Sets EVT2 reset value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: 200 mV
Page 30
30
Character Setting Item, Function, Setting Range Factory Default
EVT2 hysteresis type Reference Value
• Selects EVT2 output hysteresis type (Medium or Reference Value). (Fig. 7.4-1, p.34)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the P control action.
: Medium Value Sets the same value for both ON and OFF sides in relation to EVT2 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to EVT2 value.
Both ON and OFF sides need to be set individually.
EVT2 ON side 10 mV
• Sets the span of EVT2 ON side. (Fig. 7.4-1, p.34) If (Medium Value) is selected in [EVT2 hysteresis type], the span of ON/OFF side will be the same value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the P control action.
• Setting range: 0 to 200 mV
EVT2 OFF side 10 mV
• Sets the span of EVT2 OFF side. (Fig. 7.4-1, p.34)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the P control action, or if (Medium Value) is selected in [EVT2 hysteresis type].
• Setting range: 0 to 200 mV
EVT2 ON delay time 0 sec.
• Sets EVT2 delay time. The EVT2 output does not turn ON after the input value exceeds the EVT2 value until the time set in [EVT2 ON delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
Page 31
31
Character Setting Item, Function, Setting Range Factory Default
EVT2 OFF delay time 0 sec.
• Sets EVT2 delay time. The EVT2 output does not turn OFF after the input value exceeds the EVT2 value until the time set in [EVT2 OFF delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
EVT2 proportional cycle 30 sec.
• Sets EVT2 proportional cycle.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: 1 to 300 sec.
EVT2 output high limit 100%
• Sets EVT2 output high limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: EVT2 output low limit to 100%
EVT2 output low limit 0%
• Sets EVT2 output low limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: 0% to EVT2 output high limit
Output ON time when EVT2 output ON 0 sec
• Sets Output ON time when EVT2 output is ON. If ON time and OFF time are set, EVT2 output can be turned ON/OFF in a configured cycle when EVT2 output is turned ON.(Fig. 7.4-2, p.34)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
Page 32
32
Character Setting Item, Function, Setting Range Factory Default
Output OFF time when EVT2 output ON 0 sec.
• Sets Output OFF time when EVT2 output is ON. If ON time and OFF time are set, EVT2 output can be turned ON/OFF in a configured cycle when EVT2 output is turned ON.(Fig. 7.4-2, p.34)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
EVT2 ORP input error alarm EVT type No action
• Selects EVT output type (except EVT2 type) in order to assess
EVT2 ORP input error alarm.
• Available only when (ORP input error alarm output) is selected in [EVT2 type]
: EVT1 action : No action : EVT3 action : EVT4 action
EVT2 ORP input error alarm span when EVT output ON
0 mV
• Sets span to assess EVT2 ORP input error alarm when EVT output (selected in [EVT2 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT2 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT2 ORP input error alarm time when EVT output ON
0 sec.
• Sets time to assess EVT2 ORP input error alarm when EVT output (selected in [EVT2 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT2 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 33
33
Character Setting Item, Function, Setting Range Factory Default
EVT2 ORP input error alarm span
when EVT output OFF
0 mV
• Sets span to assess EVT2 ORP input error alarm when EVT output (selected in [EVT2 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT2 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT2 ORP input error alarm time when EVT output OFF
0 sec.
• Sets time to assess EVT2 ORP input error alarm when EVT output (selected in [EVT2 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT2 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
EVT2 cycle variable range 50.0%
• Sets EVT2 cycle variable range.
• Available when (ORP input low limit action) or
(ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: 1.0 to 100.0%
EVT2 cycle extended time 0 sec.
• Sets EVT2 cycle extended time.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type].
• Not available for the ON/OFF control action.
• Setting range: 0 to 300 sec.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 34
34
EVT2 Action
EVT2 Type P Control Action ON/OFF Control Action
ORP input low limit action
EVT2 proportional band
ON
OFF
EVT2 value
If Medium Value is selected in [EVT2 hysteresis type]:
EVT2 ON sides
ON
OFF
EVT2 value If Reference Value is selected in [EVT2 hysteresis type]:
EVT2 ON side EVT2 OFF side
ON
OFF
EVT2 value
ORP input high limit action
EVT2 proportional band
ON
OFF
EVT2 value
If Medium Value is selected in [EVT2 hysteresis type]:
EVT2 ON sides
ON
OFF
EVT2 value
If Reference Value is selected in [EVT2 hysteresis type]:
ON
OFF
EVT2 value
(Fig 7.4-1)
Timing chart of EVT2 output ON time and OFF time while in EVT2 output ON
(Fig. 7.4-2)
OFF
ON
OFF
ON
Actual EVT2 output
EVT2 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual EVT2 output turning OFF.
EVT2ON side
EVT2OFF side
Page 35
35
7.5 EVT3 Action Group
EVT3 Action group is indicated only when EVT3, EVT4 Outputs (EVT3 option) are/is ordered. To enter the EVT3 Action group, follow the procedure below.
1
Press the key 4 times in ORP Display mode or Cleansing Output mode.
2
Press the key.
The unit proceeds to the EVT3 Action group, and ”EVT3 type“ will appear.
Character Setting Item, Function, Setting Range Factory Default
EVT3 type No action
• Selects an EVT3 output (Contact output 3) type. (Fig 7.5-1, p.40)
Note: If EVT3 type is changed, EVT3 value defaults to 0.
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT3 value 0 mV
• Sets EVT3 value. (Fig 7.5-1, p.40)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Setting range: Input low limit to Input high limit
EVT3 proportional band 0 mV
• Sets EVT3 proportional band. (Fig 7.5-1, p.40) ON/OFF control action when set to 0.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Setting range: 0 to 4000 mV
EVT3 reset 0 mV
• Sets EVT3 reset value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 200 mV
Page 36
36
Character Setting Item, Function, Setting Range Factory Default
EVT3 hysteresis type Reference Value
• Selects EVT3 output hysteresis type (Medium or Reference Value). (Fig 7.5-1, p.40)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the P control action.
: Medium Value Sets the same value for both ON and OFF sides in relation to EVT3 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to EVT3 value.
Both ON and OFF sides need to be set individually.
EVT3 ON side 10 mV
• Sets the span of EVT3 ON side. (Fig 7.5-1, p.40) If (Medium Value) is selected in [EVT3 hysteresis type], the span of ON/OFF side will be the same value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the P control action.
• Setting range: 0 to 200 mV
EVT3 OFF side 10 mV
• Sets the span of EVT3 OFF side. (Fig 7.5-1, p.40)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the P control action, or if (Medium Value) is selected in [EVT3 hysteresis type].
• Setting range: 0 to 200 mV
EVT3 ON delay time 0 sec.
• Sets EVT3 delay time. The EVT3 output does not turn ON after the input value exceeds the EVT3 value until the time set in [EVT3 ON delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
Page 37
37
Character Setting Item, Function, Setting Range Factory Default
EVT3 OFF delay time 0 sec.
• Sets EVT3 delay time. The EVT3 output does not turn OFF after the input value exceeds the EVT3 value until the time set in [EVT3 OFF delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
EVT3 proportional cycle 30 sec.
• Sets EVT3 proportional cycle.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 1 to 300 sec.
EVT3 output high limit 100%
• Sets EVT3 output high limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: EVT3 output low limit to 100%
EVT3 output low limit 0%
• Sets EVT3 output low limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 0% to EVT3 output high limit
Output ON time when EVT3 output ON 0 sec
• Sets Output ON time when EVT3 output is ON. If ON time and OFF time are set, EVT3 output can be turned ON/OFF in a configured cycle when EVT3 output is turned ON.(Fig. 7.5-2, p.40)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
Page 38
38
Character Setting Item, Function, Setting Range Factory Default
Output OFF time when EVT3 output ON 0 sec.
• Sets Output OFF time when EVT3 output is ON. If ON time and OFF time are set, EVT3 output can be turned ON/OFF in a configured cycle when EVT3 output is turned ON.( Fig 7.5-2, p.40)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
EVT3 ORP input error alarm EVT type No action
• Selects EVT output type (except EVT3 type) in order to assess
EVT3 ORP input error alarm.
