If any damage is apparent, do not unpack the panel
meter. Notify the shipping carrier immediately for
damage claim instructions. Refer to the label to confirm
model number ordered and record serial number for
future reference.
Items included with the DT-5TXR or DT-5TS are:
Important Safety PrecautionsStandard Accessories
Confirm that the unit is turned OFF when
connecting or disconnecting the unit to a
sensor.
Never use wet hands when connecting or
disconnecting a sensor or checking the unit.
• Mounting adapter, 2 pieces (attached)
• Mounting screw, 2 pieces (attached)
• Decal sheet
• Front button protective cover (attached)
Features and Benets
The DT-5TXR/DT-5TS panel meter offers many features and
benefits including:
• Accepts a wide range of ppr (1-9,999) to
accommodate any sensor commonly found in the
field
• Wide range of AC voltage input (85 – 264 VAC 50/
60 Hz) eliminates duplicity of models or changing
of transformers
• Scalable; able to measure in various units
• Standard 1/8 DIN mounting enables easy mounting
while saving time
• Accepts modules (DT-5TXR only) which provide
different types of outputs, i.e. analog voltage,
analog current (4-20 mA), relay closures (set points),
BCD output and others for
recording purposes or data analysis
Always use the correct voltage (AC type: 85264VAC, DC type: 9-35VDC).
Do not position the signal wires and power
cord near each other.
Remember to check terminal connections, as
they may become loose due to vibration.
Do not use or store unit in extreme
temperatures; normal operating temperature
is 32-113ºF (0-45ºC).
Do not use or store unit in oily, dusty, or high
humidity areas.
Do not use chemicals to clean the case; use a
damp, soapy cloth.
Do not operate the unit in a location where
electrical noise, static electricity or excessive
1. Press and hold the MODE button for approximately 3 seconds; “0” will begin flashing until “-01-” appears on the
display.
2. Press any of the lower buttons (NUMERIC, NEGATIVE SIGN/DECIMAL POINT, BACKSPACE) to scroll through “-02-”,
“-03-” and “-04-” until “-99-” appears on the display.
3. Press the SET button to begin the self-test. The entire display will cycle through all digits, arrows, dashes, colons,
decimal points, etc.
4. When satisfied that all LED’s are operational, press the SET button to exit.
1. Confirm that the panel thickness is 0.047” - 0.197” (1.2 mm - 5 mm).
2. Remove the two mounting adapters (provided) from the DT-5TXR/DT-5TS (diagram 2).
3. Install a water-resistant gasket on the panel (see diagram 3 & 4).
4. Hold the panel meter in a horizontal position, insert it into the cut-out and press against gasket (diagram 5).
5. Replace the two mounting adapters (diagram 6).
6. The DT-5TXR/DT-5TS is secured on the panel by using the two mounting screws (provided) (see diagram 7).
3 margaid2 margaid
Factory Settings
The following charts reflect the default parameter and function settings as programmed by the factory:
Operating Mode -01- (RPM or Rate Measurement)
PARAMETERDESCRIPTION
1Pulses per rev olution
2Sensing RPM_ 010001,000 RPM
3Display units10001,000
4Update time_ 1.0 _1 second
5Display “Hold” after input signal ceases
6Input filter
DEFAULT FACTORY SETTING
DISPLAY
00011 ppr
006.0
10
MEANING
6 seconds
10 kHz
diagram 4
7 margaid6 margaid5 margaid
FUNCTION
1The 2nd HI and 2nd LO limit setting0 0 0 0 0 0 0
2Hysteresis of 1st HI and 1st LO set_ _ _ _ 0 00
3Secures the 1st HI and 1s t LO set
4Output activation of any set limit
5Selection of secondary displa
6Main display shows 0 below this no. setting 0 0 0 0 0 0 0
7Averages the sensor output frequency _ 0 _None
8Deceleration to ZERO display value
Output modules (DT-5TXR only)
10Logic of BCD output_ 0 _Negative-true logic
11Analog output value0 0 1 0 0 0.1000
12Minimum v alue of analog output0 0 0 0 1 0.10
13Update time for analog output
Depending upon the type of application, the DT-5TXR/DT-5TS offers different modes of operation; these modes and
their functions are:
MODENOITCNUF
-01- RPM or rate measurements
-02- Ela
-03- Process time measurement
-04- Flow rate measurement
-99- Self-test (see “Self-Test” section under “Set-Up”
Select the mode of operation as follows:
