SHIMPO Instruments DT-501XD-FVT, DT-501XD, DT-501XD-CPT, DT-501XD-TRT, DT-501XA-FVT User manual

...
Page 1
98211B
Digital Tachometer
Digital Tachometer/ Elapsed timecounter/ Time width meter/ Flowmeter
Standard input series Dierential input series
DT-501X/ DT-501F
Safety Requirements
Be sure to observe
Before operation, maintenance and inspection, please carefully read this instruction manual and follow it for proper use. Please carefully read all information related to this unit and safety, and precautions before use.
This instruction manual categorizes safety precautions as “DANGER”, “WARNING”, and “CAUTION”. Each of them is an important description related to safety. Be sure to observe.
Instruction Manual
Thank you for purchasing SHIMPO's Digital Tachometer DT-501X/ DT-501F. For instructions on how to use this product properly and optimally for a long period of time, please be sure to read this manual thoroughly before use.
When you purchase the product with optional equipment: Please refer to the operation manual of the optional equipment.
Before operation, maintenance and inspection, please carefully read this instruction manual and follow it
DANGER
!
WARNING
!
CAUTION
!
●Limited Warranty
● We are not responsible for damages resulting from negligence through failure to follow the instructions set out in this manual.
We are not responsible for damages resulting from earthquake and/or re unrelated
to us, actions by third parties, or any other accidents, intentional or through customer negligence, as well as from accidents caused by misuse or improper use under abnormal conditions.
● For information regarding assurance provisions, please read the attached warranty certicate.
Improper use by neglecting the following precautions may result in the potential for re, serious injuries, and/or death.
Improper use may result in serious injuries.
Improper use may result in minor injuries or property damage.
for proper use.
After reading, be sure to store this manual in a safe, convenient place where operators can always refer to it easily.
!
CAUTION
Electric Shock. Be sure to turn the power OFF when wiring as well as inspecting the unit.
Failure to do so could result in electric shock.
DO NOT block the ventilation holes on the side of the main unit. DO NOT put any foreign objects or materials inside the unit through these holes.
Failure to follow this could result in abnormal heat generation and/or malfunctions.
DO NOT touch the unit with wet (or sweaty) hands when inspecting or for wiring.
Failure to do this could result in electric shock.
!
Precautions before use
Power
●Be sure to use the unit under the specied voltage (AC power specications: 85 - 264VAC / DC power specications: 10.8 - 25.2VDC).
●Inverter power source cannot be used.
Input signal wire
●Connection wiring from sensors shall not be kept in the same or parallel conduit or cable as the power source, power or high voltage cables. If you fail to separate the wiring, noise may be superimposed on the signal wire, resulting in malfunctions.
Use shielded wire for input power connections with the shortest possible metal conduit.
Terminal
Check that the screws have not come loose due to vibrations after a certain period of time.
Operating environment
●Do not install the unit in the following places or conditions.
●●Places exposed to direct sunlight, or places where the ambient temperature exceeds a range of 0 - 45ºC.
●●Places where the relative humidity percentage exceeds a range of 35 - 85%, or places subject to condensation due to rapid change in humidity.
●●Places subject to corrosive and/or combustible gases.
●●Places subject to a large amount of dust, salinity, and/or ferric substance.
Find Quality Products Online at: sales@GlobalTestSupply.com
●●Places susceptible to noise (including static electricity).
www.GlobalTestSupply.com
INDEX
1. Installation to the Start of Operation
2. Unit Model
3. Specications
4. Component Part Names and Functions
5. External Dimensions
6. Installation to the Panel
7. Wiring to Power Source and Sensors (DT-501XA/DT-501XD)
9. Basic Setting Procedure
10. Keys to be Used for Various Settings and Their Applications
11. Teaching Function Settings
12. About Mode
13.
When You Select Mode 1 (Digital Tachometer Mode) for Measurement
14.
When You Select Mode 2 (Elapsed Timecounter Mode) for Measurement
15.
When You Select Mode 3 (Time Width Meter Mode) for Measurement
16.
When You Select Mode 4 (Flowmeter Mode) for Measurement
17.
Setting Method of Functions (Excluding in the Test Mode, Common in Each Mode)
18. Comparator Function
19. Memory Function
20. Test Mode (Function to Check if the Unit is Operating Normally)
21. Error Display
22. Parameter List
23. Function List
24. Option -FVT/-FVC
25. Option -BCD
26. Option -TRT
27. Option -CPT
28. DT-501X / DT-501F series model list
2
2
3
3
4
4
5
6
6
6
7
8
8
10
12
14
16
17
18
19
19
20
21
23
25
27
27
28
1
Page 2
●1.
Installation to the Start of Operation
This unit is designed for use according to your measurement purposes. Before use, follow the procedures below from installation to the start of operation.
Check before use
Check before use p.2 - 4
• Unit model
• Specications
• Component part names and functions
• External dimensions
Memory function settings
(1) Press the MEM key once to display the MAX value. (2) Press the MEM key again to display the MIN value. (3) Press the MEM key once again
to display the normal measurement value. *You can use the unit without setting various functions.
p.17 - 19
Installation to the panel
View from the front
(1)
Installation of waterproof gasket
(2) Fixation to the panel
(action conguration)
Function settings
(1) Press and hold the SHIFT + FUNC keys for 5 seconds or more and start setting. (2) Press the
item, and press the ENT key to start changing the setting value. (3) Press the
another digit, and after changing the value using the
s
nish the setting value change. (4) After setting all functions, use the SET key to nish setting.
key to select the
s
key to move to
t
key, use the ENT key to
p.4
Panel
p.8 p.
16 -17
Connection with the power and sensor
View from the back
(1) Connection with the power (2) Connection with various sensors
p.5 - 6
Connecting wires
Parameter settings p.8 - 15
(1) Press and hold the SHIFT + PAR keys for 5 seconds or more and start setting. (2) Press the
item, and press the ENT key to start changing the setting value. (3) Press the key to move to another digit, and after changing the value using the
● key, use the ENT key to nish the setting value change.
(3) After setting all parameters,
use the SET key to nish setting.
key to select the
s
t
(function)
Mode
(1) Press the SHIFT + MOD keys and start setting. (2) Use the
the SET key to nish setting.
settings p.8 - 15
key to set, and use
s
Start of operation
●2.
Unit Model
Teaching function settings
●Parameter settings are not necessary.
P7
Simple setting
Please check the model number of the equipment purchased.
501X
DT
501F
Power
Symbol Power
A AC power (85 - 264VAC)
D DC power (10.8 - 25.2VDC)
Input type
Symbol Input type
501X Standard input : Supports the sensor input, including rotary encoders, and magnetic sensors
501F(AC Power only) Dierential input Supports the line driver output, including AC servo motors
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Option #2
Option #1
Symbol Output Power Source
FVC
Analog Signal Output
BCS Output
BCD
Symbol Output Function
Analog Signal Output
FVT
CPT Relay Output
TRT Transistor Output Open Collector Output. HH, H, LL, L GO and ZERO output.
Outputs voltage and current that correspond to displayed value
Open Collector, BCD Output Simple interface with the sequencer
Outputs voltage and current that correspond to displayed value
Relay Changeover contact output. H output, L output, and GO output, against the set value
* HH : High set point 2 output
●●●●●H : High set point 1 output
●●●●● L : Low set point 1 output
●●●● LL : Low set point 2 output
2
Page 3
●3.
Specications
Unit model DT-501X ● DT-501F
Action mode Digital tachometer mode Flowmeter mode Elapsed timecounter mode Time width meter mode
Display
Display 1
0 - 999999
6 digits
Display 2 -
0:00:00 - 9:59:59
(Hour:Minute:Second / base 60 display)
(Hour:Minute:Second / base 10 display)
0:00 - 999:99
0:00:00 - 0:59:59
(Hour:Minute:Second / base 60 display)
With zero suppression function
Decimal point position 0 to 5 digits after the decimal point(000000 ● 0.00000)
-
Number indicator Red 7 segment LED, Letter height 22mm, 6 digits, - display available
LED lamp 8 (SIG, LL, L, H, HH, MAX, MIN, TEA)
Operation key 5 ●SET/SHIFT ,MEM/TEACH , ● /PARA ,
/FUNC ,ENT/MODE●
Input range 0.0067Hz ● 100kHz 10ms - 3600s
Measurement accuracy ±0.008● ±1digit ±0.1● ±1digit
Filter
Note that you can switch between only 10kHz and 0.02kHz in a magnetic sensor, and its contact is only 0.02kHz.
Switches between 100kHz, 30kHz, 10kHz, and 0.02kHz using the parameter.
Display cycle 0.2, 0.5, 1, 2, 5, 10, 15, 30, 60 sec. (changeable in the parameter settings) Dependent on the input signal
Pre-scale function Parameter setting system using the front panel keys. The teaching (combination) of display values are also available.
Memory function
Comparator function
The maximum/minimum measurement values can be memorized and displayed in the indicator.
The settings of the high set point 1, low set point 1, high set point 2, and low set point 2, and judgment results can
be displayed on the LED lamp. The hysteresis setting of the high and low set point 1 values are also available.
(Switches the display using the MEM key)
Auto zero time 0.1 - 150 sec. 0.1 - 3600 sec.
Pre-arithmetic function Updates the displayed value according to the elapsed time after the pulse stops.
Teaching function
Performs scaling automatically by setting the display value with a certain signal input. (only in the tachometer and owmeter modes)
Insulation resistance 10MΩ or more (at DC500V Mega)
Voltage proof AC1500V or more 1min
Operating temperature 0 - 45ºC(No condensation)
Operating humidity 35 - 85%RH (No condensation)
Operating atmosphere No corrosive gas
Conforming standard RoHS
Protection function Front panel: IP66 (or equivalent), Rear terminal block: IP20
Casing material ABS resin
External dimensions ● 96 × ● 48 × ● 92 ● (DIN)
Weight
*There are the input specications in addition to the above specications. For details, refer to p.5 - 6.
●4.
Component Part Names and Functions
FVT● FVC● TRT● BDC options : + approx. 50g
Approx. 200g
CPT options : + approx. 100g
●Front
●●●●●●●●●●●● ● ●●●●●● ●
●●●●●●●●●●●●●● ●●
●Rear
●Side
Name Function
● SIG lamp Lights up when the sensor signal is input
● LL lamp Lights up when the low set point 2 is judged.
● L lamp Lights up when the low set point 1 is judged.
● H lamp Lights up when the high set point 1 is judged.
● HH lamp Lights up when the high set point 2 is judged.
