FG-7000T Digital Torque Gauge
Operation Manual
Operators should wear protection such as a mask and gloves in case
pieces or components break away from the unit under test.
Whether the unit is ON or OFF, DO NOT exceed the capacity of the
sensor. NEVER exceed 120% of the rated capacity, or the torque sensor
will be damaged. At 110% of the rated capacity the display will flash
a warning.
If mounting FG-7000T Series Digital Torque Gauges, use M4 mounting
screws with a maximum insertion depth of 7 mm into the gauge. Hand
tightens mounting screws, DO NOT use tools. Do not use damaged
clamp.
Measure in line torque only. DO NOT attempt to measure forces at an
angle to the sensor – damage to sensor may result.
Do not attempt to repair or alter this instrument. Warranty will be voided
and damage to the unit may result.
Use and store within the stated temperature and humidity ranges, or
damage and failure may result.
If not using this instrument for extended periods of time, remove the
batteries to prevent potential battery leakage from causing product
damage.
DO NOT use tools to over torque the connection adapter. Hand-tighten
only so damage does not occur.
The FG-7000T Series digital torque gauges provide testing flexibility with their external torque sensor input. Torque sensors
with standard chuck adapter are immediately recognized when
connected to the display base and are available for your specific
testing needs in 1 N-m (9 in-lb) 5 N-m (44 in-lb), 10 N-m (88 inlb) capacities. The multiple-language FG-7000T’s provide menu
programming for easy selection and set-up of the instrument to
your desired requirements. Four selectable modes of operation
include: Track mode for live readings, Peak mode for displaying
the maximum reading, First Peak where only the initial peak is
recorded once a decrease is sensed and Preset which initiates
user-programmed tolerance alarms for go/no go testing. The
display has two selectable operations, numerical view with directional bar graph or graphical view with directional bar graph.
In graphical view when alarm tolerances are set, the process is
plotted in relation to the upper and lower limit graph lines. Combined with the go/no go icons, a simple pass/fail determination is
recognized. These high-tech instruments can data log a reading
at the push of a button for simple data acquisition or be set to
continuous data storage. Data can be viewed on the screen, sent
to the optional printer, or loaded to be analyzed and graphed on
the free software program. The 1,000 point memory with definable
groups allows for multiple tests to be recorded and easily separated upon loading. In addition, the comparator output can be set
up for integration of the instrument into a quality system for repetitive testing such as on a production line. The FG-7000T’s robust
housings are designed to fit perfectly in the operator’s hand for
portable testing. The large back-lit, 180° auto-reversible display,
CW/CCW directional bar graph, combined with the dual labeled
key pad allows for usage of the gauge in various positions while
still being able to easily view and operate. These various features
make the FG-7000T the ideal torque instrument for various applications such as closing or opening analysis of containers, valves
and door hardware, failure or destructive torque testing, or almost
any torque testing requirement involving incoming quality inspection, finished goods testing to R&D.
SPECIFICATIONS
Accuracy: ± 0.3% F.S.
Selectable Units: N-m, N-cm, kgf-cm, lbf-in (Depending on
Range)
Overload Capacity: 120% of F.S. (LCD flashes beyond 110% of
F.S.)
Measurement method: Peak, First Peak Preset or Track Mode
Data Sampling Rate: 1000 Hz
Display: 160*128 dot matrix LCD with LED Backlight
Display Update Rate: 10 times/second
Resolution: (See page 2 Resolution Table)
Memory: 1000 data
Set Point: Programmable high and low limits
Battery Indicator: Display flashes battery icon when battery is
low
Power: 3.6VDC 800mAH Ni-MH rechargeable batteries
Battery Life: Approximately 16 hours continuous use per full
charge
Charger / Adapter: Universal USB/BM charger, Input: 110 ~
240VAC
Temperature Effects: <0.054% per °F (0.03% FS per °C)
Outputs: USB, Serial Port RS-232, High & Low Limit NPN
Operating Temperature: 14 to 104°F (-10 to 40°C)
Storage Relative Humidity: 20 to 80%
Housing: Aluminum
Storage Temperature: -4 to 122°F (-20 to 50°C)
Oper. Relative Humidity: 5 to 95%
Dimensions: 5.7 x 2.9 x 1.4” (145 x 73 x 35.5 mm)
Product Weight: 3.7 (1.7 kg)
Package Weight: 5.5 (2.5 kg)
Warranty: 1 year
Certification: CE
Included Accessories: Key Chuck, AC Adapter/Charger, USB
cable, calibration cert.