• Available only when (ORP input error alarm output) is selected in [EVT3 type]
: EVT1 action : EVT2 action : No action : EVT4 action
EVT3 ORP input error alarm span when EVT output ON
0 mV
• Sets span to assess EVT3 ORP input error alarm when EVT output (selected in [EVT3 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT3 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT3 ORP input error alarm time when EVT output ON
0 sec.
• Sets time to assess EVT3 ORP input error alarm when EVT output (selected in [EVT3 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT3 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 39
39
Character Setting Item, Function, Setting Range Factory Default
EVT3 ORP input error alarm span
when EVT output OFF
0 mV
• Sets span to assess EVT3 ORP input error alarm when EVT output (selected in [EVT3 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT3 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT3 ORP input error alarm time when EVT output OFF
0 sec.
• Sets time to assess EVT3 ORP input error alarm when EVT output (selected in [EVT3 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT3 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
EVT3 cycle variable range 50.0%
• Sets EVT3 cycle variable range.
• Available when (ORP input low limit action) or
(ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 1.0 to 100.0%
EVT3 cycle extended time 0 sec.
• Sets EVT3 cycle extended time.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 0 to 300 sec.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 40
40
EVT3 Action
EVT3 Type P Control Action ON/OFF Control Action
ORP input low limit action
EVT3 proportional band
ON
OFF
EVT3 value
If Medium Value is selected in [EVT3 hysteresis type]:
EVT3 ON sides
ON
OFF
EVT3 value If Reference Value is selected in [EVT3 hysteresis type]:
EVT3 ON side EVT3 OFF side
ON
OFF
EVT3 value
ORP input high limit action
EVT3 proportional band
ON
OFF
EVT3 value
If Medium Value is selected in [EVT3 hysteresis type]:
EVT3 ON sides
ON
OFF
EVT3 value
If Reference Value is selected in [EVT3 hysteresis type]:
ON
OFF
EVT3 value
(Fig 7.5-1)
Timing chart of EVT3 output ON time and OFF time while in EVT3 output ON
(Fig. 7.5-2)
OFF
ON
OFF
ON
Actual EVT3 output
EVT3 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual EVT3 output turning OFF.
EVT3ON side
EVT3OFF side
Page 41
41
7.6 EVT4 Action Group
EVT4 Action group is indicated only when EVT3, EVT4 Outputs (EVT3 option) are/is ordered. To enter the EVT4 Action group, follow the procedure below.
1
Press the key 5 times in ORP Display mode or Cleansing Output mode.
2
Press the key.
The unit proceeds to the EVT4 Action group, and ”EVT4 type“ will appear.
Character Setting Item, Function, Setting Range Factory Default
EVT4 type No action
• Selects an EVT4 output (Contact output 4) type. (Fig 7.6-1, p.46)
Note: If EVT4 type is changed, EVT4 value defaults to 0.
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT4 value 0 mV
• Sets EVT4 value. (Fig 7.6-1, p.46)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Setting range: Input low limit to Input high limit
EVT4 proportional band 0 mV
• Sets EVT4 proportional band. (Fig 7.6-1, p.46) ON/OFF control action when set to 0.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Setting range: 0 to 4000 mV
EVT4 reset 0 mV
• Sets EVT4 reset value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range: 200 mV
Page 42
42
Character Setting Item, Function, Setting Range Factory Default
EVT4 hysteresis type Reference Value
• Selects EVT4 output hysteresis type (Medium or Reference Value). (Fig 7.6-1, p.46)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the P control action.
: Medium Value Sets the same value for both ON and OFF sides in relation to EVT4 value. Only ON side needs to be set.
: Reference Value
Sets individual values for ON and OFF sides in relation to EVT4 value.
Both ON and OFF sides need to be set individually.
EVT4 ON side 10 mV
• Sets the span of EVT4 ON side. (Fig 7.6-1, p.46) If (Medium Value) is selected in [EVT4 hysteresis type], the span of ON/OFF side will be the same value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the P control action.
• Setting range: 0 to 200 mV
EVT4 OFF side 10 mV
• Sets the span of EVT4 OFF side. (Fig 7.6-1, p.46)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the P control action, or if (Medium Value) is selected in [EVT4 hysteresis type].
• Setting range: 0 to 200 mV
EVT4 ON delay time 0 sec.
• Sets EVT4 delay time. The EVT4 output does not turn ON after the input value exceeds the EVT4 value until the time set in [EVT4 ON delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
Page 43
43
Character Setting Item, Function, Setting Range Factory Default
EVT4 OFF delay time 0 sec.
• Sets EVT4 delay time. The EVT4 output does not turn OFF after the input value exceeds the EVT4 value until the time set in [EVT4 OFF delay time] elapses.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the P control action.
• Setting range: 0 to 10000 sec.
EVT4 proportional cycle 30 sec.
• Sets EVT4 proportional cycle.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type].
• Not available for the ON/OFF control action.
• Setting range: 1 to 300 sec.
EVT4 output high limit 100%
• Sets EVT4 output high limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range: EVT4 output low limit to 100%
EVT4 output low limit 0%
• Sets EVT4 output low limit value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range: 0% to EVT4 output high limit
Output ON time when EVT4 output ON 0 sec
• Sets Output ON time when EVT4 output is ON. If ON time and OFF time are set, EVT4 output can be turned ON/OFF in a configured cycle when EVT4 output is turned ON.(Fig. 7.6-2, p.46)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
Page 44
44
Character Setting Item, Function, Setting Range Factory Default
Output OFF time when EVT4 output ON 0 sec.
• Sets Output OFF time when EVT4 output is ON. If ON time and OFF time are set, EVT4 output can be turned ON/OFF in a configured cycle when EVT4 output is turned ON.(Fig. 7.6-2, p.46)
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for P control action
• Setting range: 0 to 10000 sec.
EVT4 ORP input error alarm EVT type No action
• Selects EVT output type (except EVT4 type) in order to assess
EVT4 ORP input error alarm.
• Available only when (ORP input error alarm output) is selected in [EVT4 type]
: EVT1 action : EVT2 action : EVT3 action : No action
EVT4 ORP input error alarm span when EVT output ON
0 mV
• Sets span to assess EVT4 ORP input error alarm when EVT output (selected in [EVT4 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT4 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT4 ORP input error alarm time when EVT output ON
0 sec.
• Sets time to assess EVT4 ORP input error alarm when EVT output (selected in [EVT4 ORP input error alarm EVT type]) is ON.
• Available only when (ORP input error alarm output) is
selected in [EVT4 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 45
45
Character Setting Item, Function, Setting Range Factory Default
EVT4 ORP input error alarm span
when EVT output OFF
0 mV
• Sets span to assess EVT4 ORP input error alarm when EVT output (selected in [EVT4 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT4 type]
• Setting range: 0 to 2000 mV
When set to 0 (zero), ORP input error alarm is disabled.
EVT4 ORP input error alarm time when EVT output OFF
0 sec.
• Sets time to assess EVT4 ORP input error alarm when EVT output (selected in [EVT4 ORP input error alarm EVT type]) is OFF.
• Available only when (ORP input error alarm output) is
selected in [EVT4 type]
• Setting range: 0 to 10000 seconds or minutes (*)
When set to 0 (zero), ORP input error alarm is disabled.
EVT4 cycle variable range 50.0%
• Sets EVT4 cycle variable range.
• Available when (ORP input low limit action) or
(ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range: 1.0 to 100.0%
EVT4 cycle extended time 0 sec.
• Sets EVT4 cycle extended time.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range: 0 to 300 sec.
(*) Time unit follows the selection in [ORP input error alarm time unit].
Page 46
46
EVT4 Action
EVT4 Type P Control Action ON/OFF Control Action
ORP input low limit action
EVT4 proportional band
ON
OFF
EVT4 value
If Medium Value is selected in [EVT4 hysteresis type]:
EVT4 ON sides
ON
OFF
EVT4 value If Reference Value is selected in [EVT4 hysteresis type]:
EVT4 ON side EVT4 OFF side
ON
OFF
EVT4 value
ORP input high limit action
EVT4 proportional band
ON
OFF
EVT4 value
If Medium Value is selected in [EVT4 hysteresis type]:
EVT4 ON sides
ON
OFF
EVT4 value
If Reference Value is selected in [EVT4 hysteresis type]:
ON
OFF
EVT4 value
(Fig 7.6-1)
Timing chart of EVT4 output ON time and OFF time while in EVT4 output ON
(Fig. 7.6-2)
OFF
ON
OFF
ON
Actual EVT4 output
EVT4 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual EVT4 output turning OFF.