● MAX amp Lights up when the maximum value is displayed
● MIN lamp Lights up when the minimum value is displayed
● TEA lamp Lights up when the teaching function is set
● PEAK lamp Not used
● BTM lamp Not used
● Unit label space Space for attaching the supplied unit labels
● Main display Displays the measurement value
● SET/SHIFT key
● MEM/TEACH key
● ●(UP)/PARA key
●●(NEXT)/FUNC key
● ENT/MODE key
Finishes the setting in various setting modes Pressing this key with other keys switches to various setting modes
Switches to the memory display Pressing this key with the SET key goes to the teaching setting mode
Changes the selected items in various setting modes, or numerical values Pressing this key with the SET key switches to the parameter setting mode
Changes the selected digit in various setting modes Pressing this key with the SET key switches to the function setting mode
Selects the changed item(s) in various setting modes Pressing this key with the SET key switches to the mode setting mode
● Terminal block
● Rear panel
● Terminal block cover 21
Mounting adapter
Unit label
cm
21
PS
kHz rpm
●:●:●
●:●●●
h
sec min rps Hz h:m:s sec min rps Hz
m●●
mm
●● h
●● s
10
m
min
min
h ℓ
min
r
●s
km h
m:s:●
min
mm min
cm
h
min
mm sℓ
m
km
min
h
PS
kHz rpm
m h
10
s
● h
ℓ min
r min
mm min
FVT CPT TRT FVC TRC
BCD RMT DRT SDT SDC
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3
Page 4
●5.
External Dimensions
●6.
Installation to the Panel
Mount this unit to the panel according to the following procedures. Check that the mounting panel is thick enough (1.0 - 5mm) before mounting operation.
Attach the provided waterproof gasket to the panel surface.
1
* If waterproong is not necessary, skip this step.
(1)
Remove the external frame with the cutting lines(backing paper + waterproof gasket) from the gasket sheet.
(Adhesive paste is applied on the both
of the waterproof gasket.)
sides
External frame (backing paper + waterproof gasket)
Backing paper (side with cutting lines)
Backing paper (side without cutting lines)
(2) Attach the provided waterproof gasket to
the edges of the unit attachment panel and opening, aligning with its left and right sides,
and then peel o the backing paper.
*
At this time, be sure not to displace or wrinkle the waterproof gasket. Also, do not expand the waterproof gasket vertically and/or horizontally past the attachment panel’s edges.
Adhesive part of the waterproof gasket
Unit attachment opening
Backing paper
●mm●
CAUTION
!
About waterproong
● Front panel: IP66 (or equivalent)
● Rear terminal block: IP20 (non-waterproof ) Do not install the unit in the following places or conditions. (1) Places regularly subject directly to water (2) Places subject to oil splashes and/or medical supplies (3) Places subject to water splashes on the rear or side face(s). * The front panel is IP66 (or equivalent) waterproofed, but if water is splashed on the unit, be sure to wipe it o the unit as soon as possible.
Place the unit horizontally, and insert the gasket into
2
the opening of the panel surface.
Push the unit until the waterproof gasket (adhesive part) securely reaches the panel surface of the unit.
Waterproof gasket
Find Quality Products Online at: sales@GlobalTestSupply.com
Panel
www.GlobalTestSupply.com
Attach the mounting adapter to the unit.
3
Slide the mounting adapter until it lightly touches the panel surface.
Mounting adapter
4
Page 5
●7.
Wiring to Power Source and Sensors (DT-501XA/DT-501XD)
Note) In order to prevent electric shock, be sure to turn the power OFF. Be sure to use the unit under the rated voltage (AC power specifications: 85 - 264VAC / DC power specifications: 10.8 - 25.2VDC). The inverter output (output to connect a motor) cannot be used as power. Connection wiring from sensors shall not be kept in the same or parallel conduit or cable as the power source, power or high voltage cables. If you fail to separate the wiring, noise may be superimposed on the signal wire, resulting in malfunctions. Use shielded wire for input power connections with the shortest possible metal conduit.
For DT-501XA/DT-501XD
• Terminal block connection diagram
Sensor
Power 1 Power 2 Earth
power output
+12V
Pull-up
Open
collector
Voltage
input
Magnetic
sensor input
0V
1 2 3 4 5 6 7 8 9 10
NC
● Wiring requirements
●●Use M3 crimp-style terminals with a width of 7 mm or less to connect wires to the terminal block.
● After wire connection to the terminal block is completed, be sure to attach the provided terminal block cover.
PWR1 PWR2 E PU
AC power
85 - 264V
DC power
10.8 - 25.2V
-
+
* Model AC power specications: DT-501XA DC power specications: DT-501XD
+12V
Shield wire
Earth
OC
MG
RE
V
oltage output sensor
NPN open collector sensor
(Adjacent switch, gear sensor, rotary encoder)
DC 2-wire sensor
NPN open collector sensor
●adjacent switch●
GND
Magnetic sensor
Relay switch
CAUTION
!
Be sure to connect the power wires to the 1st and 2nd terminals. Incorrect connection could result in damage and/or burns to the unit.
• Input specications • Connection list with sensors
Item Description
Power
AC (DT-501XA) 85 - 264VAC(50/60Hz)
DC (DT-501XD) 10.8 - 25.2VDC Consumption power 10VA Sensor power output DC+12V Max.100mA
Open collector (NPN) input
Open collector input
LO input
Load capacity 12mA or more
0 - 3V HI input Leakage current 0.5mA or less Maximum frequency
(Minimum pulse width 5micro second●
100kHz For no-voltage contact. Short-circuit ● and ● to use.
Contact input
Contact capacity Voltage 12V, Current 15mA or more Maximum frequency
(Minimum pulse width 25micro second●
20Hz LO input 0 - 1.5V
Voltage input
HI input 4.0 - 30V Input resistance Maximum frequency
10kΩ
30kHz
(Minimum pulse width 17micro second●
Input resistance 10k Ω
1Hz ● 100Hz 0.3 ● 30Vp-p
Magnetic sensor input
Input voltage
● 1kHz 1.5 ● 30Vp-p
● 10kHz 6 ● 30Vp-p
Note) Magnetic sensors cannot be used in the time wide meter mode (mode 3).
Maximum frequency
(Minimum pulse width 50micro second●
10kHz
The table below shows the sensors and connection terminal No. to be applied according to the input signal type.Check using the connection diagram on the left.
Input signal type Sensor
Contact signal Relay switch 5-6-9 Open collector Adjacent switch SE-P12-1 4-6-9
Adjacent switch SE-P12
Square wave
Gear sensor SE-G2
Note 1) Connect sensor wires to the specied terminals, and be sure to leave theother terminals empty. More than one sensor cannot be connected simultaneously. Note 2) When one sensor is connected to more than one tachometer, supply the power for the sensor from one of them.
Our product model
Connection terminal
4-7-9Rotary encoder RE-1-●F
• Input circuit
Open collector sensor
Example of external connection
NPN open collector sensor
Example of external connection
Voltage output sensor
Example of external connection Example of external connection Example of external connection
NPN voltage output sensor
Find Quality Products Online at: sales@GlobalTestSupply.com
PNP voltage output sensor
www.GlobalTestSupply.com
DC 2-wire sensor
PNP open collector sensor
Relay switch
Magnetic sensor
Example of external connection
Example of external connection
Relay switch
Magnetic sensor
5
Page 6
●8.
Wiring to Power Source and Sensors (DT-501FA)
Note) In order to prevent electric shock, be sure to turn the power OFF. Be sure to use the unit under the rated voltage (AC power specifications: 85 - 264VAC). The inverter output (output to connect a motor) cannot be used as power. Connection wiring from sensors shall not be kept in the same or parallel conduit or cable as the power source, power or high voltage cables. If you fail to separate the wiring, noise may be superimposed on the signal wire, resulting in malfunctions. Use shielded wire for input power connections with the shortest possible metal conduit.
For DT-501FA
• Terminal block connection diagram
Sensor
Power 1
Power 2 Ear th
power output
+12V
No connection No connection
Dierential
input +
Dierential
input -
0V
1 2 3 4 5 6 7 8 9 10
No connection
● Wiring requirements
● Use M3 crimp-style terminals with a width of 7 mm or less to connect wires to the terminal block.
● After wire connection to the terminal block is completed, be sure to attach the provided terminal block cover.
PWR1 PWR2 E NC NC
Earth
AC power
85 - 264V
* Model AC power specications: DT-501FA
DIF+
DIF-
Dierential output
NC
GND+12V
• Terminal block connection diagram
Item
Power AC (501FA) 85 - 264VAC(50/60Hz)
Consumption power
Sensor power output
Dierential input
●9.
Basic Setting Procedure
10VA DC+12V Max.100mA
Connection to Dierential line driver AM26LS31 etc.
Dierential input voltage
Maximum frequency
Description
DIF+
DIF
V
DIF-
Maximum voltage±5.5V(15mA)
DIF
V
Minimum voltage
(Minimum pulse width 5micro second●
100kHz
±3.0V
Conduct settings as indicated below based on the intended use.
CAUTION
!
Be sure to connect the power wires to the 1st and 2nd terminals. Incorrect connection could result in damage and/or burns to the unit.
• Line driver input circuit
Example of external connection
Line driver
Parameter settings
2
Mode (function) settings
1
Refer to p.8 - 15.
Refer to p.8 - 15.
Teaching function settings
2
Function (action configuration)
3
settings
Refer to p.16 - 17.
Measurement start
Refer to p.7.
Conduct the comparator function p.17 and memory function settings p.18, if necessary. This unit requires no complicated calculation, and implements the teaching function which enables optional changes of display values and error compensation. (excluding the elapsed timecounter and time wide meter modes)
10.
Keys to be Used for Various Settings and Their Applications
The front panel keys to be used for mode, parameter, and function settings, as well as various settings (teaching function / high and low set point 1 value setting / memory function) are described below.
• Operation during the normal measurement display / memory display
Name Function
SHIFT key MEM key
● TEACH key Pressing and holding this key with the SHIFT key for 5 seconds switches to the teaching setting mode
PARA key
FUNC key
MODE key
Pressing this key with other keys switches to various setting modes Switches to the memory display
Pressing and holding this key with the SHIFT key for 5 seconds switches to the parameter setting mode Pressing and holding with the SHIFT key for 5 seconds switches to the function setting mode Pressing and holding this key with the SHIFT key for 5 seconds switches to the parameter setting mode
• Operation in the various setting modes
Name Function
● ● ● ● ●
Find Quality Products Online at: sales@GlobalTestSupply.com
● SET key Setting completion key in various setting modes
(UP) key Changes the selected items in various setting modes, or numerical values
s
(NEXT) key Changes the selected digit in various setting modes
t
ENT key Selects the changed item in various setting modes
www.GlobalTestSupply.com
6
Page 7
●11.
Teaching Function Settings
When you can check (measure) the actual revolution speed, the following simple method (teaching function) can be used. The teaching function requires no complicated calculation.
● The teaching function is supported only in mode 1 and mode 4. Mode 2 and mode 3 have no teaching function.
● Setting method
1 0
SHIFT + MEM keys
5 seconds elapsed
Teaching function setting mode
TEA
When you press the SHIFT + MEM keys simultaneously during the normal measurement display, the main display blinks.
