Before Installation, Operation or Maintenance, be sure to read carefully this Operation Manual and use the
equipment as directed. Use the equipment after carefully reading all the caution items, safety information, and
other discriminations. In this User’s Manual, the Safety Awareness Items are divided into three different
categories: “Danger”, “Warning” and “Caution”.
Follow these precautions.
Danger
Warni n g
Caution
Protection categories are explained and separated by the following symbols.
This symbol indicates if you ignore the contents mentioned and improperly
operate, it may lead to serious injury from a potential fire.
This symbol indicates a potential dangerous situation that could produce a
Serious injury if improperly handled.
This symbol indicates a potential situation that could produce a minor injury or
Damage to material if it is improperly handled. However, depending upon
the situation, there could be a possibility that it may cause more serious
results.
Proceed with Caution.
Not Allowed-Prohibited.
Mandatory-Must Follow
Warning
Heavy! Take sufficient care while
handling.
Ensure stand is placed on a flat, level,
stable surface that can support the unit.
Do not touch the running, inner parts.
Keep appendages and loose clothing
away from running, moving parts.
Do not operate in areas where explosive
gases or vapors may be present.
Caution
Fix object surely.
There is the possibility of damage.
Hold power plug while removing the
power cord. Do not pull or put tension on
the cord. Doing so could result in cord
damage and electric shock
Do not install in a high humidity
environment or near where water may be
present. Electric shock may occur if water
is encountered.
Never carry or move the FGS-VC with the
AC code.
There is the possibility of electric shock, a fire and
injury.
Confirm that the power supply is the
same voltage rating as the displayed
voltage rating on the unit
Do not plug AC connector with wet hand.
There is the possibility of electric shock.
Attention Safety
Warning
Disconnect from the AC outlet during any maintenance of the stand. Also, unplug when not
using for long periods of time.
May cause electric shock, a fire and injury.
Attention before using FGS-VC
Avoid the following.
Place where water, oil, or medicine may splash
Dusty locations
Places where condensation occurs
Places where explosive gases or ignition sources
are present.
Place where high vibration may occur
Places outside of operating temperature range.
32-104°F (0-40°C)
Cleaning with thinner and gasoline, etc.
When digital force gauge FGV series is installed
Set "ovEr" in the F06 parameter , which is the parameter of Output Type.
If "ovEr" is not set, FGS-VC does not stop by the overload of the force gauge.
Make sure the cable is not stuck in the stand.
Install the force gauge after turning off the power
switch.
1. Prior to use ................................................................................................................................................ 1
1.1. Procedure from the Installation of a force gauge to the measurement ........................................... 1
1.2. Confirmation of product packing ...................................................................................................... 1
2. Names of components ............................................................................................................................... 2
3. Names and functions of the operation panel ........................................................................................... 2
Emergency stops when pushing.
Push the switch when FGS-VC acts abnormally or a potentially dangerous
situation is occurring. Release the emergency stop to restart operation.
Increase or decrease the speed when pushing the up or down arrows of the
Speed key. If the up or down arrows are held, the speed adjusts continuously.
Push to override the current speed setting. The speed will instantly go to the
maximum capable of the stand. Once released, the speed returns to the
previous speed setting. This key only is functional in Manu or Jog modes.
5 MODE key Change operation modes or function settings.
6 SET key The indicated data is set when the key is pushed.
7 START key Start the program mode operation
8 PUSH key Move in the PUSH direction.
9 STOP key Stop operation.
10 PULL key Move in the PULL direction.
11 ZERO key
12 LCD display Displays distance, speed and mode, etc.
LCD display
Clear distance in Manu and Jog modes.
Clear repeat time in Sing, Cont and Prog modes.
Main display
Sub display
Status display
4. Setup
4.1. Installation of force gauge
The force gauge is set in the mounting plate.
In case of FGS-220VC
The force gauge is fixed to the stand.
Force gauge that can be connected with FGS-VC is FGV series.
Force gauges up to 100kgf are compatible with the FGS-220VC.