1. Press and hold the MODE button for approximately 3 seconds; “0” will begin flashing until one of the five modes
2. Each time one of the lower buttons (NUMERIC, NEGATIVE SIGN/DECIMAL POINT, BACKSPACE) is pressed, the
3. Press the SET button to store the mode of operation and exit.
sed time measuring / monitorin
used only in Japanese market
(see table above) appears on the display.
display will scroll through these modes as indicated below:
Setting Mode Parameters
Modes -01-, -02- and -03- have several parameters that must be selected and programmed into the DT-5TXR/DT-5TS.
In addition, there are also function settings to further customize the DT-5TXR/DT-5TS.
NOTE: Because the flow rate measurement mode (-04-) is only used in the Japanese market, it is not covered in this
manual. Contact Shimpo for additional information if necessary.
NOTE: If a mistake is made entering any number, simply press the BACKSPACE button to erase the last digit on the
primary display.
Mode -01- (RPM or Rate Measurement)
The following parameters must be set in mode -01- so that the DT-5TXR/DT-5TS will display the correct readings:
Except for parameter 3, the instructions are the same for setting the parameters as in mode -01-.
Instructions for setting the parameters in mode -02- are as follows:
1. Press the PAR button.
2. Use the NUMERIC buttons to enter the number of pulses per revolution.
3. Press the PAR button.
4. Press the NUMERIC buttons to enter the sensing RPM.
5. Press the PAR button; the display will reflect:
The secondary display A reflects that the meter is in mode -02-, but parameter 3 is showing on the primary
display (10 sec [factory setting]).
6. Use the NUMERIC buttons to enter the number of seconds to the hundredth place (000.00). If time in hours,
minutes, and seconds is necessary, press the NEGATIVE/DECIMAL POINT button, then use the NUMERIC buttons
to enter the amount of time as it relates to the already programmed RPM setting.
7. Press the PAR button.
8. Use any of the lower buttons to scroll through (see table above) and select an update time.
9. Press the PAR button.
10. Use the NUMERIC buttons to enter the time in seconds that the display should hold in the final reading.
11. Press the PAR button.
12. Refer to the “Sensor Connection” table under the “Set-Up” section for the proper filter for the sensor being
used. Press any of the lower buttons to scroll through the input filters (see table above) until the desired reading
is showing.
13. Press the SET button to store the parameters and exit.
NOTE: Please see page 13 for an example of MODE -02-.
3. Press the TEA button (the TEA indicator will be
illuminated).
4. Use the NUMERIC buttons to enter the known RPM.
5. Press the SET button to save the setting.
Setting High and Low Limits
g
)
_
_
_
_
)
_
_
_
_
g
_
g
_ _ _ _
If the variation of motor speed or time measurement is important and a “range” must be monitored, then “high” and
“low” visual limits can be programmed through secondary displays A and B, respectively.
Instructions to set the high and low limits:
1. Use the lower buttons (1-6) to set the high limit on secondary display A. There are six digits that can be
programmed and each one of these buttons corresponds with the digit position. Each time one of the buttons
is pressed, the display will increment that position one number (for example, if 56000 RPM needs to be programmed
as the high limit, then NUMERIC button 2 would have to be pressed five times and NUMERIC button 3 would be
pressed six times).
2. Use the lower buttons (7 - BACKSPACE) to set the low limit on secondary display B. These six buttons also
correspond with the digit positions and will increment that position one number (for example, if 3425 RPM needs
to be programmed as the low limit, then NUMERIC button 9 should be pressed three times, NUMERIC button 0
pressed four times, NEGATIVE/DECIMAL button pressed twice and the BACKSPACE button pressed five times).
3. Press the SET button to store the limit settings.
The comparator indicators will illuminate when the primary display exceeds the high set point ( ∆ ) or the low set
point ( ∇ ).