EVT4ON side
EVT4OFF side
Page 47
47
7.7 Special Function Group
To enter the Special Function group, follow the procedure below.
1
Press the key 4 times in ORP Display mode or Cleansing Output mode. If EVT3, EVT4 Outputs (EVT3 option) are/is ordered, press the key 6 times in ORP Display mode or Cleansing Output mode.
2
Press the key.
The unit enters the Special Function group, and the “Set Value Lock” item will appear.
Character Setting Item, Function, Setting Range Factory Default
Set value lock Unlock
• Locks the set values to prevent setting errors.
(Unlock): All set values can be changed. (Lock 1) : None of the set values can be changed. (Lock 2) : Only EVT1, EVT2, EVT3, EVT4 values can be
changed.
(Lock 3): All set values – except Adjustment value, Span
sensitivity correction value, Transmission output Zero adjustment value, Transmission output Span adjustment value – can be temporarily changed. However, they revert to their previous value after the power is turned off because they are not saved in the non-volatile IC memory. Do not change setting items (EVT1, EVT2, EVT3, EVT4 types). If they are changed, they will affect other setting items. Be sure to select Lock 3 when changing the set value frequently via software communication. (If the value set by the software communication is the same as the value before the setting, the value will not be written in the non-volatile IC memory.)
Communication protocol Shinko protocol
• Selects communication protocol.
• Available when the Serial communication (C5) option is ordered.
: Shinko protocol : Modbus ASCII mode : Modbus RTU mode
Instrument number 0
• Sets the instrument number of this unit. (The instrument numbers should be set one by one when multiple instruments are connected, otherwise communication is impossible.)
• Available when the Serial communication (C5) option is ordered.
• Setting range: 0 to 95
Communication speed 9600 bps
• Selects a communication speed equal to that of the host computer.
• Available when the Serial communication (C5) option is ordered.
: 9600 bps : 19200 bps : 38400 bps
Page 48
48
Character Setting Item, Function, Setting Range Factory Default
Data bit/Parity 7 bits/Even
• Selects data bit and parity.
• Available when the Serial communication (C5) option is ordered.
: 8 bits/No parity : 7 bits/No parity : 8 bits/Even : 7 bits/Even : 8 bits/Odd : 7 bits/Odd
Stop bit 1 bit
• Selects the stop bit.
• Available when the Serial communication (C5) option is ordered.
: 1 bit : 2 bits
Transmission output type ORP transmission
• Selects the transmission output type.
: ORP transmission
: EVT1 MV transmission : EVT2 MV transmission : EVT3 MV transmission : EVT4 MV transmission
Transmission output high limit ORP transmission: 2000 mV
MV transmission: 100.0%
• Sets the Transmission output high limit value. (This value correponds to
20 mA DC output.) If Transmission output high limit and low limit are set to the same value, transmission output will be fixed at 4 mA DC.
• Setting range: ORP transmission: Transmission output low limit to 2000 mV
MV transmission: Transmission output low limit to 100.0%
Transmission output low limit ORP transmission: -2000 mV
MV transmission: 0.0%
• Sets the Transmission output low limit value. (This value correponds to
4 mA DC output.) If Transmission output high limit and low limit are set to the same value, transmission output will be fixed at 4 mA DC.
• Setting range: ORP transmission: -2000 mV to Transmission output high limit
MV transmission: 0.0% to Transmission output high limit
Page 49
49
Character Setting Item, Function, Setting Range Factory Default
Transmission output status in Adjustment mode, Span sensitivity correction mode
Last value HOLD
• Selects Transmission output status in Adjustment mode or Span
sensitivity correction mode.
• Available only when Transmission output (TA option) is ordered.
• Selection range
: Last value HOLD (Retains and outputs the last value
before adjustment or span sensitivity correction.) : Set value HOLD (Outputs the value set in [Set value HOLD].) : Measured value (Outputs the value measured in
Adjustment mode or Span sensitivity correction mode.)
Set value HOLD
ORP transmission: 0 mV
MV transmission: 0.0%
• Sets the Set value HOLD.
• Available only when (Set value HOLD) is selected in [Transmission output status in Adjustment mode, Span sensitivity correction mode].
• Setting range: ORP transmission: -2000 to 2000 mV MV transmission: 0.0 to 100.0%
Backlight selection All are backlit
• Selects the display to backlight.
: All are backlit.
: ORP Display : Setting Display : Action indicators : ORP Display + Setting Display : ORP Display + Action indicators : Setting Display + Action indicators
ORP color Red
• Selects a color for the ORP Display.
: Green
: Red : Orange : ORP color changes continuously.
The ORP Display color changes according to [ORP color reference value] and [ORP color range] settings.
• When ORP is lower than [ORP color reference value] – [ORP color range]: Orange
• When ORP is within [ORP color reference value] [ORP color range]: Green
• When ORP is higher than [ORP color reference value] + [ORP color range]: Red
(Fig. 7.7-1)
Orange
Green
Red
: ORP color reference value
: ORP color range
Hys
Hys
Hys
Page 50
50
Character Setting Item, Function, Setting Range Factory Default
ORP color reference value 0 mV
• Sets a reference value for ORP color to be green when (ORP color changes continuously) is selected in [ORP color].
• Setting range: 2000 mV
ORP color range
200 mV
• Sets a range for ORP color to be green when (ORP color changes continuously) is selected in [ORP color].
• Setting range: 2000 mV
Backlight time
0 minutes
• Sets time to backlight from no operation status until backlight is switched off.
When set to 0 (zero), the backlight remains ON.
Backlight relights by pressing any key while backlight is OFF.
• Setting range: 0 to 99 minutes
Bar graph indication
No indication
• Selects bar graph indication.
: No indication : Transmission output
Segments will light in accordance with the output.
Scale is -5 to 105%. Segments will light from left to right in accordance
with the output.
(Fig. 7.7-2)
EVT output when input errors occur Disabled
• If input errors occur, such as ORP Combined Electrode Sensor is dis-
connected or short-circuited, EVT output can be Enabled or Disabled. If “Enabled” is selected, EVT output will be maintained when input errors occur. If “Disabled” is selected, EVT output will be turned OFF when input errors occur.
: Disabled
: Enabled
Setting Display indication
No indication
• Selects an item to be indicated on the Setting Display.
: No indication : EVT1 value : EVT2 value : EVT3 value
: EVT4 value
[When output is 50%]
-5% 50%
105%
Li
ghts
from left
to
right
according to t
he outp
u
t.
Page 51
51
Character Setting Item, Function, Setting Range Factory Default
Number of cleansing cycles 0 (Continuous cleansing)
• Sets the number of cleansing outputs. (Fig. 7.7-3)
• Available for this setting item and all following items if (Cleansing output) is selected in any of [EVT1 to EVT4 type (pp. 23, 29, 35, 41)].
• Setting range: 0 to 10 (0: Continuous cleansing)
Cleansing interval 360 minutes
• Sets an interval between cleansings. (Fig. 7.7-3)
• Setting range: 60 to 3000 minutes
Cleansing time 600 sec.
• Sets the cleansing output time for the cleansing output interval.(Fig. 7.7-3)
• Setting range: 1 to 1800 seconds
Restore time after cleansing 600 sec.
• Sets the time to restore units to normal operation after cleansing output. (Fig. 7.7-3)
• Setting range: 1 to 1800 seconds
ORP input error alarm time unit Second(s)
• Selects ORP input error alarm time unit.
• Selection item : Second(s)
: Minute(s)
Ceansing Output Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing
(Fig. 7.7-3)
Programmed
action
Programmed
action
Programmed
action
Page 52
52
8. Calibration
Adjustment Mode and Span Sensitivity Correction Mode are described below.
8.1 Adjustment Mode
When using a brand-new sensor, please calibrate in Adjustment Mode.
By setting the adjustment value, calibrates ORP value indicated on the AER-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The unit cannot enter Adjustment mode in the following cases:
• When (Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.47)]
• When (Cleansing output) is selected in any of [EVT1 to EVT4 type (pp. 23, 29, 35, 41)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines the procedure for calibration.
