SHIMPO Instruments DT-2100-NIST User manual

Page 1
● Enables measurement of a rotating body at a distance using integral laser beam
Handheld Digital Tachometer with Dual Non-Contact, Contact Operation
DT-2100
Non-contact type
● Maximum of 1000 data point storage
● The organic light emitting display (OLED) allows enhanced viewing, even in dark locations
● Selectable meter or graphic display modes
● High accuracy measurement capability of +/-0.006%
● PC communication through USB with available software
99265A
Instruction Manual
Thank you for purchasing Nidec-Shimpo's Handheld High-Performance Digital Tachometer DT-2100. For instructions on how to use this product properly and optimally for a long period of time, please be sure to read this manual thoroughly before use.
Safety Requirements
Be sure to observe
Non-contact type digital tachometers are used to perform measurement by applying the supplied reflection tape to the rotating body of the object to be measured, and projecting the laser beam (red) to the tape. In addition, they can be used for contact type measurement by attaching the supplied contact adapter.
■ Inspection result certicate
We skip issuance of the inspection result certicate for this product. However, at the time of factory shipment total shipment inspection has been completed, and we have strictly conrmed that there is no error in measurement accuracy and operation.Your understanding is appreciated.
Before operation, maintenance and inspection, please carefully read this instruction manual and follow it for proper use.
After carefully reading this manual, be sure to store it in a safe and convenient place for easy reference.
Before operation, maintenance and inspection, please carefully read this instruction man­ual and follow it for proper use. Start using only after you have read about the equipment’ s functions, safety information and precautions.
This instruction manual provides three grades of safety warnings: "Danger", "Warning" and "Caution". Each of them is an important description related to safety. Be sure to fol low them.
This indicates the possibility of re, severe injury,
Danger
Warning
Caution
● Limited Warranty
・We are not responsible for damages resulting from negligence through failure to follow
the instructions set out in this manual.
・We are not responsible for damages resulting from earthquake and/or fire unrelated to
us, actions by third parties, or any other accidents, intentional or through customer neg ligence, as well as from accidents caused by misuse or improper use under abnormal conditions.
・For information regarding assurance provisions, please read the attached warranty certi
cate.
and even death if a user disregards the instruction and operates the unit improperly.
This indicates the possibility of severe injuries if a user operates the unit improperly.
This indicates the possibility of minor injury or damage if a user operates the unit improperly.
Warning
・Do not look into the laser beam
-
・Do not point the laser beam at people ・Do not allow usage by children
Do not contact the laser beam irradiation port with a rotating body
If the unit comes into contact with a rotating body, damage may result to the laser.
Do not measure using wet or oily hands, or with loosely tted clothing.
If hands slip during measurement, fingers or part of the hand may get caught in the rotating equipment.
Caution
Be careful around installed reective
-
-
tape.
The reection tape may come loose during high-speed rotation.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
INDEX
1. Part Names and Functions 2
2. Measurement Method 4
3. Measurement Display Description 6
4. Graph Display 8
5. Memory Function 9 5-1. Memory mode details 9 5-2.
6. Various Settings 11 6-1. Contents selection 11 6-2. Memory data display 11 6-3. User settings 15 6-4. System settings 22
7. Battery Replacement 34
8. Dimensional Drawing 34
9. Specications 35
Memory registration in the memory group
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1
Page 3
1. Part Names and Functions
(4) Power/Measurement key
(6) Connector
(5) USB connector
(1) Laser beam irradiation port
(7) Mounting Adapter
(2) Display screen
(3) Operation keys
Parts name Functions
(1) Laser beam
irradiation port
(2) Display screen
Operation keys 1 to 4
(3) (4) Power/
Measurement key
(5) USB connector
(6) Connector (7) Mounting Adapter
Battery Compartment
(8)
Apply the reection tape to the rotating body of the object to be measured. Project the laser beam (red) to the tape and measure the rotation speed.
Displays the measurement value. Used to change setting or store memory data. Press this key to turn the power ON. The power will be turned OFF when no operation has been
performed for a certain period of time. (Auto o󱐯 function) Power can be supplied through the USB connector.
Also, connecting the unit to a PC enables communications using using the available software. (Sold separately)
For internal production usage only Allows unit to be mounted with its x/x ? threaded connection
on tripod Uses two AA dry cell batteries
(8) Battery Compartment
■ Operation keys and power/measurement key details
Parts name Mode Function Operation key 1
(1) Operation key 1
(4) Operation key 4
(5) Power/Measurement key
Find Quality Products Online at: sales@GlobalTestSupply.com
(2) Operation key 2
(3) Operation key 3
www.GlobalTestSupply.com
(1)
Operation key 2
(2)
Operation key 3
(3)
Operation key 4
(4)
Power/
(5)
Measurement key
2
Measurement mode (measurement
standby, during measurement) ・Set mode Used to select the item. ・Memory data display mode Used to select the item. ・
Measurement mode (measurement
standby, during measurement) ・Set mode Used to select the item. ・Memory data display mode Used to select the item. ・
Measurement mode (measurement
standby, during measurement) ・Set mode Used as the enter key. ・Memory data display mode Used as the enter key. ・
Measurement mode (measurement
standby, during measurement) ・Set mode Used as the cancel key. ・Memory data display mode Used as the cancel key. ・When the power is OFF Used to turn the power ON. ・Measurement mode Used to perform measurement.
Used to store the memory data.
Used to display memory data, and congure various settings.
Used to switch the measurement display.
Used to return the measurement display to the normal display.
Page 4
■ Display screen details
(2) Rotation detection display
(3) Upper/lower limit judgment display
Key name
(1) Main display
Rotation
(2)
detection display
(4) Unit display
(5) Description display (6) Battery display
(1) Main display
Displays the measurement value. Blinks during measurement. "A" is displayed when the contact adapter has been
attached.
With contact Without contact
(3)
Upper/lower limit judgment display
(4) Unit display
Description display
(5) (6) Battery display
*For screen protection, the display becomes dark when no operation has been performed for 10 seconds. Also, the display is turned OFF when no operation has been performed for 1 minute.
Displays judgment results of the display value according to the comparator setting.
Displays the current set unit. Displays description for the measurement value currently indicated. Displays the remaining battery level.
Blinks when the remaining battery level becomes close to 0. Also, the following mark is displayed when the power is supplied through the USB cable.
  When the USB cable is connected
■ OVER display
When the measured value exceeds the display range, " Also, when it exceeds the measurement range, "OVER" is indicated in the unit display, and it blinks.
- - - - - -
" is displayed.
When the upper When the lower
limit is reached limit is reached
● Display range OVER
* When "OVER" is indicated, the measurement accuracy
is not guaranteed.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3
Measurement range OVER
Page 5
2. Measurement Method
■ Non-contact measurement
Apply the re ection tape to
1
the object to be measured
Re ection tape
Apply the supplied reflection tape to the rotating body of the object to be measured.
● Do not apply to a rotating body smaller than the re ection tape.
● If the rotating body is glossy, apply black tape, or paint it black and apply the re ection tape.
● Before applying the reflection tape, wipe off water/oil on the attachment surface of the rotating body, and apply the tape without any irregularities.
Point the laser beam irradiation
2
port to the object to be measured
Point the laser beam irradiation port to the re ection tape applied to the rotating body.
● Keep a dist ance of 50 to 4000 m m (2 in. to 13 ft.) between the reflection tape and irradiation port.
Press the power/measurement key
3
to start measurement
Power/Measurement key
Laser beam
Press the power/measurement key, and start measurement.
● When the rotating body is static, the value "0" is displayed.
● When the rotating body rotates, the actual measurement data is displayed every display update cycle.