NOTE: The DT-5TXR provides relay or transistor closures (see section “Modules Available for DT-5TXR” on page 15).
Setting Functions
The DT-5TXR/DT-5TS has a variety of functions that can be programmed to further customize the meter. Listed below
are the available functions:
MODE -01-
0
0
FACTORY SETTING
Not secured
1st high and 1st low limit set
Off
RANGEDESCRIPTIONFUNCTION
1The 2nd high and the 2nd low limit setting000000 - 9999990 0 0 0 0 0 0
99 - 0gnittes wol ts1 dna hgih ts1 fo siseretsyH2_ _ _ _ 0 00
3Secures the 1st high and 1st low settin
5Selection of secondary display0 (1st high and 1st low limit set) 1(blank)/2 (real RPM)
6Main display shows 0 below this no. setti ng000000 - 9999990 0 0 0 0 0 0
7Averages the sensor output frequency 0 (none )/1 (three times)/2 (ten times) _ 0 _None
0(not s ecured )/1 (secured
dnoces 99 - 0timil tes yna fo noitavitca tuptuO4_ _ _ _ 0 00 second
) no( 1/)ffo( 0eulav yalpsid OREZ ot noitareleceD8
DISPLAYMEANING
_ 0 _
MODE -02-
0
_ 0 _
0
FACTORY SETTING
Not secured
1st high and 1st low limit set
FUNCTIONDESCRIPTIONEGNAR
1The 2nd high and the 2nd low limit setting0:00:00 - 9:59:59 or 0:00 – 999.99_ 0 0 0 0 0 1/100 second
99 - 0tes wol ts1 dna hgih ts1 fo siseretsyH2_ _ _ _ 0 00
3Secures the 1st high and 1st low set 0 (Not secured )/1 (Secured
dnoces 99 - 0timil tes yna fo noitavitca tuptuO4_ _ _ _ 0 00 second
5Selection of secondary display0 (1st high and 1st low limit set)/1 (blank)/2 (real RPM)
6Main display shows 0 below this no. setting000000 - 9999990 0 0 0 0 0 0
7Averages the sensor output frequency 0 (none)/1 (three time s)/2 (ten times) _ 0 _None
) no( 1/)ffo( 0eulav yalpsid OREZ ot noitareleceD8
DISPLAYMEANING
MODE -03-
FUNCTIONDESCRIPTIONEGNAR
1The 2nd high and the 2 nd low limit settin
2Hysteresis of 1st high and 1st low settin
3Secure the 1st high and 1st low setting0 (not secured )/1 (secured) _ 0 _Not secured
5Selection of secondary display0 (1st high and 1st low limit set)/1 (blank)/2 (r eal RPM)_ 0 _1st high and 1st low limit setting
0:00:00 - 9:59:59 or 0:0 0 - 999.99
0 - 99
dnoces 99 - 0timil tes yna fo noitavitca tuptuO4
DISPLAYMEANING
0 0 0 0 0 1/100 second
_ _ _ _ 0 0
FACTORY SETTING
0 00
0 second
NOTE: Although the functions are rarely used in mode -03-, they are still available.