(1) When selecting (Last value HOLD) in [Transmission output status in
Adjustment Mode, Span Sensitivity Correction Mode (p.49)], select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
(2) Press and hold the key and key (in that order) together for 3 seconds in
ORP Display mode or Cleansing Output mode. The unit enters Adjustment Mode, and indicates the following.
Display Indication
ORP Display and ORP value are indicated alternately. Setting Display The adjustment value is indicated.
(3) Immerse the ORP Combined Electrode Sensor in the standard solution
(Quinhydrone potential difference 260 mV).
(4) Set an adjustment value with the or key so that ORP value is
approximately 260 mV (at 20 ). For other temperature and electric potentials, refer to the temperature characteristics of your standard solution. Adjustment range: -200 to 200 mV
(5) Press the key.
Adjustment mode is complete, and the unit reverts to ORP Display mode or Cleansing Output mode.
Page 53
53
8.2 Span Sensitivity Correction Mode
When calibrating periodically, please calibrate in Span sensitivity correction mode.
By setting the Span sensitivity correction value in percentage, calibrates ORP value indicated on the AER-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
The unit cannot enter Span sensitivity correction mode in the following cases:
• When (Lock 1), (Lock 2) or (Lock 3) is selected in [Set value lock (p.47)]
• When (Cleansing output) is selected in any of [EVT1 to EVT4 type (pp. 23,
29, 35, 41)], and cleansing action is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines the procedure for calibration.
(1) When selecting (Last value HOLD) in [Transmission output status in
Adjustment Mode, Span Sensitivity Correction Mode (p.49)], select it while the ORP Combined Electrode Sensor is being immersed in the solution currently calibrated.
(2) Press and hold the key and key (in that order) together for 3 seconds in
ORP Display mode or Cleansing Output mode. The unit enters Span sensitivity correction mode, and indicates the following.
Display Indication
ORP Display and ORP value are displayed alternately. Setting Display The Span sensitivity correction value is displayed.
(3) Immerse the ORP Combined Electrode Sensor in the standard solution
(Quinhydrone potential difference 260 mV).
(4) Set a span sensitivity correction value with the or key so that ORP value
is approximately 260 mV (at 20 ). For other temperature and electric potentials, refer to the temperature characteristics of your standard solution. Setting range: 50 to 150%
(5) Press the key.
Span sensitivity correction mode is complete, and the unit reverts to ORP Display mode or Cleansing Output mode.
Page 54
54
8.3 Transmission Output Adjustment Mode
Fine adjustment of Transmission output is performed. This ORP meter is adjusted at the factory, however, differences may occur between the indication value of the connected equipment (recorders, etc.) and output value of this instrument. In this case, perform Transmission output Zero adjustment and Span adjustment.
The unit cannot enter Transmission output Zero adjustment mode in the following cases:
• During Adjustment mode or Span sensitivity correction mode
• When (Lock 1) ( Lock 2) or (Lock 3) is selected in [Set
value lock (p.47)]
• When (Cleansing output) is selected in any of EVT1 to EVT4 type (pp. 23,
29, 35, 41) using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
The following outlines adjustment procedure. (1) Press and hold the and key (in that order) together for approx. 3 seconds
in ORP Display mode or Cleansing Output mode. The unit enters Transmission output Zero adjustment mode, and indicates the following:
Display Indication Contents
ORP Display
Setting Display Transmission output Zero adjustment value
(2) Set Transmission output Zero adjustment value with the , keys, while
viewing the value indicated on the connected equipment (recorders, etc.). Setting range: 5.00% of Transmission output span
(3) Press the key.
The unit enters Transmission output Span adjustment mode, and indicates the following:
Display Indication Contents
ORP Display
Setting Display Transmission output Span adjustment value
(4) Set Transmission output Span adjustment value with the , keys, while
viewing the value indicated on the connected equipment (recorders, etc.). Setting range: 5.00% of Transmission output span
(5) Press the key.
The unit reverts to the Transmission output Zero adjustment mode. Repeat steps (2) to (5) if necessary.
(6) To finish the Transmission output adjustment, press the key in Transmission
output Span adjustment mode.
The unit reverts to ORP Display mode or Cleansing Output mode.
Page 55
55
9. Measurement
9.1 Starting Measurement
After mounting to the control panel, wiring, setup and calibration are complete, turn the power to the instrument ON.
For approx. 4 seconds after the power is switched ON, the following characters are indicated on the ORP Display and Setting Display.
ORP Display Setting Display
Unlit
During this time, all outputs are in OFF status, and action indicators go off. After that, measurement starts, indicating the item selected in [Backlight selection (p.49)] [ORP color (p.49)] [Bar graph indication (p.50)] [Setting Display indication (p.50] in the Special Function group.
Page 56
56
9.2 EVT1 to EVT4 Outputs
If (ORP input low limit action), (ORP input high limit action) is selected in [EVT1 type (p.23)], the following action is activated. (Fig. 9.2-1) The same applies to EVT2, EVT3 and EVT4.
EVT1 Action
EVT1 Type P Control Action ON/OFF Control Action
ORP input low limit action
EVT1 proportional band
ON
OFF
EVT1 value
If Medium Value is selected in [EVT1 hysteresis type]:
EVT1 ON sides
ON
OFF
EVT1 value If Reference Value is selected in [EVT1 hysteresis type]:
EVT1 ON side EVT1 OFF side
ON
OFF
EVT1 value
ORP input high limit action
EVT1 proportional band
ON
OFF
EVT1 value
If Medium Value is selected in [EVT1 hysteresis type]:
EVT1 ON sides
ON
OFF
EVT1 value
If Reference Value is selected in [EVT1 hysteresis type]:
ON
OFF
EVT1 value
(Fig. 9.2-1)
EVT1 ON side
EVT1 OFF side
Page 57
57
P Control Action Within the proportional band, the manipulated variable is output in proportion to the deviation between the EVT1 value and ORP value.
EVT1 Type Description
ORP input low limit action
If ORP value is lower than [EVT1 value – EVT1 proportional band], EVT1 output is turned ON. If ORP value enters within the proportional band, EVT1 output is turned ON/OFF in EVT1 proportional cycles. If ORP value exceeds the EVT1 value, EVT1 output is turned OFF.
ORP input high limit action
If ORP value is higher than [EVT1 value + EVT1 proportional band], EVT1 output is turned ON. If ORP value enters within the proportional band, EVT1 output is turned ON/OFF in EVT1 proportional cycles. If ORP value drops below the EVT1 value, EVT1 output is turned OFF.
ON/OFF Control Action
EVT1 Type Description
ORP input low limit action
If
ORP
value is lower than EVT1 value, EVT1 output
isturned
ON. If ORP value exceeds the EVT1 value, EVT1 output is turned OFF.
ORP input high limit action
If
ORP
value is higher than EVT1 value, EVT1 output
is
turned
ON. If ORP value drops below the EVT1 value, EVT1 output is turned OFF.
If ON and OFF time are set in [Output ON/OFF Time when EVT1 Output ON (p.25, 26)], and when EVT1 output is turned ON, EVT1 output is turned ON/OFF in a configured cycle.
Timing chart of Output ON time and OFF time while in EVT1 output ON
(Fig. 9.2-2)
EVT output status can be read by the status flag (EVT1, EVT2, EVT3, EVT4 output bit) in Serial communication (C5 option).
EVT output status, when input errors occur, differs depending on the selection in [EVT output when input errors occur (p.50)].
• If (Disabled) is selected, EVT output is turned OFF when input errors occur.
• If (Enabled) is selected, EVT output is maintained when input errors occur.
OFF
ON
OFF
ON
Actual EVT1 output
EVT1 output to which ON time and OFF time are set.
ON time
ON time
ON time
OFF time
OFF time
ON time is turned OFF, caused by the actual EVT1 output turning OFF.
Page 58
58
9.3 Setting EVT1 to EVT4 Values
EVT1 to EVT4 values are set in Simple Setting mode. These setting items are the same as those in EVT1 to EVT4 Action groups.
To enter Simple Setting mode, follow the procedure below.
1
Press the key in ORP Display mode or Cleansing Output mode. “EVT1 value” will appear.
2
Set each setting item with the or key, and register the value with the key.
