SHIMPO DAG-2000 User Manual

DIGIT AL AIR MICROMETER
DIGITALAIRMICROMETER
DAG2000 USERS MANUAL
Second Edition 2002.10.25
NIDEC TOSOK CORPORA T ION
1
REQUEST DURING USE In order to use this product safely, please follow the items below:
[Danger] (1) The inside of this product has hundreds of volts of electric voltage.
During installation, transfer, maintenance, and inspection of this product or connections, please shut down all the power and remove from the product the power and connection cables first.
(2) Standard attached power cable is for 100V. When using power voltage exceeding 125V, please prepare
separate 250V cable.
(3) Please do not trample or pull the connection cables that will be used for connecting the power cable to
the equipment. Also, when removing power cable or connection cables, please be sure to hold the plug part. There is danger of damage to the cable. By any means please do not use damaged power cables or connection cables. This is a high voltage object and there is danger of electrical shock.
(4) Please make sure to install ground.
There is danger of electric shock during breakdown or short circuit.
[Warning] (1) This product is a precise measuring device that performs measurement through air. For the air supply,
please prepare pure air with dirt, moisture and oils removed.
(2) Please make sure not to put foreign materials through the gaps of this product or the connecting
machines. There is danger of incurring serious injury from electric shock, fire and breakdown.
[On bringing to overseas]
When bringing this product overseas, please confirm beforehand since there are various restrictions. Please understand beforehand that when this product is brought overseas and accidents occur, this company will be held responsible.
[About this manual] (1) This manual was created doubly sure but if by chance suspicious points, omissions and such are noticed,
please inquire with our sales department. (2) Diversion or reproduction of all or part of the contents without permission is not allowed. (3) For purposes of improvements, changes can be made to product specifications, contents of the manual,
and appearance without notice in the future.
[About the warranty]
Warranty will be in accordance with the warranty regulations of this company
(1) Even during warranty period, expenses required for repairs will be charged if breakdown and damages
are caused by mishandling by the customer. (2) There may be cases also when repairs for alteration made by customers cannot be accepted. (3) Claims repair is standard. For business trip repairs, separate business trip expenses will be charged
even during warranty period.
2
TABLE OF CONTENTS
CHAPTER 1 INTRODUCTION-------------------------------------------------------------------------------------1
1.1 Outline -------------------------------------------------------1
1.2 Merits ------------------------------------------------------ 1
1.3 Block Diagram ------------------------------------------------------ 2 CHAPTER 2 NAME AND FUNCTION OF EACH PART------------------------------------------------------ 3 CHAPTER 3 PROCEDURE UP TO MEASUREMENT--------------------------------------------------------- 4
3.1 For normal cases ------------------------------------------------------- 4
3.2 For work tolerance (judgment limits) change cases ------------------------------------------------------- 5
3.3 For first time measurement tool connection cases ------------------------------------------------------ 6
3.4 For several measurement tools connection cases ------------------------------------------------------- 7 CHAPTER 4 MASTER CALIBRATION --------------------------------------------------------------------------8
4.1 Master calibration by 2 masters ------------------------------------------------------ 8
4.2 Master calibration by 1 master -----------------------------------------------------10
4.3 Clearing of master data ------------------------------------------------- ------11
4.4 Master calibration error -------------------------------------------------------11 CHAPTER 5 MEASUREMENT------------------------------------------------------------------------------------12
5.1 Measurement value display -------------------------------------------------------- 12
5.2 Display color of main display ----------------------------------------------------- 12
5.3 Measurement value hold -------------------------------------------------------12 CHAPTER 6 DETAILS OF EACH PROCEDURE--------------------------------------------------------------13
6.1 Installation -------------------------------------------------------13
6.2 Preparation ------------------------------------------------------ 13
6.3 Starting ------------------------------------------------------ 14
6.4 Settings -----------------------------------------------------15
6.5 Detector Adjustment ------------------------------------------------------21
6.6 Program Switching ------------------------------------------------------24 CHAPTER 7 EXTERNAL I/O FUNCTION--------------------------------------------------------- ------------25
7.1 Serial (RS232C) communications function ------------------------------------------------------ 25
7.2 External button input -----------------------------------------------------26 CHAPTER 8 MAINTENANCE------------------------------------------------------------------------------------ 27 CHAPTER 9 CAUSES OF FAILURE AND COUNTERMEASURES --------------------------------------27 CHAPTER 10 OTHERS--------------------------------------------------------------------------------------------- 28
10.1 Model -----------------------------------------------------28
10.2 Optional ----------------------------------------------------- 28
10.3 Specifications -----------------------------------------------------29 CHAPTER 11 WORKSHEET
--------------------------------------------------------------------------------30
3
CHAPTER 1 INTRODUCTION
Thank you for purchasing the Digital Air Micrometer (DAG2000). This manual is explained such that even first time users of the Air Micrometer will be able to use the abundant functions of this product efficiently. Please read the manual thoroughly to use this product well.