Yes (with outputs, s ee modules)
± 0.008 % ± 1 dig it
0.2, 0.5, 1, 2, 5, 10, 15, 30, 60 se conds (selectabl e)
Controlled b y a 16.000 MHz c rystal
6 digit 0.59" (15 mm) high LED (main), 2-6 di git 0.02" ( 6.5 mm) high
1 - 9,999 ( programmable)
NPN open collector in put : max. freq uency 100 kHz
Contact input : max. freq uency 20 Hz
S
uare wave input : max. frequency 30 kHz
Sine wave input (m agnetic pickup) : max. frequency 1 0 kHz
12 VDC (150 mA)
85 - 264 VAC (50 /60 Hz) (9 - 35 VDC at 1 W also avai lable)
32°- 113° F ( 0- 45° C )
0.66 lb (300 g)
5.27" L x 3 .78" W x 1.89" H ( 134 mm
1 year
Sensors, mod ules
x 96 mm x 48 mm)
Elapsed TimeProcess Time
0:00:00 - 9:59:590:00:00 - 0:59:5 9
oror
0.00 - 999.99 sec 0.00 - 999.99 sec
± 0.1 % ± 1 digit
(secondary)
MODEL
ModereteM etaR / retemohcaT
Display Range
ScalableYes ( no outpu ts)
Accurac
Display Update Time
Time Base
Display
ut No. o f P/R
In
Input Si gnal Characteris tics NPN o pen col lector in put : max. frequency 100 kHz
0.2, 0.5, 1, 2, 5, 10, 15, 30, 60 se conds (selectabl e)
Controlled b y a 16.000 MHz c rystal
6 digit 0.59" (15 mm) high LED (main), 2-6 di git 0.02" ( 6.5 mm) high (second ary)
1 - 9,999 ( programmable)
Contact input : max. freq uency 20 Hz
uare wave input : max. frequency 30 kHz
S
Sine wave input (m agnetic pickup) : max. frequency 1 0 kHz
12 VDC (150 mA)
85 - 264 VAC (50 /60 Hz) (9 - 35 VDC at 1 W also avai lable)
32°- 113° F ( 0- 45° C )
0.5 lb (234 g)
3.62" L x 3 .78" W x 1.89" H ( 92 mm x 96 mm x 48 mm)
1 year
Sensors
DT-5 TS
Elapsed TimeProcess Time
0:00:00 - 9:59:590:00:00 - 0:59:5 9
oror
0.00 - 999.99 sec 0.00 - 999.99 sec
± 0.1 % ± 1 digit
19
Troubleshooting
The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments
directly for further assistance.
If display is blank:
• Check all connections: power source (AC type: 85-264VAC 50/60Hz; DC type: 9-35VDC), power cord and wires
(NOTE: For DC applications confirm that correct voltage polarity is used)
Error codes are displayed:
• Turn power OFF and then ON
• If error code remains, see table below
DISPLAY
CONDITIONACTION
Overflow of display (when
display value is over the
limit of display)
Input signal width is under
10 msec in MODE 3
Beyond the limit of TEA
function
EE PROM reading error
Change parameter
settings or input
signal
Change the input
signal width
Decrease the input
RPM
Press SET key
Warranty
LIMITED EXPRESS WARRANTY
Shimpo Instruments warrants, to the original purchaser of new products only, that this product shall be free from defects in workmanship and
materials under normal use and proper maintenance for one year from the date of original purchase. This warranty shall not be effective if
the product has been subject to overload, misuse, negligence, or accident, or if the product has been repaired or altered outside of Shimpo
Instruments’s authorized control in any respect which in Shimpo Instruments’s judgment, adversely affects its condition or operation.
DISCLAIMER OF ALL OTHER WARRANTIES
The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY, and Shimpo Instruments hereby disclaims all other warranties,
expressed, statutory or implied, applicable to the product, including, but not limited to all implied warranties of merchantability and fitness.
LIMITATION OF REMEDY
Under this warranty, Shimpo Instruments’ SOLE OBLIGATION SHALL BE TO REPAIR OR REPLACE the defective product or part, at Shimpo
Instruments’ option. Shimpo Instruments reserves the right to satisfy warranty obligation in full by reimbursing
Shimpo Instruments, whereupon, title shall pass to Shimpo Instruments upon acceptance of return goods. To obtain warranty service, Purchaser
must obtain Shimpo Instruments’s authorization before returning the product, properly repackaged, freight pre-paid to Shimpo Instruments.
INDEMNIFICATION & LIMITATION OF DAMAGES
Buyer agrees to indemnify and hold Shimpo Instruments harmless from and against all claims and damages imposed upon or incurred
arising, directly or indirectly, from Buyer’s failure to perform or satisfy any of the terms described herein. In no event shall Shimpo Instruments
be liable for injuries of any nature involving the product, including incidental or consequential damages to person or property, any economic
loss or loss of use.
MERGER CLAUSE
Any statements made by the Seller’s representative do not constitute warranties except to the extent that they also appear in writing. This
writing constitutes the entire and