●●●●●●●●●●●●●●●●●●●●●●●●(TEA lamp lights up)
The measurement value which was displayed before is displayed in the main display. In the initial status, the digit on the left blinks. Because the value in the blinking digit can be
0 0 0 0 1 0
Use the
key
to select the digit
TEA
0 1 0 0 1 0
key to
Use the ●
select the value
TEA
0 1 0 0 1 0
Use the SET key to
nish the setting
Returns to the normal measurement display
changed, use the Every time you press the digit”, and “decimal point”, then back to “1st digit”.
After selecting the digit you want to change, use the
When you select the decimal point, use the
When the numerical value setting is complete, press the SET key to finish the teaching setting. The teaching setting result is reflected, and the display returns to the normal measurement display.
key to select the digit in which you want to change the value.
t
key, the digit blinks in the order of “1st digit”, “2nd digit”, ... “6th
t
key to change the value (0 - 9).
s
key to change the decimal point position.
s
0 1 0 0 1 0
● Teaching function details
• Mode 1
The teaching function can be used to automatically change the value of parameter items (P2, P3), and calculate the present measurement value from the teaching setting value as below.
External input pulse (Hz) A
Parameter setting value
Display calculation formula
When the input revolution speed is outside the input revolution speed range, “EE-2” is displayed, and the teaching function cannot be used.
During normal display
P1●Pulse count per revolution B P2●Setting revolution speed (detection section) P3●Value to be displayed D
{ (A/B) * 60 } * (D/C)
Input pulse (Hz) in the teaching setting A Teaching function setting value F
C
Parameter setting value
Display calculation formula
When the teaching function is executed
P1●Pulse count per revolution B P2●Setting revolution speed (detection section) P3●Value to be displayed D = F
{ (A/B) * 60 } * (D/C)
C = A * 60(Hz)/ B
Input revolution speed range : 1rpm ● Input revolution speed* ● 99999rpm
●●●●●●●●●●●●●●●●●●●●●●
* The P2 calculation value is maintained by rounding o fractions after the decimal point.
Depending on input and teaching setting values, the teaching setting value for the input at the time might not be displayed.
*Input revolution speed = A × 60(Hz) / B
• Mode 4
The teaching function can be used to automatically change the value of parameter items (P3, P4), and calculate the present measurement value from the teaching setting value as below.
External input pulse (Hz) A
Parameter setting value
Display calculation formula
When the input frequency is outside the input frequency range, “EE-2” is displayed, and the teaching function cannot be used.
Input frequency range●: 1Hz
*
The P3 scaling value maintains values after decimal point up to 7 digits inside. Note that because the number of digits after decimal point that can be maintained in the parameter setting is up to 5, values in 5 digits or more after decimal point will be rounded o when you enter the value in the parameter setting.
Find Quality Products Online at: sales@GlobalTestSupply.com
* If the P3 calculation value exceeds 999999, the P3 value will be forced to be “999999”.
Mode 4 Display value calculation
Number of blades per
P1
revolution P2 Capacity per sensor blade C P3 Scaling D P4 Decimal point display
●A * B● * C * D
Input pulse (Hz) in the teaching setting A Teaching function setting value F
B
Parameter setting value
E
Display calculation formula
●●●A●●●
99999Hz
www.GlobalTestSupply.com
Mode 4 Display value calculation
Number of blade per
P1
revolution P2 Capacity per sensor blade C P3 Scaling D = F/ (A * B * C ) P4 Decimal point display Decimal point position of E = F
●A * B● * C * D
B
7
Page 8
●12.
About Mode
DT-501X/DT-501F have ve modes (functions) which can be selected according to the measurement purpose.
Mode
No.
99 Test Mode Executes the self diagnosis p.19
* The mode 1 (digital tachometer mode) is set in the factory setting.
Mode description Details
1 Digital Tachometer Mode* Digital tachometer / used as speedmeter●Displays the proportional value to the input p.8 - 9 2 Elapsed Timecounter Mode Used as the elapsed time display in a certain period●Displays the inversely proportional value to the input p.10 - 11 3 Time Width Meter Mode Measures time for the period when the signal is ON (or OFF) and displays it p.12 - 13 4 Flowmeter Mode Displays the instantaneous ow amount p.14 - 15
for the setting method
For the lists of parameters and functions to be set in each mode, refer to pages 20 and 21.
Page No.
Setting Method
13.
When You Select Mode 1 (Digital Tachometer Mode) for Measurement
Digital tachometer mode
Set each setting item for mode and parameter according to the following procedures. (For the setting method of functions, refer to pages 16 and 17.)
● Setting example (for measuring the peripheral speed of the roller)
As in the gure on the left, the rotary encoder is connected to the feed roller via coupling in the paper manufacturing process. To display the peripheral speed (m/min.) of the feed roller in DT-501under the following conditions:
[Conditions]
● Servo motor (pulse input)
● Diameter of the feed roller
● Peripheral speed of the roller diameter of 90mm ●
●0.09m×3.14)×500rpm ● 141.3m/min
Circumference
●●●●
of the roller
Revolution speed*
*Suppose that the revolution speed in the detection section is 500rpm because the revolution speed is not instructed.
● Setting method (1) Mode
• When you select mode 1 (digital tachometer mode) for measurement (The digital tachometer mode is set in the factory setting.)
Normal display
The main display shows “0”.
0
Mode setting
Finish setting
Note) If you change the mode settings, each parameter and function setting value will return to the factory default setting value.
Press and hold the SHIFT + MODE keys for 5 seconds or more. The main display shows the current set mode No. after blinking.
* When you are setting for the rst time after the
purchase of this unit, mode 1(digital tachometer
mode) in the factory default setting is displayed.
Press the SET key. The display goes to mode 1 (digital tachometer mode).
- 0 1 -
0
The numerical value does not need to be changed and you can proceed to the next step. * Every time you press the
the mode No. switches.
s
key,
(2) Parameter
Parameter setting items in mode 1 (digital tachometer mode) nd numerical values to be set according to the above example
No. Setting item Setting range
Pulse count per revolution
P1
Revolution speed in the
P2
detection section Value to be displayed
P3
(with decimal point)
P4 Display cycle
P5 Auto zero time 0.1 - 150 sec. _ _ 006.0 6.0 sec. 6 sec. Measure using the default setting value as no special instruction is provided.
P6 Input lter 10/30/100/0.02kHz _ 10_ 10kHz 30kHz As the rotary encoder is used as a sensor, select "30".
Find Quality Products Online at: sales@GlobalTestSupply.com
* For the description about the parameter functions, refer to page 20.
1 - 9999p/r _ _ 0001 1p/r 600p/r Enter the pulse count of the rotary encoder.
1 - 99999rpm _01000 1000rpm 500rpm Enter some numerical value as the revolution speed is not instructed.
0.00001 - 999999 001000 1000 141.3 Refer to the above "Setting example".
0.2/0.5/1.0/2.0/
5.0/10/15/30/60 sec.
Default setting value
Display Description
_ 1.0 _ 1 sec. 1 sec. Measure using the default setting value as no special instruction is provided.
Numerical value to be set according to the above example
www.GlobalTestSupply.com
8
Page 9
● Mode 1 (digital tachometer mode) Display value calculation equation
Mode 1 Display value calculation
External input pulse (Hz) A
P1 Pulse count per revolution B
Parameter setting
Equation ●(A/B)
P2 P3 Value to be displayed D
Revolution speed in the detection section(rpm)
*
*
60)
(D/C)
C
• Start setting according to the setting example on the previous page.
Parameter setting
Press and hold the SHIFT + PARA keys for 5 seconds or more to switch to the parameter setting mode.
0 1 - P 0 1
Parameter 1 setting
Parameter 2 setting
Parameter 3 setting
Parameter 4 setting
Parameter 5 setting
Parameter 6 setting
Press the ENT key and start setting parameters. (The main display shows t merical value of each parameter.)
Every time you press the key, the display switches to the next pa rameter.
Every time you press the ● key, the display switches to the next parameter.
he current set nu
s
0 0 0 1
--
-
0 1 0 0 0
-
-
0 0 1 0 0 0
2. 0-
-
0 1 0. 0
--
0 1 0. 0
Use the key to move to another digit
t
s
Use the key to change the numerical value to “600”. Use the ENT key to nish setting.
Use the
key to move to another digit
t
Use the key to change the numerical value
s
to “500”. Use the ENT key to nish setting.
t
key to move to another digit
Use the
s
Use the key to change the numerical value to “141.3”. Use the ENT key to nish setting. (Use the use the key to change its position.)
Every time you press the ● key, the display switches as below.
Select "1.0" and use the ENT key to nish setting.
Use the key to move to another digit Use the key to change the numerical value to “6.0”. Use the ENT key to nish setting. (Use the use the key to change its position.)
Every time you press the key, the display switches as below.
Select "30" and use the ENT key to nish setting.
key to select the decimal point, and
t
s
0.2 > 0.5 >1.0 >2.0 >5.0 60●<●30●<●15●<●10
t
s
key to select the decimal point, and
t
s
s
●10 > 30 >100 > 0.02
●●●●●●●●●●●●
--
-
-
--
-
0 6 0 0
5 0 0
1 4 1. 3
1. 0-
6. 0
3 0
---
Press the SET key.
Finish setting
*If you enter a value outside the setting range, all the input values will blink, which indicates the setting is disabled.
●When this occurs, you cannot switch to other displays until you enter a value within the setting range.
The display goes to mode 1 (digital tachometer mode).
0. 0
(3) Function (For the setting method, refer to pages 16 and 17.)
• Function setting item in mode 1 (digital tachometer mode)
No. Setting item Setting range Input range Default value
F1 High set point 1 value Sets the high set point 1 value 000000 - 999999 0 F2 Low set point 1 value Sets the high set point 1 value 000000 - 999999 0 F3 High set point 2 value Sets the high set point 2 value 000000 - 999999 0 F4 Low set point 2 value Sets the low set point 2 value 000000 - 999999 0
Hysteresis of the high and
F5
low set point 1 values
Judgment output timer at
F6
startup
F7 Minimum revolution speed Sets the revolution speed to be displayed as zero 000000 - 999999 0
Frequency of the moving
F8
average
F9 Pre-arithmetic function
Find Quality Products Online at: sales@GlobalTestSupply.com
* For the description of functions, refer to pages 21 and 22.
Sets the hysteresis of the high and low set point 1 values
Sets the time when the comparator judgment is output at startup
Used when variation of the revolution speed is large and a stable display cannot be attained
Promptly performs the deceleration display when the signal is lost
www.GlobalTestSupply.com
0 - 99 0
0 - 99 sec. 0 sec.
0(disabled)/1(3 times)/2(10 times) 0(disabled)
0(disabled)/1(enabled) 0(disabled)
9
Page 10
Setting Method
14.
