Shimpo DT-311A Instruction Manual

DT-311A & DT-315A Stroboscope Operation Manual
z
The DT-311A/DT-315A heavy duty stroboscopes, provide the il-
lusion of “stopped motion”
equipment appear to be standing still or in slow motion. The DT-
311A/DT-315A also measure rotational (RPM) or reciprocating
(strokes per minute) speeds with the same precision as an elec-
tronic digital tachometer. Ideally suited for printing inspection, mo-
repetitive motion, DT-311A/DT-315A strobes provide timely data
on vibrating parts, leaks, spray patterns and a variety of industrial
and scientific operations the DT-311A/DT-315A stroboscopes are
extremely easy to maneuver and operate via its lightweight and
compact design and simplified user interface. Its rugged metal
construction provides unequaled reliability in the harshest envi-
ronments.
SET-UP The DT-311A and DT-315A may be operated handheld or else mounted on a tripod or swing armfor added convenience.
To mount the strobe on a tripod (or any other mounting surface), use screw ¼ -20 UNC, length 0.3” (8 mm) or shorter for the tripod screw hole on the bottom.
Do not operate or store instrument in the follow­ing places: explosive areas; near water, oil, dust, or chemicals; areas where temperature is above 104°F (40°C).
Do not look at the emitted light for long periods of time; it can be harmful to the eyes.
Do not disassemble or repair unit while in operation.
To prevent getting burned, let lamp cool after use be­fore attempting replacement.
by adjusting the strobe’s flash rate,
DT-311A
DT-315A
SPECIFICATIONS Flashing Range: 40.0 - 35,000 fpm (flashes per minute) Accuracy: ±0.01 % of reading Resolution: 0.1 fpm: 40.0 - 5,000 fpm / 0.2 fpm: 5,001 - 8,000 fpm /
0.5 fpm: 8,001 - 10,000 fpm / 1 fpm: 10,001 - 35,000 fpm
Phase Shift: Use +/- push buttons (360° in 6 seconds) Display Update Time: 0.2 sec approx Output Signal:Synchronous, 400 μsec. Pulse output, 0 to +12
VDC amplitude(approx.), 4.7 KΩ impedance
Rate Multiplier/Divider: Multiply by 2, divide by 2 Flashing Range: Synchronous Mode: 60.0 ~ 3,000FPM / Delay
mode:200 ~ 3,000 FPM
External Mode Accuracy: ±0.01 % ±1 digit Phase Shift: 0 to 359° with 1° resolution Delay Time: 0 to 2,000 mSec from 40 - 10,000 fpm External Trigger Input Signal: LO level: 0 -0.8 VDC, HI level: 2.5
- 12 VDC or open collector (NPN),pulse width 50 μsec min.
Input Impedance: 4.7 KΩ to 12 V / 6.8 KΩ to 0 V Display: 5 digits, 0.4” (10 mm) high, LED Flashtube Power/Life: Xenon, 10 W max. (100 million flashes) Flash Duration: 10 to 40 μs automatically adjusts with flash rate Lux Rating: (20 cm) Distance = 4500 lx @ 1500 fpm, 3375 lx @
6000 fpm with 4.5” (110 mm) irradiation diameter. 20” (50 cm) Distance = 1300 lx @ 1500 fpm, 930 lx @ 6000 fpm with 10” (250 mm) irradiation diameter.
Sensor Power Supply: 12 VDC (40 mA) Battery Life: (DT-315A)1 - 2 hours depending on flash rate Power Requirement: DT-311A: 115 VAC or optional 220 VAC
±10 % 60/50 Hz, 30 VA (specify voltage); DT-315A: Rechargeable Battery or AC Line Power with included AC adapter
Operating Temperature: 32 - 104°F ( 0 - 40°C ) Product Weight: DT-311A: 3.35 lbs (1.52 kg); DT-315A: 4.4 lbs (2
kg) Package Weight: DT-311A: 4.4 lbs (2 kg), DT-315A: 5.5 lbs (2.5 kg) Dimensions: DT-311A: 7.28” L x 4.72” W x 4.72” H (185 mm x 120 mm x 120 mm); DT-315A: 9.84” L x 4.72” W x 4.72” H (250 x 120 x 120 mm)
Warranty: 2 year Certifications: CE Included Accessories: Handle, flash tube removal tool. AC
Charger/Adapter (DT-315A only)
OperationDisplay Panel
True RPM Measurement
To measure true revolutions per minute (RPM):
1. “Mark” the object to be measured by either visually noting an inherent distinguishing characteristic (such as a label,
scratch, etc.) or physically marking the object with a small piece of tape, pencil mark, etc.
2. Firmly plug in power cord.
3. Turn power switch on.
4. Turn setter from highest FPM downward.
1
4
2
3
5. The true RPM can be noted once the action appears frozen and the first single image of the “mark” appears (see chart
below and accompanying diagram for further explanation).
6. To verify RPM reading, press “÷2”; a single image should appear again.
1. LED
display: Displays function and value
2. EXT: External mode indicator
3. INT: Internal mode indicator
4. B-CH: Battery charge indicator (DT-315A only)
5. FPM: Flash per minute indicator
6. deg: Phase shift degree indicator
7. mSec: Millisecond delay time indicator
8. Signal switch :Switches the unit from the external mode to the internal mode (and vice-versa)
9. Display mode switch: When unit is set to the external mode, the strobe will switch to RPM (FPM)/deg/msec
each time “MODE” is depressed
RPM (FPM) Displays flashes per minute
External input 0-35,000 RPM (FPM)
deg Displays flash delay in degrees
msec Displays flash delay in msec
10. Setter: Changes the flashing rate
11. (x2) Switch: In the internal mode, pressing “x2” doubles the flashing rate
Shaft Rotation Flashes Flashes/ Stopped (RPM) (RPM) RPM Shaft Images
6,000 4 times 4
4,500 3 times 3
At 1,500 RPM
3,000 2 times 2
1,500 1 time 1
750 1/2 time 1
500 1/3 time 1
12. (+) Switch: In the internal mode, when object appears to be standing still, pressing “+” will give the illusion
that the object is moving towards the rotating direction at a speed of 1 rotation in 6 seconds
13. ( ÷2) Switch: In the internal mode, pressing “÷ 2” divides the flashing rate by two
14. (-) Switch: In the internal mode, when object appears to be standing still, pressing “-“ will give the illusion
that the object is moving in reverse at a speed of 1 rotation in 6 seconds
15. Input and output connector: PIN #1: +12V
PIN #2: Synch output signal PIN #3: Input signal
PIN #4: 0V
16. Power cord (DT-311A) / AC adapter (DT-315A)
17. Power switch
Internal Triggering Mode
To operate the stroboscope in internal triggering mode:
1. Firmly plug in power cord.
2. Turn power switch on.
3. If internal indicator is not on, press “SIG”; the INT light will then turn on.
4. Aim light beam at object under observation. The optimal distance between the strobe and moving object is
approximately 2 feet.
5. Measure RPM by turning the setter to adjust the flashing rate to the rotational speed of the object.
NOTE: To achieve a particular rate quickly, use the “x2” or “÷2” switches and then the setter for fine tuning.
NOTE: Once the internal timer has expired, the strobe will stop flashing and the display will flash rapidly. To restart the strobe,
turn power switch off, then on, and the cycle will repeat.