Power OFF when no operation has been performed for a certain period
4
of time (Auto o󱐯 function)
Left unoperated
Final measurement value held
After measurement ends, the power is turned OFF when no operation has been performed for a certain period of time.
● The period of time until the power goes OFF can be set in "Auto o󱐯 " in the system settings.
● When performing communication using the optional USB cable, the auto o󱐯 function is disabled.
* When the speed and distance units are selected in the unit setting, be sure to select "3. SET_SYSTEM" in contents selection, and set the
distance value per pulse in the "9. DIST/PULS" setting.
Power OFF
■ Precautions on measurement and handling
Never look into the laser beam.
Failure to follow this could result in injury to the eyes.
Do not point the laser beam at people.
Do not allow usage by children.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4
Page 6
■ Contact measurement
Attaching the optional contact adapter enables measurement in the contact mode.
Attach the contact
1
adapter
Put the contact on the
4
object to be measured
Select the appropriate
2
contact tting and attach.
Contact (conical)
Measurement shaft pin
For rotation speed measurement, attach the conical or funnel contact.
● When the object to be measured has a concave shaft, attach the conical contact.
● When the object to be measured has a conical shaft, attach the funnel contact.
* When attaching the contact, securely fit
the measurement shaft pin into the contact groove.
Press the power/measurement
5
key, and start measurement
Power/Measurement key
In the "5. CONTACT" setting of "3. SET_
3
SYSTEM" in contents selection, change to "1. Contact" or "4. Contact (Low)"*.
* Select "4. Contact (Low)" when using the optional
contact adapter for low speed.
Carefully put the contact on the center of the rotating body of the object to be measured.
Press the power/measurement key, and start measurement.
■ When using the speed measurement wheel
Attach the speed measurement
1 3
wheel to the measurement shaft
Speed measurement wheel
Put the speed
2
measurement wheel on the object to be
Measurement shaft pin
Note) When attaching the wheel, securely t the measurement shaft pin into the wheel groove.
*The auto o󱐯 function works similarly as in the case of non-contact measurement. * When performing the non-contact measurement again after contact measurement ends, be sure to select "3. SET_SYSTEM"
in contents selection, and set to "1. Contact" in the "5. CONTACT" setting.
* When using the speed measurement wheel, in the "8. DIAMETER" setting of "3. SET_SYSTEM" in contents selection, set
the diameter value of the peripheral speed foil. When using the supplied speed measurement wheel (circumference of 6 inches), use it with the default value (diameter value of 48.5104 mm).
measured
Carefully put the speed measurement wheel on the object to be measured (belt part, etc.).
Press the power/measurement key, and start measurement
Refer to the above (rotation speed measurement)
■ Precautions on measurement and handling
Before measurement, securely t the contact adapter to the contact. Also, during measurement, carefully put the contact on the rotating body of the object to be measured.
When attaching the speed measurement wheel for measurement, align it along the object to be measured, and do not push the wheel with excessive force. Also, for safety measurement, use with a speed of 300m/min (500cm/s, 18000m/h, ~ 18km/h, 197inch/s) or less.
When measuring a high-speed rotating body using the conical or funnel contact over a long period of time, note that the convex and concave
Find Quality Products Online at: sales@GlobalTestSupply.com
portions of the contact may produce high heat.
www.GlobalTestSupply.com
5
Page 7
Rotation speed
3. Measurement Display Description
Press the operation key 3 in the measurement mode. The measurement value description to be displayed is switched. The types of the measurement value to be displayed are as follows:
Display Description
Normal Description Displays the measurement value corresponding to the selected unit.
Display update Updates every display update cycle. Maximum value (Inactive during distance and manual measurement) Minimum value (Inactive during distance and manual measurement) Average value (Inactive during distance and manual measurement) Acceleration rate (Inactive during distance and manual measurement) Prescale Description
Prescale maximum value (Inactive during distance and manual measurement)
Prescale minimum value (Inactive during distance and manual measurement)
Prescale average value (Inactive during distance and manual measurement)
Prescale acceleration rate (Inactive during distance and manual measurement)
Description Displays the maximum value from measurement start to end. This value is reset
when the next measurement starts. Display update Updates every display update cycle. Description Displays the minimum value from measurement start to end.
the next measurement starts, or the auto zero function is performed. Display update Updates every display update cycle. Description Displays the average value within the "GRAPH_TIME" specied in the system
settings. Display update Updates every "GRAPH_TIME". Description Displays the acceleration rate within the "GRAPH_TIME" specied in the system
settings. Display update Updates every "GRAPH_TIME".
Displays the measurement value when "PRESCALE" is enabled. This feature is selected
in the system settings. When "PRESCALE" is not specied, this is not displayed. Display update Updates every display update cycle. Description Displays the maximum value among the measurement values when "PRESCALE"
is enabled. This value is reset when the next measurement starts. Display update Updates every display update cycle. Description Displays the minimum value among the measurement values when "PRESCALE"
is enabled. This value is reset when the next measurement starts. Display update Updates every display update cycle. Description Displays the average value within the "GRAPH_TIME" specied in the system
settings, among the measurement values when "PRESCALE" is enabled. Display update Updates every "GRAPH_TIME". Description Displays the acceleration rate within the "GRAPH_TIME" specied in the system
settings, among the measurement values when "PRESCALE" is enabled. Display update Updates every "GRAPH_TIME".
This value is reset when
■ About calculation of the maximum value (MAX) and minimum value (MIN)
・ As shown in the gure below, the instrument sets the measurement value when measurement starts with initial values of MAX0 and MIN0.
It then compares them to new measurement values every sampling cycle. New updated maximum and minimum values are created and displayed if they exceed (for MAX0) or are below (for MIN0).   (Updates in the order of MAX
・The maximum and minimum values are reset when measurement starts.
Sampling cycle*1
Measurement start
→ MAX1 → MAX2)
0
MAX0
MIN0
MIN
MAX
1
MAX
1
2
MAX2
MAX
MAX0
1
MIN0
MIN2
Time
Measurement end
Measurement start
Measurement end
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6
Page 8
■ About calculation of the average value (AVE)
Rotation speed
The average value is displayed by averaging the measurement values in the corresponding cycle every average value update cycle.
4
1
Sampling cycle*1
Average value update cycle*2
Average calculation of the
3
2
measurement display values 1 to 6
6
5
7
8
9
11
10
Average calculation of the measurement display values 7 to 12
■ About calculation of the acceleration rate (ACC)
The acceleration rate is calculated based on v0 and v1 in the corresponding cycle every acceleration rate update cycle.
Rotation speed
Gradient = Acceleration rate
v
0
Acceleration rate calculation time
Sampling cycle*1
Acceleration rate update cycle*2
v
1
MIN
MAX
12
Time
Time
Acceleration rate =(v1 - v0)/Acceleration rate calculation time
*1 The sampling cycle is "DISP_CYCLE" in the system settings. *2 The average value update cycle and acceleration rate update cycle are "3-2 Graph time" in the system settings. *The acceleration rate is displayed up to 5 digits, and when the speed is decreased, it is displayed as a negative value. *Minus value of the acceleration value cannot be displayed in the graph.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7
Page 9
4. Graph Display
Using the "4. GRAPH_DISP" setting of "2. SET_USER" in contents selection enables the graph to be indicated on the display.
