Sherwood Medical Kangaroo 224, Kangaroo 324 User manual

If
plug with cord
completely
CAUTION:
cleaning
and
isopropyl
from
If
cleaning
with
isopropyl
removed.
Avoid as
of
the
A.C.
receptacle alcohol.
the
pump
dry.
of
the
exposing
this
can
power
prior
A.C.
alcohol;
lead
cord
is
necessary,
plug
section
If
the
pump
to
cleaning.
receptacle
upon
completion
A.C.
receptacle
to
an
electrical
wipe
of
the
the
power
has a detachable
Do
not
reconnect
is
necessary,
ensure
and
shock
use a cotton that
power
or
exterior
cord
with a cloth
power
power
all
residue
cord
fire
hazard.
surfaces
cord,
cord
swab
and
to
excess
of
the
wall
dampened
disconnect
to
pump
dampened
liquid
is
moi
‘ire,
the until
Functional
Perform
Feeding
1.
the
Pump
Sensor
This
test
and
the
a.
Phototransistor
at
the
[note:
and
Safety
listed
Service
Circuit
must
pump
positioned
collector
adjust
tests
be
performed
bias
of
R42
Testing
at
each
Manual”
servicing,
as
with
face
voltage
the
phototransistor
as
needed]
referring
necessary.
no
drip
down away
(0.99
to
chamber
from
1.01
to
direct
VDC)
NUS.
your
“Kangaroo
installed
lighting.
K224
Com
224/324
K324
DIRECTIONS
Refer
Clean
dampened
DIRECTIONS
Refer
Loosen shaft.
Clean
Be
To output
Push
3.
PREVENTIVE
CAUTION:
Following
inspect
If
contamination isopropyl surfaces
If
the be
components
Inspect
cord
Replace
required,
FOR
to
General
the
drop
with
sure
sensor
FOR
to
General
rotor
After
rollers
sure
all
parts
replace
rotor,
shaft.
rotor
into
Disconnect
housing.
cleaning
the
alcohol.
thoroughly.
interior
returned
the
versions,
the
return
CLEANING
Cleaning
sensors
in
isopropy!
areas
are
CLEANING
Cleaning
set
screw
of
rotor
align
and
with a 5/64”
rotor,
with
set
removing
thoroughly
place
MAINTENANCE
pump
of
pump
is
to
can
power
power
the
interior
is
minor,
Assure
extensively
Sherwood
be
replaced.
cord,
for
the
and
excessive
cord
pump
DROP
DETECTOR
Directions before
the
upper
drip
chamber
alcohol.
clear.
ROTOR
are
tighten
ASSEMBLY
Directions
allen
avoid
getting
warm
soapy
completely
screw
on
set
before
dry
hub
of
screw
wrench
any
water,
rotor
INSTRUCTIONS
from
A.C.
power
pump
exterior,
for
presence
clean using a cotton
that
ali
contaminated,
Factory
the
A.C.
wear,
if
needed.
to
Sherwood
of
residue
Service
receptacle
damage,
If
repair
for
remove
foreign
is
it
so
servicing.
starting.
guide
starting.
and
moisture
or
isopropyl
before
putting
with
(Do
not
over
source
before
back
material
swab
removed
is
recommended
that
any
in
the
or
evidence
to
the
A.C.
with a cotton
gently
puil
in
the
rotor
alcohol
it
back
the
flattened
tighten).
opening
cover
of
(e.g.,
dampened
and
dry
that
damaged
case
of
of
electrical
receptacle
swab
rotor
fons--rd
shaft
if
onto
section o the
the
pump
housing
formula).
with
all
cleaned
the
pump
detachabl:
arcing.
is
off
o,
ning.
necessary.
shaft.
pump
and
Low/Dead
a.
Low
is
between
b.
Audible
c.
Pump
5.3
and
Reassemble
3.
Touch
.
“ON” “OFF”
.
INCREMENT”
.
“DECREMENT”
.
“START/HOLD”
“VOLUME”
.
“DOSE”
oonocu
.
Visual
a.
AC
Light
b.
No
Set
c.
Flow
.
Audible
a.
Volume
.
Motor
See
Motor
a.
295
b.
35
mi/hr.
.
“Volume
a.
Volume
.
Electrical
a.
Chassis
b.
Leakage
c.
Repeat
.
Operational
a.
All
units
minimum
b.
Operate
Battery
bat
should
6.8
Alarm
should
5.6
VDC
the
pump
Panel
and
Button
Button
Button
Button
Indicators
Error
Error
Alarm
Adjust
Speed
(with
Speed
ml/hr.
(4.54
(12.77
Clear”
Clear
Safety
to
Ground
Current
1.
Normal
2.
Open
3.
Open
leakage
Tests
should
of 5 minutes
pump
Detect
be
displayed
and
7.2
VDC
turn
off
when
housing
Display
Button
(1-50,
Button
(300-50,
Button
feeding
Verification
to
4.84
to
Button indicator
Tests
sec.)
13.56
Resistance
(less
than
Ground Ground
test
be
operated
Reverse
with
with a feeding
for a minimum
Circuit
when
battery
prior
50-300)
set
in
Test
sec.)