When You Select Mode 2 (Elapsed Timecounter Mode) for Measurement
Set each setting item for mode and parameter according to the following procedures. (For the setting method of functions, refer to pages 16 and 17.)
● Setting example (for the elapsed time display of oven)
As in the gure on the left, bread passes through the oven in a bread fac­tory. The rotary encoder is attached to the motor of the conveyor belt. To display the time at which the bread passes through the oven in DT-501, do so under the following conditions:
[Conditions]
● Revolution speed of the rotary encoder
1200rpm(
rotary encoder
● Speed of the conveyor belt
● Length of the oven
● Passing time through the oven
5m÷2.8m/min ● 1.786 minutes ●Approx. 1 minute 47 seconds
• Setting method (1) Mode
• Set the mode 2 (elapsed timecounter mode).
Elapsed timecounter mode
60p/r●
● 2.8m/min
● 5m
Normal display
Mode setting
Finish setting
Note) If you change the mode settings, each parameter, function, and high and low set point 1 setting value will return to the factory default setting value.
The main display shows “0”.
Press and hold the SHIFT + MODE keys for 5 seconds or more. The main display shows the current set mode No. after blinking.
* When you are setting for the rst time after the
purchase of this unit, mode 1(digital tachometer
mode) in the factory default setting is displayed.
Press the SET key. The display goes to mode 2 (elapsed timecounter mode
).
-
0
0 1
0: 0 0
Every time you press the key, the display switches to the next mode No.
1●2●3●4
-
Select "2".
0 2
-
s
99
-
(2) Parameter
• Parameter setting items in mode 2 (elapsed timecounter mode) and numerical values to be set according to the above example.
No. Setting item Setting range
P1 Pulse count per revolution 1 - 9999p/r _ _ 0001 1p/r 60p/r Enter the pulse count of the rotary encoder.
Revolution speed in the detection
P2
section
Value to be
P3
displayed
P4 Display cycle
P5 Auto zero time 0.1 - 150 sec. _ _ 006.0 6.0 sec. 6 sec.
P6 Input lter 10/30/100/0.02kHz _ 10_ 10kHz 30kHz As the rotary encoder is used as a sensor, select "30".
* For the description about the parameter functions, refer to page 20.
Hour:Minute: display system Second display system
Second
1000 - 99999rpm _01000 1000rpm 1200rpm Enter the revolution speed of the rotary encoder.
0:00:01 - 9:59:59
0:01 - 999:99
0.2/0.5/1.0/2.0/
5.0/10/15/30/60 sec.
Default setting value
Display Description
_010:00 10:00 sec.
_ 1.0 _ 1 sec. 1 sec.
Hour:Minute:Second display system 0:01:47
Numerical value to be set according to the above example
Refer to the above "Setting example".
Measure using the default setting value as no special instruction is provided. Measure using the default setting value as no special instruction is provided.
● Mode 2 (elapsed timecounter mode) Display value calculation equation
Mode 2 Display value calculation
External input pulse (Hz) A
P1 Pulse count per revolution B
Parameter setting
Equation ●1/(((A/B)
Find Quality Products Online at: sales@GlobalTestSupply.com
P2
Revolution speed in the detection section(rpm)
P3 Value to be displayed D
*
*
60)/C))
D
www.GlobalTestSupply.com
C
10
Page 11
●Start setting according to the setting example on the previous page.
Normal display
To the parameter settings
Parameter 1 setting
Parameter 2 setting
Parameter 3 setting
Parameter 4 setting
Parameter 5 setting
The main display shows “0:00”.
Press and hold the SHIFT + PARA keys for 5 seconds or more to switch to the param eter setting mode.
Press the ENT key and start setting parameters. (The main display shows the current set numerical value of each parameter.) Every time you press the ● key, the display switches to the next parameter.
0: 0 0
0 2 - P 0 1
-
0 0 0 1
--
0 1 0 0 0
-
0 1 0: 0 0
-
2. 0-
-
0 1 0. 0
--
Use the
key to move to another digit
t
s
Use the key to change the numerical value to “60”. Use the ENT key to nish setting.
Use the key to move to another digit
t
Use the key to change the numerical value
s
to “1200”. Use the ENT key to nish setting.
Use the key to move to another digit
t
s
Use the key to change the numerical value to “0:01:47”. Use the ENT key to nish setting. To switch the Hour:Minute:Second display (0:10:00) and Second display (10:00), use the
s
key with ":" selected by the key.
Every time you press the ● key, the display switches as below.
0.2 ● 0.5 ● 1.0 ● 2.0 ● 5.0
Select "1.0" and use the ENT key to nish setting.
Use the key to move to another digit Use the key to change the numerical value to “6.0”. Use the ENT key to nish setting. (Use the key to select the decimal point, and use the key to change its position.)
60●●●30●●●15●●●10
t
s
t
s
t
0 0 6 0
--
1 2 0 0
-
0: 0 1: 4 7
-
1. 0-
-
--
6. 0
Every time you press the ● key, the
Parameter 6 setting
Finish setting
*If you enter a value outside the setting range, all the input values will blink, which indicates the setting is disabled.
●When this occurs, you cannot switch to other displays until you enter a value within the setting range.
Press the SET key. The display goes to mode 2 (elapsed timecounter mode).
1 0
-
- -
0: 0 0: 0 0
display switches as below.
Select "30" and use the ENT key to nish setting.
10 ● 30 ● 100 ● 0.02
3 0
-
(3) Function (For the setting method, refer to pages 16 and 17.)
• Function setting item in mode 2 (elapsed timecounter mode)
No. Setting item Description Input range Default setting
F1 High set point 1 value
F2 Low set point 1 value
F3 High set point 2 value
F4 Low set point 2 value
Hysteresis of the high
F5
and low set point 1 values Judgment output timer
F6
at startup Minimum revolution
F7
speed
Frequency of the moving
F8
average
F9 Pre-arithmetic function Promptly performs the deceleration display when the signal is lost 0(disabled)/1(enabled) 0(disabled)
* For the description of functions, refer to page 21.
Note) When the input signal stops in the elapsed timecounter mode, the display will overow after the specied time set with the auto zero function.
Find Quality Products Online at: sales@GlobalTestSupply.com
The overow display “-.-.-.-.-.-. does not mean there is a malfunction. When the input signal is within the display range, the display returns to normal.
Sets the high set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Second display system
Sets the low set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Second display
Sets the high set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Second display
Sets the low set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Second display
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Sets the revolution speed to be displayed as zero 000000 - 999999 0
Used when variation of the revolution speed is large and a stable display cannot be attained
0(disabled)/1(3 times)/2(10 times) 0(disabled)
0:00Sets the high set point 1 value (Second display system) 0:00 - 999:99
0:00Sets the low set point 1 value (Second display system) 0:00 - 999:99
0:00Sets the high set point 2 value (Second display system) 0:00 - 999:99
0:00Sets the low set point 2 value (Second display system) 0:00 - 999:99
www.GlobalTestSupply.com
11
Page 12
Setting Method
15.
When You Select Mode 3 (Time Width Meter Mode) for Measurement
Set each setting item for mode and parameter according to the following procedures. (For the setting method of functions, refer to pages 16 and 17.)
● Setting example (for the operating time display of the press machine)
As in the figure on the left, when the press machine finishes operation and returns to the upper edge, the adjacent switch is activated. To display the operating time of the press machine (when the proximity switch is turned o)
● Setting method (1) Mode
• Set mode 3 (Time Width Meter Mode).
Time wide meter mode
Normal display
Mode setting
Finish setting
Note) If you change the mode settings, each parameter, function, and high and low set point 1 setting value will return to the factory default setting value.
The main display shows “0”.
Press and hold the SHIFT + MODE keys for 5 seconds or more. The main display shows the current set mode No. after blinking.
* When you are setting for the rst time after the purchase of this unit, mode 1 (digital tachometer mode) in the factory default setting is displayed.
Press the SET key. The display goes to mode 3 (time wide meter mode).
0
0 1-
-
0: 0 0
Every time you press the key, the display switches to the next mode No.
1●2●3●4
99
Select "3".
0 3-
-
(2) Parameter
• Parameter setting items in mode 3 (Time Width Meter Mode) and numerical values to be set according to the above example.
No. Setting item Setting range
Switch between Hour:Minute:
P1
Second and 1/100 Second system
P2 Measurement section 0(OFF)/1(ON) _ 1_ ON 0 Select "0" because the adjacent switch of the open collector is OFF. P3 Auto zero time 0.1 - 3600 sec. _3600.0 3600 sec. 60.0 sec. Set "60.0" because there is no time setting longer than 1 minute. P4 Input lter 10/0.02kHz _ 10_ 10kHz 10kHz
* For the description about the parameter functions, refer to page 20.
0:00:00/0:00 0:00 1/100 sec. 0:00
Default setting value
Display Description
Numerical value to be set according to the above example
Select the 1/100 second display based on the operating time of the press machine.
Measure using the default setting value as no special instruction is provided.
s
● About the measurement section (parameter 2)
Relay switch, open collector, DC 2-wire sensor ON (When the No.6 terminal is at L level)
Voltage pulse ON (When the No.7 terminal is at H level)
1 0
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
12
Page 13
• Start setting according to the setting example on the previous page.
Normal display
The main display shows “0:00”.
0: 0 0
To the parameter settings
Parameter 1 setting
Parameter 2 setting
Parameter 3 setting
Parameter 4 setting
Finish setting
Press and hold the SHIFT + PARA keys for 5 seconds or more to switch to the param eter setting mode.
Press the ENT key and start setting parameters. (The main display shows the current set numerical value of each parameter.) Every time you press the key, the display switches to the next parameter.
Press the SET key. The display goes to mode 3 (time wide meter mode).
s
0 3 - P 0 1
-
0: 0 0 0: 0 0
1
-
3 6 0 0. 0
-
1 0
-
0: 0 0
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical
s
value.)
Every time you press the
- -
-
switches as below.
Select "0" and use the ENT key to finish setting.
Use the Use the key to change the numerical value to “60.0”. Use the ENT key to nish setting.
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical value.)
0●1
key to move to another digit
t
s
s
key, the display
s
-
0
-
0 0 6 0. 0
1 0-
-
*If you enter a value outside the setting range, all the input values will blink, which indicates the setting is disabled.
●When this occurs, you cannot switch to other displays until you enter a value within the setting range.
(3) Function (For the setting method, refer to pages 16 and 17.) Function setting item in mode 3 (Time Width Meter Mode)
No. Setting item Description Input range Default setting
F1 High set point 1 value
F2 Low set point 1 value
F3 High set point 2 value
F4 Low set point 2 value
Hysteresis of the high and
F5
low set point 1 values Judgment output timer at
F6
startup
* For the description of functions, refer to pages 21 and 22.