Force gauges up to 200kgf are compatible with the FGS-550VC.
Remove the mounting plate using the
hex-head spanner accessory.
FGS-VC can connect to PC for taking force and displacement.
The software is able to be downloaded from Shimpo's website free of charge.
Refer to the software manual.
Cable between FGS-VC and force gauge
USB cable
Personal computer
5. Operation
5.1. Basic operation flow
The procedure of basic operation is described as follows.
Force gauge power
Turn on the power of the force gauge.
FGS-VC power
Turn on the power of the stand.
Keep hands, hair and jewelry away from stand when drive assembly is in motion.
May cause damage or injury.
Separate hands from the force gauge and FGS-VC
movable parts.
Turn on FGS-VC.
The manual operation modes are MANU, JOG, SING and CONT.
The mode is selected by the measurement purpose and the usage.
MANU mode
5.3.1. MANU mode
This mode of operation is ideal for manually recording force measurements. The test stand will only operate
between the limits that are set by the test stand user. These are manually adjusted distance limits.
Contents
The test stand will move in the downward or upward direction when the respective PUSH or PULL button is
selected. The stand will continue to move in the selected direction until one of the following occurs: STOP
button is pushed, one of the limit stop switches is tripped or the emergency reset button is pushed.
JOG mode
SING mode
CONT mode
When pressing the PUSH (PULL) key, FGS-VC goes to PUSH (PULL)
limit switch position.
FGS-VC stops if reaching the limit position.
This mode of operation is identical to MANU, except that the movement
in any direction will only occur while either the PUSH or PULL key is
depressed.
Pressing PUSH or PULL keys, FGS-VC moves for one complete cycle
between mechanical distance limit stops.
Pressing PUSH or PULL keys, FGS-VC moves repeatedly up and down
between the mechanical distance limit stops.
Speed
Speed can be changed by UP or DOWN button of the SPEED.
The speed setting of MANU and JOG mode is common.
Also the speed is able to be changed while moving.
Display
Parameter
The speed of MANU and JOG modes is adjustable. Press the SET button to enable adjustment via the UP or
DOWN buttons of the SPEED key. Then press the SET button to save the new value. If you want to cancel
the setting press the ZERO button.
5.3.2. JOG mode
Contents
This mode of operation is identical to MANU mode, except that the movement in any direction will only occur
while either the PUSH or PULL button is depressed.
Speed
Speed can be changed by UP or DOWN button of the SPEED key.
The speed setting of MANU and JOG mode is common.
Also the speed is able to be changed while moving.
Display
Present position
Setting speed
Setting speed
The UP or DOWN button of the
SPEED key can be used to
change the speed. The range is
Parameter
The parameter, which is the speed of MANU and JOG mode, is available.
Pressing SET button, and the speed can be changed by UP or DOWN key of the SPEED.
Then press SET button for saving.
If you want to cancel the setting, press ZERO button.
Setting speed
The UP or DOWN button of the
SPEED key can be used to
change the speed. The range is
This mode of operation is ideal for completing one cycle between mechanical, manually adjusted distance limit
stops. The test stand will only operate between the limits that are set by the test stand user.
Contents
The test stand will move downward or upward when the respective PUSH or PULL button is selected.
The stand will continue to move until one of the following events occurs: the STOP button is pushed, one of the
limit switches is tripped or the emergency reset button is pushed.
Speed
Speed can be changed by UP or DOWN button of the SPEED key.
The speed setting of SING and CONT mode is common.
Also the speed is able to be changed while moving.
Display
Present position
Repeat count
Incremental count at pressing PUSH or PULL button.
The count is cleared with the ZERO button. When the mode is changed, the count will also be reset to zero.
Parameter
The parameter, which is the dwell timer and the speed of pull and push, is available.
Pressing SET button, the speed can be changed by UP or DOWN button of the SPEED key. The dwell timer
can be changed by STOP.
Finally press SET button for saving.
If you want to cancel the setting, press ZERO button.