(1) Graph_1 (Line graph)
(1) Various status display
(2) Line graph display
4) Graph time display (3) Measurement value display
(2) Graph_2 (Gauge Mode)
(1) Various status display
(2) Needle graph display
(5) Graph minimum
value display
(4) Graph maximum
value display
(3) Measurement
value display
Parts name Operation
Various status
(1)
display
(2) Line graph
display
(3) Measurement
value display Graph time
(4)
display
Update cycle: Every display update cycle
*
Parts name Operation
Various status
(1)
display
(2) Gauge Display
(3) Measurement
value display Graph maximum
(4)
value display Graph minimum
(5)
value display
Update cycle: Every display update cycle
*
Same as in the normal display
Displays the line graph for the measurement value. Depending on selection one of the below will equal the graph values:  Maximum value on the vertical axis:   Graph maximum value setting  Minimum value on the vertical axis:   Graph minimum value setting  Horizontal axis: Graph time setting value   Graph ow: Graph ow setting value
Displays the current measurement value  Can switch among MAX, MIN, AVE, and ACC
Displays the graph time setting value
Same as in the normal display
Displays the needle graph for the measurement value. Depending on selection one of the below will equal the gauge values:  Graph maximum value:   Graph maximum value setting  Graph minimum value:   Graph minimum value setting
Displays the current measurement value  Can switch among MAX, MIN, AVE, and ACC
Displays the graph maximum value setting
Displays the graph minimum value setting
(3) Graph_3 (Bar graph)
(1) Various status display
(2) Bar graph display
(5) Graph minimum
value display
(4) Graph maximum
value display
Find Quality Products Online at: sales@GlobalTestSupply.com
(3) Measurement
value display
www.GlobalTestSupply.com
Parts name Operation
Various status
(1)
display
(2) Bar graph display
(3) Measurement
value display Graph maximum
(4)
value display Graph minimum
(5)
value display
Update cycle: Every display update cycle
*
Same as in the normal display
Displays the bar graph for the measurement value. Depending on selection one of the below will equal the gauge values:  Graph maximum value:   Graph maximum value setting  Graph minimum value:   Graph minimum value setting
Displays the current measurement value  Can switch among MAX, MIN, AVE, and ACC
Displays the graph maximum value setting
Displays the graph minimum value setting
8
Page 10
5. Memory Function
The measurement value can be registered and stored in the memory during measurement. The stored data can be displayed in contents selection "1. DATA". Registration contents and methods depend on the memory mode. Set the memory mode in the "2. MEMORY_MODE" of "2. SET_USER" in contents selection.
5-1 Memory mode details
(1) Continuous memory mode
・In the continuous memory mode, the measurement value continues to be registered in the memory every display update cycle. ・The number of memory sets that can be registered in the continuous memory is up to 1000. ・For the continuous memory, the value description to be registered in the memory depends on the measurement display description.
・Normal display
・MAX display
・MIN display
・AVE display
・ACC display
・PRE display
・P.MAX display
・P.MIN display
・P.AVE display
・P.ACC display
Registers the measurement value every display update cycle
Registers the measurement value every display update cycle after prescale calculation
● Memory registration operation
・Press the Operation key 1 once during measurement to start memory registration. Press it again to end memory registration. ・When turning measurement OFF, or switching to the contents selection mode during memory registration, memory registration ends at
that moment.
・The unit and "Gr ○ _ □□□□ " are alternately displayed in the unit display part during memory registration.
 ( ○ shows the number of memory groups, and □ shows the number of memory points currently registered: 0001 to 1000)
Displayed alternately every second during memory registration
Operation key 1:  Memory registration start/end
*The continuous memory mode can be used only when one of the units from "rev/min" to "miles/h" is selected in the unit setting.
(2) Each memory mode
・In the each memory mode, only one set of the measurement value currently displayed is registered in the memory. ・The number of data points that can be registered in the each memory is up to 100.
● Memory registration operation
・Press the operation key 1 once during measurement to register the display value at that moment in the memory. ・"Gr ○ _ □□□□ " is displayed for 1 second in the unit display part during memory registration.
 ( ○ shows the number of memory groups, and □ shows the number of memory sets currently registered: 001 to 100)
Displayed for 1 second during memory registration
Operation key 1:
Find Quality Products Online at: sales@GlobalTestSupply.com
 Performs memory registration
www.GlobalTestSupply.com
9
Page 11
(3) Statistics memory mode
・In the statistics memory mode, only the statistics data for the measurement value every display update cycle from memory registration
start to end is registered in the memory.
・The number of data points that can be registered in the statistics memory is up to 100.
● Memory registration operation
・Press the operation key 1 once during measurement to start memory registration, and press it again to end memory registration. ・ When turning measurement OFF, or switching to the contents selection mode during memory registration, memory registration ends at
that moment.
・The unit and "Gr ○ _ □□□□ " are alternately displayed in the unit display part during memory registration.
 ( ○ shows the number of memory groups, and □ shows the number of memory sets currently registered: 001 to 100)
Displayed alternately every second during memory registration
Operation key 1:  Memory registration start/end
 * In the statistics memory mode, the statistics data in a single registration is calculated from the measurement value for up to 100 times
of display update. When the display update frequency exceeds 100 in a single registration, memory registration automatically ends.
 *The statistics memory mode can be used only when one of the units from "rev/min" to "miles/h" is selected in the unit setting.
5-2. Memory registration in the memory group
・Start memory registration from the start memory group (that has been set in "3. START_MEM" of "2. SET_USER" in contents selection). ・When the number of memory groups reaches the number of memory sets in each group (that has been set in "4. MEM_GROUP" of "3.
SET_SYSTEM" in contents selection), the memory group switches to the next one.
・When the number of memory sets exceeds the total number of memory sets for all memory groups, "FULL" is displayed in the unit
display part. In this case, additional memory cannot be registered.
Example: When the start memory group is set to memory group 2
(1) Start memory registration
from the start memory group
Group 1 Group 2 Group 3 Group 5Group 4
(2) Go to the next memory group
when the number of memory sets exceeds the specied number
(5) Displays "FULL" when the number of memory sets exceeds the specified number for all
memory groups, and ends memory registration ("FULL" is displayed until the operation key 1 is pressed) After that, even if you press the operation key 1 and try to start memory registration, "FULL" is displayed for 1 second, and registration is disabled.
(3) Go to the next memory
group when the number of memory sets is 0
(4) Return to memory group 1
when the number of memory sets exceeds the specied number in group 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
10
Page 12
6. Various Settings
[SET_CONTENTS]
DATA
SET_USER SET_SYSTEM
[選択してください
]
メモリ表示
ユーザー設定 システム設定
[選択してください]
メモリ表示
ユーザー設定 システム設定
[連続メモリ]
1.グループ_1
2.グループ_2 3.グループ_3
[単独メモリ グループ 1]
UNIT : rev/min
MAX : 100.00 MIN : 10.12
6-1. Contents selection
Press the operation key 2 in the measurement mode to switch the display to the contents selection mode.
English Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Item
English Japanese
1 DATA メモリ表示 2 SET_USER ユーザー設定 Go to the user settings 3 SET_SYSTEM システム設定 Go to the system settings
Description
Go to the memory display
6-2. Memory data display
The memory data registered for each group can be checked in contents selection "1. DATA".
Operation key 3 Go to the selected item
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
Return to the measurement mode
(1) Operation ow
Display the memory data according to the following operation.
Measurement mode
During measurement
[SET_CONTENTS]
1. DATA
Select [1. DATA], and turn the operation key 3 ON
During measurement standby
Turn the operation key 2 ON
2. SET_USER
3. SET_SYSTEM
Contents selection
Turn the operation key 3 ON
Use the operation keys 1 and 2 to select the memory group to be displayed
Memory data display
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
11
Page 13
(2) Memory data display/details
● Continuous memory display
Select "1. DATA" in contents selection to display the memory group selection screen.
(1) Memory group selection
Select the memory group you want to display.