1.1 Outline
This product is a measurement device that detects the air pressure changes based on the dimensions of the measurement subject through sensors, and makes judgments based on digital dimension indicators and lamps.
This product is a comparison measurement device.
Measure the master for reference and then do measurements. With regular master calibrations, good precision measurements are possible.
This product judges independently if the measurement value is OK or NG.
Judgment result display can be quickly distinguished through the indicator color of the main display and judgment LED. Also, regular master calibrations can be done through the panel key.
1.2 Merits
① Judgment result (OK or NG) of measurement value can be distinguished through the number and
lamp color.
② Measurement value unit display can be switched (µm or mm) by changing the settings. ③ Master calibration is easy. ④ Peak measurement [Optional]
+PEAK, -PEAK, TIR(=+PEAK-(-PEAK)), TIR/2(=(+PEAK-(-PEAK))/2)
Auto measurement start stop function
⑤ 5 ranks (-NG, -OK, OK, +OK, +NG) judgment is possible
Maximum of 99 ranks judgment is possible
⑥ Abundant external I/O functions
Serial communications function
To PC printer via RS232C Measurement value and judgment result can be outputted. [Standard]
External button input function [Standard]
Measurement command and master calibration command input is possible via foot switch and push button.
Digimatic output function [Optional]
Capable of outputting measurement value to Digimatic printer.
DC I/O function
Capable of outputting judgment result to the lamp sequencer and also BCD output of the measurement value.
⑦ Compact
120(W) X 180 (D) X 150(H)mm, 300(D)mm when regulator is fitted
⑧ Accessories
Handle for easily carrying [Optional]
Tuner (dial) cover [Optional]
Cable for each type of external I/O [Optional]
4
1.3 Block Diagram
d
t
p
(1) Structure
Arrow represents the direction of air and electric signal.
is optional.
Measurement element/tool
Work and master
RS232C Digimatic output External input button
A/E Converter
Air circuit
Pressure sensor
Amp circuit
Input Range switch
CPU
Memory for operation software
Regulator
0.196±0.005MPa
0.3~0.7MPa
Mist separator
Filter
Parallel I/O
Memory for saving settings values and master calibration data (good for 10 types)
Optional
DC I/O base
In case industrial
ressure contains
moisture and oil
(2) Operation flow
Shown below is the software operations diagram starting from turning on power.
Power ON
Initialization mode
Reading of last setting values use
and master calibration data
Confirmation of indicator ligh (visual)
Master calibration mode
Master calibration
Adjustment of detector
Master calibration data clear
Measurement mode Program switching mode
Program switching
Setting mode
Measurement range selection
Input of master value
Input judgment limits
Can save up to 10 kinds of settings values
Compressed air
purifier
Industrial pressure
5
d
y
t
d
CHAPTER 2 NAME AND FUNCTION OF EACH PART
Sensitivity (MAG) adjustment tuner
Adjusts the sensitivity of the air circuit. No need for adjustments except during measurement tool replacement and master calibration error.
Zero adjustment tuner Adjusts the zero position of the air circuit. No need for adjustments except during measurement tool replacement and master calibration error.
RS232C Connector
Connector for serial communications used for connecting the PC to the printer.
Switch input connector Connector for measurement comman and master calibration input done b external push button. Optional foo switch and push button is available. Cannot be used expect for foot switch and push button.
DC input connector [Optional] Connector for connecting to the lamp, PLC, etc. There are 2 types judgment output and BCD output. Capable of measurement command an master calibration input.
Analog input connector [Optional] Please inquire to the maker on how to use.
Mode lamp
Displays the current mode and master calibration condition.
Main display 4 digits 3 colors digital display. Displays the measurement value and judgment result (rank no.).
Judgment lamp Displays the judgment results.
English display Displays simple bar, measurement value, settings item name, settings value, etc.