Saving, return to SING mode
SET
SET
SET
Setting speed
The UP or DOWN button of the
SPEED can be used to change.
The range is o.40-15.75 in/min(10
Setting dwell timer
The STOP button of the dwell time
can be used to change.
Pressing STOP, the dwell timer is
incremental.
When the dwell timer is 5s, it
returns to 0s pressing STOP.
This mode of operation is ideal if the user wants the test stand to repeatedly cycle up and down continuously or
for a user-programmed number of times. The stand will start in either direction depending on whether PUSH or
PULL button is selected.
Contents
The test stand will start to move downward or upward when the respective PUSH or PULL button is selected.
The stand will continue to move until one of the following events occurs: the STOP button is pushed, the
emergency reset button is pushed.
Speed
Speed can be changed by UP or DOWN button of the SPEED.
The speed setting of SING and CONT mode is common.
Also the speed is able to be changed while moving.
Display
Present position
Repeat count
Incremental count at pressing PUSH or PULL button.
The count is cleared with the ZERO button. When the mode is changed, the count will also be reset to zero.
Parameter
The parameter, which is the repeat times and the dwell timer and the speed of pull and push, is available.
Pressing SET button, the speed and the repeat times can be changed by UP or DOWN button of the SPEE D
key, the dwell timer can be changed by STOP.
Finally press SET button for saving.
If you want to cancel the setting, press ZERO button.
Saving, return to SING mode
SET
SET
SET
SET
Setting speed
The UP or DOWN button of the
SPEED can be used to change.
The range is 0.40-15.75 in/min 10
Setting dwell timer
The STOP button of the dwell time
can be used to change.
Pressing STOP, the dwell timer is
incremental.
When the dwell timer is 5s, it returns
to 0s pressing STOP.
Setting dwell timer
The UP or DOWN button of the
SPEED can be used to change.
The maximum value is 9999.
When the display is o and press
DOWN it will be 9999, and whe n the
display is 9999 and press UP it will
be 0.
This mode of operation is ideal if user wants to operate the test stand with complex programmed movements.
Even if the user does not know the length of the object, the accurate measurement is possible because the
function to detect the object to be measured is provided.
Contents
The programmed data consists of a condition setting, starting point, a method of detection of the object, five
measurement points and a method of return.
The test stand will start to move when START button is pressed.
The stand will continue to move until one of the following events occurs: the STOP button is pushed, one of the
mechanical limit switches is tripped, the emergency reset button is pushed or the overload of the force gauge
is detected. If the force limit of the condition setting is detected, the present operation is interrupted and it will
jump to next point operation.
Start point Program start point and return point
Detect point Object detected point according to the condition
Standard point Standard point for measurement
Point 1 to 5 Programmed point
Parameter
The parameters are available while stopping.
Pressing SET button, the parameter setting starts.
The contents of the parameters setting is shown in the below flowchart.
Finally press SET button for saving.
If you want to cancel the setting, press ZERO button.
Condition setting
Moving direction
The parameter is the
direction of movement
from “Start point” to
“Detect point”.
The STOP button can
be changed to Push or
Pull alternately.
Detect object
The force for detecting
object can be set by
PUSH or PULL button.
Zero force
After detecting the
object, the stand moves
to opposite direction until
sensing the force less
than the setting force
data.
The PUSH or PULL
button can be used to
change.
SET
SET
SET
SET
Next page
Force limit
The PUSH or PULL button can
be used to change.
The sign of force limit is changed
by MAX button alternately.
Repeat times
The UP or DOWN button of the
SPEED can be used to change.
The maximum value is 9999.
When the display is o and press
DOWN it will be 9999, and whe n
the display is 9999 and press UP
it will be 0.
Speed
The parameter is the speed
setting for moving from "Start
point" to "Detect point".
The UP or DOWN button of the
SPEED can be used to change.
Speed
The parameter is the speed
setting for moving from "Detect
point" to "Standard point".
The UP or DOWN button of the
SPEED can be changed.
Dwell timer
The parameter is the dwell timer
at "Standard point".