Sets the high set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59
Sets the high set point 1 value (Second display system) 0:00 - 999:99
Sets the low set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59
Sets the low set point 1 value (Second display system) 0:00 - 999:99
Sets the high set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59
Sets the high set point 2 value (Second display system) 0:00 - 999:99
Sets the low set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59
Sets the low set point 2 value (Second display system) 0:00 - 999:99
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Second display system
0:00
Second display system
0:00
Second display system
0:00
Second display system
0:00
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13
Page 14
Setting Method
16.
When You Select Mode 4 (Flowmeter Mode) for Measurement
Set each setting item for mode and parameter according to the following procedures. (For the setting method of functions, refer to pages 16 and 17.)
● Setting example (For the ow display from the tank)
To display the ow (ℓ/min) in DT-501 under the following conditions when the owmeter for which the number of blades is unknown as in the gure on the left:
[Conditions]
● Number of blades : Unknown
● Output of the ow sensor : 2.5cc/p
Flowmeter mode
*p : Pulse
DT-501●
Flow sensor
● Setting method (1) Mode
• Set mode 4 (owmeter mode).
Normal display
Mode setting
Finish setting
Note) If you change the mode settings, each parameter, function, and high and low set point 1 setting value will return to the factory default setting value.
The main display shows "0".
Press and hold the SHIFT + MODE keys for 5 seconds or more. The main display shows the current set mode No. after blinking.
* When you are setting for the rst time after the purchase of this unit, mode 1 (digital tachometer mode) in the factory default setting is displayed.
Press the SET key. The display goes to mode 4 (owmeter mode).
0
0 1-
-
0. 0
Every time you press the key, the dis­play switches to the next mode No.
1●2●3●4
99
Select "4".
0 4-
-
(2) Parameter
• Parameter setting items in mode 4 (owmeter mode) and numerical values to be set according to the above example.
No. Setting item Setting range
P1 Number of blades per revolution 1 - 99 _ _ _ _01 1 1 P2
Capacity per sensor blade (cc, ℓ , etc.)
P3 Scaling 0.00000 - 999999 00001.0 1 0.06 Refer to "About scaling" below.
P4 Decimal point display 0.00000 - 00000.0 00000.0
P5 Display cycle
P6 Auto zero time 0.1 - 150 sec. _ _006.0 6 sec. 6 sec. P7 Input lter 10/30/100/0.02kHz _ 10_ 10kHz 10kHz
* For the description about the parameter functions, refer to page 20.
0.0001 - 99999 _ 0001.0 1.0 2.5cc/p Enter the capacity per blade that the sensor can read.
0.2/0.5/1.0/2.0/
5.0/10/15/30/60 sec.
Default setting value
Display Description
Displays to one
decimal place
_1.0_ 1 sec. 1 sec.
00000.0
Numerical value to be set according to the above example
Enter the number of blades of the owmeter. If it is unknown, enter "1".
Displays to one decimal place.
Measure using the default setting value as no special instruction is provided.
Measure using the default setting value as no special instruction is provided. Measure using the default setting value as no special instruction is provided.
s
● Mode 4 (owmeter mode) Display value calculation equation
External input pulse (Hz) A
Parameter setting
Equation (A
Mode 4 Display value calculation
P1 Number of blades per revolution B P2 Capacity per sensor blade C P3 Scaling D P4 Decimal point display E
* B) * C *
D
● About scaling (parameter 3)
Assuming the ow per second to be represented in the parameter 2 ow unit (cc in the example) is a (cc/s), parameter 3 will be coecient value when a multiplier is used to change the unit of
(1) To display the value as in the original unit (cc/s)
● ● Parameter 3
(2) To display the value in the unit (ℓ /min) as in the above setting
Find Quality Products Online at: sales@GlobalTestSupply.com
Parameter 3
*
(cc/s)
1●a (cc/s)
a
*
(cc/s)
60 ÷ 1000 ●A ( ℓ /min)
a
www.GlobalTestSupply.com
(cc/s) for display.
a
*
60 : Multiply it by 60 as it is the ow per minute.
* ÷1000 : Divide the value by 1000 as parameter is set in the unit of cc.
14
Page 15
• Start setting according to the setting example on the previous page.
Normal display
To the parameter settings
Parameter 1 setting
Parameter 2 setting
Parameter 3 setting
Parameter 4 setting
The main display shows “0:00”.
Press and hold the SHIFT + PARA keys for 5 seconds or more to switch to the param eter setting mode.
Press the ENT key and start setting parameters. (The main display shows the current set numerical value of each parameter.) Every time you press the ● key, the display switches to the next parameter.
0. 0
0 4 - P 0 1
-
-- --
0 0 0 1. 0
-
0 1
0 0 0 0 1. 0
0 0 0 0 0. 0 0 0 0 0 0. 0
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical
s
value.)
s
Use the key to move to another digit Use the key to change the numerical value
s
to “2.5”. Use the ENT key to nish setting.
s
Use the Use the key to change the numerical value to “0.06”. Use the ENT key to nish setting. Use the and use the key to change its position.
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical value.)
key to move to another digit
s
s
key to select the decimal point,
s
s
-- --
0 0 0 2. 5
-
0 0 0 0. 0 6
0 1
Parameter 5 setting
Parameter 6 setting
Parameter 7 setting
Finish setting
*If you enter a value outside the setting range, all the input values will blink, which indicates the setting is disabled.
●When this occurs, you cannot switch to other displays until you enter a value within the setting range.
Press the SET key. The display goes to mode 4 (owmeter mode).
--
1. 0-
-
0 0 6. 0
-
1 0
- -
0. 0
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical
s
value.)
The numerical value does not need to be changed, and go to next step.
s
(Use the (Use the key to change the numerical value.)
The numerical value does not need to be changed and you can proceed to the next step by pressing the ENT key. (Use the key to change the numerical value.)
key to move to another digit)
s
s
-
--
1. 0
-
0 0 6. 0
1 0
-
(3) Function (For the setting method, refer to pages 16 and 17.)
• Function setting item in mode 4 (owmeter mode)
No. Setting item Description Input range Default setting
F1 High set point 1 value Sets the high set point 1 value 000000 - 999999 0 F2 Low set point 1 value Sets the high set point 1 value 000000 - 999999 0 F3 High set point 2 value Sets the high set point 2 value 000000 - 999999 0 F4 Low set point 2 value Sets the low set point 2 value 000000 - 999999 0
Hysteresis of the high and
F5
low set point 1 values
Judgment output timer at
F6
startup
Minimum ow(display value)
F7
Frequency of the moving
F8
average
Find Quality Products Online at: sales@GlobalTestSupply.com
F9 Pre-arithmetic function Promptly performs the deceleration display when the signal is lost 0(disabled)/1(enabled) 0(disabled)
* For the description of functions, refer to page 21.
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Sets the revolution speed to be displayed as zero 000000 - 999999 0 Used when variation of the revolution speed is large and a stable display cannot be attained
0(disabled)/1(3 times)/2(10 times) 0(disabled)
www.GlobalTestSupply.com
15
Page 16
17.
Setting Method of Functions (Excluding in the Test Mode, Common in Each Mode)
The setting method of the functions is common in each mode excluding the test mode. For the lists of functions in each mode, refer to page 21.
• Function setting item <in mode 1 (digital tachometer mode)>
No. Setting item Setting range
F1 High set point 1 value* 000000 - 999999 000000 0 Sets the high set point 1 value. The decimal point is not displayed.
Low set point 1 value*
F2
F3 High set point 2 value* 000000 - 999999 000000 0 Sets the high set point 2 value. The decimal point is not displayed. F4 Low set point 2 value* 000000 - 999999 000000 0 Sets the low set point 2 value. The decimal point is not displayed.
Hysteresis of the high
F5
and low set point 1 values
Judgment output timer
F6
at startup
F7 Minimum revolution speed 000000 - 999999 000000 0 Sets the revolution speed to be displayed as zero
Frequency of the moving
F8
average
F9 Pre-arithmetic function 0(disabled)/1(enabled) _ 0 _ 0(disabled) Promptly performs the deceleration display when the signal is lost
*For the settings of the high set point 1 and 2 values, as well as the low set point 1 and 2 values, refer to “Comparator Function” on page 17.
000000 - 999999 000000 0 Sets the low set point 1 value. The decimal point is not displayed.
0 - 99 _ _ _ _ 00 0
0 - 99 sec. _ _ _ _ 00 0 sec. Sets the time when the comparator judgment is output at startup
0(disabled)/1(3 times)/2(10 times)
Default setting
Display
_ 0 _ 0(disabled)
Description
Function description
Sets the hysteresis of the high and low set point 1 values. Used when the revolution variation is large. (For details, refer to "Comparator Function" on page 17.)
Used when variation of the revolution speed is large and a stable display cannot be attained
• Setting method <in mode 1 (digital tachometer mode)> For the setting methods in other modes than mode 1 (digital tachometer mode), also refer to the description below.
Normal display
To the function settings
Function 1 setting
● Setting example : 1000
The main display shows “0”.
Press the SHIFT + FUNC keys for 5 seconds or more to switch to the parameter set ting mode.
Press the ENT key and start setting functions. Every time you press the key, the display switches to the next parameter.
s
0
0 1 - F 0 1
-
s
0 1 - F 0 1 0 0 1 0 0 0
Use the key to move to another digit
s
Use the key to change the numerical value to “1000”. Use the ENT key to nish setting.
s
Use the
Function 2 setting
● Setting example : 500
Function 3 setting
● Setting example : 1200
Function 4 setting
● Setting example : 300
Function 5 setting
● Setting example : 20
Function 6 setting
● Setting example : 3
Function 7 setting
● Setting example : 100
Find Quality Products Online at: sales@GlobalTestSupply.com
Continued on next page
www.GlobalTestSupply.com
0 0 0 0 0 0
0 0 0 0 0 0
0 0 0 0 0 0
-- --
-- --
0 0
0 0
0 0 0 0 0 0
16
Use the key to change the numerical value to “500”. Use the ENT key to nish setting.
Use the key to move to another digit Use the key to change the numerical value to “1200”. Use the ENT key to nish setting.
Use the key to move to another digit Use the key to change the numerical value to “300”. Use the ENT key to nish setting.
Use the Use the key to change the numerical value to “20”. Use the ENT key to nish setting.
Use the Use the key to change the numerical value to “03”. Use the ENT key to nish setting.
Use the Use the key to change the numerical value to “100”. Use the ENT key to nish setting.
key to move to another digit
s
s
s
s
s
s
key to move to another digit
s
s
key to move to another digit
s
s
key to move to another digit
s
0 0 0 5 0 0
0 0 1 2 0 0
0 0 0 3 0 0
-- --
-- --
2 0
0 3
0 0 0 1 0 0
Page 17
Continued from previous page
Function 8 setting
● Setting example : 1
Every time you press the display switches to the next func tion. Press the ENT key and start setting functions.
Function 9 setting
● Setting example :1
Finish setting
* If you enter a value outside the setting range, all the input values will blink, which indicates the setting is disabled. When you press any of the numerical input key, the numerical value before entry is displayed, and you can redo the setting. * When the relationship among the setting values of F1 to F4 is not appropriate, "EE-4" is displayed before nishing the function setting by pressing the SET key. For details, refer to page 19.