Character Setting Item, Function, Setting Range Factory Default
EVT1 value 0 mV
• Sets EVT1 value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT1 type (p.23)].
• Setting range: Input low limit to Input high limit
EVT2 value 0 mV
• Sets EVT2 value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT2 type (p.29)].
• Setting range: Input low limit to Input high limit
EVT3 value 0 mV
• Sets EVT3 value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT3 type (p.35)]. Available when EVT3, EVT4 Outputs (EVT3 option) are/is ordered.
• Setting range: Input low limit to Input high limit
EVT4 value 0 mV
• Sets EVT4 value.
• Available when (ORP input low limit action) or (ORP input high limit action) is selected in [EVT4 type (p.41)]. Available when EVT3, EVT4 Outputs (EVT3 option) are/is ordered.
• Setting range: Input low limit to Input high limit
3
Press the key. The unit reverts to ORP Display mode or Cleansing Output mode.
Page 59
59
9.4 Cleansing Output
If (Cleansing output) is selected in any of [EVT1 to EVT4 type (pp. 23, 29, 35, 41)], the unit will enter Cleansing Output mode.
An EVT output (for which the cleansing output is selected) will turn ON during the configured cleansing time. When the cleansing interval finishes after restore time has passed, this is counted as one cleansing time period, and the configured number of cleansing cycles will be repeated.
While cleansing is being performed, other outputs are in OFF status. ORP values are retained.
When cleansing is not being performed, normal operation continues. When power is turned ON again, the unit starts from the first of the cleansing time periods. After the configured number of cleansing cycles are finished, the EVT output (for which the cleansing output is selected) is turned OFF, and other outputs perform their programmed operations, however, they are in Cleansing Output mode.
Cleansing Output Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing (Fig. 9.4-1)
• If one or more of EVT cleansing output are selected during the cleansing time period of remaining EVT(s), then the Number of cleansing cycles, Cleansing interval, Cleansing time, and Restore time after cleansing settings for the selected EVT(s) will not take effect – however, those cleansing output(s) will perform using the remaining EVT(s) settings.
• During Adjustment mode or Span sensitivity correction mode, if cleansing action
initiates after restore time has passed, the cleansing action will not be performed in the current session.
• If the number of cleansing cycles is changed in [Number of cleansing cycles] during cleansing action, the new number will be enabled from the next cleansing time period.
If any output other than (Cleansing output) is selected in [EVT1 to EVT4 type (pp. 23, 29, 35, 41)], the unit will revert to ORP Display mode.
Programmed action
Programmed action
Programmed action
Page 60
60
9.5 Manual Cleansing Mode
By pressing the and keys simultaneously for 3 seconds, the units enters Manual cleansing mode. In Manual cleansing mode, cleansing action is performed using “Cleansing time” and “Restore time after cleansing”. After cleansing action is complete, the unit automatically reverts to Cleansing Output mode.
Manual cleansing mode will not be accessible if programmed cleansing is currently being performed. During Manual cleansing mode, if programmed cleansing action initiates after restore time has passed, the programmed cleansing action will not be performed in the current session. Cleansing action by Manual cleansing mode is counted as one cleansing time period.
Manual Cleansing Mode Action
Cleansing Cleansing Cleansing
Cno: Number of cleansing cycles Cint: Cleansing interval Ctm: Cleansing time Rtm: Restore time after cleansing
(Fig. 9.5-1)
During cleansing action, Manual cleansing mode via key operation is invalidated – Programmed cleansing mode takes priority.
During Manual cleansing mode, the programmed cleansing action will be invalidated. Cleansing action by Manual cleansing mode is counted as one cleansing time period – Programmed cleansing action will perform once Manual cleansing
time period has elapsed.
Programmed action
Programmed action
Programmed action
Key operation
Page 61
61
9.6 ORP Input Error Alarm
ORP input error alarm is used for detecting actuator trouble. Even if ORP input error alarm time has elapsed, and if ORP input does not become higher than ORP input error alarm span, the unit assumes that actuator trouble has occurred, and writes Status flag 2. In Serial communication, status can be read by reading Status flag 2 (EVT1, EVT2, EVT3, EVT4 output flags). If (ORP input error alarm output) is selected in [EVT1 type (p.23)], EVT1 output will be turned ON. The same applies to EVT2, EVT3 and EVT4.
ORP input error alarm is disabled in the following cases:
• During Adjustment mode or Span sensitivity correction mode
• When (Cleansing output) is selected in any of EVT1 to EVT4 type (p. 23, 29, 35, 41), and cleansing is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
• When ORP input error alarm time is set to 0 seconds (or minutes) or ORP input error alarm span is set to 0 mV.
Page 62
62
9.7 Cycle Automatic Variable Function
If deviation between EVT value and measured value enters EVT cycle variable range, the proportional cycle will be automatically extended in accordance with the deviation. Proportional action OFF time will be extended, and ON / OFF ratio will be adjusted. However, if EVT cycle extended time is set to 0 (zero), this function will be disabled.
(Fig. 9.7-1)
Large
EVTx
EVTx cycle
Small
0 sec.
ON time
OFF
ON
C
ON tf1OFF ON
tf2
OFF
Ctf1
C
C
C C C
OFF time
?
C C
C
OFF
ON
Deviation
Control direction
EVTx proportional band
Cycle fixed range
EVTx cycle variable range
Time
(Seconds)
extended time
proportional cycle
OFF time
ON time
Cycle delay time (Ctf1, Ctf2)
Proportional cycle of usual P control action
Usual P control action
EVTx output
Cycle is constantly fixed.
Usual P control action
EVTx output
EVTx cycle variable range
Action when EVTx cycle variable range is set.
C: EVTx proportional cycle
Ctf2
Ctf1, Ctf2: Cycle delay time
tf1, tf2: Extended OFF time
EVTx value
Page 63
63
9.8 Transmission Output
Converting ORP or MV to analog signal every input sampling period, outputs in current. If Transmission output high limit and low limit are set to the same value, Transmission output will be fixed at 4 mA DC.
Resolution 12000
Current 4 to 20 mA DC (Load resistance: Max 500 )
Output accuracy Within 0.3% of Transmission output span
Page 64
64
10. Specifications
10.1 Standard Specifications Rating
Rated Scale Rated Scale Rated Scale Rated Scale
ORP Combined Electrode Sensor
-2000 to 2000 mV 1 mV
Input ORP Combined Electrode Sensor Supply Voltage Model AER-101-ORP AER-101-ORP 1
Supply voltage
100 to 240 V AC 50/60 Hz
24 V AC/DC
50/60 Hz Allowable voltage fluctuation range
85 to 264 V AC 20 to 28 V AC/DC
General Structure
External Dimensions 48 x 96 x 98.5 mm (W x H x D) Mounting Flush (Applicable panel thickness: 1 to 8 mm) Case Material: Flame-resistant resin, Color: Black Front Panel Membrane sheet Drip-proof/Dust-proof IP66 (for front panel only) Indication Structure Display
ORP Display
11-segment LCD display 5-digits Backlight: Red/Green/Orange Character size: 14.0 x 5.4 mm (H x W)
Setting Display
11-segment LCD display 5-digits Backlight: Green
Character size: 10.0 x 4.6 mm (H x W) Output Display
22-segment LCD display Bar graph
Backlight: Green
Action indicators: Backlight: Orange color
EVT1 EVT1 output (Contact output 1) ON: Lit EVT2 EVT2 output (Contact output 2) ON: Lit EVT3 EVT3 output (Contact output 3) ON: Lit EVT4 EVT4 output (Contact output 4) ON: Lit T/R Serial communication TX output
(transmitting): Lit
LOCK Lock 1, Lock 2, Lock 3 selected: Lit
Setting Structure Input system using membrane sheet key
Indication Performance
Repeatability Within 5 mV (at equivalent input) Linearity Within 5 mV (at equivalent input) Input Sampling Period 125 ms Time Accuracy 1% of setting time
Page 65
65
Standard Functions
Adjustment Mode
For successful measurement of ORP, ORP value in the sensor location, electrode performance and standard solution accuracy respectively play an important role for obtaining reliable data. By setting the adjustment value, calibrates ORP value indicated on the AER-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260
mV). Span Sensitivity Correction Mode
By setting the Span sensitivity correction value in percentage, calibrates ORP value indicated on the AER-101-ORP to read 260 mV (at 20 ) when immersing the ORP Combined Electrode Sensor in the standard solution (Quinhydrone potential difference 260 mV).
