Congratulations on your purchase of a Shimpo DRI (Digital
Remote Indicator). We trust you will enjoy many years of
professional results from your DRI when used in conjunction with
a Shimpo Remote Load Cell or Remote Torque Sensor for force
and/or torque measurement applications.
Please read the entire instruction manual thoroughly before initial
set-up and operation; the information contained herein will aid
you in operating your Shimpo DRI safely and with excellent results.
If you have any questions regarding our product(s), call your local
Shimpo representative or contact Shimpo directly for assistance.
External Control Input14-15
Overload Output16
Parameter Settings17
Low Battery Indication18
Auto Power Off18
Calibration18-19
Troubleshooting19
Dimensions and Specifications20
Warranty20
Important Safety InstructionsInspection/Standard Accessories
If any damage is apparent, do not unpack the DRI. Notify the
shipping carrier immediately for damage claim instructions.
Refer to the nameplate to confirm model number ordered and
record serial number for future reference.
Items included with the DRI are:
• AC adapter
• (4) AA batteries
• DRI output connector
• Overload protection cable
• Plastic carrying case
Features and Benefits
The DRI offers many features and benefits, including:
• Enables easy set-up when changing applications and
reduces investment in gauges and/or torque sensors by
accepting multiple interchangeable load cells/torque
sensors
• Large memory capacity (100 samples plus statistics)
enables portable data collection for multi-location usage
• High sampling rate (1000 hz) allows peak force and tor que
points to be easily captured
• Handheld and test stand mountable - allows flexibility in
multiple test set-ups, applications and fixturing
• High and low set points permit measurement of br eakaway
torque in applications such as child safety cap testing
• External tare and hold control allows hands-free zeroing
using a keyboard or foot switch
Always wear eye protection when testing materials.
Never use a remote sensor that appears to be damaged
in conjunction with the DRI.
Confirm that the DRI is turned OFF when connecting
or disconnecting a remote sensor; failure to do so may
damage the DRI.
Attachments must be properly installed to the remote
sensors (hand-tighten only! Use only the Jacob’s Chuck
- provided with torque sensor - to tighten the torque
sensors. Do not use wrenches or other tools!).
Never exceed remote sensor capacity; if “OVR 1234”
appears on the display, STOP the test immediately to
avoid permanent sensor damage.
Never apply a force or load at an angle to the remote
sensor.
When installing the DRI on a test stand, confirm that
the indicator unit is thoroughly secured.
When not in use over an extended period of time,
remove batteries to avoid battery terminal corrosion.
Only use the provided AC adapter; usage of any other
adapter may damage circuitry.
Check the batteries periodically for corrosion when
using the unit under AC power; this will avoid damage
and ensure memory backup (refer to “Low Battery
Indication” section).
Do not use or store unit in extreme temperatures;
normal operating temperature is 32-113°F (0-45°C).
Do not use or store unit in oily, dusty, or excessively
wet areas.
Do not use any chemicals to clean the case; use a wet,
soapy cloth.
2
Set-Up
Power
Every DRI unit is shipped from the factory without the batteries (provided) installed. To install the batteries:
1. Remove the battery plate located at the bottom of the handle using a coin or appropriate screwdriver.
2. Install the (4) AA batteries, observing the proper polarity.
3. Reinsert the battery plate cover, making sure the lip engages properly before tightening.
NOTE: If AC adapter operation is preferred, refer to the “Display & Keypad” section of this manual for AC outlet location.
Connection/Disconnection of DRI to Load Cells (LC) or Torque Sensors (TS)
CAUTION: Always confirm that the DRI’s power is turned OFF when connecting or disconnecting an LC or TS.
Instructions to connect a remote sensor (LC or TS) are as follows:
1. Remove the plastic cover from the DRI’s remote sensor port.
2. Align the arrows from the remote sensor’s plug with the arrows on the port of the DRI.
3. Push the plug straight onto the port until a “click” is heard.
4. Press the DRI’s ON button; the LCD will briefly display all indicators (see “Display & Keypad” section).
5. The DRI will briefly display what type of remote sensor is connected (“SEN” will be flashing on the small display). The abbreviations
on the main display are as follows. Please call Shimpo for information on additional sensors.
Load Cells (LC)
PART NUMBERMAIN DISPLAY
LC-50S50L
LC-100S100L
LC-200S200L
LC-500S500L
Instructions to disconnect a remote sensor (LC or TS) are as follows:
1. Press the DRI’s OFF button.
2. Press and hold the gray release button located towards the top of the remote sensor’s plug.
3. Pull the plug (not the cord) straight off the DRI’s port.
4. Connect another remote sensor (see instructions above) or replace the protective plastic cover on the exposed port of the DRI.