●18.
Comparator Function
Press the SET key. The display goes to mode 1 (digital tachometer mode).
s
key, the
-
0
-
-
0
-
-
0
Every time you press the key, the display switches as below. Use the ENT key to nish setting.
Every time you press the key, the display switches as below. Use the ENT key to nish setting.
s
s
• Keys to be used for the high and low set point 1 value settings and their applications
● ● ● ●
Judgment lamp
The corresponding lamp lights up according to the judgment result.
Name Function
● LL lamp Lights up when the low set point 2 is judged.
● L lamp Lights up when the low set point 1 is judged.
● H lamp Lights up when the high set point 1 is judged.
● HH lamp Lights up when the high set point 2 is judged.
1
-
-
1
-
-
• Description
● Judgment conditions
Low set point 1 value ● Measurement value ● High set point 1 value AND Low set point 2 value ● Measurement value ● High set point 2 value When either the high and low set point 1 value judgment or the high and low set point 2 value judgment is invalid, and the other is within the setting range Measurement value ● High set point 2 value HH judgment Measurement value ● High set point 1 value H judgment Measurement value ● Low set point 2 value LL judgment Measurement value ● Low set point 1 value L judgment
GO judgment
● Set the high set point 1, low set point 1, high set point 2, and low set point 2 values using the function items F01 to F04.
● Execute the judgment in the all measurement mode.
● The high and low set point 2 value judgment, as well as the high and low set point 1 value judgment, is executed independently.
● When both the high set point 2 and low set point 2 values are “0”, the high and low set point 2 value judgment is not executed.
● When both the high set point 1 and low set point 1 values are “0”, the high and low set point 1 value judgment is not executed.
● When both the high set point 2 and low set point 2 values, as well as the high set point 1 and low set point 1 values are “0”, no judgment is executed.
• About hysteresis
When hysteresis is set in function F5, provide hysteresis between the judgment ON and OFF. The hysteresis setting value is common to the high set point 1, low set point 1, high set point 2, and low set point 2 value judgment.
Output
出出
Judgment ON
出出 ON
Hysteresis
出出出
● Conditions under the high set point 1 and 2 value judgment
Judgment OFF > ON Measurement value ● Judgment value Judgment ON > OFF Measurement value ● Judgment value - Hysteresis value
● Conditions under the low set point 1 and 2 value judgment
Judgment OFF > ON Measurement value ● Judgment value
Judgment OFF
出出 OFF
Find Quality Products Online at: sales@GlobalTestSupply.com
OFF 出出 ON 出出
OFF display value ON display value
www.GlobalTestSupply.com
Judgment ON > OFF Measurement value ● Judgment value + Hysteresis value
17
Page 18
●19.
Memory Function
• Key to be used for the memory function and display
Switches the normal display, the maximum value display, and mini mum value display. Press and hold this key for 5 seconds or more to clear the memory.
MAX and MIN lamps
-
MEM key
When the current displayed value is the maximum or minimum, the MAX or MIN lamp lights up, respectively.
• During measurement, the maximum display value (MAX value) and minimum display value (MIN value) per display update cycle is always maintained.
• Pressing the MEM key allows you to check the maximum and minimum values maintained during the measurement.
Normal measurement display
Maximum value display
MAX
1 0 0
MIN
1 0
Press the MEM key once during the normal measurement to display the maximum value.
●●●●●●●●●●●●●●●●
Press the MEM key once when the maximum value is displayed to display the minimum value.
●●●●●●●●●●●●●●●●●
(MAX lamp lights up)
(MIN lamp lights up)
1
Press the MEM key once when the minimum value is displayed to return to the normal measurement display.
Normal measurement display
1 0
• Press and hold the MEM key for 5 seconds during the normal measurement and memory display ● Clears the maintained MAX and MIN values.
1 0
Press and hold the MEM key for 5 seconds during the nor­mal measurement or when the maximum/minimum value is displayed, the MAX and MIN lamps blink for 2 seconds.
MAX MIN
●●●●●●●●●●●●●●●●●
(MIN lamp lights up)
1
After the MAX and MIN lamps blink for 2 seconds, the memory is cleared and the display returns to the normal measurement display.
Normal measurement display
1 0
• As in the gure below, the maintained values are cleared when the measurement is reset (when the mode and setting value are changed, and the power is turned ON).
Measurement reset (clears the maintained values)
電電電電
MAX value
Power ON
電電 ON
MAX
Display value
電電
MAX value
MAX
電電電
Display value
MIN 電
MIN value
MIN
MIN value
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
18
Page 19
20.
Test Mode (Function to Check if the Unit is Operating Normally)
Normal display
Mode display
Test mode
LED display test
Version display
0
Press and hold the SHIFT + MODE keys for 5 seconds or more.
- 0 1 -
Press the key to change the mode number to “99”.
- 9 9 -
Press the SET key to display the LED display test.
<<
0 0 0 0 0 0
BTM
8 8 8 8 8 8
9 9 9 9 9 9
Press the ENT key to display the software version.
v
1. 0 0
Displays the measurement value.
Displays the current set mode number.
s
The LED on the main display lights up. ”000000 > 999999” is repeatedly displayed. Also, the decimal point and display lamp light up sequentially.
8. 8 :. 8. 8 :. 8. 8.
Press the ENT key to return to the LED display test.
MAX MIN TEA PEAK BTM
<<●●
>>
Press the MEN key to display all LED digits.
Press the MEN key to return to the LED display test.
Press the MEN key to display the key test display.
Key test
21.
Error Display
0 0 0
1 0 0
0 1 0
0 0 1
Press the MEN key to display “1” in the 2nd digit place from the left, and release it to display “0”.
s
Press the key to display “1” in the 3rd digit place from the left, and release it to display “0”.
s
Press the
key to display “1” in the 4th digit
place from the left, and release it to display “0”.
Press the ENT key to return to the LED display test.
Press the SET key to return to the normal display.
Displays the measurement value.
0
When an error occurs during operation, the following error codes are displayed. Take appropriate countermeasures based on the displayed code.
Display
1
-. -. -. -. -. -.
2
3
4
5
Find Quality Products Online at: sales@GlobalTestSupply.com
E E - 1
E E - 2
E E - 3
E E - 4
Mode 1, 4: Displayed when the display is overowed (the display value exceeds the available number of display digits). Mode 2, 3: Displayed when the auto zero function is executed, or the input is less than the minimum revolution speed.
Displayed when the input pulse width is 10ms or less in mode 3 (time wide meter mode).
Displayed when the value is outside the teaching function range (when the input revolution speed is 99,999 or more). *For details, refer to page 7 (detailed teaching function).
Displayed when an internal memory error occurs. Press the ENT (MODE) key to release the error.
Displayed when the setting value F01 is less than F02, or F03 is less than F04 in the function setting mode.
www.GlobalTestSupply.com
Description Countermeasure
When the input signal becomes within the measurement range (available number of display digits), the measurement value is displayed.
Change the input pulse width to within the measurement range.
Decrease the input revolution speed and execute the teaching function.
Note that the setting values of modes, parameters, and functions are initialized.
After the error code is displayed for 2 seconds, the display returns to the previous status before EE-4 is displayed. Change the settings.
19
Page 20
22.
Parameter List
The following parameters can be set in each mode.
Parameters in mode 1 (Digital Tachometer Mode)
No. Setting item Description Input range Default value
P1 Pulse count per revolution Enter the pulse count per revolution for the rotary encoder, etc. 1 - 9999 P/r 1P/r
Setting revolution speed
P2
(detection section) Value to be displayed
P3
(with decimal point)
P4 Display cycle Sets the display update cycle 0.2/0.5/1.0/2.0/5.0/10/15/30/60 sec. 1.0
P5 Auto zero time
P6 Input lter
*1
*2
Revolution speed in the detection section 1 - 99999rpm 1000rpm
Actual value to be displayed on the panel in the above revolution speed
Sets the time from when the input pulse is gone to when the display becomes "0". Selects a minimum frequency that is larger than the maximum frequency of the input signal. 0.02kHz for the contact input
0.00001 - 999999rpm 1000rpm
0.1 - 150 sec. 6.0 sec.
10/30/100/0.02kHz 10kHz●
Parameters in mode 2 (Elapsed Timecounter Mode)
No. Setting item Description Input range Default value
P1 Pulse count per revolution Enter the pulse count per revolution for the rotary encoder, etc. 1 - 9999 P/r 1P/r
Setting revolution speed
P2
(detection section) Value to be displayed
(Hour:Minute:Second display system)
P3
Value to be displayed (second display system)
P4 Display cycle Sets the display update cycle 0.2/0.5/1.0/2.0/5.0/10/15/30/60 sec. 1 sec.
P5 Auto zero time
P6 Input lter
*1
*2
Revolution speed in the detection section 1 - 99999rpm 1000rpm
Actual value to be displayed on the panel in the above revolution speed
Sets the time from when the input pulse is gone to when the display becomes "0". Selects a minimum frequency that is larger than the maximum frequency of the input signal. 0.02kHz for the contact input
0:00:01 - 9:59:59
0:01 - 999:99
0.1 - 150 sec. 6 sec.
10/30/100/0.02kHz 10kHz●
Parameters in mode 3 (Time width meter Mode)
No. Setting item Description Input range Default value
Hour/minute/second and 1/100
P1
second display systems
P2 Measurement section
P3 Auto zero time
P4 Input lter
*1
*2
Selects the time display method.
Selects the measurement time, i.e. during input signal ON or OFF. Sets the time from when the input pulse is gone to when the display becomes "0". Selects a minimum frequency that is larger than the maximum frequency of the input signal.
(hour:minute:second display system)
0:00:00 /
0:00(1/100 second display system)
0(when OFF)/1(when ON) 1 ●ON●
0.1 - 3600 sec. 3600 sec.
10/0.02kHz 10kHz●
Parameters in mode 4 (Flowmeter Mode)
No. Setting item Description Input range Default value
Number of blades per
P1
revolution Capacity per sensor blade
P2
(cc, ℓ, etc.) P3 Scaling Unit coecient value 0.00000 - 999999 1 P4 Decimal point display Designates the decimal point position 0.00000 - 00000.0 00000.0 P5 Display cycle Sets the display update cycle 0.2/0.5/1.0/2.0/5.0/10/15/30/60 sec. 1 sec.
P6 Auto zero time
P7 Input lter
*1
*2
Sets the number of blades per revolution 1 - 99●(1 when the number is unknown) 1
Enter the capacity per blade that the sensor can read. 0.0001 - 99999 1.0
Sets the time from when the input pulse is gone to when the display becomes "0". Selects a minimum frequency that is larger than the maximum frequency of the input signal.
0.02kHz for the contact input
0.1 - 150 sec. 6 sec.