English
Item
English Japanese
1. Group_1 グループ _1
2. Group_2 グループ _2 Displays memory group 2
3. Group_3 グループ _3 Displays memory group 3
4. Group_4 グループ _4 Displays memory group 4
5. Group_5 グループ _5 Displays memory group 5
Displays memory group 1
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3
Operation key 4 Return to contents selection
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously Go to memory description display
(2) Memory data display
Select the memory group in (1), and press the operation key 3 to display the memory data for the selected memory group. Use the operation keys 1 and 2 to change the display line.
English
Display item Description UNIT: Unit MAX:
MIN: Minimum value AVE: Average SD: Standard deviation
1: ~  1000:
CLEAR Go to memory clear
Maximum value
1 blank line Displays the memory registration value for 1 to 1000 sets (max.)
Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Enabled when CLEAR is
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
selected Go to memory clear Return to memory group selection
(3) Memory clear
While the memory data is displayed, the "CLEAR" column is displayed in the last line of the data. Select the "CLEAR" column, and press the operation key 3 to display the memory clear screen below. Registration description in the memory group currently displayed can be cleared as necessary.
English
Item
English Japanese
1 No いいえ
Find Quality Products Online at: sales@GlobalTestSupply.com
2 Yes はい Clears the continuous memory description
Does not clear, and returns to the continuous memory display
www.GlobalTestSupply.com
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Execute clear/Return to the
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
continuous memory display without executing clear Return to the continuous memory display
12
Page 14
● Each memory display
Select "1. DATA" in contents selection to display the memory group selection screen.
(1) Memory group selection
Select the memory group you want to display.
Operation key Operation
Item
English Japanese
1 Group_1 グループ _1 2 Group_2 グループ _2 Displays memory group 2 3 Group_3 グループ _3 Displays memory group 3 4 Group_4 グループ _4 Displays memory group 4 5 Group_5 グループ _5 Displays memory group 5
Displays memory group 1
Description
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to the memory data display Operation key 4 Return to contents selection
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
(2) Memory data display
Select the memory group in (1), and press the operation key 3 to display the memory data for the selected memory group. Use the operation keys 1 and 2 to change the display line.
English
Display item Description UNIT: Unit MAX:
MIN: Minimum value of the registration data AVE: Average of the registration data SD: Standard deviation of the registration data
1: ~  100:
CLEAR Go to memory clear
Maximum value of the registration data
1 blank line Displays the memory registration value for 1 to 100 sets (max.)
Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Enabled when CLEAR is
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
selected Go to memory clear Return to memory group selection
While the memory data is displayed, the "CLEAR" column is displayed in the last line of the data. Select the "CLEAR" column, and press the operation key 3 to display the memory clear screen below. Registration description in the memory group currently displayed can be cleared as necessary.
English
Item
English Japanese
1. No いいえ
Find Quality Products Online at: sales@GlobalTestSupply.com
2. Yes はい Clears the each memory description
Does not clear, and returns to the each memory display
Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Description
www.GlobalTestSupply.com
Operation key 3 Execute clear/Return to the
Operation key 4
13
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
each memory display without executing clear Return to the each memory display
Page 15
● Statistics memory display
Select "1. DATA" in contents selection to display the memory group selection screen.
(1) Memory group selection
Select the memory group you want to display.
English Japanese
Operation key Operation
Item
English Japanese
1 Group_1 グループ _1 2 Group_2 グループ _2 Displays memory group 2 3 Group_3 グループ _3 Displays memory group 3 4 Group_4 グループ _4 Displays memory group 4 5 Group_5 グループ _5 Displays memory group 5
Description
Displays memory group 1
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3
Operation key 4 Return to contents selection
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously Go to memory description display
(2) Memory data display
Select the memory group in (1), and press the operation key 3 to display the memory data for the selected memory group. Use the operation key 1 to change the display line. Use the operation key 2 to display the description of the next registration memory.
English
Display item Description No: 1 - 100 UNIT:
MAX: Maximum value MIN: Minimum value AVE: Average SD: Standard deviation
Unit
Japanese
(3) Memory clear
English
Select CLEAR in the following item of the last No.
Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Enabled when CLEAR is
Operation key 4
Operation key Operation Operation key 1 Disabled Operation key 2 Return to No. 1 Operation key 3 Go to memory clear Operation key 4 Return to contents selection
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
selected Go to memory clear Return to memory group selection
English
Item
English Japanese
Find Quality Products Online at: sales@GlobalTestSupply.com
1 No いいえ 2 Yes はい Clears the statistics memory description
Does not clear, and returns to the each memory display
www.GlobalTestSupply.com
Japanese
Description
14
Operation key Operation Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Execute clear/Return to the
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
statistics memory display without executing clear Return to the statistics memory display
Page 16
6-3. User settings
[ユーザー設定]
コンパレータ
メモリモード 開始メモリ
[選択してください] メモリ表示
ユーザー設定
システム設
(1) Operation ow
Go to the user setting mode in the following operation.
Measurement mode
During measurement
During measurement standby
[SET_USER]
1. COMPARATOR
2. MEMORY_MODE
3. START_MEM
[SET_CONTENTS]
1. DATA
2. SET_USER
3. SET_SYSTEM
Contents selection
2. User settings mode selection
Go to the user settings
(2) User settings - Setting item selection
Select the item to set in the following item selection screen, and perform detailed settings.
English Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously Go to the detail settings of the selected item Return to the Contents selection
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
15
Page 17
● List of the user settings
Item type
User
settings
Item
number
1
2
3
4 GRAPH_DISP グラフ表示
5 SAVE 設定保存
6 UNIT 単位
Main item title
English Japanese English Japanese English Japanese
COMPARATOR
MEMORY_MODE
START_MEM
コンパレータ
メモリモード
開始メモリ
Item
number
1-1
UPPER_LIMIT
1-2
LOWER_LIMIT
5-1 SEL_USER
5-2 R/W 読出 / 保存
Sub item title Setting description
上限値 0.00000 ~ 999999
下限値 0.00000 ~ 999999
Cont 連続
Stats 統計
Group_1 グループ _1
Group_2 グループ _2
Group_3 グループ _3
Group_4 グループ _4
Group_5 グループ _5
Normal 通常表示
Graph_1 グラフ 1
Graph_2 グラフ 2
Graph_3 グラフ 3
設定グループ
User □ Set_1
User □ Set_2
User □ Set_3
Used 読出
Over □ Write
設定 _1
設定 _2
設定 _3
保存
rev/min
m/min
cm/min
inch/min
feet/min
yards/min
rev/sec
m/sec
cm/sec
inch/sec
km/h
cm
m
km
inch
feet
yards
STP
Default
value
0.00000
0.00000
Cont Each 単独
Group_1
Normal
rev/min
Remark
6-digit numerical value setting with decimal point
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
16
Page 18
[ 1-2LOWER_LIMIT ]
[ 1
-2下限値 ]
[ 1-2下限値 ]
[ 1-2下限値 ]
[ 1-2.下限値 ]
[ 1-1UPPER_LIMIT ]
[ 1-1上限値 ]
[ 1-1上限値 ]
[ 1-1上限値 ]
[ 1-1上限値 ]
(3) Detail settings for each item
(3-1) Comparator setting
In the "COMPARATOR" setting, set the UPPER_LIMIT and LOWER_LIMIT. Compare UPPER_LIMIT and LOWER_LIMIT with the measurement display value, and display the judgment result in the screen.