EVT Output
Setting Accuracy Same as indication accuracy Output Action
P control: When setting proportional band to any value
other than 0.
ON/OFF control: When setting proportional band to 0. EVT proportional band 0 to 4000 mV EVT proportional cycle 1 to 300 sec.
EVT ON side, OFF side 0 to 200 mV
EVT output high, low limit 0 to 100%
Type
Selectable by the keypad from the following.
• No action
• ORP input low limit action
• ORP input high limit action
• Cleansing output
• ORP input error alarm output
Output Relay contact 1a
Control capacity
3 A 250 V AC (resistive load) 1 A 250 V AC (inductive load, cos =0.4)
Electrical life 100,000 cycles
EVT ON Delay Time
0 to 10000 sec.
EVT OFF Delay Time
0 to 10000 sec.
Output ON Time/ OFF Time when
EVT Output ON
If ON time and OFF time are set, the output will be turned
ON/OFF at constant intervals when EVT output is ON.
Page 66
66
Cleansing Output Cleansing Output Mode
If (Cleansing output) is selected in any of [EVT1 to EVT4 type (pp. 23, 29, 35, 41)], the unit will enter Cleansing Output mode.
An EVT output (for which the cleansing output is selected) will turn ON during the configured cleansing time. When the cleansing interval finishes after restore time has passed, this is counted as one cleansing time period, and the configured number of cleansing cycles will be repeated.
While cleansing is being performed, other outputs are in OFF status. ORP values are retained.
When cleansing is not being performed, normal operation continues. When power is turned ON again, the unit starts from the first of the cleansing time periods. After the configured number of cleansing cycles are finished, the EVT output (for which the cleansing output is selected) is turned OFF, and other outputs perform their programmed operations, however, they are in Cleansing Output mode.
Manual Cleansing Mode
By pressing the and keys simultaneously for 3 seconds, the units enters Manual cleansing mode. In Manual cleansing mode, cleansing action is performed using “Cleansing time” and “Restore time after cleansing”. After cleansing action is complete, the unit automatically reverts to Cleansing Output mode. Manual cleansing mode (by keypad operation) will not be accessible if programmed cleansing is currently being performed. During Manual cleansing mode, if programmed cleansing action initiates after restore time has passed, the programmed cleansing action will not be performed in the current session.
ORP Input Error Alarm
Detects actuator trouble. Even if ORP input error alarm time has elapsed, and if ORP input does not become higher than ORP input error alarm span, the unit assumes that actuator trouble has occurred, and writes Status flag 2. In Serial communication, status can be read by reading Status flag 2 (EVT1, EVT2, EVT3, EVT4 output flags). If (ORP input error alarm output) is selected in [EVT1 type (p.23)], EVT1 output is turned ON. The same applies to EVT2, EVT3 and EVT4.
ORP input error alarm is disabled in the following cases:
• During Adjustment mode or Span sensitivity correction mode
• When (Cleansing output) is selected in any of EVT1 to EVT4 type (p. 23, 29, 35, 41), and cleansing is performing using the ‘Cleansing time’ and ‘Restore time after cleansing’ settings.
• When ORP input error alarm time is set to 0 seconds (or minutes) or ORP
input error alarm span is set to 0 mV.
Page 67
67
Cycle Automatic Variable Function
If deviation between EVT value and measured value enters EVT cycle variable range, the proportional cycle will be automatically extended in accordance with the deviation. Proportional action OFF time will be extended, and ON / OFF ratio will be adjusted. However, if EVT cycle extended time is set to 0 (zero), this function will be disabled.
Transmission Output Function
Transmission Output
Converting ORP value or MV to analog signal every input sampling period, outputs in current. If Transmission output high limit and low limit are set to the same value, Transmission output will be fixed at 4 mA DC. Resolution
12000 Current 4 to 20 mA DC(Load resistance: Max 500 ) Output
accuracy
Within 0.3% of Transmission output span
Transmission output adjustment
Fine adjustment of Transmission output is performed via Transmission output Zero adjustment and Span adjustment.
Transmission output status in Adjustment mode, Span sensitivity correction mode
Selects Transmission output status in Adjustment mode or Span sensitivity correction mode. Last value HOLD: Retains and outputs the last value
before adjustment or span sensitivity correction.
Set value HOLD: Outputs the value set in [Set value HOLD].
Measured value: Outputs the value measured in
Adjustment mode or Span sensitivity correction mode.
Insulation, Dielectric Strength
Circuit Insulation Configuration
: When the option is ordered.
Insulation Resistance: 10 M or more, at 500 V DC
Dielectric Strength
Power terminal - ground (GND): 1.5 kV AC for 1 minute Input terminal - ground (GND): 1.5 kV AC for 1 minute Input terminal - power terminal: 1.5 kV AC for 1 minute
GND
Serial
communication
Transmssion
output
Input
EVT1 output
EVT2 output
EVT3 output
EVT4 output
Power supply
Ele
ctrically insulated
Page 68
68
Attached Functions
Set Value Lock Lock 1: None of the set values can be changed.
Lock 2: Only EVT1, EVT2, EVT3, EVT4 values can be
changed.
Lock 3: All set values – except Adjustment value, Span
sensitivity correction value, Transmission output Zero
adjustment value, Transmission output Span
adjustment value – can be temporarily changed. However, they revert to their previous value after the power is turned off because they are not saved
in the non-volatile IC memory. Outside Measurement Range
ORP value is outside the measurement range: If the value is less than -2000 mV or exceeds 2000 mV, the following will be indicated.
However, when ORP value is outside the measurement range, and if the unit proceeds to Adjustment mode or Span sensitivity correction mode, the ORP Display will be unlit, and the Setting Display will flash .
ORP Display Setting Display
Less than -2000 mV: -2000 is flashing. Exceeding 2000 mV: 2000 is flashing.
Power Failure
Countermeasure
The setting data is backed up in the non-volatile IC memory.
Self-diagnosis The CPU is monitored by a watchdog timer, and if an
abnormal status is found on the CPU, the AER-101-ORP is switched to warm-up status.
Warm-up Indication For approx. 4 seconds after the power is switched ON, the
characters below are indicated on the ORP Display. The Setting Display is unlit.
ORP Display Setting Display
Unlit
Page 69
69
ORP Color Selects ORP Display color.
Selection Item in
[ORP Color (p.49)]
ORP Display Color
Green
Red Orange
ORP color changes continuously.
ORP color changes continuously:
ORP Display color changes according to [ORP color reference value (p.50)] and [ORP color range (p.50)] settings.
• When ORP is lower than [ORP color reference value] – [ORP color range]: Orange
• When ORP is within [ORP color reference value] [ORP color range]: Green
• When ORP is higher than [ORP color reference value] +
[ORP color range]: Red
Bar Graph Indication
When (Transmission output) is selected in [Bar graph indication (p.50)], segments light in accordance with the output. Scale is -5 to 105%. Segments light from left to right in accordance with the output.
: ORP color reference value
: ORP color range
Orange
Green Red
Hys Hys
Hys
(e.g.) When output is 50%
-5%
50%
105%
Lights from left to right in accordance with the output.