1. LCD SCREEN STANDARD VIEW
Test Mode Icons:
Track: Real Time, live measuring mode
Peak: Reading will not change until a higher value is measured
Preset: Set the upper & lower limit for GO/NG testing
First Peak: Captures First Peak after a decrease has been
detected. Drop Ratio set in menus.
2. Battery Icon: Battery level or charging status. Flashes when
gauge needs to be recharged.
3. OK/OV Indicator:
Under Lower Limit
Between Low Limit & Upper Limit
Over Upper Limit
4. Torque Icons: Indicates force direction.
Clockwise (CW)
Counterclockwise (CCW)
5. Current measured value
6. Analog Bar: Indicates current position within full scale. When
the bar enters the area enclosed by the dotted line, this
signifies the full scale capacity is exceeded by an overload
condition.
7. Storage Icon: Indicates data is being saved.
8. System time
9. Units Indicator: Selected engineering unit.
10. Load Cell Capacity Icon:
4
9
2
7
Digital View
3
1
5
6
If no load cell is connected, this symbol appears & blinks
10
8
Graphic View
2. RESOLUTION TABLE
Model N-m N-cm kgf-cm lbf-ft lbf-in Interface Type
1 N-m
5 N-m
10 N-m
Capacity 1 100 10 - 8.9 Drill
Resolution 0.001 0.1 0.01 - 0.005
Capacity 5 500 50 3.68 44.3
Resolution 0.001 0.1 0.01 0.001 0.01
Capacity 10 1000 100 7.38 88.5 Drill Chuck
Resolution 0.01 1 0.1 0.001 0.01
Chuck
1-10 mm
(0.034 to 0.39 in)
2-13 mm
(0.079 to 0.512 in)
2
3. KEY FUNCTIONS
All keys are capacitive touch.
ON/OFF: Push for 2 seconds to power On or Off
During Measurement: Print the current force value or
store data, depending on the key setting. (See 4.5.8 for
key setting)
In Menus: Back or quit.
From the home screen, touch “MENU” to enter the main menu.
(Figure 4-1)
During Measurement: Enter the menus.
In Menus: Select or Enter
During Measurement: Track mode, tares weight of
attachment. In Peak & Auto Peak modes, resets the
peak value.
In Menus: Moves selection up or increases the value.
During Measurement: Changes the measure mode from
Track, Peak, Auto Peak, First Peak
In Menus: Moves selection down or decreases the
value.
4. ADVANCED MENU OPTIONS
4.1 Menu Structure
Unit
Group
Measurement
Memory
Printing
Menu
System
Language
Information
Test Mode
Alarm
Calibration
Storage Mode
Browse All
Browse Selected
Delete Selected
Delete All
Print Recent
Print Selected
Print All
Display Mode
Display Direction
Auto Power Off
Backlight
Key Tone
Date/Time
Password
Key Setting
RS-232 Baud Rate
Default Setting
Figure 4-1a Figure 4-1b
4.2 Measurement
The Measurement menu contains six selectable items: Unit,
Group, Test Mode, Alarm and Calibration. (Figure 4-2a)
Figure 4-2a
4.2.1 Unit
The measuring unit can be selected under this menu. Different range models may have different unit selection capabilities. Touch “ZERO” or “MENU” keys to shift to the next selection. Press “LOG” to cancel or touch “MENU” to confirm and
exit. (Figure 4-2c)
Figure 4-2c
4.2.2 Group
When several test samples need to be measured, the samples can be coded into groups. The range is 01-99. When set
to “00”, become, “01” automatically. Press “ZERO” to adjust
the value, touch “MODE” to shift to the next position. Touch
“LOG” to cancel; press “MENU” to confirm and exit. (Figure
4-2d)
Figure 4-2d
3