Shimpo DT-326B Instruction Manual

DT-326B LED Stroboscope Operation Manual
Use in flammable environments is prohibited. Use in this manner may result in fire or explosive.
Don’t look directly into the LED light Source. This may result in eye in­jury.
Don’t use or store in the following environments. Direct sunshine con­densation, dust or caustic.
Do not alter, or modify of improperly. Such action may cause damage and void warranty.
Operate with 0-35°C (32-95°F), 35-85% RH. Use outside of this range may alter operation of the unit.
Case may become excessively hot when used continuously for more than 2 hours. Mount unit on a tripod or other fixed device.
The DT-326B is a portable, light-weight stroboscope utilizing su­per bright CREE High Powered LED lamps. The DT-326B’s LED array provides a bright, stable strobe light over a wide measure­ment range with a lifetime far exceeding xenon lit stroboscopes. Containing a rechargeable lithium battery, a single charge can last up to a full 12 hours of operation. The DT-326B is designed for speed and frequency measurements in motion and vibration anal­ysis. It is ideal for predictive and preventive maintenance appli­cations such as: motors, shafts, roto-gravure printing, extruders, blow molding, wire lettering and striping, engraving, pulse jets, water jets, fans, cams, gear teeth, belt inspections, fuel injectors, vibration analysis, audio speaker analysis, spindle spinning, cut­ting blade timing and sharpness, plus many more in the printing, packaging, textile, automotive, cable, mining, steel, chemical, op­tical, medical and shipbuilding industries.
SPECIFICATIONS
Flash Rate Range: 60 to 120,000 FPM ; 1-2000 Hz Accuracy: 0.01%±1 digit of F.S. @ 77° F (25°C) Lux Rating (Approx.): 6000 FPM & 2.5°: Distance 8˝ (20 cm)
5800 lx with 4˝ (100 mm) irradiation dia., Distance 20˝ (50 cm) 1050 lx with 6˝ (152 mm) irradiation dia. 1500 FPM & 2.5°: Distance 8˝ (20 cm) 6400 lx with 4˝ (100 mm) irradiation dia., Distance 20˝ (50 cm) 1125 lx with 6˝ (152 mm) irradiation dia.
Lamp Lifetime: Approximately 3~5 years depending on usage. Display: Backlit LCD Resolution: 60 ~12,000 FPM = 0.1 FPM; 12,001~120,000 FPM
= 1 FPM; 1~200 HZ = 0.01 Hz; 201~2000 = 0.1 Hz
Flash Duration: 0.1°- 2.5° Phase Shift: 0-359° Power Requirement: 100-240 VAC 50/60Hz Battery: Lithium DC 12.6V 730 mAh Battery Life: Approx. 12 hours depending on settings Input Signal: 12 V Pulse Input: Mini-USB Input Signal Range: 60-120,000 FPM (1-2000 Hz) Input pulse width: over 50 µs Input Signal Flash Delay: 0-999 ms; 0-359° Temperature Limits: 32-95°F (0-35°C) Humidity Limits: 35 to 85% RH Enclosure: ABS Product Weight: 0.84 lb (380 g) Package Weight: 3.50 lb (1.6 kg) Dimensions: 8.2 x 2.8 x 2.4” (208 x 70 x 60 mm) Approvals: CE Warranty: 1 year Included Accessories: AC power adapter, Carrying Case
Operation Panel
1. POWER: Power on and off.
2. MODE: Select mode of operation: Internal/External/ Parameter Settings
3. SET: Scroll through parameter settings, Store setting values.
4. x2: Multiplies the flash rate/frequency by a factor of 2.
5. 1/2: Divides the flash rate/frequency by a factor of 2.
6. RATIO: Adjusts flash duration (flash pulse width) in Internal/
External modes.
7. “+”: Advance image forward 3 degrees at a time in internal
mode. In parameter setting mode, adjusts setting values.
8. “-”: Retard image backwards 3 degrees at a time in internal
mode. In parameter setting mode, adjusts setting values.
9. Dial: Set flash rate or frequency. CW: Increase flash rate/fre-
quency. CCW: Decrease flash rate/frequency. (Turn dial quickly to drastically change value; Turn dial slowly to change value by 1 digit.) In parameter setting mode, CW or CCW rotation changes the setting value.
10: FPM/Hz: Select between FPM or Hz units of measure
2
1
3
4
5
6
7
8
9
10
LCD Display
Battery Indicator
Display Parameter Icons
Sub Data Display
Main Data Display
•Flash rate value will be displayed in internal flashing or Exter­nal trigger mode.
Units
Main
Data
FUNCTION INSTRUCTIONS
Mode Selection - To switch between INTERNAL, EXTERNAL and
PARAMETER mode, press and release the MODE key. For details about Parameter setting mode, please refer to Parameter setting mode section.
•Display will indicate “P” or “LCD” (in ON/OFF setting of LCD backlight) in *Parameter setting mode.
*For details about Parameter setting mode, please refer to Pa­rameter setting mode.
Sub Data Display
Internal Mode
•Phase Shift will indicate degrees of phase shift in internal flash­ing mode. To change the degrees of phase shift, press + or ­keys. The display will reflect the cumulative angle of phase shift.
•The flash duration time will be displayed when the RATIO key is pressed. The flash duration time can then be increased or decreased with the dial from 0 to 2.5 (0~2.5°) degrees.
External Mode
• In external trigger mode, the display will show the settings of delay time. For details about changing these settings, please see external trigger mode section.
Battery Indication - Indicates remaining battery charge.
Note: The display will not indicate a battery charge when the unit
is being used with the AC Adapter.
Internal Flashing Mode
Changing Units of Measure in Internal Mode - To change the
measuring units from FPM to Hz, press and release the FPM/Hz key. Units will alternate with each press and release.
Flash Rate (Frequency Setting) in Internal Mode ­You can set the flash rate by turning the dial in the center of the unit. To increase the flash rate, turn the dial in the clockwise direc­tion. To decrease the flash rate, turn the dial in the counter-clock­wise direction. For small adjustments, turn the dial slowly. For quick adjustments, rotate the dial quickly. The flash rate range and resolution will be different according to the selected measur­ing range.
One function of a stroboscope is to provide a “stopped” image of a rotating target when the flash rate of the stroboscope has matched the rotational speed of the target object. The strobo­scope will show a single image when the flash rate is set to a lesser multiple of the true RPM (1/2,1/3,etc.) When the flash rate is increased to a higher multiple (2,3,etc.), multiple images will appear. To find the true RPM of the target object, reduce the flash rate to a lower multiple until only one image appears.
Batter capacity is full
The Battery capacity is diminished Near zero capacity
When battery is empty, the unit will power off
Remarks: It can be used for more than 12 hours continuously when the frequency is 3000FPM or 50Hz.
2