Page 70
70
Other
Power Consumption Approx. 12 VA Ambient Temperature 0 to 50 Ambient Humidity 35 to 85 %RH (Non-condensing) Weight Approx. 280 g Accessories Included Unit label: 1 sheet
Mounting bracket: 1 set Instruction manual: 1 copy When Serial communication (C5 option) is ordered:
Wire harness C5J (0.2 m): 1 length
Wire harness C0J (3 m): 1 length When EVT3, EVT4 Outputs (Contact outputs 3, 4) (EVT3 option) is ordered:
Wire harness HBJ (3 m): 2 lengths
Accessories Sold Separately
Terminal cover
Page 71
71
10.2 Optional Specifications
Serial Communication (Option code: C5)
Serial Communication
The following operations can be carried out from an external computer. (1) Reading and setting of various set values (2) Reading of the ORP value and status (3) Function change, adjustment (4) Reading and setting of user save area
Cable Length
1.2 km (Max), Cable resistance value: Within 50 (Terminators are not necessary, but if used, use 120 or
more on one side.) Communication Line
EIA RS-485
Communication Method
Half-duplex communication
Communication Speed
9600, 19200, 38400 bps (Selectable by keypad)
Synchronization Method
Start-stop synchronization
Code Form ASCII, Binary Communication Protocol
Shinko protocol, Modbus ASCII, Modbus RTU (Selectable by keypad)
Data Bit/Parity
8 bits/No parity, 7 bits/No parity, 8 bits/Even, 7 bits/Even,
8 bits/Odd, 7 bits/Odd (Selectable by keypad) Stop Bit 1, 2 (Selectable by keypad) Error Correction Command request repeat system Error Detection
Parity check, Checksum (Shinko protocol),
LRC (Modbus protocol ASCII),
CRC-16 (Modbus protocol RTU) Data Format
Communication
Protocol
Shinko
Protocol
Modbus
ASCII
Modbus
RTU
Start bit 1 1 1
Data bit 7 7 8
Parity Even
Even (No parity, Odd) Selectable
No parity (Even, Odd) Selectable
Stop bit 1
1 (2) Selectable
1 (2) Selectable
EVT3, EVT4 Outputs (Contact outputs 3, 4) (Option code: EVT3)
EVT3, EVT4 Outputs (Contact outputs 3, 4)
Same as EVT output (pp. 65, 66)
Page 72
72
11. Troubleshooting
If any malfunction occurs, refer to the following items after checking that power is being supplied to the AER-101-ORP.
11.1 Indication Problem Possible Cause Solution
The ORP Display is unlit.
The time set in [Backlight time (p.50)] has passed.
If any key is pressed while
displays are unlit, it will re-light.
Set the backlight time to a
suitable time-frame. Indication of the ORP Display is unstable or irregular.
Calibration may not have finished.
Perform calibration.
Specification of ORP Combined Electrode Sensor may not be suitable.
Replace the sensor with a
suitable one.
Electrode sensor terminal screws have become loose.
Tighten the screws securely.
Electrical insulation of electrode sensor terminals has deteriorated.
Clean the terminals with
alcohol, and dry completely.
The electrode is not clean. Clean the electrode. Air bubbles are attached to the electrode.
Make sure there are no
bubbles in the measurement
solution.
The electrode has not been placed in the measurement solution.
Install the electrode in the
measurement solution,
maintaining a consistent
volume.
There may be equipment that interferes with or makes noise near the AER-101-ORP.
Keep AER-101-ORP clear of
any potentially disruptive
equipment. The Setting Display is unlit.
(No indication) is selected in [Setting Display indication (p.50)].
Select any other item except
(No indication)
[ ] is flashing on the Setting Display.
This indicates that the ORP value is outside the measure­ment range (less than -2000 mV or exceeding 2000 mV).
Check the measuring environment.
[ ] is indicating on the ORP Display.
Internal memory is defective. Contact our agency or us.
Page 73
73
11.2 Key Operation Problem Possible Cause Solution
Unable to set values. The values do not change by the or
key.
(Lock 1) is selected
in [Set value lock (p.47)].
(The LOCK indicator is lit when
Lock 1 is selected.)
Select (Unlock).
Only EVT1 to EVT4 values can be set. Other settings are impossible. The values do not change by or
key.
(Lock 2) is selected
In [Set value lock (p.47)].
(The LOCK indicator is lit when Lock 2 is selected.)
Select (Unlock).
Unable to enter Manual Cleansing
Mode.
(Cleansing output) is not selected in any of [EVT1 type to EVT4 type (pp. 23, 29, 35, 41)].
Select (Cleansing output) in any of [EVT1 type to EVT4 type (pp. 23, 29, 35,
41)]. Cleansing action is performing using the ‘Cleansing Time’ and ‘Restore Time after Cleansing’
settings.
Execute Manual cleansing after cleansing action is complete.
Unable to enter a calibration mode (Adjustment mode or Span sensitivity correction mode).
(Lock 1),
(Lock 2) or (Lock 3) has been selected in [Set value lock (p.47)].
(The LOCK indicator is lit when
Lock 1, Lock 2 or Lock 3 is selected.)
Select (Unlock).
(Cleansing output) has been selected in any of [EVT1 type to EVT4 type (pp. 23, 29, 35, 41)], and cleansing action is performing using the ‘Cleansing Time’ and ‘Restore Time after Cleansing’ settings.
Perform calibration after cleansing action is complete.
Page 74
74
12. Character Tables
The following shows our character tables. Use data column for your reference.
12.1 Setting Group List Character Setting Group Reference Section
ORP Input group Section 12.6 (p.75) EVT1 Action group Section 12.7 (p.75,76) EVT2 Action group Section 12.8 (p.77, 78) EVT3 Action group Section 12.9 (p.78, 79) EVT4 Action group Section 12.10 (p.80, 81) Special Function group Section 12.11 (pp.81, 83)
12.2 Adjustment Mode
Character Setting Item, Setting Range Factory Default Data
(*) Adjustment value 0 mV
-200 to 200 mV
(*) and ORP are lit alternately.
12.3 Span Sensitivity Correction Mode
Character Setting Item, Setting Range Factory Default Data
(*) Span sensitivity correction value 100%
50 to 150%
(*) and ORP value are lit alternately.
12.4 Transmission Output Adjustment Mode
Character Setting Item, Setting Range Factory Default Data
Transmission output Zero
adjustment value
0.00%
5.00% of Transmission output span
Transmission output Span
adjustment value
0.00%
5.00% of Transmission output span
12.5 Simple Setting Mode
Character Setting Item, Setting Range Factory Default Data
EVT1 value 0 mV
Setting range: Input low limit to Input high limit
EVT2 value 0 mV
Setting range: Input low limit to Input high limit
EVT3 value 0 mV
Setting range: Input low limit to Input high limit
EVT4 value 0 mV
Setting range: Input low limit to Input high limit
Page 75
75
12.6 ORP Input Group
Character Setting Item, Setting Range Factory Default Data
Moving average data amount 3
1 to 20
Input high limit 2000 mV
Input low limit to 2000 mV
Input low limit -2000 mV
-2000 mV to Input high limit
ORP input filter time constant 0.0 sec.
0.0 to 60.0 sec.
12.7 EVT1 Action Group
Character Setting Item, Setting Range Factory Default Data
EVT1 type No action
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT1 value 0 mV
Input low limit to Input high limit
EVT1 proportional band 0 mV
0 to 4000 mV
EVT1 reset 0 mV
200 mV
EVT1 hysteresis type Reference Value
: Medium Value : Reference Value
EVT1 ON side 10 mV
0 to 200 mV
EVT1 OFF side 10 mV
0 to 200 mV
EVT1 ON delay time 0 sec.
0 to 10000 sec.
EVT1 OFF delay time 0 sec.
0 to 10000 sec.
EVT1 proportional cycle 30 sec.
1 to 300 sec.
EVT1 output high limit 100%
EVT1 output low limit to 100%
EVT1 output low limit 0%
0% to EVT1 output high limit
Page 76
76
Character Setting Item, Setting Range Factory Default Data
Output ON time when EVT1 output ON
0 sec.
0 to 10000 sec.
Output OFF time when EVT1 output ON
0 sec.
0 to 10000 sec.
EVT1 ORP input error alarm EVT type
No action
: No action : EVT2 type : EVT3 type : EVT4 type
EVT1 ORP input error alarm span when EVT output ON
0 mV
0 to 2000 mV
EVT1 ORP input error alarm time when EVT output ON
0 sec.
0 to 10000 seconds or minutes
EVT1 ORP input error alarm span when EVT output OFF
0 mV
0 to 2000 mV
EVT1 ORP input error alarm time when EVT output OFF
0 sec.
0 to 10000 seconds or minutes
EVT1 cycle variable range 50.0%
1.0 to 100.0%
EVT1 cycle extended time 0 sec.
0 to 300 sec.
Page 77
77
12.8 EVT2 Action Group
Character Setting Item, Setting Range Factory Default Data
EVT2 type No action
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT2 value 0 mV
Input low limit to Input high limit
EVT2 proportional band 0 mV
0 to 4000 mV
EVT2 reset 0 mV
200 mV
EVT2 hysteresis type Reference Value
: Medium Value : Reference Value
EVT2 ON side 10 mV
0 to 200 mV
EVT2 OFF side 10 mV
0 to 200 mV
EVT2 ON delay time 0 sec.