・UPPER_LIMIT ≧ Measurement display value ≧ LOWER_LIMIT : OK judgment ・UPPER_LIMIT < Measurement display value : NG judgment ・LOWER_LIMIT > Measurement display value : NG judgment
(1) Upper limit value
The selected digit blinks
Operation
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
English
Japanese
Operation
key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
When setting the numerical value
Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously
Moves the digit Press and hold for 1 second to move continuously
Go to 1-2
Return to the user setting
Go to the rst digit of the numerical value setting
[1-1. UPPER_LIMIT] [1-1. UPPER_LIMIT] [1-1. UPPER_LIMIT]
Moves the digit
▲:The value in the selected
digit increases 0 - 9
Decimal point setting
The blinking decimal point moves by ▶
(2) Lower limit value
Operation
key
English
The continuous memory mode can be used only when one of "rev/min" to "km/h" is selected in the unit setting.
Japanese
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
When setting the numerical value
Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously
Moves the digit Press and hold for 1 second to move continuously
Enables the setting, and returns to the user settings.
Return to 1-1
Go to the rst digit of the numerical value setting
Operation
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
The selected digit blinks
Find Quality Products Online at: sales@GlobalTestSupply.com
[1-2. LOWER_LIMIT] [1-2. LOWER_LIMIT] [1-2. LOWER_LIMIT]
Moves the digit
▲:The value in the selected
digit increases 0 - 9
*
When both "UPPER_LIMIT" and "LOWER_LIMIT" are 0, judgment will be disabled, and the upper and lower limit judgment is not performed. *The display value is judged. *Values that represent UPPER_LIMIT<LOWER_LIMIT cannot be set.
www.GlobalTestSupply.com
Decimal point setting
The blinking decimal point moves by ▶
17
Page 19
(3-2) Memory mode
Set the memory mode.
English
Item
English Japanese
1 Cont 連続メモリ 2 Each 単独メモリ 3 Stats 統計メモリ Statistics memory mode
Japanese
Description
Continuous memory mode Each memory mode
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to memory clear Operation key 4 Return to the user setting
Memory clear conrmation
When changing the memory mode, the memory data currently registered will be cleared. When selecting "2. Yes" in the conrmation screen below, the memory data will be cleared. When selecting "1. No", the change will not be reected, and the display will return to the previous screen.
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
English
Item
English Japanese
1 No いいえ 2 Yes はい Memory data all clear permitted
* When changing the memory mode, the setting value of the number of registered memory sets for each memory group ("4. MEM_
GROUP" of "3. SET_SYSTEM" in contents selection) is initialized to the value that corresponds to the memory mode.
Memory data all clear not permitted
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Execute clear/Return to the
memory mode setting without executing clear
Operation key 4 Return to the memory mode
setting
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
18
Page 20
(3-3) Start memory group setting
Select the memory group to start memory registration.
English
Item
English Japanese
1 Group_1 グループ 1
2 Group_2 グループ 2
3 Group_3 グループ 3
4 Group_4 グループ 4
5 Group_5 グループ 5
Japanese
Description
Species the memory group to start memory registration as group 1
Species the memory group to start memory registration as group 2
Species the memory group to start memory registration as group 3
Species the memory group to start memory registration as group 4
Species the memory group to start memory registration as group 5
(3-4) Graph display
The display can be switched to the graph display.
English
Item
English Japanese
1 Normal 通常表示 2 Graph_1 グラフ _1 3 Graph_2 グラフ _2 4 Graph_3 グラフ _3 Display change: Graph 3 (Bar graph)
Japanese
Description
Display change: Normal display Display change: Graph 1 (Line graph) Display change: Graph 2 (Needle graph)
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the user settings.
Operation key 4 Return to the user setting
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the user settings.
Operation key 4 Return to the user setting
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
19
Page 21
(3-5) Setting save
(1) Setting group
The setting description can be saved in three patterns. The saved setting description can be read by selecting "Used".
English
Item
English Japanese
1 User_Set_1 設定 1 2 User_Set_2 設定 2 3 User_Set_3 設定 3 Go to READ/WRITE in the saved group 3
Japanese
Description
Go to READ/WRITE in the saved group 1 Go to READ/WRITE in the saved group 2
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to 5-2 Operation key 4 Return to the user setting
(2) Read/Write
English
Item
English Japanese
1. Used 読出
2. Over Write 保存
*The setting description to be saved includes the contents in "3. SET_SYSTEM" in contents selection.
Reads the setting description from the settings 1 to 3 selected in (1)
Writes the setting description to the settings 1 to 3 selected in (1)
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Perform read/write, and
Operation key 4 Return to 5-1
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
return to the user settings
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
20
Page 22
(3-6) Unit
The display unit of the measurement value can be selected.
English
Item (common to Japanese and English)
1 rev/min 11 2 m/min 12 3 cm/min 13 4 inch/min 14 5 feet/min 15 6 yards/min 16 7 rev/sec 17 8 m/sec 18 9 cm/sec 19
10 inch/sec
km/h miles/h cm m km inch feet yards STP
Japanese
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item,
and return to the user settings.
Operation key 4 Return to the user settings
● Unit type
・The units of items 1 to 12 are for speed measurement. They are used to measure the rotation speed and distance speed. ・The units of items 13 to 18 are for distance measurement. They are used to measure the distance value. ・The unit of item 19 STP is for manual measurement. It can be used to manually measure time required for one rotation, like using
the stopwatch, and display the converted value in the rev/min unit.
● About operation using the STP unit (manual measurement)
・When measuring in the STP unit, the measurement display is as follows. The display can be switched using the operation key 3.
(1) Time display
Displays time in second.
(2) rev/min conversion display
Displays the converted value into rev/min when measurement time is for one rotation.
(3) rev/min conversion display (after prescale)
Return to (1)
Displays the prescaled value of (2).
● Limitations for each function due to unit selection
When selecting units other than the speed measurement units, there are di󱐯erences in operation as follows.
Cases when the display can be switched
Speed
measurement
Distance
measurement
Manual
measurement
* For distance measurement and manual measurement, the graph display, continuous memory mode, and statistics memory mode
Find Quality Products Online at: sales@GlobalTestSupply.com
cannot be used. Also, the auto zero function is enabled only when the speed measurement unit is used.
Normal display Maximum value display Minimum value display
Average value display Acceleration rate display Prescale display Prescale maximum value display Prescale minimum value display Prescale average value display Prescale acceleration rate display Normal display Prescale display Normal display (time)
rev/min conversion display (after prescale)
www.GlobalTestSupply.com
Graph
display
Continuous memory Statistics memory Each memory
Stores the normal display value in the memory
Stores the prescale display value in the memory
× × ×
× × ×
Memory registration
Calculates statistics from the normal display value
Calculates statistics from the prescale display value
Registers the current display value
Registers the current display value
Registers the current display value
Registers the current display value
Auto zero
Enabled
Disabled
Disabledrev/min conversion display
21
Page 23
[システム設定]
オートオフ
オートゼロ グラフ
[選択してください] メモリ表示 ユーザー設定
システム設定
6-4. System setting mode
[SET_SYSTEM]
AUTO_OFF
AUTO_ZERO GRAPH SET
[システム設定]
オートオフ
オートゼロ グラフ
(1) Operation ow
Go to the system setting mode in the following operation.
Measurement mode
During measurement
During measurement standby
Contents selection
Go to the system settings
(2) System settings - Setting item selection
Select the item to set in the following item selection screen, and perform detailed settings.