10/30/100/0.02kHz 10kHz●
10:00
(second display
system)
1/100
second display
system
*1 About the auto zero time
When the input pulse becomes 0Hz during the measurement, and the auto zero time is elapsed, the display is reset to zero.
Display value
表表表表
Auto zero time
表表
Input pulse
表表
*In mode 2 and mode 3, the zero reset display is “-.-.:-.-.:-.-.”.
Input pulse 0
表表 0
表表表表
Zero reset
0
*When a smaller value than the input pulse cycle is set as the auto zero time, the auto zero function is activated for every pulse, which disables the normal measurement.
*2 Input lter
Set the input lter with the larger value than the input signal frequency.
Example) When the 15kHz signal is input, set the 30kHz lter.
Note) When the duty of the input signal (proportion of the ON time in one cycle) is low, the signal may be attenuated and the pulse may not be received
Find Quality Products Online at: sales@GlobalTestSupply.com
normally even if you set the lter with a larger value than the input frequency. When this occurs, set the lter with an even larger value.
www.GlobalTestSupply.com
20
Page 21
23.
判判判判
Function List
The following functions can be set in each mode.
Function in mode 1 (Digital Tachometer Mode)
No. Setting item Description Input range Default value
F1 High set point 1 value F2 Low set point 1 value F3 High set point 2 value F4 Low set point 2 value
Hysteresis of the high and
F5
low set point 1 values Judgment output timer
F6
at startup
*2
F7 Minimum revolution speed Sets the revolution speed to be displayed as zero 000000 - 999999 0
Frequency of the moving
F8
average
F9 Pre-arithmetic function Promptly performs the deceleration display when the signal is lost 0(disabled)/1(enabled) 0(disabled)
*1
*1
*1
*1
Sets the high set point 1 value 000000 - 999999 0 Sets the low set point 1 value 000000 - 999999 0 Sets the high set point 2 value 000000 - 999999 0 Sets the low set point 2 value 000000 - 999999 0
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
*1
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Used when variation of the revolution speed is large and a stable display cannot be attained
0(disabled)/1(3 times)/2(10 times)
Function setting item in mode 2 (Elapsed Timecounter Mode)
No. Setting item Description Input range Default value
F1 High set point 1 value
F2 Low set point 1 value
F3 High set point 2 value
F4 Low set point 2 value
Hysteresis of the high and
F5
low set point 1 values Judgment output timer
F6
at startup
*2
*1
*1
*1
*1
*1
Sets the high set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the high set point 1 value (Second display system) 0:00 - 999:99 Sets the low set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the low set point 1 value (Second display system) 0:00 - 999:99 Sets the high set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the high set point 2 value (Second display system) 0:00 - 999:99 Sets the low set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the low set point 2 value (Second display system) 0:00 - 999:99
Sets the hysteresis of the high and low set point 1 values 0 - 99
Sets the time when the comparator judgment is output at startup 0 - 99 sec.
F7 Minimum revolution speed Sets the revolution speed to be displayed as zero 000000 - 999999
Frequency of the moving
F8
average
Used when variation of the revolution speed is large and a stable display cannot be attained
0(disabled)/1(3 times)/2(10 times)
F9 Pre-arithmetic function Promptly performs the deceleration display when the signal is lost 0(disabled)/1(enabled)
Second display system
Second display system
Second display system
Second display system
Function setting item in mode 3 (Time width meter Mode)
No. Setting item Description Input range Default value
F1 High set point 1 value
F2 Low set point 1 value
F3 High set point 2 value
F4 Low set point 2 value
Hysteresis of the high and
F5
low set point 1 values Judgment output timer
F6
at startup
*2
*1
*1
*1
*1
*1
Sets the high set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the high set point 1 value (Second display system) 0:00 - 999:99 Sets the low set point 1 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the low set point 1 value (Second display system) 0:00 - 999:99 Sets the high set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the high set point 2 value (Second display system) 0:00 - 999:99 Sets the low set point 2 value (Hour:Minute:Second display system) 0:00:00 - 9:59:59 Sets the low set point 2 value (Second display system) 0:00 - 999:99
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Second display system
Second display system
Second display system
Second display system
0(disabled)
0:00
0:00
0:00
0:00
0
0 sec.
0
0(disabled)
0(disabled)
0:00
0:00
0:00
0:00
Function in mode 4 (Flowmeter Mode)
No. Setting item Description Input range Default value
F1 High set point 1 value F2 Low set point 1 value F3 High set point 2 value F4 Low set point 2 value
Hysteresis of the high and
F5
low set point 1 values Judgment output timer
F6
F7
F8
*2
at startup Minimum ow (display value) Frequency of the moving average
F9 Pre-arithmetic function Promptly performs the deceleration display when the signal is lost 0(disabled)/1(enabled) 0(disabled)
*1
*1
*1
*1
*1
Sets the high set point 1 value 000000 - 999999 0 Sets the low set point 1 value 000000 - 999999 0 Sets the high set point 2 value 000000 - 999999 0 Sets the low set point 2 value 000000 - 999999 0
Sets the hysteresis of the high and low set point 1 values 0 - 99 0
Sets the time when the comparator judgment is output at startup 0 - 99 sec. 0 sec.
Sets the revolution speed to be displayed as zero 000000 - 999999 0
Used when variation of the revolution speed is large and a stable display cannot be attained
0(disabled)/1(3 times)/2(10 times)
0(disabled)
*1 For details about the high set point 1 and 2 values, as well as the low set point 1 and 2 values, refer to “Comparator Function” on page 17.
*2 About the judgment output timer at startup
When you set the value larger than 1 in the function item “F6 Judgment output timer at startup”, the judgment output timer function can be used. After the power is turned on, judgment starts when the judgment output timer setting time has elapsed after the start of measurement.
Judgment output timer
Judgment
判判
Find Quality Products Online at: sales@GlobalTestSupply.com
Power ON
www.GlobalTestSupply.com
判判判判
Judgment startMeasurement start
判判判判 判判 ON
21
Page 22
*3 About the minimum revolution speed (minimum ow)
When you set a value larger than 1 in the function item “F7 Minimum revolution speed (minimum ow)”, the following functions can be used.
● In mode 1 and mode 2, the display value is “0” when the measurement revolution speed becomes less than the minimum revolution speed(In mode 2, the display shows “-.-.-.-.-.-.”).●●
● In mode 4, the display value is “0” when the measurement display value becomes less than the minimum ow.
Mode
1: Judgment condition: (Input frequency/P1 setting value)*60 ● Minimum revolution speed ● Display value
Mode
2: Judgment condition: (Input frequency/P1 setting value)*60 ● Minimum revolution speed ● Display value = Mode 3: No judgment Mode 4: Judgment condition: Display value < Minimum ow ● Display value = “0”
= “0”
“-.-.-.-.-.-.”
Revolution speed
Minimum
表表表表表表
revolution speed
Display value
表表表 0
0
Display value
表表
Input pulse
表表
Measurement speed
表表表表
Time
*4 About the frequency of the moving average
When you select values other than “0” in the function item “F8 Frequency of moving average”, the frequency of the moving average function can be used. Displays the measurement value per display cycle averaged by the frequency of the moving average.
● When the moving average function is not set (F8 setting value “0”)
Measurement value ● Measurement value ● Measurement value ● Measurement value ● Measurement value ●
Display update Measurement value ●
● When the moving average function is set (F8 setting value “1”, “2”)
Displays the measurement value per display cycle averaged by the frequency of the moving average. The gure below shows the relationship between the display update in the F8 setting “0” (moving average: 3 times) and averaging.
Display update Measurement value ●
Display update Measurement value ●
Display update Measurement value ●
Display update Measurement value ●
●*The measurement values ● to ● represent the average of the measurement values per display update cycle.
Measurement value ● Measurement value ● Measurement value ●
Measurement value ● Measurement value ● Measurement value ●
Display update
Measurement value ● Measurement value ● Measurement value ●
Display update Average of the measurement
Display update Average of the measurement values ● , ● , and ●
values ● , ● , and ●
Average of the measurement values ● , ● , and ●
●*The measurement values ● to ● represent the average of the measurement values per display update cycle.
*5 About the pre-arithmetic function
When you select “1” in the function item “F9 Pre-arithmetic function”, the pre-arithmetic function can be used. Promptly performs the deceleration display when the input signal is lost.
Display value
Pre-arithmetic
表表表表表表
function display
Input pulse
表表 0
0
Find Quality Products Online at: sales@GlobalTestSupply.com
When the auto zero function is activated, the display becomes “0” (“-.-.-.-.-.-. in mode 2 and mode 3).
www.GlobalTestSupply.com
Display value
表表
Input pulse
表表
Time
22
Page 23
24. Option -FVT/-FVC
When equipped with -FVT/-FVC option, analog signal output (voltage/current) is available according to the displayed data.
1. Specications for -FVT/-FVC Option
Model -FVT/-FVC
Current output 4 ● 20mA
Output
Load
Connector
●FVC)*
*Cables are to be connected by the user.
2. Connection for -FVT/-FVC Option
Voltage output
Output current below 500Ω Output voltage above 1kΩ Meter :PCS-E36LMD Attachment : ●Plug● PCS-E36SF●●Cover● PCS-E36LA (Both manufactured by HONDA TSUSHIN KOGYO CO., LTD.)
0 ● 10V 1 ● 5V 0 ● 1V
Select one of these four output options
●●
-FVT Option (Terminal block)
C+ C- NC NC
V10+ V10- V5+ V5- V1+
11 12 13 14 15 16 17 18 19 20
+ + + +- - - -
4 ● 20mA 0 ● 10V 1●5V 0●1V
*Select one out of these output options
V1-
● ●
-FVC Option (Connector)
Code Pin number Code
C+
4 ● 20mA+
NC 3 21 NC NC 4 22 NC NC 5 23 NC NC 6 24 NC NC 7 25 NC NC 8 26 NC
V10+
0 ● 10V+
NC 11 29 NC NC 12 30 NC
V5+
1 ● 5V+
NC 15 33 NC NC 16 34 NC
V1+
0 ● 1V+
*Select one out of these output options *1 and 2 are interconnected to each other. 9-10, 13-14, 17-18, 19-20, 27-28, 31-32, 35-36 are also interconnected respectively.
1 19 2 20
9 27
10 28
13 31 14 32
17 35 18 36
4 ● 20mA-
0 ● 10V-
1 ● 5V-
0 ● 1V-
Connector numbering (as the plug is viewed from wire connection side)
C-
V10-
V5-
V1-
1 2 3 4 ● 17 18
19 20 35 36
● The connector next to “1”marking is #2 terminal.
● Suggested wiring order is 1● 3● 5 ● 20● 22● 24 ● for the ease of nding the right numbering.
● Please use extra caution in soldering, as the clearance is small.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
35 36
23
Page 24
3. -FVT/-FVC Option Setting
When equipped with -FVT/-FVC option, following setting options are enabled from function setting feature.
