If
plug
with
cord
completely
CAUTION:
cleaning
and
isopropyl
from
If
cleaning
with
isopropyl
removed.
Avoid
as
of
the
A.C.
receptacle
alcohol.
the
pump
dry.
of
the
exposing
this
can
power
prior
A.C.
alcohol;
lead
cord
is
necessary,
plug
section
If
the
pump
to
cleaning.
receptacle
upon
completion
A.C.
receptacle
to
an
electrical
wipe
of
the
the
power
has a detachable
Do
not
reconnect
is
necessary,
ensure
and
shock
use a cotton
that
power
or
exterior
cord
with a cloth
power
power
all
residue
cord
fire
hazard.
surfaces
cord,
cord
swab
and
to
excess
of
the
wall
dampened
disconnect
to
pump
dampened
liquid
is
moi
‘ire,
the
until
Functional
Perform
Feeding
1.
the
Pump
Sensor
This
test
and
the
a.
Phototransistor
at
the
[note:
and
Safety
listed
Service
Circuit
must
pump
positioned
collector
adjust
tests
be
performed
bias
of
R42
Testing
at
each
Manual”
servicing,
as
with
face
voltage
the
phototransistor
as
needed]
referring
necessary.
no
drip
down away
(0.99
to
chamber
from
1.01
to
direct
VDC)
NUS.
your
“Kangaroo
installed
lighting.
K224
X
Com
224/324
K324
DIRECTIONS
Refer
Clean
dampened
Be
DIRECTIONS
Refer
Loosen
shaft.
Clean
Be
To
output
Push
3.
PREVENTIVE
CAUTION:
ㆍ
Following
inspect
If
contamination
isopropyl
surfaces
If
the
be
components
Inspect
cord
Replace
required,
FOR
to
General
the
drop
with
sure
sensor
FOR
to
General
rotor
After
rollers
sure
all
parts
replace
rotor,
shaft.
rotor
into
Disconnect
housing.
cleaning
the
alcohol.
thoroughly.
interior
returned
the
versions,
the
return
CLEANING
Cleaning
sensors
in
isopropy!
areas
are
CLEANING
Cleaning
set
screw
of
rotor
align
and
with a 5/64”
rotor,
with
set
removing
thoroughly
place
MAINTENANCE
pump
of
pump
is
to
can
power
power
the
interior
is
minor,
Assure
extensively
Sherwood
be
replaced.
cord,
for
the
and
excessive
cord
pump
DROP
DETECTOR
Directions before
the
upper
drip
chamber
alcohol.
clear.
ROTOR
are
tighten
ASSEMBLY
Directions
allen
avoid
getting
warm
soapy
completely
screw
on
set
before
dry
hub
of
screw
wrench
any
water,
rotor
INSTRUCTIONS
from
A.C.
power
pump
exterior,
for
presence
clean using a cotton
that
ali
contaminated,
Factory
the
A.C.
wear,
if
needed.
to
Sherwood
of
residue
Service
receptacle
damage,
If
repair
for
remove
foreign
is
it
so
servicing.
starting.
guide
starting.
and
moisture
or
isopropyl
before
putting
with
(Do
not
over
source
before
back
material
swab
removed
is
recommended
that
any
in
the
or
evidence
to
the
A.C.
with a cotton
gently
puil
in
the
rotor
alcohol
it
back
the
flattened
tighten).
opening
cover
of
(e.g.,
dampened
and
dry
that
damaged
case
of
of
electrical
receptacle
swab
rotor
fons--rd
shaft
if
onto
section o the
the
pump
housing
formula).
with
all
cleaned
the
pump
detachabl:
arcing.
is
off
o,
ning.
necessary.
shaft.
pump
and
Low/Dead
a.
Low
is
between
b.
Audible
c.
Pump
5.3
and
Reassemble
3.
Touch
.
“ON”
“OFF”
.
INCREMENT”
.
“DECREMENT”
.
“START/HOLD”
“VOLUME”
.
“DOSE”
oonocu
.
Visual
a.
AC
Light
b.
No
Set
c.
Flow
.
Audible
a.
Volume
.
Motor
See
Motor
a.
295
b.
35
mi/hr.
.
“Volume
a.
Volume
.
Electrical
a.
Chassis
b.
Leakage
c.
Repeat
.
Operational
a.
All
units
minimum
b.
Operate
Battery
bat
should
6.8
Alarm
should
5.6
VDC
the
pump
Panel
and
Button
Button
Button
Button
Indicators
Error
Error
Alarm
Adjust
Speed
(with
Speed
ml/hr.
(4.54
(12.77
Clear”
Clear
Safety
to
Ground
Current
1.
Normal
2.
Open
3.
Open
leakage
Tests
should
of 5 minutes
pump
Detect
be
displayed
and
7.2
VDC
turn
off
when
housing
Display
Button
(1-50,
Button
(300-50,
Button
feeding
Verification
to
4.84
to
Button
indicator
Tests
sec.)
13.56
Resistance
(less
than
Ground
Ground
test
be
operated
Reverse
with
with a feeding
for a minimum
Circuit
when
battery
prior
50-300)
set
in
Test
sec.)
100
pump
at
of
the
battery
voltage
to
performing
50-1)
place)
(0.15
ohms
microamps)
Polarity
turned
off.
150
mil/hr
set
16
min.
voltage
is
between
Steps 3 through
or
less)
fora
in
place
at 5 ml/hr.
X
X
x
"MMX
xxx
хх
9.
KOK
OX
D>
KOK
xxx