English Japanese
System setting mode selection
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to the detail settings of
Operation key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
the selected item Return to the measurement mode
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
22
Page 24
● List of the system settings
Item type
System
settings
Item
number
1 AUTO_OFF オートオフ
2
3
4
5 CONTACT 接触 / 非接触
6
7 PRESCALE プリスケール 0.00001 ~ 999999. 000001.
8 DIAMETER 直径
9
10
11
12
13 MEAS_OPR 計測スイッチ
14
Main item title
English Japanese English Japanese English Japanese
AUTO_ZERO
SET_GRAPH
MEM_GROUP メモリグループ
DISP_CYCLE
DIST/PULSE
BRIGHTNESS
REVERSE_LCD
LANGUAGE
FACTY_DEFAULT
オートゼロ
グラフ
表示更新周期
距離 / パルス
画面明るさ
逆転表示
言語
初期化
Item
number
3-1
3-2
3-3 3-4
4-1
4-2 Group_ □
8-1 DIA_UNIT 直径単位
8-2 DIAMETER 直径値 0.00001 ~ 999999. 48.5104
9-1 DIST_UNIT 距離単位
9-2
Find Quality Products Online at: sales@GlobalTestSupply.com
Sub item title Setting description
1 min 1 分
30 min 30 分
1 sec 1 秒 10 sec 10 秒 30 sec 30 秒 60 sec 60 秒
None 無効
GRAPH_FLOW
GRAPH_TIME
GRAPH_MAX
GRAPH_MIN
MEM_GROUP
DIST/PULSE
グラフ方向
グラフ時間
グラフ MAX 値 グラフ MIN 値
メモリグループ
グループ _ □
距離 / パルス
www.GlobalTestSupply.com
Right 右方向
Left 左方向
1 sec 1 秒 10 sec 10 秒 50 sec 50 秒
100 sec 100 秒
0.00001 ~ 999999 000100.
0.00001 ~ 999999 0.00000 Group_1 グループ _1 Group_2 グループ _2 Group_3 グループ _3 Group_4 グループ _4 Group_5 グループ _5
0 ~ 1000 0
0 ~ 100 0
0 ~ 100 0
Contact 接触
Non Contact
Auto 自動検出
Contact(Low)
0.00001 ~ 999999. 000001.
Momemtary
Continuous
Yes はい No いいえ
非接触
接触(低速) 100 msec 500 msec
1 sec 5 sec
mm
cm
m inch feet
yards
mm
cm
m inch feet
yards
20% 40% 60% 80%
100%
OFF
ON 日本語 English
モーメンタリ オルタネイト
Default
value
1 min 10 min 10 分
None
Right
1 sec
Non
Contact
1sec
mm
mm
60%
OFF
English
Momemtary
Remark
6-digit numerical value setting with decimal point
Select the memorygroup to be
set in 4-2
when setting group other than group 1 in the cont memory mode
when setting group other than group 1 in the cont memory mode
when setting group other than group 1 in the cont memory mode
6-digit numerical value setting with decimal point
6-digit numerical value setting with decimal point
6-digit numerical value setting with decimal point
23
Page 25
(3) Detail settings for each item
(3-1) Auto OFF
Set a period of time until the power is automatically turned OFF. The power is turned OFF when the unoperated status continues for the specied auto o󱐯 time.
English
Item
English Japanese
1. 1 min 1分
2. 10 min 10 分
3. 30 min 30 分
● When performing communication using the optional USB cable, the auto o󱐯 function is disabled.
Japanese
Description
Power OFF when the unoperated status continues for 1 minute
Power OFF when the unoperated status continues for 10 minutes
Power OFF when the unoperated status continues for 30 minutes
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
(3-2) Auto zero
Set a period of time until the display is turned to 0 during measurement. During measurement, when the rotation undetected status continues longer than the specied auto zero time, the measurement value display is reset to "0".
English
Item
English Japanese
1 1 sec 1 秒
2 10 sec 10 秒
3 30 sec 30 秒
4 60 sec 60 秒
None 無効 Auto zero disabled
During measurement, when no input pulse status continues for 1 second, the display is turned to 0
During measurement, when no input pulse status continues for 10 seconds, the display is turned to 0
During measurement, when no input pulse status continues for 30 seconds, the display is turned to 0
During measurement, when no input pulse status continues for 60 seconds, the display is turned to 0
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
24
Page 26
(3-3) Graph display
Set various graph settings.
(1) Graph ow
Set the line graph direction. This setting is enabled only for Graph_1 (line graph).
English
Item
English Japanese
1. Right 右方向 Graph update direction: Right
2. Left 左方向 Graph update direction: Left
Graph ow: Right direction
New Old Old New
Japanese
Description
Graph ow: Left direction
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to
move continuously Operation key 3 Go to 3-2 Operation key 4 Return to the system settings
(2) Graph time
Set the graph time. This setting is for time width of the graph for Graph_1 (line graph). Also, it represents the update time when the measurement display "AVE", "ACC", "P.AVE", or "P.ACC" is indicated.
English
Item
English Japanese 1 1 sec 1 秒 Graph time setting: 1 second 2 10 sec 10 秒 Graph time setting: 10 seconds 3 50 sec 50 秒 Graph time setting: 50 seconds 4 100 sec 100 秒 Graph time setting: 100 seconds
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to
move continuously Operation key 3 Go to 3-3 Operation key 4 Return to 3-1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
25
Page 27
[ 3-4GRAPH_MIN ]
[ 3-4グラフ最小値 ]
[ 3-4グラフ最小値 ]
[ 3-4グラフ最小値 ]
[ 3-4グラフ最小値 ]
[ 3-3GRAPH_MAX ]
[ 3-3グラフ MAX ]
[
3-3
グラフ MAX ]
[ 3-3.グラフ MAX ]
[ 3-3グラフ MAX ]
(3) Graph maximum value
Set the maximum value in the graph display.
English
Japanese
Operation key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
When setting the numerical value Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously Moves the digit Press and hold for 1 second to move continuously
Go to 3-4
Return to 3-2
Go to the rst digit of the numerical value setting
[3-3. GRAPH_MAX] [3-3. GRAPH_MAX] [3-3. GRAPH_MAX]
Operation
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
The selected digit blinks
Moves the digit
▲:The value in the selected
digit increases 0 - 9
Decimal point setting
The blinking decimal point moves by ▶
(4) Graph minimum value
Set the minimum value in the graph display.
English
Japanese
Operation key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
When setting the numerical value Increases the numerical value in the selected digit Press and hold for 1 second to continuously increase Moves the digit Press and hold for 1 second to move continuously Enable the setting, and return to the system settings.
Return to 3-3
Go to the rst digit of the numerical value setting
[3-4. GRAPH_MIN] [3-4. GRAPH_MIN] [3-4. GRAPH_MIN]
Operation
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
The selected digit blinks
Moves the digit
▲:The value in the selected
digit increases 0 - 9
Decimal point setting
The blinking decimal point moves by ▶
About the graph maximum/minimum values
Graph 1
Graph maximum value
Graph minimum value
Find Quality Products Online at: sales@GlobalTestSupply.com
*Values that represent the graph maximum value graph minimum value cannot be set.
www.GlobalTestSupply.com
Graph minimum value
Graph 2 Graph 3
Graph maximum value
Graph minimum value
26
Graph maximum value
Page 28
(3-4) Memory group
[ 4-2GROUP_1 ]
[ 4-2グループ 1 ]
[ 4-2グループ
1 ]
[ 4-2グループ 1 ]
Set the number of memory data registration sets for each memory group.
(1) Setting memory group selection
Select the memory group to set.
English Japanese
Item
English Japanese
1. Group_1 グループ 1 Go to the setting of memory group 1
2. Group_2 グループ 2 Go to the setting of memory group 2
3. Group_3 グループ 3 Go to the setting of memory group 3
4. Group_4 グループ 4 Go to the setting of memory group 4
5. Group_5 グループ 5 Go to the setting of memory group 5
Description
(2) Group setting
Set the number of memory data registration sets for the memory group selected in (1).