No. Setting item Description Input range Default
Maximum analog
F11
Minimum analog
F12
Analog signal
F13
output timing
Analog signal
F14
output oset
signal displayed
value
signal displayed
value
Set the displayed value which corresponds to the maximum value of each analog signal.
● 4 ● 20 mA
● 0 ● 10 V
● 1 ●●5 V
● 0 ●●1 V
Set the displayed value which forces the output of the minimum value of each analog signal.
● 4 ● 20 mA
● 0 ● 10 V
● 1 ●●5 V
● 0 ●●1 V
Maximum speed when the period is zero.(10ms). When the period is 1, analog signal output is renewed at each display renewal cycle.
● when the frequency is above 200Hz
Set the output oset value as the percentage of the output range for each analog signal.
● 4 ● 20 mA
● 0 ● 10 V
● 1 ●●5 V
● 0 ●●1 V
● Maximum value●20 mA
● Maximum value●10 V
● Maximum value●●5 V
● Maximum value●●1 V
● Minimum value●4 mA
● Minimum value●0 V
● Minimum value●1 V
● Minimum value●0 V
● Output range●16mA
● Output range●10V
● Output range●●4V
● Output range●●1V
Mode 1 ●000000●●●999999● 1000
Hour:Minute:Second ●_0:00:00●●●_9:59:59● Displayed in Seconds
Mode 2
Hour:Minute:Second ●_0:00:00●●●_0:59:59● Displayed in Seconds
Mode 3
Mode 4 ●000000●●●999999● 1000
Mode 1 ●000000●●●999999● 1000
Hour:Minute:Second ●_0:00:00●●●_9:59:59●
Mode 2
Displayed in Seconds ●_000:00●●●_999:99●
Hour:Minute:Second ●_0:00:00●●●_0:59:59●
Mode 3
Displayed in Seconds ●_000:00●●●_999:99●
Mode 4 ●000000●●●999999● 1000
● _ 0_ ● (Maximum speed )/
● _ 1_ ● (In sync with display renewal)
●__-100.0●●●__100.0●● 0●
10:00Displayed in Seconds ●_000:00●●●_999:99●
10:00Displayed in Seconds ●_000:00●●●_999:99●
Displayed in Seconds
10:00
Displayed in Seconds
10:00
0
(Maximum speed)
● When oset is OFF ●o set F14 = 0●●
● When oset is ON ●o set F14 > 0●
● When oset is ON ●o set F14 < 0●
displayed value ● F11set value Maximum value ●20mA● 10V● 5V● 1V●
F12 ● displayed value ● F11 (Maximum value/ F11)× Measured value for analog output displayed value ● F12 set value Minimum value ●4mA● 0V● 1V● 0V●
displayed value ● F11´● Maximum value ●20mA● 10V● 5V● 1V●
F12 ● displayed value ● F11´
displayed value ● F12 set value
displayed value ● F11´● Maximum value ●20mA● 10V● 5V● 1V●
F12 ● displayed value ● F11´
displayed value ● F12 set value Minimum value ●4mA● 0V● 1V● 0V ) - (Oset value)
Condition Analog output value
Condition Analog output value
(Maximum value/ F11)×Measured value for analog output+
(Oset value) Minimum value ●4mA● 0V● 1V● 0V) + (MAX ●● F14(● ))
● F11' is the displayed value when the analog output value, including F14(%) of the output range, is equal to the MAX value.
Example) Assuming the output is 10V, F11= 200, F14 = 10(%), then F11'= 180,
therefore the output is 10V when the displayed value is above 180.
Condition Analog output value
(Maximum value/ F11)×Measured value for analog output-
(Oset value)
● The minimum value for MIN-F14(%) in the diagram above is 0mA, in the case of current output (4 ● 20mA)
Example) Assuming the output is 10V, F11= 200, F14 = -10(%), then F11'= 220,
therefore the output is 10V when the displayed value is above 220.
Find Quality Products Online at: sales@GlobalTestSupply.com
The minimum value for MIN-F14(%) in the diagram above is 0mA, in the case of current output (4 20mA)
www.GlobalTestSupply.com
24
Page 25
25. Option -BCD
When equipped with -BCD option, Binary Coded Decimal output is possible
1. Specications for -BCD Option
NPN Open collector output Output capacity 30VDC●20mA
Open collector input
Data output 6 digits BCD code Decimal point output DP1 ● 4 ●1 ● 4 digits after decimal point●
Control output
Control input
Connector
Positive-logic-negative-logic available for BCD and decimal point output (opted at Function 10)
Model -BCD
Open collector (NPN) input
LO input
HI input Leakage current : maximum 0.1mA
PLUS When output data is positive, PLUS turns to LO DT OUT When DATA OUT is HI, output signal is set OVR When displayed value overows, OVR turns to LO HOLD While HOLD is LO, output data does not renew ENABLE While ENABLE is LO, output has high impedance
Meter :PCS-E36LMD/ Accessory side: ●Plug● PCS-E36FS●●Cover● PCS-E36LA (Both manufactured by HONDA TSUSHIN KOGYO CO., LTD.)
Load capacity : minimum 5mA 0 ● 1.5V
● Cable connection is to be done by users.
● Timing chart
Measured data
HOLD signal
BCD output
DT OUT signal
ENABLE signal
“H”
“L”
“H”
“L”
“H”
“L”
“00000” data 1
hold
“00000” “00000” data 1 data 2 data
data out
renewal
about 6 msec
data 2
about 100 msec
data 3 data 4
3
high impedance
data
4
data
4
2. Connection for -BCD Option
in/out Code ●●Pin number Code in/out
1 1 19 1
×10
0
2 2 20 2 4 3 21 4
×10
3
8 4 22 8 1 5 23 1 2 6 24 2 4 7 25 4 8 8 26 8 1 9 27 1 2 10 28 2 4 11 29 4
×10
×10
4
5
Out put
×10
×10
1
2
8 12 30 8
PLUS 13 31 DP1
DT OUT 14 32 DP2
OVR 15 33 DP3
Input
HOLD 16 34 DP4
ENABLE 17 35 GND
GND 18 36 GND
Connector numbering (as the plug is viewed from wire connection side)
1 2 3 4 ●17● 18
● The connector next to “1”marking is #2 terminal.
● Suggested wiring order is 1● 3● 5 ● 20● 22● 24 ● for the ease of nding the right numbering.
Please use extra caution in soldering,
as the clearance is small.
Out put
HOLD input circuit
Tachometer
Attachment example
Output circuit
Tachometer
Attachment example
OUTPUT
ENABLE input circuit
Tachometer
Attachment example
19 20 35 36
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
25
Page 26
3. -BCD Option Setting
When equipped with -BCD option, following setting options are enabled from function setting feature.
No. Setting item Description Input range Default
F10 BCD output logic
● F10 sets positive/negative logic for BCD output, Decimal point output, PLUS, OVR.
● Logic for DT OUT, HOLD, ENABLE signals cannot be set.
Set as 0 for negative logic, set as 1 for positive logic
(decimal point output)
● _ 0_ ● (Negative logic)/
● _ 1_ ● (Positive logic)
● BCD output display specications
Negative logic
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
26
Page 27
26. Option -TRT
When equipped with -TRT option, comparison result output is possible.(LL,L,GO,H,HH,ZERO)
1. -TRT option specications
Model -TR T Output capacity 30VDC●20mA Rsidual voltage Less than 1.5V
Measured value ● Low set point 2 value LL signal is ON
Measured value ● Low set point 1 value L signal is ON
Output signal
Output is insulated from internal circuitry
Low set point 1 value ● Measured value ● High set point 1 value GO signal is ON High set point 1 value ● Measured value H signal is ON High set point 2 value ● Measured value HH signal is ON Measured value = 0 ZERO signal is ON
Negative logic open collector output
2. -TRT option connection
COM L L HH NCCOM Z ZERO
LL
GO
COM H H
11 12 13 14 15 16 17 18 19 20
common for
27. Option -CPT
L
signal
L, LL
Power (+12V ● +24V) Power (+12V ● +24V ) Power (+12V ● +24V )
When equipped with -CPT option, comparison result output is possible.(L,GO,H)
1. -CPT option specications
Output contact 1C
Rated load
Output signal
common for
GO, ZERO
- --
ZERO signal
LL
signal
Load Load LoadLoad Load
+ + +
Model -CPT
Resistance load
Induction load cosφ =0.4
Measured value ● Low set point 1 value L signal is ON
Low set point 1 value ● Measured value ● High set point 1 value GO signal is ON
High set point 1 value ● Measured value H signal is ON
GO
signal
common for
H
signal
HH, H
Load
250VAC 5A 100K operations
30VDC 5A 100K operations
250VAC 2.5A 100K operations
30VDC 2.5A 100K operations
HH
signal
no connect
Tachometer
Attachment example
●2.●
-CPT option connection
Output circuit
GO
L COM L H NCGO
COM GO
H COM HL
11 12 13 14 15 16 17 18 19 20
B contact
for L
common
for L
A contact
for L
B contact
for GO
common
for GO
A contact
for GO
B contact
for H
common
for H
A contact
for H
no connect
Tachometer
Load
Load
Attachment example
or
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
27
Page 28
28. DT
-501X / DT-501F series model list
This operation manual is applicable to following models.
Model
Input spec AC power type DC power type
Standard
input
Dierential
input
DT-501XA-FVT DT-501XD-FVT DT-501XA-FVT-BCD DT-501XD-FVT-BCD -BCD BCDoutput DT-501XA-TRT DT-501XD-TRT DT-501XA-TRT-FVC DT-501XD-TRT-FVC -FVC Voltage output DT-501XA-TRT-BCD DT-501XD-TRT-BCD -BCD BCDoutput DT-501XA-CPT DT-501XD-CPT DT-501XA-CPT-FVC DT-501XD-CPT-FVC -FVC Voltage output DT-501XA-CPT-BCD DT-501XD-CPT-BCD -BCD BCD output DT-501XA-FVC DT-501XD-FVC - -FVC Voltage output DT-501XA-BCD DT-501XD-BCD - -BCD BCD output DT-501FA-FV T DT-501FA-FV T-BCD -BCD BCD output DT-501FA-TRT DT-501FA-TRT-FVC -FVC Voltage output DT-501FA-TRT-BCD -BCD BCD output DT-501FA-CPT DT-501FA-CPT-FVC -FVC Voltage output DT-501FA-CPT-BCD -BCD BCD output DT-501FA-FVC - -FVC Voltage output DT-501FA-BCD - -BCD BCD output
-
-FVT Voltage output
-TRT Transistor output
-CPT Relay output
-FVT Voltage output
-TRT Transistor output
-CPT Relay output
Option 1 Option 2
*For the detail of standard input and dierential input, please refer to following pages.● Standard input : P5 Input Specications Dierential input : P6 Input Specications
-
-
-
-
-
-
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com