100
pump
at
of
the
battery
voltage
to
performing
50-1)
place)
(0.15
ohms
microamps)
Polarity
turned
off.
150
mil/hr
set
16
min.
voltage
is
between
Steps 3 through
or
less)
fora
in
place
at 5 ml/hr.
X
"MMX
xxx
хх
9.
KOK OX
D>
KOK
xxx
15.
13.
14.
16.
17.
Tubing
AC
VOL
Light
CLR
Guide
Light
Latch
Pole
Pole
Clamp
Clamp
Knob
Alarm
Knob
Volume
Contro.
Direction
Power
Power
Cord
Cord
tor
Receptacle
Use
Label
Touch
Rotor
LED
Running
Display
Assembly
Panel
Indicator
and
Pump
Lights
+
Guide
Retainer
Lock
{including
Tubing
Lower
Drip
Guide
drop
Chamber
sensor)
Guide
Upper
Drip
Chamber
-
Kangaroo
Handle
324
14
| |
o
-
Kangaroo
2.
Tubing
AC
Pole
Light
Clamp
Guide
Knob
Latch
Pole
Rate
Clamp
Control
Knob
Direction
Power
Power
Cord
Cord
for
Receptacle
Use
Label
onou
Touch
LED
Running
Indicator
Lights
Rotor
Display
Assembly
Panel
and Pump
BB
Lower
Guide
Retainer
Drip
Lock
Chamber
Tubing
(including
Guide
drop
sensor)
Guide
Upper
Drip
Chamber
1.
Handle
224
DIRECTIONS
*
be
Refer
Unless
cleaned.
to
soiling
General
is observed,
Cleaning
Directions
the power
before
cord
and
starting.
pump
A.C.
receptacle
should
not
FOR
CLEANING
PUMP
A.C.
RECEPTACLE
AND
POWER
CORD
«
*
When
cleaning
Avoid
getting
cleaning
solution
any
around
get
moisture
inside
buzzer
into
pump.
opening
the A.C.
on
receptacle.
back
of pump
housing,
do not
let
!
+
Avoid
Avoid
excess
excess
moisture
moisture
near
near
handle
pole
clamp
hinges.
area.
*
+
Clean
upright
Refer
to
outside
position
General
surface
as
Cleaning
much
with
as
Directions
a
possible.
damp
(not
before
wet)
starting.
cloth
or sponge
and
keep
pump
in
DIRECTIONS
*
Vesta-Syde
DO
NOT
FOR
USE
because
CLEANING
strong
damage
cleaners
PUMP
to
such
HOUSING
the
as
pump
Spray-Nine,
case
housing
Phisohex,
can
result.
Hibiciens,
or
cleaning
applied
should
reach
areas
be
with
with
used
a
such
this
damp
sparingly
solution
as
(not
the drop
because
wet}
can
sensor
damage
cotton
repeat
and
swab
the
inside
cleaning
plastic
may
the
be
pump
may
used
A.C.
damage
case.
receptacle,
for cleaning
Isopropyl
the case.
however,
difficult
alcohol
to
it
vieaning
+ A
may
of
mild
Kangaroo
be
detergent
cleaned
224
with
should
and
a 10:1
324
be used
pumps
water
for
and
must
general
hypochlorite
be performed
cleaning.
mixture,
if
as
necessary,
follows:
however,
the
repeated
pump
electrical
completion
pump
Disconnect
and
power
cord
are thoroughly
dry.
components.
of
pump
cleaning,
from
A.C.
do not
power
connect
source
to A.C.
before
power
cleaning.
source
Upon
until
DESCRIBED
As
in
prevent
order
with
any
liquid
to avoid
A.C.
HEREIN
from
electrical
powered
entering
COULD
electrical
shock
the
RESULT
pump
hazard,
device,
IN
housing
fire
hazard,
or damage
to
HAZARDS
care
or
must
the
TO
A.C.
be
USERS.
taken
receptacle
to
2.
CAUTION:
GENERAL
DO
CLEANING
SPONGE.
FAILURE
CLEANING
NOT
IMMERSE
SOLUTION;
DIRECTIONS
PUMP
TO
CLEAN
FOLLOW
OR POWER
USING
THE
CORD
CLEANING
A DAMP
IN WATER
(NOT
PROCEDURES
WET)
OR
CLOTH
OTHER
OR
Sherwood
ramp
Prior
to
exterior
performance
Factory
as described
Service,
of functional
call
below.
1-800-448-0190.
and
safety
inspections
described
below,
clean
the
This
recommended
(12)
outside
Kangaroo
months
pump
servicing
pumps
must
per
service
JCAHO.
organization,
or by
Sherwood
Factory
Service.
To arrange
for
be
and
periodically
may
interval
Servicing
be
is
done
serviced
6 months,
at
must
your
to
be
assure
but
biomedical
performed
in no
proper
case
engineering
by
functioning
less
personnel
than
department,
once
trained
and
each
safety.
in
twelve
servicing
by
The
an
1, IMPORTANT
SERVICING
INFORMATION
Loading...