English
*No decimal point setting
Japanese
Operation key
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to delete the registration memory data
Operation key 4 Return to 4-1
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to
move continuously Operation key 3 Go to 4-2 Operation key 4 Return to the system settings
Operation When setting the numerical value
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
for each memory group
[4-2. Group 1] [4-2. Group 1]
Moves the digit
The selected digit blinks
▲:The value in the selected
digit increases 0 - 9
Moves to the top digit
*Set the value for each memory mode within the setting range in the table below.
When the total number of memory data registration sets for memory groups exceeds the maximum value in the table below, it cannot be set.
* *When the total number of memory data registration sets for memory groups is 0, it cannot be set. * When changing the memory mode, the setting value of the number of registered memory sets for each memory group ("4. MEM_
GROUP" of "3. SET_SYSTEM" in contents selection) is initialized to the value that corresponds to the memory mode.
Memory
mode
Continuous
Each 100 sets
Find Quality Products Online at: sales@GlobalTestSupply.com
Statistics 100 sets
Maximum number of registration sets
1000 sets
Memory group 1 Memory group 2 Memory group 3 Memory group 4 Memory group 5
0 - 1000
(Default value: 1000)
0 - 100
(Default value: 100)
0 - 100
(Default value: 100)
www.GlobalTestSupply.com
0 - 1000
(Default value: 0)
0 - 100
(Default value: 0)
0 - 100
(Default value: 0)
Setting range
0 - 1000
(Default value: 0)
0 - 100
(Default value: 0)
0 - 100
(Default value: 0)
27
0 - 1000
(Default value: 0)
0 - 100
(Default value: 0)
0 - 100
(Default value: 0)
0 - 1000
(Default value: 0)
0 - 100
(Default value: 0)
0 - 100
(Default value: 0)
Set the total number of data sets for memory groups 1 to 5 to be within the maximum number of registration sets
Setting condition
Page 29
● Registration memory data deletion in each memory group
When the number of memory data registration sets for each memory group has been changed, the memory data registered in the corresponding memory group needs to be deleted.
English
Japanese
グループ 1 を消去しますか?
1. いいえ
2
. はい
Item
English Japanese
1. No いいえ
2. Yes はい
* When the registration memory data is not deleted, the number of memory data registration sets in the memory group will not be
changed.
Registration memory data deletion in the current memory group: not permitted Registration memory data deletion in the current memory group: permitted
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Execute clear/Return to the
number of memory sets setting without executing clear
Operation key 4 Return to the number of
memory sets setting
(3-5) Contact/Non-contact
The measurement method can be selected from contact and non-contact. Automatic detection can be also performed.
English
Item
English Japanese 1 Contact 接触 Select when the contact adapter is attached 2 Non Contact 非接触 3 Auto 自動検出 Automatic detection mode
4 Contact(Low) 接触(低速)
*Select "4. Contact (Low)" when using the optional contact adapter for low speed.
When "3. AUTO DETECT" is selected, contact attachment may not be automatically detected correctly depending on the environment.
*
In that case, select 1 to 2 manually.In addition, for automatic detection, contact attachment is detected when measurement starts. The measurement value may not be displayed correctly from measurement start to completion of automatic detection.
*If using a low-speed contact adapter, do not select "3. Auto".
Japanese
Description
Select when the contact adapter is not attached
Select when the contact adapter for low speed is attached
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
(3-6) DISP_CYCLE Display update cycle
The display update cycle of the measurement value can be selected. The measurement value is updated every time the period of time selected in this setting is reached.
English
Item (common to
Japanese and English)
1 100msec Display update cycle: 100msec 2 500msec Display update cycle: 500msec 3 1sec Display update cycle: 1sec 4 5sec Display update cycle: 5sec
Find Quality Products Online at: sales@GlobalTestSupply.com
*When the display update cycle has been changed, the measurement value to be displayed may slightly change.
www.GlobalTestSupply.com
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
28
Page 30
(3-7) Prescale
[ 7PRESCALE ]
[ 7プリスケール ]
[ 7プリスケール ]
[ 7プリスケール ]
[ 7プリスケール
]
The prescale value can be set. Prescale is a user dened mathematical correction of the values displayed. This can be helpful when a scaling of the point measured has more signicant meaning to the user taking the measurement. The measurement value display of "PRE", "P.MAX", "P.MIN", "P.AVE", and "P.ACC" is inuenced by the prescale value.
The selected digit blinks
English Japanese
Operation key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
When setting the numerical value Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously Moves the digit Press and hold for 1 second to move continuously Enable the setting, and return to the system settings.
Return to the system settings
Go to the rst digit of the numerical value setting
[7. PRESCALE] [7. PRESCALE] [7. PRESCALE]
Moves the digit
▲:The value in the selected
digit increases 0 - 9
*The setting range is between 0.00001 and 999999. 0 cannot be set.
Decimal point setting
The blinking decimal point moves by ▶
Operation
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
・About prescale
Set the prescale value to the number of pulses per rotation (for contact measurement, the number of measurement shaft rotations), and perform the following calculation for measurement display
When the measurement value is 100.0, and the prescale value is 2.00, 100.0/2.00 = Display: 50
*The accuracy depends on the measurement value before applying prescale.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
29
Page 31
(3-8) Diameter
[ 8-2DIAMETER ]
[ 8-2直径値 ]
[ 8-2直径値 ]
[ 8-2直径値 ]
[ 8-2.直径値 ]
Set the diameter value of the speed measurement wheel when using the speed measurement in contact measurement. *These setting contents are not applied to non-contact measurement.
(1) Diameter unit
Set the diameter value unit.
English
Item (common to
Japanese and English)
1 mm Diameter unit: mm 2 cm Diameter unit: cm 3 m Diameter unit: m 4 inch Diameter unit: inch 5 feet Diameter unit: feet 6 yards Diameter unit: yards
(2) Diameter
Set the diameter value.
English
Japanese
Description
Japanese
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to 8-2 Operation key 4 Return to the system settings
Operation key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
Operation key Operation
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
Operation
When setting the numerical value Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously Moves the digit Press and hold for 1 second to move continuously Enable the setting, and return to the system settings.
Return to 8-1
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
Go to the rst digit of the numerical value setting
[8-2. DIAMETER] [8-2. DIAMETER] [8-2. DIAMETER]
Moves the digit
The selected digit blinks
▲:The value in the selected
digit increases 0 - 9
Decimal point setting
The blinking decimal point moves by ▶
・About diameter
From the diameter value of the speed measurement wheel, obtain the distance per rotation of the measurement shaft, and then obtain the peripheral speed and distance. This setting is enabled only when the speed measurement unit (excluding rev/min and rev/sec) and distance measurement unit are selected at the time of contact attachment.
Calculate the circumference distance from the diameter of the speed measurement wheel
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
30
Page 32
[ 9-2DIST/PULS ]
[ 9-2距離値 ]
[ 9-2距離値 ]
[ 9-2距離値 ]
[ 9-2.距離値 ]
(3-9) Distance/Pulse
Set the distance value per pulse for non-contact measurement. *This setting contents are not applied to contact measurement.
(1) Distance unit
Set the distance value unit per pulse.
English Japanese
Item (common to
Japanese and English) 1 mm Diameter unit: mm 2 cm Diameter unit: cm 3 m Diameter unit: m 4 inch Diameter unit: inch 5 feet Diameter unit: feet 6 yards Diameter unit: yards
Description
(2) Distance/Pulse
Set the distance value per pulse.