3. Peak Indicator (NOTE: Refer to “Changing the Measuring Mode” section for explanation)
4. Main Display
5. Small Display
6. Maximum Reading Indicator
7. Minimum Reading Indicator
8. AC Adapter Port
9. > Button
10. HI-LO Button
11. Unit Button
12. Peak Button (NOTE: Refer to “Changing the Measuring Mode” section for explanation)
13. Recall Button
14. Memo Set Button
15. Reset Button
16. On Button
17. Off Button
18. Overload Output Port
19. Data Output Port
20. Battery Compartment
4
Factory Settings
Each time the power is turned OFF and ON there are some parameters that the DRI will default to. Consult the section on “Parameter
Settings” to customize these default settings. The following chart reflects the default settings as programmed by the factory:
PARAMETERDEFAULT FACTORY SETTING
Display Update0.125 seconds
Measuring ModeTrue Peak
Auto Power Off3 Minutes
Measurement UnitEnglish (Lb, Lb-I or oz-I)
OutputRS232
Baud Rate - 4800
Data Length - 8 bits
Stop Pulse - 2 bits
Data End Code - CrLF
Software Flow Control - off
External TriggerTare
Memory ModeSingle
Analog Output Mode On
FilterOff
Operation
How to Tare and Zero the Display
After installing an attachment, it will be necessary to tare the DRI:
1. Press the RESET button while holding down the ON button to tare the display .
2. Press the RESET button by itself when the peak indicator is showing to zero the display . Repeat step 1 if the display does not return
to zero.
NOTE: The DRI can tare up to 50% of the remote sensor’s rated capacity.
Changing the Measuring Mode: True Peak Mode/Standard Mode
CAUTION: Depending on which of these two measuring modes is selected, the PEAK button performs two different functions! The DRI
will alway s display an average force when PEAK is not showing on the display. The average measurement is based upon the display
update rate chosen (see “Changing Display Update Time” section). When in the True Peak mode, pressing the PEAK button will allow
the display to scroll from average to peak “+” (compression or clockwise torque) to peak “-” (tension or counterclockwise torque). When
in the Standard mode, pressing the PEAK button will allow the display to scroll from average to maximum display reading “+” to
maximum display reading “-”. Despite showing PEAK on the display, these are not true peak readings; they are the highest averagedisplayed readings since the DRI was last “tared”.
Instructions to change the Measuring Mode are as follows:
1. Press the PEAK button while holding down the ON button. The small display will exhibit “dSP”.
3. Press the PEAK button; the small display will read “PfS” and the main display will show “on” (T rue Peak) or “oFF” (Standard).
3. Press the > button to toggle between “on” and “oFF”.
4. Press MEMO SET button to store the measuring mode and exit.
NOTE: By keeping the true peak mode “ON”, the DRI will always display the peak samples and not a maximum average.
5
Operation (continued)
Changing Units of Measure
By pressing the UNIT button, the DRI will scroll between English, Metric, and Newton units; once the DRI is turned OFF the unit will
default to the factory setting. Instructions to change the default unit of measure setting are as follows:
1. Press the PEAK button while holding down the ON button; the small display will indicate “dSP”.
2. Press the PEAK button three times; the main display will indicate “init” and the small display will reflect the current default setting.
3. Each time the UNIT button is pressed, the small display will scroll between English, Metric, and Newton units.
4. Press the MEMO SET button to store the default unit setting and exit.
Setting High and Low Limits
It is possible to program high and low limits for applications requiring tolerances. Use the diagram below as a guide for setting the limits.
Instructions to set the high and low limits are as follows:
1. Press the HI-LO button; “∆HI” will be flashing on the small display and whatever limit is already programmed will be displayed.
2. Press the > button; the entire display will be flashing.
3. Press the HI-LO button to toggle between “+” (compression or clockwise torque) and “-” (tension or counter-clockwise torque).
4. Press the > button to scroll across each digit position.
5. Press the HI-LO button to select the numerical value of each digit position.
6. Once the high limit has been programmed, press the MEMO SET button; “∇LO” will be flashing on the small display and whatever
limit is already programmed will appear on the main display.
7. To program the low limit, repeat steps 2-5.
8. Once the low limit has been programmed, press the MEMO SET button to store the set points and exit from this function.
The unit of measure will be displayed when the measurement is within the set points. Small “arrows” (
display when the measurement is above or below the set points.
The unit of measure will never be displayed if the low limit is set higher than the high limit. This is because it is impossible for the
measurement to fall between the setpoints.
Set points can be referenced by pressing the HI-LO button: the first time the button is pressed the display will show the high limit, the
second time will show the low limit, and the third time will exit.
To cancel the high and low limits, both settings must be set to zero.
NOTE: A comparator output is available for “go – no go” testing (refer to “Comparator Output” section).
or ∇) will appear on the small
∆
Changing Display Update Time
The display update time works with the sampling rate (1000 hz) to provide an average reading. All samples taken within the selected
update time are averaged (based on 1000 samples/sec) and displayed.
To change display update time:
1. Press the PEAK button while holding down the ON button. The small display will exhibit “dSP” and the main display will reflect
the current setting.
2. Each time the > button is pressed, the main display will scroll through “0.125”, “0.25”, “0.5”, “1.0” and “2.0” (these correspond
with the amount of time in seconds that the display will update).
3. Press the MEMO SET button to store the desired display update time and exit.
6
Loading...
+ 14 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.