Panel switch Used for mode switching, settings item selection, settings value input, etc.
Air pressure input
Please supply pure air at 0.196±0.005MPa. Optional precision regulator and filter is available.
Power switch
Silencer
Power connector (Input)
Can be used at AC85~264V range but for the cable included, please use within AC85~125V.
Measurement nozzle port Please connect measurement tool.
Regulator bracket Hole for fitting (M3)
6
CHAPTER 3 PROCEDURE UP TO MEASUREMENT Procedures up to the measurement corresponding to each condition are mentioned in this chapter.
3.1 For normal cases Shown here is the procedure up to measurement for normal cases.
Turn ON compressed air
Supply of compressed air to the main body
Please refer to [Chapter 6, 6.2 Preparation] for details
Turn ON power
Supply of power to the main body Please refer to [Chapter 6, 6.3 Starting] for details
Start at measurement mode Please refer to [Chapter 6, 6.2 Preparation] for details
Switch to master calibration
mode (master calibration)
Switch from measurement mode to master calibration mode Please refer to [Chapter 4 Master Calibration] for details
Master calibration
Record the master value to the main body Please refer to [Chapter 4 Master Calibration] for details
Please perform master calibration regularly (every 2 hrs.).
Please refer to Chapter 4 Master calibration for details
After finishing master calibration, switch to measurement mode.
Measurement is possible
Measurement
Please refer to [Chapter 5 Measurement] for details
7
3.2 For work tolerance (judgment limits) change cases
t
Shown here is the procedure up to measurement wherein work tolerance is changed.
Turn on compressed air
Turn ON power
Switch to settings mode
Change settings
Switch to master
calibration mode (master
calibration)
Switch from measurement mode to settings mode
Because of work tolerance change, change the judgmen limit values. Settings value change can be done in setting mode.
Please refer to [Chapter 6, 6.4 Settings] for details.
Switch to measurement mode after finishing settings change
Please refer to [Chapter 6, 6.4 Settings] for details.
Change the judgment limit values at settings mode
Please refer to [Chapter 6, 6.4 Settings] for details.
Master calibration
Measurement
8
3.3 For first time measurement tool connection cases
t
r
r
r
Shown here is the procedure up to measurement wherein measurement tool is connected to the main body for the first time. Note: There are cases wh erein detector adjustment is not needed depending on the measurement tool.
Turn ON compressed air
Turn ON power
Switch to settings mode
Switch from measurement mode to settings mode
Because of work tolerance change, change the judgmen limit values. Settings value change can be done in setting mode.
Please refer [Chapter 6, 6.4 Settings] for details
Change the settings
At settings mode, input the settings value for the
measurement tool that will be connected. Please refer to [Chapter 6, 6.4 Settings] for details
Switch to measurement mode after finishing settings
change Please refer to [Chapter 6, 6.4 Settings] for details
Switch to master calibration mode
(detector adjustment)
Switch from measurement mode to detector adjustment
Please refer to [Chapter 6, 6.5 Detector Adjustment] fo details
Detector adjustment
Adjustment of detector through ZERO and sensitivity
tuner. Please refer to [Chapter 6, 6.5 Detector Adjustment] fo details
Switch to master
calibration mode (master
calibration)
Master calibration
Measurement
Switches from detector adjustment to master calibration
Refer to [Chapter 6, 6.5 Detector Adjustment] fo details.
9
3.4 For several measurement tools connection cases
r
d
r
m
r
Shown here is the procedure up to measurement for cases wherein multiple measurement tools are connected to 1 main body.
Please refer to [Chapter 3, 3.3
For first time measurement tool connection cases] when inputting of
settings for measurement tools that will be connected are not finished.
Replace the measurement tool
Turn ON compressed air
Turn ON power
Switch to program switching
mode
Program switching
Stopping of compressed air to the main body and replace
measurement tool. Please refer to [Chapter 6, 6.2 Preparation] for details
Switch from measurement mode to program switching
mode Please refer to [Chapter 6, 6.6 Program Switching] fo details
At program switching mode, reads the settings value an
master data of the measurement tool connected. Please refer to [Chapter 6, 6.6 Program Switching] fo details
Switch to master calibration
mode (detector adjustment)
Adjustment of detector
Switch to master calibration
mode (master calibration)
Master calibration
Switch to measurement mode after finishing progra
switching. Please refer to [Chapter 6, 6.6 Program Switching] fo details
Measurement
10