0 to 10000 sec.
EVT2 OFF delay time 0 sec.
0 to 10000 sec.
EVT2 proportional cycle 30 sec.
1 to 300 sec.
EVT2 output high limit 100%
EVT2 output low limit to 100%
EVT2 output low limit 0%
0% to EVT2 output high limit
Output ON time when EVT2 output ON
0 sec.
0 to 10000 sec.
Output OFF time when EVT2 output ON
0 sec.
0 to 10000 sec.
EVT2 ORP input error alarm EVT type
No action
: EVT1 type : No action : EVT3 type : EVT4 type
Page 78
78
Character Setting Item, Setting Range Factory Default Data
EVT2 ORP input error alarm span when EVT output ON
0 mV
0 to 2000 mV
EVT2 ORP input error alarm time when EVT output ON
0 sec.
0 to 10000 seconds or minutes
EVT2 ORP input error alarm span when EVT output OFF
0 mV
0 to 2000 mV
EVT2 ORP input error alarm time when EVT output OFF
0 sec.
0 to 10000 seconds or minutes
EVT2 cycle variable range 50.0%
1.0 to 100.0%
EVT2 cycle extended time 0 sec.
0 to 300 sec.
12.9 EVT3 Action Group
Character Setting Item, Setting Range Factory Default Data
EVT3 type No action
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT3 value 0 mV
Input low limit to Input high limit
EVT3 proportional band 0 mV
0 to 4000 mV
EVT3 reset 0 mV
200 mV
EVT3 hysteresis type Reference Value
: Medium Value : Reference Value
EVT3 ON side 10 mV
0 to 200 mV
EVT3 OFF side 10 mV
0 to 200 mV
EVT3 ON delay time 0 sec.
0 to 10000 sec.
Page 79
79
Character Setting Item, Setting Range Factory Default Data
EVT3 OFF delay time 0 sec.
0 to 10000 sec.
EVT3 proportional cycle 30 sec.
1 to 300 sec.
EVT3 output high limit 100%
EVT3 output low limit to 100%
EVT3 output low limit 0%
0% to EVT3 output high limit
Output ON time when EVT3 output ON
0 sec.
0 to 10000 sec.
Output OFF time when EVT3 output ON
0 sec.
0 to 10000 sec.
EVT3 ORP input error alarm EVT type
No action
: EVT1 type : EVT2 type : No action : EVT4 type
EVT3 ORP input error alarm span when EVT output ON
0 mV
0 to 2000 mV
EVT3 ORP input error alarm time when EVT output ON
0 sec.
0 to 10000 seconds or minutes
EVT3 ORP input error alarm span when EVT output OFF
0 mV
0 to 2000 mV
EVT3 ORP input error alarm time when EVT output OFF
0 sec.
0 to 10000 seconds or minutes
EVT3 cycle variable range 50.0%
1.0 to 100.0%
EVT3 cycle extended time 0 sec.
0 to 300 sec.
Page 80
80
12.10 EVT4 Action Group
Character Setting Item, Setting Range Factory Default Data
EVT4 type No action
: No action : ORP input low limit action : ORP input high limit action : Cleansing output
: ORP input error alarm output
EVT4 value 0 mV
Input low limit to Input high limit
EVT4 proportional band 0 mV
0 to 4000 mV
EVT4 reset 0 mV
200 mV
EVT4 hysteresis type Reference Value
: Medium Value : Reference Value
EVT4 ON side 10 mV
0 to 200 mV
EVT4 OFF side 10 mV
0 to 200 mV
EVT4 ON delay time 0 sec.
0 to 10000 sec.
EVT4 OFF delay time 0 sec.
0 to 10000 sec.
EVT4 proportional cycle 30 sec.
1 to 300 sec.
EVT4 output high limit 100%
EVT4 output low limit to 100%
EVT4 output low limit 0%
0% to EVT4 output high limit
Output ON time when EVT4 output ON
0 sec.
0 to 10000 sec.
Output OFF time when EVT4 output ON
0 sec.
0 to 10000 sec.
EVT4 ORP input error alarm EVT type
No action
: EVT1 type : EVT2 type : EVT3 type : No action
Page 81
81
Character Setting Item, Setting Range Factory Default Data
EVT4 ORP input error alarm span when EVT output ON
0 mV
0 to 2000 mV
EVT4 ORP input error alarm time when EVT output ON
0 sec.
0 to 10000 seconds or minutes
EVT4 ORP input error alarm span when EVT output OFF
0 mV
0 to 2000 mV
EVT4 ORP input error alarm time when EVT output OFF
0 sec.
0 to 10000 seconds or minutes
EVT4 cycle variable range 50.0%
1.0 to 100.0%
EVT4 cycle extended time 0 sec.
0 to 300 sec.
12.11 Special Function Group
Character Setting Item, Setting Range Factory Default Data
Set value lock Unlock
: Unlock : Lock 1 : Lock 2 : Lock 3
Communication protocol Shinko protocol
: Shinko protocol
: Modbus ASCII mode : Modbus RTU mode
Instrument number 0
0 to 95
Communication speed 9600 bps
: 9600 bps : 19200 bps : 38400 bps
Data bit/Parity 7 bits/Even
: 8 bits/No parity : 7 bits/No parity : 8 bits/Even : 7 bits/Even : 8 bits/Odd : 7 bits/Odd
Page 82
82
Character Setting Item, Setting Range Factory Default Data
Stop bit 1 bit
: 1 bit : 2 bits
Transmission output type ORP transmission
: ORP transmission : EVT1 MV transmission : EVT2 MV transmission : EVT3 MV transmission : EVT4 MV transmission
Transmission output high limit
ORP transmission: 2000 mV
MV transmission: 100.0% ORP transmission: Transmission output low limit to 2000 mV MV transmission: Transmission output low limit to 100.0%
Transmission output low limit
ORP transmission: -2000 mV
MV transmission: 0.0%
ORP transmission: -2000 mV to Transmission output high limit MV transmission: 0.0% to Transmission output high limit
Transmission output status in Adjustment mode, Span sensitivity correction mode
Last value HOLD
: Last value HOLD : Set value HOLD : Measured value
Set value HOLD ORP transmission: 0 mV
MV transmission: 0.0%
ORP transmission: -2000 to 2000 mV
MV transmission: 0.0 to 100.0%
Backlight selection All are backlit
: All are backlit. : ORP Display : Setting Display : Action indicators : ORP Display + Setting Display : ORP Display + Action indicators : Setting Display + Action indicators
ORP color Red
: Green : Red : Orange : ORP color changes continuously.
ORP color reference value 0 mV
2000 mV
Page 83
83
Character Setting Item, Setting Range Factory Default Data
ORP color range 200 mV
2000 mV
Backlight time 0 min.
0 to 99 minutes
Bar graph indication No indication
: No indication : Transmission output
EVT output when input errors occur Disabled
: Disabled : Enabled
Setting Display indication
No indication : No indication : EVT1 value : EVT2 value : EVT3 value : EVT4 value
Number of cleansing cycles 0 (Continuous
cleansing)
0 to 10 (0: Continuous cleansing)
Cleansing interval 360 minutes
60 to 3000 minutes
Cleansing time 600 sec.
1 to 1800 sec.
Restore time after cleansing 600 sec.
1 to 1800 sec.
ORP input error alarm time unit Second(s)
: Second(s) : Minute(s)
Page 84
***** Inquiries *****
For any inquiries about this unit, please contact our agency or the
vendor where you purchased the unit after checking the following.
[Example]
• Model --------------------------- AER-101-ORP
• Serial number ----------------- No. 134F05000
In addition to the above, please let us know the details of the
malfunction, or discrepancy, and the operating conditions.
No.AER11ORPE2 2016.05
SHINKO TECHNOS CO., LTD.
OVERSEAS DIVISION
:Head Office URL: E-mail:
2-5-1, Senbahigashi, Minoo, Osaka, Japan http://www.shinko-technos.co.jp overseas@shinko-technos.co.jp
Tel : Fax:
+81-72-727-6100 +81-72-727-7006
Loading...