English Japanese
Operation key Operation
Operation key 1 Select the upper item
Operation key 2 Select the lower item
Operation key 3 Go to 9-2 Operation key 4 Return to the system settings
Operation key
Operation
key 1
Operation
key 2
Operation
key 3
Operation
key 4
Press and hold for 1 second to move continuously
Press and hold for 1 second to move continuously
Operation
When setting the numerical value Increases the numerical value in the selected digit Press and hold for 1 second to increase continuously Moves the digit Press and hold for 1 second to move continuously Enable the setting, and return to the system settings.
Return to 9-1
When setting the decimal point
Return to the numerical value setting
Moves the decimal point Press and hold for 1 second to move continuously
Go to the rst digit of the numerical value setting
[9.2. DIST/PULS] [9.2. DIST/PULS] [9.2. DIST/PULS]
The selected digit blinks
Moves the digit
▲:The value in the selected
digit increases 0 - 9
Decimal point setting
The blinking decimal point moves by ▶
・About distance/pulse
Obtain the peripheral speed and distance from the distance value per pulse. This setting is enabled only when the speed measurement unit (excluding rev/min and rev/sec) and distance measurement unit are selected at the time of contact attachment.
Obtain the peripheral speed and distance from the distance value per pulse.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31
Page 33
(3-10) Brightness
The brightness on the screen can be set.
English
Item (common to
Japanese and English) 1 20% 2 40% 3 60% 4 80% 5 100%
Dark
Bright
Japanese
Description
(3-11) Reverse display
The vertical reverse can be applied to the display items.
English
Item (common to
Japanese and English)
1. OFF Reverse display OFF 2. ON Reverse display ON
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
*The reverse display is enabled only in the measurement mode (including the graph display).
For various settings, and memory data display, the reverse display is disabled.
(3-12) Language
The language to be displayed can be set.
English
Item (common to
Japanese and English)
1. Japanese Language to be displayed: Changes to Japanese
Find Quality Products Online at: sales@GlobalTestSupply.com
2. English Language to be displayed: Changes to English
www.GlobalTestSupply.com
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
32
Page 34
Group_ Clear?
1. No
2. Yes
(3-13) Measurement switch
The operation mode of the measurement switch can be set.
English
Item
English Japanese
1 Momentary
2 Continuous
モーメンタリ
オルタネイト
Measures only while the measurement switch is pressed and held Starts measurement when the measurement switch is pressed once Ends measurement when it is pressed again
Japanese
Description
(3-14) Initialization
The setting value can be initialized.
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Enable the selected item, and
return to the system settings
Operation key 4 Return to the system settings
English
Item
English Japanese 1 No いいえ 2 Yes はい Initializes the setting contents
*The user setting contents are also initialized.
Note that the contents saved in "User_Set_1" to "User_Set_3" of "SEL_USER" in the user settings are not initialized.
Return to the system settings without initialization
Japanese
Description
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Execute initialization/Return
to the system settings without initialization
Operation key 4 Return to the system settings
*Memory registration when the unit, diameter, distance/pulse have been changed
When performing memory registration after the unit, diameter, distance/pulse values have been changed, the already registered memory data needs to be deleted. If the existing memory data is not deleted, new memory data cannot be registered.
Example: Start memory registration after changing the unit (when memory data is already registered in the start memory group 3, and
memory groups 1 to 5)
Memory
registration
start
Memory registration
key ON
Memory group 3 clear
グループ□ 消去しますか?
1. いいえ
はい
2
.
Item
English Japanese
1. No いいえ
Find Quality Products Online at: sales@GlobalTestSupply.com
2. Yes はい
Memory registration disabled. Return to the measurement mode
Clears the memory in the current memory group, and moves to next
Description
www.GlobalTestSupply.com
Memory group 4 clear Memory group 2 clear
Operation key Operation
Operation key 1 Select the upper item
Press and hold for 1 second to move continuously
Operation key 2 Select the lower item
Press and hold for 1 second to move continuously
Operation key 3 Execute clear/Return to the
continuous memory display without executing clear
Operation key 4 Return to the continuous
memory display
33
Page 35
7. Battery Replacement
Replace batteries according to the following procedure.
1
Replace with new batteriesWhen the battery level becomes low
2
Rear panel Battery cover
*Use AA type alkali or manganese batteries. When other types of batteries are used, operation cannot be guaranteed. * In the following cases, the settings and/or memory data may not be saved normally: when the battery level is low, when removing batteries
during operation, when disconnecting the USB cable (option) during use at the time of the power supplied only through the USB cable.
When the battery indicator blinks, open the battery cover on the rear panel, and replace batteries.
Put in two new batteries, aligning the positive and negative polarities correctly, and close the cover.
・Do not mix and use new and old
batteries, or di󱐯erent types of batteries.
・Failure to follow this may result in injury
and/or damage due to leakage. When you do not use the unit for a long period of time, remove batteries, and keep them in the carrying case.
8. Dimensional Drawing
Laser beam irradiation port
1/4-20UNC Depth 8
When a contact adapter is attached
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
34
Page 36
9. Specications
Item Description
Measurement system Non-contact  Red laser  Class2  Maximum output: 1mW  Wave length: 650nm
Display Organic light emitting display (OLED) (blue light emission) graphic display type
Operation switch 5 points
Measurement range
Measurement unit
Measurement accuracy
Measurement distance 50 ~ 4,000mm (2 in. to 13 ft.)
Measurement time cycle
(display update cycle)
Functions
Housing Polycarbonate resin
Weight 125g
Power supply AA alkali dry cell battery x 2 (non-charging type), and USB cable (option)
Continuous operating time Up to 20 hours (when the screen brightness is set to 20%)
Compliance standard CE、RoHS、PSC、FDA、IEC/EN 60825-1 : 2007
Operating/
Storage temperature
Accessories
Option
6 to 99,999rpm: Non-contact measurement
0.8 to 25,000rpm: Contact measurement Speed measuremen: rev/min, m/min, cm/min, inch/min, feet/min, yards/min, rev/sec, m/sec, cm/sec, inch/sec,
km/h, miles/h Distance measurement: m, cm, km, inch, feet, yards Manual measurement: STP
unit rev/ min : ± 0.006% of the display value, and ±1 of the least signicant digit
(Except for display update cycle :100mse
・Additional errors in the measurement accuracy may be caused by the measuring environment
(hand movement, the angle of the target, etc.).
・In the case of contact measurements, errors due to contact slippage or from the measurement wheel
may result in additional inaccuracies.
100ms, 500ms, 1sec, 5sec
Auto o󱐯 function, memory data registration function, prescale function, peripheral speed wheel diameter setting, comparator function, USB communication, graph display, reverse display, language selection (Japanese/English)
0 ~ 40℃ (32 to 104 ℉) (No condensation)
Carrying case, AA alkali dry cell battery x 2, contacts (concave and convex types), contact adapter, speed measurement wheel (circumference of 6 inches), Instruction manual, reective tape
・PC Connect Set (Software, USB Cable) ・Contact adapter Set [Contact adapter, 2 cone adapters, 1 funnel adapter, Master wheel (6”Cir.), Extension
shaft]
・Contact adapter Set (Low) [Contact adapter (Low), 2 cone adapters, 1 funnel adapter, Master wheel (6”Cir.),
Extension shaft]
・Master wheel (12”Cir.)
c & 500msec)
*Note that applying solvents such as thinner or alcohol, etc. to the main unit may result in the paint peeling and/or damage to the unit.
Only clean with cloth that is damp from water or water with a mild detergent.
*Battery maybe have been discharged.
● Regarding reective tape
There are two di󱐯erent types of reective tape included. Use tapes according to the application.
● Standard tape (plain) ● High intensity reective tape (grid pattern)
For the measurement of close range: 50mm to 1,000 mm. For longer distance, or other cases when the measurement
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
doesn't go through with the standard tape.