Pressing STOP, the dwell timer is
incremental.
When the dwell timer is 5s, it
returns to 0s pressing STOP.
The parameter is the
direction of movement
from “Standard point” to
“Point 1”.
The MAX button can be
changed to Push or Pull
Return to “Start point”
Previous page
SET
Setting P2 to P4
SET
SET
SET
Saving, return to PROG mode
Movement Distance at P1
The parameter is the distance
from "Standard point" to "P1".
Also the data should be set as
relative coordination.
The PUSH or PULL button can
be used to change.
Speed at P1
This parameter is the speed
setting for moving from "Start
point" to "Detect point".
The UP or DOWN button of the
SPEED can be used to change.
Dwell timer at P5
This parameter is the dwell timer.
Pressing STOP, the dwell timer is
incremental.
When the dwell timer is 5s, it
returns to 0s pressing STOP.
Speed at RET
This parameter is the speed
setting for moving from "Point 5"
to "Start point".
The UP or DOWN button of the
SPEED can be used to change.
Dwell timer at RET
This parameter is the dwell timer.
Pressing STOP, the dwell timer is
incremental.
When the dwell timer is 5s, it
returns to 0s pressing STOP.
The stand controls to give the force less than the "Force limit".
If the force limit is detected, the stand will stop immediately. Then the stand executes the next step.
Upper limit
Start point
Standard point
Detect point
Point 1
Point 5
Point 3
Point 2
Point 4
Lower limit
Detect object
The parameter is the threshold force level for detecting the object.
When the setting force of "Detect object" parameter is detected, the stand will stop immediately. Next
the stand moves to opposite direction until detecting "Zero force".
The recommended "Detect object" that the user should program into the stand is more than 0.2% of
the full scale of the mounted force gauge, i.e., if the programmer is using a 200lb force gauge, the
programmed "Detect object" should be 0.4lb (200lb x 0.2%).
Point 5
Point 4
Detect
the force limit
Example
5.5. Parameter setting mode
The relations function is set to the entire stand.
Contents
The parameters are shown in the below table.
Items Description Range
The sign at the position of the direction of
Sign of Push direction
Acceleration
Volume of buzzer The volume of the buzzer is adjusted. OFF, Low, Middle, High
Push is set.
If it sets "-" and the stand moves to Push, the
display of position is decrement.
The acceleration of the motor of the stand is
set.
The acceleration is the slope or rate of
increase in speed from the non-motion
starting point to the set programmed speed.
If the parameter is set at a maximum of 100,
the motor will accelerate quickly
The lower limit was reached.
“PUSH LMT” is blinking.
The upper limit was reached.
“PULL LMT” is blinking.
The over load of the mounted force gauge
occurred in push direction.
The over load of the mounted force gauge
occurred in pull direction.
The emergency stop was pushed.
The motor or the driver of the motor was
abnormal.
The communication error between the
controller and the motor driver in the stand
occurred.
The read error of the memory in the stand
occurred.
The write error of the memory in the stand
occurred.
The stand could not recognize the
mounted force gauge.
If the status is abnormal, move the
lower limit switch position.
If the status is abnormal, move the
upper limit switch position.
Confirm whether the force gauge
used is correct.
Check whether the measurement
is abnormal.
Confirm whether the force gauge
used is correct.
Check whether the measurement
is abnormal.
Confirm that a problem is not
found, then release the emergency
stop.
Turn off and wait one minute, then
turn on.
Contact TECHNICAL SUPPORT if
the alarm is not canceled.
Turn off and wait one minute, then
turn on.
Contact to our technical support if
the error alarm is not canceled.
Turn off and wait one minute, then
turn on.
If the error is not canceled,
initialize the parameter. Refer to
the below procedure of the
initialization.
Turn off and wait one minute, then
turn on.
If the error is not canceled,
initialize the parameter. Refer to
the below procedure of the
initialization.
Check:
The parameters setting of the force
gauge, the sequence of power on
and the cable.
If no problem, ask to our technical
support.