Inspection / Standard Accessories2
Features and Benefits2
Important Safety Instructions2
Set-Up3
Power3
Display and Keypad4
Factory Settings5
Operation5-8
How to Tare and Zero the Display5
Changing the Measuring Mode5
Changing Units of Measure6
Setting High and Low Limits6
Changing Display Update Time6
Changing Memory Modes7
Recalling Statistics8
Setting of Filter8
Communication with External Devices9-16
Output Selection9-13
RS23210-12
Digimatic13
Analog13
Comparator13
External Control Input14-15
Congratulations on your purchase of a Shimpo DFS (Digital
Force Gauge). We trust you will enjoy many years of
professional results from your DFS for force measurement
applications.
Please read the entire instruction manual thoroughly before
initial set-up and operation; the information contained
herein will aid you in operating your Shimpo DFS safely and
with excellent results.
If you have any questions regarding our product(s), call your
local Shimpo representative or contact Shimpo directly for
assistance.
Overload Output16
Parameter Settings17
Low Battery Indication18
Auto Power Off18
Calibration18-19
Troubleshooting19
Dimensions and Specifications20
Warranty20
If any damage is apparent, do not unpack the DFS. Notify
the shipping carrier immediately for damage claim
instructions. Refer to the nameplate to confirm model
number ordered and record serial number for future
reference.
Items included with the DFS are:
• AC adapter
• (4) AA batteries
• DFS output connector
• Overload protection cable
• Plastic carrying case
Important Safety InstructionsInspection/Standard Accessories
Always wear eye protection when testing
materials.
Attachments must be properly installed to the
force gauges (hand-tighten only! Do not use
wrenches or other tools!).
Never exceed force capacity; if “OVR 1234”
appears on the display, STOP the test immediately
to avoid permanent sensor damage.
Features and Benefits
The DFS offers many features and benefits, including:
• Large memory capacity (100 samples plus statistics)
enables portable data collection for multi-location
usage
• High sampling rate (1000 hz) allows peak force and
torque points to be easily captured
• Handheld and test stand mountable - allows flexibility
in multiple test set-ups, applications and fixturing
• High and low set points permit measurement of
breakaway torque in applications such as child safety
cap testing
• External tare and hold control allows hands-free
zeroing using a keyboard or foot switch
Never apply a force or load at an angle to the
force sensor.
When installing the DFS on a test stand, confirm
that the indicator unit is thoroughly secured.
When not in use over an extended period of
time, remove batteries to avoid battery terminal
corrosion.
Only use the provided AC adapter; usage of any
other adapter may damage circuitry.
Check the batteries periodically for corrosion
when using the unit under AC power; this will
avoid damage and ensure memory backup (refer
to “Low Battery Indication” section).
Do not use or store unit in extreme temperatures;
normal operating temperature is 32-113°F (045°C).
Do not use or store unit in oily, dusty, or
excessively wet areas.
Do not use any chemicals to clean the case; use
a wet, soapy cloth.
2
Set-Up
Power
Every DFS unit is shipped from the factory without the batteries (provided) installed. To install the batteries:
1. Remove the battery plate located at the bottom of the handle using a coin or appropriate screwdriver.
2. Install the (4) AA batteries, observing the proper polarity.
3. Reinsert the battery plate cover, making sure the lip engages properly before tightening.
NOTE: If AC adapter operation is preferred, refer to the “Display & Keypad” section of this manual for AC outlet
location.
3
Display & Keypad
1. Sensor Shaft
2. Low Battery Indicator
3.Peak Indicator (NOTE: Refer to “Changing the Measuring Mode” section for explanation)
4. Main Display
5. Small Display
6. Maximum Reading Indicator
7. Minimum Reading Indicator
8.AC Adapter Port
9. > Button
10. HI-LO Button
11. Unit Button
12. Peak Button (NOTE: Refer to “Changing the Measuring Mode” section for explanation)
13. Recall Button
14. Memo Set Button
15. Reset Button
16. On Button
17. Off Button
18. Overload Output Port
19. Data Output Port
20. Battery Compartment
4
Factory Settings
Each time the power is turned OFF and ON there are some parameters that the DFS will default to. Consult the section
on “Parameter Settings” to customize these default settings. The following chart reflects the default settings as
programmed by the factory:
PARAMETERDEFAULT FACTORY SETTING
Display Update0.125 seconds
Measuring ModeTrue Peak
Auto Power Off3 Minutes
Measurement UnitEnglish (Lb, Lb-I or oz-I)
OutputRS232
Baud Rate - 4800
Data Length - 8 bits
Stop Pulse - 2 bits
Data End Code - CrLF
Software Flow Control - off
External TriggerTare
Memory ModeSingle
Analog Output Mode On
FilterOff
Operation
How to Tare and Zero the Display
After installing an attachment, it will be necessary to tare the DFS:
1. Press the RESET button while holding down the ON button to tare the display.
2. Press the RESET button by itself when the peak indicator is showing to zero the display. Repeat step 1 if the display
does not return to zero.
NOTE: The DFS can tare up to 50% of the remote sensor’s rated capacity.
Changing the Measuring Mode: True Peak Mode/Standard Mode
CAUTION: Depending on which of these two measuring modes is selected, the PEAK button performs two different
functions! The DFS will always display an average force when PEAK is not showing on the display. The average
measurement is based upon the display update rate chosen (see “Changing Display Update Time” section). When in the
True Peak mode, pressing the PEAK button will allow the display to scroll from average to peak “+” (compression or
clockwise torque) to peak “-” (tension or counterclockwise torque). When in the Standard mode, pressing the PEAK
button will allow the display to scroll from average to maximum display reading “+” to maximum display reading “-”.
Despite showing PEAK on the display, these are not true peak readings; they are the highest average displayed
readings since the DFS was last “tared”.
Instructions to change the Measuring Mode are as follows:
1. Press the PEAK button while holding down the ON button. The small display will exhibit “dSP”.
3. Press the PEAK button; the small display will read “PfS” and the main display will show “on” (True Peak) or “oFF”
(Standard).
3. Press the > button to toggle between “on” and “oFF”.
5
Operation (continued)
4. Press MEMO SET button to store the measuring mode and exit.
NOTE: By keeping the true peak mode “ON”, the DFS will always display the peak samples and not a maximum average.
Changing Units of Measure
By pressing the UNIT button, the DFS will scroll between English, Metric, and Newton units; once the DFS is turned OFF
the unit will default to the factory setting. Instructions to change the default unit of measure setting are as follows:
1. Press the PEAK button while holding down the ON button; the small display will indicate “dSP”.
2. Press the PEAK button three times; the main display will indicate “init” and the small display will reflect the current
default setting.
3. Each time the UNIT button is pressed, the small display will scroll between English, Metric, and Newton units.
4. Press the MEMO SET button to store the default unit setting and exit.
Setting High and Low Limits
It is possible to program high and low limits for applications requiring tolerances. Use the diagram below as a guide for setting the
limits.
Instructions to set the high and low limits are as follows:
1. Press the HI-LO button; “∆HI” will be flashing on the small display and whatever limit is already programmed will
be displayed.
2. Press the > button; the entire display will be flashing.
3. Press the HI-LO button to toggle between “+” (compression or clockwise torque) and “-” (tension or counterclockwise torque).
4. Press the > button to scroll across each digit position.
5. Press the HI-LO button to select the numerical value of each digit position.
6. Once the high limit has been programmed, press the MEMO SET button; “∇LO” will be flashing on the small
display and whatever limit is already programmed will appear on the main display.
7. To program the low limit, repeat steps 2-5.
8. Once the low limit has been programmed, press the MEMO SET button to store the set points and exit from this
function.
The unit of measure will be displayed when the measurement is within the set points. Small “arrows” (
appear on the small display when the measurement is above or below the set points.
The unit of measure will never be displayed if the low limit is set higher than the high limit. This is because it is
impossible for the measurement to fall between the setpoints.
Set points can be referenced by pressing the HI-LO button: the first time the button is pressed the display will show the
high limit, the second time will show the low limit, and the third time will exit.
To cancel the high and low limits, both settings must be set to zero.
∆
or
∇
) will
NOTE: A comparator output is available for “go – no go” testing (refer to “Comparator Output” section).
6
Loading...
+ 14 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.